Sin categorizar

Method 8082A: Polychlorinated Biphenyls (PCBs) Analysis by GC

27 Oct 2020

feature EVSS1013

EPA Method 8082A details how Aroclor mixtures and polychlorinated biphenyl (PCB) congeners are to be analyzed in a parallel dual-column confirmation setup. When choosing columns, it is important to select stationary phases that have low bleed and high thermal stability. This allows the columns to be held at high temperature at the end of each analysis to prevent carryover from one injection to the next. Because many instruments used for PCB analysis also may be used for pesticide and herbicide analyses, the column pair of choice is the Rtx-CLPesticides and Rtx-CLPesticides2 columns. This column set provides low bleed, high thermal stability, and is designed for primary column analysis and secondary column confirmation.

For Aroclor mixture analysis under EPA 8082A, individual Aroclor mixes are required to be analyzed as standards, then the sample extract chromatograms are compared to the standards to qualitatively identify the Aroclor mixtures. Once identification has been made, quantification can be performed by selecting five of the largest peaks, treating them as individual compounds, and then reporting the average concentration.

When analyzing for PCB congeners using Method 8082A, each peak is to be treated as an individual component and a standard curve is made for each of the congeners of interest. Note that while many laboratories are interested in the analysis of PCBs by congener, most do not need, or desire, to analyze all 209.

VendorColumn PairAnalysis Time (min)Restek Advantage
RestekRtx-CLPesticides /
Rtx-CLPesticides2
7 / 7Fast PCB analysis times.
Competitor Acolumn 1 / column 26 / 7
Competitor Bcolumn 1 / column 224 / 21

Comparison based on published competitor data. All columns tested were 0.32 mm ID. Coelutions are excluded from comparison as they are not relevant for Aroclor analysis.

Aroclor PCB analysis on the Rtx-CLPesticides column.
Aroclor PCBs on Rtx-CLPesticides

GC_EV00993-998

Conditions

ColumnRtx-CLPesticides, 30 m, 0.32 mm ID, 0.32 µm (cat.# 11141)
Standard/Sample
PCB kit #1 diluted to 1,000 ppb in hexane (cat.# 32089)
Decachlorobiphenyl (BZ #209) diluted to 100 ppb in acetone (cat.# 32029)
Injection
Inj. Vol.:1.0 µL pulsed splitless (hold 0.3 min)
Liner:Cyclo double taper (4 mm)
Inj. Temp.:250 °C
Pulse Pressure:30 psi (206.8kPa)
Oven
Oven Temp.:120 °C to 200 °C at 45 °C/min to 230 °C at 15 °C/min to 330 °C at 30 °C/min (hold 2 min)
Carrier GasHe, constant flow
Linear Velocity:60 cm/sec
DetectorMicro-ECD @ 330 °C
NotesLiner cat.# 20895 was used to produce this chromatogram, but it has since been discontinued. We recommend cat.# 20896 [or cat.# 23310] as an alternative. For assistance choosing a replacement for this application, contact Restek Technical Service or your local Restek representative.
– – – – – –
This chromatogram was obtained using an Agilent micro-ECD. To obtain comparable results, you will need to employ a micro-ECD in addition to confirmational dual columns connected to a 5-meter guard column using a “Y” Press-Tight connector.
Aroclor PCB analysis on the Rtx-CLPesticides2 column.
Aroclor PCBs on Rtx-CLPesticides2

GC_EV00993A-998A

Conditions

ColumnRtx-CLPesticides2, 30 m, 0.32 mm ID, 0.25 µm (cat.# 11324)
Standard/Sample
PCB kit #1 diluted to 1,000 ppb in hexane (cat.# 32089)
Decachlorobiphenyl (BZ #209) diluted to 100 ppb in acetone (cat.# 32029)
Injection
Inj. Vol.:1.0 µL pulsed splitless (hold 0.3 min)
Liner:Cyclo double taper (4 mm)
Inj. Temp.:250 °C
Pulse Pressure:30 psi (206.8kPa)
Oven
Oven Temp.:120 °C to 200 °C at 45 °C/min to 230 °C at 15 °C/min to 330 °C at 30 °C/min (hold 2 min)
Carrier GasHe, constant flow
Linear Velocity:60 cm/sec
DetectorMicro-ECD @ 330 °C
NotesLiner cat.# 20895 was used to produce this chromatogram, but it has since been discontinued. We recommend cat.# 20896 [or cat.# 23310] as an alternative. For assistance choosing a replacement for this application, contact Restek Technical Service or your local Restek representative.
– – – – – –
This chromatogram was obtained using an Agilent micro-ECD. To obtain comparable results, you will need to employ a micro-ECD in addition to confirmational dual columns connected to a 5-meter guard column using a “Y” Press-Tight connector.

Products Mentioned


Rtx-CLPesticides GC Capillary Column Kit, 0.32 mm ID, 0.32 µm
Rtx-CLPesticides2 GC Capillary Column, 15 m, 0.32 mm ID, 0.25 µm
Rtx-CLPesticides GC Capillary Column, 15 m, 0.25 mm ID, 0.25 µm

Author

  • Restek Corporation

    Restek is a leading provider of chromatography columns, accessories, and certified reference materials. Trust Restek for reliable, high-quality analytical solutions.

    View all posts
EVAR4502