{"id":45269,"date":"2021-11-29T00:00:00","date_gmt":"2021-11-29T00:00:00","guid":{"rendered":"https:\/\/discover.restek.com\/uncategorized\/effective-lc-troubleshooting-symptom-based-strategies-and-solutions\/"},"modified":"2026-09-25T18:04:43","modified_gmt":"2026-09-25T18:04:43","slug":"effective-lc-troubleshooting-symptom-based-strategies-and-solutions","status":"publish","type":"post","link":"https:\/\/discover.restek.com\/de\/articles\/gnar3485\/effective-lc-troubleshooting-symptom-based-strategies-and-solutions","title":{"rendered":"Effective LC Troubleshooting: Symptom-Based Strategies and Solutions"},"content":{"rendered":"\n<ul class=\"wp-block-list\">\n<li>Tips for reestablishing a smooth baseline.<\/li>\n\n\n\n<li>Guidance on diagnosing and solving common peak problems.<\/li>\n\n\n\n<li>Best practices for managing system back pressure.<\/li>\n<\/ul>\n\n\n\n<p class=\"wp-block-paragraph\">Every instrument and every method will at some point experience a deterioration in performance that requires troubleshooting. When this occurs, the first\u2014and most essential\u2014part of LC troubleshooting is knowing what \u201cnormal\u201d looks like. By carefully documenting instrument qualification tests, maintenance events, and system suitability results, you create an invaluable source of comparison that establishes normal ranges and can reveal where problems may lie. For example, rerunning a flow rate accuracy test and comparing it to previous values can help establish whether a flow rate issue is the root cause behind retention time shifts. In this article, we explore common problems with baseline variation, chromatographic peaks, and system pressure and offer practical tactics for resolving them. Use these general guidelines in combination with equipment manuals and your laboratory\u2019s standard operating procedures to get your instrument back into service quickly and effectively.<\/p>\n\n\n\n<h2 class=\"wp-block-heading\">Reestablishing a Smooth Baseline<\/h2>\n\n\n\n<p class=\"wp-block-paragraph\">Encountering an inconsistent baseline is not unusual in LC laboratories. Variation can be caused by a number of factors, and determining if there is a pattern can help you establish the cause (Figure 1). If you are seeing an erratic baseline, it is likely caused by a leak or an air bubble, so checking all fittings, confirming the degasser is working, and purging your system with fresh mobile phase should stabilize the baseline. If you are using a UV detector, a noisy baseline can also indicate that it is time to change the detector lamp or flow cell. However, if there is regularity to the changes in the baseline, it may be a pump or piston issue, and routine maintenance of those parts should resolve the problem. Finally, if the baseline is looking bad overall, a thorough system cleaning is recommended. Note that ambient temperature changes can also cause baseline fluctuations, so using a column oven or insulating the column and tubing may also restore performance.<\/p>\n\n\n<div class=\"wp-block-custom-chromatogram-article-top\"><div class=\"chromatogram-article-placeholder\"><div class=\"figure-heading\"><strong>Figure 1:<\/strong>\u00a0Abnormal baselines that are commonly encountered in LC troubleshooting.<\/div><div class='chromatogram-article-inner-full'><div class=\"chromatogram-article-inner\">\n<div class=\"wp-block-image\">\n<figure class=\"aligncenter\"><img decoding=\"async\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-GNAR3485-01.jpg\" alt=\"decorative\" style=\"object-fit:cover\" title=\"-\"><\/figure>\n<\/div>\n<\/div><\/div><\/div>\n\n\n<figure class=\"wp-block-embed is-type-video is-provider-youtube wp-block-embed-youtube wp-embed-aspect-16-9 wp-has-aspect-ratio\"><div class=\"wp-block-embed__wrapper\">\n<iframe loading=\"lazy\" title=\"LC Troubleshooting\u2014Baseline Problems\" width=\"720\" height=\"405\" src=\"https:\/\/www.youtube.com\/embed\/JrEs89ltQUE?feature=oembed\" frameborder=\"0\" allow=\"accelerometer; autoplay; clipboard-write; encrypted-media; gyroscope; picture-in-picture; web-share\" referrerpolicy=\"strict-origin-when-cross-origin\" allowfullscreen><\/iframe>\n<\/div><\/figure>\n\n\n\n<div style=\"height:50px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n\n\n\n<h2 class=\"wp-block-heading\">Troubleshooting Peak Issues<\/h2>\n\n\n\n<p class=\"wp-block-paragraph\">Troubleshooting peak problems is necessary during both routine analysis and method development, but the changes that are allowable under those two scenarios are different. Issues with poor peak shape, sensitivity, and retention that develop over time in an established method that is otherwise performing well indicate a change has occurred (e.g., system leak, mobile phase deterioration, guard contamination, etc.) and resolving it through maintenance should restore performance. However, some of the tips discussed below, such as adding buffer to mobile phases or changing the column temperature, should not be made during routine analysis because they may require additional validation before implementation.