{"id":43417,"date":"2020-10-21T00:00:00","date_gmt":"2020-10-21T00:00:00","guid":{"rendered":"https:\/\/discover.restek.com\/uncategorized\/pfas-lc-column-anatomy-which-phase-dimensions-and-particle-type-are-best\/"},"modified":"2026-08-07T21:15:49","modified_gmt":"2026-08-07T21:15:49","slug":"pfas-lc-column-anatomy-which-phase-dimensions-and-particle-type-are-best","status":"publish","type":"post","link":"https:\/\/discover.restek.com\/es\/artikel\/evar3069\/pfas-lc-column-anatomy-which-phase-dimensions-and-particle-type-are-best","title":{"rendered":"PFAS LC Column Anatomy: Which Phase, Dimensions, and Particle Type Are Best?"},"content":{"rendered":"\n<p class=\"wp-block-paragraph\">If you work in one of the many labs routinely testing samples for per- and polyfluoroalkyl substances (PFAS), you know that awareness and interest are growing as we continue to better understand the pervasiveness, persistence, and potential health risks associated with these \u201cforever chemicals.\u201d As interest grows, the need for fast, accurate, and precise testing is expanding with it. This demand is driving the development of better methods, and LC column selection is the foundation for building an improved approach. This becomes particularly important as the list of monitored PFAS grows to include shorter and shorter alkyl chain compounds. Here, we\u2019ll examine the properties that are important to consider when choosing an LC column for PFAS analysis.<\/p>\n\n\n\n<h2 class=\"wp-block-heading\">Column Phase &amp; Dimension Selection<\/h2>\n\n\n\n<p class=\"wp-block-paragraph\">The first decision to make when determining which PFAS LC column to use is identifying an effective stationary phase. Knowing what range of PFAS you are interested in studying is crucial to this choice.<\/p>\n\n\n\n<figure class=\"wp-block-embed is-type-video is-provider-youtube wp-block-embed-youtube wp-embed-aspect-16-9 wp-has-aspect-ratio\"><div class=\"wp-block-embed__wrapper\">\n<iframe loading=\"lazy\" title=\"Choosing Your LC Column Dimension\" width=\"720\" height=\"405\" src=\"https:\/\/www.youtube.com\/embed\/qL6HiMCiJmI?feature=oembed\" frameborder=\"0\" allow=\"accelerometer; autoplay; clipboard-write; encrypted-media; gyroscope; picture-in-picture; web-share\" referrerpolicy=\"strict-origin-when-cross-origin\" allowfullscreen><\/iframe>\n<\/div><\/figure>\n\n\n\n<h3 class=\"wp-block-heading\">Short-Chain PFAS and Above<\/h3>\n\n\n\n<p class=\"wp-block-paragraph\">When it comes to short-chain PFAS (C4-C6) and above, our scouting of different phase chemistries showed the C18 phase was the best choice. As the alkyl chain on PFAS molecules gets longer, the interactions between those chains and the C18 ligand increase, providing a great mechanism for retention and resolution. Retention is strong enough that a relatively short and narrow column can be used to quickly and effectively resolve target analytes. The example in Figure 1 shows that a 50 x 2.1 mm Raptor C18 column easily elutes and separates the compounds of interest while meeting all of the EPA 537.1 method criteria for drinking water testing in less than 8 minutes (10 minute total analysis time).<\/p>\n\n\n<div class=\"wp-block-custom-chromatogram-article-top\"><div class=\"chromatogram-article-placeholder\"><div class=\"figure-heading\"><strong>Figure 1:<\/strong>\u00a0A 50 x 2.1 mm Raptor C18 column is a great PFAS LC column choice; it meets all EPA 537.1 method criteria in a fast 10-minute total cycle time.<\/div><div class='chromatogram-article-inner-full'><div class=\"chromatogram-article-inner\">\n<div class=\"wp-block-custom-chromatogram-article\"><div class=\"wp-block-custom-chromatogram-article\"><div class=\"chromatogram-image regular-image\"><img decoding=\"async\" src=\"https:\/\/ez.restek.com\/images\/cgram\/lc_ev0560.png\" alt=\"EPA Method 537.1 on Raptor C18\" title=\"-\"><\/div><p class=\"article-id\" style=\"text-align:center\"> LC_EV0560<\/p><div class=\"chromatogram-peaks\"><h4>Peaks<\/h4><table class=\"peaks col-lg-6 col-12 peak-50\">\n<thead><tr><th><\/th><th style=\"text-align: left;width: 75px\">Peaks<\/th><th style=\"text-align: center;width: 75px\">t<sub>R<\/sub> (min)<\/th><th style=\"text-align: center;width: 75px\">Conc.<br \/>(ng\/mL)<\/th><th style=\"text-align: center;width: 75px\">Precursor Ion<\/th><th style=\"text-align: center;width: 75px\">Product Ion<\/th><\/tr><\/thead>\n<tbody><tr><td class=\"num\">1.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Perfluorobutanesulfonic acid (PFBS)\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/375-73-5\/Perfluorobutanesulfonic acid\" target=\"_blank\" rel=\"noopener\">Perfluorobutanesulfonic acid (PFBS)<\/a><\/td><td class=\"oth\">2.06<\/td><td class=\"oth\">10<\/td><td class=\"oth\">298.8<\/td><td class=\"oth\">79.9<\/td><\/tr>\n<tr><td class=\"num\">2.<\/td><td class=\"cmpd\">Perfluoro-n-[1,2-<sup>13<\/sup>C<sub>2<\/sub>]hexanoic acid (<sup>13<\/sup>C<sub>2<\/sub>-PFHxA)<\/td><td class=\"oth\">3.03<\/td><td class=\"oth\">5<\/td><td class=\"oth\">314.9<\/td><td class=\"oth\">270.0<\/td><\/tr>\n<tr><td class=\"num\">3.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Perfluorohexanoic acid (PFHxA)\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/307-24-4\/Perfluorohexanoic acid\" target=\"_blank\" rel=\"noopener\">Perfluorohexanoic acid (PFHxA)<\/a><\/td><td class=\"oth\">3.04<\/td><td class=\"oth\">5<\/td><td class=\"oth\">312.9<\/td><td class=\"oth\">268.7<\/td><\/tr>\n<tr><td class=\"num\">4.<\/td><td class=\"cmpd\">Tetrafluoro-2-heptafluoropropoxy-<sup>13<\/sup>C<sub>3<\/sub>-propanoic acid (<sup>13<\/sup>C<sub>3<\/sub>-HFPO-DA)<\/td><td class=\"oth\">3.36<\/td><td class=\"oth\">5<\/td><td class=\"oth\">286.8<\/td><td class=\"oth\">168.7<\/td><\/tr>\n<tr><td class=\"num\">5.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Hexafluoropropylene oxide dimer acid (HFPO-DA)\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/13252-13-6\/Hexafluoropropylene oxide dimer acid\" target=\"_blank\" rel=\"noopener\">Hexafluoropropylene oxide dimer acid (HFPO-DA)<\/a><\/td><td class=\"oth\">3.38<\/td><td class=\"oth\">5<\/td><td class=\"oth\">285.0<\/td><td class=\"oth\">168.9<\/td><\/tr>\n<tr><td class=\"num\">6.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Perfluoroheptanoic acid (PFHpA)\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/375-85-9\/Perfluoroheptanoic acid\" target=\"_blank\" rel=\"noopener\">Perfluoroheptanoic acid (PFHpA)<\/a><\/td><td class=\"oth\">4.09<\/td><td class=\"oth\">5<\/td><td class=\"oth\">362.8<\/td><td class=\"oth\">318.8<\/td><\/tr>\n<tr><td class=\"num\">7.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Perfluorohexanesulfonic acid (PFHxS)\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/355-46-4\/Perfluorohexanesulfonic acid\" target=\"_blank\" rel=\"noopener\">Perfluorohexanesulfonic acid (PFHxS)<\/a><\/td><td class=\"oth\">4.22<\/td><td class=\"oth\">10<\/td><td class=\"oth\">398.8<\/td><td class=\"oth\">79.9<\/td><\/tr>\n<tr><td class=\"num\">8.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 4,8-Dioxa-3H-perfluorononanoic acid (ADONA)\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/919005-14-4\/4,8-Dioxa-3H-perfluorononanoic acid\" target=\"_blank\" rel=\"noopener\">4,8-Dioxa-3H-perfluorononanoic acid (ADONA)<\/a><\/td><td class=\"oth\">4.24<\/td><td class=\"oth\">5<\/td><td class=\"oth\">376.9<\/td><td class=\"oth\">250.7<\/td><\/tr>\n<tr><td class=\"num\">9.<\/td><td class=\"cmpd\">Perfluoro-[1,2-<sup>13<\/sup>C<sub>2<\/sub>]octanoic acid (<sup>13<\/sup>C<sub>2<\/sub>-PFOA)<\/td><td class=\"oth\">4.88<\/td><td class=\"oth\">5<\/td><td class=\"oth\">415.0<\/td><td class=\"oth\">370.0<\/td><\/tr>\n<tr><td class=\"num\">10.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Perfluorooctanoic acid (PFOA)\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/335-67-1\/Perfluorooctanoic acid\" target=\"_blank\" rel=\"noopener\">Perfluorooctanoic acid (PFOA)<\/a><\/td><td class=\"oth\">4.90<\/td><td class=\"oth\">5<\/td><td class=\"oth\">413.1<\/td><td class=\"oth\">368.9<\/td><\/tr>\n<tr><td class=\"num\">11.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Perfluorononanoic acid (PFNA)\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/375-95-1\/Perfluorononanoic acid\" target=\"_blank\" rel=\"noopener\">Perfluorononanoic acid (PFNA)<\/a><\/td><td class=\"oth\">5.54<\/td><td class=\"oth\">5<\/td><td class=\"oth\">462.9<\/td><td class=\"oth\">418.9<\/td><\/tr>\n<tr><td class=\"num\">12.<\/td><td class=\"cmpd\">Perfluoro-1-[1,2,3,4-<sup>13<\/sup>C<sub>4<\/sub>]octanesulfonic acid (<sup>13<\/sup>C<sub>4<\/sub>-PFOS)<\/td><td class=\"oth\">5.57<\/td><td class=\"oth\">10<\/td><td class=\"oth\">503.0<\/td><td class=\"oth\">80.0<\/td><\/tr>\n<\/tbody><\/table>\n<table class=\"peaks col-lg-6 col-12 peak-50\"><thead><tr><th><\/th><th style=\"text-align: left;width: 75px\">Peaks<\/th><th style=\"text-align: center;width: 75px\">t<sub>R<\/sub> (min)<\/th><th style=\"text-align: center;width: 75px\">Conc.