Sin categorizar

Alcoholic Beverage Analysis by GC

Increase Accuracy with Rtx-1301 Column Selectivity

18 Nov 2025

feature FFAR4494

Introduction

The volatile component profiles of alcoholic beverages consist of a wide range of compounds, including acids, alcohols, aldehydes, and other trace-level flavor compounds. Analysts trained in the sensory evaluation of distilled liquors, wines, or beers tell us no two products are exactly alike. The unique sensory properties of different types and brands of distilled liquor products often are due to minor differences among the volatile components present. By using instrumental methods to qualitatively or quantitatively evaluate these differences, in addition to sensory techniques, quality assurance analysts can obtain a wealth of information about their products.

Gas chromatography (GC) is not only a powerful tool in the analysis of alcoholic beverages, but it is also a somewhat easy technique to use for the following reasons. Minimal sample preparation is required because the samples are analyzed in their original liquid state with alcohol or alcohol/water as the matrix. The flavor compounds tend to be volatile in nature, which fulfills one of the main requirements of GC. General detectors, such as the flame ionization detector (FID), or more information-rich detectors, such as the mass selective detector (MSD), can be used. Additionally, the ability to automate the analysis makes GC a very practical tool in a QA/QC environment. In this article, we will discuss how GC can be used to monitor the volatile components of alcoholic beverages.

Analysis of Alcohols and Aldehydes in Alcoholic Beverages

Alcoholic beverages contain a wide range of volatile compounds, including alcohols and short-chain aldehydes. GC can be used to analyze these compounds without requiring sample extraction. For example, AOAC International has published methods for the analysis of fusel alcohols (also known as fusel oils), methanol, ethanol, and higher order alcohols by GC [1].

Since capillary columns offer efficient separations, capillary GC is especially useful in the analysis of structurally similar compounds, such as fusel alcohols. However, selecting a GC column stationary phase with appropriate selectivity is essential for obtaining good peak shapes and excellent resolution of a range of alcohols. The two predominant fusel alcohols, 2-methyl-1-butanol (isoamyl alcohol) and 3-methyl-1-butanol (active amyl alcohol), are critical target compounds that must be resolved. Also, since alcoholic beverage analysis involves liquid samples that are predominantly water/ethanol-based, water, acetaldehyde, methanol, and ethanol must also be separated to avoid signal suppression during sample analysis. Acetic acid presents an additional challenge because it can exhibit poor peak shapes on some stationary phases.

Tech Tip
Use a gas-tight syringe to maximize syringe lifetime with ethanol samples.

Choosing a GC Column with the Right Selectivity

To determine which GC column is best overall to use for alcoholic beverage analysis, three stationary phases with different polarity were evaluated: Rtx-502.2 (low-polarity); Rtx-VMS (unique polarity for solvents and volatiles); and Rtx-1301 (mid-polarity). Particular attention was paid to peak shapes and the resolution of specific compounds.

A standard containing many of the target compounds commonly tested for in alcoholic beverage analysis was prepared at 100 µg/mL in ethanol. Initial screening was conducted by FID. MSD analysis was then performed with a full scan method (10-400 m/z) to ensure that water was adequately resolved.

All three column phases yielded good separation of water, acetaldehyde, methanol, and ethanol. The Rtx-1301 column chromatogram shown in Figure 1 is representative of a typical separation of these compounds.

Figure 1: Typical Water/Acetaldehyde Separation on an Rtx-1301 Column Using GC-MS.

Rtx-1301 Water, Acetaldehyde, and Methanol Separation

GC_FF1389

Peaks

PeakstR (min)Conc.
(µg/mL)
1.Water4.312
2.Acetaldehyde4.483100
3.Methanol4.699100
4.Ethanol5.871100