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">In contrast, peak issues seen during method development indicate further changes to method parameters may be needed, and more opportunities for improvement are available than when they are encountered with an established method. During method development, in addition to using the LC troubleshooting guidelines below, the sample preparation method, column stationary phase, and analytical conditions can be optimized for the matrices and compounds to be analyzed. When making changes to improve chromatographic peaks, it is important to first understand what factors can or cannot be changed based on your laboratory\u2019s standard operating procedures.<\/p>\n\n\n\n<h3 class=\"wp-block-heading\">Improving Poor Peak Shape<\/h3>\n\n\n\n<p class=\"wp-block-paragraph\">Sharp, symmetrical peaks improve accuracy and sensitivity, but many factors can cause peaks to tail, front, split, or broaden. These types of peak distortions often lead to other problems, such as shifting retention times or coelution. If you observe distorted peak shapes either during routine analysis or method development, the LC troubleshooting tips in Table I will help you diagnose the problem based on the symptoms you observe in the chromatogram.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\"><strong>Table I:<\/strong>&nbsp;LC troubleshooting tips for improving peak shape.<\/p>\n\n\n\n<figure class=\"wp-block-table\"><table class=\"has-fixed-layout\"><tbody><tr><td><strong>Symptom<\/strong><\/td><td><strong>Cause<\/strong><\/td><td><strong>Solution<\/strong><\/td><\/tr><tr><td rowspan=\"7\"><strong>Peak tailing<\/strong> <img decoding=\"async\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/chart-article-GNAR3485-01.jpg\" alt=\"Peak Tailing Example\" style=\"\" title=\"-\"><\/td><td>Column overloading<\/td><td>Inject less mass (dilute the sample or decrease the injection volume). Note that injection volume is limited by column size; general guidelines for acceptable ranges are as follows: <strong>Column ID<\/strong> <strong>Volume (\u00b5L)<\/strong> 2.1 mm (30 -100 mm length) 1-3 3.0-3.2 mm (50-150 mm length) 2-12 4.6 mm (50-250 mm length) 8-40<\/td><\/tr><tr><td>Worn or degraded column<\/td><td>While proper care can maximize LC column lifetime, columns are consumables that eventually will have to be replaced. Poor chromatographic performance can indicate it is time to replace the column. But, before replacing it, try regenerating it using the procedure outlined in Table II.<\/td><\/tr><tr><td>Contamination<\/td><td>Prepare fresh solutions to rule out incorrectly prepared or degraded solutions. Flush your analytical column following the manufacturer\u2019s recommended cleaning procedure. Change guard columns regularly, making sure to match the phase of your analytical column. If using MS analyzers, make sure you use LC-MS grade solvents and additives.<\/td><\/tr><tr><td>Interactions with silanol on silica surface<\/td><td>Add buffer to your mobile phase (aqueous and organic portions should be buffered equally) to block active sites on the silica surface. A buffer can be prepared by mixing a weak acid with the salt of its conjugate base. When using formic acid, use ammonium formate to buffer the solution and when using acetic acid, use ammonium acetate.<\/td><\/tr><tr><td>Interfering\/coeluting matrix components<\/td><td>Analyze a standard prepared in solvent for comparison. Filter sample extracts. Consider adding or modifying the sample preparation procedure to produce cleaner extracts.<\/td><\/tr><tr><td>Excessive system volume<\/td><td>Reduce system volume by using shorter tubing segments or smaller internal diameter tubing to minimize peak dispersion. When replacing the tubing, it is important to make proper connections by ensuring that the tubing and ferrule are both fully seated in the column port.