Loading Events
  • This event has passed.

« All Events

SOFT 2026

September 19 - September 24

Chicago city skyline.

We’re going to SOFT 2026 in Chicago, IL! Stop by Booth 313 to talk with our team about our solutions for forensics analysis. And don’t forget to enter our raffle for a chance to win an iPad (128 GB)!

Restek’s Presentations at SOFT 2026

Oral Presentations

Evaluating Strategies for Low-Level Quantification of THC Isomers and Metabolites in Whole Blood by Conventional GC-MS Instrumentation

Thursday, September 24, 9:10-9:20 a.m. | O-118
Presenter/Authors: Amanda Rigdon, Haley Berkland

Abstract

Introduction: Liquid chromatography–tandem mass spectrometry (LC-MS/MS) is the preferred technique for drug and metabolite analysis in biological matrices due to its sensitivity, selectivity, and minimal derivatization requirements. However, analysts may need to develop methods using conventional gas chromatography–mass spectrometry (GC-MS) due to limited LC-MS/MS access and the widespread availability of GC-MS platforms.

Objectives: The objective of this work was to evaluate and mitigate the challenges of low-level drug analysis in biological matrices using conventional GC-MS through comprehensive method optimization. Analysis of THC isomers and metabolites in whole blood served as a representative test case due to the low detection limits required. Eleven analytes, including 11-hydroxy-Δ8-tetrahydrocannabinol; 9(R)-hexahydrocannabinol; and (6aR,9R)-Δ10-tetrahydrocannabinabinol were analyzed.

Methods: Thermodynamic modeling software was employed to develop an optimized chromatographic separation on a 5-type column. A selected ion monitoring (SIM) method was established to optimize dwell times and leverage shared fragment ions between analytes. Following method development, a calibration curve was prepared in ethyl acetate to evaluate the sensitivity of a conventional GC-MS system (Agilent 7890B GC coupled with a 5977B MS, standard electron ionization source). Method performance was subsequently assessed in whole blood samples prepared via liquid–liquid extraction (LLE) and spiked post-extraction to evaluate sensitivity and selectivity in a complex matrix. Whole blood was acidified with 10% acetic acid, then extracted with 9:1 hexane:ethyl acetate. Extracts were then spiked, dried, and derivatized with BSTFA + 1% TMCS.

Results: All analytes demonstrated acceptable sensitivity in solvent at a final extract concentration of 5 ng/mL, corresponding to 0.5 ng/mL in whole blood. Calibration curves exhibited acceptable linearity, with r² values greater than 0.9990 for all analytes except Δ9-THC (r² = 0.9982). Elevated back-calculated accuracies were observed for four analytes at the low calibration range. As expected, matrix interferences were significant for several monitored ions in whole blood. Acceptable sensitivity (S/N > 5) was achieved at 1 ng/mL in whole blood for at least one ion per analyte, with the exception of (6aR,9R)-Δ10-tetrahydrocannabinol and 11-nor-9-carboxy-Δ8-tetrahydrocannabinol, which showed substantial interference across all monitored ions. Although less sensitive than LC-MS/MS, this LOD is applicable to established working ranges for these analytes in whole blood. Triplicate injections of a post-extraction spiked sample (equivalent to 25 ng/mL in whole blood) yielded %RSD values of 1.0–10.6% after internal standard normalization.

Discussion: These results demonstrate that while conventional GC-MS systems cannot achieve the performance of advanced triple quadrupole platforms, they can support low-level quantification of cannabinoids and related metabolites in biological matrices when methods are carefully optimized. Key factors influencing performance include chromatographic resolution for analytes as well as matrix interferences, mass spectrometric parameter optimization, and instrument maintenance. Increased variability at the low end of the calibration range highlights the need for further refinement to achieve robust quantification for all analytes. Ongoing work building on these results will be presented, exploring the use of thermodynamic modeling to better characterize and mitigate matrix interferences as well as alternative derivatization strategies and detector optimization to enhance sensitivity.

