{"id":51735,"date":"2012-12-03T13:30:00","date_gmt":"2012-12-03T13:30:00","guid":{"rendered":"https:\/\/discover.restek.com\/uncategorized\/light-hydrocarbon-analysis-in-water-samples-not-just-a-headspace-method-anymore\/"},"modified":"2026-01-28T15:56:09","modified_gmt":"2026-01-28T15:56:09","slug":"light-hydrocarbon-analysis-in-water-samples-not-just-a-headspace-method-anymore","status":"publish","type":"post","link":"https:\/\/discover.restek.com\/de\/blogs\/gnbl4728\/light-hydrocarbon-analysis-in-water-samples-not-just-a-headspace-method-anymore","title":{"rendered":"Light hydrocarbon analysis in water samples \u2013 not just a headspace method anymore"},"content":{"rendered":"\n<p class=\"wp-block-paragraph\">Living in Pennsylvania during this natural gas boom stirs passionate, mixed emotions about the economic benefits verses the environmental costs of drilling for this resource.&nbsp;<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">When the\u00a0<a href=\"https:\/\/en.wikipedia.org\/wiki\/Marcellus_Formation\" target=\"_blank\" rel=\"noreferrer noopener\">Marcellus Shale<\/a>\u00a0natural gas drilling in Pennsylvania really started taking off several years ago, we (tech service) were taking quite a few calls about analyzing C1 through C4 (methane through butane) in water samples. Many of our customers had tried this analysis using their purge &amp; trap (p&amp;t) instruments without success, because the only available traps on the market at that time were not capable of trapping methane. As a result, the only two ways I knew how to do this analysis were to use either headspace or liquid injections. Several years later, Teledyne Tekmar published an application note using p&amp;t titled\u00a0<a href=\"https:\/\/www.epa.gov\/sites\/default\/files\/documents\/valentine.pdf\" target=\"_blank\" rel=\"noreferrer noopener\">Alternative Methods to RSK 175 Using Purge and Trap Concentration and Automated Headspace for the Analysis of Dissolved Gases in Drinking Water<\/a>.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\"><strong>Now, for the first time, I have seen a published method using purge &amp; trap (PA-DEP 9243, link below).<\/strong><\/p>\n\n\n\n<p class=\"wp-block-paragraph\"><strong><a href=\"http:\/\/files.dep.state.pa.us\/AboutDEP\/Labs\/LabsPortalFiles\/PA-DEP_9243_Rev_0.pdf\" target=\"_blank\" rel=\"noreferrer noopener\">Light Hydrocarbons in Aqueous Samples via Purge and Trap<\/a><\/strong><\/p>\n\n\n\n<p class=\"wp-block-paragraph\">So, how well does this method work? I have no idea, but if someone out there has tried it, I would like to hear back from you.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">(Note &#8211; I did find a typo in the method; the Restek U-Bond PLOT column they mentioned should be listed with a 20\u00b5mdf film-thickness (not 0.2\u00b5mdf). The Restek catalog number for this column is&nbsp;19749. The catalog number for the ShinCarbon column is&nbsp;19809.)<\/p>\n\n\n<div class=\"wp-block-image\">\n<figure class=\"aligncenter size-full\"><img loading=\"lazy\" decoding=\"async\" width=\"1005\" height=\"550\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-light-hydrocarbon-analysis-in-water-samples-not-just-a-headspace-method-anymore-01.bmp\" alt=\"\" class=\"wp-image-15646\" title=\"-\" srcset=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-light-hydrocarbon-analysis-in-water-samples-not-just-a-headspace-method-anymore-01.bmp 1005w, https:\/\/discover.restek.com\/wp-content\/uploads\/blog-light-hydrocarbon-analysis-in-water-samples-not-just-a-headspace-method-anymore-01-300x164.jpg 300w, https:\/\/discover.restek.com\/wp-content\/uploads\/blog-light-hydrocarbon-analysis-in-water-samples-not-just-a-headspace-method-anymore-01-768x420.jpg 768w\" sizes=\"auto, (max-width: 1005px) 100vw, 1005px\" \/><\/figure>\n<\/div>\n\n\n<p class=\"wp-block-paragraph\"><em>Chromatogram above from Teledyne Tekmar using the Restek U-Bond.<\/em><\/p>\n\n\n<div class=\"wp-block-custom-chromatogram-article\"><div class=\"wp-block-custom-chromatogram-article\"><h3 class=\"article-title\">Natural Gas ShinCarbon ST (Micropacked)<\/h3><div class=\"chromatogram-image regular-image\"><img decoding=\"async\" src=\"https:\/\/ez.restek.com\/images\/cgram\/gc_pc00672.png\" alt=\"Natural Gas ShinCarbon ST (Micropacked)\" title=\"-\"><\/div><p class=\"article-id\" style=\"text-align:center\"> GC_PC00672<\/p><div class=\"chromatogram-peaks\"><h4>Peaks<\/h4><table class=\"peaks col-lg-4 col-12\">\n<thead><tr><th><\/th><th style=\"text-align: left;width: 75px\">Peaks<\/th><\/tr><\/thead>\n<tbody><tr><td class=\"num\">1.