Blogs

Chiral Separations 1: How do our GC chiral columns compare?

17 Aug 2021

Predicting chiral separations is challenging. Using standard capillary columns, general predictions about separations can be made based on the solubility of compounds in different phases; however, chiral predictions are far more complicated.

To make this decision somewhat easier, we compiled resolution values for the enantiomers of 21 chiral compounds with different functional groups, such as terpenes, alcohols or lactones.

 Column Resolution Factors
 CompoundsRt-βDEX smRt-βDEX seRt-βDEX spRt-βDEX saRt-βDEX cstRt-βDEX m
Terpenes1α-pinene3.140.82nsns0.823.19
 2limonene5.608.051.22ns2.041.41
Alcohols31-octen-3-ol1.08nsns2.00nsns
 4linalool3.105.961.791.602.251.01
 5α-terpineol4.725.201.292.822.691.54
 6terpinen-4-ol1.901.92ns2.47ns1.41
 7isoborneol3.763.35ns0.791.87t2.15
 8β-citronellol0.800.89ns0.98nsns
 9menthol1.241.241.070.930.891.58
 102,3-butanediol6.447.101.462.611.272.43
 111-phenylethanol7.526.52ns6.434.875.88
Ketones12camphor1.702.130.504.202.22tns
 13α-ionone5.673.310.804.691.372.81
14menthone0.595.76ns1.162.604.11
Lactones15γ-nonalactone4.195.072.114.003.821.03
16γ-undecalactone2.393.241.203.653.18ns
 17δ-dodecalactone0.80nsns1.911.69ns
 18ethyl-2-methylbutyrate3.944.66nsnsns0.92
Esters19linalyl acetatens2.36nsnsnsns
 20styrene oxide4.5310.772.862.261.16t3.03
Epoxides21trans-linalool oxide9.712.96ns1.203.02t7.44
 21cis-linalool oxide6.064.28ns0.912.19t4.38
GNBL4618