{"id":43366,"date":"2020-10-26T00:00:00","date_gmt":"2020-10-26T00:00:00","guid":{"rendered":"https:\/\/discover.restek.com\/uncategorized\/rapid-analysis-of-steroid-hormones-by-gcms\/"},"modified":"2026-08-11T15:25:56","modified_gmt":"2026-08-11T15:25:56","slug":"rapid-analysis-of-steroid-hormones-by-gcms","status":"publish","type":"post","link":"https:\/\/discover.restek.com\/es\/articles\/cfar3228\/rapid-analysis-of-steroid-hormones-by-gcms","title":{"rendered":"Rapid Analysis of Steroid Hormones by GC\/MS"},"content":{"rendered":"\n<ul class=\"wp-block-list\">\n<li>Resolve 6 common steroid hormones in less than 25 minutes.<\/li>\n\n\n\n<li>Ultra-low bleed column greatly reduces background interferences.<\/li>\n\n\n\n<li>Stable performance at 300\u00b0C or above.<\/li>\n<\/ul>\n\n\n\n<p class=\"wp-block-paragraph\">Determinations of urinary steroid hormones are widely used for diagnosing and monitoring many health conditions, including bio-identical hormone replacement, menopause, Cushing\u2019s syndrome, Addison&#8217;s disease, adrenal fatigue, and others [1]. Many clinical laboratories use gas chromatography and mass spectrometry (GC\/MS) as the primary analytical method for identification and quantification. A capillary GC column with a thin film (0.25\u00b5m or less) of 100% dimethylpolysiloxane is the column of choice for many analysts, because this stationary phase has the highest operating temperature available. Temperatures exceeding 300\u00b0C are required to elute the high molecular weight (250-400 Dalton) hormones in a reasonable analysis time while maintaining Gaussian peak shape and resolution. A phase film thickness of 0.25\u00b5m or less minimizes column bleed at these high temperatures. Also, in order to provide reliable quantification, the column must exhibit the inertness necessary to produce symmetric peaks and reproducible results.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Our new Rxi-1ms column, designed for GC-MS applications, provides the ultra-low bleed and exceptional inertness needed for analyzing urinary steroid hormones. For this application we derivatized six sex hormones, using methoxylamine HCl and trimethylsilyl imidazole to improve chromatography. Figure 1 shows this variety of derivatized steroid sex hormones, analyzed in less than 25 minutes by using an Rxi-1ms column. Note that these compounds elute at temperatures near or above 300\u00b0C and that bleed from the Rxi-1ms column is negligible at these temperatures. The Rxi-1ms column exhibits the inertness needed to produce Gaussian peaks and excellent resolution.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Because GC\/MS analysis of urinary steroid hormones is a demanding application, it is important to use the lowest bleed, most inert column available. The new Rxi-1ms column meets these requirements better than any column we have tested, and we recommend it as the column of choice for this application.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\"><strong>Figure 1:<\/strong>&nbsp;Negligible bleed, Gaussian peaks, and fast results characterize analyses of derivatized steroids on an Rxi-1ms column.<\/p>\n\n\n<div class=\"wp-block-custom-chromatogram-article\"><div class=\"wp-block-custom-chromatogram-article\"><div class=\"chromatogram-image regular-image\"><img decoding=\"async\" src=\"https:\/\/ez.restek.com\/images\/cgram\/gc_ph00872.png\" alt=\"Steroid Sex Hormones on Rxi-1ms\" title=\"-\"><\/div><p class=\"article-id\" style=\"text-align:center\"> GC_PH00872<\/p><div class=\"chromatogram-peaks\"><h4>Peaks<\/h4><table class=\"peaks col-lg-4 col-12\">\n<thead><tr><th><\/th><th style=\"text-align: left;width: 75px\">Peaks<\/th><\/tr><\/thead>\n<tbody><tr><td class=\"num\">1.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Androsterone\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/53-41-8\/Androsterone\" target=\"_blank\" rel=\"noopener\">Androsterone<\/a><\/td><\/tr>\n<tr><td class=\"num\">2.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Dehydroepiandrosterone (DHEA)\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/53-43-0\/Dehydroepiandrosterone\" target=\"_blank\" rel=\"noopener\">Dehydroepiandrosterone (DHEA)<\/a><\/td><\/tr>\n<tr><td class=\"num\">3.