Sin categorizar

The Effects of LC Particle Choice on Column Performance: Switching from 3 and 5 µm Fully Porous Particles (FPP) to 5 µm Superficially Porous Particles (SPP)

28 Oct 2020

Introduction

Superficially porous particles (commonly referred to as SPP or “core-shell” particles) have been proven to provide fast and efficient LC separations. These particles feature a solid, impermeable core enveloped by a thin, porous layer of silica that offers significantly higher efficiency and sensitivity than traditional fully porous particles (FPP). In this technical note, we will compare the performance of Raptor 5 µm SPP LC columns to traditional 3 and 5 µm FPP LC columns. We will demonstrate how faster, more efficient analyses can be attained with greater sensitivity and reduced system pressure by switching from fully porous particles to superficially porous particles—without changing instrumentation.

Switching from FPP to SPP: Increased Efficiency, Decreased Pressure

One of the primary advantages of SPP is its ability to provide increased column efficiency, often with similar or even reduced backpressure, when compared to conventional FPP. By increasing efficiency while decreasing pressure, the user can achieve faster analysis times without changing instrumentation. As shown in Figure 1, column backpressure decreases by approximately 50% on average across the instrument flow rates tested (0.2 to 2 mL/min) when switching from a 3 µm FPP column to a Raptor 5 µm SPP column. Although the 5 µm FPP column displays comparable pressure to the Raptor 5 µm SPP column, it is at the expense of efficiency.

Figure 1: Switch from a 3 µm FPP column to a Raptor 5 µm SPP to cut backpressure in half.
chart, line chart

The ratio of theoretical plates (N) at optimal flow divided by pressure can be used as a measure of efficiency for a column. When compared for each column, the Raptor 5 µm SPP column has double the number of plates per unit pressure as the 5 µm FPP column, and four times the number of plates per unit pressure as the 3 µm FPP column (Figure 2).

Alternatively, the relationship between efficiency and pressure can be illustrated by comparing the theoretical plates of each column while holding pressure relatively constant (Figure 3). Again, the Raptor 5 µm SPP column shows a dramatic increase in efficiency at equivalent pressure.

Figure 2: Focusing on plates per unit pressure, switching from 3 and 5 µm FPP columns to Raptor 5 µm SPP columns offers clear advantages.
figure article GNAR2109 02
Figure 3: With equivalent pressures, Raptor 5 µm superficially porous particles exhibit significantly more plates than 3 and 5 µm fully porous particles.
chart, bar chart, waterfall chart

The Raptor 5 µm column has nearly twice the plates (i.e., greater efficiency) as the fully porous 3 and 5 µm particle columns.

Switching from FPP to SPP: Increased Efficiency, Faster Run Times

The invention of SPP has provided analysts with fast separations without the need for expensive Ultra High Performance Liquid Chromatography (UHPLC) instruments, thereby increasing sample throughput without capital investment. To investigate the increased speed and efficiency of superficially porous particles, the Raptor Biphenyl 5 µm SPP column was compared to columns packed with fully porous 3 and 5 µm particles by performing an assay using identical instrumentation and isocratic method conditions (Figure 4).

Figure 4: Cut analysis times almost in half by switching to 5 µm SPP with lower pressure over 3 µm FPP and far superior efficiency over 5 µm FPP.
Comparison: Efficiency and Analysis Time for Raptor™ Biphenyl 5 µm and Fully Porous 5 µm and 3 µm Columns

LC_GN0556

Peaks

PeaksConc.
(mg/mL)
1.Uracil0.02
2.Benzene3.0
3.Naphthalene0.5
4.Biphenyl0.06

Conditions

Column
Temp.:30 °C
Standard/SampleLC Reversed Phase Test Mix #1 (cat.# 35005)
Diluent:Methanol:water (75:25)
Inj. Vol.:2 µL
Mobile Phase
A:Water
B:Acetonitrile
Time (min)Flow (mL/min)%A%B
0.000.84555
200.84555
DetectorDAD @ 254 nm
InstrumentHPLC
NotesAll columns were 150 mm x 4.6 mm ID. Values for efficiency (N) were calculated from the Biphenyl peak.