Recommendations for Expanding the Scope of an Existing LC-MS/MS Method

Thursday, September 24, 11:45-11:50 a.m. | L-016
Presenter: Samantha Herbick

Introduction: Emerging drugs of abuse compounds are a consistent pain point for toxicology laboratories. There are often new analytes and novel psychoactive compounds (NPS) emerging every quarter. To stay current, it is important for labs to update their testing panel, but finding ways to keep methods up to date can be both time-consuming and challenging. This work provides recommendations to make these updates easier. Using emerging compounds from the Center for Forensic Research and Education’s (CFSRE) Scope Recommendations, 20 compounds were selected to be added to an existing LC-MS/MS method that contained 231 analytes. Recommendations range from analyte selection, metabolite considerations, instrument method modifications, and the use of a virtual chromatogram modeler.

Objectives: The objective of this work is to offer recommendations to testing laboratories on easier ways to expand their testing scope when constantly dealing with new and emerging compounds.

Methods: An existing LC-MS/MS screening method was chosen because of its extensive scope of 231 analytes. The method used a Biphenyl 100 x 2.1 mm, 2.7 um column paired with a Biphenyl Guard Column Cartridge 5 x 2.1 mm, 2.7 um. All existing analytes were separated using a 12-minute cycle time. Gradient conditions were used at 30 °C using water and methanol, both containing 0.1% formic acid and 2 mM ammonium formate. The 20 analytes were added to the method and tested under optimized method conditions first by using a virtual chromatogram modeler to test retention and separation of the new analytes. A few of the new analytes were isobars with existing analytes in the original method. For example, a-methyl acetyl fentanyl from the expansion list and fentanyl from the original list. Because of the different sets of isobars. method parameters were also tested to ensure resolution. These method changes included modifications to the column temperature and gradient conditions with the help of a virtual chromatogram modeler.

Results: Analyte selection was in accordance with the Scope Recommendations provided by the CFSRE. Twenty of the newest emerging compounds were tested on an existing screening method to provide an example of elution order as well as isobar separations. The use of a virtual chromatogram modeler can be beneficial in instances like these. Ease of use and application of the modeler was demonstrated. Due to specific isobar groups, additional modified method parameters were provided to address these separations as necessary. All additional analytes retained well on the existing method.

Discussion: Toxicology labs are constantly dealing with identification and testing of emerging compounds. It is often difficult and time-consuming to update existing methods. The recommendations included here in this work are intended to ease the burden of scope additions while retaining the integrity of the existing method. Resources, such as the CFSRE, are extremely helpful tools when considering what additions should be added to the testing scope. This resource paired with the recommendations above should help with toxicology labs future scope expansions.

Technical Posters

LC-MS/MS Methods for Alcohol Biomarkers in Urine and Whole Blood

Tuesday, September 22, 12:00-1:30 p.m. | P112
Presenter/Authors:
Samantha Herbick, Jamie York, Justin Steimling

Introduction: Several established biomarkers are available for monitoring prior alcohol consumption in biological samples using LC-MS/MS. Selection of the appropriate biomarker depends on clinical objectives, including detection of recent alcohol use (within ~3 days) or assessment of longer-term consumption (2–4 weeks). This study presents two analytical methods, one for quantifying ethyl glucuronide (EtG) and ethyl sulfate (EtS) in urine, and another for the analysis of phosphatidylethanol (PEth) in whole blood, providing complementary tools for alcohol monitoring.

Objectives: To develop and demonstrate two LC-MS/MS workflows for alcohol metabolite biomarkers in biological matrices: one workflow for quantifying short-term alcohol use biomarkers (EtG and EtS) in urine and another method for assessing drinking patterns through PEth analysis in whole blood.

Methods: Analyses were performed using a Shimadzu LC-MS 8060 triple quadrupole mass spectrometer operating in ESI negative mode.