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 17\u03b1-Estradiol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/57-91-0\/17\u03b1-Estradiol\" target=\"_blank\" rel=\"noopener\">17\u03b1-Estradiol<\/a><\/td><\/tr>\n<tr><td class=\"num\">4.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Estrone\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/53-16-7\/Estrone\" target=\"_blank\" rel=\"noopener\">Estrone<\/a><\/td><\/tr>\n<tr><td class=\"num\">5.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 17\u03b2-Estradiol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/50-28-2\/17\u03b2-Estradiol\" target=\"_blank\" rel=\"noopener\">17\u03b2-Estradiol<\/a><\/td><\/tr>\n<tr><td class=\"num\">6.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Testosterone\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/58-22-0\/Testosterone\" target=\"_blank\" rel=\"noopener\">Testosterone<\/a><\/td><\/tr>\n<tr><td class=\"num\">7.<\/td><td class=\"cmpd\">derivatization by-product<\/td><\/tr>\n<\/tbody><\/table><\/div><div class=\"chromatogram-conditions\"><h4>Conditions<\/h4><div class=\"conditions-container container-fluid\"><div class=\"row\"><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Column<\/th><td>Rxi-1ms, 30 m, 0.25 mm ID, 0.25 \u00b5m (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/13323\" rel=\"noopener\">cat.# 13323<\/a>)<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Standard\/Sample<\/th><td>Each hormone in methanol or ethanol; compounds derivatized using 2% methoxylamine HCl (CH<sub>3<\/sub>ONH<sub>2<\/sub>) in pyridine, then N-trimethylsilylimidazole (TMSI), then analyzed <\/td><\/tr>\n<tr><th class=\"sub conditions_header\" scope=\"row\">Conc.:<\/th><td>100 \u00b5g\/mL<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Injection<\/th><td><\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Inj. Vol.:<\/th><td>1.0 \u00b5L splitless (hold 0.5  min)<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Liner:<\/th><td>3.5mm single gooseneck inlet liner<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Inj. Temp.:<\/th><td>250 \u00b0C<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\" colspan=\"2\">Oven<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Oven Temp.:<\/th><td>100 &#176;C to 320 &#176;C at 10 &#176;C\/min (hold 10 min)<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Carrier Gas<\/th><td>He, constant flow<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Flow Rate:<\/th><td>1 mL\/min<\/td><\/tr><\/table><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Detector<\/th><td>MS-Scan: Shimadzu 17A with QP5000<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Transfer Line Temp.:<\/th><td>280 \u00b0C<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Ionization Mode:<\/th><td>EI <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Scan Range:<\/th><td>40-700 amu<\/td><\/tr><\/table><\/div><\/div><\/div><div class=\"chromatogram-related-skus\"><h4>Related SKUs<\/h4><ul><li>13323<\/li><\/ul><\/div><div class=\"chromatogram-related-searches\"><h4>Related Searches<\/h4><ul><li>TMSI<\/li><li>estradiol<\/li><li>hormones<\/li><li>steroids<\/li><\/ul><\/div><\/div><\/div>\n\n\n<h2 class=\"wp-block-heading\">References<\/h2>\n\n\n\n<ol class=\"wp-block-list\">\n<li><a href=\"http:\/\/www.meridianvalleylab.com\/\" target=\"_blank\" rel=\"noreferrer noopener\">www.meridianvalley lab.com.\/steroid_dept.html<\/a><\/li>\n<\/ol>\n","protected":false},"excerpt":{"rendered":"<p>GC\/MS analysis of urinary steroid hormones is a demanding application, and the Rxi-1ms column meets the requirements for low bleed and inertness better than any column we have tested. We analyzed a variety of derivatized steroid sex hormones in less than 25 minutes, with excellent resolution and symmetric peaks. At 300\u00b0C or above, bleed from the Rxi-1ms column was negligible.