The last analyte eluted on the Raptor Biphenyl 5 µm column in 7.1 minutes resulting in a 45% decrease in analysis time compared to the 3 µm and 5 µm fully porous particle columns, which had run times of 13.6 and 12.9 minutes respectively. In addition, a 43% improvement in efficiency was observed for the Raptor Biphenyl 5 µm SPP column over the 5 µm FPP column. The 3 µm FPP column displayed only slightly less efficiency but nearly double the backpressure of the Raptor Biphenyl 5 µm SPP column.

Switching from FPP to SPP: Increased Sensitivity, Better Performance

The performance of existing methods can be instantly improved by replacing conventional 3 µm or 5 µm FPP columns with the Raptor 5 µm column. Enhanced method efficiency, peak width, and analysis time can be achieved under gradient conditions, as demonstrated in Figure 5.

Figure 5: Increase resolution and decrease analysis time by switching from FPP to Raptor SPP columns.
Comparison: Fentanyl, Norfentanyl, and Two Analogues on Raptor™ Biphenyl 5 µm and Fully Porous 5 µm Columns

LC_GN0555

Conditions

Column
Temp.:30 °C
Standard/Sample
Diluent:Water
Conc.:25 µg/mL
Inj. Vol.:1 µL
Mobile Phase
A:Water + 0.1% formic acid
B:Methanol + 0.1% formic acid
Time (min)Flow (mL/min)%A%B
0.000.66535
2.000.61585
2.010.66535
4.000.66535
DetectorPDA @ 210 nm
InstrumentHPLC
NotesAll columns were 50 mm x 2.1 mm ID.

Average resolution values were 6.0 on the Raptor Biphenyl 5 µm SPP column and 4.1 on the 5 µm FPP column. In addition to a shorter run time, average analyte resolution increased 46% when switching to superficially porous particles of the same diameter.

In another example using gradient conditions and mass spectrometry, the analysis time decreased from 2.3 minutes on the 5 µm FPP column to 1.5 min on the Raptor Biphenyl SPP 5 µm column (Figure 6). Meanwhile, average sensitivity for the 5 µm SPP column increased by 64% over the 3 µm FPP column and 128% over the 5 µm FPP column.

Figure 6: Raptor 5 µm SPP columns also provide increased sensitivity over 3 or 5 µm FPP columns.
Comparison: Thyroid Hormones on Raptor™ Biphenyl 5 µm and Fully Porous 5 µm and 3 µm Columns

LC_CF0598

Peaks

Peaks
1.3,3′,5-Triiodo-L-thyronine (T3)
2.3,3′,5-Triiodo-L-thyronine (rT3)
3.L-Thyroxine (T4)

Conditions

Column
Temp.:40 °C
Standard/Sample
Diluent:Water:methanol (35:65)
Conc.:1 ng/mL
Inj. Vol.:2 µL
Mobile Phase
A:0.1% Formic acid in water
B:0.1% Formic acid in methanol
Time (min)Flow (mL/min)%A%B
0.000.43565
2.000.42080
2.010.43565
3.500.43565
DetectorMS/MS
Ion Mode:ESI+
InstrumentHPLC
NotesAll columns were 100 mm x 2.1 mm ID.

Conclusion

Raptor 5 µm LC columns with superficially porous particles (SPP) display lower backpressure and increased efficiency when compared to columns packed with 3 and 5 µm fully porous particles (FPP) of similar dimension. When used in the development of new assays, Raptor 5 µm SPP LC columns offer fast run times and excellent method performance without upgrading instrumentation. When substituted into existing methodologies, switching from conventional 3 µm and 5 µm FPP columns to Raptor 5 µm SPP LC columns has the potential to dramatically decrease analysis times while improving efficiency and sensitivity. Certain assays may require some degree of method development to achieve optimal results, but Raptor 5 µm LC columns are compatible with most assays and offer an excellent way to increase performance without extra cost or labor.

GNAR2109