{"id":43727,"date":"2025-10-03T21:35:58","date_gmt":"2025-10-03T21:35:58","guid":{"rendered":"https:\/\/discover.restek.com\/uncategorized\/fast-analysis-of-alkylfurans-by-lpgc-ms\/"},"modified":"2026-08-11T15:17:01","modified_gmt":"2026-08-11T15:17:01","slug":"fast-analysis-of-alkylfurans-by-lpgc-ms","status":"publish","type":"post","link":"https:\/\/discover.restek.com\/de\/articles\/gnfa4238\/fast-analysis-of-alkylfurans-by-lpgc-ms","title":{"rendered":"Fast Analysis of Alkylfurans by LPGC-MS"},"content":{"rendered":"\n<ul class=\"wp-block-list\">\n<li>2.3x faster than conventional GC-MS analysis.<\/li>\n\n\n\n<li>Reduces helium usage by 72% compared to conventional GC-MS methods.&nbsp;<\/li>\n\n\n\n<li>Rxi-624Sil MS kits are factory connected with a proven leak-free connector making LPGC as simple as a column change.<\/li>\n<\/ul>\n\n\n\n<p class=\"wp-block-paragraph\">A known toxin in food products generated in heat or sterilization, alkylfurans are an important class of compounds in food safety testing. These potential carcinogens are a global health concern and are commonly analyzed by static headspace or solid phase microextraction coupled with GC-MS. Laboratories analyzing for alkylfurans in food products often work with a high volume of diverse sample types that can make conventional GC-MS analysis costly and time-consuming. Additionally, there are several critical separations that can be troublesome and difficult to resolve.&nbsp;<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">We developed a novel method to significantly cut down on extended analysis times with our new Rxi-624Sil MS low-pressure GC (LPGC) column kit. This unique column setup lowers the pressure inside the column, allowing for drastic reductions in analysis time and helium usage. Analyzing eight common alkylfurans on the LPGC Rxi-624 Sil MS kit (cat.# 11804), analysis time was reduced from ~12 minutes to ~5 minutes compared to the conventional GC method (see Figure 1 below). Additionally, good resolution was achieved for the critical separations of 2-methylfuran and 3-methylfuran (Peak 2 and Peak 3) along with 2,3-dimethylfuran and 2,5-dimethylfuran (Peak 5 and Peak 6).<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">A simple column change can lead to significant gains for food safety labs analyzing a high volume of samples. Utilizing a LPGC column significantly reduces analysis time allowing for increased sample throughput while conserving helium. Each of our factory-coupled LPGC kits are individually tested, so you can have the assurance of a leak-free connection. To learn more about the LPGC technique, visit <a href=\"articles\/an-introduction-to-low-pressure-GC-MS-LPGC-MS&quot;\">www.restek.com\/lpgc<\/a><\/p>\n\n\n<div class=\"wp-block-custom-chromatogram-article-top\"><div class=\"chromatogram-article-placeholder\"><div class=\"figure-heading\"><strong>Figure 1:\u00a0<\/strong>Fast GC-MS Analysis of Alkylfurans on LPGC Rxi-624Sil MS<\/div><div class='chromatogram-article-inner-full'><div class=\"chromatogram-article-inner\">\n<div class=\"wp-block-custom-chromatogram-article\"><div class=\"wp-block-custom-chromatogram-article\"><div class=\"chromatogram-image regular-image\"><img decoding=\"async\" src=\"https:\/\/ez.restek.com\/images\/cgram\/gc_fs0596.png\" alt=\"Comparison of Conventional and LPGC-MS Analysis of Alkylfurans\" title=\"-\"><\/div><p class=\"article-id\" style=\"text-align:center\"> GC_FS0596<\/p><div class=\"chromatogram-peaks\"><h4>Peaks<\/h4><table class=\"peaks col-lg-6 col-12\">\n<thead><tr><th><\/th><th style=\"text-align: left;width: 75px\">Peaks<\/th><th style=\"text-align: center;width: 75px\">t<sub>R<\/sub> (30 m)<\/th><th style=\"text-align: center;width: 75px\">t<sub>R<\/sub> (LPGC)<\/th><th style=\"text-align: center;width: 75px\">Conc. (ppm)<\/th><th style=\"text-align: center;width: 75px\">Quan. Ion<\/th><th style=\"text-align: center;width: 75px\">Qual. Ion<\/th><\/tr><\/thead>\n<tbody><tr><td class=\"num\">1.