{"id":43686,"date":"2020-10-22T00:00:00","date_gmt":"2020-10-22T00:00:00","guid":{"rendered":"https:\/\/discover.restek.com\/uncategorized\/totally-reliable-column-to-column-performance\/"},"modified":"2026-01-03T16:26:42","modified_gmt":"2026-01-03T16:26:42","slug":"totally-reliable-column-to-column-performance","status":"publish","type":"post","link":"https:\/\/discover.restek.com\/de\/articles\/gnar3295\/totally-reliable-column-to-column-performance","title":{"rendered":"Totally Reliable Column-to-Column Performance"},"content":{"rendered":"\n<p class=\"wp-block-paragraph\">Chromatographers today need assurances that every column they receive is going to perform in the same way as the column it replaces. The processes used by Restek to make Rxi columns lead the industry in manufacturing precision and quality control, resulting in the most consistently performing columns available. To demonstrate this, test data for one of the most popular columns (Rxi-5ms, 30m x 0.25mm, 0.25\u00b5m) are shown below. Statistics are based on data for more than 2,000 columns manufactured over a 2-year period.<\/p>\n\n\n\n<div style=\"height:20px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n\n\n\n<h2 class=\"wp-block-heading\">Efficiency (plates\/meter)<\/h2>\n\n\n\n<p class=\"wp-block-paragraph\">Rxi columns will separate analytes better than competitor columns because they have more plates\/meter. Typical values for columns of the same dimensions from some vendors (Agilent, J&amp;W, Varian) are listed at 3,900 plates\/meter; other suppliers do not even specify a value, meaning there is no efficiency guarantee whatsoever. In contrast, the Rxi-5ms column has an average of 4,270 plates\/meter with a very tight distribution (Figure 1). This means that the Rxi coating process is well-controlled and highly reproducible.<\/p>\n\n\n<div class=\"wp-block-custom-chromatogram-article-top\"><div class=\"chromatogram-article-placeholder\"><div class=\"figure-heading\"><strong>Figure 1<\/strong>\u00a0\u00a0Rxi columns have higher efficiency and reliably produce better separations than non-Restek columns. (Average plates\/meter: Rxi-5ms = 4,270; competitors = 3,900.)<\/div><div class='chromatogram-article-inner-full'><div class=\"chromatogram-article-inner\">\n\n<figure class=\"wp-block-image\"><img decoding=\"async\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-GNAR33295-01.jpg\" alt=\"figure-article-GNAR33295-01.jpg\" title=\"-\"><\/figure>\n\n<\/div><\/div><\/div>\n\n\n<div style=\"height:20px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n\n\n\n<p class=\"wp-block-paragraph\">In practical terms, this means peaks will be better resolved on a Restek column than on a competitor column, simply because the Rxi process consistently produces a more efficient column (Figure 2). This allows you to run analyses at higher than optimal linear velocity and reduce analysis time\u2014easily by 10%\u2014without sacrificing separation power.<\/p>\n\n\n<div class=\"wp-block-custom-chromatogram-article-top\"><div class=\"chromatogram-article-placeholder\"><div class=\"figure-heading\"><strong>Figure 2<\/strong>\u00a0\u00a0More plates-per-meter mean better separations on an Rxi column.<\/div><div class='chromatogram-article-inner-full'><div class=\"chromatogram-article-inner\">\n\n<figure class=\"wp-block-image\"><img decoding=\"async\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-GNAR33295-02.jpg\" alt=\"figure-article-GNAR33295-02.jpg\" title=\"-\"><\/figure>\n\n<\/div><\/div><\/div>\n\n\n<div style=\"height:20px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n\n\n\n<h2 class=\"wp-block-heading\">Retention Factor (k)<\/h2>\n\n\n\n<p class=\"wp-block-paragraph\">Restek precisely determines the amount of stationary phase that is coated in each column. As a result, retention factor is tightly controlled with an average of 6.44 and standard deviation of 0.11 for pentadecane on the Rxi-5ms column (Figure 3). Narrow distribution ensures consistent polymer film thickness, which means retention times will be consistent and predictable, column-to-column.