EPA Method 8270E now allows for the use of GC tandem mass spectrometry (GC-MS/MS), which can extend detection levels for semivolatiles down into the low picogram range. Previous versions of the method only allowed GC-MS analysis, which is much less sensitive. Much of the method has remained the same, but the lower levels of detection provide new challenges, including a dynamic calibration range and retention time shifts for reactive compounds. The calibration range is limited by lower levels of detection at the lower end as well as by overloading and MS saturation at the higher end. Using an Rxi-SVOCms column (30 m, 0.25 mm ID, 0.25 µm [cat.# 16623]) and the conditions shown in Figure 1, the upper limit was found to be 0.5 ng on-column. This shifts the typical range of calibration (0.1-100 µg/mL) to concentrations 100x to 1000x lower. Note that a wider range could be obtained by adjusting the split ratio.
Figure 1: The main chromatogram is an analysis of semivolatiles at 100 pg on an Rxi-SVOCms column. A calibration curve was also performed, and the inset figures represent the lowest calibration point for selected compounds.

GC_EV1522
Peaks
| Peaks | tR (min) | Conc. (µg/g) | SRM 1 | SRM 2 | SRM 3 | |
|---|---|---|---|---|---|---|
| 1. | 1,4-Dioxane | 2.79 | 1 | 88 to 28 @14V | 58 to 28 @6V | 88 to 57 @8V |
| 2. | N-Nitrosodimethylamine | 3.08 | 1 | 74 to 42 @16V | 74 to 43 @12V | 74 to 44 @6V |
| 3. | Pyridine | 3.13 | 1 | 79 to 51 @26V | 52 to 26 @18V | 79 to 52 @12V |
| 4. | Ethyl methacrylate | 3.62 | 1 | 99 to 43 @14V | 69 to 41 @6V | 99 to 53 @12V |
| 5. | 2-Picoline | 3.96 | 1 | 93 to 65 @18V | 66 to 39 @18V | 93 to 66 @12V |
| 6. | N-Nitrosomethylethylamine | 4.12 | 1 | 88 to 42 @16V | 88 to 43 @8V | 88 to 71 @6V |
| 7. | Acrylamine | 4.5 | 1 | 71 to 44 @22V | 55 to 27 @6V | 71 to 55 @6V |
| 8. | Methyl methanesulfonate | 4.52 | 1 | 80 to 48 @28V | 65 to 48 @18V | 80 to 65 @8V |
| 9. | 2-Fluorophenol | 4.73 | 1 | 112 to 63 @22V | 92 to 63 @14V | 112 to 92 @6V |
| 10. | N-Nitrosodiethylamine | 5.05 | 1 | 102 to 44 @10V | 102 to 56 @12V | 102 to 85 @6V |
| 11. | Ethyl methanesulfonate | 5.51 | 1 | 109 to 45 @10V | 79 to 48 @34V | 109 to 79 @6V |
| 12. | Benzaldehyde | 6.01 | 1 | 105 to 51 @26V | 77 to 51 @14V | 105 to 77 @10V |
| 13. | Phenol | 6.13 | 1 | 94 to 39 @30V | 94 to 65 @16V | 94 to 66 @10V |
| 14. | Phenol-d6 | 6.14 | 1 | 99 to 69 @18V | 71 to 42 @16V | 99 to 71 @8V |
| 15. | Aniline | 6.17 | 1 | 93 to 65 @20V | 66 to 39 @18V | 93 to 66 @10V |
| 16. | Pentachloroethane | 6.26 | 1 | 167 to 95 @32V | 117 to 82 @30V | 167 to 132 @14V |
| 17. | bis-(2-Chloroethyl)ether | 6.28 | 1 | 93 to 27 @22V | 63 to 27 @10V | 93 to 63 @6V |