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Furan\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/110-00-9\/Furan\" target=\"_blank\" rel=\"noopener\">Furan<\/a><\/td><td class=\"oth\">3.03<\/td><td class=\"oth\">0.71<\/td><td class=\"oth\">10<\/td><td class=\"oth\">68<\/td><td class=\"oth\">39<\/td><\/tr>\n<tr><td class=\"num\">2.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 2-Methylfuran\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/534-22-5\/2-Methylfuran\" target=\"_blank\" rel=\"noopener\">2-Methylfuran<\/a><\/td><td class=\"oth\">5.393<\/td><td class=\"oth\">1.276<\/td><td class=\"oth\">10<\/td><td class=\"oth\">82<\/td><td class=\"oth\">53<\/td><\/tr>\n<tr><td class=\"num\">3.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 3-Methylfuran\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/930-27-8\/3-Methylfuran\" target=\"_blank\" rel=\"noopener\">3-Methylfuran<\/a><\/td><td class=\"oth\">5.727<\/td><td class=\"oth\">1.374<\/td><td class=\"oth\">10<\/td><td class=\"oth\">82<\/td><td class=\"oth\">53<\/td><\/tr>\n<tr><td class=\"num\">4.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 2-Ethylfuran\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/3208-16-0\/2-Ethylfuran\" target=\"_blank\" rel=\"noopener\">2-Ethylfuran<\/a><\/td><td class=\"oth\">8.072<\/td><td class=\"oth\">2.172<\/td><td class=\"oth\">10<\/td><td class=\"oth\">81<\/td><td class=\"oth\">96<\/td><\/tr>\n<tr><td class=\"num\">5.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 2,5-Dimethylfuran\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/625-86-5\/2,5-Dimethylfuran\" target=\"_blank\" rel=\"noopener\">2,5-Dimethylfuran<\/a><\/td><td class=\"oth\">8.208<\/td><td class=\"oth\">2.23<\/td><td class=\"oth\">10<\/td><td class=\"oth\">95<\/td><td class=\"oth\">81<\/td><\/tr>\n<tr><td class=\"num\">6.<\/td><td class=\"cmpd\">2,3-Dimethylfuran<\/td><td class=\"oth\">8.515<\/td><td class=\"oth\">2.357<\/td><td class=\"oth\">10<\/td><td class=\"oth\">95<\/td><td class=\"oth\">81<\/td><\/tr>\n<tr><td class=\"num\">7.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 2-Butylfuran\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/4466-24-4\/2-Butylfuran\" target=\"_blank\" rel=\"noopener\">2-Butylfuran<\/a><\/td><td class=\"oth\">11.21<\/td><td class=\"oth\">4.199<\/td><td class=\"oth\">10<\/td><td class=\"oth\">81<\/td><td class=\"oth\">124<\/td><\/tr>\n<tr><td class=\"num\">8.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 2-Pentylfuran\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/3777-69-3\/2-Pentylfuran\" target=\"_blank\" rel=\"noopener\">2-Pentylfuran<\/a><\/td><td class=\"oth\">12.179<\/td><td class=\"oth\">4.896<\/td><td class=\"oth\">10<\/td><td class=\"oth\">81<\/td><td class=\"oth\">138<\/td><\/tr>\n<\/tbody><\/table><\/div><div class=\"chromatogram-conditions\"><h4>Conditions<\/h4><div class=\"conditions-container container-fluid\"><div class=\"row\"><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Column<\/th><td>See notes<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Standard\/Sample<\/th><td>Custom furans standard<\/td><\/tr><tr><td><\/td><\/td><\/tr>\n<tr><th class=\"sub conditions_header\" scope=\"row\">Diluent:<\/th><td>Methanol (PT)<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Conc.:<\/th><td>10 \u00b5g\/mL<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Injection<\/th><td><\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Inj. Vol.:<\/th><td>1 \u00b5L split (split ratio 100:1)<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Liner:<\/th><td>Topaz, Precision inlet liner, 4.0 mm x 6.3 x 78.5 (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/23305\" rel=\"noopener\">cat.