{"id":51716,"date":"2013-05-09T14:30:00","date_gmt":"2013-05-09T14:30:00","guid":{"rendered":"https:\/\/discover.restek.com\/uncategorized\/how-to-connect-18-inch-tubing-to-a-restek-gas-regulator\/"},"modified":"2026-01-20T16:37:36","modified_gmt":"2026-01-20T16:37:36","slug":"how-to-connect-18-inch-tubing-to-a-restek-gas-regulator","status":"publish","type":"post","link":"https:\/\/discover.restek.com\/es\/blogs\/gnbl4707\/how-to-connect-18-inch-tubing-to-a-restek-gas-regulator","title":{"rendered":"How to connect 1\/8 inch tubing to a Restek gas regulator"},"content":{"rendered":"\n<p class=\"wp-block-paragraph\">Because the majority of gas regulators which Restek sells are designed to connect to&nbsp;1\/4 inch tubing, we (tech service) frequently get asked which additional fittings are needed to connect to 1\/8 inch tubing. This post will summarize which additional fittings are needed to accomplish this task.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">First, let me explain which gas regulators I will be discussing in this post. They can be manufactured of either chrome-plated brass or stainless steel, they can be single or dual stage, and they will have a specific CGA (Compressed Gas Association) fitting on one end. The other end will be a&nbsp;1\/4 inch MNPT (male National Pipe Thread) fitting, which screws into the shut-off valve on the regulator. The outlet end of this fitting is&nbsp;1\/4 inch compression, which can connect to&nbsp;1\/4 inch OD (outside diameter) tubing. This fitting is shown by the&nbsp;red&nbsp;arrow in the photo below.<\/p>\n\n\n<div class=\"wp-block-image\">\n<figure class=\"aligncenter size-full\"><img loading=\"lazy\" decoding=\"async\" width=\"861\" height=\"521\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-how-to-connect-18-inch-tubing-to-a-restek-gas-regulator-01.jpg\" alt=\"gas regulator\" class=\"wp-image-15190\" title=\"-\" srcset=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-how-to-connect-18-inch-tubing-to-a-restek-gas-regulator-01.jpg 861w, https:\/\/discover.restek.com\/wp-content\/uploads\/blog-how-to-connect-18-inch-tubing-to-a-restek-gas-regulator-01-300x182.jpg 300w, https:\/\/discover.restek.com\/wp-content\/uploads\/blog-how-to-connect-18-inch-tubing-to-a-restek-gas-regulator-01-768x465.jpg 768w\" sizes=\"auto, (max-width: 861px) 100vw, 861px\" \/><\/figure>\n<\/div>\n\n\n<p class=\"wp-block-paragraph\">In order to use 1\/8 inch tubing with these gas regulators, you have two options:<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">1. Leave the end-fitting (shown by the&nbsp;red&nbsp;arrow) attached to the regulator and install a tube-end reducer to this end fitting. If you have a chrome-plated brass regulator, use Restek catalog number&nbsp;<a href=\"https:\/\/www.restek.com\/p\/23129\" target=\"_blank\" rel=\"noopener\">23129<\/a>. If you have a stainless steel regulator, use&nbsp;<a href=\"https:\/\/www.restek.com\/p\/23179\" target=\"_blank\" rel=\"noopener\">23179<\/a>.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">2. Because this extra fitting can be an additional source of leaks, some customers would prefer to remove the current end fitting and replace it with one that has a 1\/8 inch compression outlet. The catalog numbers for these male connectors are&nbsp;<a href=\"https:\/\/www.restek.com\/p\/23136\" target=\"_blank\" rel=\"noopener\">23136<\/a>&nbsp;for brass and&nbsp;<a href=\"https:\/\/www.restek.com\/p\/23186\" target=\"_blank\" rel=\"noopener\">23186<\/a> for stainless steel. When replacing the end fitting, don\u2019t forget to use PTFE tape.&nbsp;Restek PTFE tape catalog number&nbsp;<a href=\"https:\/\/www.restek.com\/p\/22486\" target=\"_blank\" rel=\"noopener\">22486<\/a>&nbsp;is commonly used for brass fittings and&nbsp;<a href=\"https:\/\/www.restek.com\/p\/22487\" target=\"_blank\" rel=\"noopener\">22487<\/a>&nbsp;for stainless steel fittings.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Note: If purchasing one of our&nbsp;Silcosteel regulators, don\u2019t forget to purchase&nbsp;treated fittings&nbsp;and&nbsp;treated tubing.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">I hope this post clearly explains which fittings you will need in order to connect one of our gas regulators to 1\/8 inch OD tubing. To read more about fittings, please review my previous post\u00a0<a href=\"https:\/\/discover.restek.com\/blogs\/gnbl4714\/i-need-a-fitting-but-which-one\" data-type=\"link\" data-id=\"https:\/\/discover.restek.com\/blogs\/gnbl4714\/i-need-a-fitting-but-which-one\">I need a fitting, but which one?<\/a>\u00a0\u00a0 If you are looking for the appropriate gas regulator,\u00a0review my post\u00a0Need some help in choosing the correct gas regulator?<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">If you have any questions, give us a call. Thanks for reading.<\/p>\n\n\n\n<div style=\"height:100px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n\n\n        <div class=\"cpb\">\n            <h3 class=\"cpb-heading\">Products Mentioned<\/h3>\n            <hr class=\"cpb-heading-underline\" \/>\n            <div class=\"cpb-list\">\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/23129\">                                Catalog No. 23129                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">Swagelok Fitting, 1\/4&quot; to 1\/8&quot; Tube End Reducer, Brass, 5-pk.<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/23129\">View Product<\/a>\n                                                    <\/div>\n                    <\/div>\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/23179\">                                Catalog No. 23179                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">Swagelok Fitting, 1\/4&quot; to 1\/8&quot; Tube End Reducer, Stainless Steel, 2-pk.<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/23179\">View Product<\/a>\n                                                    <\/div>\n                    <\/div>\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/23136\">                                Catalog No. 23136                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">Swagelok Fitting, 1\/8&quot; to 1\/4&quot; NPT Male Connector, Brass, 10-pk.