| 18. | 2-Chlorophenol | 6.35 | 1 | 128 to 63 @24V | 64 to 38 @12V | 128 to 64 @14V |
| 19. | Decane | 6.49 | 1 | 85 to 41 @12V | 71 to 43 @6V | 85 to 43 @6V |
| 20. | 1,3-Dichlorobenzene | 6.61 | 1 | 146 to 75 @24V | 111 to 75 @12V | 146 to 111 @12V |
| 21. | 1,4-Dichlorobenzene-d4 | 6.71 | 1 | 150 to 76 @38V | 150 to 78 @26V | 150 to 115 @14V |
| 22. | 1,4-Dichlorobenzene | 6.74 | 1 | 146 to 75 @24V | 111 to 75 @12V | 146 to 111 @14V |
| 23. | Benzyl alcohol | 6.94 | 1 | 108 to 77 @22V | 79 to 51 @20V | 108 to 79 @12V |
| 24. | 1,2-Dichlorobenzene | 6.98 | 1 | 146 to 111 @12V | 111 to 50 @36V | 111 to 75 @10V |
| 25. | Indene | 7.12 | 1 | 116 to 89 @28V | 115 to 63 @26V | 115 to 89 @16V |
| 26. | 2-Methylphenol | 7.14 | 1 | 108 to 77 @24V | 79 to 51 @20V | 108 to 80 @8V |
| 27. | 2,2′-oxybis(1-chloropropane) | 7.2 | 1 | 121 to 45 @6V | 77 to 51 @12V | 121 to 77 @8V |
| 28. | N-Nitrosopyrrolidine | 7.34 | 1 | 100 to 43 @8V | 100 to 55 @6V | 100 to 68 @8V |
| 29. | Acetophenone | 7.37 | 1 | 105 to 51 @26V | 77 to 51 @14V | 105 to 77 @12V |
| 30. | 3- and 4-Methylphenol | 7.41 | 1 | 107 to 51 @26V | 70 to 43 @6V | 107 to 77 @12V |
| 31. | N-Nitrosomorpholine | 7.42 | 1 | 116 to 56 @10V | 86 to 56 @6V | 116 to 86 @6V |
| 32. | o-Toluidine | 7.44 | 1 | 107 to 77 @26V | 106 to 77 @16V | 106 to 79 @8V |
| 33. | Hexachloroethane | 7.56 | 1 | 201 to 131 @30V | 117 to 82 @24V | 201 to 166 @12V |
| 34. | Nitrobenzene-d5 | 7.63 | 1 | 128 to 54 @28V | 82 to 54 @12V | 128 to 82 @12V |
| 35. | Nitrobenzene | 7.68 | 1 | 123 to 51 @30V | 77 to 51 @12V | 123 to 77 @10V |
| 36. | N-Nitrosopiperidine | 7.94 | 1 | 114 to 41 @10V | 114 to 84 @6V | 114 to 97 @6V |
| 37. | Isophorone | 8.12 | 1 | 138 to 82 @8V | 82 to 39 @12V | 82 to 54 @6V |
| 38. | 2-Nitrophenol | 8.25 | 1 | 139 to 81 @12V | 109 to 81 @8V | 139 to 109 @8V |
| 39. | 2,4-Dimethylphenol | 8.36 | 1 | 122 to 77 @26V | 107 to 77 @14V | 122 to 107 @12V |
| 40. | O,O,O-Triethyl phosphorothioate | 8.51 | 1 | 198 to 114 @12V | 121 to 65 @10V | 198 to 121 @8V |
| 41. | Benzoic acid | 8.53 | 1 | 122 to 77 @20V | 105 to 77 @10V | 122 to 105 @8V |
| 42. | Bis(2-chloroethoxy)methane | 8.55 | 1 | 93 to 63 @6V | 63 to 26 @36V | 63 to 27 @12V |
| 43. | Phentermine | 8.64 | 1 | 134 to 91 @20V | 58 to 42 @22V | 134 to 115 @12V |
| 44. | 2,4-Dichlorophenol | 8.68 | 1 | 162 to 63 @24V | 98 to 63 @8V | 162 to 98 @12V |
| 45. | 1,2,4-Trichlorobenzene | 8.84 | 1 | 180 to 109 @24V | 145 to 109 @10V | 180 to 145 @12V |
| 46. | Naphthalene-d8 | 8.91 | 1 | 136 to 84 @20V | 136 to 108 @18V | 136 to 134 @14V |
| 47. | Naphthalene | 8.97 | 1 | 129 to 103 @14V | 128 to 78 @20V | 128 to 102 @16V |