For urine analysis, nine EtG and EtS calibrators (50–2,000 ng/mL) were prepared in synthetic urine. Quality control samples were generated at 100, 400, and 1600 ng/mL using six unique lots of analyte-free human urine, including two kidney disease patients in order to evaluate samples with enhanced potential for matrix effects. Samples were diluted 20-fold with water containing 0.1% formic acid, vortexed, centrifuged, and injected (10 μL). Chromatographic separation was achieved using a Force Biphenyl column (100 × 3 mm, 3 μm) with water and methanol (both containing 0.1% formic acid) as mobile phases. The flow rate was 0.8 mL/min, the column oven was set to 30 °C, and the overall run time was 5 minutes utilizing gradient conditions.

For whole blood analysis, nine PEth homologue 16:0/18:1 and 16:0/18:2 calibrators were prepared in bovine whole blood at a range of 10-1000 ng/mL with quality controls prepared at 15, 100, and 500 ng/mL in analyte-free human whole blood. Samples (50 μL) were extracted with 150 μL of 4:1 isopropanol:tetrahydrofuran, vortexed, centrifuged, and injected (5 μL). Separation was performed using a Raptor C8 column (50 × 2.1 mm, 2.7 μm) with mobile phases consisting of water with 5 mM ammonium acetate and acetonitrile/isopropanol (90:10). The flow rate was 0.6 mL/min, the column oven was set to 30 °C, and the overall run time was 5 minutes using gradient conditions.

Results: Both methods demonstrated excellent linearity, with correlation coefficients (r²) ≥ 0.995. Calibration for EtG and EtS utilized 1/x-weighted quadratic regression, whereas PEth was calibrated using 1/x-weighted linear regression. Intraday precision and accuracy for EtG and EtS showed percent recoveries ranging from 85.1-114.6%, with %RSD values ≤9.6%. For PEth, percent recoveries were within 10% of the nominal concentration, with %RSD values ≤14.2%. Interday performance for both methods showed percent recoveries within 11.9% of the nominal range, with %RSD values ≤10.1%.

Discussion: Two robust LC-MS/MS methods were successfully developed for the quantification of alcohol biomarkers in urine and whole blood. Both workflows are rapid, sensitive, and reliable, meeting the performance requirements for clinical and forensic applications, and enabling flexible monitoring of both short- and long-term alcohol consumption.

Semisynthetic Cannabinoids: The Newest Challenge in Analyzing THC Isomers

Tuesday, September 22, 12:00-1:30 p.m. | P117
Presenter/Authors:
Jared Burkhart, Haley Berkland

Introduction: Over the past several years, Δ8-THC has made headlines as a popular “legal” alternative to Δ9-THC. The rise in popularity of Δ8-THC led to a new analytical challenge for toxicology laboratories—differentiating between the Δ8 and Δ9 isomer forms. For labs performing this testing by LC-MS/MS, complete separation of these isomers and their metabolites is necessary for accurate quantitation. Recently, a new group of compounds referred to as “semisynthetic cannabinoids”, have emerged on the market. Drug monitoring groups in the U.S. have suggested that these compounds be added to toxicology testing scopes due to increased use and detection. Many toxicology laboratories have already implemented LC-MS/MS methods to simultaneously analyze for Δ8-THC and Δ9-THC in casework. In this work, eight emerging semisynthetic cannabinoids (9(R)-HHC; 9(S)-HHC; 9(R)-HHC-COOH; 9(S)-HHC-COOH; Δ10-THC; Δ9-THC-O-Acetate; Δ9-THCP’ and CBDP) were added to an existing LC-MS/MS method used for the analysis of Δ8-THC, Δ9-THC, and their hydroxy and carboxy metabolites in whole blood.

Objectives: Incorporate eight emerging semisynthetic cannabinoids into an existing method developed for analysis of THC isomers in whole blood.

Methods: The original method was developed for the quantitative analysis of Δ8-THC, Δ9-THC and their isomeric metabolites in whole blood by LC-MS/MS. The method used a Raptor FluoroPhenyl 100 x 3 mm, 2.7 µm column and mobile phases of water and methanol, both acidified with 0.1% formic acid. The column temperature was 40 °C, the flow rate was 0.8 mL/min, and gradient elution was used. A sample containing the eight semisynthetic cannabinoids and the original six cannabinoids was prepared at a concentration of 100 ng/mL. The sample was run on the original method to determine separation and elution order.

