{"id":48515,"date":"2023-04-03T14:30:00","date_gmt":"2023-04-03T14:30:00","guid":{"rendered":"https:\/\/discover.restek.com\/uncategorized\/pfas-epa-1633-and-bile-acids\/"},"modified":"2026-04-06T16:33:30","modified_gmt":"2026-04-06T16:33:30","slug":"pfas-epa-1633-and-bile-acids","status":"publish","type":"post","link":"https:\/\/discover.restek.com\/de\/blogs\/gnbl4765\/pfas-epa-1633-and-bile-acids","title":{"rendered":"PFAS &#8211; EPA 1633 and Bile Acids"},"content":{"rendered":"\n<p class=\"wp-block-paragraph\">If you\u2019re here, it\u2019s likely you\u2019ve heard about or read the EPA draft titled \u201c<a href=\"https:\/\/www.epa.gov\/system\/files\/documents\/2022-07\/2nd%20Draft%20of%20Method%201633%20June%202022%20508-compliant.pdf\" target=\"_blank\" rel=\"noreferrer noopener\">Analysis of Per- and Polyfluoroalkyl Substances (PFAS) in Aqueous, Solid, Biosolids, and Tissue Samples by LC-MS\/MS<\/a>.\u201d In it, the interference from bile acids to PFOS is outlined along with guidelines to follow for the analysis of 41 PFAS compounds. This potential interference from the bile acids with PFOS is affected by the organic modifier used in the analytical method. The document states that when using acetonitrile as the organic modifier, the bile acid taurodeoxycholic acid (TDCA) needs to be monitored and resolved with at least 1 minute or more between the bile acid and PFOS. If using any other organic modifier, such as methanol, it is necessary to also include taurochenodeoxycholic acid (TCDCA) and tauroursodeoxycholic acid (TUDCA) and achieve 1 minute or greater retention between PFOS and all three bile acids. In addition to the retention time requirement, the bile acids may also not coelute with any of the PFAS analytes. A method was developed to meet these guidelines using acetonitrile.<\/p>\n\n\n<div class=\"wp-block-custom-chromatogram-article\"><div class=\"wp-block-custom-chromatogram-article\"><h3 class=\"article-title\">EPA 1633 PFAS on Force C18 by LC-MS\/MS using Acetonitrile<\/h3><div class=\"chromatogram-image regular-image\"><img decoding=\"async\" src=\"https:\/\/ez.restek.com\/images\/cgram\/lc_ev0585.png\" alt=\"EPA 1633 PFAS on Force C18 by LC-MS\/MS using Acetonitrile\" title=\"-\"><\/div><p class=\"article-id\" style=\"text-align:center\"> LC_EV0585<\/p><div class=\"chromatogram-peaks\"><h4>Peaks<\/h4><table class=\"peaks col-12\">\n<thead><tr><th><\/th><th style=\"text-align: left;width: 75px\">Peaks<\/th><th style=\"text-align: center;width: 75px\">t<sub>R<\/sub> (min)<\/th><th style=\"text-align: center;width: 75px\">Conc.<br \/>(ng\/mL)<\/th><th style=\"text-align: center;width: 75px\">Precursor 1<\/th><th style=\"text-align: center;width: 75px\">Product 1<\/th><th style=\"text-align: center;width: 75px\">Product 2<\/th><th style=\"text-align: center;width: 75px\">Precursor 2<\/th><th style=\"text-align: center;width: 75px\">Product 1<\/th><\/tr><\/thead>\n<tbody><tr><td class=\"num\">1.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Perfluorobutanoic acid (PFBA)\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/375-22-4\/Perfluorobutanoic acid\" target=\"_blank\" rel=\"noopener\">Perfluorobutanoic acid (PFBA)<\/a><\/td><td class=\"oth\">1.13<\/td><td class=\"oth\">200<\/td><td class=\"oth\">212.8<\/td><td class=\"oth\">169.0<\/td><td class=\"oth\">&#8211;<\/td><td class=\"oth\">&#8211;<\/td><td class=\"oth\">&#8211;<\/td><\/tr>\n<tr><td class=\"num\">2.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Perfluoro-3-methoxypropanoic acid (PFMPA)\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/377-73-1\/Perfluoro-3-methoxypropanoic acid\" target=\"_blank\" rel=\"noopener\">Perfluoro-3-methoxypropanoic acid (PFMPA)<\/a><\/td><td class=\"oth\">1.49<\/td><td class=\"oth\">100<\/td><td class=\"oth\">229.0<\/td><td class=\"oth\">85.0<\/td><td class=\"oth\">&#8211;<\/td><td class=\"oth\">&#8211;<\/td><td class=\"oth\">&#8211;<\/td><\/tr>\n<tr><td class=\"num\">3.<\/td><td class=\"cmpd\">3-Perfluoropropyl propanoic acid (3:3FTCA)<\/td><td class=\"oth\">1.75<\/td><td class=\"oth\">200<\/td><td class=\"oth\">241.0<\/td><td class=\"oth\">177.0<\/td><td class=\"oth\">117.0<\/td><td class=\"oth\">&#8211;<\/td><td class=\"oth\">&#8211;<\/td><\/tr>\n<tr><td class=\"num\">4.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Perfluoropentanoic acid (PFPeA)\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/2706-90-3\/Perfluoropentanoic acid\" target=\"_blank\" rel=\"noopener\">Perfluoropentanoic acid (PFPeA)<\/a><\/td><td class=\"oth\">2.39<\/td><td class=\"oth\">100<\/td><td class=\"oth\">263.0<\/td><td class=\"oth\">219.0<\/td><td class=\"oth\">68.9<\/td><td class=\"oth\">&#8211;<\/td><td class=\"oth\">&#8211;<\/td><\/tr>\n<tr><td class=\"num\">5.<\/td><td class=\"cmpd\">Perfluoro-4-methoxybutanoic acid (PFMBA)<\/td><td class=\"oth\">2.86<\/td><td class=\"oth\">100<\/td><td class=\"oth\">279.0<\/td><td class=\"oth\">85.0<\/td><td class=\"oth\">&#8211;<\/td><td class=\"oth\">&#8211;<\/td><td class=\"oth\">&#8211;<\/td><\/tr>\n<tr><td class=\"num\">6.