# 23305<\/a>)<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Inj. Temp.:<\/th><td>250 \u00b0C<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Carrier Gas<\/th><td>He<\/td><\/tr><\/table><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Detector<\/th><td>MS<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Mode:<\/th><td>SIM<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Transfer Line Temp.:<\/th><td>280 \u00b0C<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Analyzer Type:<\/th><td>Quadrupole <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Source Temp.:<\/th><td>230 \u00b0C<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Quad Temp.:<\/th><td>150 \u00b0C<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Tune Type:<\/th><td>PFTBA <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Ionization Mode:<\/th><td>EI <\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Instrument<\/th><td>Agilent 7890B GC &amp; 5977A MSD<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Sample Preparation<\/th><td>The sample was put in a 2 mL short-cap, screw-thread vial (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/21143\" rel=\"noopener\">cat.# 21143<\/a>) and capped with a short-cap, screw-vial closure (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/24495\" rel=\"noopener\">cat.# 24495<\/a>).<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Notes<\/th><td><b>Conventional (30 m) Analysis: <\/b><br \/>Column: Rxi-624Sil MS, 30 m, 0.25 mm ID, 1.4 \u00b5m (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/13868\" rel=\"noopener\">cat.# 13868<\/a>)<br \/>Temp. program: 35 \u00b0C (hold 3 min) to 75 \u00b0C at 8 \u00b0C\/min to 280 \u00b0C at 25 \u00b0C\/min (hold 5 min)<br \/>Flow: 1.4 mL\/min<br \/><br \/><b>LPGC-MS Analysis:<\/b><br \/>Column: LPGC  Rxi-624Sil MS column kit, includes 10 m x 0.32 mm ID x 1.8 \u03bcm Rxi-624Sil MS analytical column and 5 m x 0.15 mm ID Rxi restrictor factory connected via SilTite connector (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/11804\" rel=\"noopener\">cat.# 11804<\/a>)<br \/>Temp. program: 35 \u00b0C (hold 1 min) to 75 \u00b0C at 16 \u00b0C\/min (hold 0 min) to 280 \u00b0C at 35 \u00b0C\/min (hold 5 min)<br \/>Flow: 0.9 mL\/min<br \/><br \/>The injections were performed on different instruments under different head pressures, resulting in different analyte responses.<br \/><br \/>A furan\/alkylfurans standard (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/33334\" rel=\"noopener\">cat.# 33334<\/a>) is now available.<\/td><\/tr><\/table><\/div><\/div><\/div><div class=\"chromatogram-related-skus\"><h4>Related SKUs<\/h4><ul><li>23305<\/li><li>13868<\/li><li>11804<\/li><li>33334<\/li><li>21143<\/li><li>24495<\/li><\/ul><\/div><div class=\"chromatogram-pdf-link\"><a href=\"https:\/\/ez.restek.com\/images\/cgram\/gc_fs0596.pdf\" target=\"_blank\" rel=\"noopener noreferrer\"><svg xmlns=\"http:\/\/www.w3.org\/2000\/svg\" width=\"18\" height=\"18\" viewBox=\"0 0 18 18\"><g data-name=\"Group 2996\"><path data-name=\"Rectangle 1246\" d=\"M0 0h18v18H0z\" style=\"fill: none;\"><\/path><\/g><g data-name=\"Group 2997\"><path data-name=\"Path 729\" d=\"M13.412 11.4v2.017H5.345V11.4H4v2.017a1.349 1.349 0 0 0 1.345 1.345h8.068a1.349 