{"id":48520,"date":"2019-01-21T13:30:00","date_gmt":"2019-01-21T13:30:00","guid":{"rendered":"https:\/\/discover.restek.com\/uncategorized\/sorry-for-the-delay-here-is-your-chromatogram-with-and-without-the-pfas-delay-column\/"},"modified":"2026-04-08T19:09:43","modified_gmt":"2026-04-08T19:09:43","slug":"sorry-for-the-delay-here-is-your-chromatogram-with-and-without-the-pfas-delay-column","status":"publish","type":"post","link":"https:\/\/discover.restek.com\/de\/blogs\/gnbl4800\/sorry-for-the-delay-here-is-your-chromatogram-with-and-without-the-pfas-delay-column","title":{"rendered":"Sorry for the delay. Here is your chromatogram with and without the PFAS Delay Column."},"content":{"rendered":"\n<p class=\"wp-block-paragraph\">Sometimes I hear this from customers: I don&#8217;t need a delay column. I haven&#8217;t had any trouble without it so far.&#8221;<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">It might be true. You may not have any issues without the delay column, especially when your samples are, for example, highly PFAS contaminated soil samples because the interference is too small to affect your end result.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">I&#8217;ll show you the difference between PFAS analysis with and without the delay column.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Don&#8217;t be surprised. The chromatogram on the left is a blank sample injection! As you can see, there are peaks at the retention times you would expect the compounds from your sample to come out. Imagine what can possibly happen if you don&#8217;t have the delay column and you inject your sample containing low levels of PFAS.<\/p>\n\n\n<div class=\"wp-block-custom-chromatogram-article\"><div class=\"wp-block-custom-chromatogram-article\"><h3 class=\"article-title\">Background Perfluoroalkyl Substances in Waters ACQUITY UPLC I-Class Xevo TQ-S System<\/h3><div class=\"chromatogram-image regular-image\"><img decoding=\"async\" src=\"https:\/\/ez.restek.com\/images\/cgram\/lc_ev0549.png\" alt=\"Background Perfluoroalkyl Substances in Waters ACQUITY UPLC I-Class Xevo TQ-S System\" title=\"-\"><\/div><p class=\"article-id\" style=\"text-align:center\"> LC_EV0549<\/p><div class=\"chromatogram-peaks\"><h4>Peaks<\/h4><table class=\"peaks col-lg-6 col-12\">\n<thead><tr><th><\/th><th style=\"text-align: left;width: 75px\">Peaks<\/th><th style=\"text-align: center;width: 75px\">t<sub>R<\/sub> (min)<\/th><th style=\"text-align: center;width: 75px\">Precursor Ion<\/th><th style=\"text-align: center;width: 75px\">Product Ion<\/th><\/tr><\/thead>\n<tbody><tr><td class=\"num\">1.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Perfluorooctanoic acid (PFOA)\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/335-67-1\/Perfluorooctanoic acid\" target=\"_blank\" rel=\"noopener\">Perfluorooctanoic acid (PFOA)<\/a><\/td><td class=\"oth\">3.75<\/td><td class=\"oth\">413.2<\/td><td class=\"oth\">369.1<\/td><\/tr>\n<tr><td class=\"num\">2.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Perfluorononanoic acid (PFNA)\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/375-95-1\/Perfluorononanoic acid\" target=\"_blank\" rel=\"noopener\">Perfluorononanoic acid (PFNA)<\/a><\/td><td class=\"oth\">4.15<\/td><td class=\"oth\">463.2<\/td><td class=\"oth\">419.2<\/td><\/tr>\n<tr><td class=\"num\">3.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Perfluorodecanoic acid (PFDA)\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/335-76-2\/Perfluorodecanoic acid\" target=\"_blank\" rel=\"noopener\">Perfluorodecanoic acid (PFDA)<\/a><\/td><td class=\"oth\">4.48<\/td><td class=\"oth\">513.2<\/td><td class=\"oth\">469.2<\/td><\/tr>\n<tr><td class=\"num\">4.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Perfluoroundecanoic acid (PFUnA)\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/2058-94-8\/Perfluoroundecanoic acid\" target=\"_blank\" rel=\"noopener\">Perfluoroundecanoic acid (PFUnA)<\/a><\/td><td class=\"oth\">4.77<\/td><td class=\"oth\">563.2<\/td><td class=\"oth\">519.2<\/td><\/tr>\n<tr><td class=\"num\">5.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Perfluorododecanoic acid (PFDoA)\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/307-55-1\/Perfluorododecanoic acid\" target=\"_blank\" rel=\"noopener\">Perfluorododecanoic acid (PFDoA)<\/a><\/td><td class=\"oth\">5.02<\/td><td class=\"oth\">613.2<\/td><td class=\"oth\">569.2<\/td><\/tr>\n<\/tbody><\/table><\/div><div class=\"chromatogram-conditions\"><h4>Conditions<\/h4><div class=\"conditions-container container-fluid\"><div class=\"row\"><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Column<\/th><td>Raptor C18  (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/9304252\" rel=\"noopener\">cat.