{"id":39968,"date":"2023-05-18T00:00:00","date_gmt":"2023-05-18T00:00:00","guid":{"rendered":"https:\/\/discover.restek.com\/uncategorized\/lc-msms-analysis-of-25-underivatized-acylcarnitines-for-differential-diagnosis\/"},"modified":"2025-12-12T17:00:17","modified_gmt":"2025-12-12T17:00:17","slug":"lc-msms-analysis-of-25-underivatized-acylcarnitines-for-differential-diagnosis","status":"publish","type":"post","link":"https:\/\/discover.restek.com\/es\/application-notes\/cfan4029\/lc-msms-analysis-of-25-underivatized-acylcarnitines-for-differential-diagnosis","title":{"rendered":"LC-MS\/MS Analysis of 25 Underivatized Acylcarnitines for Differential Diagnosis"},"content":{"rendered":"\n<h2 class=\"wp-block-heading\">Abstract<\/h2>\n\n\n\n<p class=\"wp-block-paragraph\">Acylcarnitines are important indicators in the diagnosis of many metabolic disorders. This rapid, 9-minute acylcarnitines LC-MS\/MS analysis allows testing of 25 underivatized acylcarnitines, including several isobaric compounds that are essential for differential diagnosis. A Raptor ARC-18 column (100 x 2.1 mm, 2.7 \u00b5m) used under the established conditions provided chromatographic separation of key isobars that shared MRM transitions.<\/p>\n\n\n\n<h2 class=\"wp-block-heading\">Introduction<\/h2>\n\n\n\n<p class=\"wp-block-paragraph\">Carnitine (Figure 1) is a quaternary ammonium compound that is required for fatty acid \u03b2-oxidation. In the body, it can exist either in free form or bind with fatty acids to form acylcarnitines. It can also vary in chain length from shorter acetyl chains to longer palmitoyl chains (Figure 2) [1].<\/p>\n\n\n<div class=\"wp-block-custom-chromatogram-article-top\"><div class=\"chromatogram-article-placeholder\"><div class=\"figure-heading\"><strong>Figure 1:<\/strong>\u00a0Chemical structure of carnitine.<\/div><div class='chromatogram-article-inner-full'><div class=\"chromatogram-article-inner\">\n<style>.kb-image39968_9fb629-c8 .kb-image-has-overlay:after{opacity:0.3;}<\/style>\n<div class=\"wp-block-kadence-image kb-image39968_9fb629-c8\"><figure class=\"aligncenter size-large\"><img loading=\"lazy\" decoding=\"async\" width=\"1024\" height=\"114\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/figure_article_cfan4029_01-1024x114.jpg\" alt=\"\" class=\"kb-img wp-image-18736\" title=\"-\" srcset=\"https:\/\/discover.restek.com\/wp-content\/uploads\/figure_article_cfan4029_01-1024x114.jpg 1024w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure_article_cfan4029_01-300x33.jpg 300w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure_article_cfan4029_01-768x85.jpg 768w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure_article_cfan4029_01-1536x171.jpg 1536w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure_article_cfan4029_01.jpg 1800w\" sizes=\"auto, (max-width: 1024px) 100vw, 1024px\" \/><\/figure><\/div>\n\n<\/div><\/div><\/div>\n\n<div class=\"wp-block-custom-chromatogram-article-top\"><div class=\"chromatogram-article-placeholder\"><div class=\"figure-heading\"><strong>Figure 2:<\/strong>\u00a0Acylcarnitines exist in two forms: short chain and long chain.<\/div><div class='chromatogram-article-inner-full'><div class=\"chromatogram-article-inner\">\n<style>.kb-image39968_100b51-2c .kb-image-has-overlay:after{opacity:0.3;}<\/style>\n<div class=\"wp-block-kadence-image kb-image39968_100b51-2c\"><figure class=\"aligncenter size-large\"><img loading=\"lazy\" decoding=\"async\" width=\"1024\" height=\"327\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/figure_article_cfan4029_02-1024x327.jpg\" alt=\"\" class=\"kb-img wp-image-18742\" title=\"-\" srcset=\"https:\/\/discover.restek.com\/wp-content\/uploads\/figure_article_cfan4029_02-1024x327.jpg 1024w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure_article_cfan4029_02-300x96.jpg 300w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure_article_cfan4029_02-768x245.jpg 768w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure_article_cfan4029_02-1536x491.jpg 1536w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure_article_cfan4029_02.jpg 1800w\" sizes=\"auto, (max-width: 1024px) 100vw, 1024px\" \/><\/figure><\/div>\n\n<\/div><\/div><\/div>\n\n\n<div style=\"height:50px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n\n\n\n<p class=\"wp-block-paragraph\">Many different organic acid and fatty acid oxidation disorders can be diagnosed using acylcarnitine profiles in blood or plasma samples. Acylcarnitines are especially important in newborn testing, which has become mandatory in many places [2]. It is important for testing to be done as soon as possible because if proper treatment is prescribed early, it can greatly reduce the risk of long-term effects. Some of these disorders present later in life, but in most cases, accurate newborn screening and early treatment can minimize their impact.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Restek previously developed a rapid screening method for acylcarnitines using a Raptor HILIC-Si EXP guard column cartridge [3]. That method can improve productivity by allowing combined screening of acylcarnitines and amino acids without derivatization, but it does not provide chromatographic separation of several compounds that are used for differential diagnosis. While high-throughput screening methods are valuable, the method developed here takes a different approach and focuses on the chromatographic separation of several key acylcarnitines that are used for diagnosing different disorders [2]. In addition, while methods have been developed using either underivatized or derivatized sample preparation techniques [4], this method was developed without derivatization in order to simplify and speed up sample preparation.<\/p>\n\n\n\n<h2 class=\"wp-block-heading\">Experimental<\/h2>\n\n\n\n<h3 class=\"wp-block-heading\"><em>Plasma Samples<\/em><\/h3>\n\n\n\n<p class=\"wp-block-paragraph\">For plasma, 100 \u00b5L of sample was added to a microcentrifuge tube along with 5 \u00b5L of internal standard (5 ng\/mL of internal standard working stock containing the D3 internal standards shown in Table I). Next the sample was vortexed for 10 seconds. After vortexing, the samples were incubated for 10 minutes at ambient temperature. Following incubation, 300 \u00b5L of methanol was added to each sample, and the samples were vortexed again for 10 seconds. The samples were then centrifuged for 10 minutes at 4000 rpm. Finally, 100 \u00b5L of the supernatant was aliquoted into a vial containing 900 \u00b5L of mobile phase A (0.1% formic acid in water) and vortexed for 10 seconds before being injected into the LC-MS\/MS for analysis. While a 10x final dilution was appropriate for the instrument used here, a different dilution factor may be necessary for instruments with lower or higher sensitivity.