{"id":56953,"date":"2025-07-05T17:59:00","date_gmt":"2025-07-05T17:59:00","guid":{"rendered":"https:\/\/discover.restek.com\/?p=56953"},"modified":"2026-10-01T17:58:24","modified_gmt":"2026-10-01T17:58:24","slug":"analyze-picogram-level-semivolatiles-accurately-with-gc-ms-ms-and-epa-method-8270e","status":"publish","type":"post","link":"https:\/\/discover.restek.com\/de\/articles\/evfa4492\/analyze-picogram-level-semivolatiles-accurately-with-gc-ms-ms-and-epa-method-8270e","title":{"rendered":"Analyze Picogram-Level Semivolatiles Accurately with GC-MS\/MS and EPA Method 8270E"},"content":{"rendered":"\n<p class=\"wp-block-paragraph\">EPA Method 8270E now allows for the use of GC tandem mass spectrometry (GC-MS\/MS), which can extend detection levels for semivolatiles down into the low picogram range. Previous versions of the method only allowed GC-MS analysis, which is much less sensitive. Much of the method has remained the same, but the lower levels of detection provide new challenges, including a dynamic calibration range and retention time shifts for reactive compounds. The calibration range is limited by lower levels of detection at the lower end as well as by overloading and MS saturation at the higher end. Using an Rxi-SVOCms column (30 m, 0.25 mm ID, 0.25 \u00b5m [cat.#\u00a016623]) and the conditions shown in Figure 1, the upper limit was found to be 0.5 ng on-column. This shifts the typical range of calibration (0.1-100 \u00b5g\/mL) to concentrations 100x to 1000x lower. Note that a wider range could be obtained by adjusting the split ratio.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\"><strong>Figure 1:<\/strong>\u00a0The main chromatogram is an analysis of semivolatiles at 100 pg on an Rxi-SVOCms column. A calibration curve was also performed, and the inset figures represent the lowest calibration point for selected compounds.<\/p>\n\n\n<div class=\"wp-block-custom-chromatogram-article\"><div class=\"wp-block-custom-chromatogram-article\"><div class=\"chromatogram-image wide-image\"><img decoding=\"async\" src=\"https:\/\/ez.restek.com\/images\/cgram\/gc_ev1522.png\" alt=\"Semivolatiles on Rxi-SVOCms by GC-MS\/MS\" title=\"-\"><\/div><p class=\"article-id\" style=\"text-align:center\"> GC_EV1522<\/p><div class=\"chromatogram-peaks\"><h4>Peaks<\/h4><table class=\"peaks col-lg-6 col-12 peak-50\">\n<thead><tr><th><\/th><th style=\"text-align: left;width: 75px\">Peaks<\/th><th style=\"text-align: center;width: 75px\">t<sub>R<\/sub> (min)<\/th><th style=\"text-align: center;width: 75px\">Conc.<br \/>(\u00b5g\/g)<\/th><th style=\"text-align: center;width: 75px\">SRM 1<\/th><th style=\"text-align: center;width: 75px\">SRM 2<\/th><th style=\"text-align: center;width: 75px\">SRM 3<\/th><\/tr><\/thead>\n<tbody><tr><td class=\"num\">1.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 1,4-Dioxane\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/123-91-1\/1,4-Dioxane\" target=\"_blank\" rel=\"noopener\">1,4-Dioxane<\/a><\/td><td class=\"oth\">2.79<\/td><td class=\"oth\">1<\/td><td class=\"oth\">88 to 28 @14V<\/td><td class=\"oth\">58 to 28 @6V<\/td><td class=\"oth\">88 to 57 @8V<\/td><\/tr>\n<tr><td class=\"num\">2.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for N-Nitrosodimethylamine\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/62-75-9\/N-Nitrosodimethylamine\" target=\"_blank\" rel=\"noopener\">N-Nitrosodimethylamine<\/a><\/td><td class=\"oth\">3.08<\/td><td class=\"oth\">1<\/td><td class=\"oth\">74 to 42 @16V<\/td><td class=\"oth\">74 to 43 @12V<\/td><td class=\"oth\">74 to 44 @6V<\/td><\/tr>\n<tr><td class=\"num\">3.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Pyridine\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/110-86-1\/Pyridine\" target=\"_blank\" rel=\"noopener\">Pyridine<\/a><\/td><td class=\"oth\">3.13<\/td><td class=\"oth\">1<\/td><td class=\"oth\">79 to 51 @26V<\/td><td class=\"oth\">52 to 26 @18V<\/td><td class=\"oth\">79 to 52 @12V<\/td><\/tr>\n<tr><td class=\"num\">4.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Ethyl methacrylate\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/97-63-2\/Ethyl methacrylate\" target=\"_blank\" rel=\"noopener\">Ethyl methacrylate<\/a><\/td><td class=\"oth\">3.62<\/td><td class=\"oth\">1<\/td><td class=\"oth\">99 to 43 @14V<\/td><td class=\"oth\">69 to 41 @6V<\/td><td class=\"oth\">99 to 53 @12V<\/td><\/tr>\n<tr><td class=\"num\">5.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 2-Picoline\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/109-06-8\/2-Picoline\" target=\"_blank\" rel=\"noopener\">2-Picoline<\/a><\/td><td class=\"oth\">3.96<\/td><td class=\"oth\">1<\/td><td class=\"oth\">93 to 65 @18V<\/td><td class=\"oth\">66 to 39 @18V<\/td><td class=\"oth\">93 to 66 @12V<\/td><\/tr>\n<tr><td class=\"num\">6.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for N-Nitrosomethylethylamine\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/10595-95-6\/N-Nitrosomethylethylamine\" target=\"_blank\" rel=\"noopener\">N-Nitrosomethylethylamine<\/a><\/td><td class=\"oth\">4.12<\/td><td class=\"oth\">1<\/td><td class=\"oth\">88 to 42 @16V<\/td><td class=\"oth\">88 to 43 @8V<\/td><td class=\"oth\">88 to 71 @6V<\/td><\/tr>\n<tr><td class=\"num\">7.<\/td><td class=\"cmpd\">Acrylamine<\/td><td class=\"oth\">4.5<\/td><td class=\"oth\">1<\/td><td class=\"oth\">71 to 44 @22V<\/td><td class=\"oth\">55 to 27 @6V<\/td><td class=\"oth\">71 to 55 @6V<\/td><\/tr>\n<tr><td class=\"num\">8.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Methyl methanesulfonate\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/66-27-3\/Methyl methanesulfonate\" target=\"_blank\" rel=\"noopener\">Methyl methanesulfonate<\/a><\/td><td class=\"oth\">4.52<\/td><td class=\"oth\">1<\/td><td class=\"oth\">80 to 48 @28V<\/td><td class=\"oth\">65 to 48 @18V<\/td><td class=\"oth\">80 to 65 @8V<\/td><\/tr>\n<tr><td class=\"num\">9.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 2-Fluorophenol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/367-12-4\/2-Fluorophenol\" target=\"_blank\" rel=\"noopener\">2-Fluorophenol<\/a><\/td><td class=\"oth\">4.73<\/td><td class=\"oth\">1<\/td><td class=\"oth\">112 to 63 @22V<\/td><td class=\"oth\">92 to 63 @14V<\/td><td class=\"oth\">112 to 92 @6V<\/td><\/tr>\n<tr><td class=\"num\">10.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for N-Nitrosodiethylamine\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/55-18-5\/N-Nitrosodiethylamine\" target=\"_blank\" rel=\"noopener\">N-Nitrosodiethylamine<\/a><\/td><td class=\"oth\">5.05<\/td><td class=\"oth\">1<\/td><td class=\"oth\">102 to 44 @10V<\/td><td class=\"oth\">102 to 56 @12V<\/td><td class=\"oth\">102 to 85 @6V<\/td><\/tr>\n<tr><td class=\"num\">11.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Ethyl methanesulfonate\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/62-50-0\/Ethyl methanesulfonate\" target=\"_blank\" rel=\"noopener\">Ethyl methanesulfonate<\/a><\/td><td class=\"oth\">5.51<\/td><td class=\"oth\">1<\/td><td class=\"oth\">109 to 45 @10V<\/td><td class=\"oth\">79 to 48 @34V<\/td><td class=\"oth\">109 to 79 @6V<\/td><\/tr>\n<tr><td class=\"num\">12.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Benzaldehyde\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/100-52-7\/Benzaldehyde\" target=\"_blank\" rel=\"noopener\">Benzaldehyde<\/a><\/td><td class=\"oth\">6.01<\/td><td class=\"oth\">1<\/td><td class=\"oth\">105 to 51 @26V<\/td><td class=\"oth\">77 to 51 @14V<\/td><td class=\"oth\">105 to 77 @10V<\/td><\/tr>\n<tr><td class=\"num\">13.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Phenol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/108-95-2\/Phenol\" target=\"_blank\" rel=\"noopener\">Phenol<\/a><\/td><td class=\"oth\">6.13<\/td><td class=\"oth\">1<\/td><td class=\"oth\">94 to 39 @30V<\/td><td class=\"oth\">94 to 65 @16V<\/td><td class=\"oth\">94 to 66 @10V<\/td><\/tr>\n<tr><td class=\"num\">14.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Phenol-d6\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/13127-88-3\/Phenol-d6\" target=\"_blank\" rel=\"noopener\">Phenol-d6<\/a><\/td><td class=\"oth\">6.14<\/td><td class=\"oth\">1<\/td><td class=\"oth\">99 to 69 @18V<\/td><td class=\"oth\">71 to 42 @16V<\/td><td class=\"oth\">99 to 71 @8V<\/td><\/tr>\n<tr><td class=\"num\">15.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Aniline\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/62-53-3\/Aniline\" target=\"_blank\" rel=\"noopener\">Aniline<\/a><\/td><td class=\"oth\">6.17<\/td><td class=\"oth\">1<\/td><td class=\"oth\">93 to 65 @20V<\/td><td class=\"oth\">66 to 39 @18V<\/td><td class=\"oth\">93 to 66 @10V<\/td><\/tr>\n<tr><td class=\"num\">16.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Pentachloroethane\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/76-01-7\/Pentachloroethane\" target=\"_blank\" rel=\"noopener\">Pentachloroethane<\/a><\/td><td class=\"oth\">6.26<\/td><td class=\"oth\">1<\/td><td class=\"oth\">167 to 95 @32V<\/td><td class=\"oth\">117 to 82 @30V<\/td><td class=\"oth\">167 to 132 @14V<\/td><\/tr>\n<tr><td class=\"num\">17.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for bis-(2-Chloroethyl)ether\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/111-44-4\/bis-(2-Chloroethyl)ether\" target=\"_blank\" rel=\"noopener\">bis-(2-Chloroethyl)ether<\/a><\/td><td class=\"oth\">6.28<\/td><td class=\"oth\">1<\/td><td class=\"oth\">93 to 27 @22V<\/td><td class=\"oth\">63 to 27 @10V<\/td><td class=\"oth\">93 to 63 @6V<\/td><\/tr>\n<tr><td class=\"num\">18.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 2-Chlorophenol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/95-57-8\/2-Chlorophenol\" target=\"_blank\" rel=\"noopener\">2-Chlorophenol<\/a><\/td><td class=\"oth\">6.35<\/td><td class=\"oth\">1<\/td><td class=\"oth\">128 to 63 @24V<\/td><td class=\"oth\">64 to 38 @12V<\/td><td class=\"oth\">128 to 64 @14V<\/td><\/tr>\n<tr><td class=\"num\">19.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Decane\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/124-18-5\/Decane\" target=\"_blank\" rel=\"noopener\">Decane<\/a><\/td><td class=\"oth\">6.49<\/td><td class=\"oth\">1<\/td><td class=\"oth\">85 to 41 @12V<\/td><td class=\"oth\">71 to 43 @6V<\/td><td class=\"oth\">85 to 43 @6V<\/td><\/tr>\n<tr><td class=\"num\">20.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 1,3-Dichlorobenzene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/541-73-1\/1,3-Dichlorobenzene\" target=\"_blank\" rel=\"noopener\">1,3-Dichlorobenzene<\/a><\/td><td class=\"oth\">6.61<\/td><td class=\"oth\">1<\/td><td class=\"oth\">146 to 75 @24V<\/td><td class=\"oth\">111 to 75 @12V<\/td><td class=\"oth\">146 to 111 @12V<\/td><\/tr>\n<tr><td class=\"num\">21.