Anwendungshinweise

High-Throughput Analysis of Immunosuppressive Drugs from Whole Blood by LC-MS/MS

19 Oct 2020

feature CFAN2666

Abstract

The success of organ transplant therapy depends in large part on the accurate and timely analysis of immunosuppressive drugs. In this study, we developed a fast, accurate method for the analysis of cyclosporin A, tacrolimus, sirolimus, and everolimus in whole blood. The method pairs a single precipitation step with LC-MS/MS analysis using a Raptor Biphenyl column. A fast, 3-minute analysis time was obtained with no interference from matrix components. Excellent results were obtained for linearity, robustness, accuracy, and precision, demonstrating that the method is suitable for high-throughput therapeutic drug monitoring.

Introduction

Immunosuppressive drugs are used to suppress the body’s immune response and are typically administered to prevent the rejection of transplanted organs or tissues. Cyclosporin A, tacrolimus, sirolimus, and everolimus are four of the most commonly used drugs in the therapy of organ transplantation. Their structures are shown in Figure 1. Cyclosporin A and tacrolimus are classified as calcineurin inhibitors, and sirolimus and everolimus are grouped as mTOR inhibitors. These two classes of drugs can be used in combination for synergistic blocking of T cell activation and proliferation. Due to their pharmacokinetic variabilities and narrow therapeutic indexes, time-sensitive and highly accurate therapeutic drug monitoring is necessary, not only to prevent rejection but also to minimize toxic side effects. Therefore, a fast and accurate measurement of drug concentration is critical to assist clinicians with timely and proper treatment of patients.

Several commercial products are available for the simultaneous analysis of cyclosporin A, tacrolimus, sirolimus, and everolimus in whole blood. However, most of these commercial kits require an additional in-line trap column or SPE sample cleanup prior to chromatographic analysis. Here, we reduced sample preparation time and effort by pairing a single precipitation step with a Raptor Biphenyl analytical column. This analytical column was chosen because it provides suitable retention and selectivity for these compounds so that a short analysis time (3 minutes) can be achieved without encountering matrix interference. By combining a simple sample preparation step and a fast chromatographic elution, a high-throughput analysis was established for the simultaneous analysis of four immunosuppressive drugs in human whole blood.

Figure 1: Analyte Structures.
figure article CFAN2666 01

Experimental

Calibration Standards and Blood Control Samples

Human whole blood was fortified with four immunosuppressive drugs to prepare the calibration standards and QC samples. For quantitation, cyclosporin D was used as the internal standard for cyclosporin A, and ascomycin was used as the internal standard for tacrolimus, sirolimus, and everolimus. The general therapeutic ranges are 25–400 ng/mL for cyclosporin A, 5–20 ng/mL for tacrolimus and sirolimus, and 3–8 ng/mL for everolimus. Accordingly, the calibration standard ranges were evaluated from 10 to 1,000 ng/mL for cyclosporin A and from 1 to 100 ng/mL for tacrolimus, sirolimus, and everolimus. Three QC levels were prepared: 15, 150, and 800 ng/mL for cyclosporin A and 5, 15, and 80 ng/mL for tacrolimus, sirolimus, and everolimus.

Sample Preparation

Each blood sample (100 µL) was mixed with 200 µL of precipitation solution [1:4 (v/v) 0.2 M ZnSO4:methanol] containing 50 ng/mL of cyclosporin D and 5 ng/mL of ascomycin. The mixture was vortexed for 20 seconds at 3,000 rpm and then centrifuged for 10 minutes at 4,300 rpm. The supernatant was directly injected (5 µL) onto a Raptor Biphenyl (2.7 µm, 50 mm x 2.1 mm) column equipped with a Raptor Biphenyl EXP (2.7 µm, 5 mm x 2.1 mm) guard column for analysis.

LC-MS/MS Analysis

Analysis of immunosuppressive drugs (cyclosporin A, tacrolimus, sirolimus, and everolimus) was conducted using a Waters Xevo TQ-S mass spectrometer with an ACQUITY UPLC under the conditions shown below. The analyte transitions that were monitored are presented in Table I.