<\/p>\n\n\n\n<h3 class=\"wp-block-heading\"><em>Calibration Standards and Quality Control Samples<\/em><\/h3>\n\n\n\n<p class=\"wp-block-paragraph\">Preliminary experiments determined that plasma would not be a suitable matrix for the standards and QC samples used for method evaluation due to the endogenous levels of acylcarnitine. For this reason, multiple surrogate matrices were investigated and 100 \u00b5g\/mL of bovine serum albumin (BSA) in water was found to be the best surrogate matrix. Calibration and QC stock solutions were then made across a 25\u20135000 ng\/mL concentration range in 1000 \u00b5L of BSA, and 100 \u00b5L of each solution was added to a microcentrifuge tube and processed following the same sample preparation procedure that was previously described for the plasma samples. Note that not all compounds were included when evaluating method performance. Instead, C5-valeryl-L-carnitine was chosen as a representative analyte for the quantitative and standard addition experiments due to its mid-range polarity and presence in a group of isomers.<\/p>\n\n\n\n<h3 class=\"wp-block-heading\"><em>Instrument Conditions<\/em><\/h3>\n\n\n\n<p class=\"wp-block-paragraph\">Analysis of acylcarnitines by LC-MS\/MS was performed on a Shimadzu Nexera UHPLC paired with a SCIEX 4500 MS\/MS using the following instrument conditions and the analyte transitions presented in Table I.<\/p>\n\n\n\n<figure class=\"wp-block-table\"><table style=\"border-style:none;border-width:0px\"><tbody><tr><td>Column:<\/td><td>Raptor ARC-18 (2.7 \u00b5m, 100 mm x 2.1 mm ID [cat.# 9314A12])<\/td><\/tr><tr><td>Guard column:<\/td><td>Raptor ARC-18 EXP guard column cartridge (2.7 \u00b5m, 5 mm x 2.1 mm [cat.# 9314A0252])<\/td><\/tr><tr><td>Column temp.:<\/td><td>35 \u00b0C<\/td><\/tr><tr><td>Injection volume:<\/td><td>3 \u00b5L<\/td><\/tr><tr><td>Mobile phase A:<\/td><td>0.1% Formic acid in water<\/td><\/tr><tr><td>Mobile phase B:<\/td><td>0.1% Formic acid in acetonitrile:isopropanol (90:10)<\/td><\/tr><tr><td>&nbsp;<\/td><td>Time (min)   |   %B<\/td><\/tr><tr><td><\/td><td>0.00 | 2<\/td><\/tr><tr><td><\/td><td>4.00 | 12<\/td><\/tr><tr><td><\/td><td>6.00 | 100<\/td><\/tr><tr><td><\/td><td>7.50 | 100<\/td><\/tr><tr><td><\/td><td>8.00 | 2<\/td><\/tr><tr><td><\/td><td>9.00 | stop<\/td><\/tr><tr><td>Flow rate:<\/td><td>0.6 mL\/min<\/td><\/tr><tr><td>Ion mode:<\/td><td>Positive ESI<\/td><\/tr><\/tbody><\/table><\/figure>\n\n\n\n<p class=\"wp-block-paragraph\"><strong>Table I: <\/strong>Ion transitions for acylcarnitines LC-MS\/MS analysis.<\/p>\n\n\n\n<figure class=\"wp-block-table\"><table style=\"border-width:1px\"><tbody><tr><td><strong>Analyte<\/strong><\/td><td><strong>Precursor<br>Ion<\/strong><\/td><td><strong>Product<br>Ion<\/strong><\/td><\/tr><tr><td>C0-L-carnitine<\/td><td>162.2<\/td><td>85.1<\/td><\/tr><tr><td>C0-L-carnitine-D3<\/td><td>165.2<\/td><td>85.1<\/td><\/tr><tr><td>C2-Acetyl-L-carnitine<\/td><td>204.2<\/td><td>85.1<\/td><\/tr><tr><td>C2-Acetyl-L-carnitine-D3<\/td><td>207.2<\/td><td>85.1<\/td><\/tr><tr><td>Methyl-malonyl-L-carnitine<\/td><td>262.3<\/td><td>85.1<\/td><\/tr><tr><td>Methylmalonyl-L-carnitine-D3<\/td><td>265.3<\/td><td>85.1<\/td><\/tr><tr><td>C3-Propionyl-L-carnitine<\/td><td>218.1<\/td><td>85.1<\/td><\/tr><tr><td>C3-Propionyl-L-carnitine-D3<\/td><td>221.1<\/td><td>85.1<\/td><\/tr><tr><td>3-Hydroxyisovaleryl-L-carnitine<\/td><td>262.4<\/td><td>85.1<\/td><\/tr><tr><td>3-Hydroxyisovaleryl-L-carnitine-D3<\/td><td>265.1<\/td><td>85.1<\/td><\/tr><tr><td>C4-Isobutyryl-L-carnitine<\/td><td>232.1<\/td><td>85.1<\/td><\/tr><tr><td>C4-Isobutyryl-L-carnitine-D3<\/td><td>235.1<\/td><td>85.1<\/td><\/tr><tr><td>C4-Butyryl-L-carnitine<\/td><td>232.2<\/td><td>85.1<\/td><\/tr><tr><td>C4-Butyryl-L-carnitine-D3<\/td><td>235.2<\/td><td>85.1<\/td><\/tr><tr><td>3-Methylcrotonyl-L-carnitine<\/td><td>244.1<\/td><td>85.1<\/td><\/tr><tr><td>3-Methylcrotonyl-L-carnitine-D3<\/td><td>247.3<\/td><td>85.1<\/td><\/tr><tr><td>C5:1-Tigyl-L-carnitine<\/td><td>244.2<\/td><td>85.1<\/td><\/tr><tr><td>2-Methylbutyryl-L-carnitine<\/td><td>246.1<\/td><td>85.1<\/td><\/tr><tr><td>C5-Isovaleryl-L-carnitine<\/td><td>246.1<\/td><td>85.1<\/td><\/tr><tr><td>C5-Isovaleryl-L-carnitine-D3<\/td><td>249.2<\/td><td>85.1<\/td><\/tr><tr><td>C5-Valeryl-L-carnitine<\/td><td>246.2<\/td><td>85.1<\/td><\/tr><tr><td>C5-Valeryl-L-carnitine-D3<\/td><td>249.1<\/td><td>85.1<\/td><\/tr><tr><td>C6-Hexanoyl-L-carnitine<\/td><td>260.3<\/td><td>85.1<\/td><\/tr><tr><td>C6-Hexanoyl-L-carnitine-D3<\/td><td>263.2<\/td><td>85.1<\/td><\/tr><tr><td>C7-Heptanoyl-L-carnitine<\/td><td>274.2<\/td><td>85.1<\/td><\/tr><tr><td>C8-Octanoyl-L-carnitine<\/td><td>288.4<\/td><td>85.1<\/td><\/tr><tr><td>C8-Octanoyl-L-carnitine-D3<\/td><td>291.2<\/td><td>85.1<\/td><\/tr><tr><td>C10-Decanoyl-L-carnitine<\/td><td>316.3<\/td><td>85.1<\/td><\/tr><tr><td>C10-Decanoyl-L-carnitine-D3<\/td><td>319.2<\/td><td>85.1<\/td><\/tr><tr><td>C12-Lauroyl-L-carnitine<\/td><td>344.5<\/td><td>85.1<\/td><\/tr><tr><td>C12-Lauroyl-L-carnitine-D3<\/td><td>347.3<\/td><td>85.1<\/td><\/tr><tr><td>C14:2-Tetradecadienoyl-L-carnitine<\/td><td>368.5<\/td><td>85.1<\/td><\/tr><tr><td>C14:1 Tetradecanoyl-L-carnitine<\/td><td>369.8<\/td><td>85.1<\/td><\/tr><tr><td>C14-Myristoyl-L-carnitine<\/td><td>372.4<\/td><td>85.1<\/td><\/tr><tr><td>C14-Myristoyl-L-carnitine-D3<\/td><td>375.3<\/td><td>85.1<\/td><\/tr><tr><td>C16:1 Palmitolelyl-L-carnitine<\/td><td>398.4<\/td><td>85.1<\/td><\/tr><tr><td>C16-Palmitoyl-L-carnitine<\/td><td>400.5<\/td><td>85.1<\/td><\/tr><tr><td>C16-Palmitoyl-L-carnitine-D3<\/td><td>403.3<\/td><td>85.1<\/td><\/tr><tr><td>C18:2 Linoleoyl-L-carnitine<\/td><td>424.1<\/td><td>85.1<\/td><\/tr><tr><td>C18:1 Oleoyl-L-carnitine<\/td><td>426.4<\/td><td>85.1<\/td><\/tr><tr><td>C18-Oleoyl-L-carnitine-D3<\/td><td>429.4<\/td><td>85.1<\/td><\/tr><tr><td>C18 &#8211; Stearoyl-L-carnitine<\/td><td>429.1<\/td><td>85.1<\/td><\/tr><tr><td>C18 &#8211; Stearoyl-L-carnitine-D3<\/td><td>431.9<\/td><td>85.1<\/td><\/tr><\/tbody><\/table><\/figure>\n\n\n\n<h2 class=\"wp-block-heading\">Results and Discussion<\/h2>\n\n\n\n<h3 class=\"wp-block-heading\"><em>Chromatographic Performance<\/em><\/h3>\n\n\n\n<p class=\"wp-block-paragraph\">Simultaneous LC-MS\/MS analysis of 25 underivatized acylcarnitines was achieved in a fast 9-minute cycle time on a Raptor ARC-18 column as demonstrated in Figure 3. This method uses a quick and simple sample preparation procedure and provides the chromatographic separation of multiple isobars that is essential for differential diagnosis.<\/p>\n\n\n<div class=\"wp-block-custom-chromatogram-article-top\"><div class=\"chromatogram-article-placeholder\"><div class=\"figure-heading\"><strong>Figure 3:<\/strong>\u00a025 Underivatized acylcarnitines were chromatographically separated, allowing for positive identification of compounds that are critical for differential diagnosis (internal standards not shown).