{"id":51727,"date":"2024-08-19T00:00:00","date_gmt":"2024-08-19T00:00:00","guid":{"rendered":"https:\/\/discover.restek.com\/uncategorized\/is-inert-lc-column-hardware-beneficial-for-pesticide-or-mycotoxin-analysis\/"},"modified":"2026-06-17T19:14:52","modified_gmt":"2026-06-17T19:14:52","slug":"is-inert-lc-column-hardware-beneficial-for-pesticide-or-mycotoxin-analysis","status":"publish","type":"post","link":"https:\/\/discover.restek.com\/es\/blogs\/gnbl4720\/is-inert-lc-column-hardware-beneficial-for-pesticide-or-mycotoxin-analysis","title":{"rendered":"Is Inert LC Column Hardware Beneficial for Pesticide or Mycotoxin Analysis?"},"content":{"rendered":"\n<p class=\"wp-block-paragraph\">For some years now, the development departments of high-pressure liquid chromatography companies have been dealing with the question of what influence the surfaces of the instrument and component hardware can have on peak shape and sensitivity in the detection of polar, chelating molecules. Biomolecules in particular, with their complex acid\/base properties, but also compounds such as glyphosate and AMPA, with their organophosphate groups, show that the surface properties of the components in the sample path have a decisive influence on peak shape and sensitivity.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Recently, Restek developed a inert LC column line to address this issue more efficiently and effectively than short-term solutions, such as priming or medronic acid deactivation. Reports from several independent laboratories indicate that these new columns can offer substantial benefits for pesticides and mycotoxins analysis.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">In June 2024, Maria Antonietta Carrera from the Department of Desertification and Geo-ecology, Experimental Station of Arid Zones in Almeria, under the supervision of Amadeo Fernandez-Alba from the European Union Reference Laboratory for Pesticide Residues in Fruit &amp; Vegetables (EURL-FV), published a remarkable paper on Simultaneous analysis of pesticides and mycotoxins in primary processed foods: The case of bee pollen&#8221;, in which she compared a column commonly used for routine analysis of the analytes under consideration with a Raptor Inert ARC-18 column from Restek.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">In her publication she describes \u201cAs reported in Fig. 3, no substantial differences were revealed by this test, since almost the same number of compounds could be detected at a concentration of 1 \u03bcg\/kg with both columns. However, by looking at peaks obtained for about 15 % of analytes at 1 \u03bcg\/kg, it was possible to notice a visible improvement in the peak shape (Fig. 4) when using the Raptor Inert column. This particular feature can be sometimes decisive for the identification of the analytes in real samples.\u201d<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Please find the full publication here: <a href=\"https:\/\/www.cell.com\/heliyon\/fulltext\/S2405-8440(24)09543-4\" target=\"_blank\" rel=\"noreferrer noopener\">https:\/\/www.cell.com\/heliyon\/fulltext\/S2405-8440(24)09543-4<\/a>&nbsp;<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">In addition to influential labs, such as EURL, that assess the application of new analytical developments within their specific fields, method developers in instrument companies must ensure that the performance statement about their devices is not countered by poor chromatography, for which they bear no responsibility.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">LCTech GmbH has made a name for itself in automated sample preparation for the analysis of mycotoxins, PFAS, and other multicomponent analyte panels. Frederik Wuppermann, Product Manager Biotechnology at LCTech, describes his experience with the inert columns in mycotoxin analysis as follows.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">&#8220;The separation of aflatoxin B1, G1, B2, G2, OTA, zearalenone, deoxynivalenol, T2, H-T2, fumonisin B1 and B2 after CrossTOX cleanup in high-throughput mycotoxin analysis was very good. The shorter conditioning times were also very beneficial. I can certainly also imagine the Inert Biphenyl column as an alternative for other applications.&#8221;<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">In addition to these independent reports, Restek has assessed the impact of the new inert LC columns on pesticides and mycotoxins. We selected these groups of analytes because they include a variety of diverse compound chemistries and are often analyzed together in multi-class screening analyses.&nbsp;In such screening analyses, it is very important to ensure reliable identification and quantification of the individual components at trace levels (low LOQs), and an inert LC column can help with this by providing proper peak shape and good sensitivity.