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 1H,1H, 2H, 2H-Perfluorohexane sulfonic acid (4:2 FTS)\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/757124-72-4\/1H,1H, 2H, 2H-Perfluorohexane sulfonic acid\" target=\"_blank\" rel=\"noopener\">1H,1H, 2H, 2H-Perfluorohexane sulfonic acid (4:2 FTS)<\/a><\/td><td class=\"oth\">3.44<\/td><td class=\"oth\">200<\/td><td class=\"oth\">327.0<\/td><td class=\"oth\">307.0<\/td><td class=\"oth\">81.0<\/td><td class=\"oth\">&#8211;<\/td><td class=\"oth\">&#8211;<\/td><\/tr>\n<tr><td class=\"num\">7.<\/td><td class=\"cmpd\">Nonafluoro-3,6-dioxaheptanoic acid (NFDHA)<\/td><td class=\"oth\">3.71<\/td><td class=\"oth\">100<\/td><td class=\"oth\">294.9<\/td><td class=\"oth\">201.1<\/td><td class=\"oth\">85.0<\/td><td class=\"oth\">&#8211;<\/td><td class=\"oth\">&#8211;<\/td><\/tr>\n<tr><td class=\"num\">8.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Perfluorobutanesulfonic acid (PFBS)\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/375-73-5\/Perfluorobutanesulfonic acid\" target=\"_blank\" rel=\"noopener\">Perfluorobutanesulfonic acid (PFBS)<\/a><\/td><td class=\"oth\">3.87<\/td><td class=\"oth\">50<\/td><td class=\"oth\">299.0<\/td><td class=\"oth\">80.0<\/td><td class=\"oth\">99.0<\/td><td class=\"oth\">&#8211;<\/td><td class=\"oth\">&#8211;<\/td><\/tr>\n<tr><td class=\"num\">9.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Perfluorohexanoic acid (PFHxA)\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/307-24-4\/Perfluorohexanoic acid\" target=\"_blank\" rel=\"noopener\">Perfluorohexanoic acid (PFHxA)<\/a><\/td><td class=\"oth\">3.88<\/td><td class=\"oth\">50<\/td><td class=\"oth\">313.0<\/td><td class=\"oth\">269.0<\/td><td class=\"oth\">119.0<\/td><td class=\"oth\">&#8211;<\/td><td class=\"oth\">&#8211;<\/td><\/tr>\n<tr><td class=\"num\">10.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Hexafluoropropylene oxide dimer acid (HFPO-DA)\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/13252-13-6\/Hexafluoropropylene oxide dimer acid\" target=\"_blank\" rel=\"noopener\">Hexafluoropropylene oxide dimer acid (HFPO-DA)<\/a><\/td><td class=\"oth\">4.37<\/td><td class=\"oth\">100<\/td><td class=\"oth\">285.0<\/td><td class=\"oth\">169.0<\/td><td class=\"oth\">185.0<\/td><td class=\"oth\">&#8211;<\/td><td class=\"oth\">&#8211;<\/td><\/tr>\n<tr><td class=\"num\">11.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Perfluoro(2-ethoxyethane)sulfonic acid (PFEESA)\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/113507-82-7\/Perfluoro(2-ethoxyethane)sulfonic acid\" target=\"_blank\" rel=\"noopener\">Perfluoro(2-ethoxyethane)sulfonic acid (PFEESA)<\/a><\/td><td class=\"oth\">4.58<\/td><td class=\"oth\">100<\/td><td class=\"oth\">315.0<\/td><td class=\"oth\">135.0<\/td><td class=\"oth\">83.0<\/td><td class=\"oth\">&#8211;<\/td><td class=\"oth\">&#8211;<\/td><\/tr>\n<tr><td class=\"num\">12.<\/td><td class=\"cmpd\">2H,2H,3H,3H-Perfluorooctanoic acid (5:3FTCA)<\/td><td class=\"oth\">4.83<\/td><td class=\"oth\">20<\/td><td class=\"oth\">341.0<\/td><td class=\"oth\">237.0<\/td><td class=\"oth\">217.0<\/td><td class=\"oth\">&#8211;<\/td><td class=\"oth\">&#8211;<\/td><\/tr>\n<tr><td class=\"num\">13.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Perfluoroheptanoic acid (PFHpA)\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/375-85-9\/Perfluoroheptanoic acid\" target=\"_blank\" rel=\"noopener\">Perfluoroheptanoic acid (PFHpA)<\/a><\/td><td class=\"oth\">5.25<\/td><td class=\"oth\">50<\/td><td class=\"oth\">363.1<\/td><td class=\"oth\">319.0<\/td><td class=\"oth\">169.0<\/td><td class=\"oth\">&#8211;<\/td><td class=\"oth\">&#8211;<\/td><\/tr>\n<tr><td class=\"num\">14.<\/td><td class=\"cmpd\">Perfluoropentansulfonic acid (PFPeS)<\/td><td class=\"oth\">5.40<\/td><td class=\"oth\">50<\/td><td class=\"oth\">349.0<\/td><td class=\"oth\">80.0<\/td><td class=\"oth\">99.0<\/td><td class=\"oth\">&#8211;<\/td><td class=\"oth\">&#8211;<\/td><\/tr>\n<tr><td class=\"num\">15.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 4,8-Dioxa-3H-perfluorononanoic acid (ADONA)\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/919005-14-4\/4,8-Dioxa-3H-perfluorononanoic acid\" target=\"_blank\" rel=\"noopener\">4,8-Dioxa-3H-perfluorononanoic acid (ADONA)<\/a><\/td><td class=\"oth\">5.70<\/td><td class=\"oth\">100<\/td><td class=\"oth\">377.0<\/td><td class=\"oth\">251.0<\/td><td class=\"oth\">84.8<\/td><td class=\"oth\">&#8211;<\/td><td class=\"oth\">&#8211;<\/td><\/tr>\n<tr><td class=\"num\">16.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Taurodeoxycholic acid (TDCA)\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/516-50-7 \/Taurodeoxycholic acid\" target=\"_blank\" rel=\"noopener\">Taurodeoxycholic acid (TDCA)<\/a><\/td><td class=\"oth\">5.90<\/td><td class=\"oth\">100<\/td><td class=\"oth\">498.4<\/td><td class=\"oth\">80.0<\/td><td class=\"oth\">&#8211;<\/td><td class=\"oth\">&#8211;<\/td><td class=\"oth\">&#8211;<\/td><\/tr>\n<tr><td class=\"num\">17.