Sin categorizar

Analyze Picogram-Level Semivolatiles Accurately with GC-MS/MS and EPA Method 8270E

Featured Application: Semivolatiles by GC-MS/MS on Rxi-SVOCms

05 Jul 2025

feature EVFA4492

EPA Method 8270E now allows for the use of GC tandem mass spectrometry (GC-MS/MS), which can extend detection levels for semivolatiles down into the low picogram range. Previous versions of the method only allowed GC-MS analysis, which is much less sensitive. Much of the method has remained the same, but the lower levels of detection provide new challenges, including a dynamic calibration range and retention time shifts for reactive compounds. The calibration range is limited by lower levels of detection at the lower end as well as by overloading and MS saturation at the higher end. Using an Rxi-SVOCms column (30 m, 0.25 mm ID, 0.25 µm [cat.# 16623]) and the conditions shown in Figure 1, the upper limit was found to be 0.5 ng on-column. This shifts the typical range of calibration (0.1-100 µg/mL) to concentrations 100x to 1000x lower. Note that a wider range could be obtained by adjusting the split ratio.

Figure 1: The main chromatogram is an analysis of semivolatiles at 100 pg on an Rxi-SVOCms column. A calibration curve was also performed, and the inset figures represent the lowest calibration point for selected compounds.

Semivolatiles on Rxi-SVOCms by GC-MS/MS

GC_EV1522

Peaks

PeakstR (min)Conc.
(µg/g)
SRM 1SRM 2SRM 3
1.1,4-Dioxane2.79188 to 28 @14V58 to 28 @6V88 to 57 @8V
2.N-Nitrosodimethylamine3.08174 to 42 @16V74 to 43 @12V74 to 44 @6V
3.Pyridine3.13179 to 51 @26V52 to 26 @18V79 to 52 @12V
4.Ethyl methacrylate3.62199 to 43 @14V69 to 41 @6V99 to 53 @12V
5.2-Picoline3.96193 to 65 @18V66 to 39 @18V93 to 66 @12V
6.N-Nitrosomethylethylamine4.12188 to 42 @16V88 to 43 @8V88 to 71 @6V
7.Acrylamine4.5171 to 44 @22V55 to 27 @6V71 to 55 @6V
8.Methyl methanesulfonate4.52180 to 48 @28V65 to 48 @18V80 to 65 @8V
9.2-Fluorophenol4.731112 to 63 @22V92 to 63 @14V112 to 92 @6V
10.N-Nitrosodiethylamine5.051102 to 44 @10V102 to 56 @12V102 to 85 @6V
11.Ethyl methanesulfonate5.511109 to 45 @10V79 to 48 @34V109 to 79 @6V
12.Benzaldehyde6.011105 to 51 @26V77 to 51 @14V105 to 77 @10V
13.Phenol6.13194 to 39 @30V94 to 65 @16V94 to 66 @10V
14.Phenol-d66.14199 to 69 @18V71 to 42 @16V99 to 71 @8V
15.Aniline6.17193 to 65 @20V66 to 39 @18V93 to 66 @10V
16.Pentachloroethane6.261167 to 95 @32V117 to 82 @30V167 to 132 @14V
17.bis-(2-Chloroethyl)ether6.28193 to 27 @22V63 to 27 @10V93 to 63 @6V
