Anwendungshinweise

Fast, 3.5 Minute Analysis of Psilocin and Psilocybin in Urine by LC-MS/MS

09 Aug 2021

feature CFAN3694

Abstract

In this rapid LC-MS/MS analysis of psilocin and psilocybin in urine, supernatant from a simple protein precipitation was analyzed directly (no derivatization) on a Raptor Biphenyl column. Good quantitative results were obtained for linearity, accuracy, precision, and robustness in just 3.5 minutes. In addition, these compounds were analyzed using an established method for drugs of abuse/pain management medications, providing labs with an option to improve efficiency.

Introduction

Psilocybin-containing mushrooms, commonly called “magic mushrooms,” contain two hallucinogenic indole alkaloids, psilocin and psilocybin, which possess mind-altering properties. While magic mushrooms are currently illegal in the U.S., psilocybin is a potential therapeutic for cluster headaches, anxiety, depression, and other disorders, so several municipalities have decriminalized microdosing of mushroom extracts. As therapeutic use is being explored, the need for routine diagnostic testing is growing. However, typical LC-MS/MS methods for analysis of psilocin and psilocybin can be problematic for several reasons. The primary chromatographic issue is that these alkaloids are highly polar and require a column with retention mechanisms beyond traditional hydrophobic interactions to adequately retain and separate them from matrix interferences. In addition, analysis is further complicated because psilocybin breaks down into psilocin via cleavage of the phosphate moiety (Figure 1) during in-source fragmentation.

Due to increased interest in the analysis of psilocin and psilocybin, we developed a simple and robust LC-MS/MS analytical method that pairs a quick solvent protein precipitation with a fast chromatographic separation. For this work, a Raptor Biphenyl column was selected because the phenyl-based stationary phase has been shown to provide good retention of hydrophilic compounds that elute early on alkyl (C18) phases due to enhanced pi-pi interactions. Here, in addition to evaluating method performance, we demonstrated that both compounds could be successfully added to an established method for 231 drugs of abuse [1]. Adding psilocin and psilocybin to an existing drug panel method instead of setting up an independent method can allow labs to operate much more efficiently.

Figure 1:  Chemical Structures
figure article CFAN3694 01

Experimental

Calibration Standards and Quality Control Samples

First, 10 individual lots of human urine (BioreclamationIVT) were screened to confirm that both psilocin and psilocybin were not present. Then, the lots were pooled and fortified with psilocin and psilocybin to prepare calibration and QC standards. The linear range of the calibration curve was 50-5000 ng/mL, and four QC levels were prepared at 50 (LLOQ), 125 (LQC), 700 (MQC), and 4000 (HQC) ng/mL, respectively. For quantitation, psilocin D-10 (200 ng/mL) was used as the internal standard.

Sample Preparation

A 500 ng/mL sample of psilocin and psilocybin was prepared in pooled urine. A 50 µL aliquot was taken from the spiked urine sample and mixed with 10 µL of internal standard (psilocin-D10, 20 µg/mL) and 100 µL of methanol. The mixture was then vortexed at 3000 rpm for 10 seconds and centrifuged at 4300 rpm for 10 minutes at 10 °C. After centrifugation, 100 µL of the supernatant was diluted with 900 µL (20-fold dilution) of mobile phase A (0.1% formic acid and 2 mM ammonium formate in water) and injected for LC-MS/MS analysis.

Chromatographic Method:

The chromatographic conditions used on a Sciex 4000 coupled with a Shimadzu Prominence HPLC for this LC-MS/MS analysis of psilocin and psilocybin in urine are detailed below. The ion transitions and internal standards used for each analyte are provided in Table I.

