- Sample Matrix: emissions from stationary sources
- Sample Preparation: WAX solid-phase extraction
- Detection Technique: LC-MS/MS
Other Test Method 45 (OTM-45) Measurement of Selected Per- and Polyfluorinated Alkyl Substances from Stationary Sources
U.S. EPA Method OTM-45 is a performance-based draft method that can be used for the analysis of 50 PFAS in air emissions from stationary sources. It combines solid-phase extraction sample preparation using weak anion exchange (WAX) cartridges with LC-MS/MS analysis on a C18 column. Draft Method OTM-45 was released to encourage a consistent process for facilities, stationary source test teams, research laboratories, and other stakeholders and to promote discussion and further evaluation of the method. The full analyte list and recommended products for PFAS analysis in emissions by Method OTM-45 are provided below. In addition, you can register and directly compare analyte lists and matrices across PFAS analysis methods here or visit www.restek.com/PFAS to explore our technical resources for PFAS analysis.
Analyte List for EPA Method OTM-45 for PFAS Analysis
PFAS acronyms, compound names, and CAS numbers vary across methods and guidelines, and they also depend on compound form (e.g., neutral vs. salt). The following list uses standardized terminology to allow easy comparison across the methods and guidelines in our comprehensive guide. Please refer to U.S. EPA Method OTM-45 for method-specific nomenclature.
| Acronym | Compound Name | CAS# |
| NEtFOSAA | N-Ethyl perfluorooctanesulfonamidoacetic acid | 2991-50-6 |
| NMeFOSAA | N-Methyl perfluorooctanesulfonamidoacetic acid | 2355-31-9 |
| PFBS | Perfluoro-1-butanesulfonic acid | 375-73-5 |
| PFDA | Perfluorodecanoic acid | 335-76-2 |
| PFDoA | Perfluorododecanoic acid | 307-55-1 |
| PFHpA | Perfluoroheptanoic acid | 375-85-9 |
| PFHxS | Perfluoro-1-hexanesulfonic acid | 355-46-4 |
| PFHxA | Perfluorohexanoic acid | 307-24-4 |
| PFNA | Perfluorononanoic acid | 375-95-1 |
| PFOS | Perfluorooctanesulfonic acid | 1763-23-1 |
| PFOA | Perfluorooctanoic acid | 335-67-1 |
| PFTeDA | Perfluorotetradecanoic acid | 376-06-7 |
| PFTrDA | Perfluorotridecanoic acid | 72629-94-8 |
| PFUnA | Perfluoroundecanoic acid | 2058-94-8 |
| HFPO-DA, GenX | Hexafluoropropylene oxide dimer acid | 13252-13-6 |
| PFBA | Perfluorobutanoic acid | 375-22-4 |
| PFPeA | Perfluoropentanoic acid | 2706-90-3 |
| PFDS | Perfluoro-1-decanesulfonic acid | 335-77-3 |
| PFHpS | Perfluoro-1-heptanesulfonic acid | 375-92-8 |
| PFPeS | Perfluoro-1-pentanesulfonic acid | 2706-91-4 |
| FOSA | Perfluoro-1-octanesulfonamide | 754-91-6 |
| 4:2 FTS | 1H,1H,2H,2H-Perfluorohexane sulfonic acid | 757124-72-4 |
| 6:2 FTS | 1H,1H,2H,2H-Perfluorooctane sulfonic acid | 27619-97-2 |
| 8:2 FTS | 1H,1H,2H,2H-Perfluorodecane sulfonic acid | 39108-34-4 |
| 10:2 FTS | 1H,1H,2H,2H-perfluorododecane sulfonate (10:2) | 120226-60-0 |
| 8:2 FTUCA, FOUEA | 2H-Perfluoro-2-decenoic acid (8:2) | 70887-84-2 |
| NMeFOSA | N-Methylperfluorooctanesulfonamide | 31506-32-8 |
| NEtFOSA | N-Ethylperfluorooctanesulfonamide | 4151-50-2 |
| PFHxDA | Perfluoro-n-hexadecanoic acid | 67905-19-5 |
| PFODA | Perfluoro-n-octadecanoic acid | 16517-11-6 |
| 9Cl-PF3ONS | 9-Chlorohexadecafluoro-3-oxanonane-1-sulfonic acid (F-53B Major) | 756426-58-1 |
| 11Cl-PF3OUdS | 11-Chloroeicosafluoro-3-oxaundecane-1-sulfonic acid (F-53B Minor) | 763051-92-9 |
| ADONA | 4,8-Dioxa-3H-perfluorononanoic acid | 919005-14-4 |
| PFNS | Perfluoro-1-nonanesulfonic acid | 68259-12-1 |
| PFecHS | Decafluoro-4-(pentafluoroethyl)cyclohexanesulfonate | 646-83-3 (67584-42-3) |
| 6:2FTCA, 6:2 FHEA | 2-Perfluorohexyl ethanoic acid | 53826-12-3 |
| 8:2 FOEA | 2-Perfluorooctyl ethanoic acid | 27854-31-5 |
| 10:2 FDEA | 2-Perfluorodecyl ethanoic acid | 53826-13-4 |
| 6:2 FHUEA | 2H-Perfluoro-2-octenoic acid | 70887-88-6 |
| PFEESA | Perfluoro(2-ethoxyethane)sulfonic acid | 113507-82-7 |
| NFDHA | Nonafluoro-3,6-dioxaheptanoic acid | 151772-58-6 |
| PFMPA | Perfluoro-3-methoxypropanoic acid | 377-73-1 |
| PFMBA | Perfluoro-4-methoxybutanoic acid | 863090-89-5 |
| NMeFOSE | 2-(N-methylperfluoro-1-octanesulfonamido)-ethanol | 24448-09-7 |
| NEtFOSE | 2-(N-ethylperfluoro-1-octanesulfonamido)-ethanol | 1691-99-2 |
| PFDoS | Perfluorododecane sulfonate | 79780-39-5 |
| PFDoS | Sodium perfluoro-1-dodecanesulfonate | 1260224-54-1 |
| 3:3 FTCA | 3:3 Fluorotelomer carboxylic acid | 356-02-5 |
| 5:3 FTCA | 5:3 Fluorotelomer carboxylic acid | 914637-49-3 |
| 7:3 FTCA, FHpPA | 7:3 Fluorotelomer carboxylic acid or 3-perfluoroheptyl propanoic acid | 812-70-4 |
Recommended Products for U.S. EPA Method OTM-45 for PFAS Analysis
Analytical LC Columns:
Guard Columns:
Sample Prep:
Vacuum Manifolds:
Vacuum Pumps:
General Lab Supplies:

Need Help with Your PFAS Analysis?
From method selection to product recommendations, our PFAS experts are here to help you navigate
complex workflows and achieve accurate, reliable results.
Discover More Solutions for PFAS Analysis!
Start Saving
Time Today
Develop PFAS Methods in Minutes with our Pro EZLC Chromatogram Modeler
Start Saving Time Today
Develop PFAS Methods in Minutes with our Pro EZLC Chromatogram Modeler