18.2-Chlorophenol6.351128 to 63 @24V64 to 38 @12V128 to 64 @14V
19.Decane6.49185 to 41 @12V71 to 43 @6V85 to 43 @6V
20.1,3-Dichlorobenzene6.611146 to 75 @24V111 to 75 @12V146 to 111 @12V
21.1,4-Dichlorobenzene-d46.711150 to 76 @38V150 to 78 @26V150 to 115 @14V
22.1,4-Dichlorobenzene6.741146 to 75 @24V111 to 75 @12V146 to 111 @14V
23.Benzyl alcohol6.941108 to 77 @22V79 to 51 @20V108 to 79 @12V
24.1,2-Dichlorobenzene6.981146 to 111 @12V111 to 50 @36V111 to 75 @10V
25.Indene7.121116 to 89 @28V115 to 63 @26V115 to 89 @16V
26.2-Methylphenol7.141108 to 77 @24V79 to 51 @20V108 to 80 @8V
27.2,2′-oxybis(1-chloropropane)7.21121 to 45 @6V77 to 51 @12V121 to 77 @8V
28.N-Nitrosopyrrolidine7.341100 to 43 @8V100 to 55 @6V100 to 68 @8V
29.Acetophenone7.371105 to 51 @26V77 to 51 @14V105 to 77 @12V
30.3- and 4-Methylphenol7.411107 to 51 @26V70 to 43 @6V107 to 77 @12V
31.N-Nitrosomorpholine7.421116 to 56 @10V86 to 56 @6V116 to 86 @6V
32.o-Toluidine7.441107 to 77 @26V106 to 77 @16V106 to 79 @8V
33.Hexachloroethane7.561201 to 131 @30V117 to 82 @24V201 to 166 @12V
34.Nitrobenzene-d57.631128 to 54 @28V82 to 54 @12V128 to 82 @12V
35.Nitrobenzene7.681123 to 51 @30V77 to 51 @12V123 to 77 @10V
36.N-Nitrosopiperidine7.941114 to 41 @10V114 to 84 @6V114 to 97 @6V
37.Isophorone8.121138 to 82 @8V82 to 39 @12V82 to 54 @6V
38.2-Nitrophenol8.251139 to 81 @12V109 to 81 @8V139 to 109 @8V
39.2,4-Dimethylphenol8.361122 to 77 @26V107 to 77 @14V122 to 107 @12V
40.O,O,O-Triethyl phosphorothioate8.511198 to 114 @12V121 to 65 @10V198 to 121 @8V
41.Benzoic acid8.531122 to 77 @20V105 to 77 @10V122 to 105 @8V
42.Bis(2-chloroethoxy)methane8.55193 to 63 @6V63 to 26 @36V63 to 27 @12V
43.Phentermine8.641134 to 91 @20V58 to 42 @22V134 to 115 @12V
44.2,4-Dichlorophenol8.681162 to 63 @24V98 to 63 @8V162 to 98 @12V
45.1,2,4-Trichlorobenzene8.841180 to 109 @24V145 to 109 @10V180 to 145 @12V
46.Naphthalene-d88.911136 to 84 @20V136 to 108 @18V136 to 134 @14V
47.Naphthalene8.971129 to 103 @14V128 to 78 @20V128 to 102 @16V
48.a-Terpineol9.041136 to 93 @10V121 to 93 @8V136 to 121 @8V
49.4-Chloroaniline9.091127 to 65 @20V65 to 39 @12V127 to 100 @10V
50.2,6-Dichlorophenol9.111164 to 63 @26V162 to 63 @26V162 to 126 @8V
51.Hexachloropropene9.151215 to 119 @18V213 to 119 @18V215 to 121 @18V
52.Hexachlorobutadiene9.231260 to 190 @26V225 to 190 @14V260 to 225 @12V
53.Quinoline9.61129 to 76 @26V102 to 76 @12V129 to 102 @16V
54.Caprolactam9.71113 to 56 @10V85 to 67 @6V113 to 85 @6V
55.N-Nitrosodibutylamine9.761116 to 74 @8V84 to 56 @14V116 to 99 @6V
56.1,4-Phenylenediamine9.771108 to 80 @20V80 to 53 @12V108 to 81 @10V