| 48. | a-Terpineol | 9.04 | 1 | 136 to 93 @10V | 121 to 93 @8V | 136 to 121 @8V |
| 49. | 4-Chloroaniline | 9.09 | 1 | 127 to 65 @20V | 65 to 39 @12V | 127 to 100 @10V |
| 50. | 2,6-Dichlorophenol | 9.11 | 1 | 164 to 63 @26V | 162 to 63 @26V | 162 to 126 @8V |
| 51. | Hexachloropropene | 9.15 | 1 | 215 to 119 @18V | 213 to 119 @18V | 215 to 121 @18V |
| 52. | Hexachlorobutadiene | 9.23 | 1 | 260 to 190 @26V | 225 to 190 @14V | 260 to 225 @12V |
| 53. | Quinoline | 9.6 | 1 | 129 to 76 @26V | 102 to 76 @12V | 129 to 102 @16V |
| 54. | Caprolactam | 9.7 | 1 | 113 to 56 @10V | 85 to 67 @6V | 113 to 85 @6V |
| 55. | N-Nitrosodibutylamine | 9.76 | 1 | 116 to 74 @8V | 84 to 56 @14V | 116 to 99 @6V |
| 56. | 1,4-Phenylenediamine | 9.77 | 1 | 108 to 80 @20V | 80 to 53 @12V | 108 to 81 @10V |
| 57. | 4-Chloro-3-methylphenol | 10.04 | 1 | 142 to 77 @26V | 107 to 77 @12V | 142 to 107 @12V |
| 58. | Isosafrole I | 10.16 | 1 | 162 to 104 @12V | 131 to 103 @8V | 162 to 131 @12V |
| 59. | 2-Methylnaphthalene | 10.28 | 1 | 141 to 89 @30V | 115 to 89 @16V | 141 to 115 @16V |
| 60. | 1-Methylnaphthalene | 10.46 | 1 | 141 to 89 @30V | 115 to 89 @14V | 141 to 115 @16V |
| 61. | Hexachloropentadiene | 10.59 | 1 | 272 to 237 @12V | 237 to 119 @20V | 237 to 143 @22V |
| 62. | 1,2,4,5-Tetrachlorobenzene | 10.61 | 1 | 216 to 108 @36V | 179 to 143 @8V | 216 to 181 @14V |
| 63. | Isosafrole II | 10.73 | 1 | 162 to 104 @12V | 131 to 103 @10V | 162 to 131 @12V |
| 64. | 2,3-Dichloroaniline | 10.83 | 1 | 163 to 90 @18V | 161 to 90 @16V | 161 to 99 @18V |
| 65. | 2,4,6-Trichlorophenol | 10.83 | 1 | 196 to 97 @24V | 132 to 97 @10V | 160 to 97 @16V |
| 66. | 2,4,5-Trichlorophenol | 10.88 | 1 | 196 to 97 @24V | 132 to 97 @10V | 196 to 132 @12V |
| 67. | 2-Fluorobiphenyl | 10.99 | 1 | 172 to 151 @22V | 172 to 170 @22V | 172 to 171 @12V |
| 68. | Safrole | 11.16 | 1 | 162 to 77 @30V | 131 to 103 @10V | 162 to 127 @16V |
| 69. | 2-Chloronaphthalene | 11.2 | 1 | 162 to 77 @32V | 127 to 77 @16V | 162 to 127 @16V |
| 70. | Biphenyl | 11.2 | 1 | 153 to 126 @32V | 154 to 115 @24V | 154 to 152 @22V |
| 71. | 1-Chloronaphthalene | 11.25 | 1 | 162 to 77 @30V | 127 to 77 @16V | 162 to 127 @16V |
| 72. | o-Nitroaniline | 11.41 | 1 | 138 to 65 @22V | 108 to 80 @8V | 138 to 92 @12V |
| 73. | Diphenyl ether | 11.42 | 1 | 170 to 141 @16V | 141 to 115 @14V | 170 to 142 @10V |
| 74. | 1,4-Naphthoquinone | 11.56 | 1 | 158 to 102 @14V | 102 to 76 @12V | 158 to 130 @8V |
| 75. | 1,2-Dinitrobenzene | 11.7 | 1 | 168 to 75 @20V | 122 to 75 @12V | 168 to 122 @8V |
| 76. | 1,3-Dinitrobenzene | 11.83 | 1 | 168 to 75 @22V | 76 to 50 @10V | 168 to 122 @8V |
| 77. | Diphenyl phthalate | 11.83 | 1 | 163 to 77 @20V | 163 to 92 @24V | 163 to 133 @8V |