<\/p>\n","protected":false},"author":35,"featured_media":0,"comment_status":"closed","ping_status":"open","sticky":false,"template":"","format":"standard","meta":{"_acf_changed":false,"_kad_blocks_custom_css":"","_kad_blocks_head_custom_js":"","_kad_blocks_body_custom_js":"","_kad_blocks_footer_custom_js":"","_kadence_starter_templates_imported_post":false,"_kad_post_transparent":"","_kad_post_title":"","_kad_post_layout":"","_kad_post_sidebar_id":"","_kad_post_content_style":"","_kad_post_vertical_padding":"","_kad_post_feature":"","_kad_post_feature_position":"","_kad_post_header":false,"_kad_post_footer":false,"footnotes":"","_ppma_block_editor_authors":""},"categories":[7],"tags":[],"industries-application":[],"post-badge":[],"resource-type":[],"product-library":[],"resource-technique":[],"ppma_author":[455],"class_list":["post-43366","post","type-post","status-publish","format-standard","hentry","category-articles"],"acf":[],"taxonomy_info":{"category":[{"value":7,"label":"Articles"}]},"featured_image_src_large":false,"author_info":{"display_name":"Kristi Sellers","author_link":"https:\/\/discover.restek.com\/es\/author\/kristi-sellers\/"},"comment_info":0,"category_info":[{"term_id":7,"name":"Articles","slug":"articles","term_group":0,"term_taxonomy_id":7,"taxonomy":"category","description":"","parent":0,"count":510,"filter":"raw","cat_ID":7,"category_count":510,"category_description":"","cat_name":"Articles","category_nicename":"articles","category_parent":0}],"tag_info":false,"authors":[{"term_id":455,"user_id":35,"is_guest":0,"slug":"kristi-sellers","display_name":"Kristi Sellers","avatar_url":{"url":"https:\/\/discover.restek.com\/wp-content\/uploads\/people-sellers-kristi.png","url2x":"https:\/\/discover.restek.com\/wp-content\/uploads\/people-sellers-kristi.png"},"author_category":"1","first_name":"Kristi","last_name":"Sellers","user_url":"","job_title":"","description":"With over 30 years of experience in chromatography, Kristi has worked for Restek as an applications chemist with experience in both GC and LC method development. She has also worked as a QC chemist and a process chemist in addition to her managerial positions in GC Column manufacturing, QC, the Applied Science Group, and R&amp;D. Kristi is currently a senior product marketing manager for GC columns. She holds a BS in chemistry from Lock Haven University, an MBA from Regis University, and has been a Restek seminar instructor for more than three decades."}],"_links":{"self":[{"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/posts\/43366","targetHints":{"allow":["GET"]}}],"collection":[{"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/posts"}],"about":[{"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/types\/post"}],"author":[{"embeddable":true,"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/users\/35"}],"replies":[{"embeddable":true,"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/comments?post=43366"}],"version-history":[{"count":2,"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/posts\/43366\/revisions"}],"predecessor-version":[{"id":63293,"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/posts\/43366\/revisions\/63293"}],"wp:attachment":[{"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/media?parent=43366"}],"wp:term":[{"taxonomy":"category","embeddable":true,"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/categories?post=43366"},{"taxonomy":"post_tag","embeddable":true,"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/tags?post=43366"},{"taxonomy":"industries-application","embeddable":true,"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/industries-application?post=43366"},{"taxonomy":"post-badge","embeddable":true,"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/post-badge?post=43366"},{"taxonomy":"resource-type","embeddable":true,"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/resource-type?post=43366"},{"taxonomy":"product-library","embeddable":true,"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/product-library?post=43366"},{"taxonomy":"resource-technique","embeddable":true,"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/resource-technique?post=43366"},{"taxonomy":"author","embeddable":true,"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/ppma_author?post=43366"}],"curies":[{"name":"wp","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}