Conditions

ColumnRtx-1301, 60 m, 0.25 mm ID, 1.00 µm (cat.# 16056)
Standard/SampleCustom acetaldehyde, methanol, and ethanol standard in water
Diluent:Water
Conc.:100 µg/mL
Injection
Inj. Vol.:1 µL split (split ratio 100:1)
Liner:Topaz Precision inlet liner with wool, 4.0 mm x 6.3 x 78.5 (cat.# 23305)
Inj. Temp.:250 °C
Split Vent Flow Rate:141.4 mL/min
Oven
Oven Temp.:35 °C (hold 2.45 min) to 220 °C at 7 °C/min (hold 5 min)
Carrier GasHe, constant flow
Flow Rate:1.4 mL/min
Linear Velocity:30.261 cm/sec @ 35 °C
DetectorMS
Mode:Scan
Scan Program:
GroupStart Time
(min)
Scan Range
(amu)
Scan Rate
(scans/sec)
1010-4006.84
Transfer Line Temp.:280 °C
Analyzer Type:Quadrupole
Source Temp.:230 °C
Quad Temp.:150 °C
Electron Energy:70 eV
Solvent Delay Time:0.0 min
Tune Type:PFTBA
Ionization Mode:EI
InstrumentAgilent 7890A GC & 5975C MSD
NotesThe EZGC method translator was used to establish the MS conditions.
Did You Know?
Fusel alcohols are higher order (i.e., secondary or tertiary) alcohols, traces of which usually are present in all beers. They are produced through a pathway very similar to the pathway for ethanol, the preferred alcohol for beer. Fusel alcohols contribute a hot, spicy, solvent-like flavor and an alcohol “burn.” Small amounts of these components can be desirable in a strong ale or barley wine, but they can be offensive and therefore are unwanted in a Pilsner or other lager. In addition to their influence on flavor, they usually cause low carbonation and poor head retention in bottle-conditioned beers because they are deadly to yeast. Higher fermentation temperatures can produce excessively rapid yeast growth and yeast mutations, which, in turn, stimulate the formation of these components.
picture of beer bottles

The first phase evaluated for resolution and peak shape was the Rtx-VMS phase in a 30 m x 0.25 mm ID x 1.40 µm column format (cat.# 19915). While the separation of water, acetaldehyde, and methanol was adequate (similar to the Rtx-1301 column), the Rtx-VMS column does not resolve 2-methyl-1-butanol (isoamyl alcohol) and 3-methyl-1-butanol (active amyl alcohol) as shown in Figure 2. Even with the oven set to 35 °C and held isothermally, the amyl alcohols coelute regardless of the Rtx-VMS column dimensions.

Figure 2: Common alcoholic beverage target compounds 2-methyl-1-butanol and 3-methyl-1-butanol are not resolved on an Rtx-VMS column (peaks 12 and 13).

Custom Beverage Analysis Mix on Rtx-VMS

GC_FF1385

Peaks

PeakstR (min)Conc.
(µg/mL)
1.Acetaldehyde2.607100
2.Methyl alcohol2.685100
3.Ethanol3.917
4.Acetone4.521100
5.Ethyl formate4.627100
6.1-Propanol6.040100
7.Ethyl acetate6.874100
8.2-Butanol6.959100
PeakstR (min)Conc.
(µg/mL)
9.1-Propanol, 2-methyl-7.799100
10.Acetic acid8.497100
11.Ethane, 1,1-diethoxy-9.112100
12.3-Methyl-1-butanol 10.614100
13.2-Methyl-1-butanol 10.632100
14.1-Pentanol11.328100
15.Propylene glycol12.375100
16.Isoamyl acetate13.478100
17.Furfural13.839100
Final on-column concentration is 1 ppm due to 100:1 split. Ethanol is the diluent solvent.

Conditions

ColumnRtx-VMS, 30 m, 0.25 mm ID, 1.40 µm (cat.# 19915)
Standard/SampleCustom beverage analysis mix
Diluent:Ethanol
Conc.:100 µg/mL Final on-column concentration is 1 ppm.
Injection
Inj. Vol.:1 µL split (split ratio 100:1)
Liner:Topaz, precision inlet liner, 4.0 mm x 6.3 x 78.5, for Agilent GCs, w/quartz wool (cat.# 23305)
Inj. Temp.:250 °C
Split Vent Flow Rate:144.4 mL/min
Oven
Oven Temp.:35 °C (hold 3 min) to 220 °C at 8 °C/min (hold 5 min)
Carrier GasHe, constant flow
Flow Rate:1.4 mL/min
Linear Velocity:31.993 cm/sec @ 35 °C
DetectorFID @ 250 °C
Make-up Gas Flow Rate:25 mL/min
Make-up Gas Type:N2
Hydrogen flow:35 mL/min
Air flow:400 mL/min
Data Rate:50 Hz
InstrumentAgilent 7890A GC
Sample PreparationSingle component compounds combined to make final custom mix in ethanol.

The Rtx-502.2 phase in a 60 m x 0.25 mm ID x 1.40 µm column format (cat.# 10916) offered excellent resolution and an acceptable analysis time for all target compounds. However, the acetic acid peak shape shows fronting, which indicates poor solubility of acetic acid in the low-polarity stationary phase (Figure 3).