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Competitor columns typically have variations of \u00b17%. If we apply that standard to the Restek column, we calculate a window of \u00b10.45 around the average for pentadecane. As shown in Figure 3, this tolerance window is much wider than the entire distribution for the Rxi column, meaning that there will be much less variation in retention times on an Rxi column than on a competitor column. This is confirmed in Figure 4, which illustrates that retention times are much more consistent\u2014and thus predictable\u2014on the Rxi column.<\/p>\n\n\n<div class=\"wp-block-custom-chromatogram-article-top\"><div class=\"chromatogram-article-placeholder\"><div class=\"figure-heading\"><strong>Figure 3<\/strong>\u00a0Retention factor data for the Rxi-5ms, 30m x 0.25mm, 0.25\u00b5m column. Industry standard minimum and maximum for quality control is noted at the 7% lines.<\/div><div class='chromatogram-article-inner-full'><div class=\"chromatogram-article-inner\">\n\n<figure class=\"wp-block-image\"><img decoding=\"async\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-GNAR33295-03.jpg\" alt=\"figure-article-GNAR33295-03.jpg\" title=\"-\"><\/figure>\n\n<\/div><\/div><\/div>\n\n<div class=\"wp-block-custom-chromatogram-article-top\"><div class=\"chromatogram-article-placeholder\"><div class=\"figure-heading\"><strong>Figure 4<\/strong>\u00a0\u00a0A tightly controlled Rxi coating process ensures more consistent and predictable retention times on Restek columns than on columns produced by other suppliers.<\/div><div class='chromatogram-article-inner-full'><div class=\"chromatogram-article-inner\">\n\n<figure class=\"wp-block-image\"><img decoding=\"async\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-GNAR33295-04.jpg\" alt=\"figure-article-GNAR33295-04.jpg\" title=\"-\"><\/figure>\n\n<\/div><\/div><\/div>\n\n\n<div style=\"height:20px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n\n\n\n<h2 class=\"wp-block-heading\">Retention Indices (RI)<\/h2>\n\n\n\n<p class=\"wp-block-paragraph\">Most commercial columns have a tolerance on retention indices which is typically within \u00b10.5 units for acenaphthalene and undecanol. As shown in Figure 5 the variation in RI for the Rxi-5ms column is much less than the industry standard (1-undecanol: average = 1372.8; standard deviation \u00b10.15). This narrow distribution ensures consistent column selectivity, which means the components always elute in the same position relative to the other components. Resolution is, therefore, more consistent column-to-column on the Restek column than on columns from other manufacturers.<\/p>\n\n\n<div class=\"wp-block-custom-chromatogram-article-top\"><div class=\"chromatogram-article-placeholder\"><div class=\"figure-heading\"><strong>Figure 5<\/strong>\u00a0Relative retention is much more consistent on Restek columns than competitor columns because the variance is more tightly controlled. (Industry standard minimum and maximum for quality control is indicated by the \u00b10.5 lines.)<\/div><div class='chromatogram-article-inner-full'><div class=\"chromatogram-article-inner\">\n\n<figure class=\"wp-block-image\"><img decoding=\"async\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-GNAR33295-05.jpg\" alt=\"figure-article-GNAR33295-05.jpg\" title=\"-\"><\/figure>\n\n<\/div><\/div><\/div>\n\n\n<div style=\"height:20px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n\n\n\n<h2 class=\"wp-block-heading\">Bleed<\/h2>\n\n\n\n<p class=\"wp-block-paragraph\">The industry standard for acceptable bleed is 4pA for the 30m x 0.25mm, 0.25\u00b5m column. The Rxi-5ms column consistently performs below this criterion. As shown in Figure 6, the average bleed is 1.4pA with a standard deviation of 0.5.