Results: While resolution was achieved for all compounds, including isomers, using the original method conditions, Δ10-THC, Δ9-THCP, and Δ9-THC-O-acetate eluted at 100% B, where matrix interferences are known to wash off the column. To prevent interference, the method conditions were altered to allow these compounds more time on the column. Additional resolution for the isomers was also achieved by dropping the column temperature from 40 °C to 30 °C. All critical pairs had a calculated resolution of 1.5 or better, demonstrating baseline resolution was achieved.

Discussion: All eight semisynthetic cannabinoids were successfully integrated into a preexisting LC-MS/MS panel for the analysis of THC isomers in whole blood. Because different isomer configurations can have different psychoactive activities, it is important to differentiate between isomer forms in toxicology casework. This work highlights the importance of choosing a stationary phase that shows optimal selectivity for the analyte class being analyzed, allowing new analytes to easily be integrated into the method as they emerge.

Analyzing the Impact of Centrifugation and Extraction Aides on Oral Fluid Sample Recovery

Wednesday, September 23, 12:00-1:30 p.m. | P265
Presenter/Authors:
Samantha Herbick, Jared Burkhart

Introduction: Oral fluid is a relatively simple matrix to collect and prepare; however, the collection kits can cause issues downstream. Quantisal™ collection kits consist of an applicator with attached sponge, a tube with buffer solution, and a cap. The total volume for testing should be 4 mL (1 mL of oral fluid and 3 mL of buffer); however, it is difficult to recover the full amount. There are different techniques to manipulate the sponge and improve sample recovery. Popular techniques include manual compression with and without and aide and centrifugation. This work also tested and compared an oral fluid extraction accessory (OFEA) accompanied with a centrifugation step to previously mentioned techniques. This study compared these four extraction techniques by examining volume recovery and analyte recovery.

Objectives: Demonstrate the advantages of incorporating a centrifugation step and the use of extraction aides to sample preparation to determine its impacts on volume and analyte recovery when performing analysis of drugs of abuse and (DoA) in oral fluids by LC-MS/MS.

Methods: Thirty-one common DoA compounds were separated by LC-MS/MS using a Biphenyl column under gradient conditions using water and methanol, both containing 0.1% formic acid, with a total cycle time of 7 minutes. Samples were prepared in synthetic oral fluid and combined with Quantisal™ buffer. Four recovery techniques were compared and tested in triplicate: manual compression, centrifugation, manual compression with an aide, and centrifugation with OFEA. Samples were prepared using a salt-assisted liquid-liquid extraction (SALLE).

Results: Four techniques were evaluated by examining total volume removed and analyte recovery. Volume recovery was tested by pouring the solution into a graduated cylinder and recording the total volume recovered. When using centrifugation with the OFEA, an additional 500 µL was collected compared to the other techniques. Analyte recovery was compared by spiking a known concentration, 50 ng/mL, into each of the samples. Samples were evaluated using a calibration curve prepared without using the kits. Peak areas increased for all analytes when using the centrifugation and centrifugation with the OFEA. Accuracy and precision were assessed by analyzing whether the results fell within +/- 15% of the target as well as comparing %RSD values. Manual compression with the aide and centrifugation with the OFEA showed the best accuracy and precision with comparable results to each other. A study was also conducted to assess efficiency of each of the techniques. Four samples were prepared by each technique, and the time was then multiplied to give an indication of time for 12, 48, and 96 samples. When preparing 96 samples, the centrifugation with the OFEA is the fastest technique by up to 32 minutes.

Discussion: Four techniques were compared to extract liquid from the sponge of oral fluid collection kits. Through examining volume and analyte recovery, it was clear that the centrifugation step paired with the oral fluid extraction accessory showed the best results for total volume recovered as well as increased analyte recovery when analyzed quantitatively compared to the other techniques.

Enter Our Raffle!

Enter our SOFT raffle for a chance to win an Apple iPad (128 GB)!

Details

  • Start: September 19
  • End: September 24

Venue