Figure 3: Acetic acid exhibits poor peak shape on an Rtx-502.2 column, indicating poor solubility in the stationary phase (peak 7).

Custom Beverage Analysis Mix on Rtx-502.2

GC_FF1387

Peaks

PeakstR (min)Conc.
(µg/mL)
1.Acetaldehyde5.785100
2.Methanol5.508100
3.Ethanol6.971100
4.Acetone8.103100
5.Ethyl formate8.523100
6.1-Propanol9.559100
7.Acetic acid10.244100
8.2-Butanol10.750100
PeakstR (min)Conc.
(µg/mL)
9.Ethyl acetate11.318100
10.Isobutanol11.544100
11.Acetal14.260100
12.3-Methyl-1-butanol 14.730100
13.2-Methyl-1-butanol 14.850100
14.Propylene glycol15.670100
15.n-Amyl alcohol16.286100
16.Isoamyl acetate18.511100
17.Furfural18.708100

Conditions

ColumnRtx-502.2, 60 m, 0.25 mm ID, 1.4 µm (cat.# 10916)
Standard/SampleCustom beverage analysis
Diluent:Ethanol
Conc.:100 µg/mL 100:1 split
Injection
Inj. Vol.:1.0 µL split (split ratio 100:1)
Liner:Topaz precision split liner (cat.# 23305)
Inj. Temp.:250 °C
Split Vent Flow Rate:140 mL/min
Oven
Oven Temp.:35 °C (hold 3 min) to 220 °C at 8 °C/min (hold 5 min)
Carrier GasHe, constant flow
Flow Rate:1.4 mL/min
Linear Velocity:25.503 cm/sec @ 35 °C
DetectorFID @ 250 °C
Make-up Gas Flow Rate:25 mL/min
Make-up Gas Type:N2
Hydrogen flow:35 mL/min
Air flow:400 mL/min
Data Rate:50 Hz
InstrumentAgilent 7890A GC

The last column evaluated was an Rtx-1301 phase in a 60 m x 0.25 mm ID x 1.00 µm format (cat.# 16056). This column offered the best overall chromatography, particularly in terms of peak shape and resolution of the critical pairs. Importantly, the 2-methyl-1-butanol and 3-methyl-1-butanol peaks are adequately resolved. In addition, by using the higher polarity Rtx-1301 stationary phase (a cyanopropylphenyl-type polysiloxane phase) compared to the lower polarity Rtx-502.2 (a phenylmethyl-type polysiloxane phase), the acetic acid peak shape was much improved (Figure 4). This occurs because acetic acid is more soluble in the Rtx-1301 phase. Because the Rtx-1301 column produced the best overall chromatography in less than 20 minutes, it is the recommended column for GC analysis of alcoholic beverages.

Figure 4: The Rtx-1301 column shows the best overall chromatography for targeted alcoholic beverage compounds.

Custom Beverage Analysis Mix on Rtx-1301

GC_FF1388

Peaks

PeakstR (min)Conc.
(µg/mL)
1.Acetaldehyde5.230100
2.Methyl alcohol5.431100
3.Ethanol6.911100
4.Acetone7.476100
5.Ethyl formate7.674100
6.1-Propanol9.517100
7.Ethyl acetate10.400100
8.2-Butanol10.620100
PeakstR (min)Conc.
(µg/mL)
9.Isobutanol11.590100
10.Acetic acid11.753100
11.Acetal13.517100
12.3-Methyl-1-butanol 14.884100
13.2-Methyl-1-butanol 14.975100
14.1-Pentanol15.802100
15.Propylene glycol16.490100
16.Isoamyl acetate18.246100
17.Furfural18.363100

Conditions

ColumnRtx-1301, 60 m, 0.25 mm ID, 1.0 µm (cat.# 16056)
Standard/SampleCustom beverage analysis standard
Diluent:Ethanol
Conc.:100 µg/mL
Injection
Inj. Vol.:1 µL split (split ratio 100:1)
Liner:Topaz precision split liner (cat.# 23305)
Inj. Temp.:250 °C
Split Vent Flow Rate:140 mL/min
Oven
Oven Temp.:35 °C (hold 3 min) to 220 °C at 7 °C/min (hold 5 min)
Carrier GasHe, constant flow
Flow Rate:1.4 mL/min
Linear Velocity:25.457 cm/sec @ 35 °C
DetectorFID @ 250 °C
Make-up Gas Flow Rate:25 mL/min
Make-up Gas Type:N2
Hydrogen flow:35 mL/min
Air flow:400 mL/min
Data Rate:50 Hz
InstrumentAgilent 7890A GC