Analytical column:Raptor Biphenyl, 2.7 µm, 50 mm x 2.1 mm (cat.# 9309A52)
Guard column:Raptor Biphenyl EXP, 2.7 µm, 5 mm x 2.1 mm (cat.# 9309A0252)
Mobile phase A:0.05% Formic acid, 5 mM ammonium formate in water
Mobile phase B:Methanol
GradientTime (min)%B
 0.0060
 2.00100
 2.0160
 3.0060
Flow rate:0.5 mL/min
Injection volume:5 µL
Column temp.:70 °C
Ion mode:Positive ESI

Table I: Analyte Transitions

AnalytePrecursor IonProduct Ion QuantifierProduct Ion Qualifier
Cyclosporin A1219.831202.871184.58
Everolimus975.68908.62926.57
Sirolimus931.63864.57882.58
Tacrolimus821.61768.51786.52
Cyclosporin D1233.911216.88
Ascomycin809.53756.50

Results and Discussion

Chromatographic Performance

A fast 3-minute chromatographic analysis of all four immunosuppressive drugs (Figure 2) was achieved with direct injection of the supernatant without any further in- or off-line sample treatment. No matrix interference was observed for all four analytes and internal standards (Figure 3) with the established sample preparation and chromatographic methods. The Raptor Biphenyl column provided more retention than a C18 phase and less retention than a fluorophenyl phase, resulting in optimum retention and selectivity for cyclosporin A, tacrolimus, sirolimus, and everolimus.

Figure 2: Fortified Human Whole Blood at 10 ng/mL.
Immunosuppressive Drugs in Blood on Raptor Biphenyl by LC-MS/MS

LC_CF0686

Peaks

PeakstR (min)Conc.
(ng/mL)
Precursor IonProduct Ion
1.Ascomycin1.4510809.53756.50
2.Tacrolimus1.4710821.61768.51
3.Sirolimus1.6410931.63864.57
4.Everolimus1.6910975.68908.62
5.Cyclosporin A1.83101219.831202.87
6.Cyclosporin D1.891001233.911216.88
Figure 3: Blank Human Whole Blood.
Blank Blood Analysis for Immunosuppressive Drugs Method Validation

LC_CF0687

Peaks

PeaksConc.
(ng/mL)
Precursor IonProduct Ion
1.Ascomycin0809.53756.50
2.Tacrolimus0821.61768.51
3.Sirolimus0931.63864.57
4.Everolimus0975.68908.62
5.Cyclosporin A01219.831202.87
6.Cyclosporin D01233.911216.88

Method Robustness

Following 500 injections of a prepared whole blood standard (25 ng/mL), the chromatographic peaks of all four immunosuppressive drug analytes maintained the initial peak shape, retention time, and intensity (Figure 4). The maximum system pressure also remained at the same level, indicating no column clogging had occurred.

Figure 4: Robust Column Performance Over 500 Injections.
Column Lifetime Evaluation for Immunosuppressive Drug Analysis

LC_CF0688

Peaks

PeaksConc.
(ng/mL)
Precursor IonProduct Ion
1.Ascomycin10809.53756.50
2.Tacrolimus25821.61768.51
3.Sirolimus25931.63864.57
4.Everolimus25975.68908.62
5.Cyclosporin A251219.831202.87
6.Cyclosporin D1001233.911216.88
See chromatograms for retention times.