<\/td><\/tr><tr><td>Poor connections<\/td><td>Ensure that the tubing and ferrule are both fully seated in the column port. <img loading=\"lazy\" decoding=\"async\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/chart-article-GNAR3485-01a.jpg\" alt=\"decorative\" width=\"1800\" height=\"760\" title=\"-\"><\/td><\/tr><tr><td rowspan=\"6\"><strong>Peak fronting<\/strong> <img loading=\"lazy\" decoding=\"async\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/chart-article-GNAR3485-02.jpg\" alt=\"decorative\" width=\"1000\" height=\"817\" title=\"-\"><\/td><td>Solvent incompatibility (particularly likely if early eluting peaks are fronting)<\/td><td>Dilute the sample in the same solvent composition (or weaker) as the initial mobile phase composition. Try to match both the aqueous:organic ratio and the buffer strength.<\/td><\/tr><tr><td>Worn or degraded column<\/td><td>While proper care can maximize LC column lifetime, columns are consumables that eventually will have to be replaced. Poor chromatographic performance can indicate it is time to replace the column. But, before replacing it, try regenerating it using the procedure outlined in Table II.<\/td><\/tr><tr><td>Contamination<\/td><td>Prepare fresh solutions to rule out incorrectly prepared or degraded solutions. Flush your analytical column following the manufacturer\u2019s recommended cleaning procedure. Change guard columns regularly, making sure to match the phase of your analytical column. If using MS analyzers, make sure you use LC-MS grade solvents and additives.<\/td><\/tr><tr><td>Column overloading<\/td><td>Inject less mass (dilute the sample or decrease the injection volume). Note that injection volume is limited by column size; general guidelines for acceptable ranges are as follows: <strong>Column ID<\/strong> <strong>Volume (\u00b5L)<\/strong> 2.1 mm (30 -100 mm length) 1-3 3.0-3.2 mm (50-150 mm length) 2-12 4.6 mm (50-250 mm length) 8-40<\/td><\/tr><tr><td>Interfering\/coeluting matrix components<\/td><td>Analyze a standard prepared in solvent for comparison. Filter sample extracts. Consider adding or modifying the sample preparation procedure to produce cleaner extracts.<\/td><\/tr><tr><td>Poor connections<\/td><td>Ensure that the tubing and ferrule are both fully seated in the column port. <img loading=\"lazy\" decoding=\"async\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/chart-article-GNAR3485-01a.jpg\" alt=\"decorative\" width=\"1800\" height=\"760\" title=\"-\"><\/td><\/tr><tr><td rowspan=\"3\"><strong>Peak splitting<\/strong> <img loading=\"lazy\" decoding=\"async\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/chart-article-GNAR3485-03.jpg\" alt=\"shape, arrow, rectangle\" width=\"1000\" height=\"967\" title=\"-\"><\/td><td>Solvent incompatibility<\/td><td>Dilute the sample in the same solvent composition (or weaker) as the initial mobile phase composition.&nbsp;<\/td><\/tr><tr><td>Solubility issue<\/td><td>Ensure that the sample is fully soluble in both the sample solvent and the mobile phase to prevent precipitation.<\/td><\/tr><tr><td>Contamination<\/td><td>Prepare fresh solutions to rule out incorrectly prepared or degraded solutions. Flush your analytical column following the manufacturer\u2019s recommended cleaning procedure. Change guard columns regularly, making sure to match the phase of your analytical column. If using MS analyzers, make sure you use LC-MS grade solvents and additives.<\/td><\/tr><tr><td rowspan=\"12\"><strong>Broad Peaks<\/strong> <img loading=\"lazy\" decoding=\"async\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/chart-article-GNAR3485-04.jpg\" alt=\"decorative\" width=\"1000\" height=\"1700\" title=\"-\"><\/td><td>Column overloading<\/td><td>Inject less mass (dilute the sample or decrease the injection volume). Note that injection volume is limited by column size; general guidelines for acceptable ranges are as follows: <strong>Column ID<\/strong> <strong>Volume (\u00b5L)<\/strong> 2.1 mm (30 -100 mm length) 1-3 3.0-3.2 mm (50-150 mm length) 2-12 4.6 mm (50-250 mm length) 8-40<\/td><\/tr><tr><td>Change in mobile phase concentration<\/td><td>Make fresh mobile phase and keep capped to prevent evaporation or contamination.<\/td><\/tr><tr><td>Flow rate too low<\/td><td>Increase mobile phase flow rate.<\/td><\/tr><tr><td>Peak dispersion in the injector<\/td><td>Use a smaller sample loop.<\/td><\/tr><tr><td>Too much retention<\/td><td>Use less retentive stationary phase. Try gradient elution. If running an isocratic method, increase mobile phase strength. Adjust buffer composition.<\/td><\/tr><tr><td>Detector cell volume too large or response time too long<\/td><td>Use smaller cell and\/or decrease response time.<\/td><\/tr><tr><td>Extra column volume is too large<\/td><td>Use shorter tubing with a smaller internal diameter. Use zero-dead-volume end fittings.