Column: Raptor Biphenyl 2.7 µm, 50 mm x 2.1 mm (cat.# 9309A52)
Guard column: Raptor Biphenyl EXP guard column cartridge 2.7 µm, 5 mm x 2.1 mm (cat.# 9309A0252)
Column temp.: 35 °C
Injection volume: 5 µL
Mobile phase A: 0.1% Formic acid, 2 mM ammonium formate in water
Mobile phase B: 0.1% Formic acid, 2 mM ammonium formate in methanol
 
Time (min) %A %B
0.00 95 5
0.20 95 5
2.50 5 95
2.51 95 5
3.50 95 5
Flow rate: 0.5 mL/min
Ion mode: Positive ESI
Table I: Ion Transitions for LC-MS/MS Analysis of Psilocin and Psilocybin in Urine
Peak IdentificationPrecursor IonProduct Ion QuantifierProduct Ion Qualifier
Psilocybin  285.1205.1240.0
Psilocin205.1160.1115.0
Psilocin-D10 (IS)215.166.1

Results and Discussion

Chromatographic Performance and Robustness

In this analysis of psilocin and psilocybin, both compounds were sufficiently retained on the Raptor Biphenyl column, and good baseline resolution was achieved in a fast, 3.5-minute total cycle time (Figure 2). In addition, the assay was free of interference from the urine matrix and other ions (Figures 2 and 3). Carryover was assessed by injecting a blank sample directly after injection of the HLOQ sample, and no carryover was observed. As shown in Figure 4, good robustness was demonstrated by the consistent peak shapes and retention times that were still observed after 500 injections of a pooled urine extract fortified at 500 ng/mL. The maximum system pressure also remained consistent indicating that no column clogging had occurred.

Linearity, Accuracy, and Precision

Using a 1/x weighted linear regression for both analytes, psilocin and psilocybin showed acceptable linearity with r2 values of 0.995 or greater (Figure 5). In addition, the signal-to-noise values of the lowest calibration standards ranged from 16 to 65, indicating that this method could be used for the detection of much lower concentrations, if needed.

Precision and accuracy analysis of QC samples was performed on three different days. Method accuracy was demonstrated by percent recovery values being within 20% of the nominal concentration for all QC levels on all three days. The %RSD was 1.07-16.0% and 0.2-14.2% for intraday and interday evaluations, respectively, indicating acceptable method precision was achieved (Table II).

Easy Addition of Psilocin and Psilocybin to Existing Methods

In addition to developing a standalone method, we also explored adding both psilocin and psilocybin to an existing method to avoid the need for additional instrumentation and time for this analysis alone. As demonstrated in Figure 6, both psilocin and psilocybin in urine were run using the exact conditions of our previously published pain panel method that accommodates the analysis of 231 analytes [1]. Both the analytes were well separated from each other with a retention factor >5. In addition, no compounds from the pain panel method with the same masses eluted prior to 6 minutes, ensuring good separation from the earlier eluting psilocin and psilocybin. However, if coelution did occur under different conditions, psilocin and psilocybin could still be specifically detected by MS/MS based on their Q1 and Q3 masses. The selectivity of the Raptor Biphenyl column allowed these new compounds to be added to an existing method, providing labs with an important technique for improving efficiency and productivity.

Figure 2: Analysis of Psilocin and Psilocybin in Fortified Human Urine (500 ng/mL)
Psilocin and Psilocybin in Urine on Raptor Biphenyl by LC-MS/MS

LC_CF0769

Peaks

PeakstR (min)Conc.
(ng/mL)
Precursor IonProduct Ion 1Product Ion 2
1.Psilocybin1.36500285.1205.1240.0
2.Psilocin-D101.58200215.166.1
3.Psilocin1.60500205.1160.1115.0

Conditions

ColumnRaptor Biphenyl (cat.# 9309A52)
Dimensions:50 mm x 2.1 mm ID
Particle Size:2.7 µm
Pore Size:90 Å
Guard Column:Raptor Biphenyl EXP guard column cartridge 5.0 mm, 2.1 mm ID, 2.7 µm (cat.# 9309A0252)
Temp.:35 °C
Standard/Sample
Diluent:Water, 0.1% formic acid + 2 mM ammonium formate
Conc.:500 ng/mL
Inj. Vol.:5 µL
Mobile Phase
A:Water, 0.1% formic acid + 2 mM ammonium formate
B:Methanol, 0.1% formic acid + 2 mM ammonium formate
Time (min)Flow (mL/min)%A%B
0.000.5955
0.200.5955
2.500.5595
2.510.5955
3.500.5955
DetectorMS/MS
Ion Source:Electrospray
Ion Mode:ESI+
Mode:MRM
InstrumentHPLC
Sample PreparationA 500 ng/mL standard mix of psilocin and psilocybin was prepared in pooled urine. A 50 µL aliquot was taken from the standard and mixed with 10 µL of internal standard (psilocin-D10, 20 µg/mL) and 100 µL of methanol. The mixture was vortexed at 3000 rpm for 10 seconds and then centrifuged at 4300 rpm for 10 minutes at 10 °C. After centrifugation, 100 µL of the supernatant was diluted with 900 µL (20-fold dilution) of water containing 0.1% formic acid and 2 mM ammonium formate (mobile phase A) and injected for LC-MS/MS analysis.
Figure 3: Blank Human Urine Spiked with Internal Standard Only
Blank Human Urine Spiked with Internal Standard Only on Raptor Biphenyl