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 1,4-Dichlorobenzene-d4\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/3855-82-1\/1,4-Dichlorobenzene-d4\" target=\"_blank\" rel=\"noopener\">1,4-Dichlorobenzene-d4<\/a><\/td><td class=\"oth\">6.71<\/td><td class=\"oth\">1<\/td><td class=\"oth\">150 to 76 @38V<\/td><td class=\"oth\">150 to 78 @26V<\/td><td class=\"oth\">150 to 115 @14V<\/td><\/tr>\n<tr><td class=\"num\">22.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 1,4-Dichlorobenzene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/106-46-7\/1,4-Dichlorobenzene\" target=\"_blank\" rel=\"noopener\">1,4-Dichlorobenzene<\/a><\/td><td class=\"oth\">6.74<\/td><td class=\"oth\">1<\/td><td class=\"oth\">146 to 75 @24V<\/td><td class=\"oth\">111 to 75 @12V<\/td><td class=\"oth\">146 to 111 @14V<\/td><\/tr>\n<tr><td class=\"num\">23.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Benzyl alcohol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/100-51-6\/Benzyl alcohol\" target=\"_blank\" rel=\"noopener\">Benzyl alcohol<\/a><\/td><td class=\"oth\">6.94<\/td><td class=\"oth\">1<\/td><td class=\"oth\">108 to 77 @22V<\/td><td class=\"oth\">79 to 51 @20V<\/td><td class=\"oth\">108 to 79 @12V<\/td><\/tr>\n<tr><td class=\"num\">24.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 1,2-Dichlorobenzene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/95-50-1\/1,2-Dichlorobenzene\" target=\"_blank\" rel=\"noopener\">1,2-Dichlorobenzene<\/a><\/td><td class=\"oth\">6.98<\/td><td class=\"oth\">1<\/td><td class=\"oth\">146 to 111 @12V<\/td><td class=\"oth\">111 to 50 @36V<\/td><td class=\"oth\">111 to 75 @10V<\/td><\/tr>\n<tr><td class=\"num\">25.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Indene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/95-13-6\/Indene\" target=\"_blank\" rel=\"noopener\">Indene<\/a><\/td><td class=\"oth\">7.12<\/td><td class=\"oth\">1<\/td><td class=\"oth\">116 to 89 @28V<\/td><td class=\"oth\">115 to 63 @26V<\/td><td class=\"oth\">115 to 89 @16V<\/td><\/tr>\n<tr><td class=\"num\">26.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 2-Methylphenol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/95-48-7\/2-Methylphenol\" target=\"_blank\" rel=\"noopener\">2-Methylphenol<\/a><\/td><td class=\"oth\">7.14<\/td><td class=\"oth\">1<\/td><td class=\"oth\">108 to 77 @24V<\/td><td class=\"oth\">79 to 51 @20V<\/td><td class=\"oth\">108 to 80 @8V<\/td><\/tr>\n<tr><td class=\"num\">27.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 2,2&amp;apos;-oxybis(1-chloropropane)\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/108-60-1\/2,2&amp;apos;-oxybis(1-chloropropane)\" target=\"_blank\" rel=\"noopener\">2,2&#8242;-oxybis(1-chloropropane)<\/a><\/td><td class=\"oth\">7.2<\/td><td class=\"oth\">1<\/td><td class=\"oth\">121 to 45 @6V<\/td><td class=\"oth\">77 to 51 @12V<\/td><td class=\"oth\">121 to 77 @8V<\/td><\/tr>\n<tr><td class=\"num\">28.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for N-Nitrosopyrrolidine\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/930-55-2\/N-Nitrosopyrrolidine\" target=\"_blank\" rel=\"noopener\">N-Nitrosopyrrolidine<\/a><\/td><td class=\"oth\">7.34<\/td><td class=\"oth\">1<\/td><td class=\"oth\">100 to 43 @8V<\/td><td class=\"oth\">100 to 55 @6V<\/td><td class=\"oth\">100 to 68 @8V<\/td><\/tr>\n<tr><td class=\"num\">29.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Acetophenone\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/98-86-2\/Acetophenone\" target=\"_blank\" rel=\"noopener\">Acetophenone<\/a><\/td><td class=\"oth\">7.37<\/td><td class=\"oth\">1<\/td><td class=\"oth\">105 to 51 @26V<\/td><td class=\"oth\">77 to 51 @14V<\/td><td class=\"oth\">105 to 77 @12V<\/td><\/tr>\n<tr><td class=\"num\">30.<\/td><td class=\"cmpd\">3- and 4-Methylphenol<\/td><td class=\"oth\">7.41<\/td><td class=\"oth\">1<\/td><td class=\"oth\">107 to 51 @26V<\/td><td class=\"oth\">70 to 43 @6V<\/td><td class=\"oth\">107 to 77 @12V<\/td><\/tr>\n<tr><td class=\"num\">31.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for N-Nitrosomorpholine\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/59-89-2\/N-Nitrosomorpholine\" target=\"_blank\" rel=\"noopener\">N-Nitrosomorpholine<\/a><\/td><td class=\"oth\">7.42<\/td><td class=\"oth\">1<\/td><td class=\"oth\">116 to 56 @10V<\/td><td class=\"oth\">86 to 56 @6V<\/td><td class=\"oth\">116 to 86 @6V<\/td><\/tr>\n<tr><td class=\"num\">32.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for o-Toluidine\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/95-53-4\/o-Toluidine\" target=\"_blank\" rel=\"noopener\">o-Toluidine<\/a><\/td><td class=\"oth\">7.44<\/td><td class=\"oth\">1<\/td><td class=\"oth\">107 to 77 @26V<\/td><td class=\"oth\">106 to 77 @16V<\/td><td class=\"oth\">106 to 79 @8V<\/td><\/tr>\n<tr><td class=\"num\">33.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Hexachloroethane\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/67-72-1\/Hexachloroethane\" target=\"_blank\" rel=\"noopener\">Hexachloroethane<\/a><\/td><td class=\"oth\">7.56<\/td><td class=\"oth\">1<\/td><td class=\"oth\">201 to 131 @30V<\/td><td class=\"oth\">117 to 82 @24V<\/td><td class=\"oth\">201 to 166 @12V<\/td><\/tr>\n<tr><td class=\"num\">34.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Nitrobenzene-d5\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/4165-60-0\/Nitrobenzene-d5\" target=\"_blank\" rel=\"noopener\">Nitrobenzene-d5<\/a><\/td><td class=\"oth\">7.63<\/td><td class=\"oth\">1<\/td><td class=\"oth\">128 to 54 @28V<\/td><td class=\"oth\">82 to 54 @12V<\/td><td class=\"oth\">128 to 82 @12V<\/td><\/tr>\n<tr><td class=\"num\">35.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Nitrobenzene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/98-95-3\/Nitrobenzene\" target=\"_blank\" rel=\"noopener\">Nitrobenzene<\/a><\/td><td class=\"oth\">7.68<\/td><td class=\"oth\">1<\/td><td class=\"oth\">123 to 51 @30V<\/td><td class=\"oth\">77 to 51 @12V<\/td><td class=\"oth\">123 to 77 @10V<\/td><\/tr>\n<tr><td class=\"num\">36.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for N-Nitrosopiperidine\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/100-75-4\/N-Nitrosopiperidine\" target=\"_blank\" rel=\"noopener\">N-Nitrosopiperidine<\/a><\/td><td class=\"oth\">7.94<\/td><td class=\"oth\">1<\/td><td class=\"oth\">114 to 41 @10V<\/td><td class=\"oth\">114 to 84 @6V<\/td><td class=\"oth\">114 to 97 @6V<\/td><\/tr>\n<tr><td class=\"num\">37.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Isophorone\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/78-59-1\/Isophorone\" target=\"_blank\" rel=\"noopener\">Isophorone<\/a><\/td><td class=\"oth\">8.12<\/td><td class=\"oth\">1<\/td><td class=\"oth\">138 to 82 @8V<\/td><td class=\"oth\">82 to 39 @12V<\/td><td class=\"oth\">82 to 54 @6V<\/td><\/tr>\n<tr><td class=\"num\">38.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 2-Nitrophenol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/88-75-5\/2-Nitrophenol\" target=\"_blank\" rel=\"noopener\">2-Nitrophenol<\/a><\/td><td class=\"oth\">8.25<\/td><td class=\"oth\">1<\/td><td class=\"oth\">139 to 81 @12V<\/td><td class=\"oth\">109 to 81 @8V<\/td><td class=\"oth\">139 to 109 @8V<\/td><\/tr>\n<tr><td class=\"num\">39.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 2,4-Dimethylphenol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/105-67-9\/2,4-Dimethylphenol\" target=\"_blank\" rel=\"noopener\">2,4-Dimethylphenol<\/a><\/td><td class=\"oth\">8.36<\/td><td class=\"oth\">1<\/td><td class=\"oth\">122 to 77 @26V<\/td><td class=\"oth\">107 to 77 @14V<\/td><td class=\"oth\">122 to 107 @12V<\/td><\/tr>\n<tr><td class=\"num\">40.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for O,O,O-Triethyl phosphorothioate\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/126-68-1\/O,O,O-Triethyl phosphorothioate\" target=\"_blank\" rel=\"noopener\">O,O,O-Triethyl phosphorothioate<\/a><\/td><td class=\"oth\">8.51<\/td><td class=\"oth\">1<\/td><td class=\"oth\">198 to 114 @12V<\/td><td class=\"oth\">121 to 65 @10V<\/td><td class=\"oth\">198 to 121 @8V<\/td><\/tr>\n<tr><td class=\"num\">41.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Benzoic acid\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/65-85-0\/Benzoic acid\" target=\"_blank\" rel=\"noopener\">Benzoic acid<\/a><\/td><td class=\"oth\">8.53<\/td><td class=\"oth\">1<\/td><td class=\"oth\">122 to 77 @20V<\/td><td class=\"oth\">105 to 77 @10V<\/td><td class=\"oth\">122 to 105 @8V<\/td><\/tr>\n<tr><td class=\"num\">42.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Bis(2-chloroethoxy)methane\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/111-91-1\/Bis(2-chloroethoxy)methane\" target=\"_blank\" rel=\"noopener\">Bis(2-chloroethoxy)methane<\/a><\/td><td class=\"oth\">8.55<\/td><td class=\"oth\">1<\/td><td class=\"oth\">93 to 63 @6V<\/td><td class=\"oth\">63 to 26 @36V<\/td><td class=\"oth\">63 to 27 @12V<\/td><\/tr>\n<tr><td class=\"num\">43.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Phentermine\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/122-09-8\/Phentermine\" target=\"_blank\" rel=\"noopener\">Phentermine<\/a><\/td><td class=\"oth\">8.64<\/td><td class=\"oth\">1<\/td><td class=\"oth\">134 to 91 @20V<\/td><td class=\"oth\">58 to 42 @22V<\/td><td class=\"oth\">134 to 115 @12V<\/td><\/tr>\n<tr><td class=\"num\">44.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 2,4-Dichlorophenol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/120-83-2\/2,4-Dichlorophenol\" target=\"_blank\" rel=\"noopener\">2,4-Dichlorophenol<\/a><\/td><td class=\"oth\">8.68<\/td><td class=\"oth\">1<\/td><td class=\"oth\">162 to 63 @24V<\/td><td class=\"oth\">98 to 63 @8V<\/td><td class=\"oth\">162 to 98 @12V<\/td><\/tr>\n<tr><td class=\"num\">45.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 1,2,4-Trichlorobenzene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/120-82-1\/1,2,4-Trichlorobenzene\" target=\"_blank\" rel=\"noopener\">1,2,4-Trichlorobenzene<\/a><\/td><td class=\"oth\">8.84<\/td><td class=\"oth\">1<\/td><td class=\"oth\">180 to 109 @24V<\/td><td class=\"oth\">145 to 109 @10V<\/td><td class=\"oth\">180 to 145 @12V<\/td><\/tr>\n<tr><td class=\"num\">46.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Naphthalene-d8\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/1146-65-2\/Naphthalene-d8\" target=\"_blank\" rel=\"noopener\">Naphthalene-d8<\/a><\/td><td class=\"oth\">8.91<\/td><td class=\"oth\">1<\/td><td class=\"oth\">136 to 84 @20V<\/td><td class=\"oth\">136 to 108 @18V<\/td><td class=\"oth\">136 to 134 @14V<\/td><\/tr>\n<tr><td class=\"num\">47.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Naphthalene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/91-20-3\/Naphthalene\" target=\"_blank\" rel=\"noopener\">Naphthalene<\/a><\/td><td class=\"oth\">8.97<\/td><td class=\"oth\">1<\/td><td class=\"oth\">129 to 103 @14V<\/td><td class=\"oth\">128 to 78 @20V<\/td><td class=\"oth\">128 to 102 @16V<\/td><\/tr>\n<tr><td class=\"num\">48.