| 78. | 1,6-Dinitrotoluene | 11.91 | 1 | 165 to 63 @20V | 148 to 90 @10V | 165 to 148 @8V |
| 79. | Acenaphthylene | 11.98 | 1 | 152 to 76 @36V | 152 to 102 @26V | 152 to 126 @24V |
| 80. | 1,4-Dinitrobenzene | 11.99 | 1 | 168 to 51 @14V | 76 to 63 @6V | 168 to 63 @26V |
| 81. | 3-Nitroaniline | 12.21 | 1 | 138 to 65 @20V | 92 to 65 @8V | 138 to 92 @12V |
| 82. | Acenaphthene-d10 | 12.24 | 1 | 164 to 160 @32V | 162 to 160 @18V | 164 to 162 @14V |
| Peaks | tR (min) | Conc. (µg/g) | SRM 1 | SRM 2 | SRM 3 | |
|---|---|---|---|---|---|---|
| 83. | Acenaphthene | 12.32 | 1 | 153 to 77 @38V | 153 to 126 @36V | 154 to 127 @34V |
| 84. | 2,4-Dinitrophenol | 12.41 | 1 | 184 to 107 @10V | 154 to 79 @12V | 184 to 154 @6V |
| 85. | 4-Nitrophenol | 12.57 | 1 | 139 to 81 @14V | 109 to 81 @8V | 139 to 109 @6V |
| 86. | Pentachlorobenzene | 12.6 | 1 | 250 to 142 @38V | 215 to 179 @10V | 250 to 215 @16V |
| 87. | 2,4-Dinitrotoluene | 12.66 | 1 | 165 to 63 @22V | 89 to 63 @12V | 165 to 119 @6V |
| 88. | Dibenzofuran | 12.66 | 1 | 168 to 139 @22V | 139 to 63 @30V | 139 to 89 @18V |
| 89. | 1-Naphthylamine | 12.81 | 1 | 143 to 115 @22V | 115 to 89 @16V | 143 to 116 @10V |
| 90. | 2,3,5,6-Tetrachlorophenol | 12.83 | 1 | 234 to 133 @26V | 232 to 131 @26V | 232 to 168 @12V |
| 91. | 2,3,4,6-Tetrachlorophenol | 12.91 | 1 | 234 to 131 @24V | 232 to 133 @26V | 232 to 168 @12V |
| 92. | 2-Naphthylamine | 12.97 | 1 | 143 to 115 @24V | 115 to 89 @16V | 143 to 116 @10V |
| 93. | Diethyl Phthalate | 13.21 | 1 | 177 to 149 @8V | 149 to 65 @20V | 149 to 93 @12V |
| 94. | Hexadecane | 13.29 | 1 | 99 to 41 @14V | 85 to 43 @6V | 99 to 57 @6V |
| 95. | Fluorene | 13.31 | 1 | 167 to 166 @14V | 165 to 115 @24V | 165 to 139 @26V |
| 96. | Zinophos | 13.36 | 1 | 143 to 79 @10V | 107 to 52 @20V | 107 to 79 @14V |
| 97. | 5-Nitro-o-toluidine | 13.37 | 1 | 152 to 77 @24V | 106 to 77 @12V | 152 to 106 @10V |
| 98. | 4-Chlorophenyl phenyl ether | 13.37 | 1 | 204 to 77 @22V | 141 to 115 @14V | 204 to 141 @12V |
| 99. | 4-Nitroaniline | 13.37 | 1 | 138 to 80 @18V | 108 to 80 @10V | 138 to 108 @8V |
| 100. | 4,6-Dinitro-2-methylphenol | 13.45 | 1 | 198 to 121 @10V | 168 to 121 @6V | 198 to 168 @6V |
| 101. | Diphenylamine | 13.6 | 1 | 169 to 66 @22V | 169 to 77 @30V | 170 to 66 @22V |
| 102. | Azobenzene | 13.67 | 1 | 182 to 77 @12V | 77 to 51 @14V | 182 to 105 @6V |
| 103. | 2,4,6-Tribromophenol | 13.77 | 1 | 332 to 143 @34V | 330 to 141 @36V | 330 to 222 @20V |
| 104. | Sulfotepp | 14 | 1 | 322 to 146 @24V | 202 to 146 @10V | 322 to 202 @10V |
| 105. | 1,3,5-Trinitrobenzene | 14.14 | 1 | 213 to 74 @38V | 213 to 120 @22V | 213 to 167 @8V |
