{"id":51597,"date":"2024-03-26T00:00:00","date_gmt":"2024-03-26T00:00:00","guid":{"rendered":"https:\/\/discover.restek.com\/uncategorized\/a-nitrosamines-crm-and-method-development-utilizing-resteks-pro-ezlc\/"},"modified":"2026-08-10T18:50:51","modified_gmt":"2026-08-10T18:50:51","slug":"a-nitrosamines-crm-and-method-development-utilizing-resteks-pro-ezlc","status":"publish","type":"post","link":"https:\/\/discover.restek.com\/de\/blogs-de\/gnbl4569\/a-nitrosamines-crm-and-method-development-utilizing-resteks-pro-ezlc","title":{"rendered":"A Nitrosamines CRM and Method Development Utilizing Restek\u2019s Pro <em>EZ<\/em>LC"},"content":{"rendered":"\n<p class=\"wp-block-paragraph\">We can be exposed to nitrosamines through our daily routines. They are largely unavoidable because they can exist in air, water, food, and pharmaceuticals to name a few routes of exposure. There is data that supports a link between elevated nitrosamine exposure and cancer. Therefore, action must be taken to limit our exposure to these harmful compounds. Focusing on pharmaceuticals, regulatory agencies, such as the FDA and EMA, have issued guidance drawn from internal research ICH recommendations regarding drug manufacturing and acceptable intake (AI) limits.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Simply put, what makes a compound a nitrosamine is the presence of the nitrosoamino group shown here:<\/p>\n\n\n\n<figure class=\"wp-block-image\"><img decoding=\"async\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-nitrosamines-crm-method-01.png\" alt=\"Nitrosamine structure\" title=\"-\"><\/figure>\n\n\n\n<p class=\"wp-block-paragraph\">There are countless compounds we can call nitrosamines. They can form from amines (secondary, tertiary, and quaternary) in the presence of nitrite salts under acidic conditions. These types of amines are also referred to as nitrosamine precursors. See below an example of nitrosamine formation from the FDA\u2019s Nitrosamine Guidelines[1]:<\/p>\n\n\n\n<figure class=\"wp-block-image\"><img decoding=\"async\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-nitrosamines-crm-method-02.png\" alt=\"Structure of nitrosamine formation.\" title=\"-\"><\/figure>\n\n\n\n<p class=\"wp-block-paragraph\">Nitrosamines can also form in the presence of an active pharmaceutical ingredient (API) during the drug manufacturing process as byproducts of the intended synthesis. These byproducts are called Nitroso-Drug Substance Related Impurities (NDSRI\u2019s).<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">A recent example of the need to monitor for all possible formations of nitrosamines is from July 2021. An API (Varenicline tartate, Chantix) that is prescribed as a smoking cessation aid had specific lots recalled due to this inadvertent creation of an NDSRI, N-nitroso-varenicline[4]. The FDA went so far as to publish a method using LC-ESI-HRMS titled <strong>Liquid Chromatography-High Resolution Mass Spectrometry (LC-ESI-HRMS) Method for the Determination of Varenicline Nitroso-Drug Substance Related Impurity (NDSRI) in Chantix\u2122 Drug Product and Varenicline Drug Substance<\/strong>[5]. The complication with NDSRI\u2019s is that there is not always a reference standard available specific to the parent compound, so other means need to be utilized to monitor NDSRI\u2019s, including LC\/QTOF. Beyond detection, synthesis of a reference standard would be useful for quantitation and regular quality testing by LC\/MS\/MS.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">The FDA guidelines are found in the <em>Control of Nitrosamine Impurities in Human Drugs<\/em>[1]. This <em>Guidance for Industry<\/em> lists seven potential nitrosamine impurities that can theoretically exist in API\u2019s and drug products. It is because of this guidance that we at Restek sought to develop a reference material that would streamline the analysis of these 7 nitrosamines (P\/N: 36025).<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">The seven theoretical nitrosamine impurities and structures are[1]:<\/p>\n\n\n\n<figure class=\"wp-block-image\"><img decoding=\"async\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-nitrosamines-crm-method-03.png\" alt=\"The structures of seven nitrosamine impurities.\" title=\"-\"><\/figure>\n\n\n\n<p class=\"wp-block-paragraph\"><em>*Please note the stereocenter around the secondary nitrogen.<\/em><\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Curious about chromatographic separation and overall workflow, I utilized <a href=\"https:\/\/ez.restek.com\/proezlc\" target=\"_blank\" rel=\"noreferrer noopener\">Pro <em>EZ<\/em>LC<\/a> to see how they might resolve on our Raptor Biphenyl.