57.4-Chloro-3-methylphenol10.041142 to 77 @26V107 to 77 @12V142 to 107 @12V
58.Isosafrole I10.161162 to 104 @12V131 to 103 @8V162 to 131 @12V
59.2-Methylnaphthalene10.281141 to 89 @30V115 to 89 @16V141 to 115 @16V
60.1-Methylnaphthalene10.461141 to 89 @30V115 to 89 @14V141 to 115 @16V
61.Hexachloropentadiene10.591272 to 237 @12V237 to 119 @20V237 to 143 @22V
62.1,2,4,5-Tetrachlorobenzene10.611216 to 108 @36V179 to 143 @8V216 to 181 @14V
63.Isosafrole II10.731162 to 104 @12V131 to 103 @10V162 to 131 @12V
64.2,3-Dichloroaniline10.831163 to 90 @18V161 to 90 @16V161 to 99 @18V
65.2,4,6-Trichlorophenol10.831196 to 97 @24V132 to 97 @10V160 to 97 @16V
66.2,4,5-Trichlorophenol10.881196 to 97 @24V132 to 97 @10V196 to 132 @12V
67.2-Fluorobiphenyl10.991172 to 151 @22V172 to 170 @22V172 to 171 @12V
68.Safrole11.161162 to 77 @30V131 to 103 @10V162 to 127 @16V
69.2-Chloronaphthalene11.21162 to 77 @32V127 to 77 @16V162 to 127 @16V
70.Biphenyl11.21153 to 126 @32V154 to 115 @24V154 to 152 @22V
71.1-Chloronaphthalene11.251162 to 77 @30V127 to 77 @16V162 to 127 @16V
72.o-Nitroaniline11.411138 to 65 @22V108 to 80 @8V138 to 92 @12V
73.Diphenyl ether11.421170 to 141 @16V141 to 115 @14V170 to 142 @10V
74.1,4-Naphthoquinone11.561158 to 102 @14V102 to 76 @12V158 to 130 @8V
75.1,2-Dinitrobenzene11.71168 to 75 @20V122 to 75 @12V168 to 122 @8V
76.1,3-Dinitrobenzene11.831168 to 75 @22V76 to 50 @10V168 to 122 @8V
77.Diphenyl phthalate11.831163 to 77 @20V163 to 92 @24V163 to 133 @8V
78.1,6-Dinitrotoluene11.911165 to 63 @20V148 to 90 @10V165 to 148 @8V
79.Acenaphthylene11.981152 to 76 @36V152 to 102 @26V152 to 126 @24V
80.1,4-Dinitrobenzene11.991168 to 51 @14V76 to 63 @6V168 to 63 @26V
81.3-Nitroaniline12.211138 to 65 @20V92 to 65 @8V138 to 92 @12V
82.Acenaphthene-d1012.241164 to 160 @32V162 to 160 @18V164 to 162 @14V
PeakstR (min)Conc.
(µg/g)
SRM 1SRM 2SRM 3
83.Acenaphthene12.321153 to 77 @38V153 to 126 @36V154 to 127 @34V
84.2,4-Dinitrophenol12.411184 to 107 @10V154 to 79 @12V184 to 154 @6V
85.4-Nitrophenol12.571139 to 81 @14V109 to 81 @8V139 to 109 @6V
86.Pentachlorobenzene12.61250 to 142 @38V215 to 179 @10V250 to 215 @16V
87.2,4-Dinitrotoluene12.661165 to 63 @22V89 to 63 @12V165 to 119 @6V
88.Dibenzofuran12.661168 to 139 @22V139 to 63 @30V139 to 89 @18V
89.1-Naphthylamine12.811143 to 115 @22V115 to 89 @16V143 to 116 @10V
90.2,3,5,6-Tetrachlorophenol12.831234 to 133 @26V232 to 131 @26V232 to 168 @12V
91.2,3,4,6-Tetrachlorophenol12.911234 to 131 @24V232 to 133 @26V232 to 168 @12V