1.349 0 0 0 1.345-1.345V11.4zm-.672-2.694-.948-.948-1.741 1.735V4H8.706v5.493L6.965 7.758l-.948.948 3.361 3.361z\" transform=\"translate(-.437 -.414)\" style=\"fill: rgb(13, 123, 196);\"><\/path><\/g><\/svg>Download PDF<\/a><\/div><\/div><\/div>\n<\/div><\/div><\/div>\n\n\n<div style=\"height:30px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n\n\n<style>.kb-row-layout-id45254_2e5dc7-30 > .kt-row-column-wrap{align-content:start;}:where(.kb-row-layout-id45254_2e5dc7-30 > .kt-row-column-wrap) > .wp-block-kadence-column{justify-content:start;}.kb-row-layout-id45254_2e5dc7-30 > .kt-row-column-wrap{column-gap:var(--global-kb-gap-md, 2rem);row-gap:var(--global-kb-gap-md, 2rem);padding-top:var(--global-kb-spacing-sm, 1.5rem);padding-bottom:var(--global-kb-spacing-sm, 1.5rem);grid-template-columns:minmax(0, 2fr) minmax(0, 1fr);}.kb-row-layout-id45254_2e5dc7-30{border-top-left-radius:10px;border-top-right-radius:10px;border-bottom-right-radius:10px;border-bottom-left-radius:10px;overflow:clip;isolation:isolate;}.kb-row-layout-id45254_2e5dc7-30 > .kt-row-layout-overlay{border-top-left-radius:10px;border-top-right-radius:10px;border-bottom-right-radius:10px;border-bottom-left-radius:10px;}.kb-row-layout-id45254_2e5dc7-30{background-color:#fffaf1;}.kb-row-layout-id45254_2e5dc7-30 > .kt-row-layout-overlay{opacity:0.30;}@media all and (max-width: 1024px){.kb-row-layout-id45254_2e5dc7-30 > .kt-row-column-wrap{grid-template-columns:minmax(0, 2fr) minmax(0, 1fr);}}@media all and (max-width: 767px){.kb-row-layout-id45254_2e5dc7-30 > .kt-row-column-wrap{grid-template-columns:minmax(0, 1fr);}}<\/style><div class=\"kb-row-layout-wrap kb-row-layout-id45254_2e5dc7-30 alignnone kt-row-has-bg wp-block-kadence-rowlayout\"><div class=\"kt-row-column-wrap kt-has-2-columns kt-row-layout-left-golden kt-tab-layout-inherit kt-mobile-layout-row kt-row-valign-top\">\n<style>.kadence-column45254_f24283-93 > .kt-inside-inner-col{display:flex;}.kadence-column45254_f24283-93 > .kt-inside-inner-col{padding-top:var(--global-kb-spacing-xs, 1rem);padding-right:var(--global-kb-spacing-xs, 1rem);padding-bottom:var(--global-kb-spacing-xs, 1rem);padding-left:var(--global-kb-spacing-xs, 1rem);}.kadence-column45254_f24283-93 > .kt-inside-inner-col,.kadence-column45254_f24283-93 > .kt-inside-inner-col:before{border-top-left-radius:0px;border-top-right-radius:0px;border-bottom-right-radius:0px;border-bottom-left-radius:0px;}.kadence-column45254_f24283-93 > .kt-inside-inner-col{column-gap:var(--global-kb-gap-sm, 1rem);}.kadence-column45254_f24283-93 > .kt-inside-inner-col{flex-direction:column;justify-content:center;}.kadence-column45254_f24283-93 > .kt-inside-inner-col > .aligncenter{width:100%;}.kt-row-column-wrap > .kadence-column45254_f24283-93{align-self:center;}.kt-inner-column-height-full:not(.kt-has-1-columns) > .wp-block-kadence-column.kadence-column45254_f24283-93{align-self:auto;}.kt-inner-column-height-full:not(.kt-has-1-columns) > .wp-block-kadence-column.kadence-column45254_f24283-93 > .kt-inside-inner-col{flex-direction:column;justify-content:center;}.kadence-column45254_f24283-93 > .kt-inside-inner-col:before{opacity:0.3;}.kadence-column45254_f24283-93{position:relative;}@media all and (max-width: 1024px){.kt-row-column-wrap > .kadence-column45254_f24283-93{align-self:center;}}@media all and (max-width: 1024px){.kt-inner-column-height-full:not(.kt-has-1-columns) > .wp-block-kadence-column.kadence-column45254_f24283-93{align-self:auto;}}@media all and (max-width: 1024px){.kt-inner-column-height-full:not(.kt-has-1-columns) > .wp-block-kadence-column.kadence-column45254_f24283-93 > .kt-inside-inner-col{flex-direction:column;justify-content:center;}}@media