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for a-Terpineol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/98-55-5\/a-Terpineol\" target=\"_blank\" rel=\"noopener\">a-Terpineol<\/a><\/td><td class=\"oth\">9.04<\/td><td class=\"oth\">1<\/td><td class=\"oth\">136 to 93 @10V<\/td><td class=\"oth\">121 to 93 @8V<\/td><td class=\"oth\">136 to 121 @8V<\/td><\/tr>\n<tr><td class=\"num\">49.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 4-Chloroaniline\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/106-47-8\/4-Chloroaniline\" target=\"_blank\" rel=\"noopener\">4-Chloroaniline<\/a><\/td><td class=\"oth\">9.09<\/td><td class=\"oth\">1<\/td><td class=\"oth\">127 to 65 @20V<\/td><td class=\"oth\">65 to 39 @12V<\/td><td class=\"oth\">127 to 100 @10V<\/td><\/tr>\n<tr><td class=\"num\">50.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 2,6-Dichlorophenol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/87-65-0\/2,6-Dichlorophenol\" target=\"_blank\" rel=\"noopener\">2,6-Dichlorophenol<\/a><\/td><td class=\"oth\">9.11<\/td><td class=\"oth\">1<\/td><td class=\"oth\">164 to 63 @26V<\/td><td class=\"oth\">162 to 63 @26V<\/td><td class=\"oth\">162 to 126 @8V<\/td><\/tr>\n<tr><td class=\"num\">51.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Hexachloropropene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/1888-71-7\/Hexachloropropene\" target=\"_blank\" rel=\"noopener\">Hexachloropropene<\/a><\/td><td class=\"oth\">9.15<\/td><td class=\"oth\">1<\/td><td class=\"oth\">215 to 119 @18V<\/td><td class=\"oth\">213 to 119 @18V<\/td><td class=\"oth\">215 to 121 @18V<\/td><\/tr>\n<tr><td class=\"num\">52.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Hexachlorobutadiene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/87-68-3\/Hexachlorobutadiene\" target=\"_blank\" rel=\"noopener\">Hexachlorobutadiene<\/a><\/td><td class=\"oth\">9.23<\/td><td class=\"oth\">1<\/td><td class=\"oth\">260 to 190 @26V<\/td><td class=\"oth\">225 to 190 @14V<\/td><td class=\"oth\">260 to 225 @12V<\/td><\/tr>\n<tr><td class=\"num\">53.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Quinoline\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/91-22-5\/Quinoline\" target=\"_blank\" rel=\"noopener\">Quinoline<\/a><\/td><td class=\"oth\">9.6<\/td><td class=\"oth\">1<\/td><td class=\"oth\">129 to 76 @26V<\/td><td class=\"oth\">102 to 76 @12V<\/td><td class=\"oth\">129 to 102 @16V<\/td><\/tr>\n<tr><td class=\"num\">54.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Caprolactam\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/105-60-2\/Caprolactam\" target=\"_blank\" rel=\"noopener\">Caprolactam<\/a><\/td><td class=\"oth\">9.7<\/td><td class=\"oth\">1<\/td><td class=\"oth\">113 to 56 @10V<\/td><td class=\"oth\">85 to 67 @6V<\/td><td class=\"oth\">113 to 85 @6V<\/td><\/tr>\n<tr><td class=\"num\">55.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for N-Nitrosodibutylamine\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/924-16-3\/N-Nitrosodibutylamine\" target=\"_blank\" rel=\"noopener\">N-Nitrosodibutylamine<\/a><\/td><td class=\"oth\">9.76<\/td><td class=\"oth\">1<\/td><td class=\"oth\">116 to 74 @8V<\/td><td class=\"oth\">84 to 56 @14V<\/td><td class=\"oth\">116 to 99 @6V<\/td><\/tr>\n<tr><td class=\"num\">56.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 1,4-Phenylenediamine\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/106-50-3\/1,4-Phenylenediamine\" target=\"_blank\" rel=\"noopener\">1,4-Phenylenediamine<\/a><\/td><td class=\"oth\">9.77<\/td><td class=\"oth\">1<\/td><td class=\"oth\">108 to 80 @20V<\/td><td class=\"oth\">80 to 53 @12V<\/td><td class=\"oth\">108 to 81 @10V<\/td><\/tr>\n<tr><td class=\"num\">57.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 4-Chloro-3-methylphenol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/59-50-7\/4-Chloro-3-methylphenol\" target=\"_blank\" rel=\"noopener\">4-Chloro-3-methylphenol<\/a><\/td><td class=\"oth\">10.04<\/td><td class=\"oth\">1<\/td><td class=\"oth\">142 to 77 @26V<\/td><td class=\"oth\">107 to 77 @12V<\/td><td class=\"oth\">142 to 107 @12V<\/td><\/tr>\n<tr><td class=\"num\">58.<\/td><td class=\"cmpd\">Isosafrole I<\/td><td class=\"oth\">10.16<\/td><td class=\"oth\">1<\/td><td class=\"oth\">162 to 104 @12V<\/td><td class=\"oth\">131 to 103 @8V<\/td><td class=\"oth\">162 to 131 @12V<\/td><\/tr>\n<tr><td class=\"num\">59.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 2-Methylnaphthalene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/91-57-6\/2-Methylnaphthalene\" target=\"_blank\" rel=\"noopener\">2-Methylnaphthalene<\/a><\/td><td class=\"oth\">10.28<\/td><td class=\"oth\">1<\/td><td class=\"oth\">141 to 89 @30V<\/td><td class=\"oth\">115 to 89 @16V<\/td><td class=\"oth\">141 to 115 @16V<\/td><\/tr>\n<tr><td class=\"num\">60.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 1-Methylnaphthalene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/90-12-0\/1-Methylnaphthalene\" target=\"_blank\" rel=\"noopener\">1-Methylnaphthalene<\/a><\/td><td class=\"oth\">10.46<\/td><td class=\"oth\">1<\/td><td class=\"oth\">141 to 89 @30V<\/td><td class=\"oth\">115 to 89 @14V<\/td><td class=\"oth\">141 to 115 @16V<\/td><\/tr>\n<tr><td class=\"num\">61.<\/td><td class=\"cmpd\">Hexachloropentadiene<\/td><td class=\"oth\">10.59<\/td><td class=\"oth\">1<\/td><td class=\"oth\">272 to 237 @12V<\/td><td class=\"oth\">237 to 119 @20V<\/td><td class=\"oth\">237 to 143 @22V<\/td><\/tr>\n<tr><td class=\"num\">62.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 1,2,4,5-Tetrachlorobenzene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/95-94-3\/1,2,4,5-Tetrachlorobenzene\" target=\"_blank\" rel=\"noopener\">1,2,4,5-Tetrachlorobenzene<\/a><\/td><td class=\"oth\">10.61<\/td><td class=\"oth\">1<\/td><td class=\"oth\">216 to 108 @36V<\/td><td class=\"oth\">179 to 143 @8V<\/td><td class=\"oth\">216 to 181 @14V<\/td><\/tr>\n<tr><td class=\"num\">63.<\/td><td class=\"cmpd\">Isosafrole II<\/td><td class=\"oth\">10.73<\/td><td class=\"oth\">1<\/td><td class=\"oth\">162 to 104 @12V<\/td><td class=\"oth\">131 to 103 @10V<\/td><td class=\"oth\">162 to 131 @12V<\/td><\/tr>\n<tr><td class=\"num\">64.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 2,3-Dichloroaniline\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/608-27-5\/2,3-Dichloroaniline\" target=\"_blank\" rel=\"noopener\">2,3-Dichloroaniline<\/a><\/td><td class=\"oth\">10.83<\/td><td class=\"oth\">1<\/td><td class=\"oth\">163 to 90 @18V<\/td><td class=\"oth\">161 to 90 @16V<\/td><td class=\"oth\">161 to 99 @18V<\/td><\/tr>\n<tr><td class=\"num\">65.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 2,4,6-Trichlorophenol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/88-06-2\/2,4,6-Trichlorophenol\" target=\"_blank\" rel=\"noopener\">2,4,6-Trichlorophenol<\/a><\/td><td class=\"oth\">10.83<\/td><td class=\"oth\">1<\/td><td class=\"oth\">196 to 97 @24V<\/td><td class=\"oth\">132 to 97 @10V<\/td><td class=\"oth\">160 to 97 @16V<\/td><\/tr>\n<tr><td class=\"num\">66.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 2,4,5-Trichlorophenol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/95-95-4\/2,4,5-Trichlorophenol\" target=\"_blank\" rel=\"noopener\">2,4,5-Trichlorophenol<\/a><\/td><td class=\"oth\">10.88<\/td><td class=\"oth\">1<\/td><td class=\"oth\">196 to 97 @24V<\/td><td class=\"oth\">132 to 97 @10V<\/td><td class=\"oth\">196 to 132 @12V<\/td><\/tr>\n<tr><td class=\"num\">67.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 2-Fluorobiphenyl\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/321-60-8\/2-Fluorobiphenyl\" target=\"_blank\" rel=\"noopener\">2-Fluorobiphenyl<\/a><\/td><td class=\"oth\">10.99<\/td><td class=\"oth\">1<\/td><td class=\"oth\">172 to 151 @22V<\/td><td class=\"oth\">172 to 170 @22V<\/td><td class=\"oth\">172 to 171 @12V<\/td><\/tr>\n<tr><td class=\"num\">68.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Safrole\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/94-59-7\/Safrole\" target=\"_blank\" rel=\"noopener\">Safrole<\/a><\/td><td class=\"oth\">11.16<\/td><td class=\"oth\">1<\/td><td class=\"oth\">162 to 77 @30V<\/td><td class=\"oth\">131 to 103 @10V<\/td><td class=\"oth\">162 to 127 @16V<\/td><\/tr>\n<tr><td class=\"num\">69.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 2-Chloronaphthalene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/91-58-7\/2-Chloronaphthalene\" target=\"_blank\" rel=\"noopener\">2-Chloronaphthalene<\/a><\/td><td class=\"oth\">11.2<\/td><td class=\"oth\">1<\/td><td class=\"oth\">162 to 77 @32V<\/td><td class=\"oth\">127 to 77 @16V<\/td><td class=\"oth\">162 to 127 @16V<\/td><\/tr>\n<tr><td class=\"num\">70.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Biphenyl\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/92-52-4\/Biphenyl\" target=\"_blank\" rel=\"noopener\">Biphenyl<\/a><\/td><td class=\"oth\">11.2<\/td><td class=\"oth\">1<\/td><td class=\"oth\">153 to 126 @32V<\/td><td class=\"oth\">154 to 115 @24V<\/td><td class=\"oth\">154 to 152 @22V<\/td><\/tr>\n<tr><td class=\"num\">71.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 1-Chloronaphthalene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/90-13-1\/1-Chloronaphthalene\" target=\"_blank\" rel=\"noopener\">1-Chloronaphthalene<\/a><\/td><td class=\"oth\">11.25<\/td><td class=\"oth\">1<\/td><td class=\"oth\">162 to 77 @30V<\/td><td class=\"oth\">127 to 77 @16V<\/td><td class=\"oth\">162 to 127 @16V<\/td><\/tr>\n<tr><td class=\"num\">72.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for o-Nitroaniline\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/88-74-4\/o-Nitroaniline\" target=\"_blank\" rel=\"noopener\">o-Nitroaniline<\/a><\/td><td class=\"oth\">11.41<\/td><td class=\"oth\">1<\/td><td class=\"oth\">138 to 65 @22V<\/td><td class=\"oth\">108 to 80 @8V<\/td><td class=\"oth\">138 to 92 @12V<\/td><\/tr>\n<tr><td class=\"num\">73.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Diphenyl ether\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/101-84-8\/Diphenyl ether\" target=\"_blank\" rel=\"noopener\">Diphenyl ether<\/a><\/td><td class=\"oth\">11.42<\/td><td class=\"oth\">1<\/td><td class=\"oth\">170 to 141 @16V<\/td><td class=\"oth\">141 to 115 @14V<\/td><td class=\"oth\">170 to 142 @10V<\/td><\/tr>\n<tr><td class=\"num\">74.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 1,4-Naphthoquinone\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/130-15-4\/1,4-Naphthoquinone\" target=\"_blank\" rel=\"noopener\">1,4-Naphthoquinone<\/a><\/td><td class=\"oth\">11.56<\/td><td class=\"oth\">1<\/td><td class=\"oth\">158 to 102 @14V<\/td><td class=\"oth\">102 to 76 @12V<\/td><td class=\"oth\">158 to 130 @8V<\/td><\/tr>\n<tr><td class=\"num\">75.