<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/23136\">View Product<\/a>\n                                                    <\/div>\n                    <\/div>\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/23186\">                                Catalog No. 23186                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">Swagelok Fitting, 1\/8&quot; to 1\/4&quot; NPT Male Connector, Stainless Steel, 2-pk.<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/23186\">View Product<\/a>\n                                                    <\/div>\n                    <\/div>\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/22486\">                                Catalog No. 22486                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">ResTape, Yellow PTFE, 1\/2&quot; x 260&quot;, Thread Tape for Gas Service<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/22486\">View Product<\/a>\n                                                    <\/div>\n                    <\/div>\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/22487\">                                Catalog No. 22487                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">ResTape, Gray PTFE,  1\/2&quot; x 260&quot;, Thread Tape for Stainless Steel<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/22487\">View Product<\/a>\n                                                    <\/div>\n                    <\/div>\n                            <\/div>\n        <\/div>\n        \n\n\n<div style=\"height:100px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n","protected":false},"excerpt":{"rendered":"<p>Because the majority of gas regulators which Restek sells are designed to connect to 1\/4 inch tubing, we (tech service) frequently get asked which additional fittings are needed to connect to 1\/8 inch tubing. This post will summarize which additional&#8230;<\/p>\n","protected":false},"author":5,"featured_media":0,"comment_status":"closed","ping_status":"open","sticky":false,"template":"","format":"standard","meta":{"_acf_changed":false,"_kad_blocks_custom_css":"","_kad_blocks_head_custom_js":"","_kad_blocks_body_custom_js":"","_kad_blocks_footer_custom_js":"","_kadence_starter_templates_imported_post":false,"_kad_post_transparent":"","_kad_post_title":"","_kad_post_layout":"","_kad_post_sidebar_id":"","_kad_post_content_style":"","_kad_post_vertical_padding":"","_kad_post_feature":"","_kad_post_feature_position":"","_kad_post_header":false,"_kad_post_footer":false,"footnotes":"","_ppma_block_editor_authors":""},"categories":[9],"tags":[],"industries-application":[],"post-badge":[],"resource-type":[],"product-library":[2385,2372],"resource-technique":[],"ppma_author":[582],"class_list":["post-51716","post","type-post","status-publish","format-standard","hentry","category-blogs","product-library-ferrules-fittings-connectors","product-library-gas-chromatography-products"],"acf":[],"taxonomy_info":{"category":[{"value":9,"label":"Blogs"}],"product-library":[{"value":2385,"label":"Ferrules, Fittings, Connectors"},{"value":2372,"label":"Gas Chromatography Products"}]},"featured_image_src_large":false,"author_info":{"display_name":"Brenda Seprish","author_link":"https:\/\/discover.restek.com\/es\/author\/brenda-coganrestek-com\/"},"comment_info":0,"category_info":[{"term_id":9,"name":"Blogs","slug":"blogs","term_group":0,"term_taxonomy_id":9,"taxonomy":"category","description":"","parent":0,"count":458,"filter":"raw","cat_ID":9,"category_count":458,"category_description":"","cat_name":"Blogs","category_nicename":"blogs","category_parent":0}],"tag_info":false,"authors":[{"term_id":582,"user_id":0,"is_guest":1,"slug":"alan-sensue","display_name":"Alan Sensue","avatar_url":"https:\/\/secure.gravatar.com\/avatar\/?s=96&d=mm&r=g","author_category":"1","first_name":"Alan","last_name":"Sensue","user_url":"","job_title":"","description":""}],"_links":{"self":[{"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/posts\/51716","targetHints":{"allow":["GET"]}}],"collection":[{"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/posts"}],"about":[{"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/types\/post"}],"author":[{"embeddable":true,"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/users\/5"}],"replies":[{"embeddable":true,"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/comments?post=51716"}],"version-history":[{"count":3,"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/posts\/51716\/revisions"}],"predecessor-version":[{"id":83490,"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/posts\/51716\/revisions\/83490"}],"wp:attachment":[{"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/media?parent=51716"}],"wp:term":[{"taxonomy":"category","embeddable":true,"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/categories?post=51716"},{"taxonomy":"post_tag","embeddable":true,"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/tags?post=51716"},{"taxonomy":"industries-application","embeddable":true,"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/industries-application?post=51716"},{"taxonomy":"post-badge","embeddable":true,"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/post-badge?post=51716"},{"taxonomy":"resource-type","embeddable":true,"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/resource-type?post=51716"},{"taxonomy":"product-library","embeddable":true,"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/product-library?post=51716"},{"taxonomy":"resource-technique","embeddable":true,"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/resource-technique?post=51716"},{"taxonomy":"author","embeddable":true,"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/ppma_author?post=51716"}],"curies":[{"name":"wp","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}