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">I started by selecting Nitrosamines as a compound class.<\/p>\n\n\n\n<figure class=\"wp-block-image\"><img decoding=\"async\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-nitrosamines-crm-method-04.jpg\" alt=\"Screenshot of Pro EZLC.\" title=\"-\"><\/figure>\n\n\n\n<p class=\"wp-block-paragraph\">Then, I selected all the nitrosamines in this certified reference material and chose Raptor Biphenyl from the drop-down menu under Phase. Note that they are many other common nitrosamines present for exploration and method development.<\/p>\n\n\n\n<figure class=\"wp-block-image\"><img decoding=\"async\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-nitrosamines-crm-method-05.jpg\" alt=\"Screenshot of Pro EZLC.\" title=\"-\"><\/figure>\n\n\n\n<p class=\"wp-block-paragraph\">Next, I clicked <em>Generate Model<\/em>.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">I tabbed over to <em>Conditions <\/em>to play with column dimensions, column temperature and flow rates. Keep in mind the <em>Back Pressure<\/em> calculation to make sure you aren\u2019t exceeding your max pressure for your instrumentation.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">There are many directions you can take in method development. Utilize the tools in Pro <em>EZ<\/em>LC to achieve your goals and find balance for what is most important to you and your lab. These are the goals I had in mind while utilizing <em>EZ<\/em>LC:<\/p>\n\n\n\n<ol class=\"wp-block-list\">\n<li><strong>Compound Retention:<\/strong> I didn\u2019t want any compound eluting before 1 minute.<\/li>\n\n\n\n<li><strong>Reproducibility:<\/strong> I wanted to add enough volume for at least 5 column rinses for re-equilibration.<\/li>\n\n\n\n<li><strong>Green Thumb:<\/strong> I wanted to lower my flow rates as much as possible while still maintaining back pressure for sharp resolution.<\/li>\n\n\n\n<li><strong>Fast LC:<\/strong> Toggling temperatures to maintain backpressure while balancing elution times.<\/li>\n\n\n\n<li><strong>Accessibility:<\/strong> I wanted to keep my max pressure under 6000psi to show separations without UHPLC.<\/li>\n<\/ol>\n\n\n\n<p class=\"wp-block-paragraph\">Here are the conditions that I landed on for my goals with this assay:<\/p>\n\n\n\n<figure class=\"wp-block-image\"><img decoding=\"async\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-nitrosamines-crm-method-06.png\" alt=\"Screenshot of Pro EZLC.\" title=\"-\"><\/figure>\n\n\n\n<p class=\"wp-block-paragraph\">Here is the chromatogram I would expect following the conditions and column selection using this reference standard:<\/p>\n\n\n\n<figure class=\"wp-block-image\"><img decoding=\"async\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-nitrosamines-crm-method-07.png\" alt=\"Screenshot of Pro EZLC.\" title=\"-\"><\/figure>\n\n\n\n<p class=\"wp-block-paragraph\">Here are the calculated RT\u2019s and MS conditions:<\/p>\n\n\n\n<figure class=\"wp-block-image\"><img decoding=\"async\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-nitrosamines-crm-method-08.png\" alt=\"Screenshot of Pro EZLC.\" title=\"-\"><\/figure>\n\n\n\n<p class=\"wp-block-paragraph\">Time to test our Pro <em>EZ<\/em>LC software against a real experiment. We plumbed the Raptor Biphenyl (cat.# 9309A12) and equilibrated the conditions on a Shimadzu Nexera X2. The compounds were optimized on a SCIEX 4500 MS\/MS. The ionization technique was APCI+. APCI is preferred for nitrosamines over ESI. It has been reported to provide better sensitivity and reproducibility.[2]<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Here are the results:<\/p>\n\n\n<div class=\"wp-block-custom-chromatogram-article\"><div class=\"wp-block-custom-chromatogram-article\"><div class=\"chromatogram-image regular-image\"><img decoding=\"async\" src=\"https:\/\/ez.restek.com\/images\/cgram\/lc_gn0698.png\" alt=\"Nitrosamines 7 Standard on Raptor Biphenyl\" title=\"-\"><\/div><p class=\"article-id\" style=\"text-align:center\"> LC_GN0698<\/p><div class=\"chromatogram-peaks\"><h4>Peaks<\/h4><table class=\"peaks col-lg-6 col-12\">\n<thead><tr><th><\/th><th style=\"text-align: left;width: 75px\">Peaks<\/th><th style=\"text-align: center;width: 75px\">t<sub>R<\/sub> (min)<\/th><th style=\"text-align: center;width: 75px\">Precursor Ion<\/th><th style=\"text-align: center;width: 75px\">Product Ion<\/th><\/tr><\/thead>\n<tbody><tr><td class=\"num\">1.