all and (max-width: 1024px){.kadence-column45254_f24283-93 > .kt-inside-inner-col{flex-direction:column;justify-content:center;}}@media all and (max-width: 767px){.kt-row-column-wrap > .kadence-column45254_f24283-93{align-self:center;}.kt-inner-column-height-full:not(.kt-has-1-columns) > .wp-block-kadence-column.kadence-column45254_f24283-93{align-self:auto;}.kt-inner-column-height-full:not(.kt-has-1-columns) > .wp-block-kadence-column.kadence-column45254_f24283-93 > .kt-inside-inner-col{flex-direction:column;justify-content:center;}.kadence-column45254_f24283-93 > .kt-inside-inner-col{flex-direction:column;justify-content:center;}}<\/style>\n<div class=\"wp-block-kadence-column kadence-column45254_f24283-93\"><div class=\"kt-inside-inner-col\">\n<h3 class=\"wp-block-heading has-text-color has-link-color wp-elements-a63340580c5d302b34f73ac16aec943d\" style=\"color:#de772c\">Looking for furan\/alkylfurans standards?<\/h3>\n\n\n\n<p class=\"wp-block-paragraph\">A native furan\/alkylfurans standard (<a href=\"https:\/\/www.restek.com\/p\/33334\" target=\"_blank\" rel=\"noreferrer noopener\">cat.# 33334<\/a>) is now available. Two labeled furan\/alkylfurans standard mixes are also available (<a href=\"https:\/\/www.restek.com\/p\/33335\" data-type=\"link\" data-id=\"https:\/\/www.restek.com\/p\/33335\" target=\"_blank\" rel=\"noopener\">cat.# 33335<\/a> &amp; <a href=\"https:\/\/www.restek.com\/p\/33336\" data-type=\"link\" data-id=\"https:\/\/www.restek.com\/p\/33336\" target=\"_blank\" rel=\"noopener\">cat.# 33336<\/a>).<\/p>\n<\/div><\/div>\n\n\n<style>.kadence-column45254_df90e6-d5 > .kt-inside-inner-col{display:flex;}.kadence-column45254_df90e6-d5 > .kt-inside-inner-col,.kadence-column45254_df90e6-d5 > .kt-inside-inner-col:before{border-top-left-radius:0px;border-top-right-radius:0px;border-bottom-right-radius:0px;border-bottom-left-radius:0px;}.kadence-column45254_df90e6-d5 > .kt-inside-inner-col{column-gap:var(--global-kb-gap-sm, 1rem);}.kadence-column45254_df90e6-d5 > .kt-inside-inner-col{flex-direction:column;justify-content:center;}.kadence-column45254_df90e6-d5 > .kt-inside-inner-col > .aligncenter{width:100%;}.kt-row-column-wrap > .kadence-column45254_df90e6-d5{align-self:center;}.kt-inner-column-height-full:not(.kt-has-1-columns) > .wp-block-kadence-column.kadence-column45254_df90e6-d5{align-self:auto;}.kt-inner-column-height-full:not(.kt-has-1-columns) > .wp-block-kadence-column.kadence-column45254_df90e6-d5 > .kt-inside-inner-col{flex-direction:column;justify-content:center;}.kadence-column45254_df90e6-d5 > .kt-inside-inner-col:before{opacity:0.3;}.kadence-column45254_df90e6-d5{position:relative;}@media all and (max-width: 1024px){.kt-row-column-wrap > .kadence-column45254_df90e6-d5{align-self:center;}}@media all and (max-width: 1024px){.kt-inner-column-height-full:not(.kt-has-1-columns) > .wp-block-kadence-column.kadence-column45254_df90e6-d5{align-self:auto;}}@media all and (max-width: 1024px){.kt-inner-column-height-full:not(.kt-has-1-columns) > .wp-block-kadence-column.kadence-column45254_df90e6-d5 > .kt-inside-inner-col{flex-direction:column;justify-content:center;}}@media all and (max-width: 1024px){.kadence-column45254_df90e6-d5 > .kt-inside-inner-col{flex-direction:column;justify-content:center;}}@media all and (max-width: 767px){.kt-row-column-wrap > .kadence-column45254_df90e6-d5{align-self:center;}.kt-inner-column-height-full:not(.kt-has-1-columns) > .wp-block-kadence-column.kadence-column45254_df90e6-d5{align-self:auto;}.kt-inner-column-height-full:not(.kt-has-1-columns) > .wp-block-kadence-column.kadence-column45254_df90e6-d5 > .kt-inside-inner-col{flex-direction:column;justify-content:center;}.kadence-column45254_df90e6-d5 > .kt-inside-inner-col{flex-direction:column;justify-content:center;}}<\/style>\n<div class=\"wp-block-kadence-column kadence-column45254_df90e6-d5\"><div class=\"kt-inside-inner-col\"><style>.kb-image45254_51474a-14:not(.kb-image-is-ratio-size) .kb-img, .kb-image45254_51474a-14.kb-image-is-ratio-size{padding-top:var(--global-kb-spacing-xs, 1rem);padding-right:var(--global-kb-spacing-xs, 1rem);padding-bottom:var(--global-kb-spacing-xs, 1rem);padding-left:var(--global-kb-spacing-xs, 1rem);}.kb-image45254_51474a-14 .kb-image-has-overlay:after{opacity:0.3;}<\/style>\n<div class=\"wp-block-kadence-image kb-image45254_51474a-14\"><figure class=\"aligncenter size-medium\"><img loading=\"lazy\" decoding=\"async\" width=\"300\" height=\"92\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/teaser-product-reference-standards-furan-alkylfurans-300x92.png\" alt=\"\" class=\"kb-img wp-image-29782\" title=\"-\" srcset=\"https:\/\/discover.restek.com\/wp-content\/uploads\/teaser-product-reference-standards-furan-alkylfurans-300x92.png 300w, https:\/\/discover.restek.com\/wp-content\/uploads\/teaser-product-reference-standards-furan-alkylfurans.png 384w\" sizes=\"auto, (max-width: 300px) 100vw, 300px\" \/><\/figure><\/div>\n<\/div><\/div>\n\n<\/div><\/div>\n\n\n<div style=\"height:50px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n\n\n        <div class=\"cpb\">\n            <h3 class=\"cpb-heading\">Erw\u00e4hnte Produkte<\/h3>\n            <hr class=\"cpb-heading-underline\" \/>\n            <div class=\"cpb-list\">\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/de.restek.com\/p\/33334\">                                Art.-Nr. 33334                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">Furan\/Alkylfuran Standard, 100 \u00b5g\/ml, P&amp;T Methanol, 1 ml\/Ampulle<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/de.restek.com\/p\/33334\">Produkt anzeigen<\/a>\n                                                    <\/div>\n                    <\/div>\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/de.restek.com\/p\/11804\">                                Art.-Nr. 11804                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">LPGC Rxi-624Sil MS S\u00e4ulenkit, 0.32 mm ID<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/de.restek.com\/p\/11804\">Produkt anzeigen<\/a>\n                                                    <\/div>\n                    <\/div>\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/de.restek.com\/p\/23305\">                                Art.-Nr. 23305                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">Topaz, Precision Inlet Liner, 4.0 mm x 6.3 x 78.5, f\u00fcr Agilent GCs, mit Quarzwolle, Premium-Deaktivierung, 5er Pack<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/de.restek.com\/p\/23305\">Produkt anzeigen<\/a>\n                                                    <\/div>\n                    <\/div>\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/de.restek.com\/p\/23849\">                                Art.-Nr. 23849                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">GC Accelerator-Ofeneinsatz-Kit, f\u00fcr Agilent 5890, 6890, 7890 und 8890 GCs<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/de.restek.com\/p\/23849\">Produkt anzeigen<\/a>\n                                                    <\/div>\n                    <\/div>\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/de.restek.com\/p\/26552\">                                Art.