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 1H,1H, 2H, 2H-Perfluorooctane sulfonic acid (6:2 FTS)\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/27619-97-2\/1H,1H, 2H, 2H-Perfluorooctane sulfonic acid\" target=\"_blank\" rel=\"noopener\">1H,1H, 2H, 2H-Perfluorooctane sulfonic acid (6:2 FTS)<\/a><\/td><td class=\"oth\">6.00<\/td><td class=\"oth\">200<\/td><td class=\"oth\">427.0<\/td><td class=\"oth\">407.0<\/td><td class=\"oth\">81.0<\/td><td class=\"oth\">&#8211;<\/td><td class=\"oth\">&#8211;<\/td><\/tr>\n<tr><td class=\"num\">18.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Perfluorooctanoic acid (PFOA)\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/335-67-1\/Perfluorooctanoic acid\" target=\"_blank\" rel=\"noopener\">Perfluorooctanoic acid (PFOA)<\/a><\/td><td class=\"oth\">6.46<\/td><td class=\"oth\">50<\/td><td class=\"oth\">413.0<\/td><td class=\"oth\">369.0<\/td><td class=\"oth\">169.0<\/td><td class=\"oth\">&#8211;<\/td><td class=\"oth\">&#8211;<\/td><\/tr>\n<tr><td class=\"num\">19.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Perfluorohexanesulfonic acid (PFHxS)\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/355-46-4\/Perfluorohexanesulfonic acid\" target=\"_blank\" rel=\"noopener\">Perfluorohexanesulfonic acid (PFHxS)<\/a><\/td><td class=\"oth\">6.74<\/td><td class=\"oth\">50<\/td><td class=\"oth\">399.0<\/td><td class=\"oth\">80.0<\/td><td class=\"oth\">99.0<\/td><td class=\"oth\">&#8211;<\/td><td class=\"oth\">&#8211;<\/td><\/tr>\n<tr><td class=\"num\">20.<\/td><td class=\"cmpd\">3-Perfluoroheptyl propanoic acid (7:3FTCA)<\/td><td class=\"oth\">7.42<\/td><td class=\"oth\">20<\/td><td class=\"oth\">441.0<\/td><td class=\"oth\">317.0<\/td><td class=\"oth\">337.0<\/td><td class=\"oth\">&#8211;<\/td><td class=\"oth\">&#8211;<\/td><\/tr>\n<tr><td class=\"num\">21.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Perfluorononanoic acid (PFNA)\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/375-95-1\/Perfluorononanoic acid\" target=\"_blank\" rel=\"noopener\">Perfluorononanoic acid (PFNA)<\/a><\/td><td class=\"oth\">7.57<\/td><td class=\"oth\">50<\/td><td class=\"oth\">463.0<\/td><td class=\"oth\">419.0<\/td><td class=\"oth\">219.0<\/td><td class=\"oth\">&#8211;<\/td><td class=\"oth\">&#8211;<\/td><\/tr>\n<tr><td class=\"num\">22.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Perfluoroheptanesulfonic acid (PFHpS)\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/375-92-8\/Perfluoroheptanesulfonic acid\" target=\"_blank\" rel=\"noopener\">Perfluoroheptanesulfonic acid (PFHpS)<\/a><\/td><td class=\"oth\">7.94<\/td><td class=\"oth\">50<\/td><td class=\"oth\">449.0<\/td><td class=\"oth\">80.0<\/td><td class=\"oth\">99.0<\/td><td class=\"oth\">&#8211;<\/td><td class=\"oth\">&#8211;<\/td><\/tr>\n<tr><td class=\"num\">23.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 1H,1H, 2H, 2H-Perfluorodecane sulfonic acid (8:2 FTS)\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/39108-34-4\/1H,1H, 2H, 2H-Perfluorodecane sulfonic acid\" target=\"_blank\" rel=\"noopener\">1H,1H, 2H, 2H-Perfluorodecane sulfonic acid (8:2 FTS)<\/a><\/td><td class=\"oth\">8.15<\/td><td class=\"oth\">200<\/td><td class=\"oth\">527.0<\/td><td class=\"oth\">507.0<\/td><td class=\"oth\">81.0<\/td><td class=\"oth\">&#8211;<\/td><td class=\"oth\">&#8211;<\/td><\/tr>\n<tr><td class=\"num\">24.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Perfluorodecanoic acid (PFDA)\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/335-76-2\/Perfluorodecanoic acid\" target=\"_blank\" rel=\"noopener\">Perfluorodecanoic acid (PFDA)<\/a><\/td><td class=\"oth\">8.53<\/td><td class=\"oth\">50<\/td><td class=\"oth\">513.0<\/td><td class=\"oth\">469.0<\/td><td class=\"oth\">219.0<\/td><td class=\"oth\">&#8211;<\/td><td class=\"oth\">&#8211;<\/td><\/tr>\n<tr><td class=\"num\">25.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for N-methyl perfluorooctanesulfonamidoacetic acid (NMeFOSAA)\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/2355-31-9\/N-methyl perfluorooctanesulfonamidoacetic acid\" target=\"_blank\" rel=\"noopener\">N-methyl perfluorooctanesulfonamidoacetic acid (NMeFOSAA)<\/a><\/td><td class=\"oth\">8.57<\/td><td class=\"oth\">50<\/td><td class=\"oth\">570.0<\/td><td class=\"oth\">419.0<\/td><td class=\"oth\">483.0<\/td><td class=\"oth\">&#8211;<\/td><td class=\"oth\">&#8211;<\/td><\/tr>\n<tr><td class=\"num\">26.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Perfluorooctanesulfonic acid (PFOS)\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/1763-23-1\/Perfluorooctanesulfonic acid\" target=\"_blank\" rel=\"noopener\">Perfluorooctanesulfonic acid (PFOS)<\/a><\/td><td class=\"oth\">8.80<\/td><td class=\"oth\">50<\/td><td class=\"oth\">499.0<\/td><td class=\"oth\">80.0<\/td><td class=\"oth\">99.0<\/td><td class=\"oth\">&#8211;<\/td><td class=\"oth\">&#8211;<\/td><\/tr>\n<tr><td class=\"num\">27.