# 9304252<\/a>)<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Dimensions:<\/th><td>50 mm x 2.1 mm ID<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Particle Size:<\/th><td>1.8 \u00b5m<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Pore Size:<\/th><td>90 \u00c5<\/td><\/tr><tr><td><\/td><tr><th class=\"sub conditions_header\" scope=\"row\">Temp.:<\/th><td>40 \u00b0C<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Standard\/Sample<\/th><td>Blank injection without delay column after 10 minutes of equilibration.<\/td><\/tr>\n<tr><th class=\"sub conditions_header\" scope=\"row\">Diluent:<\/th><td>80:20 Water:methanol<\/td><\/tr><td><\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Inj. Vol.:<\/th><td>5 \u00b5L <\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Mobile Phase<\/th><td><\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">A:<\/th><td>Water, 5mM ammonium acetate <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">B:<\/th><td>Methanol <\/td><\/tr><tr><td><\/td><td><table class=\"cgram_ramp\"><thead><tr><th>Time (min)<\/th><th>Flow (mL\/min)<\/th><th>%A<\/th><th>%B<\/th><\/tr><\/thead><tbody><tr><td>0.00<\/td><td>0.4<\/td><td>80<\/td><td>20<\/td><\/tr><tr><td>6.00<\/td><td>0.4<\/td><td>5<\/td><td>95<\/td><\/tr><tr><td>6.50<\/td><td>0.4<\/td><td>5<\/td><td>95<\/td><\/tr><tr><td>6.51<\/td><td>0.4<\/td><td>80<\/td><td>20<\/td><\/tr><tr><td>8.00<\/td><td>0.4<\/td><td>80<\/td><td>20<\/td><\/tr><\/tbody><\/table><\/td><\/tr><\/table><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Detector<\/th><td>MS\/MS<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Ion Mode:<\/th><td>ESI- <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Mode:<\/th><td>MRM <\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Instrument<\/th><td>UHPLC<\/td><\/tr><\/table><\/div><\/div><\/div><div class=\"chromatogram-related-skus\"><h4>Related SKUs<\/h4><ul><li>9304252<\/li><\/ul><\/div><div class=\"chromatogram-related-searches\"><h4>Related Searches<\/h4><ul><li>PFAS<\/li><li>PFC<\/li><li>delay column<\/li><\/ul><\/div><div class=\"chromatogram-pdf-link\"><a href=\"https:\/\/ez.restek.com\/images\/cgram\/lc_ev0549.pdf\" target=\"_blank\" rel=\"noopener noreferrer\"><svg xmlns=\"http:\/\/www.w3.org\/2000\/svg\" width=\"18\" height=\"18\" viewBox=\"0 0 18 18\"><g data-name=\"Group 2996\"><path data-name=\"Rectangle 1246\" d=\"M0 0h18v18H0z\" style=\"fill: none;\"><\/path><\/g><g data-name=\"Group 2997\"><path data-name=\"Path 729\" d=\"M13.412 11.4v2.017H5.345V11.4H4v2.017a1.349 1.349 0 0 0 1.345 1.345h8.068a1.349 1.349 0 0 0 1.345-1.345V11.4zm-.672-2.694-.948-.948-1.741 1.735V4H8.706v5.493L6.965 7.758l-.948.948 3.361 3.361z\" transform=\"translate(-.437 -.414)\" style=\"fill: rgb(13, 123, 196);\"><\/path><\/g><\/svg>Download PDF<\/a><\/div><\/div><\/div>\n\n<div class=\"wp-block-custom-chromatogram-article\"><div class=\"wp-block-custom-chromatogram-article\"><h3 class=\"article-title\">Delayed Background Perfluoroalkyl Substances in Waters ACQUITY UPLC I-Class Xevo TQ-S System<\/h3><div class=\"chromatogram-image regular-image\"><img decoding=\"async\" src=\"https:\/\/ez.restek.com\/images\/cgram\/lc_ev0550.png\" alt=\"Delayed Background Perfluoroalkyl Substances in Waters ACQUITY UPLC I-Class Xevo TQ-S System\" title=\"-\"><\/div><p class=\"article-id\" style=\"text-align:center\"> LC_EV0550<\/p><div class=\"chromatogram-peaks\"><h4>Peaks<\/h4><table class=\"peaks col-lg-6 col-12\">\n<thead><tr><th><\/th><th style=\"text-align: left;width: 75px\">Peaks<\/th><th style=\"text-align: center;width: 75px\">Precursor Ion<\/th><th style=\"text-align: center;width: 75px\">Product Ion<\/th><\/tr><\/thead>\n<tbody><tr><td class=\"num\">1.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Perfluorooctanoic acid (PFOA)\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/335-67-1\/Perfluorooctanoic acid\" target=\"_blank\" rel=\"noopener\">Perfluorooctanoic acid (PFOA)<\/a><\/td><td class=\"oth\">413.2<\/td><td class=\"oth\">369.1<\/td><\/tr>\n<tr><td class=\"num\">2.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Perfluorononanoic acid (PFNA)\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/375-95-1\/Perfluorononanoic acid\" target=\"_blank\" rel=\"noopener\">Perfluorononanoic acid (PFNA)<\/a><\/td><td class=\"oth\">463.2<\/td><td class=\"oth\">419.2<\/td><\/tr>\n<tr><td class=\"num\">3.