Blogs

Maximum Temperatures of Packed Columns Part 1 – Porous Polymers

15 Jun 2015

We often get asked about the maximum temperature limit of our packed columns, so I decided to write a post series that will provide this information for the majority of the packed columns which we sell.  I thought I would start with the least complicated packed columns, those whose packings are typically not coated with a liquid phase.  These packings include porous polymers.

For this post, I will only be focusing on HayeSep, Porapak, and Tenax packings.  I decided not to include Chromosorb porous polymers because they are either in short supply or currently unavailable.

I am going to make the following assumption:  The column’s maximum temperature is limited to the packing, and not the column’s tubing; end plugs; fittings/adaptors; and/or ferrules used for installation. If you have a packed column that includes nonmetal tubing or end plugs other than glass wool or metal, contact Restek.

Please note:  The catalog (part) numbers shown below are just for the packing and not any packed column.

Keep watching for future posts on maximum temperatures of packed columns.  Thank you.

HayeSep Packing Materials

Descriptiontemp. limitsg/btl.Mesh 60/80
cat.#
Mesh 80/100
cat.#
Mesh 100/120
cat.#
HayeSep A165 °C24 gN/A2503225033
HayeSep B190 °C24 gN/AN/AN/A
HayeSep C250 °C24 gNot Available–Contact your local Restek Representative for suggested replacement.
HayeSep D290 °C24 g255632503825039
HayeSep DIP290 °C24 gN/AN/AN/A
HayeSep DB290 °C24 gN/A25568N/A
HayeSep DOX(Use HayeSep DB)
HayeSep N165 °C24 g255702504525046
HayeSep P250 °C24 gNot Available–Restek recommends Porapak P as a replacement.
HayeSep Q275 °C24 g25572N/A25050
HayeSep R250 °C24 gNot Available–Restek recommends Porapak R as a replacement.
HayeSep S250 °C24 gN/AN/A25054
HayeSep T165 °C24 g25575N/A25056

Porapak Packing Materials

Descriptiontemp. limitsg/btl.Mesh 50/80
cat.#
Mesh 80/100
cat.#
Mesh 100/120
cat.#
Porapak P250 °C20 gAvailable as Custom–Contact your local Restek Representative for ordering information.
Porapak PS250 °C20 g
Porapak Q250 °C26 g
Porapak QS250 °C26 g
Porapak R250 °C24 g
Porapak S250 °C26 g
Porapak N190 °C29 g
Porapak T190 °C31 g

Tenax Packing Materials

Descriptiontemp. limitsg/btl.Mesh 60/80
cat.#
Mesh 80/100
cat.#
Tenax-TA350 °C10 g2555025551
Tenax-GR350 °C10 gNot Available–Contact your local Restek Representative for suggested replacement.