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 1,2-Dinitrobenzene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/528-29-0\/1,2-Dinitrobenzene\" target=\"_blank\" rel=\"noopener\">1,2-Dinitrobenzene<\/a><\/td><td class=\"oth\">11.7<\/td><td class=\"oth\">1<\/td><td class=\"oth\">168 to 75 @20V<\/td><td class=\"oth\">122 to 75 @12V<\/td><td class=\"oth\">168 to 122 @8V<\/td><\/tr>\n<tr><td class=\"num\">76.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 1,3-Dinitrobenzene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/99-65-0\/1,3-Dinitrobenzene\" target=\"_blank\" rel=\"noopener\">1,3-Dinitrobenzene<\/a><\/td><td class=\"oth\">11.83<\/td><td class=\"oth\">1<\/td><td class=\"oth\">168 to 75 @22V<\/td><td class=\"oth\">76 to 50 @10V<\/td><td class=\"oth\">168 to 122 @8V<\/td><\/tr>\n<tr><td class=\"num\">77.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Diphenyl phthalate\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/84-62-8\/Diphenyl phthalate\" target=\"_blank\" rel=\"noopener\">Diphenyl phthalate<\/a><\/td><td class=\"oth\">11.83<\/td><td class=\"oth\">1<\/td><td class=\"oth\">163 to 77 @20V<\/td><td class=\"oth\">163 to 92 @24V<\/td><td class=\"oth\">163 to 133 @8V<\/td><\/tr>\n<tr><td class=\"num\">78.<\/td><td class=\"cmpd\">1,6-Dinitrotoluene<\/td><td class=\"oth\">11.91<\/td><td class=\"oth\">1<\/td><td class=\"oth\">165 to 63 @20V<\/td><td class=\"oth\">148 to 90 @10V<\/td><td class=\"oth\">165 to 148 @8V<\/td><\/tr>\n<tr><td class=\"num\">79.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Acenaphthylene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/208-96-8\/Acenaphthylene\" target=\"_blank\" rel=\"noopener\">Acenaphthylene<\/a><\/td><td class=\"oth\">11.98<\/td><td class=\"oth\">1<\/td><td class=\"oth\">152 to 76 @36V<\/td><td class=\"oth\">152 to 102 @26V<\/td><td class=\"oth\">152 to 126 @24V<\/td><\/tr>\n<tr><td class=\"num\">80.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 1,4-Dinitrobenzene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/100-25-4\/1,4-Dinitrobenzene\" target=\"_blank\" rel=\"noopener\">1,4-Dinitrobenzene<\/a><\/td><td class=\"oth\">11.99<\/td><td class=\"oth\">1<\/td><td class=\"oth\">168 to 51 @14V<\/td><td class=\"oth\">76 to 63 @6V<\/td><td class=\"oth\">168 to 63 @26V<\/td><\/tr>\n<tr><td class=\"num\">81.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 3-Nitroaniline\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/99-09-2\/3-Nitroaniline\" target=\"_blank\" rel=\"noopener\">3-Nitroaniline<\/a><\/td><td class=\"oth\">12.21<\/td><td class=\"oth\">1<\/td><td class=\"oth\">138 to 65 @20V<\/td><td class=\"oth\">92 to 65 @8V<\/td><td class=\"oth\">138 to 92 @12V<\/td><\/tr>\n<tr><td class=\"num\">82.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Acenaphthene-d10\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/15067-26-2\/Acenaphthene-d10\" target=\"_blank\" rel=\"noopener\">Acenaphthene-d10<\/a><\/td><td class=\"oth\">12.24<\/td><td class=\"oth\">1<\/td><td class=\"oth\">164 to 160 @32V<\/td><td class=\"oth\">162 to 160 @18V<\/td><td class=\"oth\">164 to 162 @14V<\/td><\/tr>\n<\/tbody><\/table>\n<table class=\"peaks col-lg-6 col-12 peak-50\"><thead><tr><th><\/th><th style=\"text-align: left;width: 75px\">Peaks<\/th><th style=\"text-align: center;width: 75px\">t<sub>R<\/sub> (min)<\/th><th style=\"text-align: center;width: 75px\">Conc.<br \/>(\u00b5g\/g)<\/th><th style=\"text-align: center;width: 75px\">SRM 1<\/th><th style=\"text-align: center;width: 75px\">SRM 2<\/th><th style=\"text-align: center;width: 75px\">SRM 3<\/th><\/tr><\/thead><tbody>\n<tr><td class=\"num\">83.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Acenaphthene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/83-32-9\/Acenaphthene\" target=\"_blank\" rel=\"noopener\">Acenaphthene<\/a><\/td><td class=\"oth\">12.32<\/td><td class=\"oth\">1<\/td><td class=\"oth\">153 to 77 @38V<\/td><td class=\"oth\">153 to 126 @36V<\/td><td class=\"oth\">154 to 127 @34V<\/td><\/tr>\n<tr><td class=\"num\">84.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 2,4-Dinitrophenol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/51-28-5\/2,4-Dinitrophenol\" target=\"_blank\" rel=\"noopener\">2,4-Dinitrophenol<\/a><\/td><td class=\"oth\">12.41<\/td><td class=\"oth\">1<\/td><td class=\"oth\">184 to 107 @10V<\/td><td class=\"oth\">154 to 79 @12V<\/td><td class=\"oth\">184 to 154 @6V<\/td><\/tr>\n<tr><td class=\"num\">85.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 4-Nitrophenol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/100-02-7\/4-Nitrophenol\" target=\"_blank\" rel=\"noopener\">4-Nitrophenol<\/a><\/td><td class=\"oth\">12.57<\/td><td class=\"oth\">1<\/td><td class=\"oth\">139 to 81 @14V<\/td><td class=\"oth\">109 to 81 @8V<\/td><td class=\"oth\">139 to 109 @6V<\/td><\/tr>\n<tr><td class=\"num\">86.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Pentachlorobenzene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/608-93-5\/Pentachlorobenzene\" target=\"_blank\" rel=\"noopener\">Pentachlorobenzene<\/a><\/td><td class=\"oth\">12.6<\/td><td class=\"oth\">1<\/td><td class=\"oth\">250 to 142 @38V<\/td><td class=\"oth\">215 to 179 @10V<\/td><td class=\"oth\">250 to 215 @16V<\/td><\/tr>\n<tr><td class=\"num\">87.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 2,4-Dinitrotoluene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/121-14-2\/2,4-Dinitrotoluene\" target=\"_blank\" rel=\"noopener\">2,4-Dinitrotoluene<\/a><\/td><td class=\"oth\">12.66<\/td><td class=\"oth\">1<\/td><td class=\"oth\">165 to 63 @22V<\/td><td class=\"oth\">89 to 63 @12V<\/td><td class=\"oth\">165 to 119 @6V<\/td><\/tr>\n<tr><td class=\"num\">88.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Dibenzofuran\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/132-64-9\/Dibenzofuran\" target=\"_blank\" rel=\"noopener\">Dibenzofuran<\/a><\/td><td class=\"oth\">12.66<\/td><td class=\"oth\">1<\/td><td class=\"oth\">168 to 139 @22V<\/td><td class=\"oth\">139 to 63 @30V<\/td><td class=\"oth\">139 to 89 @18V<\/td><\/tr>\n<tr><td class=\"num\">89.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 1-Naphthylamine\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/134-32-7\/1-Naphthylamine\" target=\"_blank\" rel=\"noopener\">1-Naphthylamine<\/a><\/td><td class=\"oth\">12.81<\/td><td class=\"oth\">1<\/td><td class=\"oth\">143 to 115 @22V<\/td><td class=\"oth\">115 to 89 @16V<\/td><td class=\"oth\">143 to 116 @10V<\/td><\/tr>\n<tr><td class=\"num\">90.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 2,3,5,6-Tetrachlorophenol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/935-95-5\/2,3,5,6-Tetrachlorophenol\" target=\"_blank\" rel=\"noopener\">2,3,5,6-Tetrachlorophenol<\/a><\/td><td class=\"oth\">12.83<\/td><td class=\"oth\">1<\/td><td class=\"oth\">234 to 133 @26V<\/td><td class=\"oth\">232 to 131 @26V<\/td><td class=\"oth\">232 to 168 @12V<\/td><\/tr>\n<tr><td class=\"num\">91.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 2,3,4,6-Tetrachlorophenol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/58-90-2\/2,3,4,6-Tetrachlorophenol\" target=\"_blank\" rel=\"noopener\">2,3,4,6-Tetrachlorophenol<\/a><\/td><td class=\"oth\">12.91<\/td><td class=\"oth\">1<\/td><td class=\"oth\">234 to 131 @24V<\/td><td class=\"oth\">232 to 133 @26V<\/td><td class=\"oth\">232 to 168 @12V<\/td><\/tr>\n<tr><td class=\"num\">92.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 2-Naphthylamine\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/91-59-8\/2-Naphthylamine\" target=\"_blank\" rel=\"noopener\">2-Naphthylamine<\/a><\/td><td class=\"oth\">12.97<\/td><td class=\"oth\">1<\/td><td class=\"oth\">143 to 115 @24V<\/td><td class=\"oth\">115 to 89 @16V<\/td><td class=\"oth\">143 to 116 @10V<\/td><\/tr>\n<tr><td class=\"num\">93.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Diethyl Phthalate\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/84-66-2\/Diethyl Phthalate\" target=\"_blank\" rel=\"noopener\">Diethyl Phthalate<\/a><\/td><td class=\"oth\">13.21<\/td><td class=\"oth\">1<\/td><td class=\"oth\">177 to 149 @8V<\/td><td class=\"oth\">149 to 65 @20V<\/td><td class=\"oth\">149 to 93 @12V<\/td><\/tr>\n<tr><td class=\"num\">94.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Hexadecane\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/544-76-3\/Hexadecane\" target=\"_blank\" rel=\"noopener\">Hexadecane<\/a><\/td><td class=\"oth\">13.29<\/td><td class=\"oth\">1<\/td><td class=\"oth\">99 to 41 @14V<\/td><td class=\"oth\">85 to 43 @6V<\/td><td class=\"oth\">99 to 57 @6V<\/td><\/tr>\n<tr><td class=\"num\">95.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Fluorene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/86-73-7\/Fluorene\" target=\"_blank\" rel=\"noopener\">Fluorene<\/a><\/td><td class=\"oth\">13.31<\/td><td class=\"oth\">1<\/td><td class=\"oth\">167 to 166 @14V<\/td><td class=\"oth\">165 to 115 @24V<\/td><td class=\"oth\">165 to 139 @26V<\/td><\/tr>\n<tr><td class=\"num\">96.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Zinophos\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/297-97-2\/Zinophos\" target=\"_blank\" rel=\"noopener\">Zinophos<\/a><\/td><td class=\"oth\">13.36<\/td><td class=\"oth\">1<\/td><td class=\"oth\">143 to 79 @10V<\/td><td class=\"oth\">107 to 52 @20V<\/td><td class=\"oth\">107 to 79 @14V<\/td><\/tr>\n<tr><td class=\"num\">97.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 5-Nitro-o-toluidine\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/99-55-8\/5-Nitro-o-toluidine\" target=\"_blank\" rel=\"noopener\">5-Nitro-o-toluidine<\/a><\/td><td class=\"oth\">13.37<\/td><td class=\"oth\">1<\/td><td class=\"oth\">152 to 77 @24V<\/td><td class=\"oth\">106 to 77 @12V<\/td><td class=\"oth\">152 to 106 @10V<\/td><\/tr>\n<tr><td class=\"num\">98.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 4-Chlorophenyl phenyl ether\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/7005-72-3\/4-Chlorophenyl phenyl ether\" target=\"_blank\" rel=\"noopener\">4-Chlorophenyl phenyl ether<\/a><\/td><td class=\"oth\">13.37<\/td><td class=\"oth\">1<\/td><td class=\"oth\">204 to 77 @22V<\/td><td class=\"oth\">141 to 115 @14V<\/td><td class=\"oth\">204 to 141 @12V<\/td><\/tr>\n<tr><td class=\"num\">99.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 4-Nitroaniline\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/100-01-6\/4-Nitroaniline\" target=\"_blank\" rel=\"noopener\">4-Nitroaniline<\/a><\/td><td class=\"oth\">13.37<\/td><td class=\"oth\">1<\/td><td class=\"oth\">138 to 80 @18V<\/td><td class=\"oth\">108 to 80 @10V<\/td><td class=\"oth\">138 to 108 @8V<\/td><\/tr>\n<tr><td class=\"num\">100.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 4,6-Dinitro-2-methylphenol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/534-52-1\/4,6-Dinitro-2-methylphenol\" target=\"_blank\" rel=\"noopener\">4,6-Dinitro-2-methylphenol<\/a><\/td><td class=\"oth\">13.45<\/td><td class=\"oth\">1<\/td><td class=\"oth\">198 to 121 @10V<\/td><td class=\"oth\">168 to 121 @6V<\/td><td class=\"oth\">198 to 168 @6V<\/td><\/tr>\n<tr><td class=\"num\">101.