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Perfluorodecanoic acid (PFDA)\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/335-76-2\/Perfluorodecanoic acid\" target=\"_blank\" rel=\"noopener\">Perfluorodecanoic acid (PFDA)<\/a><\/td><td class=\"oth\">513.2<\/td><td class=\"oth\">469.2<\/td><\/tr>\n<tr><td class=\"num\">4.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Perfluoroundecanoic acid (PFUnA)\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/2058-94-8\/Perfluoroundecanoic acid\" target=\"_blank\" rel=\"noopener\">Perfluoroundecanoic acid (PFUnA)<\/a><\/td><td class=\"oth\">563.2<\/td><td class=\"oth\">519.2<\/td><\/tr>\n<tr><td class=\"num\">5.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Perfluorododecanoic acid (PFDoA)\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/307-55-1\/Perfluorododecanoic acid\" target=\"_blank\" rel=\"noopener\">Perfluorododecanoic acid (PFDoA)<\/a><\/td><td class=\"oth\">613.2<\/td><td class=\"oth\">569.2<\/td><\/tr>\n<\/tbody><\/table><\/div><div class=\"chromatogram-conditions\"><h4>Conditions<\/h4><div class=\"conditions-container container-fluid\"><div class=\"row\"><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Column<\/th><td>Raptor C18  (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/9304252\" rel=\"noopener\">cat.# 9304252<\/a>)<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Dimensions:<\/th><td>50 mm x 2.1 mm ID<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Particle Size:<\/th><td>1.8 \u00b5m<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Pore Size:<\/th><td>90 \u00c5<\/td><\/tr><tr><td><\/td><tr><th class=\"sub conditions_header\" scope=\"row\">Temp.:<\/th><td>40 \u00b0C<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Standard\/Sample<\/th><td>Blank injection with delay column (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/27854\" rel=\"noopener\">cat.# 27854<\/a>) after 10 minutes of equilibration.<\/td><\/tr>\n<tr><th class=\"sub conditions_header\" scope=\"row\">Diluent:<\/th><td>80:20 Water:methanol<\/td><\/tr><td><\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Inj. Vol.:<\/th><td>5 \u00b5L <\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Mobile Phase<\/th><td><\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">A:<\/th><td>Water, 5mM ammonium acetate <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">B:<\/th><td>Methanol <\/td><\/tr><tr><td><\/td><td><table class=\"cgram_ramp\"><thead><tr><th>Time (min)<\/th><th>Flow (mL\/min)<\/th><th>%A<\/th><th>%B<\/th><\/tr><\/thead><tbody><tr><td>0.00<\/td><td>0.4<\/td><td>80<\/td><td>20<\/td><\/tr><tr><td>6.00<\/td><td>0.4<\/td><td>5<\/td><td>95<\/td><\/tr><tr><td>6.50<\/td><td>0.4<\/td><td>5<\/td><td>95<\/td><\/tr><tr><td>6.51<\/td><td>0.4<\/td><td>80<\/td><td>20<\/td><\/tr><tr><td>8.00<\/td><td>0.4<\/td><td>80<\/td><td>20<\/td><\/tr><\/tbody><\/table><\/td><\/tr><\/table><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Detector<\/th><td>MS\/MS<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Ion Mode:<\/th><td>ESI- <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Mode:<\/th><td>MRM <\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Instrument<\/th><td>UHPLC<\/td><\/tr><\/table><\/div><\/div><\/div><div class=\"chromatogram-related-skus\"><h4>Related SKUs<\/h4><ul><li>9304252<\/li><li>27854<\/li><\/ul><\/div><div class=\"chromatogram-related-searches\"><h4>Related Searches<\/h4><ul><li>PFAS<\/li><li>PFC<\/li><li>delay column<\/li><\/ul><\/div><div class=\"chromatogram-pdf-link\"><a href=\"https:\/\/ez.restek.com\/images\/cgram\/lc_ev0550.pdf\" target=\"_blank\" rel=\"noopener noreferrer\"><svg xmlns=\"http:\/\/www.w3.org\/2000\/svg\" width=\"18\" height=\"18\" viewBox=\"0 0 18 18\"><g data-name=\"Group 2996\"><path data-name=\"Rectangle 1246\" d=\"M0 0h18v18H0z\" style=\"fill: none;\"><\/path><\/g><g data-name=\"Group 2997\"><path data-name=\"Path 729\" d=\"M13.412 11.4v2.017H5.345V11.4H4v2.017a1.349 1.349 0 0 0 1.345 1.345h8.068a1.349 1.349 0 0 0 1.345-1.345V11.4zm-.672-2.694-.948-.948-1.741 1.735V4H8.706v5.493L6.965 7.758l-.948.948 3.361 3.361z\" transform=\"translate(-.437 -.414)\" style=\"fill: rgb(13, 123, 196);\"><\/path><\/g><\/svg>Download PDF<\/a><\/div><\/div><\/div>\n\n\n<p class=\"wp-block-paragraph\">The chromatogram on the right also shows a blank sample injection, but this time the PFAS delay column was installed. You can see that the sharp peaks disappeared, but where are those delayed \u201dpeaks\u201d? &nbsp;&nbsp;The elevated signals coming AFTER the expected retention times of the compounds of interest are the system-related PFAS that were held up by the delay column.