Articles

PFAS Analysis by U.S. EPA Method 537.1 for Drinking Water

03 Oct 2025

feature EVAR4284
  • Sample Matrix: drinking water
  • Sample Preparation: S-DVB solid-phase extraction
  • Detection Technique: LC-MS/MS

Determination of Selected Per- and Polyfluorinated Alkyl Substances in Drinking Water by Solid Phase Extraction and Liquid Chromatography/Tandem Mass Spectrometry (LC-MS/MS)

U.S. EPA Method 537.1 is used for the analysis of 18 PFAS in drinking water. It combines solid-phase extraction sample preparation using styrene-divinylbenzene cartridges with LC-MS/MS analysis on a C18 column. The method allows flexibility in some aspects, such as evaporation technique, separation technique, mobile phase composition, LC column and conditions, and detector conditions, but changes to sample collection, preservation, and extraction and quality control requirements are prohibited. The full analyte list and recommended products for PFAS analysis in drinking water by Method 537.1 are provided below. In addition, you can register and directly compare analyte lists and matrices across PFAS analysis methods here or visit www.restek.com/PFAS to explore our technical resources for PFAS analysis.

Analyte List for U.S. EPA Method 537.1 for PFAS Analysis

PFAS acronyms, compound names, and CAS numbers vary across methods and guidelines, and they also depend on compound form (e.g., neutral vs. salt). The following list uses standardized terminology to allow easy comparison across the methods and guidelines in our comprehensive guide. Please refer to U.S. EPA Method 537.1 for method-specific nomenclature.

AcronymCompound NameCAS#
NEtFOSAAN-ethyl perfluorooctanesulfonamidoacetic acid2991-50-6
NMeFOSAAN-methyl perfluorooctanesulfonamidoacetic acid2355-31-9
PFBSPerfluoro-1-butanesulfonic acid375-73-5
PFDAPerfluorodecanoic acid335-76-2
PFDoAPerfluorododecanoic acid307-55-1
PFHpAPerfluoroheptanoic acid375-85-9
PFHxSPerfluoro-1-hexanesulfonic acid355-46-4
PFHxAPerfluorohexanoic acid307-24-4
PFNAPerfluorononanoic acid375-95-1
PFOSPerfluorooctanesulfonic acid1763-23-1
PFOAPerfluorooctanoic acid335-67-1
PFTeDAPerfluorotetradecanoic acid376-06-7
PFTrDAPerfluorotridecanoic acid72629-94-8
PFUnAPerfluoroundecanoic acid2058-94-8
HFPO-DA, GenXHexafluoropropylene oxide dimer acid13252-13-6
9Cl-PF3ONS9-Chlorohexadecafluoro-3-oxanonane-1-sulfonic acid (F-53B Major)756426-58-1
11Cl-PF3OUdS11-Chloroeicosafluoro-3-oxaundecane-1-sulfonic acid (F-53B Minor)763051-92-9
ADONA4,8-Dioxa-3H-perfluorononanoic acid919005-14-4

Recommended Products for U.S. EPA Method 537.1 for PFAS Analysis

Analytical LC Columns:


Raptor C18, 2.7 µm, 50 x 2.1 mm HPLC Column
Force C18, 1.8 µm, 50 x 2.1 mm LC Column

Delay Columns:


PFAS Delay Column, 5 µm, 50 x 2.1 mm HPLC Column

PFAS delay column: 5 µm, 50 x 3.0 mm (cat.# 580959 [custom product])

Guard Columns:


UltraShield UHPLC PreColumn Filter, 0.2 µm Frit, ea.
EXP Direct Connect Holder for EXP Guard Cartridges, Includes Fitting & Ferrules
Raptor C18 EXP Guard Column Cartridge, 2.7 µm, 5 x 2.1 mm, 3-pk.

Standards:


PFAS EPA Method 537.1 Analyte Primary Dilution Standard (PDS) 2 µg/mL, Methanol (1 mM NaOH)/2-Propanol (98:2), 1 mL/ampul

Sample Prep:


Resprep S-DVB SPE Cartridge, 6 mL/500 mg, 30-pk.

Sample Reservoirs:


SPE Empty Tubes, Sample Reservoir, 75 mL, Polypropylene, 12-pk.
SPE Connector, 1, 3, 6, 10, or 15 mL, Polypropylene, 15-pk.

Vacuum Manifolds:


Resprep QR-24 Vacuum Manifold
Resprep QR-12 Vacuum Manifold

Vacuum Pumps:


Chemker 300 PTFE Vacuum Pump, 20 L/min, AC110 V, 60 Hz
Chemker 300 PTFE Vacuum Pump, 18 L/min, AC220 V, 50 Hz

General Lab Supplies:


Limited-Volume Screw-Thread Polypropylene Vials, 9 mm, 700 µL, 12 x 32 mm, 100-pk.
Limited-Volume Screw-Thread Polypropylene Vials, 9 mm, 700 µL, 12 x 32 mm, 1000-pk.
2.0 mL, 9 mm Solid-Top Polyethylene Caps, Screw-Thread, 10 mil thick membrane, Clear, 100-pk.
2.0 mL, 9 mm Solid-Top Polyethylene Caps, Screw-Thread, 10 mil thick membrane, Clear, 1000-pk.
Empty Centrifuge Tubes, 50 mL, Polypropylene w/Cap, 50-pk.
Empty Centrifuge Tubes, 50 mL, Polypropylene w/Cap, 500-pk.
Norm-Ject Plastic Syringe, 10 mL Luer Lock Tip, 100-pk.
Norm-Ject Plastic Syringe, 5 mL Luer Lock Tip, 100-pk.

Discover More Solutions for PFAS Analysis!

Start Saving
Time Today

Develop PFAS Methods in Minutes with our Pro EZLC Chromatogram Modeler

Start Saving Time Today

Develop PFAS Methods in Minutes with our Pro EZLC Chromatogram Modeler

Tags

Author

  • Restek Corporation

    Restek is a leading provider of chromatography columns, accessories, and certified reference materials. Trust Restek for reliable, high-quality analytical solutions.

    View all posts
EVAR4284
Version: B