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for N-ethyl perfluorooctanesulfonamidoacetic acid (NEtFOSAA)\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/2991-50-6\/N-ethyl perfluorooctanesulfonamidoacetic acid\" target=\"_blank\" rel=\"noopener\">N-ethyl perfluorooctanesulfonamidoacetic acid (NEtFOSAA)<\/a><\/td><td class=\"oth\">8.83<\/td><td class=\"oth\">50<\/td><td class=\"oth\">584.0<\/td><td class=\"oth\">419.0<\/td><td class=\"oth\">526.0<\/td><td class=\"oth\">&#8211;<\/td><td class=\"oth\">&#8211;<\/td><\/tr>\n<tr><td class=\"num\">28.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Perfluoroundecanoic acid (PFUnA)\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/2058-94-8\/Perfluoroundecanoic acid\" target=\"_blank\" rel=\"noopener\">Perfluoroundecanoic acid (PFUnA)<\/a><\/td><td class=\"oth\">9.08<\/td><td class=\"oth\">50<\/td><td class=\"oth\">563.0<\/td><td class=\"oth\">519.0<\/td><td class=\"oth\">269.0<\/td><td class=\"oth\">&#8211;<\/td><td class=\"oth\">&#8211;<\/td><\/tr>\n<tr><td class=\"num\">29.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 9-Chlorohexadecafluoro-3-oxanonane-1-sulfonic acid (9Cl-PF3ONS)\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/756426-58-1\/9-Chlorohexadecafluoro-3-oxanonane-1-sulfonic acid\" target=\"_blank\" rel=\"noopener\">9-Chlorohexadecafluoro-3-oxanonane-1-sulfonic acid (9Cl-PF3ONS)<\/a><\/td><td class=\"oth\">9.17<\/td><td class=\"oth\">100<\/td><td class=\"oth\">531.0<\/td><td class=\"oth\">351.0<\/td><td class=\"oth\">&#8211;<\/td><td class=\"oth\">533.0<\/td><td class=\"oth\">353.0<\/td><\/tr>\n<tr><td class=\"num\">30.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Perfluorononanesulfonic acid (PFNS)\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/68259-12-1\/Perfluorononanesulfonic acid\" target=\"_blank\" rel=\"noopener\">Perfluorononanesulfonic acid (PFNS)<\/a><\/td><td class=\"oth\">9.30<\/td><td class=\"oth\">50<\/td><td class=\"oth\">549.0<\/td><td class=\"oth\">80.0<\/td><td class=\"oth\">99.0<\/td><td class=\"oth\">&#8211;<\/td><td class=\"oth\">&#8211;<\/td><\/tr>\n<tr><td class=\"num\">31.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Perfluorododecanoic acid (PFDoA)\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/307-55-1\/Perfluorododecanoic acid\" target=\"_blank\" rel=\"noopener\">Perfluorododecanoic acid (PFDoA)<\/a><\/td><td class=\"oth\">9.50<\/td><td class=\"oth\">50<\/td><td class=\"oth\">612.7<\/td><td class=\"oth\">568.8<\/td><td class=\"oth\">319.0<\/td><td class=\"oth\">&#8211;<\/td><td class=\"oth\">&#8211;<\/td><\/tr>\n<tr><td class=\"num\">32.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Perfluorodecanesulfonic acid (PFDS)\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/335-77-3\/Perfluorodecanesulfonic acid\" target=\"_blank\" rel=\"noopener\">Perfluorodecanesulfonic acid (PFDS)<\/a><\/td><td class=\"oth\">9.70<\/td><td class=\"oth\">50<\/td><td class=\"oth\">598.6<\/td><td class=\"oth\">80.0<\/td><td class=\"oth\">99.0<\/td><td class=\"oth\">&#8211;<\/td><td class=\"oth\">&#8211;<\/td><\/tr>\n<tr><td class=\"num\">33.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Perfluorotridecanoic acid (PFTrDA)\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/72629-94-8\/Perfluorotridecanoic acid\" target=\"_blank\" rel=\"noopener\">Perfluorotridecanoic acid (PFTrDA)<\/a><\/td><td class=\"oth\">9.87<\/td><td class=\"oth\">50<\/td><td class=\"oth\">662.8<\/td><td class=\"oth\">618.8<\/td><td class=\"oth\">169.0<\/td><td class=\"oth\">&#8211;<\/td><td class=\"oth\">&#8211;<\/td><\/tr>\n<tr><td class=\"num\">34.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 11-Chloroeicosafluoro-3-oxaundecane-1-sulfonic acid (11Cl-PF3OUdS)\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/763051-92-9\/11-Chloroeicosafluoro-3-oxaundecane-1-sulfonic acid\" target=\"_blank\" rel=\"noopener\">11-Chloroeicosafluoro-3-oxaundecane-1-sulfonic acid (11Cl-PF3OUdS)<\/a><\/td><td class=\"oth\">9.97<\/td><td class=\"oth\">100<\/td><td class=\"oth\">631.0<\/td><td class=\"oth\">451.0<\/td><td class=\"oth\">&#8211;<\/td><td class=\"oth\">633.0<\/td><td class=\"oth\">453.0<\/td><\/tr>\n<tr><td class=\"num\">35.<\/td><td class=\"cmpd\">Perfluorooctanesulfonamide (PFOSA)<\/td><td class=\"oth\">10.19<\/td><td class=\"oth\">50<\/td><td class=\"oth\">498.0<\/td><td class=\"oth\">78.0<\/td><td class=\"oth\">478.0<\/td><td class=\"oth\">&#8211;<\/td><td class=\"oth\">&#8211;<\/td><\/tr>\n<tr><td class=\"num\">36.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Perfluorotetradecanoic acid (PFTeDA)\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/376-06-7\/Perfluorotetradecanoic acid\" target=\"_blank\" rel=\"noopener\">Perfluorotetradecanoic acid (PFTeDA)<\/a><\/td><td class=\"oth\">10.21<\/td><td class=\"oth\">50<\/td><td class=\"oth\">712.7<\/td><td class=\"oth\">668.7<\/td><td class=\"oth\">169.0<\/td><td class=\"oth\">&#8211;<\/td><td class=\"oth\">&#8211;<\/td><\/tr>\n<tr><td class=\"num\">37.