<\/td><\/tr><tr><td>Low column temperature<\/td><td>Raise temperature.<\/td><\/tr><tr><td>End of guard or analytical column lifetime<\/td><td>Replace guard and\/or column.<\/td><\/tr><tr><td>Insufficient column efficiency<\/td><td>Try a smaller particle column.<\/td><\/tr><tr><td>Coelution<\/td><td>Loss of resolution can be cause by degraded mobile phase or contaminated guard columns. Prepare fresh mobile phase and replace the guard column. If this does not separate the coeluting peak, restore or replace the analytical column. Lack of initial resolution during method development indicates column selectivity or chromatographic parameters are not suitable. Adjust mobile phase, column temperature, or try column with a different stationary phase selectivity.<\/td><\/tr><tr><td>Leak<\/td><td>Check for loose fittings and tighten or replace. If leaks or precipitates are found on pump, may need to replace pump seals.<\/td><\/tr><\/tbody><\/table><\/figure>\n\n\n\n<h3 class=\"wp-block-heading\">Dealing With a Decrease in Sensitivity<\/h3>\n\n\n<style>.kb-image45269_0739fd-ed .kb-image-has-overlay:after{opacity:0.3;}<\/style>\n<div class=\"wp-block-kadence-image kb-image45269_0739fd-ed\"><figure class=\"aligncenter size-medium_large\"><img loading=\"lazy\" decoding=\"async\" width=\"768\" height=\"915\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-GNAR3485-02-768x915.jpg\" alt=\"\" class=\"kb-img wp-image-21316\" title=\"-\" srcset=\"https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-GNAR3485-02-768x915.jpg 768w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-GNAR3485-02-252x300.jpg 252w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-GNAR3485-02-859x1024.jpg 859w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-GNAR3485-02.jpg 1000w\" sizes=\"auto, (max-width: 768px) 100vw, 768px\" \/><\/figure><\/div>\n\n\n\n<p class=\"wp-block-paragraph\">A loss in sensitivity can indicate problems with either the sample preparation or the analytical system. First, confirm that the sample preparation procedure was followed correctly, including storage temperature and time, and verify that system parameters are set correctly and functioning properly. Always check for obvious problems (e.g., calculation\/dilution error, autosampler needle not reaching sample, incorrect injection volume, wrong detector settings, no mobile phase flow, detector or lamp turned off, integrity of the reference standard, etc.) before looking for more complex issues. Having a coworker double check your work can be invaluable because it is often easier for an independent person to spot problems than it is for someone who is already familiar with the work.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Once simple problems have been ruled out, analyzing a known standard is a helpful diagnostic tool for LC troubleshooting. If results are within the expected range, the system is performing well, and the problem is in the sample preparation or handling; however, if low response is also seen for the standard, the problem is likely with the instrument. It is also possible that analytes are simply outside of the instrument\u2019s sensitivity range. In this case, sample concentration may bring them within range, but care must be taken to prevent the loss of more volatile analytes.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">If poor response is seen only in the first few injections, it is possible that the sample is adsorbing to active sites in the sample loop or column. Using a passivation solution or making a few preliminary sample injections to condition the system and reduce adsorption prior to actual sample analysis may help restore analyte response. If response has decreased across all peaks, this is likely caused by calculation errors, incorrect dilutions, or a decrease in injection volume that may result from leaks, system malfunctions, programming errors, or use of the wrong size sample loop. If a catastrophic loss of retention is observed, it is possible that phase dewetting has occurred and the column must be regenerated or replaced, but this is a relatively rare event compared to other causes.