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for N-Nitrosodimethylamine (NDMA)\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/62-75-9\/N-Nitrosodimethylamine\" target=\"_blank\" rel=\"noopener\">N-Nitrosodimethylamine (NDMA)<\/a><\/td><td class=\"oth\">1.62<\/td><td class=\"oth\">75.1<\/td><td class=\"oth\">58.0<\/td><\/tr>\n<tr><td class=\"num\">2.<\/td><td class=\"cmpd\">N-Nitroso-N-methyl-4-aminobutyric acid (NMBA)<\/td><td class=\"oth\">2.79<\/td><td class=\"oth\">147.0<\/td><td class=\"oth\">117.1<\/td><\/tr>\n<tr><td class=\"num\">3.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for N-Nitrosodiethylamine (NDEA)\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/55-18-5\/N-Nitrosodiethylamine\" target=\"_blank\" rel=\"noopener\">N-Nitrosodiethylamine (NDEA)<\/a><\/td><td class=\"oth\">4.05<\/td><td class=\"oth\">103.1<\/td><td class=\"oth\">75.0<\/td><\/tr>\n<tr><td class=\"num\">4.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for N-Nitrosoethylisopropylamine (NEIPA)\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/16339-04-1\/N-Nitrosoethylisopropylamine\" target=\"_blank\" rel=\"noopener\">N-Nitrosoethylisopropylamine (NEIPA)<\/a><\/td><td class=\"oth\">5.10<\/td><td class=\"oth\">117.1<\/td><td class=\"oth\">75.0<\/td><\/tr>\n<tr><td class=\"num\">5.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for N-Nitrosodiisopropylamine (NDIPA)\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/601-77-4\/N-Nitrosodiisopropylamine\" target=\"_blank\" rel=\"noopener\">N-Nitrosodiisopropylamine (NDIPA)<\/a><\/td><td class=\"oth\">5.99<\/td><td class=\"oth\">131.2<\/td><td class=\"oth\">43.1<\/td><\/tr>\n<tr><td class=\"num\">6.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for N-Nitrosomethylphenylamine (NMPA)\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/614-00-6\/N-Nitrosomethylphenylamine\" target=\"_blank\" rel=\"noopener\">N-Nitrosomethylphenylamine (NMPA)<\/a><\/td><td class=\"oth\">6.55<\/td><td class=\"oth\">137.0<\/td><td class=\"oth\">65.9<\/td><\/tr>\n<tr><td class=\"num\">7.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for N-Nitrosodibutylamine (NDBA)\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/924-16-3\/N-Nitrosodibutylamine\" target=\"_blank\" rel=\"noopener\">N-Nitrosodibutylamine (NDBA)<\/a><\/td><td class=\"oth\">7.64<\/td><td class=\"oth\">159.1<\/td><td class=\"oth\">57.0<\/td><\/tr>\n<\/tbody><\/table><\/div><div class=\"chromatogram-conditions\"><h4>Conditions<\/h4><div class=\"conditions-container container-fluid\"><div class=\"row\"><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Column<\/th><td>Raptor Biphenyl  (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/9309A12\" rel=\"noopener\">cat.# 9309A12<\/a>)<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Dimensions:<\/th><td>100 mm x 2.1 mm ID<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Particle Size:<\/th><td>2.7 \u00b5m<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Pore Size:<\/th><td>90 \u00c5<\/td><\/tr><tr><td><\/td><tr><th class=\"sub conditions_header\" scope=\"row\">Temp.:<\/th><td>40 \u00b0C<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Standard\/Sample<\/th><td>Nitrosamines 7 standard (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/36025\" rel=\"noopener\">cat.# 36025<\/a>)<\/td><\/tr>\n<tr><th class=\"sub conditions_header\" scope=\"row\">Diluent:<\/th><td>Water<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Conc.:<\/th><td>100 ng\/mL<\/td><\/tr><td><\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Inj. Vol.:<\/th><td>5 \u00b5L <\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Mobile Phase<\/th><td><\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">A:<\/th><td>Water, 0.1% formic acid <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">B:<\/th><td>Methanol, 0.1% formic acid <\/td><\/tr><tr><td><\/td><td><table class=\"cgram_ramp\"><thead><tr><th>Time (min)<\/th><th>Flow (mL\/min)<\/th><th>%A<\/th><th>%B<\/th><\/tr><\/thead><tbody><tr><td>0.00<\/td><td>0.4<\/td><td>95<\/td><td>5<\/td><\/tr><tr><td>9.00<\/td><td>0.4<\/td><td>5<\/td><td>95<\/td><\/tr><tr><td>9.01<\/td><td>0.4<\/td><td>95<\/td><td>5<\/td><\/tr><tr><td>11.00<\/td><td>0.4<\/td><td>95<\/td><td>5<\/td><\/tr><\/tbody><\/table><\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Max Pressure:<\/th><td>340 bar<\/td><\/tr><\/table><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Detector<\/th><td>SCIEX 4500 MS\/MS<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Ion Source:<\/th><td>APCI <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Ion Mode:<\/th><td>APCI+ <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Mode:<\/th><td>Scheduled MRM <\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Instrument<\/th><td>Shimadzu Nexera X2<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Sample Preparation<\/th><td>A 100 ng\/mL stock was prepared by adding 1 \u00b5L of nitrosamines 7 standard to 999 \u00b5L of water in 2 mL, screw-thread amber vials (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/21143\" rel=\"noopener\">cat.