Alcoholic Beverage Sample Analysis

To confirm the performance of the Rtx-1301 column for targeted alcoholic beverage compounds, real-world alcohol samples were analyzed. Data was collected with both an FID and an MS detector with the method initially developed by GC-FID and then transferred to GC-MS. To ease this transition, Restek’s EZGC method translator was used. The translation provided by this tool proved to be an excellent starting point for the MS method—it was generated in seconds and needed only minor optimization and adjustment of the temperature program to ensure that isobutanol and acetic acid were adequately resolved. For example, Figures 5 and 6 show the same chromatography for rye whiskey analysis by GC-FID and GC-MS, respectively; note the excellent resolution between the isobutanol and acetic acid as well as the separation of the 2-methyl-1-butanol and 3-methyl-1-butanol peaks. The acetic acid peak shape is satisfactory, and the analysis time is less than 15 minutes.

Figure 5: Rye whiskey by GC-FID on an Rtx-1301 column exhibits excellent resolution, satisfactory acetic acid peak shape, and a fast <15-minute analysis time.

Rye Whiskey on Rtx-1301

GC_FF1395

Peaks

PeakstR (min)
1.Acetaldehyde5.405
2.Methanol5.631
3.Ethanol7.032
4.1-Propanol9.489
5.Ethyl acetate10.318
6.Isobutanol11.385
7.Acetic acid11.472
8.Acetal13.137
9.3-Methyl-1-butanol14.334
10.2-Methyl-1-butanol14.419

Conditions

ColumnRtx-1301, 60 m, 0.25 mm ID, 1.0 µm (cat.# 16056)
Standard/SampleRye whiskey (neat)
Diluent:n/a
Injection
Inj. Vol.:1 µL split (split ratio 100:1)
Liner:Topaz precision inlet liner, 4.0 mm x 6.3 x 78.5, with wool (cat.# 23305)
Inj. Temp.:250 °C
Split Vent Flow Rate:144.4 mL/min
Oven
Oven Temp.:35 °C (hold 3 min) to 220 °C at 8 °C/min (hold 5 min)
Carrier GasHe, constant flow
Flow Rate:1.4 mL/min
Linear Velocity:25.457 cm/sec @ 35 °C
DetectorFID @ 250 °C
Make-up Gas Flow Rate:25 mL/min
Make-up Gas Type:N2
Hydrogen flow:35 mL/min
Air flow:400 mL/min
Data Rate:50 Hz
InstrumentAgilent 7890A GC

Figure 6: GC-MS analysis of rye whiskey shows the same chromatography as the GC-FID analysis in Figure 5.

Rye Whiskey on Rtx-1301

GC_FF1396

Peaks

PeakstR (min)
1.Acetaldehyde4.493
2.Methanol4.684
3.Ethanol5.988
4.1-Propanol8.392
5.Ethyl acetate9.234
6.Isobutanol10.373
7.Acetic acid10.521
8.Acetal12.265
9.3-Methyl-1-butanol 13.595
10.2-Methyl-1-butanol 13.667

Conditions

ColumnRtx-1301, 60 m, 0.25 mm ID, 1.0 µm (cat.# 16056)
Standard/SampleRye whiskey
Diluent:n/a
Injection
Inj. Vol.:1 µL split (split ratio 100:1)
Liner:Topaz precision inlet liner w/wool, 4.0 mm x 6.3 x 78.5 (cat.# 23305)
Inj. Temp.:250 °C
Split Vent Flow Rate:140 mL/min
Oven
Oven Temp.:35 °C (hold 2.45 min) to 220 °C at 7 °C/min (hold 5 min)
Carrier GasHe, constant flow
Flow Rate:1.4 mL/min
Linear Velocity:30.261 cm/sec @ 35 °C
DetectorMS
Mode:Scan
Transfer Line Temp.:280 °C
Analyzer Type:Quadrupole
Source Temp.:230 °C
Quad Temp.:150 °C
Electron Energy:70 eV
Solvent Delay Time:0.0 min
Tune Type:PFTBA
Ionization Mode:EI
InstrumentAgilent 7890A GC & 5975C MSD

Additives are commonly encountered when analyzing liquors. In fact, up to 2% of any spirit can be unreported and not disclosed on the label, so unlabeled additives are frequently present. For example, propylene glycol and glycerin are sometimes added to liquor. Rum may have propylene glycol added to give the product a mouthfeel similar to an aged rum. Similarly, glycerin may be added for smoothness. Some of these compounds appeared in the range of liquors that were analyzed in Figures 7-11. Note that for these real-world samples, the Rtx-1301 column again provided excellent resolution of critical compounds, good peak shape for acetic acid, and fast analysis times, making it an outstanding column choice for alcoholic beverage analysis by GC.