<br \/>(ng\/mL)<\/th><th style=\"text-align: center;width: 75px\">Precursor Ion<\/th><th style=\"text-align: center;width: 75px\">Product Ion<\/th><\/tr><\/thead><tbody>\n<tr><td class=\"num\">13.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Perfluorooctanesulfonic acid (PFOS)\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/1763-23-1\/Perfluorooctanesulfonic acid\" target=\"_blank\" rel=\"noopener\">Perfluorooctanesulfonic acid (PFOS)<\/a><\/td><td class=\"oth\">5.58<\/td><td class=\"oth\">10<\/td><td class=\"oth\">498.9<\/td><td class=\"oth\">80.0<\/td><\/tr>\n<tr><td class=\"num\">14.<\/td><td class=\"cmpd\">9-Chlorohexadecafluoro-3-oxanone-1-sulfonic acid (9Cl-PF3ONS)<\/td><td class=\"oth\">5.88<\/td><td class=\"oth\">5<\/td><td class=\"oth\">530.8<\/td><td class=\"oth\">350.7<\/td><\/tr>\n<tr><td class=\"num\">15.<\/td><td class=\"cmpd\">Perfluoro-n-[1,2-<sup>13<\/sup>C<sub>2<\/sub>]decanoic acid (<sup>13<\/sup>C<sub>2<\/sub>-PFDA)<\/td><td class=\"oth\">6.08<\/td><td class=\"oth\">5<\/td><td class=\"oth\">514.9<\/td><td class=\"oth\">469.9<\/td><\/tr>\n<tr><td class=\"num\">16.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Perfluorodecanoic acid (PFDA)\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/335-76-2\/Perfluorodecanoic acid\" target=\"_blank\" rel=\"noopener\">Perfluorodecanoic acid (PFDA)<\/a><\/td><td class=\"oth\">6.08<\/td><td class=\"oth\">5<\/td><td class=\"oth\">512.9<\/td><td class=\"oth\">469.0<\/td><\/tr>\n<tr><td class=\"num\">17.<\/td><td class=\"cmpd\">N-deuteriomethylperfluoro-1-octanesulfonamidoacetic acid (d3-NMeFOSAA)<\/td><td class=\"oth\">6.28<\/td><td class=\"oth\">10<\/td><td class=\"oth\">572.9<\/td><td class=\"oth\">418.8<\/td><\/tr>\n<tr><td class=\"num\">18.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for N-methyl perfluorooctanesulfonamidoacetic acid (NMeFOSAA)\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/2355-31-9\/N-methyl perfluorooctanesulfonamidoacetic acid\" target=\"_blank\" rel=\"noopener\">N-methyl perfluorooctanesulfonamidoacetic acid (NMeFOSAA)<\/a><\/td><td class=\"oth\">6.30<\/td><td class=\"oth\">10<\/td><td class=\"oth\">569.8<\/td><td class=\"oth\">418.8<\/td><\/tr>\n<tr><td class=\"num\">19.<\/td><td class=\"cmpd\">N-deuterioethylperfluoro-1-octanesulfonamidoacetic acid (d5-NEtFOSAA)<\/td><td class=\"oth\">6.51<\/td><td class=\"oth\">10<\/td><td class=\"oth\">588.9<\/td><td class=\"oth\">418.8<\/td><\/tr>\n<tr><td class=\"num\">20.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for N-ethyl perfluorooctanesulfonamidoacetic acid (NEtFOSAA)\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/2991-50-6\/N-ethyl perfluorooctanesulfonamidoacetic acid\" target=\"_blank\" rel=\"noopener\">N-ethyl perfluorooctanesulfonamidoacetic acid (NEtFOSAA)<\/a><\/td><td class=\"oth\">6.52<\/td><td class=\"oth\">10<\/td><td class=\"oth\">583.8<\/td><td class=\"oth\">418.8<\/td><\/tr>\n<tr><td class=\"num\">21.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Perfluoroundecanoic acid (PFUnA)\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/2058-94-8\/Perfluoroundecanoic acid\" target=\"_blank\" rel=\"noopener\">Perfluoroundecanoic acid (PFUnA)<\/a><\/td><td class=\"oth\">6.55<\/td><td class=\"oth\">5<\/td><td class=\"oth\">562.9<\/td><td class=\"oth\">518.8<\/td><\/tr>\n<tr><td class=\"num\">22.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 11-chloroeicosafluoro-3-oxaundecane-1-sulfonic acid (11Cl-PF3OUdS)\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/763051-92-9\/11-chloroeicosafluoro-3-oxaundecane-1-sulfonic acid\" target=\"_blank\" rel=\"noopener\">11-chloroeicosafluoro-3-oxaundecane-1-sulfonic acid (11Cl-PF3OUdS)<\/a><\/td><td class=\"oth\">6.77<\/td><td class=\"oth\">5<\/td><td class=\"oth\">630.7<\/td><td class=\"oth\">451.0<\/td><\/tr>\n<tr><td class=\"num\">23.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Perfluorododecanoic acid (PFDoA)\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/307-55-1\/Perfluorododecanoic acid\" target=\"_blank\" rel=\"noopener\">Perfluorododecanoic acid (PFDoA)<\/a><\/td><td class=\"oth\">6.95<\/td><td class=\"oth\">5<\/td><td class=\"oth\">612.7<\/td><td class=\"oth\">568.9<\/td><\/tr>\n<tr><td class=\"num\">24.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Perfluorotridecanoic acid (PFTrDA)\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/72629-94-8\/Perfluorotridecanoic acid\" target=\"_blank\" rel=\"noopener\">Perfluorotridecanoic acid (PFTrDA)<\/a><\/td><td class=\"oth\">7.30<\/td><td class=\"oth\">5<\/td><td class=\"oth\">662.7<\/td><td class=\"oth\">618.8<\/td><\/tr>\n<tr><td class=\"num\">25.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Perfluorotetradecanoic acid (PFTA)\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/376-06-7\/Perfluorotetradecanoic acid\" target=\"_blank\" rel=\"noopener\">Perfluorotetradecanoic acid (PFTA)<\/a><\/td><td class=\"oth\">7.60<\/td><td class=\"oth\">5<\/td><td class=\"oth\">712.7<\/td><td class=\"oth\">668.7<\/td><\/tr>\n<\/tbody><\/table><\/div><div class=\"chromatogram-conditions\"><h4>Conditions<\/h4><div class=\"conditions-container container-fluid\"><div class=\"row\"><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Column<\/th><td>Raptor C18  (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/9304A52\" rel=\"noopener\">cat.# 9304A52<\/a>)<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Dimensions:<\/th><td>50 mm x 2.1 mm ID<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Particle Size:<\/th><td>2.7 \u00b5m<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Pore Size:<\/th><td>90 \u00c5<\/td><\/tr><tr><td><\/td><tr><th class=\"sub conditions_header\" scope=\"row\">Temp.:<\/th><td>40 \u00b0C<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Standard\/Sample<\/th><td><\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Diluent:<\/th><td>96:4 Methanol:water<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Conc.:<\/th><td>5-10 ng\/mL<\/td><\/tr><td><\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Inj. Vol.:<\/th><td>2 \u00b5L <\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Mobile Phase<\/th><td><\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">A:<\/th><td>Water, 5 mM ammonium acetate <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">B:<\/th><td>Methanol <\/td><\/tr><tr><td><\/td><td><table class=\"cgram_ramp\"><thead><tr><th>Time (min)<\/th><th>Flow (mL\/min)<\/th><th>%A<\/th><th>%B<\/th><\/tr><\/thead><tbody><tr><td>0.00<\/td><td>0.4<\/td><td>70<\/td><td>30<\/td><\/tr><tr><td>8.00<\/td><td>0.4<\/td><td>10<\/td><td>90<\/td><\/tr><tr><td>8.01<\/td><td>0.4<\/td><td>70<\/td><td>30<\/td><\/tr><tr><td>10.0<\/td><td>0.4<\/td><td>70<\/td><td>30<\/td><\/tr><\/tbody><\/table><\/td><\/tr><\/table><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Detector<\/th><td>MS\/MS<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Ion Mode:<\/th><td>ESI- <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Mode:<\/th><td>MRM <\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Instrument<\/th><td>HPLC<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Notes<\/th><td>Want even better performance when analyzing metal-sensitive compounds? Check out Inert LC columns at <a href=\"https:\/\/www.restek.com\/inert\" target=\"_blank\" rel=\"noopener\">www.restek.com\/inert<\/a>.<\/td><\/tr><\/table><\/div><\/div><\/div><div class=\"chromatogram-related-skus\"><h4>Related SKUs<\/h4><ul><li>9304A52<\/li><\/ul><\/div><div class=\"chromatogram-related-searches\"><h4>Related Searches<\/h4><ul><li>ADONA<\/li><li>EPA537<\/li><li>GenX<\/li><li>HFPO-DA<\/li><li>PFAS<\/li><li>PFAS-library<\/li><li>PFOA<\/li><li>PFOS<\/li><li>Raptor<\/li><\/ul><\/div><div class=\"chromatogram-pdf-link\"><a href=\"https:\/\/ez.restek.com\/images\/cgram\/lc_ev0560.pdf\" target=\"_blank\" rel=\"noopener noreferrer\"><svg xmlns=\"http:\/\/www.w3.org\/2000\/svg\" width=\"18\" height=\"18\" viewBox=\"0 0 18 18\"><g data-name=\"Group 2996\"><path data-name=\"Rectangle 1246\" d=\"M0 0h18v18H0z\" style=\"fill: none;\"><\/path><\/g><g data-name=\"Group 2997\"><path data-name=\"Path 729\" d=\"M13.412 11.4v2.017H5.345V11.4H4v2.017a1.349 1.349 0 0 0 1.345 1.345h8.068a1.349 1.349 0 0 0 1.345-1.345V11.4zm-.672-2.694-.948-.948-1.741 1.735V4H8.706v5.493L6.965 7.758l-.948.948 3.361 3.361z\" transform=\"translate(-.437 -.414)\" style=\"fill: rgb(13, 123, 196);\"><\/path><\/g><\/svg>Download PDF<\/a><\/div><\/div><\/div>\n<\/div><\/div><\/div>\n\n\n<div style=\"height:20px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n\n\n\n<h3 class=\"wp-block-heading\">Ultrashort-Chain PFAS<\/h3>\n\n\n\n<p class=\"wp-block-paragraph\">When C8 PFAS were banned, other compounds with shorter alkyl chains were commercially adopted, and as the list of PFAS of interest grows, compounds with chains shorter than four carbons, or &#8220;ultrashort chain&#8221; (C2 and C3) are getting more attention. As the length of the carbon chain decreases, the influence of the polar head increases, ultimately decreasing retention on a C18 column, which has a retention mechanism based primarily on hydrophobic interaction.