Articles

PFAS Analysis by U.S. FDA C-010.03 for Food and Feed

03 Oct 2025

feature FSAR4292
  • Sample Matrix: food and feed (validated in lettuce, chocolate milk, salmon, bread, eggs, clams, blueberries, silage, and corn snaplage)
  • Sample Preparation: QuEChERS, filtration, solid-phase extraction
  • Detection Technique: LC-MS/MS

Determination of 30 Per- and Polyfluoroalkyl Substances (PFAS) in Food and Feed using Liquid Chromatography-Tandem Mass Spectrometry (LC-MS/MS)

U.S. FDA Method C-010.03 is used for the analysis of 30 PFAS in food and feed. It combines QuEChERS, filtration, and solid phase extraction sample preparation with LC-MS/MS analysis on a C18 column. The method also uses LC-HRMS for verification of PFBA and PFPeA detection in order to address the potential for false positives due to matrix interferences since these compounds each have only one MS/MS transition. The full analyte list and recommended products for PFAS analysis in food and feed by Method C-010.03 are provided below. In addition, you can register and directly compare analyte lists and matrices across PFAS analysis methods here or visit www.restek.com/PFAS to explore our technical resources for PFAS analysis.

Analyte List for FDA C-010.03 for PFAS Samples

PFAS acronyms, compound names, and CAS numbers vary across methods and guidelines, and they also depend on compound form (e.g., neutral vs. salt). The following list uses standardized terminology to allow easy comparison across the methods and guidelines in our comprehensive guide. Please refer to U.S. FDA C-010.03 for method-specific nomenclature.

AcronymCompound NameCAS#
PFBSPerfluoro-1-butanesulfonic acid375-73-5
PFDAPerfluorodecanoic acid335-76-2
PFDoAPerfluorododecanoic acid307-55-1
PFHpAPerfluoroheptanoic acid375-85-9
PFHxSPerfluoro-1-hexanesulfonic acid355-46-4
PFHxAPerfluorohexanoic acid307-24-4
PFNAPerfluorononanoic acid375-95-1
PFOSPerfluorooctanesulfonic acid1763-23-1
PFOAPerfluorooctanoic acid335-67-1
PFTeDAPerfluorotetradecanoic acid376-06-7
PFTrDAPerfluorotridecanoic acid72629-94-8
PFUnAPerfluoroundecanoic acid2058-94-8
HFPO-DA, GenXHexafluoropropylene oxide dimer acid13252-13-6
PFBAPerfluorobutanoic acid375-22-4
PFPeAPerfluoropentanoic acid2706-90-3
PFDSPerfluoro-1-decanesulfonic acid335-77-3
PFHpSPerfluoro-1-heptanesulfonic acid375-92-8
PFPeSPerfluoro-1-pentanesulfonic acid2706-91-4
FOSAPerfluoro-1-octanesulfonamide754-91-6
4:2 FTS1H,1H,2H,2H-Perfluorohexane sulfonic acid757124-72-4
6:2 FTS1H,1H,2H,2H-Perfluorooctane sulfonic acid27619-97-2
8:2 FTS1H,1H,2H,2H-Perfluorodecane sulfonic acid39108-34-4
9Cl-PF3ONS9-Chlorohexadecafluoro-3-oxanonane-1-sulfonic acid (F-53B Major)756426-58-1
11Cl-PF3OUdS11-Chloroeicosafluoro-3-oxaundecane-1-sulfonic acid (F-53B Minor)763051-92-9
ADONA4,8-Dioxa-3H-perfluorononanoic acid919005-14-4
PFNSPerfluoro-1-nonanesulfonic acid68259-12-1
PFDoSPerfluorododecane sulfonate79780-39-5
10:2 FTS1H,1H,2H,2H-perfluorododecane sulfonate (10:2)120226-60-0
PFUnDSPerfluoroundecane sulfonic acid749786-16-1
PFTrDSPerfluorotridecane sulfonic acid791563-89-8

Recommended Products for FDA C-010.03 for PFAS Analysis

Analytical LC Columns:


Raptor C18, 2.7 µm, 50 x 2.1 mm HPLC Column
Force C18, 1.8 µm, 50 x 2.1 mm LC Column

Delay Columns:


PFAS Delay Column, 5 µm, 50 x 2.1 mm HPLC Column

PFAS delay column: 5 µm, 50 x 3.0 mm (cat.# 580959 [custom product])

Guard Columns:


UltraShield UHPLC PreColumn Filter, 0.2 µm Frit, ea.
EXP Direct Connect Holder for EXP Guard Cartridges, Includes Fitting & Ferrules
Raptor C18 EXP Guard Column Cartridge, 2.7 µm, 5 x 2.1 mm, 3-pk.

Sample Prep:


30 mm Syringe Filter, 0.45 µm, Nylon, Pink, 100-pk.
30 mm Syringe Filter, 0.22 µm, Nylon, Pink, 100-pk.
Resprep Polymeric SPE Cartridge, WAX, 6 mL/500 mg, 30 μm, 30-pk.
Resprep Polymeric SPE Cartridge, WAX, 6 mL/200 mg, 30 μm, 30-pk.
Q-sep QuEChERS Extraction Salt Packets Only, Unbuffered FDA Method, 6 g MgSO4, 1.5 g NaCl, 50 Packets

Vacuum Manifolds:


Resprep QR-24 Vacuum Manifold
Resprep QR-12 Vacuum Manifold

Vacuum Pumps:


Chemker 300 PTFE Vacuum Pump, 20 L/min, AC110 V, 60 Hz
Chemker 300 PTFE Vacuum Pump, 18 L/min, AC220 V, 50 Hz

General Lab Supplies:


Limited-Volume Screw-Thread Polypropylene Vials, 9 mm, 700 µL, 12 x 32 mm, 100-pk.
Limited-Volume Screw-Thread Polypropylene Vials, 9 mm, 700 µL, 12 x 32 mm, 1000-pk.
2.0 mL, 9 mm Solid-Top Polyethylene Caps, Screw-Thread, 10 mil thick membrane, Clear, 100-pk.
2.0 mL, 9 mm Solid-Top Polyethylene Caps, Screw-Thread, 10 mil thick membrane, Clear, 1000-pk.
Empty Centrifuge Tubes, 50 mL, Polypropylene w/Cap, 50-pk.
Empty Centrifuge Tubes, 50 mL, Polypropylene w/Cap, 500-pk.
Norm-Ject Plastic Syringe, 10 mL Luer Lock Tip, 100-pk.
Norm-Ject Plastic Syringe, 5 mL Luer Lock Tip, 100-pk.
Icon depicting customer service person.

Need Help with Your PFAS Analysis?

Contact a PFAS Expert

Discover More Solutions for PFAS Analysis!

Start Saving
Time Today

Develop PFAS Methods in Minutes with our Pro EZLC Chromatogram Modeler

Start Saving Time Today

Develop PFAS Methods in Minutes with our Pro EZLC Chromatogram Modeler

Author

  • Restek Corporation

    Restek is a leading provider of chromatography columns, accessories, and certified reference materials. Trust Restek for reliable, high-quality analytical solutions.

    View Articles
FSAR4292
Version: E