<\/div><div class='chromatogram-article-inner-full'><div class=\"chromatogram-article-inner\">\n<div class=\"wp-block-custom-chromatogram-article\"><div class=\"wp-block-custom-chromatogram-article\"><div class=\"chromatogram-image wide-image\"><img decoding=\"async\" src=\"https:\/\/ez.restek.com\/images\/cgram\/lc_cf0782.png\" alt=\"Separation of 25 Acylcarnitines on Raptor ARC-18\" title=\"-\"><\/div><p class=\"article-id\" style=\"text-align:center\"> LC_CF0782<\/p><div class=\"chromatogram-peaks\"><h4>Peaks<\/h4><table class=\"peaks col-lg-6 col-12 peak-50\">\n<thead><tr><th><\/th><th style=\"text-align: left;width: 75px\">Peaks<\/th><th style=\"text-align: center;width: 75px\">t<sub>R<\/sub> (min)<\/th><th style=\"text-align: center;width: 75px\">Precursor Ion<\/th><th style=\"text-align: center;width: 75px\">Product Ion<\/th><\/tr><\/thead>\n<tbody><tr><td class=\"num\">1.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for C0-L-Carnitine\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/6645-46-1\/C0-L-Carnitine\" target=\"_blank\" rel=\"noopener\">C0-L-Carnitine<\/a><\/td><td class=\"oth\">0.40<\/td><td class=\"oth\">162.2<\/td><td class=\"oth\">85.1<\/td><\/tr>\n<tr><td class=\"num\">2.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for C2-Acetyl-L-carnitine\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/5080-50-2\/C2-Acetyl-L-carnitine\" target=\"_blank\" rel=\"noopener\">C2-Acetyl-L-carnitine<\/a><\/td><td class=\"oth\">0.59<\/td><td class=\"oth\">204.2<\/td><td class=\"oth\">85.1<\/td><\/tr>\n<tr><td class=\"num\">3.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Methyl-malonyl-L-carnitine\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/251976-17-7\/Methyl-malonyl-L-carnitine\" target=\"_blank\" rel=\"noopener\">Methyl-malonyl-L-carnitine<\/a><\/td><td class=\"oth\">1.15<\/td><td class=\"oth\">262.3<\/td><td class=\"oth\">85.1<\/td><\/tr>\n<tr><td class=\"num\">4.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for C3-Propionyl-L-carnitine\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/20064-19-1\/C3-Propionyl-L-carnitine\" target=\"_blank\" rel=\"noopener\">C3-Propionyl-L-carnitine<\/a><\/td><td class=\"oth\">1.29<\/td><td class=\"oth\">218.1<\/td><td class=\"oth\">85.1<\/td><\/tr>\n<tr><td class=\"num\">5.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 3-Hydroxyisovaleryl-L-carnitine\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/99159-87-2\/3-Hydroxyisovaleryl-L-carnitine\" target=\"_blank\" rel=\"noopener\">3-Hydroxyisovaleryl-L-carnitine<\/a><\/td><td class=\"oth\">1.60<\/td><td class=\"oth\">262.4<\/td><td class=\"oth\">85.1<\/td><\/tr>\n<tr><td class=\"num\">6.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for C4-Isobutyrl-L-carnitine\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/25518-49-4\/C4-Isobutyrl-L-carnitine\" target=\"_blank\" rel=\"noopener\">C4-Isobutyrl-L-carnitine<\/a><\/td><td class=\"oth\">2.31<\/td><td class=\"oth\">232.1<\/td><td class=\"oth\">85.1<\/td><\/tr>\n<tr><td class=\"num\">7.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for C4-Butyryl-L-carnitine\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/25576-40-3\/C4-Butyryl-L-carnitine\" target=\"_blank\" rel=\"noopener\">C4-Butyryl-L-carnitine<\/a><\/td><td class=\"oth\">2.42<\/td><td class=\"oth\">232.2<\/td><td class=\"oth\">85.1<\/td><\/tr>\n<tr><td class=\"num\">8.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 3-Methylcrotonyl-L-carnitine\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/64656-41-3\/3-Methylcrotonyl-L-carnitine\" target=\"_blank\" rel=\"noopener\">3-Methylcrotonyl-L-carnitine<\/a><\/td><td class=\"oth\">3.31<\/td><td class=\"oth\">244.1<\/td><td class=\"oth\">85.1<\/td><\/tr>\n<tr><td class=\"num\">9.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for C5:1-Tigyl-L-carnitine\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/64681-36-3\/C5:1-Tigyl-L-carnitine\" target=\"_blank\" rel=\"noopener\">C5:1-Tigyl-L-carnitine<\/a><\/td><td class=\"oth\">3.49<\/td><td class=\"oth\">244.2<\/td><td class=\"oth\">85.1<\/td><\/tr>\n<tr><td class=\"num\">10.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 2-Methylbutyryl-L-carnitine\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/256928-75-3\/2-Methylbutyryl-L-carnitine\" target=\"_blank\" rel=\"noopener\">2-Methylbutyryl-L-carnitine<\/a><\/td><td class=\"oth\">3.61<\/td><td class=\"oth\">246.1<\/td><td class=\"oth\">85.1<\/td><\/tr>\n<tr><td class=\"num\">11.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for C5-Isovaleryl-L-carnitine\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/31023-24-2\/C5-Isovaleryl-L-carnitine\" target=\"_blank\" rel=\"noopener\">C5-Isovaleryl-L-carnitine<\/a><\/td><td class=\"oth\">3.71<\/td><td class=\"oth\">246.1<\/td><td class=\"oth\">85.1<\/td><\/tr>\n<tr><td class=\"num\">12.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for C5-Valeryl-L-carnitine\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/40225-14-7\/C5-Valeryl-L-carnitine\" target=\"_blank\" rel=\"noopener\">C5-Valeryl-L-carnitine<\/a><\/td><td class=\"oth\">3.87<\/td><td class=\"oth\">246.2<\/td><td class=\"oth\">85.1<\/td><\/tr>\n<\/tbody><\/table>\n<table class=\"peaks col-lg-6 col-12 peak-50\"><thead><tr><th><\/th><th style=\"text-align: left;width: 75px\">Peaks<\/th><th style=\"text-align: center;width: 75px\">t<sub>R<\/sub> (min)<\/th><th style=\"text-align: center;width: 75px\">Precursor Ion<\/th><th style=\"text-align: center;width: 75px\">Product Ion<\/th><\/tr><\/thead><tbody>\n<tr><td class=\"num\">13.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for C6-Hexanoyl-L-carnitine\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/22671-29-0\/C6-Hexanoyl-L-carnitine\" target=\"_blank\" rel=\"noopener\">C6-Hexanoyl-L-carnitine<\/a><\/td><td class=\"oth\">5.08<\/td><td class=\"oth\">260.3<\/td><td class=\"oth\">85.1<\/td><\/tr>\n<tr><td class=\"num\">14.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for C7-Heptanoyl-L-carnitine\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/22671-29-0\/C7-Heptanoyl-L-carnitine\" target=\"_blank\" rel=\"noopener\">C7-Heptanoyl-L-carnitine<\/a><\/td><td class=\"oth\">5.27<\/td><td class=\"oth\">274.2<\/td><td class=\"oth\">85.1<\/td><\/tr>\n<tr><td class=\"num\">15.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for C8-Octanoyl-L-carnitine\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/25243-95-2\/C8-Octanoyl-L-carnitine\" target=\"_blank\" rel=\"noopener\">C8-Octanoyl-L-carnitine<\/a><\/td><td class=\"oth\">5.41<\/td><td class=\"oth\">288.4<\/td><td class=\"oth\">85.1<\/td><\/tr>\n<tr><td class=\"num\">16.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for C10-Decanoyl-L-carnitine\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/3992-45-8\/C10-Decanoyl-L-carnitine\" target=\"_blank\" rel=\"noopener\">C10-Decanoyl-L-carnitine<\/a><\/td><td class=\"oth\">5.67<\/td><td class=\"oth\">316.3<\/td><td class=\"oth\">85.1<\/td><\/tr>\n<tr><td class=\"num\">17.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for C14:2-Tetradecadienoyl-L-carnitine\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/9999999-07-6\/C14:2-Tetradecadienoyl-L-carnitine\" target=\"_blank\" rel=\"noopener\">C14:2-Tetradecadienoyl-L-carnitine<\/a><\/td><td class=\"oth\">5.90<\/td><td class=\"oth\">368.5<\/td><td class=\"oth\">85.1<\/td><\/tr>\n<tr><td class=\"num\">18.