LC_CF0770

Peaks

PeakstR (min)Conc.
(ng/mL)
Precursor IonProduct Ion 1
1.Psilocin-D101.58200215.166.1

Conditions

ColumnRaptor Biphenyl (cat.# 9309A52)
Dimensions:50 mm x 2.1 mm ID
Particle Size:2.7 µm
Pore Size:90 Å
Guard Column:Raptor Biphenyl EXP guard column cartridge 5.0 mm, 2.1 mm ID, 2.7 µm (cat.# 9309A0252)
Temp.:35 °C
Standard/Sample
Diluent:Water, 0.1% formic acid + 2 mM ammonium formate
Inj. Vol.:5 µL
Mobile Phase
A:Water, 0.1% formic acid + 2 mM ammonium formate
B:Methanol, 0.1% formic acid + 2 mM ammonium formate
Time (min)Flow (mL/min)%A%B
0.000.5955
0.200.5955
2.500.5595
2.510.5955
3.500.5955
DetectorMS/MS
Ion Source:Electrospray
Ion Mode:ESI+
Mode:MRM
InstrumentHPLC
Sample PreparationA 50 µL aliquot was taken from blank pooled human urine and mixed with 10 µL of internal standard (psilocin-D10, 20 µg/mL) and 100 µL of methanol. The mixture was vortexed at 3000 rpm for 10 seconds and then centrifuged at 4300 rpm for 10 minutes at 10 °C. After centrifugation, 100 µL of the supernatant was diluted with 900 µL (20-fold dilution) of water containing 0.1% formic acid and 2 mM ammonium formate (mobile phase A) and injected for LC-MS/MS analysis.
Figure 4: Robust column performance over 500 urine sample injections was obtained on a Raptor Biphenyl column. Good retention, consistent peak shapes, and stable retention times were seen over the course of the experiment.
Column Lifetime Test: Psilocin and Psilocybin in Urine on Raptor Biphenyl

LC_CF0771

Peaks

PeaksConc.
(ng/mL)
Precursor IonProduct Ion 1Product Ion 2
1.Psilocybin500285.1205.1240.0
2.Psilocin-D10200215.166.1
3.Psilocin500205.1160.1115.0

Conditions

ColumnRaptor Biphenyl (cat.# 9309A52)
Dimensions:50 mm x 2.1 mm ID
Particle Size:2.7 µm
Pore Size:90 Å
Guard Column:Raptor Biphenyl EXP guard column cartridge 5.0 mm, 2.1 mm ID, 2.7 µm (cat.# 9309A0252)
Temp.:35 °C
Standard/Sample
Diluent:Water, 0.1% formic acid + 2 mM ammonium formate
Conc.:500 ng/mL
Inj. Vol.:5 µL
Mobile Phase
A:Water, 0.1% formic acid + 2 mM ammonium formate
B:Methanol, 0.1% formic acid + 2 mM ammonium formate
Time (min)Flow (mL/min)%A%B
0.000.5955
0.200.5955
2.500.5595
2.510.5955
3.500.5955
DetectorMS/MS
Ion Source:Electrospray
Ion Mode:ESI+
Mode:MRM
InstrumentHPLC
Sample PreparationA 500 ng/mL standard mix of psilocin and psilocybin was prepared in pooled urine. A 50 µL aliquot was taken from the standard and mixed with 10 µL of internal standard (psilocin-D10, 20 µg/mL) and 100 µL of methanol. The mixture was vortexed at 3000 rpm for 10 seconds and centrifuged at 4300 rpm for 10 minutes at 10 °C. After centrifugation, 100 µL of the supernatant was diluted with 900 µL (20-fold dilution) of water containing 0.1% formic acid and 2 mM ammonium formate (mobile phase A) and injected for LC-MS/MS analysis.
Figure 5: Standard Curves for the Analysis of Psilocin and Psilocybin
figure article CFAN3694 05
Figure 6: Analysis of psilocin and psilocybin in urine using the existing “Big Pain” multi-analyte method.
Psilocin and Psilocybin in Urine on Raptor Biphenyl by Restek's 'Big Pain' Method (CFAR2309-UNV)