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">For LC-MS analysis, there could be additional causes responsible for a decrease in sensitivity. Issues with the MS analyzer can be a common source, so verifying that the MS performance is at its best is essential. Directly infusing the tuning solution into the MS (bypassing the LC system) is an easy step that provides useful information for troubleshooting. Since there are several possible causes of loss in MS sensitivity (e.g. cleanliness of the optics, buffer precipitate in the MS entrance, need to change an electron multiplier, ionization source performance, compounds tuning, etc.) we recommend checking with your manufacturer for specific troubleshooting information. Ion suppression due to matrix interferences can also cause poor response in LC-MS systems, but this typically must be addressed with major method changes that may require revalidation, such as better sample preparation procedures, use of internal or matrix-matched standards, and changes to the mobile phase composition.<\/p>\n\n\n\n<h3 class=\"wp-block-heading\">Stabilizing Shifting Retention Times<\/h3>\n\n\n\n<figure class=\"wp-block-image\"><img decoding=\"async\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-GNAR3485-03.jpg\" alt=\"decorative\" title=\"-\"><\/figure>\n\n\n\n<p class=\"wp-block-paragraph\">For accurate identification and quantification, it is critical that analytes always elute at the same retention time. However, it is not uncommon to observe retention time shifts during routine analysis, and there are a number of possible causes. As a first step in LC troubleshooting, verify that all method settings are correct and that the mobile phase was properly prepared (particularly if you see a big change instead of a slight drift). Preparing fresh mobile phase is an easy way to eliminate problems caused by composition changes due to evaporation, reactions, contamination, or incomplete mixing. Also, if you are transferring an existing method to a new instrument, be sure to account for any difference in dwell volume between the instruments because that can significantly affect retention time. Other possible causes include the following:<\/p>\n\n\n\n<ol class=\"wp-block-list\">\n<li>Incorrect or variable flow rate\u2014manually confirm that the flow rate is correct and verify that there are no leaks in the system.<\/li>\n\n\n\n<li>Temperature fluctuations\u2014always use a thermostatically controlled column compartment.<\/li>\n\n\n\n<li>Solvent mismatch\u2014make sure that the sample solvent is compatible with the starting mobile phase (try to match both the aqueous:organic ratio and the buffer strength).<\/li>\n\n\n\n<li>Sample overloading\u2014don\u2019t inject more sample than the column can hold.<\/li>\n\n\n\n<li>Column contamination\/clogging\u2014flush the column to remove any particulates, particularly if back pressure is also elevated. If this does not work, it may be time to replace the column. Using a guard column is recommended to minimize contamination and maximize column lifetime.<\/li>\n\n\n\n<li>Matrix interferences\u2014analyze a solvent standard and compare it to your sample to see if the shifting peaks are actually matrix components.<\/li>\n\n\n\n<li>Insufficient column equilibration\u2014allow more time between runs, sufficient time for passing 10 column volumes is recommended. Note: Column equilibration may take longer for HILIC analyses.<\/li>\n\n\n\n<li>Air bubbles\u2014purge air from pump and ensure the mobile phase is degassed. Inspect pump seals and change if leaks are evident.<\/li>\n<\/ol>\n\n\n\n<p class=\"wp-block-paragraph\">Occasionally, ghost peaks will appear at unexpected times during analysis. This can be tested by making a solvent injection after a sample injection. If a peak appears, it is likely a late-eluting peak carried over from the previous sample injection. In this situation, increasing the flow rate, extending the run time, or adding a gradient step may help eliminate the carryover problem. The risk of carryover can also be reduced by minimizing extra system volume by plumbing the LC with shorter, smaller diameter tubing and zero-dead-volume connections because this will reduce the places that sample can build up. In addition, needle wash solvents and needle wash parameters are important to consider in order to reduce carryover. Ghost peaks can also be caused by contamination in the mobile phase, guard column, or analytical column. This can be resolved by preparing fresh, degassed mobile phase; replacing the guard column; and replacing your analytical column (but first try flushing it according to the manufacturer\u2019s recommended cleaning procedures).