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for C1\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/74-82-8\/C1\" target=\"_blank\" rel=\"noopener\">C1<\/a><\/td><\/tr>\n<tr><td class=\"num\">2.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for C2\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/74-84-0\/C2\" target=\"_blank\" rel=\"noopener\">C2<\/a><\/td><\/tr>\n<tr><td class=\"num\">3.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for C3\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/74-98-6\/C3\" target=\"_blank\" rel=\"noopener\">C3<\/a><\/td><\/tr>\n<tr><td class=\"num\">4.<\/td><td class=\"cmpd\">iC4-isobutane<\/td><\/tr>\n<tr><td class=\"num\">5.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for C4\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/106-97-8\/C4\" target=\"_blank\" rel=\"noopener\">C4<\/a><\/td><\/tr>\n<tr><td class=\"num\">6.<\/td><td class=\"cmpd\">iC5-isopentane<\/td><\/tr>\n<tr><td class=\"num\">7.<\/td><td class=\"cmpd\">nC5<\/td><\/tr>\n<tr><td class=\"num\">8.<\/td><td class=\"cmpd\">Hexanes<\/td><\/tr>\n<\/tbody><\/table><\/div><div class=\"chromatogram-conditions\"><h4>Conditions<\/h4><div class=\"conditions-container container-fluid\"><div class=\"row\"><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Column<\/th><td>100\/120 mesh on ShinCarbon ST, SilcoSmooth tubing, 1 m, 1\/16 OD, 1.0 mm ID (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/19809\" rel=\"noopener\">cat.# 19809<\/a>)<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Standard\/Sample<\/th><td>C1-C6 hydrocarbons<\/td><\/tr>\n<tr><th class=\"sub conditions_header\" scope=\"row\">Conc.:<\/th><td>  0.5-8% varied<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Injection<\/th><td>packed on-column<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Inj. Temp.:<\/th><td>200 \u00b0C<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\" colspan=\"2\">Oven<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Oven Temp.:<\/th><td>125 &#176;C to 250 &#176;C at 16 &#176;C\/min (hold 25.2 min)<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Carrier Gas<\/th><td>He, constant flow<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Flow Rate:<\/th><td>10 mL\/min<\/td><\/tr><\/table><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Detector<\/th><td>FID @ 250 \u00b0C<\/td><\/tr><\/table><\/div><\/div><\/div><div class=\"chromatogram-related-skus\"><h4>Related SKUs<\/h4><ul><li>19809<\/li><\/ul><\/div><div class=\"chromatogram-related-searches\"><h4>Related Searches<\/h4><ul><li>c1-c6<\/li><li>hydrocarbon gases<\/li><li>micropacked<\/li><li>natural gas<\/li><li>packed column<\/li><li>shincarbon<\/li><li>shincarbon ST<\/li><\/ul><\/div><div class=\"chromatogram-pdf-link\"><a href=\"https:\/\/ez.restek.com\/images\/cgram\/gc_pc00672.pdf\" target=\"_blank\" rel=\"noopener noreferrer\"><svg xmlns=\"http:\/\/www.w3.org\/2000\/svg\" width=\"18\" height=\"18\" viewBox=\"0 0 18 18\"><g data-name=\"Group 2996\"><path data-name=\"Rectangle 1246\" d=\"M0 0h18v18H0z\" style=\"fill: none;\"><\/path><\/g><g data-name=\"Group 2997\"><path data-name=\"Path 729\" d=\"M13.412 11.4v2.017H5.345V11.4H4v2.017a1.349 1.349 0 0 0 1.345 1.345h8.068a1.349 1.349 0 0 0 1.345-1.345V11.4zm-.672-2.694-.948-.948-1.741 1.735V4H8.706v5.493L6.965 7.758l-.948.948 3.361 3.361z\" transform=\"translate(-.437 -.414)\" style=\"fill: rgb(13, 123, 196);\"><\/path><\/g><\/svg>Download PDF<\/a><\/div><\/div><\/div>\n\n\n<p class=\"wp-block-paragraph\">As a final note, I would like to thank co-workers Chris English for a previous post which laid the groundwork for this post, and Chris Rattray for contacting Teledyne Tekmar to confirm that the trap used in this method was their proprietary trap P\/N: 15-0885-403.<\/p>\n\n\n\n<div style=\"height:100px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n\n\n        <div class=\"cpb\">\n            <h3 class=\"cpb-heading\">Erw\u00e4hnte Produkte<\/h3>\n            <hr class=\"cpb-heading-underline\" \/>\n            <div class=\"cpb-list\">\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/de.restek.com\/p\/19749\">                                Art.-Nr. 19749                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">Rt-U-BOND PLOT S\u00e4ule 15m, 0.53 mm, ID, 20 \u00b5m<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/de.restek.com\/p\/19749\">Produkt anzeigen<\/a>\n                                                    <\/div>\n                    <\/div>\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/de.restek.com\/p\/19809\">                                Art.