-Nr. 26552                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">Spritze, Hamilton 1001 (1 mL\/N\/23\/2 Zoll\/5pt), High Dynamic (HD) Headspace, f\u00fcr CTC CombiPAL Autosampler<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/de.restek.com\/p\/26552\">Produkt anzeigen<\/a>\n                                                    <\/div>\n                    <\/div>\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/de.restek.com\/p\/28903-1\">                                Art.-Nr. 28903-1                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">PAL Smart SPME Arrow 1.10 mm: Carbon-WR\/PDMS, Phasendicke 120 \u00b5m, Phasenl\u00e4nge 20 mm, Farbcode: hellblau, 1 St.<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/de.restek.com\/p\/28903-1\">Produkt anzeigen<\/a>\n                                                    <\/div>\n                    <\/div>\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/de.restek.com\/p\/21143\">                                Art.-Nr. 21143                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">Short-Cap-Vial mit graduiertem Markierungsfeld, 9-425 Screw-Thread, 2.0 mL, 9 mm, 12 x 32 (nur Vial), braun, 1000er Pack<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/de.restek.com\/p\/21143\">Produkt anzeigen<\/a>\n                                                    <\/div>\n                    <\/div>\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/de.restek.com\/p\/24495\">                                Art.-Nr. 24495                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">Short Screw Kappen, Polypropylen, Screw-Thread, PTFE\/Silikon\/PTFE Septen, schwarz, vormontiert, 2.0 mL, 9 mm, 100er Pack<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/de.restek.com\/p\/24495\">Produkt anzeigen<\/a>\n                                                    <\/div>\n                    <\/div>\n                            <\/div>\n        <\/div>\n        \n\n\n<div style=\"height:100px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n","protected":false},"excerpt":{"rendered":"<p>This fast analysis of alkylfurans by low-pressure GC-MS is 2.3x faster than conventional GC-MS analysis and offers up to 72% helium savings.<\/p>\n","protected":false},"author":11,"featured_media":7261,"comment_status":"closed","ping_status":"open","sticky":false,"template":"","format":"standard","meta":{"_acf_changed":false,"_kad_blocks_custom_css":"","_kad_blocks_head_custom_js":"","_kad_blocks_body_custom_js":"","_kad_blocks_footer_custom_js":"","_kadence_starter_templates_imported_post":false,"_kad_post_transparent":"","_kad_post_title":"","_kad_post_layout":"","_kad_post_sidebar_id":"","_kad_post_content_style":"","_kad_post_vertical_padding":"","_kad_post_feature":"","_kad_post_feature_position":"","_kad_post_header":false,"_kad_post_footer":false,"footnotes":"","_ppma_block_editor_authors":""},"categories":[7],"tags":[],"industries-application":[2167,2169],"post-badge":[],"resource-type":[2623],"product-library":[2372,2384],"resource-technique":[2294,2300,2301],"ppma_author":[414],"class_list":["post-43727","post","type-post","status-publish","format-standard","has-post-thumbnail","hentry","category-articles","industries-application-food-beverage","industries-application-food-safety","resource-type-featured-application","product-library-gas-chromatography-products","product-library-gc-columns","resource-technique-gas-chromatography-gc","resource-technique-low-pressure-gc-lpgc","resource-technique-mass-spectrometry-ms"],"acf":[],"taxonomy_info":{"category":[{"value":7,"label":"Articles"}],"industries-application":[{"value":2167,"label":"Food &amp; Beverage"},{"value":2169,"label":"Food