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for C12-Lauroyl-L-carnitine\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/25518-54-1\/C12-Lauroyl-L-carnitine\" target=\"_blank\" rel=\"noopener\">C12-Lauroyl-L-carnitine<\/a><\/td><td class=\"oth\">5.94<\/td><td class=\"oth\">344.5<\/td><td class=\"oth\">85.1<\/td><\/tr>\n<tr><td class=\"num\">19.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for C14:1-Tetradecanoyl-L-carnitine\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/1469901-01-6\/C14:1-Tetradecanoyl-L-carnitine\" target=\"_blank\" rel=\"noopener\">C14:1-Tetradecanoyl-L-carnitine<\/a><\/td><td class=\"oth\">6.17<\/td><td class=\"oth\">369.8<\/td><td class=\"oth\">85.1<\/td><\/tr>\n<tr><td class=\"num\">20.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for C14-Myristoyl-L-carnitine\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/25597-07-3\/C14-Myristoyl-L-carnitine\" target=\"_blank\" rel=\"noopener\">C14-Myristoyl-L-carnitine<\/a><\/td><td class=\"oth\">6.25<\/td><td class=\"oth\">372.4<\/td><td class=\"oth\">85.1<\/td><\/tr>\n<tr><td class=\"num\">21.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for C18:2 Linoleoyl-L-carnitine\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/173686-75-4\/C18:2 Linoleoyl-L-carnitine\" target=\"_blank\" rel=\"noopener\">C18:2 Linoleoyl-L-carnitine<\/a><\/td><td class=\"oth\">6.42<\/td><td class=\"oth\">424.1<\/td><td class=\"oth\">85.1<\/td><\/tr>\n<tr><td class=\"num\">22.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for C16:1 Palmitolelyl-L-carnitine\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/162067-53-0\/C16:1 Palmitolelyl-L-carnitine\" target=\"_blank\" rel=\"noopener\">C16:1 Palmitolelyl-L-carnitine<\/a><\/td><td class=\"oth\">6.58<\/td><td class=\"oth\">398.4<\/td><td class=\"oth\">85.1<\/td><\/tr>\n<tr><td class=\"num\">23.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for C16-Palmitoyl-L-carnitine\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/2364-67-2\/C16-Palmitoyl-L-carnitine\" target=\"_blank\" rel=\"noopener\">C16-Palmitoyl-L-carnitine<\/a><\/td><td class=\"oth\">6.69<\/td><td class=\"oth\">400.5<\/td><td class=\"oth\">85.1<\/td><\/tr>\n<tr><td class=\"num\">24.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for C18:1 Oleoyl-L-carnitine\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/31062-78-9\/C18:1 Oleoyl-L-carnitine\" target=\"_blank\" rel=\"noopener\">C18:1 Oleoyl-L-carnitine<\/a><\/td><td class=\"oth\">6.77<\/td><td class=\"oth\">426.4<\/td><td class=\"oth\">85.1<\/td><\/tr>\n<tr><td class=\"num\">25.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for C18 - Stearoyl-L-carnitine\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/25597-09-5\/C18 - Stearoyl-L-carnitine\" target=\"_blank\" rel=\"noopener\">C18 &#8211; Stearoyl-L-carnitine<\/a><\/td><td class=\"oth\">7.13<\/td><td class=\"oth\">429.1<\/td><td class=\"oth\">85.1<\/td><\/tr>\n<\/tbody><\/table><\/div><div class=\"chromatogram-conditions\"><h4>Conditions<\/h4><div class=\"conditions-container container-fluid\"><div class=\"row\"><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Column<\/th><td>Raptor ARC-18  (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/9314A12\" rel=\"noopener\">cat.# 9314A12<\/a>)<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Dimensions:<\/th><td>100 mm x 2.1 mm ID<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Particle Size:<\/th><td>2.7 \u00b5m<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Pore Size:<\/th><td>90 \u00c5<\/td><\/tr><tr><td><\/td><tr><th class=\"sub conditions_header\" scope=\"row\">Guard Column:<\/th><td>Raptor ARC-18 EXP Guard Column Cartridge 5 mm, 2.1 mm ID, 2.7 \u00b5m (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/9314A0252\" rel=\"noopener\">cat.# 9314A0252<\/a>)<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Temp.:<\/th><td>35 \u00b0C<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Standard\/Sample<\/th><td><\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Diluent:<\/th><td>Water, 0.1% formic acid<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Conc.:<\/th><td>100 ng\/mL<\/td><\/tr><td><\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Inj. Vol.:<\/th><td>3 \u00b5L <\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Mobile Phase<\/th><td><\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">A:<\/th><td>Water, 0.1% formic acid <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">B:<\/th><td>90:10 Acetonitrile:isopropanol, 0.1% formic acid <\/td><\/tr><tr><td><\/td><td><table class=\"cgram_ramp\"><thead><tr><th>Time (min)<\/th><th>Flow (mL\/min)<\/th><th>%A<\/th><th>%B<\/th><\/tr><\/thead><tbody><tr><td>0.00<\/td><td>0.6<\/td><td>98<\/td><td>2<\/td><\/tr><tr><td>1.00<\/td><td>0.6<\/td><td>98<\/td><td>2<\/td><\/tr><tr><td>4.00<\/td><td>0.6<\/td><td>88<\/td><td>12<\/td><\/tr><tr><td>6.00<\/td><td>0.6<\/td><td>0<\/td><td>100<\/td><\/tr><tr><td>7.50<\/td><td>0.6<\/td><td>0<\/td><td>100<\/td><\/tr><tr><td>8.00<\/td><td>0.6<\/td><td>98<\/td><td>2<\/td><\/tr><tr><td>9.00<\/td><td>0.6<\/td><td>98<\/td><td>2<\/td><\/tr><\/tbody><\/table><\/td><\/tr><\/table><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Detector<\/th><td>SCIEX 4500 MS\/MS <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Ion Source:<\/th><td>Electrospray <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Ion Mode:<\/th><td>ESI+ <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Mode:<\/th><td>MRM <\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Instrument<\/th><td>Shimadzu Nexera UHPLC <\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Sample Preparation<\/th><td>A 100 ng\/mL standard mix of all of the acylcarnitines was prepared in plasma. A 100 \u00b5L aliquot was taken from the standard and mixed with 300 \u00b5L of methanol, vortexed for 10 seconds, and then centrifuged for 10 minutes at 4000 rpm. 100 \u00b5L of the supernatant was added to a 2 mL vial (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/24619\" rel=\"noopener\">cat.# 24619<\/a>) containing 900 \u00b5L of MPA (0.1% formic acid in water), capped with a short screw cap (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/24498\" rel=\"noopener\">cat.# 24498<\/a>), and injected for LC-MS\/MS analysis.