# 21143<\/a>) and capped with 9 mm, short-cap, screw-vial closures (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/24498\" rel=\"noopener\">cat.# 24498<\/a>).<\/td><\/tr><\/table><\/div><\/div><\/div><div class=\"chromatogram-related-skus\"><h4>Related SKUs<\/h4><ul><li>36025<\/li><li>9309A12<\/li><li>21143<\/li><li>24498<\/li><\/ul><\/div><div class=\"chromatogram-related-searches\"><h4>Related Searches<\/h4><ul><li>nitrosamines<\/li><\/ul><\/div><div class=\"chromatogram-pdf-link\"><a href=\"https:\/\/ez.restek.com\/images\/cgram\/lc_gn0698.pdf\" target=\"_blank\" rel=\"noopener noreferrer\"><svg xmlns=\"http:\/\/www.w3.org\/2000\/svg\" width=\"18\" height=\"18\" viewBox=\"0 0 18 18\"><g data-name=\"Group 2996\"><path data-name=\"Rectangle 1246\" d=\"M0 0h18v18H0z\" style=\"fill: none;\"><\/path><\/g><g data-name=\"Group 2997\"><path data-name=\"Path 729\" d=\"M13.412 11.4v2.017H5.345V11.4H4v2.017a1.349 1.349 0 0 0 1.345 1.345h8.068a1.349 1.349 0 0 0 1.345-1.345V11.4zm-.672-2.694-.948-.948-1.741 1.735V4H8.706v5.493L6.965 7.758l-.948.948 3.361 3.361z\" transform=\"translate(-.437 -.414)\" style=\"fill: rgb(13, 123, 196);\"><\/path><\/g><\/svg>Download PDF<\/a><\/div><\/div><\/div>\n\n\n<p class=\"wp-block-paragraph\">Comparing the two tables, we see that there is very little difference between retention times in the Pro <em>EZ<\/em>LC tool and the actual results using identical conditions.<\/p>\n\n\n\n<figure class=\"wp-block-table is-style-stripes\"><table class=\"has-fixed-layout\"><tbody><tr><td><strong>Compound<\/strong><\/td><td><strong>Pro <em>EZ<\/em>LC<\/strong><\/td><td><strong>Actual<\/strong><\/td><td><strong>Percent Difference<\/strong><\/td><\/tr><tr><td>NDMA<\/td><td>1.56<\/td><td>1.62<\/td><td>3.84<\/td><\/tr><tr><td>NMBA<\/td><td>2.79<\/td><td>2.79<\/td><td>0<\/td><\/tr><tr><td>NDEA<\/td><td>4.06<\/td><td>4.05<\/td><td>0.24<\/td><\/tr><tr><td>NEIPA<\/td><td>5.1<\/td><td>5.1<\/td><td>0<\/td><\/tr><tr><td>NDIPA<\/td><td>5.97<\/td><td>5.99<\/td><td>0.33<\/td><\/tr><tr><td>NMPA<\/td><td>6.49<\/td><td>6.55<\/td><td>0.92<\/td><\/tr><tr><td>NBDA<\/td><td>7.61<\/td><td>7.64<\/td><td>0.39<\/td><\/tr><\/tbody><\/table><\/figure>\n\n\n\n<p class=\"wp-block-paragraph\"><\/p>\n\n\n\n<p class=\"wp-block-paragraph\">When I first analyzed this chromatogram, it was concerning to me that there is some peak splitting. Investigating this further, we can expect a mixture of chiral compounds, and thus peak doublets, from NMBA, NEIPA and NMPA[2]. These compounds have stereocenters around their secondary nitrogen. It is impossible to isolate these nitrogenic stereocenters due to their rapid interconversion of their enantiomers.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">This poses a great question: Is the additional selectivity of our Raptor biphenyl the ideal column for this analysis?<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">It makes the most sense to avoid this separation of enantiomers. If nitrosamines are present in an unknown, we want to account for the total nitrosamine. This means we would need to integrate both peaks in all samples, calibrators, and controls. This creates potential possibility of error because it introduces a factor of interpretation alongside co-eluting peaks that can adversely affect accuracy and precision. It also leads to less sensitivity, given that these compounds are not stacked on top of each other contributing to a higher S\/N. Referencing USP General Chapter &lt;1469&gt; Nitrosamine Impurities, Procedure 3, USP recommends an L1 column[3]. This is more commonly known as a C18 column. Going back to the drawing board, and running the exact same assay using our Raptor C18 we observe the following chromatogram:<\/p>\n\n\n<div class=\"wp-block-custom-chromatogram-article\"><div class=\"wp-block-custom-chromatogram-article\"><div class=\"chromatogram-image wide-image\"><img decoding=\"async\" src=\"https:\/\/ez.restek.com\/images\/cgram\/lc_gn0700.png\" alt=\"Nitrosamines 7 Standard on Raptor C18\" title=\"-\"><\/div><p class=\"article-id\" style=\"text-align:center\"> LC_GN0700<\/p><div class=\"chromatogram-peaks\"><h4>Peaks<\/h4><table class=\"peaks col-lg-6 col-12\">\n<thead><tr><th><\/th><th style=\"text-align: left;width: 75px\">Peaks<\/th><th style=\"text-align: center;width: 75px\">t<sub>R<\/sub> (min)<\/th><th style=\"text-align: center;width: 75px\">Precursor Ion<\/th><th style=\"text-align: center;width: 75px\">Product Ion<\/th><\/tr><\/thead>\n<tbody><tr><td class=\"num\">1.