Figure 7: Brandy analysis using an Rtx-1301 column and GC-FID shows excellent resolution.

Brandy on Rtx-1301

GC_FF1391

Peaks

PeakstR (min)
1.Acetaldehyde5.449
2.Methanol5.634
3.Ethanol7.028
4.1-Propanol9.495
5.Ethyl acetate10.321
6.Isobutanol11.387
7.Acetal13.137
8.3-Methyl-1-butanol14.333
9.2-Methyl-1-butanol14.420

Conditions

ColumnRtx-1301, 60 m, 0.25 mm ID, 1.0 µm (cat.# 16056)
Standard/SampleBrandy (neat)
Diluent:n/a
Injection
Inj. Vol.:1 µL split (split ratio 100:1)
Liner:Topaz precision inlet liner, 4.0 mm x 6.3 x 78.5, with wool (cat.# 23305)
Inj. Temp.:250 °C
Split Vent Flow Rate:144.4 mL/min
Oven
Oven Temp.:35 °C (hold 3 min) to 220 °C at 8 °C/min (hold 5 min)
Carrier GasHe, constant flow
Flow Rate:1.4 mL/min
Linear Velocity:25.457 cm/sec @ 35 °C
DetectorFID @ 250 °C
Make-up Gas Flow Rate:25 mL/min
Make-up Gas Type:N2
Hydrogen flow:35 mL/min
Air flow:400 mL/min
Data Rate:50 Hz
InstrumentAgilent 7890A GC

Figure 8: Sherry analysis using an Rtx-1301 column and GC-FID not only shows excellent resolution but also exhibits good peak shape for the additive propylene glycol.

Sherry on Rtx-1301

GC_FF1392

Peaks

PeakstR (min)
1.Acetaldehyde5.405
2.Methanol5.636
3.Ethanol6.963
4.1-Propanol9.490
5.Ethyl acetate10.319
6.Isobutanol11.385
7.3-Methyl-1-butanol14.333
8.2-Methyl-1-butanol14.419
9.Propylene glycol15.752
10.Furfural17.486

Conditions

ColumnRtx-1301, 60 m, 0.25 mm ID, 1.0 µm (cat.# 16056)
Standard/SampleSherry (neat)
Diluent:n/a
Injection
Inj. Vol.:1 µL split (split ratio 100:1)
Liner:Topaz precision inlet liner, 4.0 mm x 6.3 x 78.5, with wool (cat.# 23305)
Inj. Temp.:250 °C
Split Vent Flow Rate:144.4 mL/min
Oven
Oven Temp.:35 °C (hold 3 min) to 220 °C at 8 °C/min (hold 5 min)
Carrier GasHe, constant flow
Flow Rate:1.4 mL/min
Linear Velocity:25.457 cm/sec @ 35 °C
DetectorFID @ 250 °C
Make-up Gas Flow Rate:25 mL/min
Make-up Gas Type:N2
Hydrogen flow:35 mL/min
Air flow:400 mL/min
Data Rate:50 Hz
InstrumentAgilent 7890A GC

Figure 9: Irish whiskey analysis using an Rtx-1301 column and GC-FID shows good resolution and a <15-minute analysis time.

Irish Whiskey on Rtx-1301

GC_FF1393

Peaks

PeakstR (min)
1.Acetaldehyde5.410
2.Methanol5.638
3.Ethanol7.029
4.1-Propanol9.492
5.Ethyl acetate10.322
6.Isobutanol11.388
7.Acetal13.137
8.3-Methyl-1-butanol14.334
9.2-Methyl-1-butanol14.420

Conditions

ColumnRtx-1301, 60 m, 0.25 mm ID, 1.0 µm (cat.# 16056)
Standard/SampleIrish whiskey
Diluent:n/a
Injection
Inj. Vol.:1 µL split (split ratio 100:1)
Liner:Topaz precision split liner, 4.0 mm x 6.3 x 78.5, with wool (cat.# 23305)
Inj. Temp.:250 °C
Split Vent Flow Rate:144.4 mL/min
Oven
Oven Temp.:35 °C (hold 3 min) to 220 °C at 8 °C/min (hold 5 min)
Carrier GasHe, constant flow
Flow Rate:1.4 mL/min
Linear Velocity:25.457 cm/sec @ 35 °C
DetectorFID @ 250 °C
Make-up Gas Flow Rate:25 mL/min
Make-up Gas Type:N2
Hydrogen flow:35 mL/min
Air flow:400 mL/min
Data Rate:50 Hz
InstrumentAgilent 7890A GC

Figure 10: Marsala analysis using an Rtx-1301 column and GC-FID shows good resolution.