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Diphenylamine\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/122-39-4\/Diphenylamine\" target=\"_blank\" rel=\"noopener\">Diphenylamine<\/a><\/td><td class=\"oth\">13.6<\/td><td class=\"oth\">1<\/td><td class=\"oth\">169 to 66 @22V<\/td><td class=\"oth\">169 to 77 @30V<\/td><td class=\"oth\">170 to 66 @22V<\/td><\/tr>\n<tr><td class=\"num\">102.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Azobenzene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/103-33-3\/Azobenzene\" target=\"_blank\" rel=\"noopener\">Azobenzene<\/a><\/td><td class=\"oth\">13.67<\/td><td class=\"oth\">1<\/td><td class=\"oth\">182 to 77 @12V<\/td><td class=\"oth\">77 to 51 @14V<\/td><td class=\"oth\">182 to 105 @6V<\/td><\/tr>\n<tr><td class=\"num\">103.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 2,4,6-Tribromophenol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/118-79-6\/2,4,6-Tribromophenol\" target=\"_blank\" rel=\"noopener\">2,4,6-Tribromophenol<\/a><\/td><td class=\"oth\">13.77<\/td><td class=\"oth\">1<\/td><td class=\"oth\">332 to 143 @34V<\/td><td class=\"oth\">330 to 141 @36V<\/td><td class=\"oth\">330 to 222 @20V<\/td><\/tr>\n<tr><td class=\"num\">104.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Sulfotepp\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/3689-24-5\/Sulfotepp\" target=\"_blank\" rel=\"noopener\">Sulfotepp<\/a><\/td><td class=\"oth\">14<\/td><td class=\"oth\">1<\/td><td class=\"oth\">322 to 146 @24V<\/td><td class=\"oth\">202 to 146 @10V<\/td><td class=\"oth\">322 to 202 @10V<\/td><\/tr>\n<tr><td class=\"num\">105.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 1,3,5-Trinitrobenzene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/99-35-4\/1,3,5-Trinitrobenzene\" target=\"_blank\" rel=\"noopener\">1,3,5-Trinitrobenzene<\/a><\/td><td class=\"oth\">14.14<\/td><td class=\"oth\">1<\/td><td class=\"oth\">213 to 74 @38V<\/td><td class=\"oth\">213 to 120 @22V<\/td><td class=\"oth\">213 to 167 @8V<\/td><\/tr>\n<tr><td class=\"num\">106.<\/td><td class=\"cmpd\">Diallate 1<\/td><td class=\"oth\">14.2<\/td><td class=\"oth\">1<\/td><td class=\"oth\">234 to 150 @16V<\/td><td class=\"oth\">86 to 43 @6V<\/td><td class=\"oth\">234 to 192 @12V<\/td><\/tr>\n<tr><td class=\"num\">107.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Phorate\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/298-02-2\/Phorate\" target=\"_blank\" rel=\"noopener\">Phorate<\/a><\/td><td class=\"oth\">14.22<\/td><td class=\"oth\">1<\/td><td class=\"oth\">121 to 47 @26V<\/td><td class=\"oth\">75 to 47 @8V<\/td><td class=\"oth\">121 to 65 @8V<\/td><\/tr>\n<tr><td class=\"num\">108.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Phenacetin\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/62-44-2\/Phenacetin\" target=\"_blank\" rel=\"noopener\">Phenacetin<\/a><\/td><td class=\"oth\">14.25<\/td><td class=\"oth\">1<\/td><td class=\"oth\">179 to 109 @14V<\/td><td class=\"oth\">108 to 80 @12V<\/td><td class=\"oth\">179 to 137 @8V<\/td><\/tr>\n<tr><td class=\"num\">109.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 4-Bromophenyl phenyl ether\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/101-55-3\/4-Bromophenyl phenyl ether\" target=\"_blank\" rel=\"noopener\">4-Bromophenyl phenyl ether<\/a><\/td><td class=\"oth\">14.31<\/td><td class=\"oth\">1<\/td><td class=\"oth\">250 to 141 @16V<\/td><td class=\"oth\">141 to 115 @12V<\/td><td class=\"oth\">248 to 141 @14V<\/td><\/tr>\n<tr><td class=\"num\">110.<\/td><td class=\"cmpd\">Diallate 2<\/td><td class=\"oth\">14.37<\/td><td class=\"oth\">1<\/td><td class=\"oth\">234 to 150 @16V<\/td><td class=\"oth\">86 to 43 @6V<\/td><td class=\"oth\">234 to 192 @10V<\/td><\/tr>\n<tr><td class=\"num\">111.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Hexachlorobenzene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/118-74-1\/Hexachlorobenzene\" target=\"_blank\" rel=\"noopener\">Hexachlorobenzene<\/a><\/td><td class=\"oth\">14.37<\/td><td class=\"oth\">1<\/td><td class=\"oth\">288 to 216 @26V<\/td><td class=\"oth\">249 to 214 @12V<\/td><td class=\"oth\">284 to 249 @14V<\/td><\/tr>\n<tr><td class=\"num\">112.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Dimethoate\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/60-51-5\/Dimethoate\" target=\"_blank\" rel=\"noopener\">Dimethoate<\/a><\/td><td class=\"oth\">14.51<\/td><td class=\"oth\">1<\/td><td class=\"oth\">125 to 47 @14V<\/td><td class=\"oth\">93 to 63 @8V<\/td><td class=\"oth\">125 to 79 @8V<\/td><\/tr>\n<tr><td class=\"num\">113.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Atrazine\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/1912-24-9\/Atrazine\" target=\"_blank\" rel=\"noopener\">Atrazine<\/a><\/td><td class=\"oth\">14.7<\/td><td class=\"oth\">1<\/td><td class=\"oth\">215 to 58 @10V<\/td><td class=\"oth\">200 to 122 @8V<\/td><td class=\"oth\">215 to 200 @8V<\/td><\/tr>\n<tr><td class=\"num\">114.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Pentachlorophenol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/87-86-5\/Pentachlorophenol\" target=\"_blank\" rel=\"noopener\">Pentachlorophenol<\/a><\/td><td class=\"oth\">14.77<\/td><td class=\"oth\">1<\/td><td class=\"oth\">270 to 169 @22V<\/td><td class=\"oth\">228 to 165 @14V<\/td><td class=\"oth\">266 to 167 @22V<\/td><\/tr>\n<tr><td class=\"num\">115.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 4-Aminobiphenyl\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/92-67-1\/4-Aminobiphenyl\" target=\"_blank\" rel=\"noopener\">4-Aminobiphenyl<\/a><\/td><td class=\"oth\">14.79<\/td><td class=\"oth\">1<\/td><td class=\"oth\">169 to 141 @22V<\/td><td class=\"oth\">168 to 141 @8V<\/td><td class=\"oth\">168 to 167 @10V<\/td><\/tr>\n<tr><td class=\"num\">116.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Pentachloronitrobenzene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/82-68-8\/Pentachloronitrobenzene\" target=\"_blank\" rel=\"noopener\">Pentachloronitrobenzene<\/a><\/td><td class=\"oth\">14.81<\/td><td class=\"oth\">1<\/td><td class=\"oth\">295 to 237 @14V<\/td><td class=\"oth\">237 to 143 @20V<\/td><td class=\"oth\">295 to 265 @8V<\/td><\/tr>\n<tr><td class=\"num\">117.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Propyzamide\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/23950-58-5\/Propyzamide\" target=\"_blank\" rel=\"noopener\">Propyzamide<\/a><\/td><td class=\"oth\">15<\/td><td class=\"oth\">1<\/td><td class=\"oth\">254 to 191 @16V<\/td><td class=\"oth\">173 to 145 @14V<\/td><td class=\"oth\">254 to 226 @10V<\/td><\/tr>\n<tr><td class=\"num\">118.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Octadecane\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/593-45-3\/Octadecane\" target=\"_blank\" rel=\"noopener\">Octadecane<\/a><\/td><td class=\"oth\">15.1<\/td><td class=\"oth\">1<\/td><td class=\"oth\">99 to 41 @14V<\/td><td class=\"oth\">85 to 43 @6V<\/td><td class=\"oth\">99 to 57 @6V<\/td><\/tr>\n<tr><td class=\"num\">119.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Phenanthrene-d10\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/1517-22-2\/Phenanthrene-d10\" target=\"_blank\" rel=\"noopener\">Phenanthrene-d10<\/a><\/td><td class=\"oth\">15.1<\/td><td class=\"oth\">1<\/td><td class=\"oth\">188 to 160 @20V<\/td><td class=\"oth\">184 to 154 @32V<\/td><td class=\"oth\">184 to 156 @22V<\/td><\/tr>\n<tr><td class=\"num\">120.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Phenanthrene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/85-01-8\/Phenanthrene\" target=\"_blank\" rel=\"noopener\">Phenanthrene<\/a><\/td><td class=\"oth\">15.16<\/td><td class=\"oth\">1<\/td><td class=\"oth\">178 to 151 @32V<\/td><td class=\"oth\">176 to 150 @20V<\/td><td class=\"oth\">178 to 152 @20V<\/td><\/tr>\n<tr><td class=\"num\">121.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Dinoseb\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/88-85-7\/Dinoseb\" target=\"_blank\" rel=\"noopener\">Dinoseb<\/a><\/td><td class=\"oth\">15.21<\/td><td class=\"oth\">1<\/td><td class=\"oth\">240 to 117 @24V<\/td><td class=\"oth\">163 to 116 @14V<\/td><td class=\"oth\">240 to 211 @8V<\/td><\/tr>\n<tr><td class=\"num\">122.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Disulfoton\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/298-04-4\/Disulfoton\" target=\"_blank\" rel=\"noopener\">Disulfoton<\/a><\/td><td class=\"oth\">15.22<\/td><td class=\"oth\">1<\/td><td class=\"oth\">97 to 65 @16V<\/td><td class=\"oth\">88 to 60 @6V<\/td><td class=\"oth\">97 to 79 @12V<\/td><\/tr>\n<tr><td class=\"num\">123.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Anthracene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/120-12-7\/Anthracene\" target=\"_blank\" rel=\"noopener\">Anthracene<\/a><\/td><td class=\"oth\">15.26<\/td><td class=\"oth\">1<\/td><td class=\"oth\">178 to 151 @32V<\/td><td class=\"oth\">177 to 151 @18V<\/td><td class=\"oth\">178 to 152 @18V<\/td><\/tr>\n<tr><td class=\"num\">124.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Carbazole\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/86-74-8\/Carbazole\" target=\"_blank\" rel=\"noopener\">Carbazole<\/a><\/td><td class=\"oth\">15.6<\/td><td class=\"oth\">1<\/td><td class=\"oth\">167 to 139 @24V<\/td><td class=\"oth\">166 to 140 @14V<\/td><td class=\"oth\">167 to 166 @12V<\/td><\/tr>\n<tr><td class=\"num\">125.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Methyl Parathion\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/298-00-0\/Methyl Parathion\" target=\"_blank\" rel=\"noopener\">Methyl Parathion<\/a><\/td><td class=\"oth\">15.95<\/td><td class=\"oth\">1<\/td><td class=\"oth\">263 to 79 @24V<\/td><td class=\"oth\">125 to 47 @12V<\/td><td class=\"oth\">263 to 109 @10V<\/td><\/tr>\n<tr><td class=\"num\">126.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for di-n-Butyl phthalate\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/84-74-2\/di-n-Butyl phthalate\" target=\"_blank\" rel=\"noopener\">di-n-Butyl phthalate<\/a><\/td><td class=\"oth\">16.41<\/td><td class=\"oth\">1<\/td><td class=\"oth\">149 to 65 @20V<\/td><td class=\"oth\">149 to 93 @14V<\/td><td class=\"oth\">149 to 121 @12V<\/td><\/tr>\n<tr><td class=\"num\">127.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 4-Nitroquinoline-N-oxide\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/56-57-5\/4-Nitroquinoline-N-oxide\" target=\"_blank\" rel=\"noopener\">4-Nitroquinoline-N-oxide<\/a><\/td><td class=\"oth\">16.69<\/td><td class=\"oth\">1<\/td><td class=\"oth\">190 to 77 @20V<\/td><td class=\"oth\">101 to 75 @10V<\/td><td class=\"oth\">190 to 160 @8V<\/td><\/tr>\n<tr><td class=\"num\">128.