Conditions

ColumnRaptor Biphenyl (cat.# 9309A52)
Dimensions:50 mm x 2.1 mm ID
Particle Size:2.7 µm
Pore Size:90 Å
Guard Column:Raptor Biphenyl EXP guard column cartridge 5 mm, 2.1 mm ID, 2.7 µm (cat.# 9309A0252)
Temp.:70 °C
Standard/Sample
Inj. Vol.:5 µL
Mobile Phase
A:0.05% Formic acid, 5 mM ammonium formate in water
B:Methanol
Time (min)Flow (mL/min)%A%B
0.000.54060
2.000.50100
2.010.54060
3.000.54060
Max Pressure:305 bar
DetectorMS/MS
Ion Mode:ESI+
Mode:MRM
Instrument57I
Sample PreparationHuman whole blood was fortified with 4 analytes at 25 ng/mL. For quantitation, cyclosporin D was used as the internal standard for cyclosporin A and ascomycin was used as the internal standard for tacrolimus, sirolimus, and everolimus. The blood sample (2.5 mL) was mixed with 5 mL of precipitation solution (1:4 v/v 0.2 M ZnSO4:methanol) containing 50 ng/mL of cyclosporin D and 5 ng/mL of ascomycin. The mixture was vortexed for 20 seconds at 3000 rpm and then centrifuged for 10 minutes at 4300 rpm. The supernatant was directly injected (5 μL) for analysis.

Linearity

Using a 1/x2 weighted linear regression for cyclosporin A and a 1/x weighted linear regression for tacrolimus, sirolimus, and everolimus, all four immunosuppressive drug compounds showed very good linearity with r values of 0.999 or greater, and with deviations of <10% (Figure 5). The signal-to-noise values of the lowest standard samples were from 100 to 300, indicating that this method could be used for the detection of much lower concentrations if necessary.

Figure 5: Standard Curves.
figure article CFAN2666 05

Accuracy and Precision

Precision and accuracy analyses were performed on three different days. Method accuracy was demonstrated with recovery values that were within 10% of the nominal concentration for all QC levels. The %RSD across all analytes and fortification levels was 0.2–4.0% and 1.2–5.4% for intraday and interday evaluations respectively, indicating good method precision (Table II).

Table II: Accuracy and Precision of QC Samples

 QC Level 1 (5.0 ng/mL)
(Cyclosporin A: 15 ng/mL)
QC Level 2 (15.0 ng/mL)
(Cyclosporin A: 150 ng/mL)
QC Level 3 (80 ng/mL)
(Cyclosporin A: 800 ng/mL)
AnalyteAverage Conc. (ng/mL)Average Accuracy (%)%RSDAverage Conc. (ng/mL)Average Accuracy (%)%RSDAverage Conc. (ng/mL)Average Accuracy (%)%RSD
Cyclosporin A15.099.95.4152.3101.51.2795.699.52.1
Tacrolimus5.2103.61.814.395.11.981.8102.31.3
Sirolimus5.2103.94.314.496.22.783.0103.72.1
Everolimus5.1101.12.014.294.82.180.8101.02.8

Conclusion

It was demonstrated that the Raptor Biphenyl column provides excellent chromatographic performance for the rapid and accurate analysis of cyclosporin A, tacrolimus, sirolimus, and everolimus in human whole blood. With a fast and simple sample preparation procedure and a 3-minute analysis time, the method established here provides high-throughput therapeutic drug monitoring for these commonly used immunosuppressive drugs.

Products Mentioned


Raptor Biphenyl, 2.7 µm, 50×2.1 mm HPLC-Säule
Raptor Biphenyl EXP Vorsäulenkartusche, 2.7 µm, 5 x 2.1 mm, 3er Pack

Authors

  • Shun-Hsin Liang, PhD

    Shun-Hsin is a senior principal scientist in LC Solutions at Restek. He received his bachelor’s degree from the National Taiwan University in 1988 and obtained his PhD from Michigan State University in 1996. He performed postdoctoral research at the University of Michigan from 1996 to 2000 for oncology studies. In 2001, he was appointed as research faculty at The Pennsylvania State University and focused on molecular toxicology research. In 2006, he joined MPI Research Inc. as a senior research scientist and was a study director for GLP analytical projects. In 2013, Dr. Liang joined the LC Solutions department at Restek and specialized in developing application methods across the fields of environmental, food safety, and life sciences.

    View all posts
  • Frances Carroll
CFAN2666