<\/td><td class=\"cmpd\">&lt;a class=&quot;cmpd_link&quot; title=&quot;View compound information for <i>N<\/i>-Nitrosodimethylamine (NDMA)&#8221; title=&#8221;View compound information for <i>N<\/i>-Nitrosodimethylamine (NDMA)&#8221; href=&#8221;https:\/\/ez.restek.com\/compound\/view\/en\/62-75-9\/<i>N-Nitrosodimethylamine&#8221;&gt;<i>N<\/i>-Nitrosodimethylamine (NDMA)<\/a><\/td><td class=\"oth\">0.93<\/td><td class=\"oth\">75.1<\/td><td class=\"oth\">58.0<\/td><\/tr>\n<tr><td class=\"num\">2.<\/td><td class=\"cmpd\"><i>N<\/i>-Nitroso-N-methyl-4-aminobutyric acid (NMBA)<\/td><td class=\"oth\">1.58<\/td><td class=\"oth\">147.0<\/td><td class=\"oth\">117.1<\/td><\/tr>\n<tr><td class=\"num\">3.<\/td><td class=\"cmpd\">&lt;a class=&quot;cmpd_link&quot; title=&quot;View compound information for <i>N<\/i>-Nitrosodiethylamine (NDEA)&#8221; title=&#8221;View compound information for <i>N<\/i>-Nitrosodiethylamine (NDEA)&#8221; href=&#8221;https:\/\/ez.restek.com\/compound\/view\/en\/55-18-5\/<i>N-Nitrosodiethylamine&#8221;&gt;<i>N<\/i>-Nitrosodiethylamine (NDEA)<\/a><\/td><td class=\"oth\">2.65<\/td><td class=\"oth\">103.1<\/td><td class=\"oth\">75.0<\/td><\/tr>\n<tr><td class=\"num\">4.<\/td><td class=\"cmpd\">&lt;a class=&quot;cmpd_link&quot; title=&quot;View compound information for <i>N<\/i>-Nitrosoethylisopropylamine (NEIPA)&#8221; title=&#8221;View compound information for <i>N<\/i>-Nitrosoethylisopropylamine (NEIPA)&#8221; href=&#8221;https:\/\/ez.restek.com\/compound\/view\/en\/16339-04-1\/<i>N-Nitrosoethylisopropylamine&#8221;&gt;<i>N<\/i>-Nitrosoethylisopropylamine (NEIPA)<\/a><\/td><td class=\"oth\">3.66<\/td><td class=\"oth\">117.1<\/td><td class=\"oth\">75.0<\/td><\/tr>\n<tr><td class=\"num\">5.<\/td><td class=\"cmpd\">&lt;a class=&quot;cmpd_link&quot; title=&quot;View compound information for <i>N<\/i>-Nitrosodiisopropylamine (NDIPA)&#8221; title=&#8221;View compound information for <i>N<\/i>-Nitrosodiisopropylamine (NDIPA)&#8221; href=&#8221;https:\/\/ez.restek.com\/compound\/view\/en\/601-77-4\/<i>N-Nitrosodiisopropylamine&#8221;&gt;<i>N<\/i>-Nitrosodiisopropylamine (NDIPA)<\/a><\/td><td class=\"oth\">4.61<\/td><td class=\"oth\">131.2<\/td><td class=\"oth\">43.1<\/td><\/tr>\n<tr><td class=\"num\">6.<\/td><td class=\"cmpd\">&lt;a class=&quot;cmpd_link&quot; title=&quot;View compound information for <i>N<\/i>-Nitrosomethylphenylamine (NMPA)&#8221; title=&#8221;View compound information for <i>N<\/i>-Nitrosomethylphenylamine (NMPA)&#8221; href=&#8221;https:\/\/ez.restek.com\/compound\/view\/en\/614-00-6\/<i>N-Nitrosomethylphenylamine&#8221;&gt;<i>N<\/i>-Nitrosomethylphenylamine (NMPA)<\/a><\/td><td class=\"oth\">4.82<\/td><td class=\"oth\">137.0<\/td><td class=\"oth\">65.9<\/td><\/tr>\n<tr><td class=\"num\">7.<\/td><td class=\"cmpd\">&lt;a class=&quot;cmpd_link&quot; title=&quot;View compound information for <i>N<\/i>-Nitrosodibutylamine (NDBA)&#8221; title=&#8221;View compound information for <i>N<\/i>-Nitrosodibutylamine (NDBA)&#8221; href=&#8221;https:\/\/ez.restek.com\/compound\/view\/en\/924-16-3\/<i>N-Nitrosodibutylamine&#8221;&gt;<i>N<\/i>-Nitrosodibutylamine (NDBA)<\/a><\/td><td class=\"oth\">6.84<\/td><td class=\"oth\">159.1<\/td><td class=\"oth\">57.0<\/td><\/tr>\n<\/tbody><\/table><\/div><div class=\"chromatogram-conditions\"><h4>Conditions<\/h4><div class=\"conditions-container container-fluid\"><div class=\"row\"><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Column<\/th><td>Raptor C18  (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/9304A12\" rel=\"noopener\">cat.# 9304A12<\/a>)<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Dimensions:<\/th><td>100 mm x 2.1 mm ID<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Particle Size:<\/th><td>2.7 \u00b5m<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Pore Size:<\/th><td>90 \u00c5<\/td><\/tr><tr><td><\/td><tr><th class=\"sub conditions_header\" scope=\"row\">Temp.:<\/th><td>40 \u00b0C<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Standard\/Sample<\/th><td>Nitrosamines 7 Standard (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/36025\" rel=\"noopener\">cat.