- Separate d- and l- enantiomers without chiral columns and dedicated instruments.
- Simple precolumn derivatization and dilution sample preparation.
- Accurately distinguish and quantify licit vs. illicit methamphetamine.
Amphetamine and methamphetamine are psychostimulant drugs and occur as two enantiomers, dextrorotary and levorotary, as a result of their chiral center. The dextro-methamphetamine (d-isomer) form is highly abused and is typically found in illicit preparations. However, detection of abuse is complicated because consumption of over-the-counter and prescription medications may yield positive results if the analytical method used cannot distinguish between the enantiomers. Chiral separation of d- and l-methamphetamine and their metabolites d- and l-amphetamine can help determine whether the source was licit or illicit, but chiral columns can be expensive, may necessitate a dedicated instrument, and are not as broadly useful as ubiquitous C18 columns.
In order to provide labs with a high-throughput assay that effectively separates d- and l-amphetamine and methamphetamine enantiomers in urine without the use of a costly and specialized chiral column, we developed the LC-MS/MS method shown here on a standard reversed-phase Raptor C18 column. Our method employs a simple precolumn derivatization followed by dilution and results in a selective, specific analysis of d- and l-amphetamine and methamphetamine enantiomers that is free from sample matrix interferences. Separation was achieved within a total run time of seven minutes, and quantitation in urine was performed across a linear range of 50-5000 ng/mL. Validation across this range demonstrated that this method provides reliable analysis of d- and l-amphetamine and methamphetamine enantiomers in a workflow and time frame suitable for high-throughput clinical and forensic toxicology labs.
LC_CF0737
Peaks
*Analytes are DNPA derivatives.
Conditions
| Raptor C18 (cat.# 9304A12) |
| 100 mm x 2.1 mm ID |
| 2.7 µm |
| 90 Å |
|
| Raptor C18 EXP guard column cartridge 5 mm, 2.1 mm ID, 2.7 µm (cat.# 9304A0252) |
| 35 °C |
| 500 ng/mL in urine |
| | 10 µL |
| 0.1% Formic acid in water |
| 0.1% Formic acid in methanol |
| | Time (min) | Flow (mL/min) | %A | %B |
|---|
| 0.00 | 0.5 | 40 | 60 | | 5.00 | 0.5 | 40 | 60 | | 5.01 | 0.5 | 10 | 90 | | 5.50 | 0.5 | 10 | 90 | | 5.51 | 0.5 | 40 | 60 | | 7.00 | 0.5 | 40 | 60 |
|
Products Mentioned
Raptor C18, 2.7 µm, 100 x 2.1 mm HPLC Column
Raptor C18 EXP Guard Column Cartridge, 2.7 µm, 5 x 2.1 mm, 3-pk.
EXP Direct Connect Holder for EXP Guard Cartridges, Includes Fitting & Ferrules
Thomson SINGLE StEP Standard Filter Vials with Screw-Top Caps, 0.2 µm, PTFE w/preslit cap, Green Cap, 100-pk.
-
Restek is a leading provider of chromatography columns, accessories, and certified reference materials. Trust Restek for reliable, high-quality analytical solutions.
View Articles
CFSS2974
Version: A