92.2-Naphthylamine12.971143 to 115 @24V115 to 89 @16V143 to 116 @10V
93.Diethyl Phthalate13.211177 to 149 @8V149 to 65 @20V149 to 93 @12V
94.Hexadecane13.29199 to 41 @14V85 to 43 @6V99 to 57 @6V
95.Fluorene13.311167 to 166 @14V165 to 115 @24V165 to 139 @26V
96.Zinophos13.361143 to 79 @10V107 to 52 @20V107 to 79 @14V
97.5-Nitro-o-toluidine13.371152 to 77 @24V106 to 77 @12V152 to 106 @10V
98.4-Chlorophenyl phenyl ether13.371204 to 77 @22V141 to 115 @14V204 to 141 @12V
99.4-Nitroaniline13.371138 to 80 @18V108 to 80 @10V138 to 108 @8V
100.4,6-Dinitro-2-methylphenol13.451198 to 121 @10V168 to 121 @6V198 to 168 @6V
101.Diphenylamine13.61169 to 66 @22V169 to 77 @30V170 to 66 @22V
102.Azobenzene13.671182 to 77 @12V77 to 51 @14V182 to 105 @6V
103.2,4,6-Tribromophenol13.771332 to 143 @34V330 to 141 @36V330 to 222 @20V
104.Sulfotepp141322 to 146 @24V202 to 146 @10V322 to 202 @10V
105.1,3,5-Trinitrobenzene14.141213 to 74 @38V213 to 120 @22V213 to 167 @8V
106.Diallate 114.21234 to 150 @16V86 to 43 @6V234 to 192 @12V
107.Phorate14.221121 to 47 @26V75 to 47 @8V121 to 65 @8V
108.Phenacetin14.251179 to 109 @14V108 to 80 @12V179 to 137 @8V
109.4-Bromophenyl phenyl ether14.311250 to 141 @16V141 to 115 @12V248 to 141 @14V
110.Diallate 214.371234 to 150 @16V86 to 43 @6V234 to 192 @10V
111.Hexachlorobenzene14.371288 to 216 @26V249 to 214 @12V284 to 249 @14V
112.Dimethoate14.511125 to 47 @14V93 to 63 @8V125 to 79 @8V
113.Atrazine14.71215 to 58 @10V200 to 122 @8V215 to 200 @8V
114.Pentachlorophenol14.771270 to 169 @22V228 to 165 @14V266 to 167 @22V
115.4-Aminobiphenyl14.791169 to 141 @22V168 to 141 @8V168 to 167 @10V
116.Pentachloronitrobenzene14.811295 to 237 @14V237 to 143 @20V295 to 265 @8V
117.Propyzamide151254 to 191 @16V173 to 145 @14V254 to 226 @10V
118.Octadecane15.1199 to 41 @14V85 to 43 @6V99 to 57 @6V
119.Phenanthrene-d1015.11188 to 160 @20V184 to 154 @32V184 to 156 @22V
120.Phenanthrene15.161178 to 151 @32V176 to 150 @20V178 to 152 @20V
121.Dinoseb15.211240 to 117 @24V163 to 116 @14V240 to 211 @8V
122.Disulfoton15.22197 to 65 @16V88 to 60 @6V97 to 79 @12V
123.Anthracene15.261178 to 151 @32V177 to 151 @18V178 to 152 @18V
124.Carbazole15.61167 to 139 @24V166 to 140 @14V167 to 166 @12V
125.Methyl Parathion15.951263 to 79 @24V125 to 47 @12V263 to 109 @10V
126.di-n-Butyl phthalate16.411149 to 65 @20V149 to 93 @14V149 to 121 @12V
127.4-Nitroquinoline-N-oxide16.691190 to 77 @20V101 to 75 @10V190 to 160 @8V
128.Ethyl Parathion16.731291 to 81 @26V109 to 81 @10V291 to 109 @10V
129.Methapyrilene16.91190 to 97 @14V97 to 53 @16V190 to 157 @8V