<\/p>\n\n\n\n<h2 class=\"wp-block-heading\">Restoring Normal System Pressure<\/h2>\n\n\n\n<h3 class=\"wp-block-heading\">Diagnosing the Cause of Increased Back Pressure<\/h3>\n\n\n\n<p class=\"wp-block-paragraph\">Increased back pressure is one of the most common of all LC troubleshooting issues, but it is important to remember that not all pressure changes indicate a problem. For example, pressure will change during gradient analysis as mobile phase viscosity increases (maximum viscosities are at ~50:50 for methanol:water and ~20:80 for acetonitrile:water). So, it is important to know what is normal for your setup in order to diagnose a problem. It is also important to work within the pressure limits of your LC (e.g., do not use &lt;2.7 \u00b5m columns in instruments with a 400 bar limit). When abnormally high back pressure is observed, it indicates a blockage exists somewhere in the system, and the first step in resolving it is to identify the location by isolating one potential source at a time following the steps in Figure 2.<\/p>\n\n\n<div class=\"wp-block-custom-chromatogram-article-top\"><div class=\"chromatogram-article-placeholder\"><div class=\"figure-heading\"><strong>Figure 2:<\/strong>\u00a0LC troubleshooting steps for locating system blockages that cause increased back pressure.<\/div><div class='chromatogram-article-inner-full'><div class=\"chromatogram-article-inner\">\n<style>.kb-image45269_f817a5-b4 .kb-image-has-overlay:after{opacity:0.3;}<\/style>\n<div class=\"wp-block-kadence-image kb-image45269_f817a5-b4\"><figure class=\"aligncenter size-large\"><img loading=\"lazy\" decoding=\"async\" width=\"1024\" height=\"437\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-GNAR3485-04-1024x437.jpg\" alt=\"\" class=\"kb-img wp-image-21328\" title=\"-\" srcset=\"https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-GNAR3485-04-1024x437.jpg 1024w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-GNAR3485-04-300x128.jpg 300w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-GNAR3485-04-768x328.jpg 768w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-GNAR3485-04-1536x655.jpg 1536w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-GNAR3485-04.jpg 1800w\" sizes=\"auto, (max-width: 1024px) 100vw, 1024px\" \/><\/figure><\/div>\n\n<\/div><\/div><\/div>\n\n\n<div style=\"height:20px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n\n\n\n<h3 class=\"wp-block-heading\">Removing Column Clogs and Contaminants<\/h3>\n\n\n\n<p class=\"wp-block-paragraph\">Increased back pressure and decreased chromatographic performance can indicate column contamination or blockage. Flushing the column and preventing future blockage using the following guidelines can solve the problem and extend the life of your analytical column.<\/p>\n\n\n\n<h4 class=\"wp-block-heading\">Treatment<\/h4>\n\n\n\n<p class=\"wp-block-paragraph\">Flushing HPLC and UHPLC columns in the direction shown on the column with a series of solvents can be an effective way to remove both chemical contaminants and particulates that have formed clogs. Use a minimum of 20 column volumes of each solvent and flush in the order shown in Table II below. Following regeneration, always store your column in the solvent recommended by the column manufacturer.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\"><strong>Table II:<\/strong>&nbsp;Solvent Sequence for Column Flushing<\/p>\n\n\n\n<figure class=\"wp-block-table\"><table class=\"has-fixed-layout\"><tbody><tr><td><strong>Reversed-Phase Columns<\/strong><\/td><td><strong>Normal Phase Columns<\/strong><\/td><\/tr><tr><td>1.&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp; Water:methanol (95:5 v\/v)<\/td><td>1. Isopropyl alcohol<\/td><\/tr><tr><td>2.&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp; Methanol<\/td><td>2.&nbsp;<em>n<\/em>-Hexane<\/td><\/tr><tr><td>3.&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp; Isopropyl alcohol<\/td><td>3. Ethanol<\/td><\/tr><tr><td>4.&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;<em>n<\/em>-Hexane<\/td><td>4. Original mobile phase<\/td><\/tr><tr><td>5.&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp; Isopropyl alcohol<\/td><td>&nbsp;<\/td><\/tr><tr><td>6.&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp; Methanol<\/td><td>&nbsp;<\/td><\/tr><tr><td>7.&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp; Water:methanol (95:5 v\/v)<\/td><td>&nbsp;<\/td><\/tr><tr><td>8.&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp; Original mobile phase<\/td><td>&nbsp;<\/td><\/tr><\/tbody><\/table><\/figure>\n\n\n\n<h4 class=\"wp-block-heading\">Prevention<\/h4>\n\n\n\n<ul class=\"wp-block-list\">\n<li>Always use fresh, filtered mobile phases to remove particulates and prevent bacterial growth. Keep your mobile phase bottles capped to avoid contamination. If using MS analyzers, make sure you use LC-MS grade solvents and additives.