# 36025<\/a>)<\/td><\/tr>\n<tr><th class=\"sub conditions_header\" scope=\"row\">Diluent:<\/th><td>Water<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Conc.:<\/th><td>100 ng\/mL<\/td><\/tr><td><\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Inj. Vol.:<\/th><td>5 \u00b5L <\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Mobile Phase<\/th><td><\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">A:<\/th><td>Water, 0.1% formic acid <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">B:<\/th><td>Methanol, 0.1% formic acid <\/td><\/tr><tr><td><\/td><td><table class=\"cgram_ramp\"><thead><tr><th>Time (min)<\/th><th>Flow (mL\/min)<\/th><th>%A<\/th><th>%B<\/th><\/tr><\/thead><tbody><tr><td>0.00<\/td><td>0.4<\/td><td>95<\/td><td>5<\/td><\/tr><tr><td>9.00<\/td><td>0.4<\/td><td>5<\/td><td>95<\/td><\/tr><tr><td>9.01<\/td><td>0.4<\/td><td>95<\/td><td>5<\/td><\/tr><tr><td>11.00<\/td><td>0.4<\/td><td>95<\/td><td>5<\/td><\/tr><\/tbody><\/table><\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Max Pressure:<\/th><td>340 bar<\/td><\/tr><\/table><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Detector<\/th><td>SCIEX 4500 MS\/MS<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Ion Source:<\/th><td>APCI <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Ion Mode:<\/th><td>APCI+ <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Mode:<\/th><td>Scheduled MRM <\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Instrument<\/th><td>Shimadzu Nexera X2<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Sample Preparation<\/th><td>A 100 ng\/mL stock was prepared by adding 1 \u00b5L of Nitrosamines 7 Standard to 999 \u00b5L of water in 2 mL, screw-thread amber vials (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/21143\" rel=\"noopener\">cat.# 21143<\/a>) and capped with 9 mm, short-cap, screw-vial closures (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/24498\" rel=\"noopener\">cat.# 24498<\/a>).<\/td><\/tr><\/table><\/div><\/div><\/div><div class=\"chromatogram-related-skus\"><h4>Related SKUs<\/h4><ul><li>36025<\/li><li>9304A12<\/li><li>21143<\/li><li>24498<\/li><\/ul><\/div><div class=\"chromatogram-related-searches\"><h4>Related Searches<\/h4><ul><li>nitrosamines<\/li><\/ul><\/div><div class=\"chromatogram-pdf-link\"><a href=\"https:\/\/ez.restek.com\/images\/cgram\/lc_gn0700.pdf\" target=\"_blank\" rel=\"noopener noreferrer\"><svg xmlns=\"http:\/\/www.w3.org\/2000\/svg\" width=\"18\" height=\"18\" viewBox=\"0 0 18 18\"><g data-name=\"Group 2996\"><path data-name=\"Rectangle 1246\" d=\"M0 0h18v18H0z\" style=\"fill: none;\"><\/path><\/g><g data-name=\"Group 2997\"><path data-name=\"Path 729\" d=\"M13.412 11.4v2.017H5.345V11.4H4v2.017a1.349 1.349 0 0 0 1.345 1.345h8.068a1.349 1.349 0 0 0 1.345-1.345V11.4zm-.672-2.694-.948-.948-1.741 1.735V4H8.706v5.493L6.965 7.758l-.948.948 3.361 3.361z\" transform=\"translate(-.437 -.414)\" style=\"fill: rgb(13, 123, 196);\"><\/path><\/g><\/svg>Download PDF<\/a><\/div><\/div><\/div>\n\n\n<p class=\"wp-block-paragraph\">As you can see, the C18 still has some peak splitting for the compounds that can form enantiomers. However, there is more overlap and enough resolution between all compounds. There is some co-elution between NDIPA and NMPA. However, these compounds are not isobars, and the MS can easily differentiate between the two.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Pro <em>EZ<\/em>LC is a great tool for beginners to experts in LC\/MS. It\u2019s especially useful for new applications that you may not be familiar with. Be aware of the possibility of separating out racemic mixtures when using biphenyl.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">If you find Pro <em>EZ<\/em>LC helpful and utilize GC, be sure to check out <a href=\"https:\/\/ez.restek.com\/proezgc\" target=\"_blank\" rel=\"noopener\">Pro <em>EZ<\/em>GC<\/a> too.<\/p>\n\n\n\n<h3 class=\"wp-block-heading\">References:<\/h3>\n\n\n\n<ol class=\"wp-block-list\">\n<li><a href=\"https:\/\/www.fda.gov\/media\/141720\/download#:~:text=Because%20nitrosamines%20are%20probable%20or,their%20APIs%20and%20drug%20products\" target=\"_blank\" rel=\"noreferrer noopener\">https:\/\/www.fda.gov\/media\/141720\/download#:~:text=Because%20nitrosamines%20are%20probable%20or,their%20APIs%20and%20drug%20products.<\/a><\/li>\n\n\n\n<li><a href=\"https:\/\/pubmed.ncbi.nlm.nih.gov\/35696119\/\" target=\"_blank\" rel=\"noreferrer noopener\">https:\/\/pubmed.ncbi.nlm.nih.gov\/35696119\/<\/a><\/li>\n\n\n\n<li><a href=\"https:\/\/www.usp.org\/sites\/default\/files\/usp\/document\/stakeholder-forum\/pnp\/highlights-of-1469-nitrosamine-impurities.pdf\" target=\"_blank\" rel=\"noreferrer noopener\">https:\/\/www.usp.org\/sites\/default\/files\/usp\/document\/stakeholder-forum\/pnp\/highlights-of-1469-nitrosamine-impurities.pdf<\/a><\/li>\n\n\n\n<li><a href=\"https:\/\/www.fda.gov\/drugs\/drug-safety-and-availability\/fda-updates-and-press-announcements-nitrosamine-varenicline-chantix\" target=\"_blank\" rel=\"noreferrer noopener\">https:\/\/www.fda.gov\/drugs\/drug-safety-and-availability\/fda-updates-and-press-announcements-nitrosamine-varenicline-chantix<\/a><\/li>\n\n\n\n<li><a href=\"https:\/\/www.fda.gov\/media\/151470\/download?attachment\" target=\"_blank\" rel=\"noreferrer noopener\">https:\/\/www.fda.gov\/media\/151470\/download?attachment<\/a><\/li>\n<\/ol>\n\n\n\n<div style=\"height:100px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n\n\n        <div class=\"cpb\">\n            <h3 class=\"cpb-heading\">Erw\u00e4hnte Produkte<\/h3>\n            <hr class=\"cpb-heading-underline\" \/>\n            <div class=\"cpb-list\">\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/de.restek.com\/p\/36025\">                                Art.