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">For C3 PFAS compounds like perfluoropropanoic acid (PFPrA) and perfluoropropanesulfonic acid (PFPrS), a C18 phase will still work when appropriate column dimensions are selected. In Figure 2, for example, a 100 x 3 mm Raptor C18 shows great performance, easily incorporating C3-length PFAS in a quick 11-minute analysis.<\/p>\n\n\n<div class=\"wp-block-custom-chromatogram-article-top\"><div class=\"chromatogram-article-placeholder\"><div class=\"figure-heading\"><strong>Figure 2:<\/strong>\u00a0The Raptor C18 stationary phase will also effectively separate short-chain PFAS, but column dimensions must be increased to ensure adequate retention.<\/div><div class='chromatogram-article-inner-full'><div class=\"chromatogram-article-inner\">\n<div class=\"wp-block-custom-chromatogram-article\"><div class=\"wp-block-custom-chromatogram-article\"><div class=\"chromatogram-image regular-image\"><img decoding=\"async\" src=\"https:\/\/ez.restek.com\/images\/cgram\/lc_ev0555.png\" alt=\"Ultra-short Chain, Legacy, and Alternative PFAS on Raptor C18 in Reagent Water\" title=\"-\"><\/div><p class=\"article-id\" style=\"text-align:center\"> LC_EV0555<\/p><div class=\"chromatogram-peaks\"><h4>Peaks<\/h4><table class=\"peaks col-lg-6 col-12 peak-50\">\n<thead><tr><th><\/th><th style=\"text-align: left;width: 75px\">Peaks<\/th><th style=\"text-align: center;width: 75px\">t<sub>R<\/sub> (min)<\/th><th style=\"text-align: center;width: 75px\">Conc.<br \/>(ng\/L)<\/th><th style=\"text-align: center;width: 75px\">Precursor Ion<\/th><th style=\"text-align: center;width: 75px\">Product Ion<\/th><\/tr><\/thead>\n<tbody><tr><td class=\"num\">1.<\/td><td class=\"cmpd\">Perfluoropropanoic acid (PFPrA)<\/td><td class=\"oth\">2.74<\/td><td class=\"oth\">80<\/td><td class=\"oth\">162.9<\/td><td class=\"oth\">119.0<\/td><\/tr>\n<tr><td class=\"num\">2.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Perfluorobutanoic acid (PFBA)\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/375-22-4\/Perfluorobutanoic acid\" target=\"_blank\" rel=\"noopener\">Perfluorobutanoic acid (PFBA)<\/a><\/td><td class=\"oth\">4.69<\/td><td class=\"oth\">80<\/td><td class=\"oth\">212.8<\/td><td class=\"oth\">169.0<\/td><\/tr>\n<tr><td class=\"num\">3.<\/td><td class=\"cmpd\">Perfluoropropanesulfonic acid (PFPrS)<\/td><td class=\"oth\">5.13<\/td><td class=\"oth\">80<\/td><td class=\"oth\">248.8<\/td><td class=\"oth\">79.6<\/td><\/tr>\n<tr><td class=\"num\">4.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Perfluorobutanesulfonic acid (PFBS)\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/375-73-5\/Perfluorobutanesulfonic acid\" target=\"_blank\" rel=\"noopener\">Perfluorobutanesulfonic acid (PFBS)<\/a><\/td><td class=\"oth\">6.14<\/td><td class=\"oth\">80<\/td><td class=\"oth\">298.8<\/td><td class=\"oth\">79.9<\/td><\/tr>\n<tr><td class=\"num\">5.<\/td><td class=\"cmpd\">Perfluoro-n-[1,2-<sup>13<\/sup>C<sub>2<\/sub>]hexanoic acid (<sup>13<\/sup>C<sub>2<\/sub>-PFHxA)<\/td><td class=\"oth\">6.75<\/td><td class=\"oth\">50<\/td><td class=\"oth\">314.9<\/td><td class=\"oth\">270.0<\/td><\/tr>\n<tr><td class=\"num\">6.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Hexafluoropropylene oxide-dimer acid (HFPO-DA)\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/13252-13-6\/Hexafluoropropylene oxide-dimer acid\" target=\"_blank\" rel=\"noopener\">Hexafluoropropylene oxide-dimer acid (HFPO-DA)<\/a><\/td><td class=\"oth\">6.92<\/td><td class=\"oth\">80<\/td><td class=\"oth\">285.0<\/td><td class=\"oth\">168.9<\/td><\/tr>\n<\/tbody><\/table>\n<table class=\"peaks col-lg-6 col-12 peak-50\"><thead><tr><th><\/th><th style=\"text-align: left;width: 75px\">Peaks<\/th><th style=\"text-align: center;width: 75px\">t<sub>R<\/sub> (min)<\/th><th style=\"text-align: center;width: 75px\">Conc.<br \/>(ng\/L)<\/th><th style=\"text-align: center;width: 75px\">Precursor Ion<\/th><th style=\"text-align: center;width: 75px\">Product Ion<\/th><\/tr><\/thead><tbody>\n<tr><td class=\"num\">7.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Ammonium 4,8-dioxa-3H-perfluorononanoate (ADONA)\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/919005-14-4\/Ammonium 4,8-dioxa-3H-perfluorononanoate\" target=\"_blank\" rel=\"noopener\">Ammonium 4,8-dioxa-3H-perfluorononanoate (ADONA)<\/a><\/td><td class=\"oth\">7.33<\/td><td class=\"oth\">80<\/td><td class=\"oth\">376.9<\/td><td class=\"oth\">250.7<\/td><\/tr>\n<tr><td class=\"num\">8.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Perfluorooctanoic acid (PFOA)\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/335-67-1\/Perfluorooctanoic acid\" target=\"_blank\" rel=\"noopener\">Perfluorooctanoic acid (PFOA)<\/a><\/td><td class=\"oth\">7.70<\/td><td class=\"oth\">80<\/td><td class=\"oth\">413.1<\/td><td class=\"oth\">368.9<\/td><\/tr>\n<tr><td class=\"num\">9.<\/td><td class=\"cmpd\">Perfluoro-[1,2-<sup>13<\/sup>C<sub>2<\/sub>]octanoic acid (<sup>13<\/sup>C<sub>2<\/sub>-PFOA)<\/td><td class=\"oth\">7.70<\/td><td class=\"oth\">50<\/td><td class=\"oth\">415.0<\/td><td class=\"oth\">370.0<\/td><\/tr>\n<tr><td class=\"num\">10.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Perfluorooctanesulfonic acid (PFOS)\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/1763-23-1\/Perfluorooctanesulfonic acid\" target=\"_blank\" rel=\"noopener\">Perfluorooctanesulfonic acid (PFOS)<\/a><\/td><td class=\"oth\">8.01<\/td><td class=\"oth\">80<\/td><td class=\"oth\">498.8<\/td><td class=\"oth\">80.0<\/td><\/tr>\n<tr><td class=\"num\">11.<\/td><td class=\"cmpd\">Perfluoro-[1,2,3,4-<sup>13<\/sup>C<sub>4<\/sub>]octanesulfonic acid (<sup>13<\/sup>C<sub>4<\/sub>-PFOS)<\/td><td class=\"oth\">8.01<\/td><td class=\"oth\">50<\/td><td class=\"oth\">503.0<\/td><td class=\"oth\">80.0<\/td><\/tr>\n<tr><td class=\"num\">12.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 9-Chlorohexadecafluoro-3-oxanonane-1-sulfonate (9Cl-PF3ONS)\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/756426-58-1\/9-Chlorohexadecafluoro-3-oxanonane-1-sulfonate\" target=\"_blank\" rel=\"noopener\">9-Chlorohexadecafluoro-3-oxanonane-1-sulfonate (9Cl-PF3ONS)<\/a><\/td><td class=\"oth\">8.15<\/td><td class=\"oth\">80<\/td><td class=\"oth\">530.8<\/td><td class=\"oth\">350.7<\/td><\/tr>\n<tr><td class=\"num\">13.<\/td><td class=\"cmpd\">11-Chloroeicosafluoro-3-oxanonane-1-sulfonate (11Cl-PF3OUdS)<\/td><td class=\"oth\">8.61<\/td><td class=\"oth\">80<\/td><td class=\"oth\">630.7<\/td><td class=\"oth\">451.0<\/td><\/tr>\n<\/tbody><\/table><\/div><div class=\"chromatogram-conditions\"><h4>Conditions<\/h4><div class=\"conditions-container container-fluid\"><div class=\"row\"><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Column<\/th><td>Raptor C18  (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/9304A1E\" rel=\"noopener\">cat.# 9304A1E<\/a>)<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Dimensions:<\/th><td>100 mm x 3 mm ID<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Particle Size:<\/th><td>2.7 \u00b5m<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Pore Size:<\/th><td>90 \u00c5<\/td><\/tr><tr><td><\/td><tr><th class=\"sub conditions_header\" scope=\"row\">Temp.:<\/th><td>40 \u00b0C<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Standard\/Sample<\/th><td><\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Conc.:<\/th><td>80  ppt<\/td><\/tr><td><\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Inj. Vol.:<\/th><td>10 \u00b5L <\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Mobile Phase<\/th><td><\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">A:<\/th><td>Water, 5 mM ammonium acetate <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">B:<\/th><td>Methanol <\/td><\/tr><tr><td><\/td><td><table class=\"cgram_ramp\"><thead><tr><th>Time (min)<\/th><th>Flow (mL\/min)<\/th><th>%A<\/th><th>%B<\/th><\/tr><\/thead><tbody><tr><td>0.00<\/td><td>0.25<\/td><td>80<\/td><td>20<\/td><\/tr><tr><td>7.00<\/td><td>0.25<\/td><td>5<\/td><td>95<\/td><\/tr><tr><td>9.00<\/td><td>0.25<\/td><td>5<\/td><td>95<\/td><\/tr><tr><td>9.01<\/td><td>0.25<\/td><td>80<\/td><td>20<\/td><\/tr><tr><td>11.0<\/td><td>0.25<\/td><td>80<\/td><td>20<\/td><\/tr><\/tbody><\/table><\/td><\/tr><\/table><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Detector<\/th><td>MS\/MS<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Ion Mode:<\/th><td>ESI- <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Mode:<\/th><td>MRM <\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Instrument<\/th><td>UHPLC<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Sample Preparation<\/th><td>In a polypropylene vial, 250 \u00b5L of reagent water (fortified at 80 ppt) was mixed with 250 \u00b5L of 40:60 reagent water:methanol and 5 \u00b5L of internal standard solution (5 ng\/mL of <sup>13<\/sup>C<sub>2<\/sub>-PFHxA, <sup>13<\/sup>C<sub>2<\/sub>-PFOA, <sup>13<\/sup>C<sub>4<\/sub>-PFOS in methanol). The vial was capped with a polyethylene cap prior to analysis.  <\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Notes<\/th><td>A PFAS delay column (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/27854\" rel=\"noopener\">cat.# 27854<\/a>) was installed between the pump mixer and the injector.<br \/><br \/>Want even better performance when analyzing metal-sensitive compounds? Check out Inert LC columns at <a href=\"https:\/\/www.restek.com\/inert\" target=\"_blank\" rel=\"noopener\">www.restek.com\/inert<\/a>.<br \/><\/td><\/tr><\/table><\/div><\/div><\/div><div class=\"chromatogram-related-skus\"><h4>Related SKUs<\/h4><ul><li>9304A1E<\/li><li>27854<\/li><\/ul><\/div><div class=\"chromatogram-related-searches\"><h4>Related Searches<\/h4><ul><li>11Cl-PF3OUdS<\/li><li>9Cl-PF3ONS<\/li><li>ADONA<\/li><li>HFPO-DA<\/li><li>PFAS<\/li><li>PFAS-library<\/li><li>PFPrA<\/li><li>PFPrS<\/li><li>Ultrashort Chain<\/li><\/ul><\/div><div class=\"chromatogram-pdf-link\"><a href=\"https:\/\/ez.restek.com\/images\/cgram\/lc_ev0555.pdf\" target=\"_blank\" rel=\"noopener noreferrer\"><svg xmlns=\"http:\/\/www.w3.org\/2000\/svg\" width=\"18\" height=\"18\" viewBox=\"0 0 18 18\"><g data-name=\"Group 2996\"><path data-name=\"Rectangle 1246\" d=\"M0 0h18v18H0z\" style=\"fill: none;\"><\/path><\/g><g data-name=\"Group 2997\"><path data-name=\"Path 729\" d=\"M13.412 11.4v2.017H5.345V11.4H4v2.017a1.349 1.349 0 0 0 1.345 1.345h8.068a1.349 1.349 0 0 0 1.345-1.345V11.4zm-.672-2.694-.948-.948-1.741 1.735V4H8.706v5.493L6.965 7.758l-.948.948 3.361 3.361z\" transform=\"translate(-.437 -.414)\" style=\"fill: rgb(13, 123, 196);\"><\/path><\/g><\/svg>Download PDF<\/a><\/div><\/div><\/div>\n<\/div><\/div><\/div>\n\n\n<div style=\"height:20px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n\n\n\n<p class=\"wp-block-paragraph\">If C2 PFAS (e.g., trifluoroacetic acid) eventually make it on a list of monitored compounds, an alternate phase chemistry that targets the polar moiety of a PFAS molecule is required. Moving from the mainly hydrophobic interactions of a C18 column to a Raptor Polar X column\u2014which has a stationary phase that is capable of both ion-exchange and HILIC separation modes\u2014permits the retention of ultrashort-chain PFAS. Along with ultrashort-chain PFAS, the Raptor Polar X column is also capable of retaining and separating short-chain, legacy, and alternate PFAS compounds in the same analysis, providing the most comprehensive single-analysis PFAS method, as shown in Figure 3.<\/p>\n\n\n<div class=\"wp-block-custom-chromatogram-article-top\"><div class=\"chromatogram-article-placeholder\"><div class=\"figure-heading\"><strong>Figure 3:<\/strong>\u00a0Raptor Polar X columns utilize multiple retention modes, making them the best choice for analyzing ultrashort-chain, legacy, and alternative PFAS with a single method.<\/div><div class='chromatogram-article-inner-full'><div class=\"chromatogram-article-inner\">\n<div class=\"wp-block-custom-chromatogram-article\"><div class=\"wp-block-custom-chromatogram-article\"><div class=\"chromatogram-image regular-image\"><img decoding=\"async\" src=\"https:\/\/ez.restek.com\/images\/cgram\/lc_ev0569.png\" alt=\"Ultrashort- Through Long-Chain and Alternative PFAS on Raptor Polar X\" title=\"-\"><\/div><p class=\"article-id\" style=\"text-align:center\"> LC_EV0569<\/p><div class=\"chromatogram-peaks\"><h4>Peaks<\/h4><table class=\"peaks col-lg-6 col-12 peak-50\">\n<thead><tr><th><\/th><th style=\"text-align: left;width: 75px\">Peaks<\/th><th style=\"text-align: center;width: 75px\">t<sub>R<\/sub> (min)<\/th><th style=\"text-align: center;width: 75px\">Conc.<br \/>(ng\/L)<\/th><th style=\"text-align: center;width: 75px\">Precursor Ion<\/th><th style=\"text-align: center;width: 75px\">Product Ion<\/th><\/tr><\/thead>\n<tbody><tr><td class=\"num\">1.<\/td><td class=\"cmpd\">11-Chloroeicosafluoro-3-oxanonane-1-sulfonate (11CL-PF3OUdS)<\/td><td class=\"oth\">1.25<\/td><td class=\"oth\">400<\/td><td class=\"oth\">630.78<\/td><td class=\"oth\">450.80<\/td><\/tr>\n<tr><td class=\"num\">2.<\/td><td class=\"cmpd\">9-Chlorohexadecafluoro-3-oxanonane-1-sulfonate (9Cl-PF3ONS)<\/td><td class=\"oth\">1.34<\/td><td class=\"oth\">400<\/td><td class=\"oth\">530.78<\/td><td class=\"oth\">350.85<\/td><\/tr>\n<tr><td class=\"num\">3.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Perfluorooctanesulfonic acid (PFOS)\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/1763-23-1\/Perfluorooctanesulfonic acid\" target=\"_blank\" rel=\"noopener\">Perfluorooctanesulfonic acid (PFOS)<\/a><\/td><td class=\"oth\">1.38<\/td><td class=\"oth\">400<\/td><td class=\"oth\">498.84<\/td><td class=\"oth\">79.97<\/td><\/tr>\n<tr><td class=\"num\">4.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Perfluorohexanesulfonic acid (PFHxS)\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/355-46-4\/Perfluorohexanesulfonic acid\" target=\"_blank\" rel=\"noopener\">Perfluorohexanesulfonic acid (PFHxS)<\/a><\/td><td class=\"oth\">1.49<\/td><td class=\"oth\">400<\/td><td class=\"oth\">398.90<\/td><td class=\"oth\">79.97<\/td><\/tr>\n<tr><td class=\"num\">5.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Perfluorobutanesulfonic acid (PFBS)\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/375-73-5\/Perfluorobutanesulfonic acid\" target=\"_blank\" rel=\"noopener\">Perfluorobutanesulfonic acid (PFBS)<\/a><\/td><td class=\"oth\">1.64<\/td><td class=\"oth\">400<\/td><td class=\"oth\">298.97<\/td><td class=\"oth\">79.97<\/td><\/tr>\n<tr><td class=\"num\">6.<\/td><td class=\"cmpd\">Perfluoropropanesulfonic acid (PFPrS)<\/td><td class=\"oth\">1.73<\/td><td class=\"oth\">400<\/td><td class=\"oth\">248.97<\/td><td class=\"oth\">79.98<\/td><\/tr>\n<tr><td class=\"num\">7.<\/td><td class=\"cmpd\">Perfluoroethanesulfonic acid (PFEtS)<\/td><td class=\"oth\">1.86<\/td><td class=\"oth\">400<\/td><td class=\"oth\">198.98<\/td><td class=\"oth\">79.92<\/td><\/tr>\n<\/tbody><\/table>\n<table class=\"peaks col-lg-6 col-12 peak-50\"><thead><tr><th><\/th><th style=\"text-align: left;width: 75px\">Peaks<\/th><th style=\"text-align: center;width: 75px\">t<sub>R<\/sub> (min)<\/th><th style=\"text-align: center;width: 75px\">Conc.<br \/>(ng\/L)<\/th><th style=\"text-align: center;width: 75px\">Precursor Ion<\/th><th style=\"text-align: center;width: 75px\">Product Ion<\/th><\/tr><\/thead><tbody>\n<tr><td class=\"num\">8.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Hexafluoropropylene oxide dimer acid (HFPO-DA)\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/13252-13-6\/Hexafluoropropylene oxide dimer acid\" target=\"_blank\" rel=\"noopener\">Hexafluoropropylene oxide dimer acid (HFPO-DA)<\/a><\/td><td class=\"oth\">2.06<\/td><td class=\"oth\">400<\/td><td class=\"oth\">284.97<\/td><td class=\"oth\">168.92<\/td><\/tr>\n<tr><td class=\"num\">9.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Perfluorooctanoic acid (PFOA)\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/335-67-1\/Perfluorooctanoic acid\" target=\"_blank\" rel=\"noopener\">Perfluorooctanoic acid (PFOA)<\/a><\/td><td class=\"oth\">2.11<\/td><td class=\"oth\">400<\/td><td class=\"oth\">412.90<\/td><td class=\"oth\">368.91<\/td><\/tr>\n<tr><td class=\"num\">10.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Ammonium 4,8-dioxa-3H-perfluorononanoate (ADONA)\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/919005-14-4\/Ammonium 4,8-dioxa-3H-perfluorononanoate\" target=\"_blank\" rel=\"noopener\">Ammonium 4,8-dioxa-3H-perfluorononanoate (ADONA)<\/a><\/td><td class=\"oth\">2.15<\/td><td class=\"oth\">400<\/td><td class=\"oth\">376.90<\/td><td class=\"oth\">250.93<\/td><\/tr>\n<tr><td class=\"num\">11.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Perfluorohexanoic acid (PFHxA)\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/307-24-4\/Perfluorohexanoic acid\" target=\"_blank\" rel=\"noopener\">Perfluorohexanoic acid (PFHxA)<\/a><\/td><td class=\"oth\">2.36<\/td><td class=\"oth\">400<\/td><td class=\"oth\">312.97<\/td><td class=\"oth\">268.90<\/td><\/tr>\n<tr><td class=\"num\">12.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Perfluorobutanoic acid (PFBA)\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/375-22-4\/Perfluorobutanoic acid\" target=\"_blank\" rel=\"noopener\">Perfluorobutanoic acid (PFBA)<\/a><\/td><td class=\"oth\">2.76<\/td><td class=\"oth\">400<\/td><td class=\"oth\">212.97<\/td><td class=\"oth\">168.97<\/td><\/tr>\n<tr><td class=\"num\">13.<\/td><td class=\"cmpd\">Perfluoropropionic acid (PFPrA)<\/td><td class=\"oth\">3.06<\/td><td class=\"oth\">400<\/td><td class=\"oth\">163.03<\/td><td class=\"oth\">119.01<\/td><\/tr>\n<tr><td class=\"num\">14.