-Nr. 19809                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">Mikrogepackte GC S\u00e4ule ShinCarbon ST, 100\/120, 1 m, 1 mm ID, 1\/16&quot; AD, Silco<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/de.restek.com\/p\/19809\">Produkt anzeigen<\/a>\n                                                    <\/div>\n                    <\/div>\n                            <\/div>\n        <\/div>\n        \n\n\n<div style=\"height:100px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n","protected":false},"excerpt":{"rendered":"<p>Living in Pennsylvania during this natural gas boom stirs passionate, mixed emotions about the economic benefits verses the environmental costs of drilling for this resource. While it is tempting to provide you all my thoughts on the subject, I&rsquo;ll keep&#8230;<\/p>\n","protected":false},"author":2,"featured_media":0,"comment_status":"closed","ping_status":"open","sticky":false,"template":"","format":"standard","meta":{"_acf_changed":false,"_kad_blocks_custom_css":"","_kad_blocks_head_custom_js":"","_kad_blocks_body_custom_js":"","_kad_blocks_footer_custom_js":"","_kadence_starter_templates_imported_post":false,"_kad_post_transparent":"","_kad_post_title":"","_kad_post_layout":"","_kad_post_sidebar_id":"","_kad_post_content_style":"","_kad_post_vertical_padding":"","_kad_post_feature":"","_kad_post_feature_position":"","_kad_post_header":false,"_kad_post_footer":false,"footnotes":"","_ppma_block_editor_authors":""},"categories":[9],"tags":[],"industries-application":[2204],"post-badge":[],"resource-type":[],"product-library":[],"resource-technique":[2316,2358],"ppma_author":[582],"class_list":["post-51735","post","type-post","status-publish","format-standard","hentry","category-blogs","industries-application-energie","resource-technique-chromatographie-en-phase-gazeuse-gc","resource-technique-headspace-fr"],"acf":[],"taxonomy_info":{"category":[{"value":9,"label":"Blogs"}],"industries-application":[{"value":2204,"label":"Energie"}],"resource-technique":[{"value":2316,"label":"Chromatographie en phase gazeuse (GC)"},{"value":2358,"label":"Headspace"}]},"featured_image_src_large":false,"author_info":{"display_name":"Patrick Gallagher","author_link":"https:\/\/discover.restek.com\/de\/author\/patrick-gallagherrestek-com\/"},"comment_info":0,"category_info":[{"term_id":9,"name":"Blogs","slug":"blogs","term_group":0,"term_taxonomy_id":9,"taxonomy":"category","description":"","parent":0,"count":457,"filter":"raw","cat_ID":9,"category_count":457,"category_description":"","cat_name":"Blogs","category_nicename":"blogs","category_parent":0}],"tag_info":false,"authors":[{"term_id":582,"user_id":0,"is_guest":1,"slug":"alan-sensue","display_name":"Alan Sensue","avatar_url":"https:\/\/secure.gravatar.com\/avatar\/?s=96&d=mm&r=g","author_category":"1","first_name":"Alan","last_name":"Sensue","user_url":"","job_title":"","description":""}],"_links":{"self":[{"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/posts\/51735","targetHints":{"allow":["GET"]}}],"collection":[{"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/posts"}],"about":[{"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/types\/post"}],"author":[{"embeddable":true,"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/users\/2"}],"replies":[{"embeddable":true,"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/comments?post=51735"}],"version-history":[{"count":3,"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/posts\/51735\/revisions"}],"predecessor-version":[{"id":85380,"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/posts\/51735\/revisions\/85380"}],"wp:attachment":[{"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/media?parent=51735"}],"wp:term":[{"taxonomy":"category","embeddable":true,"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/categories?post=51735"},{"taxonomy":"post_tag","embeddable":true,"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/tags?post=51735"},{"taxonomy":"industries-application","embeddable":true,"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/industries-application?post=51735"},{"taxonomy":"post-badge","embeddable":true,"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/post-badge?post=51735"},{"taxonomy":"resource-type","embeddable":true,"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/resource-type?post=51735"},{"taxonomy":"product-library","embeddable":true,"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/product-library?post=51735"},{"taxonomy":"resource-technique","embeddable":true,"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/resource-technique?post=51735"},{"taxonomy":"author","embeddable":true,"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/ppma_author?post=51735"}],"curies":[{"name":"wp","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}