Method 8081B: Organochlorine Pesticides Analysis by GC (Extended)

26 Oct 2020

feature EVSS1013

Many labs extend the list of compounds tested using Method 8081B to include other organochlorine pesticides. Rtx-CLPesticides and Rtx-CLPesticides2 columns are ideal for these determinations and can be used in a parallel dual-column configuration to reliably quantify 38 commonly analyzed organochlorine pesticides, as well as many other compounds. Full separations are achieved between both columns and the unique, optimized selectivity allows additional compounds to be included. (Note that since extended compound lists vary among laboratories the lists compared here are not identical.)

VendorColumn PairAnalysis Time (min)CoelutionsRestek Advantage
RestekRtx-CLPesticides /
Rtx-CLPesticides2
24 / 231 / 2Best balance of speed and selectivity. All compounds are resolved on at least one column.
Competitor Acolumn 1 / column 227 / 290 / 3
Competitor Bcolumn 1 / column 2NDP / 16NDP / 3

Comparison based on published competitor data. All columns tested were 0.32 mm ID. NDP = no data published

The unique selectivity of the Rtx-CLPesticides column pair allows easy separation of extended lists of organochlorine pesticides.
Organochlorine Pesticides by EPA Method 8081B on Rtx-CLPesticides/Rtx-CLPesticides2 (extended list)

GC_EV1193

Conditions

ColumnsRtx-CLPesticides2 30 m, 0.32 mm ID, 0.25 µm (cat.# 11324)
and Rtx-CLPesticides 30 m, 0.32 mm ID, 0.32 µm (cat.# 11141)
using Rxi guard column 5 m, 0.32 mm ID (cat.# 10039)
with deactivated universal “Y” Press-Tight connector (cat.# 20405-261)
Standard/Sample
Organochlorine pesticide mix AB #2 (cat.# 32292)
Organochlorine pesticide mix C #1 (cat.# 32296)
2,4,5,6-Tetrachloro-m-xylene (cat.# 32027)
Decachlorobiphenyl (BZ #209) (cat.# 32029)
Chlorpyrifos (cat.# 32212)
2,4′-DDE (cat.# 32099)
2,4′-DDD (cat.# 32098)
Chlorobenzilate
Captan (custom)
cis-Nonachlor (custom)
Mirex (custom)
Oxychlordone (custom)
Diallate (custom)
Diluent:Hexane
Injection
Inj. Vol.:1 µL splitless (hold 0.3 min)
Liner:Cyclo double taper (4 mm) (cat.# 20896)
Inj. Temp.:250 °C
Oven
Oven Temp.:80 °C to 175 °C at 8 °C/min to 195 °C at 11 °C/min to 235 °C at 7 °C/min to 320 °C at 25 °C/min (hold 3 min)
Carrier GasHe, constant flow
Linear Velocity:28 cm/sec
DetectorMicro-ECD @ 330 °C
Make-up Gas Flow Rate:60 mL/min
Make-up Gas Type:N2
InstrumentAgilent/HP6890 GC
NotesThis chromatogram was obtained using an Agilent micro-ECD. To obtain comparable results, you will need to employ a micro-ECD in addition to confirmational dual columns connected to a 5-meter guard column using a “Y” Press-Tight connector.

Reference standard cat.# 32211 was used to produce this chromatogram, but has since been discontinued. For assistance choosing a replacement for this application, contact Restek Technical Service or your local Restek representative.

Products Mentioned


Rtx-CLPesticides GC Capillary Column, 15 m, 0.25 mm ID, 0.25 µm
Rtx-CLPesticides2 GC Capillary Column, 15 m, 0.32 mm ID, 0.25 µm
Rtx-CLPesticides GC Capillary Column Kit, 0.32 mm ID, 0.32 µm

Author

  • Restek Corporation

    Restek is a leading provider of chromatography columns, accessories, and certified reference materials. Trust Restek for reliable, high-quality analytical solutions.

    View Articles
EVAR4500