When starting method development for potency testing, it’s important to choose the right column dimension for the target analysis. In this work, different column dimensions of the Raptor ARC-18 phase were utilized to develop methods to meet various labs’ needs using HPLC-UV. To demonstrate the powerful resolving capabilities of Raptor ARC-18, a 50 x 3 mm, 2.7 µm column was used to analyze 7 cannabinoids including CBD, CBDA, delta-9-THC, delta-8-THC, (6aR, 9S)-delta-10-THC, (6aR, 9R)-delta-10-THC, and THCA. Next, additional cannabinoids including CBDV, THCV, CBG, CBN, CBGA, and CBC were added to the previous analytes for a total of 13 cannabinoids. Using the same column dimension and mobile phases, a method was developed to resolve all analytes in 10 minutes. Finally, to include exo-THC and CBNA, a 150 x 3 mm, 2.7 µm column dimension was used to demonstrate the utility of a longer column dimension. Each of these methods was applied to hemp matrix to demonstrate the applicability of these methods in real-world samples.
Cannabis Potency Testing—Which Column Dimension is Right for You?
05 Nov 2025
GNOT5282