<\/p>\n\n\n<div class=\"wp-block-custom-chromatogram-article-top\"><div class=\"chromatogram-article-placeholder\"><div class=\"figure-heading\"><strong>Figure 6<\/strong>\u00a0Bleed for the Rxi-5ms column (30m x 0.25mm, 0.25\u00b5m) is consistently less than the industry standard of 4pA.<\/div><div class='chromatogram-article-inner-full'><div class=\"chromatogram-article-inner\">\n\n<figure class=\"wp-block-image\"><img decoding=\"async\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-GNAR33295-06.jpg\" alt=\"figure-article-GNAR33295-06.jpg\" title=\"-\"><\/figure>\n\n<\/div><\/div><\/div>\n\n\n<div style=\"height:20px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n\n\n\n<p class=\"wp-block-paragraph\">All Rxi columns are guaranteed to be low bleed because Restek does not batch test for bleed to reduce testing costs. To assess bleed, first every Rxi column is conditioned for 16 hours at maximum isothermal temperature. After this, each column is individually tested for bleed using calibrated equipment. Bleed values obtained are listed on the test report for every Rxi column.&nbsp;<\/p>\n\n\n\n<h2 class=\"wp-block-heading\">Inertness<\/h2>\n\n\n\n<p class=\"wp-block-paragraph\">Column inertness has been the biggest challenge in making fused silica columns until now\u2014Rxi column technology has made it possible to make highly inert columns in a remarkably consistent fashion. Inertness is critical as it affects peak shape, response, retention time stability, integration, and data accuracy (a detailed discussion of inertness can be found <a href=\"https:\/\/www.restek.com\/articles\/superior-inertness-ensures-accurate-trace-level-analysis\/\" target=\"_blank\" rel=\"noopener\">here<\/a>).&nbsp;Each Rxi column is evaluated with a rigorous test mixture designed to indicate different types of column activity. As shown in Table I and Figure 7, peaks for all test probes are highly symmetric and standard deviations are very low, indicating extremely consistent column-to-column inertness.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\"><strong>Table 1<\/strong>&nbsp;&nbsp;Rxi columns are extremely inert and rigorously tested for different types of activity.<\/p>\n\n\n\n<figure class=\"wp-block-table\"><table class=\"has-fixed-layout\"><tbody><tr><td><strong>Test Probe&nbsp;&nbsp;&nbsp;<\/strong><\/td><td><strong>Activity Indicated<\/strong><\/td><td><strong>Peak Asymmetry<br>(average)<\/strong><\/td><td><strong>Peak Asymmetry<br>(standard deviation)<\/strong><\/td><\/tr><tr><td>1,6-hexanediol<\/td><td>silanol groups (hidden)<\/td><td>1.04<\/td><td>0.062<\/td><\/tr><tr><td>1-undecanol<\/td><td>silanol groups<\/td><td>1.00<\/td><td>0.029<\/td><\/tr><tr><td>4-chlorophenol<\/td><td>basic active sites<\/td><td>1.02<\/td><td>0.041<\/td><\/tr><tr><td>1-decylamine<\/td><td>acidic active sites<\/td><td>1.14<\/td><td>0.055<\/td><\/tr><\/tbody><\/table><\/figure>\n\n\n\n<p class=\"wp-block-paragraph\">Data shown are for Rxi-5ms columns (30m x 0.25mm, 0.25\u00b5m).<\/p>\n\n\n<div class=\"wp-block-custom-chromatogram-article-top\"><div class=\"chromatogram-article-placeholder\"><div class=\"figure-heading\"><strong>Figure 7<\/strong>\u00a0\u00a0Tight distributions and excellent peak asymmetry values assure consistently inert columns<\/div><div class='chromatogram-article-inner-full'><div class=\"chromatogram-article-inner\">\n\n<p class=\"wp-block-paragraph\">A. 1,6-Hexanediol asymmetry plot<\/p>\n\n\n\n<figure class=\"wp-block-image size-large\"><img loading=\"lazy\" decoding=\"async\" width=\"1024\" height=\"570\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-GNAR33295-10-1024x570.jpg\" alt=\"D. 1-Decylamine asymmetry plot\" class=\"wp-image-21646\" title=\"-\" srcset=\"https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-GNAR33295-10-1024x570.jpg 1024w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-GNAR33295-10-300x167.jpg 