<\/li>\n\n\n\n<li>Filter samples to remove matrix particulates; <a href=\"https:\/\/www.restek.com\/c\/1338\" target=\"_blank\" rel=\"noopener\">syringe filters<\/a> and <a href=\"https:\/\/www.restek.com\/c\/1337\" target=\"_blank\" rel=\"noopener\">filter vials<\/a> simplify this task.<\/li>\n\n\n\n<li>Confirm the sample solvent and mobile phases are compatible. Solvent mismatch can cause sample components or buffer salts to precipitate and form clogs.<\/li>\n\n\n\n<li>Use a guard column or precolumn filter to protect your analytical column. Always select a guard system and phase that matches the analytical column.<\/li>\n\n\n\n<li>Regularly inspect and replace pump seals, auto-injector rotors, inline filters, and needle seats before normal wear and tear causes them to break down and shed particles into the flow stream.<\/li>\n<\/ul>\n\n\n\n<p class=\"wp-block-paragraph\">Although less common, it is also possible to encounter pressures that are too low. In this scenario, leaks are a usually the culprit, so check your fittings and pump carefully for signs of moisture or precipitates. If tightening or replacing the fittings does not solve the problem, it may be time to replace the pump seals. Trapped air bubbles or a stuck check valve can also cause pressure drops, but both issues can generally be fixed by flushing the system with degassed mobile phase or an appropriate solvent, such as isopropyl alcohol. If this does not resolve this problem, it may be time to replace the check valves. Some leaks could be due to scratches in the rotor seal in six-port valves. The leak may not be visible, but it will cause a drop in the pressure and flow. When looking for leaks or measuring flow rate accuracy, be sure to utilize an old column or a flow restrictor. In addition to pressure drops, leaks can cause other chromatographic problems, such as increased retention, decreased signal, or even automatic instrument shut down. To prevent this, make it a daily practice to inspect fittings for residue, and wipe tubing with a lab tissue to detect moisture. Gently retightening fittings and regularly replacing tubing, fittings, and seals as part of your routine maintenance plan is a good way to prevent catastrophic leaks and unexpected downtime. <a href=\"articles\/routine-LC-maintenance\/\">Restek\u2019s Routine LC Maintenance guide<\/a> is a helpful starting place when developing or refining a maintenance plan.<\/p>\n\n\n\n<h2 class=\"wp-block-heading\">Summary<\/h2>\n\n\n\n<p class=\"wp-block-paragraph\">LC troubleshooting is an essential part of generating accurate and reliable data. While there are many potential causes of poor performance, careful observation of the chromatographic symptoms, testing of one possible cause at a time, and comparison to an established \u201cnormal\u201d state will help you quickly and effectively identify the root cause. Finally, working any lessons learned into a routine maintenance plan will help prevent future problems and keep your instrument running more efficiently.<\/p>\n\n\n\n<figure class=\"wp-block-embed is-type-video is-provider-youtube wp-block-embed-youtube wp-embed-aspect-16-9 wp-has-aspect-ratio\"><div class=\"wp-block-embed__wrapper\">\n<iframe loading=\"lazy\" title=\"LC Leak Checking\" width=\"720\" height=\"405\" src=\"https:\/\/www.youtube.com\/embed\/eJZDdd6SkE8?feature=oembed\" frameborder=\"0\" allow=\"accelerometer; autoplay; clipboard-write; encrypted-media; gyroscope; picture-in-picture; web-share\" referrerpolicy=\"strict-origin-when-cross-origin\" allowfullscreen><\/iframe>\n<\/div><\/figure>\n\n\n\n<div style=\"height:100px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n\n\n\n<p class=\"wp-block-paragraph\"><\/p>\n","protected":false},"excerpt":{"rendered":"<p>This article about LC troubleshooting reviews common causes of performance issues, the symptoms associated with them, and effective ways to resolve them.<\/p>\n","protected":false},"author":11,"featured_media":26253,"comment_status":"closed","ping_status":"open","sticky":false,"template":"","format":"standard","meta":{"_acf_changed":false,"_kad_blocks_custom_css":"","_kad_blocks_head_custom_js":"","_kad_blocks_body_custom_js":"","_kad_blocks_footer_custom_js":"","_kadence_starter_templates_imported_post":false,"_kad_post_transparent":"","_kad_post_title":"","_kad_post_layout":"","_kad_post_sidebar_id":"","_kad_post_content_style":"","_kad_post_vertical_padding":"","_kad_post_feature":"","_kad_post_feature_position":"","_kad_post_header":false,"_kad_post_footer":false,"footnotes":"","_ppma_block_editor_authors":""},"categories":[7,786],"tags":[],"industries-application":[],"post-badge":[],"resource-type":[],"product-library":[2393,2395,2397,2396,2392,2373],"resource-technique":[2299],"ppma_author":[414],"class_list":["post-45269","post","type-post","status-publish","format-standard","has-post-thumbnail","hentry","category-articles","category-articles-zh","product-library-lc-autosampler-supplies","product-library-lc-detector-supplies","product-library-lc-pump-supplies","product-library-lc-tubing-fittings","product-library-lc-ms-supplies","product-library-liquid-chromatography-products","resource-technique-liquid-chromatography"],"acf":[],"taxonomy_info":{"category":[{"value":7,"label":"Articles"},{"value":786,"label":"\u6587\u7ae0"}],"product-library":[{"value":2393,"label":"LC