-Nr. 36025                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">Nitrosamin 7 Standard, 100 \u00b5g\/mL in Acetonitril, 1 mL\/Ampulle<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/de.restek.com\/p\/36025\">Produkt anzeigen<\/a>\n                                                    <\/div>\n                    <\/div>\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/de.restek.com\/p\/36026\">                                Art.-Nr. 36026                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">N-Nitrosodimethyl-d6-amin Standard, 1000 \u00b5g\/mL in Acetonitrile, 1 mL\/Ampulle<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/de.restek.com\/p\/36026\">Produkt anzeigen<\/a>\n                                                    <\/div>\n                    <\/div>\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/de.restek.com\/p\/9304A12\">                                Art.-Nr. 9304A12                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">Raptor C18, 2.7 \u00b5m, 100&#215;2.1 mm HPLC-S\u00e4ule<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/de.restek.com\/p\/9304A12\">Produkt anzeigen<\/a>\n                                                    <\/div>\n                    <\/div>\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/de.restek.com\/p\/9309A12\">                                Art.-Nr. 9309A12                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">Raptor Biphenyl, 2.7 \u00b5m, 100&#215;2.1 mm HPLC-S\u00e4ule<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/de.restek.com\/p\/9309A12\">Produkt anzeigen<\/a>\n                                                    <\/div>\n                    <\/div>\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/de.restek.com\/p\/21143\">                                Art.-Nr. 21143                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">Short-Cap-Vial mit graduiertem Markierungsfeld, 9-425 Screw-Thread, 2.0 mL, 9 mm, 12 x 32 (nur Vial), braun, 1000er Pack<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/de.restek.com\/p\/21143\">Produkt anzeigen<\/a>\n                                                    <\/div>\n                    <\/div>\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/de.restek.com\/p\/24498\">                                Art.-Nr. 24498                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">Short Screw Kappen, Polypropylen, Screw-Thread, PTFE\/Silikon\/PTFE Septen, blau, vormontiert, 2.0 mL, 9 mm, 1000er Pack<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/de.restek.com\/p\/24498\">Produkt anzeigen<\/a>\n                                                    <\/div>\n                    <\/div>\n                            <\/div>\n        <\/div>\n        \n\n\n<div style=\"height:100px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n","protected":false},"excerpt":{"rendered":"<p>In this blog, we discuss how to use Restek\u2019s Pro EZLC software to develop and optimize your nitrosamines testing method.<\/p>\n","protected":false},"author":51,"featured_media":0,"comment_status":"closed","ping_status":"open","sticky":false,"template":"","format":"standard","meta":{"_acf_changed":false,"_kad_blocks_custom_css":"","_kad_blocks_head_custom_js":"","_kad_blocks_body_custom_js":"","_kad_blocks_footer_custom_js":"","_kadence_starter_templates_imported_post":false,"_kad_post_transparent":"","_kad_post_title":"","_kad_post_layout":"","_kad_post_sidebar_id":"","_kad_post_content_style":"","_kad_post_vertical_padding":"","_kad_post_feature":"","_kad_post_feature_position":"","_kad_post_header":false,"_kad_post_footer":false,"footnotes":"","_ppma_block_editor_authors":""},"categories":[792],"tags":[],"industries-application":[2668],"post-badge":[],"resource-type":[],"product-library":[],"resource-technique":[],"ppma_author":[592,432],"class_list":["post-51597","post","type-post","status-publish","format-standard","hentry","category-blogs-de","industries-application-semivolatiles"],"acf":[],"taxonomy_info":{"category":[{"value":792,"label":"Blogs"}],"industries-application":[{"value":2668,"label":"Semivolatiles"}]},"featured_image_src_large":false,"author_info":{"display_name":"Jared