<\/td><\/tr><\/table><\/div><\/div><\/div><div class=\"chromatogram-related-skus\"><h4>Related SKUs<\/h4><ul><li>9314A0252<\/li><li>9314A12<\/li><li>24619<\/li><li>24498<\/li><\/ul><\/div><div class=\"chromatogram-related-searches\"><h4>Related Searches<\/h4><ul><li>Acylcarnitines<\/li><li>Isomer<\/li><li>Plasma<\/li><li>Separation<\/li><\/ul><\/div><div class=\"chromatogram-pdf-link\"><a href=\"https:\/\/ez.restek.com\/images\/cgram\/lc_cf0782.pdf\" target=\"_blank\" rel=\"noopener noreferrer\"><svg xmlns=\"http:\/\/www.w3.org\/2000\/svg\" width=\"18\" height=\"18\" viewBox=\"0 0 18 18\"><g data-name=\"Group 2996\"><path data-name=\"Rectangle 1246\" d=\"M0 0h18v18H0z\" style=\"fill: none;\"><\/path><\/g><g data-name=\"Group 2997\"><path data-name=\"Path 729\" d=\"M13.412 11.4v2.017H5.345V11.4H4v2.017a1.349 1.349 0 0 0 1.345 1.345h8.068a1.349 1.349 0 0 0 1.345-1.345V11.4zm-.672-2.694-.948-.948-1.741 1.735V4H8.706v5.493L6.965 7.758l-.948.948 3.361 3.361z\" transform=\"translate(-.437 -.414)\" style=\"fill: rgb(13, 123, 196);\"><\/path><\/g><\/svg>Download PDF<\/a><\/div><\/div><\/div>\n<\/div><\/div><\/div>\n\n\n<div style=\"height:50px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n\n\n\n<h3 class=\"wp-block-heading\"><em>Key Separations<\/em><\/h3>\n\n\n\n<p class=\"wp-block-paragraph\">There are many isomeric separations that are crucial for differential diagnosis. Some of these include the separation of C4-butyryl-L-carnitine from C4-isobutyryl-L-carnitine for the diagnosis of butyryl-CoA dehydrogenase deficiency versus isobutyryl-CoA dehydrogenase deficiency. This separation can be seen in Figure 4 along with the separation of C5-valeryl-L-carnitine, C5-isovaleryl-L-carnitine, and 2-methyl-butyryl-L-carnitine. The separation of these critical isobars is easily achieved on a Raptor ARC-18 column in a 9-minute run, which allows higher sample throughput compared to the 20+ minute runs observed in some literature for this type of isobaric separation.<\/p>\n\n\n<div class=\"wp-block-custom-chromatogram-article-top\"><div class=\"chromatogram-article-placeholder\"><div class=\"figure-heading\"><strong>Figure 4:<\/strong>\u00a0Separation of C4-butyryl-L-carnitine and C4-isobutyryl-L-carntine as well as C5-valeryl-L-carnitine, C5-isovaleryl-L-carnitine, and 2-methyl-butyryl-L-carnitine.<\/div><div class='chromatogram-article-inner-full'><div class=\"chromatogram-article-inner\">\n<div class=\"wp-block-custom-chromatogram-article\"><div class=\"wp-block-custom-chromatogram-article\"><div class=\"chromatogram-image wide-image\"><img decoding=\"async\" src=\"https:\/\/ez.restek.com\/images\/cgram\/lc_cf0783.png\" alt=\"Key Separation of Acylcarnitines for Differential Diagnosis Part 1\" title=\"-\"><\/div><p class=\"article-id\" style=\"text-align:center\"> LC_CF0783<\/p><div class=\"chromatogram-peaks\"><h4>Peaks<\/h4><table class=\"peaks col-lg-6 col-12\">\n<thead><tr><th><\/th><th style=\"text-align: left;width: 75px\">Peaks<\/th><th style=\"text-align: center;width: 75px\">t<sub>R<\/sub> (min)<\/th><th style=\"text-align: center;width: 75px\">Precursor Ion<\/th><th style=\"text-align: center;width: 75px\">Product Ion<\/th><\/tr><\/thead>\n<tbody><tr><td class=\"num\">1.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for C4-Isobutyrl-L-carnitine\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/25518-49-4\/C4-Isobutyrl-L-carnitine\" target=\"_blank\" rel=\"noopener\">C4-Isobutyrl-L-carnitine<\/a><\/td><td class=\"oth\">2.33<\/td><td class=\"oth\">232.1<\/td><td class=\"oth\">85.1<\/td><\/tr>\n<tr><td class=\"num\">2.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for C4-Butyryl-L-carnitine\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/25576-40-3\/C4-Butyryl-L-carnitine\" target=\"_blank\" rel=\"noopener\">C4-Butyryl-L-carnitine<\/a><\/td><td class=\"oth\">2.47<\/td><td class=\"oth\">232.1<\/td><td class=\"oth\">85.1<\/td><\/tr>\n<tr><td class=\"num\">3.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 2-Methylbutyryl-L-carnitine\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/256928-75-3\/2-Methylbutyryl-L-carnitine\" target=\"_blank\" rel=\"noopener\">2-Methylbutyryl-L-carnitine<\/a><\/td><td class=\"oth\">3.64<\/td><td class=\"oth\">246.1<\/td><td class=\"oth\">85.1<\/td><\/tr>\n<tr><td class=\"num\">4.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for C5-Isovaleryl-L-carnitine\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/31023-24-2\/C5-Isovaleryl-L-carnitine\" target=\"_blank\" rel=\"noopener\">C5-Isovaleryl-L-carnitine<\/a><\/td><td class=\"oth\">3.74<\/td><td class=\"oth\">246.1<\/td><td class=\"oth\">85.1<\/td><\/tr>\n<tr><td class=\"num\">5.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for C5-Valeryl-L-carnitine\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/40225-14-7\/C5-Valeryl-L-carnitine\" target=\"_blank\" rel=\"noopener\">C5-Valeryl-L-carnitine<\/a><\/td><td class=\"oth\">3.90<\/td><td class=\"oth\">246.2<\/td><td class=\"oth\">85.1<\/td><\/tr>\n<\/tbody><\/table><\/div><div class=\"chromatogram-conditions\"><h4>Conditions<\/h4><div class=\"conditions-container container-fluid\"><div class=\"row\"><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Column<\/th><td>Raptor ARC-18  (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/9314A12\" rel=\"noopener\">cat.# 9314A12<\/a>)<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Dimensions:<\/th><td>100 mm x 2.1 mm ID<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Particle Size:<\/th><td>2.7 \u00b5m<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Pore Size:<\/th><td>90 \u00c5<\/td><\/tr><tr><td><\/td><tr><th class=\"sub conditions_header\" scope=\"row\">Guard Column:<\/th><td>Raptor ARC-18 EXP Guard Column Cartridge 5 mm, 2.1 mm ID, 2.7 \u00b5m (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/9314A0252\" rel=\"noopener\">cat.# 9314A0252<\/a>)<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Temp.:<\/th><td>35 \u00b0C<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Standard\/Sample<\/th><td><\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Diluent:<\/th><td>Water, 0.1% formic acid<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Conc.:<\/th><td>100 ng\/mL<\/td><\/tr><td><\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Inj. Vol.:<\/th><td>3 \u00b5L <\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Mobile Phase<\/th><td><\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">A:<\/th><td>Water, 0.1% formic acid <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">B:<\/th><td>90:10 Acetonitrile:isopropanol, 0.1% formic acid <\/td><\/tr><tr><td><\/td><td><table class=\"cgram_ramp\"><thead><tr><th>Time (min)<\/th><th>Flow (mL\/min)<\/th><th>%A<\/th><th>%B<\/th><\/tr><\/thead><tbody><tr><td>0.00<\/td><td>0.6<\/td><td>98<\/td><td>2<\/td><\/tr><tr><td>1.00<\/td><td>0.6<\/td><td>98<\/td><td>2<\/td><\/tr><tr><td>4.00<\/td><td>0.6<\/td><td>88<\/td><td>12<\/td><\/tr><tr><td>6.00<\/td><td>0.6<\/td><td>0<\/td><td>100<\/td><\/tr><tr><td>7.50<\/td><td>0.6<\/td><td>0<\/td><td>100<\/td><\/tr><tr><td>8.00<\/td><td>0.6<\/td><td>98<\/td><td>2<\/td><\/tr><tr><td>9.00<\/td><td>0.6<\/td><td>98<\/td><td>2<\/td><\/tr><\/tbody><\/table><\/td><\/tr><\/table><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Detector<\/th><td>SCIEX 4500 MS\/MS <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Ion Source:<\/th><td>Electrospray <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Ion Mode:<\/th><td>ESI+ <\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Instrument<\/th><td>Shimadzu Nexera UHPLC <\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Notes<\/th><td>All analytes spiked at 100 ng\/mL<\/td><\/tr><\/table><\/div><\/div><\/div><div class=\"chromatogram-related-skus\"><h4>Related SKUs<\/h4><ul><li>9314A0252<\/li><li>9314A12<\/li><\/ul><\/div><div class=\"chromatogram-related-searches\"><h4>Related Searches<\/h4><ul><li>Acylcarnitines<\/li><li>Plasma<\/li><li>Separation<\/li><li>isomer<\/li><\/ul><\/div><div class=\"chromatogram-pdf-link\"><a href=\"https:\/\/ez.restek.com\/images\/cgram\/lc_cf0783.pdf\" target=\"_blank\" rel=\"noopener noreferrer\"><svg xmlns=\"http:\/\/www.w3.org\/2000\/svg\" width=\"18\" height=\"18\" viewBox=\"0 0 18 18\"><g data-name=\"Group 2996\"><path data-name=\"Rectangle 1246\" d=\"M0 0h18v18H0z\" style=\"fill: none;\"><\/path><\/g><g data-name=\"Group 2997\"><path data-name=\"Path 729\" d=\"M13.412 11.4v2.017H5.345V11.4H4v2.017a1.349 1.349 0 0 0 1.345 1.345h8.068a1.349 1.349 0 0 0 1.345-1.345V11.4zm-.672-2.694-.948-.948-1.741 1.735V4H8.706v5.493L6.965 7.758l-.948.948 3.361 3.361z\" transform=\"translate(-.437 -.414)\" style=\"fill: rgb(13, 123, 196);\"><\/path><\/g><\/svg>Download PDF<\/a><\/div><\/div><\/div>\n<\/div><\/div><\/div>\n\n\n<div style=\"height:50px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n\n\n\n<h3 class=\"wp-block-heading\"><em>Accuracy and Precision<\/em><\/h3>\n\n\n\n<p class=\"wp-block-paragraph\">Accuracy and precision were evaluated using QC samples that were analyzed over a total of three days, and the results are presented in Table II. Method accuracy was demonstrated by recovery values being within 15% of the nominal concentrations for all QC samples. Precision was shown by the %RSD values being \u226414% for all QC samples.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\"><strong>Table II: <\/strong>Interday accuracy and precision for underivatized C5-valeryl-L-carnitine in quality control samples prepared in BSA.<\/p>\n\n\n\n<figure class=\"wp-block-table\"><table><tbody><tr><td><\/td><td><strong>Avg. Conc. (ng\/mL)<\/strong><\/td><td><strong>Avg. Accuracy (%)<\/strong><\/td><td><strong>% RSD<\/strong><\/td><\/tr><tr><td><strong>QC 25 ng\/mL<\/strong><\/td><td>27.7<\/td><td>110.9<\/td><td>13.5<\/td><\/tr><tr><td><strong>QC 150 ng\/mL<\/strong><\/td><td>146.5<\/td><td>97.8<\/td><td>3.1<\/td><\/tr><tr><td><strong>QC 600 ng\/mL<\/strong><\/td><td>679.0<\/td><td>113.0<\/td><td>3.8<\/td><\/tr><tr><td><strong>QC 1500 ng\/mL<\/strong><\/td><td>1500.0<\/td><td>100.0<\/td><td>2.0<\/td><\/tr><tr><td><strong>QC 3000 ng\/mL<\/strong><\/td><td>2900.0<\/td><td>96.6<\/td><td>0.0<\/td><\/tr><\/tbody><\/table><\/figure>\n\n\n\n<h3 class=\"wp-block-heading\"><em>Column Robustness<\/em><\/h3>\n\n\n\n<p class=\"wp-block-paragraph\">Column robustness was tested by making 250 injections onto the same column of a 100 ppb acylcarnitines standard prepared in BSA. First injection and last injection retention times were consistent for all 25 of the monitored acylcarnitines as demonstrated in Table III below. Stable retention times help ensure accurate selectivity and identification over longer column lifetimes.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\"><strong>Table III: <\/strong>Retention times in this acylcarnitines LC-MS\/MS analysis were highly consistent over 250 injections for all compounds, demonstrating good robustness.<\/p>\n\n\n\n<figure class=\"wp-block-table\"><table><tbody><tr><td><strong>Analyte<\/strong><\/td><td><strong>Inj. 1 (min)<\/strong><\/td><td><strong>Inj. 250 (min)<\/strong><\/td><td><strong>% Difference<\/strong><\/td><\/tr><tr><td>C0-L-carnitine<\/td><td>0.40<\/td><td>0.40<\/td><td>0.0<\/td><\/tr><tr><td>C2-Acetyl-L-carnitine<\/td><td>0.59<\/td><td>0.59<\/td><td>0.0<\/td><\/tr><tr><td>Methyl-malonyl-L-carnitine<\/td><td>1.15<\/td><td>1.15<\/td><td>0.0<\/td><\/tr><tr><td>C3-Propionyl-L-carnitine<\/td><td>1.29<\/td><td>1.28<\/td><td>0.8<\/td><\/tr><tr><td>3-Hydroxyisovaleryl-L-carnitine<\/td><td>1.60<\/td><td>1.58<\/td><td>1.3<\/td><\/tr><tr><td>C4-Isobutyryl-L-carnitine<\/td><td>2.34<\/td><td>2.31<\/td><td>1.3<\/td><\/tr><tr><td>C4-Butyryl-L-carnitine<\/td><td>2.49<\/td><td>2.43<\/td><td>2.4<\/td><\/tr><tr><td>3-Methylcrotonyl-L-carnitine<\/td><td>3.32<\/td><td>3.31<\/td><td>0.3<\/td><\/tr><tr><td>C5:1-Tigyl-L-carnitine<\/td><td>3.32<\/td><td>3.31<\/td><td>0.3<\/td><\/tr><tr><td>2-Methylbutyryl-L-carnitine<\/td><td>3.63<\/td><td>3.63<\/td><td>0.0<\/td><\/tr><tr><td>C5-Isovaleryl-L-carnitine<\/td><td>3.73<\/td><td>3.72<\/td><td>0.3<\/td><\/tr><tr><td>C5-Valeryl-L-carnitine<\/td><td>3.89<\/td><td>3.89<\/td><td>0.0<\/td><\/tr><tr><td>C6-Hexanoyl-L-carnitine<\/td><td>5.07<\/td><td>5.07<\/td><td>0.0<\/td><\/tr><tr><td>C7-Heptanoyl-L-carnitine<\/td><td>5.28<\/td><td>5.28<\/td><td>0.0<\/td><\/tr><tr><td>C8-Octanoyl-L-carnitine<\/td><td>5.41<\/td><td>5.41<\/td><td>0.0<\/td><\/tr><tr><td>C10-Decanoyl-L-carnitine<\/td><td>5.67<\/td><td>5.67<\/td><td>0.0<\/td><\/tr><tr><td>C14:2-Tetradecadienoyl-L-carnitine<\/td><td>5.92<\/td><td>5.92<\/td><td>0.0<\/td><\/tr><tr><td>C12-Lauroyl-L-carnitine<\/td><td>5.94<\/td><td>5.94<\/td><td>0.0<\/td><\/tr><tr><td>C14:1 Tetradecanoyl-L-carnitine<\/td><td>6.18<\/td><td>6.18<\/td><td>0.0<\/td><\/tr><tr><td>C14-Myristoyl-L-carnitine<\/td><td>6.25<\/td><td>6.25<\/td><td>0.0<\/td><\/tr><tr><td>C18:2 Linoleoyl-L-carnitine<\/td><td>6.40<\/td><td>6.41<\/td><td>0.2<\/td><\/tr><tr><td>C16:1 Palmitolelyl-L-carnitine<\/td><td>6.56<\/td><td>6.57<\/td><td>0.2<\/td><\/tr><tr><td>C16-Palmitoyl-L-carnitine<\/td><td>6.69<\/td><td>6.69<\/td><td>0.0<\/td><\/tr><tr><td>C18:1 Oleoyl-L-carnitine<\/td><td>6.77<\/td><td>6.78<\/td><td>0.1<\/td><\/tr><tr><td>C18 &#8211; Stearoyl-L-carnitine<\/td><td>7.14<\/td><td>7.15<\/td><td>0.1<\/td><\/tr><\/tbody><\/table><\/figure>\n\n\n\n<h3 class=\"wp-block-heading\"><em>Linearity<\/em><\/h3>\n\n\n\n<p class=\"wp-block-paragraph\">Using 1\/x<sup>2<\/sup> weighted linear regression, C5-valeryl-L-carnitine showed acceptable linearity with R<sup>2<\/sup> of 0.9952 or greater (Figure 5).<\/p>\n\n\n<div class=\"wp-block-custom-chromatogram-article-top\"><div class=\"chromatogram-article-placeholder\"><div class=\"figure-heading\"><strong>Figure 5:<\/strong>\u00a0C5-Valeryl-L-carnitine showed good linear response over the entire calibration range.