Comprehensive Per- and Polyfluorinated Substances Profiling in Beverages: Simultaneous Quantification of Ultrashort-Chain to Long-Chain Compounds in Ready-to-Drink Teas and Fruit Juices

26 May 2026

Author(s): Shun-Hsin Liang, Justin A. Steimling
Restek Corporation

Published By: Toxics  

Issue: Volume 14, Issue 5   

Year of Publication: 2026

Link: https://doi.org/10.3390/toxics14050422


Abstract:
Ultrashort-chain (USC) per- and polyfluoroalkyl substances (PFAS) are highly polar, mobile, and persistent emerging pollutants. While the environmental distribution of USC species is well-documented, their presence in widely consumed beverages remains undercharacterized due to the analytical difficulty of capturing such highly polar species. This study established a robust workflow for the simultaneous determination of C1 to C14 perfluoroalkyl carboxylic and sulfonic acids, alongside other PFAS classes, in diverse beverage matrices including teas and fruit juices. Chromatographic separation was achieved using a mixed-mode inert-coated alkyl-phase LC column to enhance USC retention while maintaining performance for longer-chain analytes. A high-throughput, minimal-handling sample preparation was optimized to mitigate matrix effects and contamination. Method performance was evaluated using fortified beverage samples across 2–500 ng/L, with
calibration ranges of 1–2000 ng/L and incorporation of 13 isotopically labeled internal standards. Results demonstrated acceptable accuracy (recoveries within 30% of nominal values) and optimal precision (%RSD < 12%). Application to commercial samples revealed frequent PFAS occurrence, specifically highlighting the prevalence of previously overlooked USC species in the human diet. These results demonstrate that ready-to-drink beverages are a significant pathway for human exposure, necessitating the inclusion of USC compounds in future food safety monitoring and risk assessments.

Autores

  • Shun-Hsin Liang, PhD

    Shun-Hsin is a senior principal scientist in LC Solutions at Restek. He received his bachelor’s degree from the National Taiwan University in 1988 and obtained his PhD from Michigan State University in 1996. He performed postdoctoral research at the University of Michigan from 1996 to 2000 for oncology studies. In 2001, he was appointed as research faculty at The Pennsylvania State University and focused on molecular toxicology research. In 2006, he joined MPI Research Inc. as a senior research scientist and was a study director for GLP analytical projects. In 2013, Dr. Liang joined the LC Solutions department at Restek and specialized in developing application methods across the fields of environmental, food safety, and life sciences.

    View Articles
  • Justin Steimling
GNOT5727