- Identifying vial cap septum bleed prevents lengthy inlet troubleshooting.
- Reduce interference with correct solvent-septum compatibility.
When septum bleed occurs, it is observed as sharp, repetitive peaks in high temperature portions of an analysis. Bleed peaks can come from either the inlet septum or the vial cap septum. Interfering peaks and inaccurate data can result, so it is important to correctly identify the source and to understand how to eliminate or minimize the bleed.
Diagnose GC Septum Bleed: What’s the Source?
The bleed from either septum shows a similar pattern (Figure 1), but it is easy to determine the source with a simple test. Isolate the inlet by setting the instrument to perform a run without an injection. Perform an analysis; if the bleed disappears, then the vial cap septum was the source. Determining if the vial cap septum is the source of the bleed can save time by preventing unnecessary troubleshooting and maintenance. If the problem is in the inlet, review our technical article before replacing the septum. However, if the vial cap septum is causing the bleed, the issue can be eliminated or minimized with the following considerations.

GC_EX00980
Conditions
| Column | Rtx-5MS, 30 m, 0.25 mm ID, 0.25 µm (cat.# 12623) |
|---|---|
| Standard/Sample | Methylene chloride blank |
| Injection | |
| Inj. Vol.: | 1.0 µL split (split ratio 10:1) |
| Liner: | 4.0 mm ID straight inlet liner w/wool (standard deactivation) (cat.# 20782) |
| Inj. Temp.: | 240 °C |
| Oven | |
| Oven Temp.: | 70 °C (hold 1 min) to 290 °C at 20 °C/min (hold 13 min) |
| Carrier Gas | He, constant flow |
| Flow Rate: | 1.2 mL/min |
| Detector | FID @ 250 °C |
|---|
Check Solvent-Septum Compatibility
Once you’ve diagnosed the GC septum bleed source as the vial cap, the first step in resolving the issue is to make sure that the vial cap septum material is compatible with your injection solvent. While septum bleed is not excessive most of the time, when a solvent and vial cap septum are incompatible, extreme contamination can occur. Figure 2 compares the first and fifth injections from a vial containing a derivatized amphetamine sample. In this case, the septum bleed peaks are almost as large as the analyte peaks. This level of bleed can interfere with analyses, especially those geared for trace levels. Reduce the risk of septum bleed by using a compatibility chart to determine which septum material is compatible with the sample solvent used.

GC_EX00981
Peaks
| Peaks | |
|---|---|
| 1. | Amphetamine |
| 2. | Methamphetamine |
| 3. | MDA |
| 4. | MDMA |
| 5. | MDEA |
Conditions
| Column | Rtx-5MS, 30 m, 0.25 mm ID, 0.25 µm (cat.# 12623) |
|---|---|
| Standard/Sample | Amphetamine, methamphetamine, MDA, MDMA, and MDEA extracted from methanol and HFAA derivatized |
| Conc.: | 100 µg/mL each |
| Injection | |
| Inj. Vol.: | 1 µL splitless (hold 0.5 min) |
| Liner: | 3.5 mm Custom splitless inlet liner w/ standard deactivated wool |
| Inj. Temp.: | 220 °C |
| Oven | |
| Oven Temp.: | 70 °C (hold 1 min) to 290 °C at 15 °C/min (hold 4 min) |
| Carrier Gas | He, constant flow |
| Flow Rate: | 1.25 mL/min |
| Detector | MS |
|---|---|
| Mode: | Scan |
| Transfer Line Temp.: | 280 °C |
| Ionization Mode: | EI |
| Scan Range: | 43-450 amu |
Use Lined Septa
Most vial cap septa are lined with a protective layer of polytetrafluoroethylene (PTFE) to prevent solvent attack. As shown in Figure 3, PTFE effectively prevents septum breakdown due to solvent exposure. In comparison, unlined vial cap septa exhibit bleed after just 24 hours at room temperature. Bleed levels for unlined vial cap septa varied by material, but even a low level of bleed can interfere with integration and is of particular concern for trace analyses (Figure 4).


Consider Resealability
Multiple injections can core the vial cap septum and lead to significant GC septum bleed. Resistance to coring varies by septum material (Figure 5). Coring can be minimized by preparing separate vials for replicate injections, when feasible, and by carefully considering the type of septum material when multiple injections are necessary. Septum resealability also affects evaporative loss, which can be a significant source of error for low volume samples. For example, a relatively nonvolatile analyte in a volatile solvent can concentrate significantly due to evaporative loss (Figure 6). Vials should be recapped when necessary for extended runs or long-term storage.


Conclusion
When septum bleed occurs, it is easy to assume the inlet septum is the source because the vial cap septum often is not considered. However, correctly identifying the GC septum bleed source can save time and effort by preventing unnecessary inlet maintenance. Effectively and efficiently reducing interfering peaks by controlling septum bleed can significantly improve analytical performance, particularly for trace analyses.