<\/div><div class='chromatogram-article-inner-full'><div class=\"chromatogram-article-inner\">\n<style>.kb-image39968_c38765-e1 .kb-image-has-overlay:after{opacity:0.3;}<\/style>\n<div class=\"wp-block-kadence-image kb-image39968_c38765-e1\"><figure class=\"aligncenter size-large\"><img loading=\"lazy\" decoding=\"async\" width=\"1024\" height=\"404\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/figure_article_cfan4029_05-1024x404.jpg\" alt=\"\" class=\"kb-img wp-image-18748\" title=\"-\" srcset=\"https:\/\/discover.restek.com\/wp-content\/uploads\/figure_article_cfan4029_05-1024x404.jpg 1024w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure_article_cfan4029_05-300x118.jpg 300w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure_article_cfan4029_05-768x303.jpg 768w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure_article_cfan4029_05-1536x606.jpg 1536w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure_article_cfan4029_05.jpg 1800w\" sizes=\"auto, (max-width: 1024px) 100vw, 1024px\" \/><\/figure><\/div>\n\n<\/div><\/div><\/div>\n\n\n<div style=\"height:50px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n\n\n\n<h3 class=\"wp-block-heading\"><em>Accuracy of Surrogate Matrix<\/em><\/h3>\n\n\n\n<p class=\"wp-block-paragraph\">Fortified plasma samples were prepared from three different lots of plasma and analyzed using calibrators prepared in BSA to test the accuracy of using BSA as a surrogate matrix for underivatized acylcarnitines analysis. First, the endogenous concentration of C5-valeryl-L-carnitine was quantitated for each lot. Then, standard addition was used to test the accuracy of the surrogate matrix. For standard addition, the samples were fortified with an additional 1000 ng\/mL. All three lots of plasma showed acceptable results that were less than 11% different from the expected values for both intra- and interday repeatability studies as demonstrated in Table IV and Table V below.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\"><strong>Table IV:<\/strong> Intraday repeatability for standard addition spiking of C5-valeryl-L-carnitine in human plasma.<\/p>\n\n\n\n<figure class=\"wp-block-table\"><table><tbody><tr><td><strong>Plasma Lot<\/strong><\/td><td><strong>Expected (ng\/mL)<\/strong><\/td><td><strong>Average (ng\/mL)<\/strong><\/td><td><strong>% Difference<\/strong><\/td><td><strong>% RSD<\/strong><\/td><\/tr><tr><td>Lot 1<\/td><td>1006<\/td><td>983<\/td><td>2.3<\/td><td>2.3<\/td><\/tr><tr><td>Lot 2<\/td><td>1005<\/td><td>1100<\/td><td>9.0<\/td><td>3.4<\/td><\/tr><tr><td>Lot 3<\/td><td>1002<\/td><td>1072<\/td><td>6.8<\/td><td>3.3<\/td><\/tr><\/tbody><\/table><\/figure>\n\n\n\n<p class=\"wp-block-paragraph\"><strong>Table V: <\/strong>Interday repeatability for standard addition spiking of C5-valeryl-L-carnitine in human plasma.<\/p>\n\n\n\n<figure class=\"wp-block-table\"><table><tbody><tr><td><strong>Plasma Lot<\/strong><\/td><td><strong>Expected (ng\/mL)<\/strong><\/td><td><strong>Average (ng\/mL)<\/strong><\/td><td><strong>% Difference<\/strong><\/td><td><strong>% RSD<\/strong><\/td><\/tr><tr><td>Lot 1<\/td><td>1006<\/td><td>989<\/td><td>1.7<\/td><td>2.6<\/td><\/tr><tr><td>Lot 2<\/td><td>1006<\/td><td>1120<\/td><td>10.7<\/td><td>2.7<\/td><\/tr><tr><td>Lot 3<\/td><td>1004<\/td><td>1109<\/td><td>9.9<\/td><td>2.3<\/td><\/tr><\/tbody><\/table><\/figure>\n\n\n\n<h2 class=\"wp-block-heading\">Conclusion<\/h2>\n\n\n\n<p class=\"wp-block-paragraph\">The acylcarnitines LC-MS\/MS analysis developed here provides a quick, efficient approach for the preparation and analysis of underivatized acylcarnitines in plasma samples. Separation of critical pairs was achieved in a fast, 9-minute run, allowing high throughput analysis that can support differential diagnosis. Method performance testing demonstrated acceptable method precision, accuracy, and linearity, and the standard addition experiment showed that BSA is a suitable surrogate matrix.<\/p>\n\n\n\n<h2 class=\"wp-block-heading\">References<\/h2>\n\n\n\n<ol class=\"wp-block-list\">\n<li>Giesbertz, J. Ecker, A. Haag, B. Spanier, H. Daniel, An LC-MS\/MS method to quantify acylcarnitine species including isomeric and odd-numbered forms in plasma and tissues, Journal of Lipid Research 56 (2015) 2029-2039. DOI: <a href=\"https:\/\/doi.org\/10.1194\/jlr.D061721\" target=\"_blank\" rel=\"noreferrer noopener\">https:\/\/doi.org\/10.1194\/jlr.D061721<\/a><\/li>\n\n\n\n<li>E. Minkler, M.S.K. Stoll, S.T. Ingalls, S. Yang, J. Kerner, C.L. Hoppel, Quantification of carnitine and acylcarnitines in biological matrices by HPLC electrospray ionization-mass spectrometry, Clinical Chemistry 54 (9) (2008) 1451-1462. DOI: <a href=\"https:\/\/doi.org\/10.1373\/clinchem.2007.099226\" target=\"_blank\" rel=\"noreferrer noopener\">https:\/\/doi.org\/10.1373\/clinchem.2007.099226<\/a><\/li>\n\n\n\n<li>R. Alagandula, Improved screening method for acylcarnitines and amino acids in dried blood spots by LC-MS\/MS, Application note, CFAN3216A-UNV, Restek Corporation, 2020 <a href=\"https:\/\/discover.restek.com\/CFAN3216\/improved-screening-method-for-acylcarnitines-and-amino-acids-in-dried-blood-spots-by-ls-msms\/\" data-type=\"post\" data-id=\"43354\">https:\/\/www.restek.com\/technical-literature-library\/articles\/improved-screening-method-for-acylcarnitines-and-amino-acids-in-dried-blood-spots-by-LS-MSMS\/<\/a><\/li>\n\n\n\n<li>R. De Jes\u00fas, D.H. Chace, T.H. Lim, J.V. Mei, W.H. Hannon, Comparison of amino acids and acylcarnitines assay methods used in newborn screening assays by tandem mass spectrometry, Clinica Chemica Acta 411 (2010) 684-689. DOI: <a href=\"https:\/\/doi.org\/10.1016\/j.cca.2010.01.034\" target=\"_blank\" rel=\"noreferrer noopener\">https:\/\/doi.org\/10.1016\/j.cca.2010.01.034<\/a><\/li>\n<\/ol>\n\n\n\n<p class=\"wp-block-paragraph\"><em>This method has been developed for research use only; it is not suitable for use in diagnostic procedures without further evaluation.<\/em><\/p>\n\n\n\n<div style=\"height:30px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n\n\n        <div class=\"cpb\">\n            <h3 class=\"cpb-heading\">Products Mentioned<\/h3>\n            <hr class=\"cpb-heading-underline\" \/>\n            <div class=\"cpb-list\">\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/9314A12\">                                Catalog No. 9314A12                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">Raptor ARC-18, 2.7 \u00b5m, 100 x 2.1 mm HPLC Column<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/9314A12\">View Product<\/a>\n                                                    <\/div>\n                    <\/div>\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/9314A0252\">                                Catalog No. 9314A0252                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">Raptor ARC-18 EXP Guard Column Cartridge, 2.7 \u00b5m, 5 x 2.1 mm, 3-pk.<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/9314A0252\">View Product<\/a>\n                                                    <\/div>\n                    <\/div>\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/24619\">                                Catalog No. 24619                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">Screw-Thread Vials with Grad Marking Spot, 2 mL, 8 mm (vial only), Clear, 100-pk.<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/24619\">View Product<\/a>\n                                                    <\/div>\n                    <\/div>\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/24498\">                                Catalog No. 24498                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">Short Screw Cap, Polypropylene, Screw-Thread, PTFE\/Silicone\/PTFE Septa, Blue, Preassembled, 2.0 mL, 9 mm, 1000-pk.