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Perfluorododecanesulfonic acid (PFDoS)\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/79780-39-5\/Perfluorododecanesulfonic acid\" target=\"_blank\" rel=\"noopener\">Perfluorododecanesulfonic acid (PFDoS)<\/a><\/td><td class=\"oth\">10.40<\/td><td class=\"oth\">50<\/td><td class=\"oth\">698.7<\/td><td class=\"oth\">80.0<\/td><td class=\"oth\">99.0<\/td><td class=\"oth\">&#8211;<\/td><td class=\"oth\">&#8211;<\/td><\/tr>\n<tr><td class=\"num\">38.<\/td><td class=\"cmpd\">N-methyl perfluorooctanesulfonamidoethanol (NMeFOSE)<\/td><td class=\"oth\">11.11<\/td><td class=\"oth\">500<\/td><td class=\"oth\">616.0<\/td><td class=\"oth\">59.0<\/td><td class=\"oth\">&#8211;<\/td><td class=\"oth\">&#8211;<\/td><td class=\"oth\">&#8211;<\/td><\/tr>\n<tr><td class=\"num\">39.<\/td><td class=\"cmpd\">N-methyl perfluorooctanesulfonamide (NMeFOSA)<\/td><td class=\"oth\">11.23<\/td><td class=\"oth\">50<\/td><td class=\"oth\">511.8<\/td><td class=\"oth\">219.0<\/td><td class=\"oth\">169.0<\/td><td class=\"oth\">&#8211;<\/td><td class=\"oth\">&#8211;<\/td><\/tr>\n<tr><td class=\"num\">40.<\/td><td class=\"cmpd\">N-ethyl perfluorooctanesulfonamidoethanol (NEtFOSE)<\/td><td class=\"oth\">11.41<\/td><td class=\"oth\">500<\/td><td class=\"oth\">630.0<\/td><td class=\"oth\">59.0<\/td><td class=\"oth\">&#8211;<\/td><td class=\"oth\">&#8211;<\/td><td class=\"oth\">&#8211;<\/td><\/tr>\n<tr><td class=\"num\">41.<\/td><td class=\"cmpd\">N-ethyl perfluorooctanesulfonamide (NEtFOSA)<\/td><td class=\"oth\">11.52<\/td><td class=\"oth\">50<\/td><td class=\"oth\">526.0<\/td><td class=\"oth\">219.0<\/td><td class=\"oth\">169.0<\/td><td class=\"oth\">&#8211;<\/td><td class=\"oth\">&#8211;<\/td><\/tr>\n<\/tbody><\/table><\/div><div class=\"chromatogram-conditions\"><h4>Conditions<\/h4><div class=\"conditions-container container-fluid\"><div class=\"row\"><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Column<\/th><td>Force C18  (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/9634252\" rel=\"noopener\">cat.# 9634252<\/a>)<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Dimensions:<\/th><td>50 mm x 2.1 mm ID<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Particle Size:<\/th><td>1.8 \u00b5m<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Pore Size:<\/th><td>100 \u00c5<\/td><\/tr><tr><td><\/td><tr><th class=\"sub conditions_header\" scope=\"row\">Guard Column:<\/th><td>Force C18 EXP guard column cartridge 5 mm, 2.1 mm ID, (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/963450252\" rel=\"noopener\">cat.# 963450252<\/a>)<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Temp.:<\/th><td>40 \u00b0C<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Standard\/Sample<\/th><td><\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Diluent:<\/th><td>Water, 5 mM ammonium acetate<\/td><\/tr><td><\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Inj. Vol.:<\/th><td>1 \u00b5L <\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Mobile Phase<\/th><td><\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">A:<\/th><td>Water, 5 mM ammonium acetate <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">B:<\/th><td>Acetonitrile <\/td><\/tr><tr><td><\/td><td><table class=\"cgram_ramp\"><thead><tr><th>Time (min)<\/th><th>Flow (mL\/min)<\/th><th>%A<\/th><th>%B<\/th><\/tr><\/thead><tbody><tr><td>0.00<\/td><td>0.4<\/td><td>80<\/td><td>20<\/td><\/tr><tr><td>7.00<\/td><td>0.4<\/td><td>55<\/td><td>45<\/td><\/tr><tr><td>12.00<\/td><td>0.4<\/td><td>0<\/td><td>100<\/td><\/tr><tr><td>12.01<\/td><td>0.4<\/td><td>80<\/td><td>20<\/td><\/tr><tr><td>14.00<\/td><td>0.4<\/td><td>80<\/td><td>20<\/td><\/tr><\/tbody><\/table><\/td><\/tr><\/table><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Detector<\/th><td>Shimadzu 8060 MS\/MS<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Ion Source:<\/th><td>Electrospray <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Ion Mode:<\/th><td>ESI- <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Mode:<\/th><td>MRM <\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Instrument<\/th><td>Shimadzu Nexera X2<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Sample Preparation<\/th><td>Samples were aliquoted into a 2.0 mL, 9 mm screw-thread polypropylene vials and capped with 9 mm solid-top polyethylene caps (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/23244\" rel=\"noopener\">cat.# 23244<\/a>).<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Notes<\/th><td>A PFAS delay column (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/27854\" rel=\"noopener\">cat.# 27854<\/a>) was installed before the injector.