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Trifluoroacetic acid (TFA)\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/76-05-1\/Trifluoroacetic acid\" target=\"_blank\" rel=\"noopener\">Trifluoroacetic acid (TFA)<\/a><\/td><td class=\"oth\">3.77<\/td><td class=\"oth\">400<\/td><td class=\"oth\">113.03<\/td><td class=\"oth\">69.01<\/td><\/tr>\n<\/tbody><\/table><\/div><div class=\"chromatogram-conditions\"><h4>Conditions<\/h4><div class=\"conditions-container container-fluid\"><div class=\"row\"><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Column<\/th><td>Raptor Polar X  (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/9311A52\" rel=\"noopener\">cat.# 9311A52<\/a>)<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Dimensions:<\/th><td>50 mm x 2.1 mm ID<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Particle Size:<\/th><td>2.7 \u00b5m<\/td><\/tr><tr><td><\/td><tr><th class=\"sub conditions_header\" scope=\"row\">Temp.:<\/th><td>40 \u00b0C<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Standard\/Sample<\/th><td><\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Diluent:<\/th><td>50:50 Water:methanol<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Conc.:<\/th><td>400  ng\/L<\/td><\/tr><td><\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Inj. Vol.:<\/th><td>10 \u00b5L <\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Mobile Phase<\/th><td><\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">A:<\/th><td>Water, 10 mM ammonium formate, 0.05% formic acid   <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">B:<\/th><td>60:40 Acetonitrile:methanol, 0.05% formic acid <\/td><\/tr><tr><td><\/td><td><table class=\"cgram_ramp\"><thead><tr><th>Time (min)<\/th><th>Flow (mL\/min)<\/th><th>%A<\/th><th>%B<\/th><\/tr><\/thead><tbody><tr><td>0.00<\/td><td>0.5<\/td><td>15<\/td><td>85<\/td><\/tr><tr><td>8.00<\/td><td>0.5<\/td><td>15<\/td><td>85<\/td><\/tr><\/tbody><\/table><\/td><\/tr><\/table><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Detector<\/th><td>MS\/MS<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Ion Mode:<\/th><td>ESI- <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Mode:<\/th><td>MRM <\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Instrument<\/th><td>UHPLC<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Notes<\/th><td>To achieve more stable and robust performance with Polar X columns, the recommended LC conditions have changed from those that were used here. The new recommended conditions are an aqueous mobile phase (mobile phase A) consisting of 5 mM or 10 mM ammonium formate with 0.1% formic acid in water and an organic mobile phase (mobile phase B) of 100% acetonitrile. Higher ammonium formate concentrations will result in reduced analyte retention. Therefore, laboratories can optimize retention by adjusting both the ammonium formate concentration and the flow rate to best meet their analytical requirements. In addition, it is recommended that labs evaluate different isocratic conditions by varying the amount of mobile phase B between 70% and 85% acetonitrile. This approach can help labs identify the most suitable separation conditions for their specific target analytes and sample matrices of interest.<\/td><\/tr><\/table><\/div><\/div><\/div><div class=\"chromatogram-related-skus\"><h4>Related SKUs<\/h4><ul><li>9311A52<\/li><\/ul><\/div><div class=\"chromatogram-related-searches\"><h4>Related Searches<\/h4><ul><li>PFAS<\/li><li>PFAS-library<\/li><li>ultrashort-chain<\/li><\/ul><\/div><div class=\"chromatogram-pdf-link\"><a href=\"https:\/\/ez.restek.com\/images\/cgram\/lc_ev0569.pdf\" target=\"_blank\" rel=\"noopener noreferrer\"><svg xmlns=\"http:\/\/www.w3.org\/2000\/svg\" width=\"18\" height=\"18\" viewBox=\"0 0 18 18\"><g data-name=\"Group 2996\"><path data-name=\"Rectangle 1246\" d=\"M0 0h18v18H0z\" style=\"fill: none;\"><\/path><\/g><g data-name=\"Group 2997\"><path data-name=\"Path 729\" d=\"M13.412 11.4v2.017H5.345V11.4H4v2.017a1.349 1.349 0 0 0 1.345 1.345h8.068a1.349 1.349 0 0 0 1.345-1.345V11.4zm-.672-2.694-.948-.948-1.741 1.735V4H8.706v5.493L6.965 7.758l-.948.948 3.361 3.361z\" transform=\"translate(-.437 -.414)\" style=\"fill: rgb(13, 123, 196);\"><\/path><\/g><\/svg>Download PDF<\/a><\/div><\/div><\/div>\n<\/div><\/div><\/div>\n\n\n<div style=\"height:20px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n\n\n\n<h2 class=\"wp-block-heading\">Column Particle Size and Particle Type Selection<\/h2>\n\n\n\n<p class=\"wp-block-paragraph\">In addition to phase chemistry and column dimensions, choices about particle size and type have to be made. Ultimately, a 2.7 \u00b5m superficially porous particle (SPP), such as those used in Raptor columns, is the most versatile choice. PFAS LC columns made with these particles produce efficient chromatography that rivals the sub-2 \u00b5m fully-porous particles (FPP) without the associated ultra-high pressure, which allows labs using either UHPLC or HPLC instruments to obtain fast, efficient analyses.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">However, choice in particle size and type will have a greater or lesser effect depending on your instrument setup. For instance, labs operating traditional HPLC instruments and regularly using 5 \u00b5m FPP columns may be surprised at the improvement an SPP column can offer, regardless of the particle size. With SPP columns, you could easily see improvements in chromatographic efficiency and better peak shapes while still operating in the pressure region of most HPLC instruments. Faster runs with the same equipment are a winning combination for many labs interested in increasing sample throughput without the capital expense of a UHPLC instrument.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">For those labs already using UHPLC instruments, the choice of particle size, especially for SPP-packed columns, makes less of a difference in efficiency and analysis speed. Figure 4 illustrates the effect of particle size using three Raptor SPP columns and a panel of PFAS on a UHPLC system. Even though the peaks are narrowest for the 1.8 \u00b5m particle column, as dictated by general chromatographic principles, the difference in the observed efficiency of 5 \u00b5m and 1.8 \u00b5m columns was not significant. However, the difference in the back pressure generated while achieving those very similar results is significant, with pressures being in the range of traditional HPLC instruments for the 5 and 2.7 \u00b5m particle columns. Using a PFAS LC column packed with these particles, you can get UHPLC performance without UHPLC pressure.<\/p>\n\n\n<div class=\"wp-block-custom-chromatogram-article-top\"><div class=\"chromatogram-article-placeholder\"><div class=\"figure-heading\"><strong>Figure 4:<\/strong>\u00a0For SPP columns, similar chromatographic performance can be obtained with different particle sizes, but choosing 2.7 or 5 \u00b5m particle columns keeps back pressure within the limits of traditional HPLC systems. (All columns are 50 mm x 2.1 mm.)Figure Heading<\/div><div class='chromatogram-article-inner-full'><div class=\"chromatogram-article-inner\">\n\n<p class=\"wp-block-paragraph\"><strong>1.8 \u00b5m Raptor C18<\/strong><br>6500\u20138000 psi<\/p>\n\n\n<div class=\"wp-block-custom-chromatogram-article\"><div class=\"wp-block-custom-chromatogram-article\"><div class=\"chromatogram-image regular-image\"><img decoding=\"async\" src=\"https:\/\/ez.restek.com\/images\/cgram\/lc_ev0540.png\" alt=\"Perfluorinated Compounds (PFCs) on Raptor C18 (1.8 \u03bcm)\" title=\"-\"><\/div><p class=\"article-id\" style=\"text-align:center\"> LC_EV0540<\/p><div class=\"chromatogram-related-skus\"><h4>Related SKUs<\/h4><ul><li>9304252<\/li><\/ul><\/div><div class=\"chromatogram-related-searches\"><h4>Related Searches<\/h4><ul><li>PFAS<\/li><li>PFAS-library<\/li><li>PFC<\/li><li>PFOA<\/li><li>PFOS<\/li><li>fluorotelomer<\/li><li>perfluorinated compound<\/li><li>perfluoroalkyl<\/li><\/ul><\/div><div class=\"chromatogram-pdf-link\"><a href=\"https:\/\/ez.restek.com\/images\/cgram\/lc_ev0540.pdf\" target=\"_blank\" rel=\"noopener noreferrer\"><svg xmlns=\"http:\/\/www.w3.org\/2000\/svg\" width=\"18\" height=\"18\" viewBox=\"0 0 18 18\"><g data-name=\"Group 2996\"><path data-name=\"Rectangle 1246\" d=\"M0 0h18v18H0z\" style=\"fill: none;\"><\/path><\/g><g data-name=\"Group 2997\"><path data-name=\"Path 729\" d=\"M13.412 11.4v2.017H5.345V11.4H4v2.017a1.349 1.349 0 0 0 1.345 1.345h8.068a1.349 1.349 0 0 0 1.345-1.345V11.4zm-.672-2.694-.948-.948-1.741 1.735V4H8.706v5.493L6.965 7.758l-.948.948 3.361 3.361z\" transform=\"translate(-.437 -.414)\" style=\"fill: rgb(13, 123, 196);\"><\/path><\/g><\/svg>Download PDF<\/a><\/div><\/div><\/div>\n\n\n<p class=\"wp-block-paragraph\"><strong>2.7 \u00b5m Raptor C18<\/strong><br>4000\u20135500 psi<\/p>\n\n\n<div class=\"wp-block-custom-chromatogram-article\"><div class=\"wp-block-custom-chromatogram-article\"><div class=\"chromatogram-image regular-image\"><img