130.Isodrin17.191261 to 191 @28V193 to 123 @28V261 to 226 @16V
131.Fluoranthene17.471202 to 152 @30V200 to 174 @22V202 to 176 @26V
132.Benzidine17.81184 to 130 @26V184 to 156 @18V184 to 166 @16V
133.Pyrene-d1017.841212 to 208 @36V212 to 210 @20V0
134.Pyrene17.881200 to 149 @34V200 to 150 @26V200 to 174 @24V
135.p-Terphenyl-d1418.31244 to 226 @18V244 to 240 @22V244 to 242 @14V
136.Aramite 118.331319 to 123 @10V175 to 107 @14V319 to 185 @6V
137.Aramite 218.491319 to 107 @36V185 to 63 @12V319 to 185 @6V
138.p-Dimethylaminoazobenzene18.571225 to 120 @10V120 to 77 @16V225 to 148 @6V
139.Chlorobenzilate18.711251 to 111 @30V139 to 111 @12V251 to 139 @12V
140.Famphur19.171218 to 109 @14V125 to 47 @14V125 to 79 @6V
141.3,3′-Dimethylbenzidine19.211212 to 180 @22V211 to 195 @16V211 to 196 @8V
142.Kepone19.241272 to 237 @12V237 to 119 @24V237 to 143 @22V
143.Benzyl butyl phthalate19.311206 to 149 @8V149 to 65 @20V149 to 121 @10V
144.Bis(2-ethylhexyl) adipate19.551129 to 55 @14V111 to 83 @6V129 to 83 @8V
145.2-Acetylaminofluorene19.691223 to 180 @26V181 to 152 @34V223 to 181 @10V
146.Chrysene-d1220.231240 to 212 @22V240 to 236 @30V240 to 238 @14V
147.Benz[a]anthracene20.231228 to 202 @24V226 to 199 @36V226 to 200 @28V
148.3,3′-Dichlorobenzidine20.251252 to 154 @26V252 to 181 @24V252 to 182 @20V
149.4,4′-Methylenebis(2-chloroaniline)20.281266 to 195 @28V231 to 195 @16V266 to 231 @12V
150.Chrysene20.31228 to 202 @22V226 to 200 @28V228 to 201 @36V
151.Bis(2-ethylhexyl) phthalate20.491167 to 149 @8V149 to 93 @16V149 to 121 @14V
152.6-Methylchrysene21.191242 to 226 @28V239 to 213 @26V239 to 237 @32V
153.Di-n-octyl phthalate21.861149 to 65 @22V149 to 93 @16V149 to 121 @12V
154.7,12-Dimethylbenz[a]anthracene22.451256 to 239 @38V241 to 226 @14V256 to 241 @12V
155.Benzo[b]fluoranthene22.461252 to 226 @24V250 to 223 @38V250 to 224 @24V
156.Benzo[k]fluoranthene22.531252 to 226 @22V250 to 223 @38V250 to 224 @26V
157.Benzo[a]pyrene23.211252 to 226 @26V250 to 223 @38V250 to 224 @26V
158.Perylene-d1223.331264 to 236 @24V264 to 260 @34V264 to 262 @20V
159.3-Methylcholanthrene24.161268 to 252 @32V252 to 226 @22V268 to 253 @14V
160.Dibenz(a,h)acridine25.551279 to 252 @34V278 to 250 @24V278 to 275 @36V
161.Dibenz[a,j]acridine25.741279 to 250 @38V278 to 250 @26V279 to 277 @30V
162.Indeno[1,2,3-cd]pyrene26.161276 to 250 @30V276 to 274 @34V138 to 125 @12V
163.Dibenz[a,h]anthracene26.241278 to 252 @26V139 to 113 @14V139 to 126 @8V
164.Benzo[ghi]perylene26.881276 to 274 @38V138 to 124 @28V138 to 125 @12V

