What ferrules do I use with my column?

01 Aug 2025

one-piece ferrules

One-piece ferrules are available in two designs: the standard 116’’ compression type and the short, compact design. Agilent GCs typically use the compact design at the injector and detector, but we always recommend using the 116’’ compression type ferrule at the detector if you are doing GC-MS analysis. Whichever configuration you select, it is essential that you use the corresponding standard or compact nut.

In addition, ferrules are generally available in three compositions: graphite, Vespel, and Vespel/graphite. Composition is usually a matter of preference, but we recommend Vespel/graphite ferrules for GC-MS because they seal with minimum torque, maintain sealing performance under high pressure and vacuum, and are reusable.

The hole in the ferrule should always have a slightly larger inner diameter (ID) than the column’s outer diameter (OD) so that it can compress and seal around the column. For example, a 0.53 mm ID Restek column, which has an OD of approximately 0.7 mm, would require a ferrule with an ID of 0.8 mm.

For more on ferrule selection, see our blogs on ferrule selection and our GC nut/ferrule compatibility chart. You’ll find specific information there for all the major instrument brands.