| 106. | Diallate 1 | 14.2 | 1 | 234 to 150 @16V | 86 to 43 @6V | 234 to 192 @12V |
| 107. | Phorate | 14.22 | 1 | 121 to 47 @26V | 75 to 47 @8V | 121 to 65 @8V |
| 108. | Phenacetin | 14.25 | 1 | 179 to 109 @14V | 108 to 80 @12V | 179 to 137 @8V |
| 109. | 4-Bromophenyl phenyl ether | 14.31 | 1 | 250 to 141 @16V | 141 to 115 @12V | 248 to 141 @14V |
| 110. | Diallate 2 | 14.37 | 1 | 234 to 150 @16V | 86 to 43 @6V | 234 to 192 @10V |
| 111. | Hexachlorobenzene | 14.37 | 1 | 288 to 216 @26V | 249 to 214 @12V | 284 to 249 @14V |
| 112. | Dimethoate | 14.51 | 1 | 125 to 47 @14V | 93 to 63 @8V | 125 to 79 @8V |
| 113. | Atrazine | 14.7 | 1 | 215 to 58 @10V | 200 to 122 @8V | 215 to 200 @8V |
| 114. | Pentachlorophenol | 14.77 | 1 | 270 to 169 @22V | 228 to 165 @14V | 266 to 167 @22V |
| 115. | 4-Aminobiphenyl | 14.79 | 1 | 169 to 141 @22V | 168 to 141 @8V | 168 to 167 @10V |
| 116. | Pentachloronitrobenzene | 14.81 | 1 | 295 to 237 @14V | 237 to 143 @20V | 295 to 265 @8V |
| 117. | Propyzamide | 15 | 1 | 254 to 191 @16V | 173 to 145 @14V | 254 to 226 @10V |
| 118. | Octadecane | 15.1 | 1 | 99 to 41 @14V | 85 to 43 @6V | 99 to 57 @6V |
| 119. | Phenanthrene-d10 | 15.1 | 1 | 188 to 160 @20V | 184 to 154 @32V | 184 to 156 @22V |
| 120. | Phenanthrene | 15.16 | 1 | 178 to 151 @32V | 176 to 150 @20V | 178 to 152 @20V |
| 121. | Dinoseb | 15.21 | 1 | 240 to 117 @24V | 163 to 116 @14V | 240 to 211 @8V |
| 122. | Disulfoton | 15.22 | 1 | 97 to 65 @16V | 88 to 60 @6V | 97 to 79 @12V |
| 123. | Anthracene | 15.26 | 1 | 178 to 151 @32V | 177 to 151 @18V | 178 to 152 @18V |
| 124. | Carbazole | 15.6 | 1 | 167 to 139 @24V | 166 to 140 @14V | 167 to 166 @12V |
| 125. | Methyl Parathion | 15.95 | 1 | 263 to 79 @24V | 125 to 47 @12V | 263 to 109 @10V |
| 126. | di-n-Butyl phthalate | 16.41 | 1 | 149 to 65 @20V | 149 to 93 @14V | 149 to 121 @12V |
| 127. | 4-Nitroquinoline-N-oxide | 16.69 | 1 | 190 to 77 @20V | 101 to 75 @10V | 190 to 160 @8V |
| 128. | Ethyl Parathion | 16.73 | 1 | 291 to 81 @26V | 109 to 81 @10V | 291 to 109 @10V |
| 129. | Methapyrilene | 16.9 | 1 | 190 to 97 @14V | 97 to 53 @16V | 190 to 157 @8V |
| 130. | Isodrin | 17.19 | 1 | 261 to 191 @28V | 193 to 123 @28V | 261 to 226 @16V |
| 131. | Fluoranthene | 17.47 | 1 | 202 to 152 @30V | 200 to 174 @22V | 202 to 176 @26V |
| 132. | Benzidine | 17.8 | 1 | 184 to 130 @26V | 184 to 156 @18V | 184 to 166 @16V |
| 133. | Pyrene-d10 | 17.84 | 1 | 212 to 208 @36V | 212 to 210 @20V | 0 |
| 134. | Pyrene | 17.88 | 1 | 200 to 149 @34V | 200 to 150 @26V | 200 to 174 @24V |
| 135. | p-Terphenyl-d14 | 18.3 | 1 | 244 to 226 @18V | 244 to 240 @22V | 244 to 242 @14V |
| 136. | Aramite 1 | 18.33 | 1 | 319 to 123 @10V | 175 to 107 @14V | 319 to 185 @6V |