decoding=\"async\" src=\"https:\/\/ez.restek.com\/images\/cgram\/lc_ev0551.png\" alt=\"Perfluorinated Compounds (PFCs) on Raptor C18 2.7 \u03bcm\" title=\"-\"><\/div><p class=\"article-id\" style=\"text-align:center\"> LC_EV0551<\/p><div class=\"chromatogram-related-skus\"><h4>Related SKUs<\/h4><ul><li>9304A52<\/li><\/ul><\/div><div class=\"chromatogram-related-searches\"><h4>Related Searches<\/h4><ul><li>PFAS<\/li><li>PFAS-library<\/li><li>PFC<\/li><li>PFOA<\/li><li>PFOS<\/li><li>fluorotelomer<\/li><li>perfluorinated compound<\/li><li>perfluoroalkyl<\/li><\/ul><\/div><div class=\"chromatogram-pdf-link\"><a href=\"https:\/\/ez.restek.com\/images\/cgram\/lc_ev0551.pdf\" target=\"_blank\" rel=\"noopener noreferrer\"><svg xmlns=\"http:\/\/www.w3.org\/2000\/svg\" width=\"18\" height=\"18\" viewBox=\"0 0 18 18\"><g data-name=\"Group 2996\"><path data-name=\"Rectangle 1246\" d=\"M0 0h18v18H0z\" style=\"fill: none;\"><\/path><\/g><g data-name=\"Group 2997\"><path data-name=\"Path 729\" d=\"M13.412 11.4v2.017H5.345V11.4H4v2.017a1.349 1.349 0 0 0 1.345 1.345h8.068a1.349 1.349 0 0 0 1.345-1.345V11.4zm-.672-2.694-.948-.948-1.741 1.735V4H8.706v5.493L6.965 7.758l-.948.948 3.361 3.361z\" transform=\"translate(-.437 -.414)\" style=\"fill: rgb(13, 123, 196);\"><\/path><\/g><\/svg>Download PDF<\/a><\/div><\/div><\/div>\n\n\n<p class=\"wp-block-paragraph\"><strong>5 \u00b5m Raptor C18<\/strong><br>2000\u20133500 psi<\/p>\n\n\n<div class=\"wp-block-custom-chromatogram-article\"><div class=\"wp-block-custom-chromatogram-article\"><div class=\"chromatogram-image regular-image\"><img decoding=\"async\" src=\"https:\/\/ez.restek.com\/images\/cgram\/lc_ev0541.png\" alt=\"Perfluorinated Compounds (PFCs) on Raptor C18 (5 \u03bcm)\" title=\"-\"><\/div><p class=\"article-id\" style=\"text-align:center\"> LC_EV0541<\/p><div class=\"chromatogram-related-skus\"><h4>Related SKUs<\/h4><ul><li>9304552<\/li><\/ul><\/div><div class=\"chromatogram-related-searches\"><h4>Related Searches<\/h4><ul><li>PFAS<\/li><li>PFAS-library<\/li><li>PFC<\/li><li>PFOA<\/li><li>PFOS<\/li><li>fluorotelomer<\/li><li>perfluorinated compound<\/li><li>perfluoroalkyl<\/li><\/ul><\/div><div class=\"chromatogram-pdf-link\"><a href=\"https:\/\/ez.restek.com\/images\/cgram\/lc_ev0541.pdf\" target=\"_blank\" rel=\"noopener noreferrer\"><svg xmlns=\"http:\/\/www.w3.org\/2000\/svg\" width=\"18\" height=\"18\" viewBox=\"0 0 18 18\"><g data-name=\"Group 2996\"><path data-name=\"Rectangle 1246\" d=\"M0 0h18v18H0z\" style=\"fill: none;\"><\/path><\/g><g data-name=\"Group 2997\"><path data-name=\"Path 729\" d=\"M13.412 11.4v2.017H5.345V11.4H4v2.017a1.349 1.349 0 0 0 1.345 1.345h8.068a1.349 1.349 0 0 0 1.345-1.345V11.4zm-.672-2.694-.948-.948-1.741 1.735V4H8.706v5.493L6.965 7.758l-.948.948 3.361 3.361z\" transform=\"translate(-.437 -.414)\" style=\"fill: rgb(13, 123, 196);\"><\/path><\/g><\/svg>Download PDF<\/a><\/div><\/div><\/div>\n<\/div><\/div><\/div>\n\n\n<div style=\"height:20px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n\n\n\n<p class=\"wp-block-paragraph\">When it comes to generating the most efficiency\u2014which often translates into fast run times\u2014while keeping instrument back pressure as low as possible, a PFAS LC column containing SPP particles is the clear winner. However, many labs that have a long history using FPP columns may prefer to continue using them. The good news is that an FPP column can be used to successfully analyze PFAS as well. The higher surface area and carbon load in an FPP column, such as a Force C18 column, results in increased chromatographic retention compared to an SPP column (Raptor C18) of similar particle size (Figure 5).<\/p>\n\n\n<div class=\"wp-block-custom-chromatogram-article-top\"><div class=\"chromatogram-article-placeholder\"><div class=\"figure-heading\"><strong>Figure 5<\/strong>: If an FPP column is preferred, Force C18 columns provide effective separations for PFAS analysis. (All columns are 50 mm x 2.1 mm.)<\/div><div class='chromatogram-article-inner-full'><div class=\"chromatogram-article-inner\">\n\n<p class=\"wp-block-paragraph\"><strong>1.8 \u00b5m Raptor C18<\/strong><br>6500\u20138000 psi<\/p>\n\n\n<div class=\"wp-block-custom-chromatogram-article\"><div class=\"wp-block-custom-chromatogram-article\"><div class=\"chromatogram-image regular-image\"><img decoding=\"async\" src=\"https:\/\/ez.restek.com\/images\/cgram\/lc_ev0540.png\" alt=\"Perfluorinated Compounds (PFCs) on Raptor C18 (1.8 \u03bcm)\" title=\"-\"><\/div><p class=\"article-id\" style=\"text-align:center\"> LC_EV0540<\/p><div class=\"chromatogram-related-skus\"><h4>Related SKUs<\/h4><ul><li>9304252<\/li><\/ul><\/div><div class=\"chromatogram-related-searches\"><h4>Related Searches<\/h4><ul><li>PFAS<\/li><li>PFAS-library<\/li><li>PFC<\/li><li>PFOA<\/li><li>PFOS<\/li><li>fluorotelomer<\/li><li>perfluorinated compound<\/li><li>perfluoroalkyl<\/li><\/ul><\/div><div class=\"chromatogram-pdf-link\"><a href=\"https:\/\/ez.restek.com\/images\/cgram\/lc_ev0540.pdf\" target=\"_blank\" rel=\"noopener noreferrer\"><svg xmlns=\"http:\/\/www.w3.org\/2000\/svg\" width=\"18\" height=\"18\" viewBox=\"0 0 18 18\"><g data-name=\"Group 2996\"><path data-name=\"Rectangle 1246\" d=\"M0 0h18v18H0z\" style=\"fill: none;\"><\/path><\/g><g data-name=\"Group 2997\"><path data-name=\"Path 729\" d=\"M13.412 11.4v2.017H5.345V11.4H4v2.017a1.349 1.349 0 0 0 1.345 1.345h8.068a1.349 1.349 0 0 0 1.345-1.345V11.4zm-.672-2.694-.948-.948-1.741 1.735V4H8.706v5.493L6.965 7.758l-.948.948 3.361 3.361z\" transform=\"translate(-.437 -.414)\" style=\"fill: rgb(13, 123, 196);\"><\/path><\/g><\/svg>Download PDF<\/a><\/div><\/div><\/div>\n\n\n<p class=\"wp-block-paragraph\"><strong>1.8 \u00b5m Force C18<\/strong><br>7550\u20139250 psi<\/p>\n\n\n<div class=\"wp-block-custom-chromatogram-article\"><div class=\"wp-block-custom-chromatogram-article\"><div class=\"chromatogram-image regular-image\"><img decoding=\"async\" src=\"https:\/\/ez.restek.com\/images\/cgram\/lc_ev0552.png\" alt=\"Perfluorinated Compounds (PFCs) on Force C18 (1.8 \u03bcm)\" title=\"-\"><\/div><p class=\"article-id\" style=\"text-align:center\"> LC_EV0552<\/p><div class=\"chromatogram-related-skus\"><h4>Related SKUs<\/h4><ul><li>9634252<\/li><\/ul><\/div><div class=\"chromatogram-related-searches\"><h4>Related Searches<\/h4><ul><li>PFAS<\/li><li>PFAS-library<\/li><li>PFC<\/li><li>PFOA<\/li><li>PFOS<\/li><li>fluorotelomer<\/li><li>perfluorinated compound<\/li><li>perfluoroalkyl<\/li><\/ul><\/div><div class=\"chromatogram-pdf-link\"><a href=\"https:\/\/ez.restek.com\/images\/cgram\/lc_ev0552.pdf\" target=\"_blank\" rel=\"noopener noreferrer\"><svg xmlns=\"http:\/\/www.w3.org\/2000\/svg\" width=\"18\" height=\"18\" viewBox=\"0 0 18 18\"><g data-name=\"Group 2996\"><path data-name=\"Rectangle 1246\" d=\"M0 0h18v18H0z\" style=\"fill: none;\"><\/path><\/g><g data-name=\"Group 2997\"><path data-name=\"Path 729\" d=\"M13.412 11.4v2.017H5.345V11.4H4v2.017a1.349 1.349 0 0 0 1.345 1.345h8.068a1.349 1.349 0 0 0 1.345-1.345V11.4zm-.672-2.694-.948-.948-1.741 1.735V4H8.706v5.493L6.965 7.758l-.948.948 3.361 3.361z\" transform=\"translate(-.437 -.414)\" style=\"fill: rgb(13, 123, 196);\"><\/path><\/g><\/svg>Download PDF<\/a><\/div><\/div><\/div>\n<\/div><\/div><\/div>\n\n\n<div style=\"height:20px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n\n\n\n<h3 class=\"wp-block-heading\">Conclusion<\/h3>\n\n\n\n<p class=\"wp-block-paragraph\">When choosing a PFAS LC column, a 2.7 \u00b5m Raptor SPP column will provide excellent resolution, short run times, and compatibility with both UHPLC and HPLC instruments. The choice of stationary phase depends on the range of PFAS being monitored, and a Raptor C18 column is an excellent choice for PFAS down to C3 chain lengths. However, for the most comprehensive single analysis, including the full range of ultrashort-chain (C2-C3) PFAS and above, as well as alternate PFAS, a Raptor Polar X column is the best choice.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Note that for trace-level analysis using a Raptor C18 column, false positives or elevated responses can be mitigated by adding a PFAS delay column. Use of a PFAS delay column will eliminate instrument-related background PFAS contaminants that can coelute with sample analytes.