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Ethyl Parathion\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/56-38-2\/Ethyl Parathion\" target=\"_blank\" rel=\"noopener\">Ethyl Parathion<\/a><\/td><td class=\"oth\">16.73<\/td><td class=\"oth\">1<\/td><td class=\"oth\">291 to 81 @26V<\/td><td class=\"oth\">109 to 81 @10V<\/td><td class=\"oth\">291 to 109 @10V<\/td><\/tr>\n<tr><td class=\"num\">129.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Methapyrilene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/91-80-5\/Methapyrilene\" target=\"_blank\" rel=\"noopener\">Methapyrilene<\/a><\/td><td class=\"oth\">16.9<\/td><td class=\"oth\">1<\/td><td class=\"oth\">190 to 97 @14V<\/td><td class=\"oth\">97 to 53 @16V<\/td><td class=\"oth\">190 to 157 @8V<\/td><\/tr>\n<tr><td class=\"num\">130.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Isodrin\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/465-73-6\/Isodrin\" target=\"_blank\" rel=\"noopener\">Isodrin<\/a><\/td><td class=\"oth\">17.19<\/td><td class=\"oth\">1<\/td><td class=\"oth\">261 to 191 @28V<\/td><td class=\"oth\">193 to 123 @28V<\/td><td class=\"oth\">261 to 226 @16V<\/td><\/tr>\n<tr><td class=\"num\">131.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Fluoranthene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/206-44-0\/Fluoranthene\" target=\"_blank\" rel=\"noopener\">Fluoranthene<\/a><\/td><td class=\"oth\">17.47<\/td><td class=\"oth\">1<\/td><td class=\"oth\">202 to 152 @30V<\/td><td class=\"oth\">200 to 174 @22V<\/td><td class=\"oth\">202 to 176 @26V<\/td><\/tr>\n<tr><td class=\"num\">132.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Benzidine\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/92-87-5\/Benzidine\" target=\"_blank\" rel=\"noopener\">Benzidine<\/a><\/td><td class=\"oth\">17.8<\/td><td class=\"oth\">1<\/td><td class=\"oth\">184 to 130 @26V<\/td><td class=\"oth\">184 to 156 @18V<\/td><td class=\"oth\">184 to 166 @16V<\/td><\/tr>\n<tr><td class=\"num\">133.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Pyrene-d10\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/1718-52-1\/Pyrene-d10\" target=\"_blank\" rel=\"noopener\">Pyrene-d10<\/a><\/td><td class=\"oth\">17.84<\/td><td class=\"oth\">1<\/td><td class=\"oth\">212 to 208 @36V<\/td><td class=\"oth\">212 to 210 @20V<\/td><td class=\"oth\">0<\/td><\/tr>\n<tr><td class=\"num\">134.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Pyrene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/129-00-0\/Pyrene\" target=\"_blank\" rel=\"noopener\">Pyrene<\/a><\/td><td class=\"oth\">17.88<\/td><td class=\"oth\">1<\/td><td class=\"oth\">200 to 149 @34V<\/td><td class=\"oth\">200 to 150 @26V<\/td><td class=\"oth\">200 to 174 @24V<\/td><\/tr>\n<tr><td class=\"num\">135.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for p-Terphenyl-d14\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/1718-51-0\/p-Terphenyl-d14\" target=\"_blank\" rel=\"noopener\">p-Terphenyl-d14<\/a><\/td><td class=\"oth\">18.3<\/td><td class=\"oth\">1<\/td><td class=\"oth\">244 to 226 @18V<\/td><td class=\"oth\">244 to 240 @22V<\/td><td class=\"oth\">244 to 242 @14V<\/td><\/tr>\n<tr><td class=\"num\">136.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Aramite 1\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/140-57-8\/Aramite 1\" target=\"_blank\" rel=\"noopener\">Aramite 1<\/a><\/td><td class=\"oth\">18.33<\/td><td class=\"oth\">1<\/td><td class=\"oth\">319 to 123 @10V<\/td><td class=\"oth\">175 to 107 @14V<\/td><td class=\"oth\">319 to 185 @6V<\/td><\/tr>\n<tr><td class=\"num\">137.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Aramite 2\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/140-57-8\/Aramite 2\" target=\"_blank\" rel=\"noopener\">Aramite 2<\/a><\/td><td class=\"oth\">18.49<\/td><td class=\"oth\">1<\/td><td class=\"oth\">319 to 107 @36V<\/td><td class=\"oth\">185 to 63 @12V<\/td><td class=\"oth\">319 to 185 @6V<\/td><\/tr>\n<tr><td class=\"num\">138.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for p-Dimethylaminoazobenzene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/60-11-7\/p-Dimethylaminoazobenzene\" target=\"_blank\" rel=\"noopener\">p-Dimethylaminoazobenzene<\/a><\/td><td class=\"oth\">18.57<\/td><td class=\"oth\">1<\/td><td class=\"oth\">225 to 120 @10V<\/td><td class=\"oth\">120 to 77 @16V<\/td><td class=\"oth\">225 to 148 @6V<\/td><\/tr>\n<tr><td class=\"num\">139.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Chlorobenzilate\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/510-15-6\/Chlorobenzilate\" target=\"_blank\" rel=\"noopener\">Chlorobenzilate<\/a><\/td><td class=\"oth\">18.71<\/td><td class=\"oth\">1<\/td><td class=\"oth\">251 to 111 @30V<\/td><td class=\"oth\">139 to 111 @12V<\/td><td class=\"oth\">251 to 139 @12V<\/td><\/tr>\n<tr><td class=\"num\">140.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Famphur\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/52-85-7\/Famphur\" target=\"_blank\" rel=\"noopener\">Famphur<\/a><\/td><td class=\"oth\">19.17<\/td><td class=\"oth\">1<\/td><td class=\"oth\">218 to 109 @14V<\/td><td class=\"oth\">125 to 47 @14V<\/td><td class=\"oth\">125 to 79 @6V<\/td><\/tr>\n<tr><td class=\"num\">141.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 3,3&amp;apos;-Dimethylbenzidine\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/119-93-7\/3,3&amp;apos;-Dimethylbenzidine\" target=\"_blank\" rel=\"noopener\">3,3&#8242;-Dimethylbenzidine<\/a><\/td><td class=\"oth\">19.21<\/td><td class=\"oth\">1<\/td><td class=\"oth\">212 to 180 @22V<\/td><td class=\"oth\">211 to 195 @16V<\/td><td class=\"oth\">211 to 196 @8V<\/td><\/tr>\n<tr><td class=\"num\">142.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Kepone\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/143-50-0\/Kepone\" target=\"_blank\" rel=\"noopener\">Kepone<\/a><\/td><td class=\"oth\">19.24<\/td><td class=\"oth\">1<\/td><td class=\"oth\">272 to 237 @12V<\/td><td class=\"oth\">237 to 119 @24V<\/td><td class=\"oth\">237 to 143 @22V<\/td><\/tr>\n<tr><td class=\"num\">143.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Benzyl butyl phthalate\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/85-68-7\/Benzyl butyl phthalate\" target=\"_blank\" rel=\"noopener\">Benzyl butyl phthalate<\/a><\/td><td class=\"oth\">19.31<\/td><td class=\"oth\">1<\/td><td class=\"oth\">206 to 149 @8V<\/td><td class=\"oth\">149 to 65 @20V<\/td><td class=\"oth\">149 to 121 @10V<\/td><\/tr>\n<tr><td class=\"num\">144.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Bis(2-ethylhexyl) adipate\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/103-23-1\/Bis(2-ethylhexyl) adipate\" target=\"_blank\" rel=\"noopener\">Bis(2-ethylhexyl) adipate<\/a><\/td><td class=\"oth\">19.55<\/td><td class=\"oth\">1<\/td><td class=\"oth\">129 to 55 @14V<\/td><td class=\"oth\">111 to 83 @6V<\/td><td class=\"oth\">129 to 83 @8V<\/td><\/tr>\n<tr><td class=\"num\">145.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 2-Acetylaminofluorene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/53-96-3\/2-Acetylaminofluorene\" target=\"_blank\" rel=\"noopener\">2-Acetylaminofluorene<\/a><\/td><td class=\"oth\">19.69<\/td><td class=\"oth\">1<\/td><td class=\"oth\">223 to 180 @26V<\/td><td class=\"oth\">181 to 152 @34V<\/td><td class=\"oth\">223 to 181 @10V<\/td><\/tr>\n<tr><td class=\"num\">146.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Chrysene-d12\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/1719-03-5\/Chrysene-d12\" target=\"_blank\" rel=\"noopener\">Chrysene-d12<\/a><\/td><td class=\"oth\">20.23<\/td><td class=\"oth\">1<\/td><td class=\"oth\">240 to 212 @22V<\/td><td class=\"oth\">240 to 236 @30V<\/td><td class=\"oth\">240 to 238 @14V<\/td><\/tr>\n<tr><td class=\"num\">147.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Benz[a]anthracene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/56-55-3\/Benz[a]anthracene\" target=\"_blank\" rel=\"noopener\">Benz[a]anthracene<\/a><\/td><td class=\"oth\">20.23<\/td><td class=\"oth\">1<\/td><td class=\"oth\">228 to 202 @24V<\/td><td class=\"oth\">226 to 199 @36V<\/td><td class=\"oth\">226 to 200 @28V<\/td><\/tr>\n<tr><td class=\"num\">148.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 3,3&amp;apos;-Dichlorobenzidine\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/91-94-1\/3,3&amp;apos;-Dichlorobenzidine\" target=\"_blank\" rel=\"noopener\">3,3&#8242;-Dichlorobenzidine<\/a><\/td><td class=\"oth\">20.25<\/td><td class=\"oth\">1<\/td><td class=\"oth\">252 to 154 @26V<\/td><td class=\"oth\">252 to 181 @24V<\/td><td class=\"oth\">252 to 182 @20V<\/td><\/tr>\n<tr><td class=\"num\">149.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 4,4&amp;apos;-Methylenebis(2-chloroaniline)\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/101-14-4\/4,4&amp;apos;-Methylenebis(2-chloroaniline)\" target=\"_blank\" rel=\"noopener\">4,4&#8242;-Methylenebis(2-chloroaniline)<\/a><\/td><td class=\"oth\">20.28<\/td><td class=\"oth\">1<\/td><td class=\"oth\">266 to 195 @28V<\/td><td class=\"oth\">231 to 195 @16V<\/td><td class=\"oth\">266 to 231 @12V<\/td><\/tr>\n<tr><td class=\"num\">150.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Chrysene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/218-01-9\/Chrysene\" target=\"_blank\" rel=\"noopener\">Chrysene<\/a><\/td><td class=\"oth\">20.3<\/td><td class=\"oth\">1<\/td><td class=\"oth\">228 to 202 @22V<\/td><td class=\"oth\">226 to 200 @28V<\/td><td class=\"oth\">228 to 201 @36V<\/td><\/tr>\n<tr><td class=\"num\">151.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Bis(2-ethylhexyl) phthalate\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/117-81-7\/Bis(2-ethylhexyl) phthalate\" target=\"_blank\" rel=\"noopener\">Bis(2-ethylhexyl) phthalate<\/a><\/td><td class=\"oth\">20.49<\/td><td class=\"oth\">1<\/td><td class=\"oth\">167 to 149 @8V<\/td><td class=\"oth\">149 to 93 @16V<\/td><td class=\"oth\">149 to 121 @14V<\/td><\/tr>\n<tr><td class=\"num\">152.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 6-Methylchrysene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/1705-85-7\/6-Methylchrysene\" target=\"_blank\" rel=\"noopener\">6-Methylchrysene<\/a><\/td><td class=\"oth\">21.19<\/td><td class=\"oth\">1<\/td><td class=\"oth\">242 to 226 @28V<\/td><td class=\"oth\">239 to 213 @26V<\/td><td class=\"oth\">239 to 237 @32V<\/td><\/tr>\n<tr><td class=\"num\">153.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Di-n-octyl phthalate\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/117-84-0\/Di-n-octyl phthalate\" target=\"_blank\" rel=\"noopener\">Di-n-octyl phthalate<\/a><\/td><td class=\"oth\">21.86<\/td><td class=\"oth\">1<\/td><td class=\"oth\">149 to 65 @22V<\/td><td class=\"oth\">149 to 93 @16V<\/td><td class=\"oth\">149 to 121 @12V<\/td><\/tr>\n<tr><td class=\"num\">154.