<\/td><\/tr><\/table><\/div><\/div><\/div><div class=\"chromatogram-related-skus\"><h4>Related SKUs<\/h4><ul><li>963450252<\/li><li>9634252<\/li><li>27854<\/li><li>23244<\/li><\/ul><\/div><div class=\"chromatogram-related-searches\"><h4>Related Searches<\/h4><ul><li>EPA 1633<\/li><li>PFAS<\/li><li>PFAS-library<\/li><\/ul><\/div><div class=\"chromatogram-pdf-link\"><a href=\"https:\/\/ez.restek.com\/images\/cgram\/lc_ev0585.pdf\" target=\"_blank\" rel=\"noopener noreferrer\"><svg xmlns=\"http:\/\/www.w3.org\/2000\/svg\" width=\"18\" height=\"18\" viewBox=\"0 0 18 18\"><g data-name=\"Group 2996\"><path data-name=\"Rectangle 1246\" d=\"M0 0h18v18H0z\" style=\"fill: none;\"><\/path><\/g><g data-name=\"Group 2997\"><path data-name=\"Path 729\" d=\"M13.412 11.4v2.017H5.345V11.4H4v2.017a1.349 1.349 0 0 0 1.345 1.345h8.068a1.349 1.349 0 0 0 1.345-1.345V11.4zm-.672-2.694-.948-.948-1.741 1.735V4H8.706v5.493L6.965 7.758l-.948.948 3.361 3.361z\" transform=\"translate(-.437 -.414)\" style=\"fill: rgb(13, 123, 196);\"><\/path><\/g><\/svg>Download PDF<\/a><\/div><\/div><\/div>\n\n\n<p class=\"wp-block-paragraph\">In this method, a Force C18 column is implemented using water containing 5 mM ammonium acetate and neat acetonitrile as mobile phases to achieve &gt; 1 minute resolution between PFOS (8.8 min.) and TDCA (5.8 min.) while also resolving TDCA from all other PFAS analytes in a quick, 14-minute cycle time. Even when all three bile acids are analyzed, resolution greater than 1 minute between the bile acids and PFOS is achieved when using ACN but the resolution between TUDCA (RT = 5.5 min.) and ADONA (RT = 5.7 min.) narrows.<\/p>\n\n\n\n<figure class=\"wp-block-image size-large\"><img loading=\"lazy\" decoding=\"async\" width=\"1024\" height=\"879\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-pfas-epa-1633-and-bile-acids-01-1024x879.png\" alt=\"\" class=\"wp-image-16456\" title=\"-\" srcset=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-pfas-epa-1633-and-bile-acids-01-1024x879.png 1024w, https:\/\/discover.restek.com\/wp-content\/uploads\/blog-pfas-epa-1633-and-bile-acids-01-300x258.png 300w, https:\/\/discover.restek.com\/wp-content\/uploads\/blog-pfas-epa-1633-and-bile-acids-01-768x659.png 768w, https:\/\/discover.restek.com\/wp-content\/uploads\/blog-pfas-epa-1633-and-bile-acids-01-1536x1319.png 1536w, https:\/\/discover.restek.com\/wp-content\/uploads\/blog-pfas-epa-1633-and-bile-acids-01.png 2000w\" sizes=\"auto, (max-width: 1024px) 100vw, 1024px\" \/><\/figure>\n\n\n\n<p class=\"wp-block-paragraph\">But what happens when we switch to methanol as the organic modifier? Now, instead of requiring the monitoring of only one bile acid interference, there are three that need to be resolved from PFOS by at least one minute while also not coeluting with any other PFAS compound per the requirements of EPA 1633. In addition to the EPA 1633 requirements that need to be met, the method needs to be kept as short as possible to not waste valuable instrument time. Let\u2019s look at some method scouting chromatograms using methanol as the organic modifier.<\/p>\n\n\n\n<figure class=\"wp-block-image size-large\"><img loading=\"lazy\" decoding=\"async\" width=\"773\" height=\"1024\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-pfas-epa-1633-and-bile-acids-02b-773x1024.png\" alt=\"\" class=\"wp-image-16474\" title=\"-\" srcset=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-pfas-epa-1633-and-bile-acids-02b-773x1024.png 773w, https:\/\/discover.restek.com\/wp-content\/uploads\/blog-pfas-epa-1633-and-bile-acids-02b-226x300.png 226w, https:\/\/discover.restek.com\/wp-content\/uploads\/blog-pfas-epa-1633-and-bile-acids-02b-768x1017.png 768w, https:\/\/discover.restek.com\/wp-content\/uploads\/blog-pfas-epa-1633-and-bile-acids-02b-1160x1536.png 1160w, https:\/\/discover.restek.com\/wp-content\/uploads\/blog-pfas-epa-1633-and-bile-acids-02b.png 1500w\" sizes=\"auto, (max-width: 773px) 100vw, 773px\" \/><\/figure>\n\n\n\n<figure class=\"wp-block-image size-large\"><img loading=\"lazy\" decoding=\"async\" width=\"1024\" height=\"633\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-pfas-epa-1633-and-bile-acids-02-1024x633.png\" alt=\"\" class=\"wp-image-16462\" title=\"-\" srcset=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-pfas-epa-1633-and-bile-acids-02-1024x633.png 1024w, https:\/\/discover.restek.com\/wp-content\/uploads\/blog-pfas-epa-1633-and-bile-acids-02-300x186.png 300w, https:\/\/discover.restek.com\/wp-content\/uploads\/blog-pfas-epa-1633-and-bile-acids-02-768x475.png 768w, https:\/\/discover.restek.com\/wp-content\/uploads\/blog-pfas-epa-1633-and-bile-acids-02-1536x950.png 1536w, https:\/\/discover.restek.com\/wp-content\/uploads\/blog-pfas-epa-1633-and-bile-acids-02.png 2000w\" sizes=\"auto, (max-width: 1024px) 100vw, 1024px\" \/><\/figure>\n\n\n\n<p class=\"wp-block-paragraph\">Using a scouting gradient, the 1633 standard and a mixture of bile acids were analyzed and their retention times compared. When using methanol, it is difficult to achieve the one-minute separation that is required for PFOS, TCDCA, and TDCA with only 0.263 minutes between PFOS and the closest eluting bile acid in the provided example. Next, the method was tested again using a longer run time and extending the gradient from 40-75% by two minutes.