300w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-GNAR33295-10-768x428.jpg 768w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-GNAR33295-10-1536x855.jpg 1536w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-GNAR33295-10.jpg 1800w\" sizes=\"auto, (max-width: 1024px) 100vw, 1024px\" \/><\/figure>\n\n\n\n<p class=\"wp-block-paragraph\">B. 1-Undecanol asymmetry plot<\/p>\n\n\n\n<figure class=\"wp-block-image size-large\"><img loading=\"lazy\" decoding=\"async\" width=\"1024\" height=\"571\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-GNAR33295-07-1024x571.jpg\" alt=\"\" class=\"wp-image-21628\" title=\"-\" srcset=\"https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-GNAR33295-07-1024x571.jpg 1024w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-GNAR33295-07-300x167.jpg 300w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-GNAR33295-07-768x428.jpg 768w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-GNAR33295-07-1536x857.jpg 1536w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-GNAR33295-07.jpg 1800w\" sizes=\"auto, (max-width: 1024px) 100vw, 1024px\" \/><\/figure>\n\n\n\n<p class=\"wp-block-paragraph\">C. 4-Chlorophenol asymmetry plot<\/p>\n\n\n\n<figure class=\"wp-block-image size-large\"><img loading=\"lazy\" decoding=\"async\" width=\"1024\" height=\"571\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-GNAR33295-08-1024x571.jpg\" alt=\"B. 1-Undecanol asymmetry plot\" class=\"wp-image-21634\" title=\"-\" srcset=\"https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-GNAR33295-08-1024x571.jpg 1024w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-GNAR33295-08-300x167.jpg 300w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-GNAR33295-08-768x428.jpg 768w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-GNAR33295-08-1536x857.jpg 1536w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-GNAR33295-08.jpg 1800w\" sizes=\"auto, (max-width: 1024px) 100vw, 1024px\" \/><\/figure>\n\n\n\n<p class=\"wp-block-paragraph\">D. 1-Decylamine asymmetry plot<\/p>\n\n\n\n<figure class=\"wp-block-image size-large\"><img loading=\"lazy\" decoding=\"async\" width=\"1024\" height=\"570\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-GNAR33295-09-1024x570.jpg\" alt=\"C. 4-Chlorophenol asymmetry plot\" class=\"wp-image-21640\" title=\"-\" srcset=\"https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-GNAR33295-09-1024x570.jpg 1024w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-GNAR33295-09-300x167.jpg 300w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-GNAR33295-09-768x428.jpg 768w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-GNAR33295-09-1536x855.jpg 1536w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-GNAR33295-09.jpg 1800w\" sizes=\"auto, (max-width: 1024px) 100vw, 1024px\" \/><\/figure>\n\n<\/div><\/div><\/div>\n\n\n<div style=\"height:20px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n\n\n\n<p class=\"wp-block-paragraph\">Rxi columns deliver more accurate, reliable trace-level results than any other fused silica column on the market. Take variation out of the equation by switching to highly reproducible Rxi columns.<\/p>\n\n\n\n<div style=\"height:100px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n","protected":false},"excerpt":{"rendered":"<p>Rxi columns demonstrate outstanding column-to-column reproducibility.<\/p>\n","protected":false},"author":11,"featured_media":0,"comment_status":"closed","ping_status":"open","sticky":false,"template":"","format":"standard","meta":{"_acf_changed":false,"_kad_blocks_custom_css":"","_kad_blocks_head_custom_js":"","_kad_blocks_body_custom_js":"","_kad_blocks_footer_custom_js":"","_kadence_starter_templates_imported_post":false,"_kad_post_transparent":"","_kad_post_title":"","_kad_post_layout":"","_kad_post_sidebar_id":"","_kad_post_content_style":"","_kad_post_vertical_padding":"","_kad_post_feature":"","_kad_post_feature_position":"","_kad_post_header":false,"_kad_post