Conditions

ColumnRxi-SVOCms, 30 m, 0.25 mm ID, 0.25 µm (cat.# 16623)
Standard/Sample
SVOC MegaMix 150 kit (cat.# 31907)
SV internal standard mix (cat.# 31206)
Acid surrogate mix (4/89 SOW) (cat.# 31063)
Revised B/N surrogate mix (cat.# 31888)
Diluent:Dichloromethane
Conc.:1 µg/mL (0.1 ng on-column)
Injection
Inj. Vol.:1 µL split (split ratio 10:1)
Liner:Topaz 4.0 mm ID single taper inlet liner w/ wool (cat.# 23303)
Inj. Temp.:250 °C
Oven
Oven Temp.:40 °C (hold 1 min) to 280 °C at 12.4 °C/min to 330 °C at 3.3 °C/min (hold 3.65 min)
Carrier GasHe, constant flow
Flow Rate:1.2 mL/min
DetectorTSQ8000
Transfer Line Temp.:280 °C
Analyzer Type:Quadrupole
Source Temp.:330 °C
Quad Temp.:180 °C
Electron Energy:70 eV
Solvent Delay Time:2 min
Tune Type:PFTBA
Ionization Mode:EI
InstrumentThermo Trace GC
Sample PreparationSolutions were mixed and diluted to 1 ppm or lower with dichloromethane in amber 2 mL, 9 mm short-cap, screw-thread vials (cat.# 21143) and sealed with 2.0 mL, 9 mm short-cap, screw-vial closures (cat.# 23842).
NotesThe compounds in the inserts are at lower concentrations than the full chromatogram and are showing specific transitions as noted in the inserts.

The SVOC MegaMix 150 kit (cat.# 31907) contains one ampul each of the following standards:
• 8270 MegaMix (cat.# 31850)
• SVOC additions standard (cat.# 31909)
• Appendix IX mix #1, revised (cat.# 32459)
• Methapyrilene (cat.# 32460)
• Appendix IX mix #2 (cat.# 31806)
• Benzoic acid (cat.# 31879)

The second challenge presented by GC-MS/MS analysis of semivolatiles using EPA Method 8270E is the response stability of reactive compounds, such as pyridine, benzoic acid, and acrylamide. These compounds exhibit a shift in retention time across different concentrations, which can make detection challenging when using SRM windows. As the concentration of active analytes decreases, the retention time increases (Figure 2). We found that the best concentration range for these compounds is 1-10 ng on-column.

In summary, GC-MS/MS is a great technique for detecting picograms on-column, and it significantly extends the sensitivity range for semivolatile analysis using EPA Method 8270E, albeit with some caveats. Analysts must be aware of calibration range changes, and they must also watch for retentions time shifts for reactive compounds and adjust SRM windows accordingly, if necessary.

Figure 2: Pyridine retention time shifts with decreased concentration.

figure article EVFA4492 02

Products Mentioned


Rxi-SVOCms GC Capillary Column, 30 m, 0.25 mm ID, 0.25 µm
Topaz, Single Taper Inlet Liner, 4.0 mm x 6.5 x 78.5, for Agilent GCs, w/Quartz Wool, Premium Deactivation, 5-pk.
SVOC MegaMix 150 Kit, 1 mL/ampul; 6 ampuls/kit
SV Internal Standard Mix, 2000 µg/mL, Methylene Chloride, 1 mL/ampul
Acid Surrogate Mix (4/89 SOW), 10,000 µg/mL, Methanol, 1 mL/ampul
Revised B/N Surrogate Mix, 5000 µg/mL, Methylene Chloride, 1 mL/ampul
Short-Cap Vial with Grad Marking Spot, 9-425 Screw-Thread, 2 mL, 9 mm, 12 x 32 (vial only), Amber, 1000-pk.
Short Screw Caps, Polypropylene, Ribbed, Screw-Thread, PTFE/Silicone Lined, for Agilent 7693A, Blue, Preassembled, 2.0 mL, 9 mm, 1000-pk.

Author

  • Restek Corporation

    Restek is a leading provider of chromatography columns, accessories, and certified reference materials. Trust Restek for reliable, high-quality analytical solutions.

    View Articles
EVFA4492