Safety"}],"resource-type":[{"value":2623,"label":"Featured Application"}],"product-library":[{"value":2372,"label":"Gas Chromatography Products"},{"value":2384,"label":"GC Columns"}],"resource-technique":[{"value":2294,"label":"Gas Chromatography (GC)"},{"value":2300,"label":"Low-Pressure GC (LPGC)"},{"value":2301,"label":"Mass Spectrometry (MS)"}]},"featured_image_src_large":["https:\/\/discover.restek.com\/wp-content\/uploads\/feature-GNFA4238-1024x568.jpg",1024,568,true],"author_info":{"display_name":"Restek Corporation","author_link":"https:\/\/discover.restek.com\/de\/author\/restek-corporation\/"},"comment_info":0,"category_info":[{"term_id":7,"name":"Articles","slug":"articles","term_group":0,"term_taxonomy_id":7,"taxonomy":"category","description":"","parent":0,"count":511,"filter":"raw","cat_ID":7,"category_count":511,"category_description":"","cat_name":"Articles","category_nicename":"articles","category_parent":0}],"tag_info":false,"authors":[{"term_id":414,"user_id":11,"is_guest":0,"slug":"restek-corporation","display_name":"Restek Corporation","avatar_url":{"url":"https:\/\/discover.restek.com\/wp-content\/uploads\/Restek_Favicon_300x300.jpg","url2x":"https:\/\/discover.restek.com\/wp-content\/uploads\/Restek_Favicon_300x300.jpg"},"author_category":"1","first_name":"Restek","last_name":"Corporation","user_url":"https:\/\/www.linkedin.com\/company\/restek-corporation\/posts\/?feedView=all","job_title":"Chromatography Products and Solutions","description":"Restek is a leading provider of chromatography columns, accessories, and certified reference materials. Trust Restek for reliable, high-quality analytical solutions."}],"_links":{"self":[{"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/posts\/43727","targetHints":{"allow":["GET"]}}],"collection":[{"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/posts"}],"about":[{"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/types\/post"}],"author":[{"embeddable":true,"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/users\/11"}],"replies":[{"embeddable":true,"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/comments?post=43727"}],"version-history":[{"count":9,"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/posts\/43727\/revisions"}],"predecessor-version":[{"id":98429,"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/posts\/43727\/revisions\/98429"}],"wp:featuredmedia":[{"embeddable":true,"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/media\/7261"}],"wp:attachment":[{"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/media?parent=43727"}],"wp:term":[{"taxonomy":"category","embeddable":true,"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/categories?post=43727"},{"taxonomy":"post_tag","embeddable":true,"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/tags?post=43727"},{"taxonomy":"industries-application","embeddable":true,"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/industries-application?post=43727"},{"taxonomy":"post-badge","embeddable":true,"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/post-badge?post=43727"},{"taxonomy":"resource-type","embeddable":true,"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/resource-type?post=43727"},{"taxonomy":"product-library","embeddable":true,"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/product-library?post=43727"},{"taxonomy":"resource-technique","embeddable":true,"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/resource-technique?post=43727"},{"taxonomy":"author","embeddable":true,"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/ppma_author?post=43727"}],"curies":[{"name":"wp","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}