| 137. | Aramite 2 | 18.49 | 1 | 319 to 107 @36V | 185 to 63 @12V | 319 to 185 @6V |
| 138. | p-Dimethylaminoazobenzene | 18.57 | 1 | 225 to 120 @10V | 120 to 77 @16V | 225 to 148 @6V |
| 139. | Chlorobenzilate | 18.71 | 1 | 251 to 111 @30V | 139 to 111 @12V | 251 to 139 @12V |
| 140. | Famphur | 19.17 | 1 | 218 to 109 @14V | 125 to 47 @14V | 125 to 79 @6V |
| 141. | 3,3′-Dimethylbenzidine | 19.21 | 1 | 212 to 180 @22V | 211 to 195 @16V | 211 to 196 @8V |
| 142. | Kepone | 19.24 | 1 | 272 to 237 @12V | 237 to 119 @24V | 237 to 143 @22V |
| 143. | Benzyl butyl phthalate | 19.31 | 1 | 206 to 149 @8V | 149 to 65 @20V | 149 to 121 @10V |
| 144. | Bis(2-ethylhexyl) adipate | 19.55 | 1 | 129 to 55 @14V | 111 to 83 @6V | 129 to 83 @8V |
| 145. | 2-Acetylaminofluorene | 19.69 | 1 | 223 to 180 @26V | 181 to 152 @34V | 223 to 181 @10V |
| 146. | Chrysene-d12 | 20.23 | 1 | 240 to 212 @22V | 240 to 236 @30V | 240 to 238 @14V |
| 147. | Benz[a]anthracene | 20.23 | 1 | 228 to 202 @24V | 226 to 199 @36V | 226 to 200 @28V |
| 148. | 3,3′-Dichlorobenzidine | 20.25 | 1 | 252 to 154 @26V | 252 to 181 @24V | 252 to 182 @20V |
| 149. | 4,4′-Methylenebis(2-chloroaniline) | 20.28 | 1 | 266 to 195 @28V | 231 to 195 @16V | 266 to 231 @12V |
| 150. | Chrysene | 20.3 | 1 | 228 to 202 @22V | 226 to 200 @28V | 228 to 201 @36V |
| 151. | Bis(2-ethylhexyl) phthalate | 20.49 | 1 | 167 to 149 @8V | 149 to 93 @16V | 149 to 121 @14V |
| 152. | 6-Methylchrysene | 21.19 | 1 | 242 to 226 @28V | 239 to 213 @26V | 239 to 237 @32V |
| 153. | Di-n-octyl phthalate | 21.86 | 1 | 149 to 65 @22V | 149 to 93 @16V | 149 to 121 @12V |
| 154. | 7,12-Dimethylbenz[a]anthracene | 22.45 | 1 | 256 to 239 @38V | 241 to 226 @14V | 256 to 241 @12V |
| 155. | Benzo[b]fluoranthene | 22.46 | 1 | 252 to 226 @24V | 250 to 223 @38V | 250 to 224 @24V |
| 156. | Benzo[k]fluoranthene | 22.53 | 1 | 252 to 226 @22V | 250 to 223 @38V | 250 to 224 @26V |
| 157. | Benzo[a]pyrene | 23.21 | 1 | 252 to 226 @26V | 250 to 223 @38V | 250 to 224 @26V |
| 158. | Perylene-d12 | 23.33 | 1 | 264 to 236 @24V | 264 to 260 @34V | 264 to 262 @20V |
| 159. | 3-Methylcholanthrene | 24.16 | 1 | 268 to 252 @32V | 252 to 226 @22V | 268 to 253 @14V |
| 160. | Dibenz(a,h)acridine | 25.55 | 1 | 279 to 252 @34V | 278 to 250 @24V | 278 to 275 @36V |
| 161. | Dibenz[a,j]acridine | 25.74 | 1 | 279 to 250 @38V | 278 to 250 @26V | 279 to 277 @30V |
| 162. | Indeno[1,2,3-cd]pyrene | 26.16 | 1 | 276 to 250 @30V | 276 to 274 @34V | 138 to 125 @12V |
| 163. | Dibenz[a,h]anthracene | 26.24 | 1 | 278 to 252 @26V | 139 to 113 @14V | 139 to 126 @8V |
| 164. | Benzo[ghi]perylene | 26.88 | 1 | 276 to 274 @38V | 138 to 124 @28V | 138 to 125 @12V |