<\/p>\n\n\n\n<figure class=\"wp-block-image size-large\"><img loading=\"lazy\" decoding=\"async\" width=\"1024\" height=\"707\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-pfas-epa-1633-and-bile-acids-03-1024x707.png\" alt=\"\" class=\"wp-image-16480\" title=\"-\" srcset=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-pfas-epa-1633-and-bile-acids-03-1024x707.png 1024w, https:\/\/discover.restek.com\/wp-content\/uploads\/blog-pfas-epa-1633-and-bile-acids-03-300x207.png 300w, https:\/\/discover.restek.com\/wp-content\/uploads\/blog-pfas-epa-1633-and-bile-acids-03-768x530.png 768w, https:\/\/discover.restek.com\/wp-content\/uploads\/blog-pfas-epa-1633-and-bile-acids-03-1536x1060.png 1536w, https:\/\/discover.restek.com\/wp-content\/uploads\/blog-pfas-epa-1633-and-bile-acids-03.png 2000w\" sizes=\"auto, (max-width: 1024px) 100vw, 1024px\" \/><\/figure>\n\n\n\n<p class=\"wp-block-paragraph\">These efforts resulted in slightly more resolution between PFOS and the closest eluting bile acid, TCDCA, but at the cost of a longer method cycle time. Several other conditions were tested to meet the requirements of the EPA 1633 document using methanol, but no conditions tested were able to meet all conditions within a reasonable analysis time. Overall, acetonitrile proves to be the superior solvent choice for this analysis of EPA 1633 and is able to achieve all outlined requirements.<\/p>\n\n\n\n<div style=\"height:100px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n\n\n        <div class=\"cpb\">\n            <h3 class=\"cpb-heading\">Erw\u00e4hnte Produkte<\/h3>\n            <hr class=\"cpb-heading-underline\" \/>\n            <div class=\"cpb-list\">\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/de.restek.com\/p\/9634252\">                                Art.-Nr. 9634252                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">Force C18, 1.8 \u00b5m, 50 x 2.1mm LC-S\u00e4ule<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/de.restek.com\/p\/9634252\">Produkt anzeigen<\/a>\n                                                    <\/div>\n                    <\/div>\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/de.restek.com\/p\/963450252\">                                Art.-Nr. 963450252                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">Force C18 Vors\u00e4ulenkartusche, 5 x 2.1 mm EXP, 3er Pack<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/de.restek.com\/p\/963450252\">Produkt anzeigen<\/a>\n                                                    <\/div>\n                    <\/div>\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/de.restek.com\/p\/23244\">                                Art.-Nr. 23244                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">2.0 mL, 9 mm Solid Top-Polyethylen Kappen, Screw-Thread, 10 mil dicke Membran, farblos, 100er Pack<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/de.restek.com\/p\/23244\">Produkt anzeigen<\/a>\n                                                    <\/div>\n                    <\/div>\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/de.restek.com\/p\/27854\">                                Art.-Nr. 27854                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">PFAS Delay-S\u00e4ule, 5 \u00b5m, 50&#215;2.1 mm HPLC-S\u00e4ule<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/de.restek.com\/p\/27854\">Produkt anzeigen<\/a>\n                                                    <\/div>\n                    <\/div>\n                            <\/div>\n        <\/div>\n        \n\n\n<div style=\"height:100px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n","protected":false},"excerpt":{"rendered":"<p>If you&rsquo;re here, it&rsquo;s likely you&rsquo;ve heard about or read the EPA draft titled &ldquo;Analysis of Per- and Polyfluoroalkyl Substances (PFAS) in Aqueous, Solid, Biosolids, and Tissue Samples by LC-MS\/MS.&rdquo; In it, the interference from bile acids to PFOS is outlined along with guidelines to follow for the analysis of 41 PFAS compounds.