- Analyze arylamines 3.3x faster than conventional GC-MS.
- Cut costs by reducing helium consumption 81%.
- Pre-connected LPGC column kit prevents leaks, simplifies installation, and allows large-volume injection.
Azo dyes are synthetically produced colorants that are commonly used in the food, cosmetic, pharmaceutical, and textile industries. They are present in a wide variety of consumer products, and some can produce carcinogenic or mutagenic arylamines as breakdown products. Due to concerns about human health and exposure, effective methods are needed for the analysis of arylamines. Testing is usually performed by conventional GC-MS or LC-MS, but long analysis times limit sample throughput.
A much faster alternative, low-pressure GC-MS (LPGC-MS) is a paired-column technique that can bring significant speed gains to labs analyzing arylamines. This approach uses the MS vacuum to lower pressure within the analytical column, thus increasing speed. For example, in the analysis of arylamines (22 of which are regulated by EU REACH legislation [1]) shown in Figure 1, the run time was 3.3x faster than the conventional method, and it also used 81% less helium, which provides a substantial cost savings.
LPGC-MS only works because the inlet pressure is maintained by pairing the analytical column to a narrow restrictor column. However, due to the difference in tubing diameters, it can be difficult to make a robust connection manually, which has limited adoption of the technique. LPGC column kits from Restek solve this problem because they are factory connected and individually leak checked, which ensures a leak-free connection and simplifies setup. Visit www.restek.com/lpgc to learn more about this powerful technique.
References
- Regulation (EC) No. 1907/2006 of the European Parliament and of the Council of 18 December 2006 Concerning the Registration, Evaluation, Authorization and Restriction of Chemicals (REACH). https://osha.europa.eu/en/legislation/directives/regulation-ec-no-1907-2006-of-the-european-parliament-and-of-the-council.

GC_GN1244
Peaks
| Peaks | tR (30 m) | tR (LPGC) | Conc. (ppm) | Quant Ion | |
|---|---|---|---|---|---|
| 1. | o-Toluidine | 3.217 | 1.207 | 2 | 107 |
| 2. | o-Anisidine | 4.42 | 1.589 | 2 | 123 |
| 3. | 4-Chloroaniline | 4.832 | 1.71 | 2 | 127 |
| 4. | p-Cresidine | 5.519 | 1.92 | 2 | 137 |
| 5. | 2,4,5-Trimethylaniline | 5.582 | 1.935 | 2 | 135 |
| 6. | 3-Chloro-o-toluidine | 5.902 | 2.026 | 2 | 141 |
| 7. | 4-Chloro-o-toluidine | 6.025 | 2.061 | 2 | 141 |
| 8. | 2,4-Diaminotoluene | 7.637 | 2.523 | 2 | 122 |
| 9. | 2,4-Diaminoanisole | 8.89 | 2.872 | 2 | 138 |
| 10. | 2-Naphthylamine | 9.773 | 3.112 | 2 | 143 |
| 11. | 2-Aminobiphenyl | 9.892 | 3.148 | 2 | 169 |
| 12. | 2-Amino-4-nitrotoluene | 10.537 | 3.325 | 2 | 152 |
| Peaks | tR (30 m) | tR (LPGC) | Conc. (ppm) | Quant Ion | |
|---|---|---|---|---|---|
| 13. | 4-Aminobiphenyl | 12.154 | 3.767 | 2 | 169 |
| 14. | p-Aminoazobenzene | 15.95 | 4.805 | 2 | 197 |
| 15. | 4,4′-Oxydianiline | 16.613 | 4.99 | 2 | 200 |
| 16. | 4,4′-Diaminodiphenylmethane | 16.714 | 5.016 | 2 | 198 |
| 17. | Benzidine | 16.787 | 5.034 | 2 | 184 |
| 18. | o-Aminoazotoluene | 17.53 | 5.24 | 2 | 225 |
| 19. | 3,3′-Dimethyl-4,4′-diaminodiphenylmethane | 18.23 | 5.436 | 2 | 226 |
| 20. | 3,3′-Dimethylbenzidine | 18.483 | 5.504 | 2 | 212 |
| 21. | 4,4’-Thiodianiline | 19.439 | 5.765 | 2 | 216 |
| 22. | 3,3′-Dichlorobenzidine | 19.795 | 5.864 | 2 | 252 |
| 23. | 4,4′-Methylenebis(2-chloroaniline) | 19.795 | 5.864 | 2 | 266 |
| 24. | 3,3′-Dimethoxybenzidine | 19.905 | 5.898 | 2 | 244 |
Conditions
| Column | See notes |
|---|---|
| Standard/Sample | |
| AccuStandard carcinogenic aryl amine mix (AE-00049-R1-10X) | |
| AccuStandard 2,4-diaminoanisole (ALR-070S-R2) | |
| Diluent: | Ethyl acetate |
| Conc.: | 20 µg/mL |
| Injection | |
| Inj. Vol.: | 1 µL split (split ratio 10:1) |
| Liner: | Topaz, splitless, single taper inlet liner, 4.0 mm x 6.5 x 78.5 (cat.# 23303) |
| Inj. Temp.: | 280 °C |
| Carrier Gas | He |
| Detector | MS |
|---|---|
| Mode: | Scan |
| Transfer Line Temp.: | 300 °C |
| Analyzer Type: | Quadrupole |
| Source Temp.: | 230 °C |
| Quad Temp.: | 150 °C |
| Electron Energy: | 70 eV |
| Tune Type: | PFTBA |
| Ionization Mode: | EI |
| Instrument | Agilent 7890B GC & 5977A MSD |
| Sample Preparation | The standards were diluted with ethyl acetate to 20 ppm; analyzed in a 2 mL, short-cap, screw-thread vial (cat.# 21143); and capped with a short-cap, screw-vial closure (cat.# 24495). |
| Notes | Conventional (30 m) Analysis: Column: Rxi-35Sil MS, 30 m, 0.25 mm ID, 0.25 µm (cat.# 13823) Temp. program: 200 °C (hold 0.5 min) to 320 °C at 9.5 °C/min (hold 5 min) Flow: 2.0 mL/min Scan start time: 2 min Scan range: 30-300 amu Scan rate: 10 scans/sec LPGC-MS Analysis: Column: LPGC Rxi-35Sil MS column kit, includes 10 m x 0.32 mm ID x 0.25 μm Rxi-35Sil MS analytical column and 5 m x 0.15 mm ID Rxi restrictor factory connected via SilTite connector (cat.# 11806) Temp. program: 100 °C (hold 0.5 min) to 300 °C at 35 °C/min (hold 5 min) Flow: 0.9 mL/min Scan start time: 1 min Scan range: 35-300 amu Scan rate: 9.7 scans/sec Compound list based on EU legislation of Registration, Evaluation, Authorisation and Restriction of Chemicals (REACH) Appendix 8 (https://reachonline.eu/reach/en/appendix-8.html). The 3-chloro-o-toluidine and 2-aminibiphenyl compounds are not part of the list. |