LC_CF0772

Peaks

PeakstR (min)Conc.
(ng/mL)
Precursor IonProduct Ion 1Product Ion 2
1.Psilocybin2.09500285.1205.1240.0
2.Psilocin-D102.74200215.166.1
3.Psilocin2.78500205.1160.1115.0

Conditions

ColumnRaptor Biphenyl (cat.# 9309A12)
Dimensions:100 mm x 2.1 mm ID
Particle Size:2.7 µm
Pore Size:90 Å
Guard Column:Raptor Biphenyl EXP guard column cartridge 5.0 mm, 2.1 mm ID, 2.7 µm (cat.# 9309A0252)
Temp.:30 °C
Standard/Sample
Diluent:Water, 0.1% formic acid + 2 mM ammonium formate
Conc.:500 ng/mL
Inj. Vol.:5 µL
Mobile Phase
A:Water, 0.1% formic acid + 2 mM ammonium formate
B:Methanol, 0.1% formic acid + 2 mM ammonium formate
Time (min)Flow (mL/min)%A%B
0.000.6955
9.000.60100
10.000.60100
10.010.6955
12.000.6955
DetectorMS/MS
Ion Source:Electrospray
Ion Mode:ESI+
Mode:MRM
InstrumentHPLC
Sample PreparationA 500 ng/mL standard mix of psilocin and psilocybin was prepared in pooled urine. A 50 µL aliquot was taken from the standard and mixed with 10 µL of internal standard (psilocin-D10, 20 µg/mL) and 100 µL of methanol. The mixture was vortexed at 3000 rpm for 10 seconds and centrifuged at 4300 rpm for 10 minutes at 10 °C. After centrifugation, 100 µL of the supernatant was diluted with 900 µL (20-fold dilution) of water containing 0.1% formic acid and 2 mM ammonium formate (mobile phase A) and injected for LC-MS/MS analysis.
Table II: Accuracy and Precision of QC Samples
AnalyteQC LLOQ (50 ng/mL)QC Low (125 ng/mL)QC Mid (700 ng/mL)QC High (4000 ng/mL)
Avg. Conc. (ng/mL)Avg. Accuracy (%)% RSDAvg. Conc. (ng/mL)Avg. Accuracy (%)% RSDAvg. Conc. (ng/mL)Avg. Accuracy (%)% RSDAvg. Conc. (ng/mL)Avg. Accuracy (%)% RSD
Psilocybin57.41153.9011188.814.262088.50.200373093.37.10
Psilocin55.71128.301471183.9065593.53.8042001062.30

Conclusion

A fast, 3.5-minute chromatographic separation employing a simple protein precipitation and a Raptor Biphenyl column was developed for the quantitative analysis of psilocin and psilocybin in human urine. The use of a Raptor Biphenyl column ensured adequate retention and provided baseline separation, consistent peak shapes, and stable retention times over the course of 500 matrix injections. In addition to this robust method, these compounds can be analyzed by an established method for drugs of abuse/pain management medications, which affords labs an important opportunity to increase productivity and efficiency.

References

  1. S. Lupo, “The Big Pain”: Development of Pain-Free Methods for Analyzing 231 Multiclass Drugs and Metabolites by LC-MS/MS, Restek Corporation, 2016 https://www.restek.com/articles/the-big-pain-development-of-pain-free-methods-for-analyzing-231-multiclass-drugs-and-metabolites-by-LC-MSMS

Products Mentioned


Raptor Biphenyl, 2.7 µm, 50 x 2.1 mm HPLC Column
Raptor Biphenyl EXP Guard Column Cartridge, 2.7 µm, 5 x 2.1 mm, 3-pk.
CFAN3694