Marsala on Rtx-1301

GC_FF1394

Peaks

PeakstR (min)
1.Acetaldehyde5.404
2.Methanol5.632
3.Ethanol6.956
4.Ethyl formate7.779
5.1-Propanol9.485
6.Ethyl acetate10.318
PeakstR (min)
7.sec-Butanol10.507
8.Isobutanol11.384
9.Acetic acid11.479
10.Acetal13.143
11.3-Methyl-1-butanol14.333
12.2-Methyl-1-butanol14.419
13.Furfural17.491

Conditions

ColumnRtx-1301, 60 m, 0.25 mm ID, 1.0 µm (cat.# 16056)
Standard/SampleMarsala (neat)
Diluent:n/a
Injection
Inj. Vol.:1 µL split (split ratio 100:1)
Liner:Topaz precision inlet liner, 4.0 mm x 6.3 x 78.5, with wool (cat.# 23305)
Inj. Temp.:250 °C
Split Vent Flow Rate:144.4 mL/min
Oven
Oven Temp.:35 °C (hold 3 min) to 220 °C at 8 °C/min (hold 5 min)
Carrier GasHe, constant flow
Flow Rate:1.4 mL/min
Linear Velocity:25.457 cm/sec @ 35 °C
DetectorFID @ 250 °C
Make-up Gas Flow Rate:25 mL/min
Make-up Gas Type:N2
Hydrogen flow:35 mL/min
Air flow:400 mL/min
Data Rate:50 Hz
InstrumentAgilent 7890A GC

Figure 11: Rum analysis using an Rtx-1301 column and GC-FID shows good resolution and good peak shape for the additive propylene glycol.

Rum on Rtx-1301

GC_FF1390

Peaks

PeakstR (min)
1.Acetaldehyde5.410
2.Methanol5.634
3.Ethanol7.026
4.1-Propanol9.492
5.Ethyl acetate10.321
6.Isobutanol11.387
7.Acetal13.138
8.3-Methyl-1-butanol14.334
9.2-Methyl-1-butanol14.420
10.Propylene glycol15.749

Conditions

ColumnRtx-1301, 60 m, 0.25 mm ID, 1.0 µm (cat.# 16056)
Standard/SampleRum (neat)
Diluent:n/a
Injection
Inj. Vol.:1 µL split (split ratio 100:1)
Liner:Topaz, precision inlet liner, 4.0 mm x 6.3 x 78.5, with wool (cat.# 23305)
Inj. Temp.:250 °C
Split Vent Flow Rate:144.4 mL/min
Oven
Oven Temp.:35 °C (hold 3 min) to 220 °C at 8 °C/min (hold 5 min)
Carrier GasHe, constant flow
Flow Rate:1.4 mL/min
Linear Velocity:25.457 cm/sec @ 35 °C
DetectorFID @ 250 °C
Make-up Gas Flow Rate:25 mL/min
Make-up Gas Type:N2
Hydrogen flow:35 mL/min
Air flow:400 mL/min
Data Rate:50 Hz
InstrumentAgilent 7890A GC

Meet the Experts

photo of corbyCorby Hilliard
As an advanced scientist in Restek’s GC lab with over 20 years of experience, Corby’s focus is on generating methodology and application data predominately for the food, petrochemical, and environmental industries. He is also involved with R&D and new product development.
photo of carolynCarolyn King
Carolyn is currently a technical service specialist at Restek. She began her career in 2012 in the QC lab conducting testing to ensure product quality before joining the technical service group in 2015. Carolyn holds a BS in animal sciences and nutritional sciences from The Pennsylvania State University.

References

  1. AOAC International, AOAC Official methods of analysis, 22nd edition, Edited by George W. Latimer, Jr., 3 vols., Oxford University Press, Oxford, UK, 2023.

Author

  • Restek Corporation

    Restek is a leading provider of chromatography columns, accessories, and certified reference materials. Trust Restek for reliable, high-quality analytical solutions.

    View all posts
FFAR4494