<\/p>\n\n\n\n<div style=\"height:20px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n\n\n        <div class=\"cpb\">\n            <h3 class=\"cpb-heading\">Products Mentioned<\/h3>\n            <hr class=\"cpb-heading-underline\" \/>\n            <div class=\"cpb-list\">\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/9304A52\">                                Catalog No. 9304A52                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">Raptor C18, 2.7 \u00b5m, 50 x 2.1 mm HPLC Column<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/9304A52\">View Product<\/a>\n                                                    <\/div>\n                    <\/div>\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/27854\">                                Catalog No. 27854                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">PFAS Delay Column, 5 \u00b5m, 50 x 2.1 mm HPLC Column<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/27854\">View Product<\/a>\n                                                    <\/div>\n                    <\/div>\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/9304A1E\">                                Catalog No. 9304A1E                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">Raptor C18, 2.7 \u00b5m, 100 x 3.0 mm HPLC Column<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/9304A1E\">View Product<\/a>\n                                                    <\/div>\n                    <\/div>\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/9311A52\">                                Catalog No. 9311A52                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">Raptor Polar X, 2.7 \u00b5m, 50 x 2.1 mm LC Column<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/9311A52\">View Product<\/a>\n                                                    <\/div>\n                    <\/div>\n                            <\/div>\n        <\/div>\n        \n\n\n<div class=\"wp-block-group hub-fullwidth-container has-theme-palette-9-background-color has-background\" style=\"margin-top:0px;margin-bottom:0px;padding-top:var(--wp--preset--spacing--50);padding-bottom:var(--wp--preset--spacing--50)\"><div class=\"wp-block-group__inner-container is-layout-constrained wp-block-group-is-layout-constrained\"><style>.kadence-column99903_3b9a6b-09 > .kt-inside-inner-col{padding-top:45px;padding-right:0px;padding-bottom:45px;padding-left:0px;}.kadence-column99903_3b9a6b-09 > .kt-inside-inner-col{column-gap:var(--global-kb-gap-sm, 1rem);}.kadence-column99903_3b9a6b-09 > .kt-inside-inner-col{flex-direction:column;}.kadence-column99903_3b9a6b-09 > .kt-inside-inner-col > .aligncenter{width:100%;}.kadence-column99903_3b9a6b-09 > .kt-inside-inner-col:before{opacity:0.3;}.kadence-column99903_3b9a6b-09{position:relative;}@media all and (max-width: 1024px){.kadence-column99903_3b9a6b-09 > .kt-inside-inner-col{flex-direction:column;justify-content:center;}}@media all and (max-width: 767px){.kadence-column99903_3b9a6b-09 > .kt-inside-inner-col{flex-direction:column;justify-content:center;}}<\/style>\n<div class=\"wp-block-kadence-column kadence-column99903_3b9a6b-09 hub-guttered-centered-container\"><div class=\"kt-inside-inner-col\"><style>.kb-row-layout-id99903_b5fa4b-ad > .kt-row-column-wrap{align-content:start;}:where(.kb-row-layout-id99903_b5fa4b-ad > .kt-row-column-wrap) > .wp-block-kadence-column{justify-content:start;}.kb-row-layout-id99903_b5fa4b-ad > .kt-row-column-wrap{column-gap:var(--global-kb-gap-md, 2rem);row-gap:var(--global-kb-gap-md, 2rem);padding-top:20px;padding-right:20px;padding-bottom:20px;padding-left:20px;grid-template-columns:minmax(0, 1fr);}.kb-row-layout-id99903_b5fa4b-ad{border-top-left-radius:16px;border-top-right-radius:16px;border-bottom-right-radius:16px;border-bottom-left-radius:16px;overflow:clip;isolation:isolate;}.kb-row-layout-id99903_b5fa4b-ad > .kt-row-layout-overlay{border-top-left-radius:16px;border-top-right-radius:16px;border-bottom-right-radius:16px;border-bottom-left-radius:16px;}.kb-row-layout-id99903_b5fa4b-ad > .kt-row-layout-overlay{opacity:0.30;}@media all and (max-width: 1024px){.kb-row-layout-id99903_b5fa4b-ad > .kt-row-column-wrap{grid-template-columns:minmax(0, 1fr);}}@media all and (max-width: 767px){.kb-row-layout-id99903_b5fa4b-ad > .kt-row-column-wrap{grid-template-columns:minmax(0, 1fr);}}<\/style><div class=\"kb-row-layout-wrap kb-row-layout-id99903_b5fa4b-ad alignnone wp-block-kadence-rowlayout\"><div class=\"kt-row-column-wrap kt-has-1-columns kt-row-layout-equal kt-tab-layout-inherit kt-mobile-layout-row kt-row-valign-top\">\n<style>.kadence-column99903_2824bd-5d > .kt-inside-inner-col{column-gap:var(--global-kb-gap-sm, 1rem);}.kadence-column99903_2824bd-5d > .kt-inside-inner-col{flex-direction:column;}.kadence-column99903_2824bd-5d > .kt-inside-inner-col > .aligncenter{width:100%;}.kadence-column99903_2824bd-5d > .kt-inside-inner-col:before{opacity:0.3;}.kadence-column99903_2824bd-5d{position:relative;}@media all and (max-width: 1024px){.kadence-column99903_2824bd-5d > .kt-inside-inner-col{flex-direction:column;justify-content:center;}}@media all and (max-width: 767px){.kadence-column99903_2824bd-5d > .kt-inside-inner-col{flex-direction:column;justify-content:center;}}<\/style>\n<div class=\"wp-block-kadence-column kadence-column99903_2824bd-5d\"><div class=\"kt-inside-inner-col\"><div class=\"wp-block-image\">\n<figure class=\"aligncenter size-large is-resized\"><img loading=\"lazy\" decoding=\"async\" width=\"1024\" height=\"1024\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/figure-gnlp4243-pfas-hub-contact-an-expert-leaf-graphic-1024x1024.png\" alt=\"Icon depicting customer service person.\" class=\"wp-image-95744\" style=\"width:100px\" title=\"-\" srcset=\"https:\/\/discover.restek.com\/wp-content\/uploads\/figure-gnlp4243-pfas-hub-contact-an-expert-leaf-graphic-1024x1024.png 1024w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure-gnlp4243-pfas-hub-contact-an-expert-leaf-graphic-300x300.png 300w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure-gnlp4243-pfas-hub-contact-an-expert-leaf-graphic-150x150.png 150w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure-gnlp4243-pfas-hub-contact-an-expert-leaf-graphic-768x768.png 768w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure-gnlp4243-pfas-hub-contact-an-expert-leaf-graphic.png 1200w\" sizes=\"auto, (max-width: 1024px) 100vw, 1024px\" \/><\/figure>\n<\/div>\n\n<style>.wp-block-kadence-advancedheading.kt-adv-heading99903_df06fd-cb, .wp-block-kadence-advancedheading.kt-adv-heading99903_df06fd-cb[data-kb-block=\"kb-adv-heading99903_df06fd-cb\"]{margin-top:20px;text-align:center;font-size:48px;line-height:55px;font-weight:400;font-style:normal;font-family:'Open Sans';}.wp-block-kadence-advancedheading.kt-adv-heading99903_df06fd-cb, .wp-block-kadence-advancedheading.kt-adv-heading99903_df06fd-cb[data-kb-block=\"kb-adv-heading99903_df06fd-cb\"] .kb-adv-text-inner{background-image:linear-gradient(270deg,rgb(4,137,214) 0%,rgb(1,94,155) 100%);background-clip:text;-webkit-box-decoration-break:clone;box-decoration-break:clone;-webkit-background-clip:text;-webkit-text-fill-color:transparent;}.wp-block-kadence-advancedheading.kt-adv-heading99903_df06fd-cb mark.kt-highlight, .wp-block-kadence-advancedheading.kt-adv-heading99903_df06fd-cb[data-kb-block=\"kb-adv-heading99903_df06fd-cb\"] mark.kt-highlight{font-style:normal;-webkit-text-fill-color:initial !important;-webkit-background-clip:initial !important;background-clip:initial !important;color:#f76a0c;-webkit-box-decoration-break:clone;box-decoration-break:clone;padding-top:0px;padding-right:0px;padding-bottom:0px;padding-left:0px;}.wp-block-kadence-advancedheading.kt-adv-heading99903_df06fd-cb img.kb-inline-image, .wp-block-kadence-advancedheading.kt-adv-heading99903_df06fd-cb[data-kb-block=\"kb-adv-heading99903_df06fd-cb\"] img.kb-inline-image{width:150px;vertical-align:baseline;}@media all and (max-width: 1024px){.wp-block-kadence-advancedheading.kt-adv-heading99903_df06fd-cb, .wp-block-kadence-advancedheading.kt-adv-heading99903_df06fd-cb[data-kb-block=\"kb-adv-heading99903_df06fd-cb\"]{font-size:40px;}}@media all and (max-width: 767px){.wp-block-kadence-advancedheading.kt-adv-heading99903_df06fd-cb, .wp-block-kadence-advancedheading.kt-adv-heading99903_df06fd-cb[data-kb-block=\"kb-adv-heading99903_df06fd-cb\"]{font-size:36px;}}<\/style><h2 class=\"kt-adv-heading99903_df06fd-cb wp-block-kadence-advancedheading\" data-kb-block=\"kb-adv-heading99903_df06fd-cb\"><span class=\"kb-adv-text-inner\">\u00bfNecesitas Ayuda con Tu An\u00e1lisis de PFAS?<\/span><\/h2>\n\n\n<p class=\"has-text-align-center has-17-font-size has-theme-palette-3-color has-text-color has-link-color wp-elements-1 wp-block-paragraph\" style=\"padding-right:var(--wp--preset--spacing--50);padding-left:var(--wp--preset--spacing--50);font-size:23px\">Desde la selecci\u00f3n del m\u00e9todo hasta las recomendaciones de productos, nuestros expertos en PFAS est\u00e1n aqu\u00ed para ayudarte a gestionar flujos de trabajo complejos y a obtener resultados precisos y fiables.