Conditions
| Column | Rxi-SVOCms, 30 m, 0.25 mm ID, 0.25 µm (cat.# 16623) |
|---|---|
| Standard/Sample | |
| SVOC MegaMix 150 kit (cat.# 31907) | |
| SV internal standard mix (cat.# 31206) | |
| Acid surrogate mix (4/89 SOW) (cat.# 31063) | |
| Revised B/N surrogate mix (cat.# 31888) | |
| Diluent: | Dichloromethane |
| Conc.: | 1 µg/mL (0.1 ng on-column) |
| Injection | |
| Inj. Vol.: | 1 µL split (split ratio 10:1) |
| Liner: | Topaz 4.0 mm ID single taper inlet liner w/ wool (cat.# 23303) |
| Inj. Temp.: | 250 °C |
| Oven | |
| Oven Temp.: | 40 °C (hold 1 min) to 280 °C at 12.4 °C/min to 330 °C at 3.3 °C/min (hold 3.65 min) |
| Carrier Gas | He, constant flow |
| Flow Rate: | 1.2 mL/min |
| Detector | TSQ8000 |
|---|---|
| Transfer Line Temp.: | 280 °C |
| Analyzer Type: | Quadrupole |
| Source Temp.: | 330 °C |
| Quad Temp.: | 180 °C |
| Electron Energy: | 70 eV |
| Solvent Delay Time: | 2 min |
| Tune Type: | PFTBA |
| Ionization Mode: | EI |
| Instrument | Thermo Trace GC |
| Sample Preparation | Solutions were mixed and diluted to 1 ppm or lower with dichloromethane in amber 2 mL, 9 mm short-cap, screw-thread vials (cat.# 21143) and sealed with 2.0 mL, 9 mm short-cap, screw-vial closures (cat.# 23842). |
| Notes | The compounds in the inserts are at lower concentrations than the full chromatogram and are showing specific transitions as noted in the inserts. The SVOC MegaMix 150 kit (cat.# 31907) contains one ampul each of the following standards: • 8270 MegaMix (cat.# 31850) • SVOC additions standard (cat.# 31909) • Appendix IX mix #1, revised (cat.# 32459) • Methapyrilene (cat.# 32460) • Appendix IX mix #2 (cat.# 31806) • Benzoic acid (cat.# 31879) |
The second challenge presented by GC-MS/MS analysis of semivolatiles using EPA Method 8270E is the response stability of reactive compounds, such as pyridine, benzoic acid, and acrylamide. These compounds exhibit a shift in retention time across different concentrations, which can make detection challenging when using SRM windows. As the concentration of active analytes decreases, the retention time increases (Figure 2). We found that the best concentration range for these compounds is 1-10 ng on-column.
In summary, GC-MS/MS is a great technique for detecting picograms on-column, and it significantly extends the sensitivity range for semivolatile analysis using EPA Method 8270E, albeit with some caveats. Analysts must be aware of calibration range changes, and they must also watch for retentions time shifts for reactive compounds and adjust SRM windows accordingly, if necessary.
Figure 2: Pyridine retention time shifts with decreased concentration.

