Articles

Fast and Rugged Direct Analysis of Polar Pesticides in Spinach

Featured Application: Polar Pesticides in Spinach on Raptor Polar X by LC-MS/MS

12 Jan 2021

feature FSFA3215
  • No time-consuming derivatization, ion-pairing steps, or long waits for column equilibration.
  • Unique stationary phase provides good retention and efficient elution, resulting in better peak shape.
  • Robust column performance ensures consistent chromatography and reliable results.

The direct analysis of polar pesticides, such as glyphosate, its main metabolite AMPA, and glufosinate, is difficult because they are not retained well on reversed phase columns, and chelation with metal surfaces in the LC-MS/MS can significantly reduce response. Derivatization and ion-pairing reagents are often used to improve performance, but they require additional sample preparation time and can produce unfavorable side effects (precipitates, long equilibration times, system contamination, etc.).

Using a Raptor Polar X column is a better solution because the novel stationary phase offers stronger interaction with polar anionic compounds. It features hybrid retention modes (HILIC and ion exchange) and provides strong retention yet allows analytes to be quickly eluted as symmetrical peaks through simple mobile phase changes. In this analysis of polar pesticides in spinach, excellent chromatographic performance is seen for all three compounds with a total analysis time of just eight minutes. Further, peak shapes and retention times remain stable over hundreds of matrix injections, and a simple guard column change restores full performance. This fast, simple, and effective workflow is a good alternative for food labs interested in the direct analysis of polar pesticides without the need for complex and time-consuming procedures.

Polar Pesticides in Spinach on Raptor Polar X by LC-MS/MS

LC_FS0547

Peaks

PeaksConc.
(ng/mL)
Precursor IonProduct Ion 1Product Ion 2Product Ion 3
1.AMPA100110.1079.0563.1081.10
2.Glufosinate100180.2085.1595.10
3.Glyphosate100168.1063.0579.05

Conditions

ColumnRaptor Polar X (cat.# 9311A32)
Dimensions:30 mm x 2.1 mm ID
Particle Size:2.7 µm
Pore Size:90 Å
Guard Column:Raptor Polar X guard column cartridge 5 mm, 2.1 mm ID, 2.7 µm (cat.# 9311A0252)
Temp.:35 °C
Standard/SampleSpinach extract
Inj. Vol.:2 µL
Mobile Phase
A:Water, 0.5% formic acid
B:Acetonitrile, 0.5% formic acid
Time (min)Flow (mL/min)%A%B
0.000.53565
5.00.59010
6.50.59010
6.510.53565
80.53565
DetectorMS/MS
Ion Mode:ESI-
Mode:MRM
InstrumentUHPLC
Sample PreparationFrozen spinach was added to a Blixer processor with dry ice (3:1-4:1 ratio) and then ground into a very fine powder. The homogenate was placed into the freezer immediately. A 5.0 gram sample of the spinach powder was weighed into a 50 mL centrifuge tube (cat.# 25846). According to the QuPPe method (Quick Polar Pesticides Method), 5.0 mL of methanol with 1.0% formic acid was added into the centrifuge tube. The tube was shaken by hand for 1 min and then by a mechanical shaker vigorously for 5 min. After centrifuging for 10 min at 4200 rpm, the supernatant was filtered through a 0.22 μm filter (cat.# 23984). The final extract was fortified with the AMPA, glufosinate, and glyphosate at a final concentration of 100 ng/mL.

Author

  • Restek Corporation

    Restek is a leading provider of chromatography columns, accessories, and certified reference materials. Trust Restek for reliable, high-quality analytical solutions.

    View Articles
FSFA3215