<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/24498\">View Product<\/a>\n                                                    <\/div>\n                    <\/div>\n                            <\/div>\n        <\/div>\n        \n\n\n<div style=\"height:20px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n\n\n        <div class=\"cpb\">\n            <h3 class=\"cpb-heading\">Related Products<\/h3>\n            <hr class=\"cpb-heading-underline\" \/>\n            <div class=\"cpb-list\">\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/21140\">                                Catalog No. 21140                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">Short-Cap Vial with Grad Marking Spot, 9-425 Screw-Thread, 2 mL, 9 mm, 12 x 32 (vial only), Clear, 100-pk.<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/21140\">View Product<\/a>\n                                                    <\/div>\n                    <\/div>\n                            <\/div>\n        <\/div>\n        \n\n\n<div style=\"height:100px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n","protected":false},"excerpt":{"rendered":"<p>In this acylcarnitines LC-MS\/MS analysis, plasma samples were prepared without derivatization and 25 acylcarnitines were separated on a Raptor ARC-18 column, allowing positive identification of key analytes needed for differential diagnosis.<\/p>\n","protected":false},"author":24,"featured_media":70197,"comment_status":"closed","ping_status":"open","sticky":false,"template":"","format":"standard","meta":{"_acf_changed":false,"_kad_blocks_custom_css":"","_kad_blocks_head_custom_js":"","_kad_blocks_body_custom_js":"","_kad_blocks_footer_custom_js":"","_kadence_starter_templates_imported_post":false,"_kad_post_transparent":"","_kad_post_title":"","_kad_post_layout":"","_kad_post_sidebar_id":"","_kad_post_content_style":"","_kad_post_vertical_padding":"","_kad_post_feature":"","_kad_post_feature_position":"","_kad_post_header":false,"_kad_post_footer":false,"footnotes":"","_ppma_block_editor_authors":""},"categories":[13],"tags":[],"industries-application":[2153,2154],"post-badge":[],"resource-type":[],"product-library":[2391,2373],"resource-technique":[2299,2301,2302],"ppma_author":[444],"class_list":["post-39968","post","type-post","status-publish","format-standard","has-post-thumbnail","hentry","category-application-notes","industries-application-clinical","industries-application-clinical-diagnostics","product-library-lc-columns","product-library-liquid-chromatography-products","resource-technique-liquid-chromatography","resource-technique-mass-spectrometry-ms","resource-technique-ms-ms"],"acf":[],"taxonomy_info":{"category":[{"value":13,"label":"Application Notes"}],"industries-application":[{"value":2153,"label":"Clinical"},{"value":2154,"label":"Clinical Diagnostics"}],"product-library":[{"value":2391,"label":"LC Columns"},{"value":2373,"label":"Liquid Chromatography Products"}],"resource-technique":[{"value":2299,"label":"Liquid Chromatography"},{"value":2301,"label":"Mass Spectrometry (MS)"},{"value":2302,"label":"MS\/MS"}]},"featured_image_src_large":["https:\/\/discover.restek.com\/wp-content\/uploads\/feature-CFAN4029-1024x536.jpg",1024,536,true],"author_info":{"display_name":"Samantha Herbick","author_link":"https:\/\/discover.restek.com\/es\/author\/samantha-herbick\/"},"comment_info":0,"category_info":[{"term_id":13,"name":"Application Notes","slug":"application-notes","term_group":0,"term_taxonomy_id":13,"taxonomy":"category","description":"","parent":0,"count":88,"filter":"raw","cat_ID":13,"category_count":88,"category_description":"","cat_name":"Application Notes","category_nicename":"application-notes","category_parent":0}],"tag_info":false,"authors":[{"term_id":444,"user_id":24,"is_guest":0,"slug":"samantha-herbick","display_name":"Samantha Herbick","avatar_url":{"url":"https:\/\/discover.restek.com\/wp-content\/uploads\/people-herbick-samantha.jpg","url2x":"https:\/\/discover.restek.com\/wp-content\/uploads\/people-herbick-samantha.jpg"},"author_category":"1","first_name":"Samantha","last_name":"Herbick","user_url":"https:\/\/www.linkedin.com\/in\/samantha-herbick-656378136\/","job_title":"Advanced Scientist, LC Solutions","description":"Samantha Herbick is an applications scientist within the LC Solutions Department at Restek. Her primary focus is on the development of applications in the toxicology and life science markets. She attended Duquesne University where she earned a bachelor\u2019s degree in Biochemistry and a master's degree in forensic science and law. Prior to joining Restek, Samantha worked as a scientist and method developer in a forensic toxicology lab. In this role, she performed analysis on toxicology casework and worked on the development and validation of new and existing assays using LC-MS\/MS, GC-MS, and GC-FID."}],"_links":{"self":[{"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/posts\/39968","targetHints":{"allow":["GET"]}}],"collection":[{"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/posts"}],"about":[{"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/types\/post"}],"author":[{"embeddable":true,"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/users\/24"}],"replies":[{"embeddable":true,"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/comments?post=39968"}],"version-history":[{"count":10,"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/posts\/39968\/revisions"}],"predecessor-version":[{"id":74008,"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/posts\/39968\/revisions\/74008"}],"wp:featuredmedia":[{"embeddable":true,"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/media\/70197"}],"wp:attachment":[{"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/media?parent=39968"}],"wp:term":[{"taxonomy":"category","embeddable":true,"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/categories?post=39968"},{"taxonomy":"post_tag","embeddable":true,"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/tags?post=39968"},{"taxonomy":"industries-application","embeddable":true,"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/industries-application?post=39968"},{"taxonomy":"post-badge","embeddable":true,"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/post-badge?post=39968"},{"taxonomy":"resource-type","embeddable":true,"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/resource-type?post=39968"},{"taxonomy":"product-library","embeddable":true,"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/product-library?post=39968"},{"taxonomy":"resource-technique","embeddable":true,"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/resource-technique?post=39968"},{"taxonomy":"author","embeddable":true,"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/ppma_author?post=39968"}],"curies":[{"name":"wp","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}