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Remember the&nbsp;PFAS delay column&nbsp;is installed BEFORE the injector in the LC-MS\/MS workflow and system-related interference will enter the delay column in a continuum from the mobile phase bottle and make their way through the analytical column to the detector. Unlike the injections of your sample which focus analytes at the head of the analytical column before the gradient starts, the interferences are continuously fed and they will be retained as they go through the delay column, diluted as they go through the injector, then retained again on analytical column not necessarily being focused much at any stage; hence resulting in broad and elevated signals.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Yes, the delayed interferences are broad because they enter the system continuously, and the signal is only elevated instead of looking like a sharp peak because the interferences exist from the very beginning of your analytical workflow (mobile phase) and are never focused on the analytical column like your sample. The magnitude of the delayed signal elevation depends on how long you equilibrate your LC column (length of column equilibration is related to how much your LC-MS\/MS system components leach out the interferences onto the delay column) and what LC-MS\/MS system you use (different LC-MS\/MS systems may have different levels of system-related interferences depending on the materials used in various component parts).<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Another way of minimizing system-related interferences is to replace all plastic parts in the workflow which may have PFAS leachates with PFAS-free parts such as PEEK. Still, the possibility of having background interferences from the workflow is endless since different LC-MS\/MS systems may need different parts to be replaced. This approach might be too time-consuming to be practical, and you can still struggle with finding unknown sources of the interferences.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Sometimes you may analyze PFAS samples with high concentration for remediation projects, &nbsp;e.g. contaminated soil from military bases or fire department practice sites, and sometimes your samples can be drinking water samples, which usually require low level detection (parts per trillion or even below).<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Regardless of the LC-MS\/MS systems you use for PFAS analysis and the level of concentrations you work on, having the&nbsp;PFAS delay column&nbsp;installed in your workflow BEFORE the injector will make you headache-free in your method development, method validation, and your day-to-day PFAS analysis.<\/p>\n\n\n\n<div style=\"height:100px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n\n\n        <div class=\"cpb\">\n            <h3 class=\"cpb-heading\">Erw\u00e4hnte Produkte<\/h3>\n            <hr class=\"cpb-heading-underline\" \/>\n            <div class=\"cpb-list\">\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/de.restek.com\/p\/9304252\">                                Art.-Nr. 9304252                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">Raptor C18, 1.8 \u00b5m, 50&#215;2.1 mm HPLC-S\u00e4ule<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/de.restek.com\/p\/9304252\">Produkt anzeigen<\/a>\n                                                    <\/div>\n                    <\/div>\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/de.restek.com\/p\/27854\">                                Art.-Nr. 27854                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">PFAS Delay-S\u00e4ule, 5 \u00b5m, 50&#215;2.1 mm HPLC-S\u00e4ule<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/de.restek.com\/p\/27854\">Produkt anzeigen<\/a>\n                                                    <\/div>\n                    <\/div>\n                            <\/div>\n        <\/div>\n        \n\n\n<div style=\"height:100px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n","protected":false},"excerpt":{"rendered":"<p>Sometimes I hear this from customers: I don&#8217;t need a delay column. I haven&#8217;t had any trouble without it so far.