Autosampler Supplies"},{"value":2395,"label":"LC Detector Supplies"},{"value":2397,"label":"LC Pump Supplies"},{"value":2396,"label":"LC Tubing &amp; Fittings"},{"value":2392,"label":"LC-MS Supplies"},{"value":2373,"label":"Liquid Chromatography Products"}],"resource-technique":[{"value":2299,"label":"Liquid Chromatography"}]},"featured_image_src_large":["https:\/\/discover.restek.com\/wp-content\/uploads\/feature-GNAR3485-1024x536.jpg",1024,536,true],"author_info":{"display_name":"Restek Corporation","author_link":"https:\/\/discover.restek.com\/de\/author\/restek-corporation\/"},"comment_info":0,"category_info":[{"term_id":7,"name":"Articles","slug":"articles","term_group":0,"term_taxonomy_id":7,"taxonomy":"category","description":"","parent":0,"count":511,"filter":"raw","cat_ID":7,"category_count":511,"category_description":"","cat_name":"Articles","category_nicename":"articles","category_parent":0},{"term_id":786,"name":"\u6587\u7ae0","slug":"articles-zh","term_group":0,"term_taxonomy_id":786,"taxonomy":"category","description":"","parent":0,"count":511,"filter":"raw","cat_ID":786,"category_count":511,"category_description":"","cat_name":"\u6587\u7ae0","category_nicename":"articles-zh","category_parent":0}],"tag_info":false,"authors":[{"term_id":414,"user_id":11,"is_guest":0,"slug":"restek-corporation","display_name":"Restek Corporation","avatar_url":{"url":"https:\/\/discover.restek.com\/wp-content\/uploads\/Restek_Favicon_300x300.jpg","url2x":"https:\/\/discover.restek.com\/wp-content\/uploads\/Restek_Favicon_300x300.jpg"},"author_category":"1","first_name":"Restek","last_name":"Corporation","user_url":"https:\/\/www.linkedin.com\/company\/restek-corporation\/posts\/?feedView=all","job_title":"Chromatography Products and Solutions","description":"Restek is a leading provider of chromatography columns, accessories, and certified reference materials. Trust Restek for reliable, high-quality analytical solutions."}],"_links":{"self":[{"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/posts\/45269","targetHints":{"allow":["GET"]}}],"collection":[{"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/posts"}],"about":[{"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/types\/post"}],"author":[{"embeddable":true,"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/users\/11"}],"replies":[{"embeddable":true,"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/comments?post=45269"}],"version-history":[{"count":13,"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/posts\/45269\/revisions"}],"predecessor-version":[{"id":103116,"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/posts\/45269\/revisions\/103116"}],"wp:featuredmedia":[{"embeddable":true,"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/media\/26253"}],"wp:attachment":[{"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/media?parent=45269"}],"wp:term":[{"taxonomy":"category","embeddable":true,"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/categories?post=45269"},{"taxonomy":"post_tag","embeddable":true,"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/tags?post=45269"},{"taxonomy":"industries-application","embeddable":true,"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/industries-application?post=45269"},{"taxonomy":"post-badge","embeddable":true,"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/post-badge?post=45269"},{"taxonomy":"resource-type","embeddable":true,"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/resource-type?post=45269"},{"taxonomy":"product-library","embeddable":true,"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/product-library?post=45269"},{"taxonomy":"resource-technique","embeddable":true,"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/resource-technique?post=45269"},{"taxonomy":"author","embeddable":true,"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/ppma_author?post=45269"}],"curies":[{"name":"wp","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}