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 7,12-Dimethylbenz[a]anthracene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/57-97-6\/7,12-Dimethylbenz[a]anthracene\" target=\"_blank\" rel=\"noopener\">7,12-Dimethylbenz[a]anthracene<\/a><\/td><td class=\"oth\">22.45<\/td><td class=\"oth\">1<\/td><td class=\"oth\">256 to 239 @38V<\/td><td class=\"oth\">241 to 226 @14V<\/td><td class=\"oth\">256 to 241 @12V<\/td><\/tr>\n<tr><td class=\"num\">155.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Benzo[b]fluoranthene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/205-99-2\/Benzo[b]fluoranthene\" target=\"_blank\" rel=\"noopener\">Benzo[b]fluoranthene<\/a><\/td><td class=\"oth\">22.46<\/td><td class=\"oth\">1<\/td><td class=\"oth\">252 to 226 @24V<\/td><td class=\"oth\">250 to 223 @38V<\/td><td class=\"oth\">250 to 224 @24V<\/td><\/tr>\n<tr><td class=\"num\">156.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Benzo[k]fluoranthene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/207-08-9\/Benzo[k]fluoranthene\" target=\"_blank\" rel=\"noopener\">Benzo[k]fluoranthene<\/a><\/td><td class=\"oth\">22.53<\/td><td class=\"oth\">1<\/td><td class=\"oth\">252 to 226 @22V<\/td><td class=\"oth\">250 to 223 @38V<\/td><td class=\"oth\">250 to 224 @26V<\/td><\/tr>\n<tr><td class=\"num\">157.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Benzo[a]pyrene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/50-32-8\/Benzo[a]pyrene\" target=\"_blank\" rel=\"noopener\">Benzo[a]pyrene<\/a><\/td><td class=\"oth\">23.21<\/td><td class=\"oth\">1<\/td><td class=\"oth\">252 to 226 @26V<\/td><td class=\"oth\">250 to 223 @38V<\/td><td class=\"oth\">250 to 224 @26V<\/td><\/tr>\n<tr><td class=\"num\">158.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Perylene-d12\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/1520-96-3\/Perylene-d12\" target=\"_blank\" rel=\"noopener\">Perylene-d12<\/a><\/td><td class=\"oth\">23.33<\/td><td class=\"oth\">1<\/td><td class=\"oth\">264 to 236 @24V<\/td><td class=\"oth\">264 to 260 @34V<\/td><td class=\"oth\">264 to 262 @20V<\/td><\/tr>\n<tr><td class=\"num\">159.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 3-Methylcholanthrene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/56-49-5\/3-Methylcholanthrene\" target=\"_blank\" rel=\"noopener\">3-Methylcholanthrene<\/a><\/td><td class=\"oth\">24.16<\/td><td class=\"oth\">1<\/td><td class=\"oth\">268 to 252 @32V<\/td><td class=\"oth\">252 to 226 @22V<\/td><td class=\"oth\">268 to 253 @14V<\/td><\/tr>\n<tr><td class=\"num\">160.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Dibenz(a,h)acridine\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/226-36-8\/Dibenz(a,h)acridine\" target=\"_blank\" rel=\"noopener\">Dibenz(a,h)acridine<\/a><\/td><td class=\"oth\">25.55<\/td><td class=\"oth\">1<\/td><td class=\"oth\">279 to 252 @34V<\/td><td class=\"oth\">278 to 250 @24V<\/td><td class=\"oth\">278 to 275 @36V<\/td><\/tr>\n<tr><td class=\"num\">161.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Dibenz[a,j]acridine\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/224-42-0\/Dibenz[a,j]acridine\" target=\"_blank\" rel=\"noopener\">Dibenz[a,j]acridine<\/a><\/td><td class=\"oth\">25.74<\/td><td class=\"oth\">1<\/td><td class=\"oth\">279 to 250 @38V<\/td><td class=\"oth\">278 to 250 @26V<\/td><td class=\"oth\">279 to 277 @30V<\/td><\/tr>\n<tr><td class=\"num\">162.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Indeno[1,2,3-cd]pyrene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/193-39-5\/Indeno[1,2,3-cd]pyrene\" target=\"_blank\" rel=\"noopener\">Indeno[1,2,3-cd]pyrene<\/a><\/td><td class=\"oth\">26.16<\/td><td class=\"oth\">1<\/td><td class=\"oth\">276 to 250 @30V<\/td><td class=\"oth\">276 to 274 @34V<\/td><td class=\"oth\">138 to 125 @12V<\/td><\/tr>\n<tr><td class=\"num\">163.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Dibenz[a,h]anthracene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/53-70-3\/Dibenz[a,h]anthracene\" target=\"_blank\" rel=\"noopener\">Dibenz[a,h]anthracene<\/a><\/td><td class=\"oth\">26.24<\/td><td class=\"oth\">1<\/td><td class=\"oth\">278 to 252 @26V<\/td><td class=\"oth\">139 to 113 @14V<\/td><td class=\"oth\">139 to 126 @8V<\/td><\/tr>\n<tr><td class=\"num\">164.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Benzo[ghi]perylene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/191-24-2\/Benzo[ghi]perylene\" target=\"_blank\" rel=\"noopener\">Benzo[ghi]perylene<\/a><\/td><td class=\"oth\">26.88<\/td><td class=\"oth\">1<\/td><td class=\"oth\">276 to 274 @38V<\/td><td class=\"oth\">138 to 124 @28V<\/td><td class=\"oth\">138 to 125 @12V<\/td><\/tr>\n<\/tbody><\/table><\/div><div class=\"chromatogram-conditions\"><h4>Conditions<\/h4><div class=\"conditions-container container-fluid\"><div class=\"row\"><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Column<\/th><td>Rxi-SVOCms, 30 m, 0.25 mm ID, 0.25 \u00b5m (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/16623\" rel=\"noopener\">cat.# 16623<\/a>)<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Standard\/Sample<\/th><td><\/td><\/tr><tr><td><\/td><td>SVOC MegaMix 150 kit  (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/31907\" rel=\"noopener\">cat.# 31907<\/a>)<\/td><\/tr><tr><td><\/td><td>SV internal standard mix (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/31206\" rel=\"noopener\">cat.# 31206<\/a>)<\/td><\/tr><tr><td><\/td><td>Acid surrogate mix (4\/89 SOW) (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/31063\" rel=\"noopener\">cat.# 31063<\/a>)<\/td><\/tr><tr><td><\/td><td>Revised B\/N surrogate mix (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/31888\" rel=\"noopener\">cat.# 31888<\/a>)<\/td><\/tr><tr><td><\/td><\/td><\/tr>\n<tr><th class=\"sub conditions_header\" scope=\"row\">Diluent:<\/th><td>Dichloromethane<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Conc.:<\/th><td>1 \u00b5g\/mL (0.1 ng on-column)<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Injection<\/th><td><\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Inj. Vol.:<\/th><td>1 \u00b5L split (split ratio 10:1)<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Liner:<\/th><td>Topaz 4.0 mm ID single taper inlet liner w\/ wool (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/23303\" rel=\"noopener\">cat.# 23303<\/a>)<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Inj. Temp.:<\/th><td>250 \u00b0C<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\" colspan=\"2\">Oven<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Oven Temp.:<\/th><td>40 &#176;C (hold 1 min) to 280 &#176;C at 12.4 &#176;C\/min to 330 &#176;C at 3.3 &#176;C\/min (hold 3.65 min)<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Carrier Gas<\/th><td>He, constant flow<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Flow Rate:<\/th><td>1.2 mL\/min<\/td><\/tr><\/table><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Detector<\/th><td>TSQ8000<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Transfer Line Temp.:<\/th><td>280 \u00b0C<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Analyzer Type:<\/th><td>Quadrupole <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Source Temp.:<\/th><td>330 \u00b0C<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Quad Temp.:<\/th><td>180 \u00b0C<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Electron Energy:<\/th><td>70 eV<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Solvent Delay Time:<\/th><td>2  min<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Tune Type:<\/th><td>PFTBA <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Ionization Mode:<\/th><td>EI <\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Instrument<\/th><td>Thermo Trace  GC<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Sample Preparation<\/th><td>Solutions were mixed and diluted to 1 ppm or lower with dichloromethane in amber 2 mL, 9 mm short-cap, screw-thread vials (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/21143\" rel=\"noopener\">cat.# 21143<\/a>) and sealed with 2.0 mL, 9 mm short-cap, screw-vial closures (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/23842\" rel=\"noopener\">cat.# 23842<\/a>).<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Notes<\/th><td>The compounds in the inserts are at lower concentrations than the full chromatogram and are showing specific transitions as noted in the inserts.<br \/><br \/>The SVOC MegaMix 150 kit (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/31907\" rel=\"noopener\">cat.# 31907<\/a>) contains one ampul each of the following standards:<br \/>\u2022 8270 MegaMix (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/31850\" rel=\"noopener\">cat.# 31850<\/a>)<br \/>\u2022 SVOC additions standard (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/31909\" rel=\"noopener\">cat.# 31909<\/a>)<br \/>\u2022 Appendix IX mix #1, revised (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/32459\" rel=\"noopener\">cat.# 32459<\/a>)<br \/>\u2022 Methapyrilene (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/32460\" rel=\"noopener\">cat.# 32460<\/a>)<br \/>\u2022 Appendix IX mix #2 (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/31806\" rel=\"noopener\">cat.# 31806<\/a>)<br \/>\u2022 Benzoic acid (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/31879\" rel=\"noopener\">cat.# 31879<\/a>)<\/td><\/tr><\/table><\/div><\/div><\/div><div class=\"chromatogram-related-skus\"><h4>Related SKUs<\/h4><ul><li>31907<\/li><li>31206<\/li><li>31063<\/li><li>31888<\/li><li>16623<\/li><li>23303<\/li><li>31907<\/li><li>31850<\/li><li>31909<\/li><li>32459<\/li><li>32460<\/li><li>31806<\/li><li>31879<\/li><li>21143<\/li><li>23842<\/li><\/ul><\/div><div class=\"chromatogram-pdf-link\"><a href=\"https:\/\/ez.restek.com\/images\/cgram\/gc_ev1522.pdf\" target=\"_blank\" rel=\"noopener noreferrer\"><svg xmlns=\"http:\/\/www.w3.org\/2000\/svg\" width=\"18\" height=\"18\" viewBox=\"0 0 18 18\"><g data-name=\"Group 2996\"><path data-name=\"Rectangle 1246\" d=\"M0 0h18v18H0z\" style=\"fill: none;\"><\/path><\/g><g data-name=\"Group 2997\"><path data-name=\"Path 729\" d=\"M13.412 11.4v2.017H5.345V11.4H4v2.017a1.349 1.349 0 0 0 1.345 1.345h8.068a1.349 1.349 0 0 0 1.345-1.345V11.4zm-.672-2.694-.948-.948-1.741 1.735V4H8.706v5.493L6.965 7.758l-.948.948 3.361 3.361z\" transform=\"translate(-.437 -.414)\" style=\"fill: rgb(13, 123, 196);\"><\/path><\/g><\/svg>Download PDF<\/a><\/div><\/div><\/div>\n\n\n<p class=\"wp-block-paragraph\">The second challenge presented by GC-MS\/MS analysis of semivolatiles using EPA Method 8270E is the response stability of reactive compounds, such as pyridine, benzoic acid, and acrylamide. These compounds exhibit a shift in retention time across different concentrations, which can make detection challenging when using SRM windows. As the concentration of active analytes decreases, the retention time increases (Figure 2). We found that the best concentration range for these compounds is 1-10 ng on-column.