_footer":false,"footnotes":"","_ppma_block_editor_authors":""},"categories":[7],"tags":[],"industries-application":[],"post-badge":[],"resource-type":[],"product-library":[2372,2384],"resource-technique":[2294,2301],"ppma_author":[414],"class_list":["post-43686","post","type-post","status-publish","format-standard","hentry","category-articles","product-library-gas-chromatography-products","product-library-gc-columns","resource-technique-gas-chromatography-gc","resource-technique-mass-spectrometry-ms"],"acf":[],"taxonomy_info":{"category":[{"value":7,"label":"Articles"}],"product-library":[{"value":2372,"label":"Gas Chromatography Products"},{"value":2384,"label":"GC Columns"}],"resource-technique":[{"value":2294,"label":"Gas Chromatography (GC)"},{"value":2301,"label":"Mass Spectrometry (MS)"}]},"featured_image_src_large":false,"author_info":{"display_name":"Restek Corporation","author_link":"https:\/\/discover.restek.com\/de\/author\/restek-corporation\/"},"comment_info":0,"category_info":[{"term_id":7,"name":"Articles","slug":"articles","term_group":0,"term_taxonomy_id":7,"taxonomy":"category","description":"","parent":0,"count":510,"filter":"raw","cat_ID":7,"category_count":510,"category_description":"","cat_name":"Articles","category_nicename":"articles","category_parent":0}],"tag_info":false,"authors":[{"term_id":414,"user_id":11,"is_guest":0,"slug":"restek-corporation","display_name":"Restek Corporation","avatar_url":{"url":"https:\/\/discover.restek.com\/wp-content\/uploads\/Restek_Favicon_300x300.jpg","url2x":"https:\/\/discover.restek.com\/wp-content\/uploads\/Restek_Favicon_300x300.jpg"},"author_category":"1","first_name":"Restek","last_name":"Corporation","user_url":"https:\/\/www.linkedin.com\/company\/restek-corporation\/posts\/?feedView=all","job_title":"Chromatography Products and Solutions","description":"Restek is a leading provider of chromatography columns, accessories, and certified reference materials. Trust Restek for reliable, high-quality analytical solutions."}],"_links":{"self":[{"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/posts\/43686","targetHints":{"allow":["GET"]}}],"collection":[{"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/posts"}],"about":[{"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/types\/post"}],"author":[{"embeddable":true,"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/users\/11"}],"replies":[{"embeddable":true,"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/comments?post=43686"}],"version-history":[{"count":4,"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/posts\/43686\/revisions"}],"predecessor-version":[{"id":79394,"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/posts\/43686\/revisions\/79394"}],"wp:attachment":[{"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/media?parent=43686"}],"wp:term":[{"taxonomy":"category","embeddable":true,"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/categories?post=43686"},{"taxonomy":"post_tag","embeddable":true,"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/tags?post=43686"},{"taxonomy":"industries-application","embeddable":true,"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/industries-application?post=43686"},{"taxonomy":"post-badge","embeddable":true,"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/post-badge?post=43686"},{"taxonomy":"resource-type","embeddable":true,"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/resource-type?post=43686"},{"taxonomy":"product-library","embeddable":true,"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/product-library?post=43686"},{"taxonomy":"resource-technique","embeddable":true,"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/resource-technique?post=43686"},{"taxonomy":"author","embeddable":true,"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/ppma_author?post=43686"}],"curies":[{"name":"wp","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}