<\/p>\n","protected":false},"author":47,"featured_media":0,"comment_status":"closed","ping_status":"open","sticky":false,"template":"","format":"standard","meta":{"_acf_changed":false,"_kad_blocks_custom_css":"","_kad_blocks_head_custom_js":"","_kad_blocks_body_custom_js":"","_kad_blocks_footer_custom_js":"","_kadence_starter_templates_imported_post":false,"_kad_post_transparent":"","_kad_post_title":"","_kad_post_layout":"","_kad_post_sidebar_id":"","_kad_post_content_style":"","_kad_post_vertical_padding":"","_kad_post_feature":"","_kad_post_feature_position":"","_kad_post_header":false,"_kad_post_footer":false,"footnotes":"","_ppma_block_editor_authors":""},"categories":[9],"tags":[2782],"industries-application":[2161,2165,2166],"post-badge":[],"resource-type":[],"product-library":[],"resource-technique":[2299,2302,2304],"ppma_author":[579],"class_list":["post-48515","post","type-post","status-publish","format-standard","hentry","category-blogs","tag-pfas","industries-application-environmental","industries-application-pfas-testing","industries-application-soil-sediment-biota-testing","resource-technique-liquid-chromatography","resource-technique-ms-ms","resource-technique-solid-phase-extraction-spe"],"acf":[],"taxonomy_info":{"category":[{"value":9,"label":"Blogs"}],"post_tag":[{"value":2782,"label":"PFAS"}],"industries-application":[{"value":2161,"label":"Environmental"},{"value":2165,"label":"PFAS Testing"},{"value":2166,"label":"Soil, Sediment, Biota Testing"}],"resource-technique":[{"value":2299,"label":"Liquid Chromatography"},{"value":2302,"label":"MS\/MS"},{"value":2304,"label":"Solid-Phase Extraction (SPE)"}]},"featured_image_src_large":false,"author_info":{"display_name":"Jamie L. York, PhD","author_link":"https:\/\/discover.restek.com\/de\/author\/jamie-l-york-phd\/"},"comment_info":0,"category_info":[{"term_id":9,"name":"Blogs","slug":"blogs","term_group":0,"term_taxonomy_id":9,"taxonomy":"category","description":"","parent":0,"count":457,"filter":"raw","cat_ID":9,"category_count":457,"category_description":"","cat_name":"Blogs","category_nicename":"blogs","category_parent":0}],"tag_info":[{"term_id":2782,"name":"PFAS","slug":"pfas","term_group":0,"term_taxonomy_id":2782,"taxonomy":"post_tag","description":"","parent":0,"count":124,"filter":"raw"}],"authors":[{"term_id":579,"user_id":47,"is_guest":0,"slug":"jamie-l-york-phd","display_name":"Jamie L. York, PhD","avatar_url":{"url":"https:\/\/discover.restek.com\/wp-content\/uploads\/york-jamie-1000x1000-1-e1765483005988.jpg","url2x":"https:\/\/discover.restek.com\/wp-content\/uploads\/york-jamie-1000x1000-1-e1765483005988.jpg"},"author_category":"1","first_name":"Jamie","last_name":"York","user_url":"https:\/\/www.linkedin.com\/in\/jamielyork\/","job_title":"Principal Scientist, LC Solutions","description":"Jamie York is a principal scientist in the Applications Lab at Restek Corporation. She leads the development of innovative analytical methods tailored to the food, clinical, environmental, and cannabis markets. Jamie earned her PhD in chemistry from The University of Texas at Arlington, where she gained extensive expertise in a range of analytical techniques, including gas chromatography\u2013vacuum ultraviolet (GC\u2013VUV); gas chromatography\u2013mass spectrometry (GC\u2013MS); matrix-assisted laser desorption\/ionization (MALDI); and liquid chromatography\u2013 mass spectrometry (LC\u2013MS\/MS); with a research emphasis on food and environmental analysis. Today, her work focuses on complex method development and advanced sample preparation strategies to support the evolving needs of the scientific community."}],"_links":{"self":[{"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/posts\/48515","targetHints":{"allow":["GET"]}}],"collection":[{"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/posts"}],"about":[{"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/types\/post"}],"author":[{"embeddable":true,"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/users\/47"}],"replies":[{"embeddable":true,"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/comments?post=48515"}],"version-history":[{"count":6,"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/posts\/48515\/revisions"}],"predecessor-version":[{"id":92207,"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/posts\/48515\/revisions\/92207"}],"wp:attachment":[{"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/media?parent=48515"}],"wp:term":[{"taxonomy":"category","embeddable":true,"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/categories?post=48515"},{"taxonomy":"post_tag","embeddable":true,"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/tags?post=48515"},{"taxonomy":"industries-application","embeddable":true,"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/industries-application?post=48515"},{"taxonomy":"post-badge","embeddable":true,"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/post-badge?post=48515"},{"taxonomy":"resource-type","embeddable":true,"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/resource-type?post=48515"},{"taxonomy":"product-library","embeddable":true,"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/product-library?post=48515"},{"taxonomy":"resource-technique","embeddable":true,"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/resource-technique?post=48515"},{"taxonomy":"author","embeddable":true,"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/ppma_author?post=48515"}],"curies":[{"name":"wp","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}