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">In summary, GC-MS\/MS is a great technique for detecting picograms on-column, and it significantly extends the sensitivity range for semivolatile analysis using EPA Method 8270E, albeit with some caveats. Analysts must be aware of calibration range changes, and they must also watch for retentions time shifts for reactive compounds and adjust SRM windows accordingly, if necessary.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\"><strong>Figure 2:<\/strong>\u00a0Pyridine retention time shifts with decreased concentration.<\/p>\n\n\n\n<figure class=\"wp-block-image size-large\"><img loading=\"lazy\" decoding=\"async\" width=\"1024\" height=\"689\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-EVFA4492-02-1024x689.jpg\" alt=\"\" class=\"wp-image-19690\" title=\"-\" srcset=\"https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-EVFA4492-02-1024x689.jpg 1024w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-EVFA4492-02-300x202.jpg 300w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-EVFA4492-02-768x517.jpg 768w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-EVFA4492-02-1536x1033.jpg 1536w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-EVFA4492-02.jpg 1800w\" sizes=\"auto, (max-width: 1024px) 100vw, 1024px\" \/><\/figure>\n\n\n\n<div style=\"height:50px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n\n\n        <div class=\"cpb\">\n            <h3 class=\"cpb-heading\">Erw\u00e4hnte Produkte<\/h3>\n            <hr class=\"cpb-heading-underline\" \/>\n            <div class=\"cpb-list\">\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/de.restek.com\/p\/16623\">                                Art.-Nr. 16623                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">Rxi-SVOCms GC Kapillars\u00e4ule, 30 m, 0.25 mm ID, 0.25 \u00b5m<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/de.restek.com\/p\/16623\">Produkt anzeigen<\/a>\n                                                    <\/div>\n                    <\/div>\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/de.restek.com\/p\/23303\">                                Art.-Nr. 23303                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">Topaz, Single Taper Inlet Liner, 4.0 mm x 6.5 x 78.5, f\u00fcr Agilent GCs, mit Quarzwolle, Premium-Deaktivierung, 5er Pack<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/de.restek.com\/p\/23303\">Produkt anzeigen<\/a>\n                                                    <\/div>\n                    <\/div>\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/de.restek.com\/p\/31907\">                                Art.-Nr. 31907                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">SVOC MegaMix 150 Kit: 1 mL\/Ampulle; 6 Ampullen\/Kit<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/de.restek.com\/p\/31907\">Produkt anzeigen<\/a>\n                                                    <\/div>\n                    <\/div>\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/de.restek.com\/p\/31206\">                                Art.-Nr. 31206                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">SV Interner Standard Mix, 2000 \u03bcg\/mL, Methylenchlorid, 1 mL\/Ampulle<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/de.restek.com\/p\/31206\">Produkt anzeigen<\/a>\n                                                    <\/div>\n                    <\/div>\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/de.restek.com\/p\/31063\">                                Art.-Nr. 31063                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">S\u00e4ure-Surrogat Mix (4\/89 SOW), 10.000 \u03bcg\/mL, Methanol, 1 mL\/Ampulle<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/de.restek.com\/p\/31063\">Produkt anzeigen<\/a>\n                                                    <\/div>\n                    <\/div>\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/de.restek.com\/p\/31888\">                                Art.-Nr. 31888                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">Revidiertes B\/N-Surrogatgemisch, 5000 \u03bcg\/mL, Methylenchlorid, 1 mL\/Ampulle<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/de.restek.com\/p\/31888\">Produkt anzeigen<\/a>\n                                                    <\/div>\n                    <\/div>\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/de.restek.com\/p\/21143\">                                Art.-Nr. 21143                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">Short-Cap-Vial mit graduiertem Markierungsfeld, 9-425 Screw-Thread, 2.0 mL, 9 mm, 12 x 32 (nur Vial), braun, 1000er Pack<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/de.restek.com\/p\/21143\">Produkt anzeigen<\/a>\n                                                    <\/div>\n                    <\/div>\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/de.restek.com\/p\/23842\">                                Art.-Nr. 23842                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">Short Screw Kappen, Polypropylen, gerippt, Screw-Thread, PTFE\/Silikon-beschichtet, f\u00fcr Agilent 7693A, blau, vormontiert, 2.0 mL, 9 mm, 1000er Pack<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/de.restek.com\/p\/23842\">Produkt anzeigen<\/a>\n                                                    <\/div>\n                    <\/div>\n                            <\/div>\n        <\/div>\n        \n\n\n<div style=\"height:100px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n\n\n\n<p class=\"wp-block-paragraph\"><\/p>\n","protected":false},"excerpt":{"rendered":"<p>The addition of a GC-MS\/MS option to EPA Method 8270E significantly improves sensitivity for semivolatiles, but it also presents challenges. Here, we explore some of the issues that analysts need to consider.<\/p>\n","protected":false},"author":11,"featured_media":6739,"comment_status":"closed","ping_status":"open","sticky":false,"template":"","format":"standard","meta":{"_acf_changed":false,"_kad_blocks_custom_css":"","_kad_blocks_head_custom_js":"","_kad_blocks_body_custom_js":"","_kad_blocks_footer_custom_js":"","_kadence_starter_templates_imported_post":false,"_kad_post_transparent":"","_kad_post_title":"","_kad_post_layout":"","_kad_post_sidebar_id":"","_kad_post_content_style":"","_kad_post_vertical_padding":"","_kad_post_feature":"","_kad_post_feature_position":"","_kad_post_header":false,"_kad_post_footer":false,"footnotes":"","_ppma_block_editor_authors":""},"categories":[7,1],"tags":[2625],"industries-application":[],"post-badge":[],"resource-type":[2623],"product-library":[],"resource-technique":[],"ppma_author":[414],"class_list":["post-56953","post","type-post","status-publish","format-standard","has-post-thumbnail","hentry","category-articles","category-uncategorized","tag-epa-method-8270e","resource-type-featured-application"],"acf":[],"taxonomy_info":{"category":[{"value":7,"label":"Articles"},{"value":1,"label":"Uncategorized"}],"post_tag":[{"value":2625,"label":"EPA Method 8270E"}],"resource-type":[{"value":2623,"label":"Featured Application"}]},"featured_image_src_large":["https:\/\/discover.restek.com\/wp-content\/uploads\/feature-EVFA4492-1024x536.jpg",1024,536,true],"author_info":{"display_name":"Restek Corporation","author_link":"https:\/\/discover.restek.com\/de\/author\/restek-corporation\/"},"comment_info":0,"category_info":[{"term_id":7,"name":"Articles","slug":"articles","term_group":0,"term_taxonomy_id":7,"taxonomy":"category","description":"","parent":0,"count":510,"filter":"raw","cat_ID":7,"category_count":510,"category_description":"","cat_name":"Articles","category_nicename":"articles","category_parent":0},{"term_id":1,"name":"Uncategorized","slug":"uncategorized","term_group":0,"term_taxonomy_id":1,"taxonomy":"category","description":"","parent":0,"count":510,"filter":"raw","cat_ID":1,"category_count":510,"category_description":"","cat_name":"Uncategorized","category_nicename":"uncategorized","category_parent":0}],"tag_info":[{"term_id":2625,"name":"EPA Method 8270E","slug":"epa-method-8270e","term_group":0,"term_taxonomy_id":2625,"taxonomy":"post_tag","description":"","parent":0,"count":1,"filter":"raw"}],"authors":[{"term_id":414,"user_id":11,"is_guest":0,"slug":"restek-corporation","display_name":"Restek Corporation","avatar_url":{"url":"https:\/\/discover.restek.com\/wp-content\/uploads\/Restek_Favicon_300x300.jpg","url2x":"https:\/\/discover.restek.com\/wp-content\/uploads\/Restek_Favicon_300x300.jpg"},"author_category":"1","first_name":"Restek","last_name":"Corporation","user_url":"https:\/\/www.linkedin.com\/company\/restek-corporation\/posts\/?feedView=all","job_title":"Chromatography Products and Solutions","description":"Restek is a leading provider of chromatography columns, accessories, and certified reference materials. Trust Restek for reliable, high-quality analytical solutions."}],"_links":{"self":[{"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/posts\/56953","targetHints":{"allow":["GET"]}}],"collection":[{"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/posts"}],"about":[{"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/types\/post"}],"author":[{"embeddable":true,"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/users\/11"}],"replies":[{"embeddable":true,"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/comments?post=56953"}],"version-history":[{"count":3,"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/posts\/56953\/revisions"}],"predecessor-version":[{"id":56985,"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/posts\/56953\/revisions\/56985"}],"wp:featuredmedia":[{"embeddable":true,"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/media\/6739"}],"wp:attachment":[{"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/media?parent=56953"}],"wp:term":[{"taxonomy":"category","embeddable":true,"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/categories?post=56953"},{"taxonomy":"post_tag","embeddable":true,"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/tags?post=56953"},{"taxonomy":"industries-application","embeddable":true,"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/industries-application?post=56953"},{"taxonomy":"post-badge","embeddable":true,"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/post-badge?post=56953"},{"taxonomy":"resource-type","embeddable":true,"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/resource-type?post=56953"},{"taxonomy":"product-library","embeddable":true,"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/product-library?post=56953"},{"taxonomy":"resource-technique","embeddable":true,"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/resource-technique?post=56953"},{"taxonomy":"author","embeddable":true,"href":"https:\/\/discover.restek.com\/de\/wp-json\/wp\/v2\/ppma_author?post=56953"}],"curies":[{"name":"wp","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}