&#8221; It might be true. You may not have any issues without the delay column, especially when your samples are,&#8230;<\/p>\n","protected":false},"author":65,"featured_media":0,"comment_status":"closed","ping_status":"open","sticky":false,"template":"","format":"standard","meta":{"_acf_changed":false,"_kad_blocks_custom_css":"","_kad_blocks_head_custom_js":"","_kad_blocks_body_custom_js":"","_kad_blocks_footer_custom_js":"","_kadence_starter_templates_imported_post":false,"_kad_post_transparent":"","_kad_post_title":"","_kad_post_layout":"","_kad_post_sidebar_id":"","_kad_post_content_style":"","_kad_post_vertical_padding":"","_kad_post_feature":"","_kad_post_feature_position":"","_kad_post_header":false,"_kad_post_footer":false,"footnotes":"","_ppma_block_editor_authors":""},"categories":[9],"tags":[2782],"industries-application":[2161,2165],"post-badge":[],"resource-type":[],"product-library":[],"resource-technique":[2299,2301,2302],"ppma_author":[601],"class_list":["post-48520","post","type-post","status-publish","format-standard","hentry","category-blogs","tag-pfas","industries-application-environmental","industries-application-pfas-testing","resource-technique-liquid-chromatography","resource-technique-mass-spectrometry-ms","resource-technique-ms-ms"],"acf":[],"taxonomy_info":{"category":[{"value":9,"label":"Blogs"}],"post_tag":[{"value":2782,"label":"PFAS"}],"industries-application":[{"value":2161,"label":"Environmental"},{"value":2165,"label":"PFAS Testing"}],"resource-technique":[{"value":2299,"label":"Liquid Chromatography"},{"value":2301,"label":"Mass Spectrometry (MS)"},{"value":2302,"label":"MS\/MS"}]},"featured_image_src_large":false,"author_info":{"display_name":"SaVanna Cano","author_link":"https:\/\/discover.restek.com\/de\/author\/savanna-cano\/"},"comment_info":0,"category_info":[{"term_id":9,"name":"Blogs","slug":"blogs","term_group":0,"term_taxonomy_id":9,"taxonomy":"category","description":"","parent":0,"count":457,"filter":"raw","cat_ID":9,"category_count":457,"category_description":"","cat_name":"Blogs","category_nicename":"blogs","category_parent":0}],"tag_info":[{"term_id":2782,"name":"PFAS","slug":"pfas","term_group":0,"term_taxonomy_id":2782,"taxonomy":"post_tag","description":"","parent":0,"count":124,"filter":"raw"}],"authors":[{"term_id":601,"user_id":0,"is_guest":1,"slug":"mike-chang","display_name":"Mike Chang","avatar_url":{"url":"https:\/\/discover.restek.com\/wp-content\/uploads\/people-square-chang-mike.jpg","url2x":"https:\/\/discover.restek.com\/wp-content\/uploads\/people-square-chang-mike.jpg"},"author_category":"1","first_name":"Mike","last_name":"Chang","user_url":"","job_title":"","description":""}],"_links":{"self":[{"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/posts\/48520","targetHints":{"allow":["GET"]}}],"collection":[{"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/posts"}],"about":[{"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/types\/post"}],"author":[{"embeddable":true,"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/users\/65"}],"replies":[{"embeddable":true,"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/comments?post=48520"}],"version-history":[{"count":5,"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/posts\/48520\/revisions"}],"predecessor-version":[{"id":92304,"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/posts\/48520\/revisions\/92304"}],"wp:attachment":[{"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/media?parent=48520"}],"wp:term":[{"taxonomy":"category","embeddable":true,"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/categories?post=48520"},{"taxonomy":"post_tag","embeddable":true,"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/tags?post=48520"},{"taxonomy":"industries-application","embeddable":true,"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/industries-application?post=48520"},{"taxonomy":"post-badge","embeddable":true,"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/post-badge?post=48520"},{"taxonomy":"resource-type","embeddable":true,"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/resource-type?post=48520"},{"taxonomy":"product-library","embeddable":true,"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/product-library?post=48520"},{"taxonomy":"resource-technique","embeddable":true,"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/resource-technique?post=48520"},{"taxonomy":"author","embeddable":true,"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/ppma_author?post=48520"}],"curies":[{"name":"wp","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}