Burkhart","author_link":"https:\/\/discover.restek.com\/de\/author\/jared-burkhart\/"},"comment_info":0,"category_info":[{"term_id":792,"name":"Blogs","slug":"blogs-de","term_group":0,"term_taxonomy_id":792,"taxonomy":"category","description":"","parent":0,"count":457,"filter":"raw","cat_ID":792,"category_count":457,"category_description":"","cat_name":"Blogs","category_nicename":"blogs-de","category_parent":0}],"tag_info":false,"authors":[{"term_id":592,"user_id":51,"is_guest":0,"slug":"jared-burkhart","display_name":"Jared Burkhart","avatar_url":{"url":"https:\/\/discover.restek.com\/wp-content\/uploads\/people-burkhart-jared.jpg","url2x":"https:\/\/discover.restek.com\/wp-content\/uploads\/people-burkhart-jared.jpg"},"author_category":"1","first_name":"Jared","last_name":"Burkhart","user_url":"https:\/\/www.linkedin.com\/in\/jaredburkhart\/","job_title":"Business Development Manager, Global Sales","description":"Jared Burkhart is the business development manager for life sciences at Restek with a BS in biology and an ASCP (C) certification. He brings 14 years of experience in LC\/MS laboratories supporting both clinical and forensic toxicology, giving him a strong understanding of the analytical and operational needs of these markets. An active member of MATT since 2023 and SOFT since 2025, Jared leverages his technical expertise and industry experience to help laboratories identify solutions that advance their testing capabilities."},{"term_id":432,"user_id":13,"is_guest":0,"slug":"haley-berkland-ms","display_name":"Haley Berkland, MS","avatar_url":{"url":"https:\/\/discover.restek.com\/wp-content\/uploads\/people-berkland-haley.jpg","url2x":"https:\/\/discover.restek.com\/wp-content\/uploads\/people-berkland-haley.jpg"},"author_category":"1","first_name":"Haley","last_name":"Berkland","user_url":"https:\/\/www.linkedin.com\/in\/haley-berkland-60b66517b\/","job_title":"LC Applications Scientist","description":"Haley is an LC applications scientist at Restek. She attended Duquesne University, receiving her bachelor's degree in biochemistry and a master's degree in forensic science and law. As a graduate student, she performed research on the detection of drugs of abuse in vitreous humor by LC-MS\/MS. Before joining Restek in 2023, Haley spent four years working as a forensic toxicologist. While in this role, she performed analysis of postmortem toxicology casework, identification of seized drug evidence, and development\/validation of new assays by LC-MS\/MS, GC-MS, and GC-FID."}],"_links":{"self":[{"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/posts\/51597","targetHints":{"allow":["GET"]}}],"collection":[{"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/posts"}],"about":[{"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/types\/post"}],"author":[{"embeddable":true,"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/users\/51"}],"replies":[{"embeddable":true,"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/comments?post=51597"}],"version-history":[{"count":6,"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/posts\/51597\/revisions"}],"predecessor-version":[{"id":85201,"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/posts\/51597\/revisions\/85201"}],"wp:attachment":[{"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/media?parent=51597"}],"wp:term":[{"taxonomy":"category","embeddable":true,"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/categories?post=51597"},{"taxonomy":"post_tag","embeddable":true,"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/tags?post=51597"},{"taxonomy":"industries-application","embeddable":true,"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/industries-application?post=51597"},{"taxonomy":"post-badge","embeddable":true,"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/post-badge?post=51597"},{"taxonomy":"resource-type","embeddable":true,"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/resource-type?post=51597"},{"taxonomy":"product-library","embeddable":true,"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/product-library?post=51597"},{"taxonomy":"resource-technique","embeddable":true,"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/resource-technique?post=51597"},{"taxonomy":"author","embeddable":true,"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/ppma_author?post=51597"}],"curies":[{"name":"wp","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}