<\/p>\n\n\n<style>.wp-block-kadence-advancedbtn.kb-btns99903_bac608-3f{gap:var(--global-kb-gap-xs, 0.5rem );justify-content:center;align-items:center;}.kt-btns99903_bac608-3f .kt-button{font-weight:normal;font-style:normal;}.kt-btns99903_bac608-3f .kt-btn-wrap-0{margin-right:5px;}.wp-block-kadence-advancedbtn.kt-btns99903_bac608-3f .kt-btn-wrap-0 .kt-button{color:#555555;border-color:#555555;}.wp-block-kadence-advancedbtn.kt-btns99903_bac608-3f .kt-btn-wrap-0 .kt-button:hover, .wp-block-kadence-advancedbtn.kt-btns99903_bac608-3f .kt-btn-wrap-0 .kt-button:focus{color:#ffffff;border-color:#444444;}.wp-block-kadence-advancedbtn.kt-btns99903_bac608-3f .kt-btn-wrap-0 .kt-button::before{display:none;}.wp-block-kadence-advancedbtn.kt-btns99903_bac608-3f .kt-btn-wrap-0 .kt-button:hover, .wp-block-kadence-advancedbtn.kt-btns99903_bac608-3f .kt-btn-wrap-0 .kt-button:focus{background:#444444;}<\/style>\n<div class=\"wp-block-kadence-advancedbtn kb-buttons-wrap kb-btns99903_bac608-3f\"><style>ul.menu .wp-block-kadence-advancedbtn .kb-btn99903_980c0c-c6.kb-button{width:initial;}.wp-block-kadence-advancedbtn .kb-btn99903_980c0c-c6.kb-button{color:var(--global-palette9, #ffffff);background:#61b522;font-size:var(--global-kb-font-size-md, 1.25rem);border-top-left-radius:10px;border-top-right-radius:10px;border-bottom-right-radius:10px;border-bottom-left-radius:10px;border-top:2px solid #61b522;border-right:2px solid #61b522;border-bottom:2px solid #61b522;border-left:2px solid #61b522;}.kb-btn99903_980c0c-c6.kb-button .kb-svg-icon-wrap{color:var(--global-palette9, #ffffff);}.kb-btn99903_980c0c-c6.kb-button:hover .kb-svg-icon-wrap, .kb-btn99903_980c0c-c6.kb-button:focus .kb-svg-icon-wrap{color:#61b522;}.wp-block-kadence-advancedbtn .kb-btn99903_980c0c-c6.kb-button:hover, .wp-block-kadence-advancedbtn .kb-btn99903_980c0c-c6.kb-button:focus{color:#61b522;background:var(--global-palette9, #ffffff);border-top:2px solid #61b522;border-right:2px solid #61b522;border-bottom:2px solid #61b522;border-left:2px solid #61b522;}@media all and (max-width: 1024px){.wp-block-kadence-advancedbtn .kb-btn99903_980c0c-c6.kb-button{border-top:2px solid #61b522;border-right:2px solid #61b522;border-bottom:2px solid #61b522;border-left:2px solid #61b522;}}@media all and (max-width: 1024px){.wp-block-kadence-advancedbtn .kb-btn99903_980c0c-c6.kb-button:hover, .wp-block-kadence-advancedbtn .kb-btn99903_980c0c-c6.kb-button:focus{border-top:2px solid #61b522;border-right:2px solid #61b522;border-bottom:2px solid #61b522;border-left:2px solid #61b522;}}@media all and (max-width: 767px){.wp-block-kadence-advancedbtn .kb-btn99903_980c0c-c6.kb-button{border-top:2px solid #61b522;border-right:2px solid #61b522;border-bottom:2px solid #61b522;border-left:2px solid #61b522;}.wp-block-kadence-advancedbtn .kb-btn99903_980c0c-c6.kb-button:hover, .wp-block-kadence-advancedbtn .kb-btn99903_980c0c-c6.kb-button:focus{border-top:2px solid #61b522;border-right:2px solid #61b522;border-bottom:2px solid #61b522;border-left:2px solid #61b522;}}<\/style><a class=\"kb-button kt-button button kb-btn99903_980c0c-c6 kt-btn-size-standard kt-btn-width-type-auto kb-btn-global-outline kt-btn-has-text-true kt-btn-has-svg-true wp-block-kadence-singlebtn kb-preset--outline\" href=\"https:\/\/explore.restek.com\/PFAS-ES\" target=\"_blank\" rel=\"noopener\"><span class=\"kt-btn-inner-text\">Contacta con un experto en PFAS<\/span><span class=\"kb-svg-icon-wrap kb-svg-icon-fe_arrowRight kt-btn-icon-side-right\"><svg viewBox=\"0 0 24 24\"  fill=\"none\" stroke=\"currentColor\" stroke-width=\"2\" stroke-linecap=\"round\" stroke-linejoin=\"round\" xmlns=\"http:\/\/www.w3.org\/2000\/svg\"  aria-hidden=\"true\"><line x1=\"5\" y1=\"12\" x2=\"19\" y2=\"12\"\/><polyline points=\"12 5 19 12 12 19\"\/><\/svg><\/span><\/a><\/div>\n<\/div><\/div>\n\n<\/div><\/div><\/div><\/div>\n<\/div><\/div>\n","protected":false},"excerpt":{"rendered":"<p>As PFAS testing expands into shorter chain compounds, new methods are being developed. This article explores the impact of phase, dimensions, and particle types and provides options that ensure fast, accurate results.<\/p>\n","protected":false},"author":11,"featured_media":6601,"comment_status":"closed","ping_status":"open","sticky":false,"template":"","format":"standard","meta":{"_acf_changed":false,"_kad_blocks_custom_css":"","_kad_blocks_head_custom_js":"","_kad_blocks_body_custom_js":"","_kad_blocks_footer_custom_js":"","_kadence_starter_templates_imported_post":false,"_kad_post_transparent":"","_kad_post_title":"","_kad_post_layout":"","_kad_post_sidebar_id":"","_kad_post_content_style":"","_kad_post_vertical_padding":"","_kad_post_feature":"","_kad_post_feature_position":"","_kad_post_header":false,"_kad_post_footer":false,"footnotes":"","_ppma_block_editor_authors":""},"categories":[787],"tags":[2782],"industries-application":[2230,2234],"post-badge":[],"resource-type":[567],"product-library":[2499,2481],"resource-technique":[2326,2328,2363],"ppma_author":[414],"class_list":["post-43417","post","type-post","status-publish","format-standard","has-post-thumbnail","hentry","category-artikel","tag-pfas","industries-application-umweltbereich","industries-application-pfas-analyse","resource-type-selection-guides","product-library-lc-saulen","product-library-flussigchromatographie-produkte","resource-technique-flussigchromatografie","resource-technique-massenspektrometrie-ms","resource-technique-ms-ms-de"],"acf":[],"taxonomy_info":{"category":[{"value":787,"label":"Artikel"}],"post_tag":[{"value":2782,"label":"PFAS"}],"industries-application":[{"value":2230,"label":"Umweltbereich"},{"value":2234,"label":"PFAS Analyse"}],"resource-type":[{"value":567,"label":"Selection Guides"}],"product-library":[{"value":2499,"label":"LC S\u00e4ulen"},{"value":2481,"label":"Fl\u00fcssigchromatographie Produkte"}],"resource-technique":[{"value":2326,"label":"Fl\u00fcssigchromatografie"},{"value":2328,"label":"Massenspektrometrie (MS)"},{"value":2363,"label":"MS\/MS"}]},"featured_image_src_large":["https:\/\/discover.restek.com\/wp-content\/uploads\/feature-EVAR3069-1024x536.jpg",1024,536,true],"author_info":{"display_name":"Restek Corporation","author_link":"https:\/\/discover.restek.com\/es\/author\/restek-corporation\/"},"comment_info":0,"category_info":[{"term_id":787,"name":"Artikel","slug":"artikel","term_group":0,"term_taxonomy_id":787,"taxonomy":"category","description":"","parent":0,"count":510,"filter":"raw","cat_ID":787,"category_count":510,"category_description":"","cat_name":"Artikel","category_nicename":"artikel","category_parent":0}],"tag_info":[{"term_id":2782,"name":"PFAS","slug":"pfas","term_group":0,"term_taxonomy_id":2782,"taxonomy":"post_tag","description":"","parent":0,"count":124,"filter":"raw"}],"authors":[{"term_id":414,"user_id":11,"is_guest":0,"slug":"restek-corporation","display_name":"Restek Corporation","avatar_url":{"url":"https:\/\/discover.restek.com\/wp-content\/uploads\/Restek_Favicon_300x300.jpg","url2x":"https:\/\/discover.restek.com\/wp-content\/uploads\/Restek_Favicon_300x300.jpg"},"author_category":"1","first_name":"Restek","last_name":"Corporation","user_url":"https:\/\/www.linkedin.com\/company\/restek-corporation\/posts\/?feedView=all","job_title":"Chromatography Products and Solutions","description":"Restek is a leading provider of chromatography columns, accessories, and certified reference materials. Trust Restek for reliable, high-quality analytical solutions."}],"_links":{"self":[{"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/posts\/43417","targetHints":{"allow":["GET"]}}],"collection":[{"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/posts"}],"about":[{"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/types\/post"}],"author":[{"embeddable":true,"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/users\/11"}],"replies":[{"embeddable":true,"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/comments?post=43417"}],"version-history":[{"count":14,"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/posts\/43417\/revisions"}],"predecessor-version":[{"id":100713,"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/posts\/43417\/revisions\/100713"}],"wp:featuredmedia":[{"embeddable":true,"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/media\/6601"}],"wp:attachment":[{"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/media?parent=43417"}],"wp:term":[{"taxonomy":"category","embeddable":true,"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/categories?post=43417"},{"taxonomy":"post_tag","embeddable":true,"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/tags?post=43417"},{"taxonomy":"industries-application","embeddable":true,"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/industries-application?post=43417"},{"taxonomy":"post-badge","embeddable":true,"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/post-badge?post=43417"},{"taxonomy":"resource-type","embeddable":true,"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/resource-type?post=43417"},{"taxonomy":"product-library","embeddable":true,"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/product-library?post=43417"},{"taxonomy":"resource-technique","embeddable":true,"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/resource-technique?post=43417"},{"taxonomy":"author","embeddable":true,"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/ppma_author?post=43417"}],"curies":[{"name":"wp","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}