&nbsp;For both groups, we selected an appropriate stationary phase, analyzed the target compounds on both conventional LC columns and the new inert LC columns, and compared the chromatographic results. Here are our findings.<\/p>\n\n\n\n<h2 class=\"wp-block-heading\">Pesticides<\/h2>\n\n\n\n<p class=\"wp-block-paragraph\">For this experiment, we compared a Raptor Inert ARC-18 column to a standard Raptor ARC-18 column. Pesticides analysis can be challenging because these panels typically contain a wide variety of compound chemistries. Phosphorylated, acidic, polar compounds, and\/or metal chelating species, such as organophosphate pesticides, can react to the metal surfaces inside standard, stainless-steel analytical columns, which can degrade chromatographic performance. In contrast, the deactivated hardware in Restek\u2019s inert LC columns significantly improved chromatographic results (Figures 1-2, Table I).<\/p>\n\n\n<div class=\"wp-block-custom-chromatogram-article-top\"><div class=\"chromatogram-article-placeholder\"><div class=\"figure-heading\"><strong>Figure 1:<\/strong> Analysis of pesticides on a Raptor Inert ARC-18 column without any preconditioning.<\/div><div class='chromatogram-article-inner-full'><div class=\"chromatogram-article-inner\">\n<div class=\"wp-block-custom-chromatogram-article\"><div class=\"wp-block-custom-chromatogram-article\"><div class=\"chromatogram-image wide-image\"><img decoding=\"async\" src=\"https:\/\/ez.restek.com\/images\/cgram\/lc_ev0596.png\" alt=\"Pesticides on Raptor Inert ARC-18\" title=\"-\"><\/div><p class=\"article-id\" style=\"text-align:center\"> LC_EV0596<\/p><div class=\"chromatogram-peaks\"><h4>Peaks<\/h4><table class=\"peaks col-12\">\n<thead><tr><th><\/th><th style=\"text-align: left;width: 75px\">Peaks<\/th><th style=\"text-align: center;width: 75px\">t<sub>R<\/sub> (min)<\/th><th style=\"text-align: center;width: 75px\">Precursor Ion<\/th><th style=\"text-align: center;width: 75px\">Product Ion 1<\/th><th style=\"text-align: center;width: 75px\">Product Ion 2<\/th><th style=\"text-align: center;width: 75px\">Peak Area<\/th><th style=\"text-align: center;width: 75px\">Peak Height<\/th><\/tr><\/thead>\n<tbody><tr><td class=\"num\">1.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Methamidophos\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/10265-92-6\/Methamidophos\" target=\"_blank\" rel=\"noopener\">Methamidophos<\/a><\/td><td class=\"oth\">1.33<\/td><td class=\"oth\">142.0<\/td><td class=\"oth\">94.0<\/td><td class=\"oth\">125.1<\/td><td class=\"oth\">428941<\/td><td class=\"oth\">105189<\/td><\/tr>\n<tr><td class=\"num\">2.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Acephate\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/30560-19-1\/Acephate\" target=\"_blank\" rel=\"noopener\">Acephate<\/a><\/td><td class=\"oth\">1.55<\/td><td class=\"oth\">184.0<\/td><td class=\"oth\">143.0<\/td><td class=\"oth\">48.9<\/td><td class=\"oth\">300642<\/td><td class=\"oth\">104729<\/td><\/tr>\n<tr><td class=\"num\">3.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Omethoate\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/1113-02-6\/Omethoate\" target=\"_blank\" rel=\"noopener\">Omethoate<\/a><\/td><td class=\"oth\">1.72<\/td><td class=\"oth\">214.0<\/td><td class=\"oth\">125.0<\/td><td class=\"oth\">182.9<\/td><td class=\"oth\">892008<\/td><td class=\"oth\">337690<\/td><\/tr>\n<tr><td class=\"num\">4.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Monocrotophos\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/6923-22-4\/Monocrotophos\" target=\"_blank\" rel=\"noopener\">Monocrotophos<\/a><\/td><td class=\"oth\">2.21<\/td><td class=\"oth\">224.1<\/td><td class=\"oth\">127.0<\/td><td class=\"oth\">193.1<\/td><td class=\"oth\">215810<\/td><td class=\"oth\">78425<\/td><\/tr>\n<tr><td class=\"num\">5.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Dicrotophos\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/141-66-2\/Dicrotophos\" target=\"_blank\" rel=\"noopener\">Dicrotophos<\/a><\/td><td class=\"oth\">2.35<\/td><td class=\"oth\">238.1<\/td><td class=\"oth\">112.1<\/td><td class=\"oth\">72.0<\/td><td class=\"oth\">404916<\/td><td class=\"oth\">159292<\/td><\/tr>\n<tr><td class=\"num\">6.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Dimethoate\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/60-51-5\/Dimethoate\" target=\"_blank\" rel=\"noopener\">Dimethoate<\/a><\/td><td class=\"oth\">2.52<\/td><td class=\"oth\">230.0<\/td><td class=\"oth\">125.0<\/td><td class=\"oth\">199.0<\/td><td class=\"oth\">807805<\/td><td class=\"oth\">342939<\/td><\/tr>\n<tr><td class=\"num\">7.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Trichlorfon\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/52-68-6\/Trichlorfon\" target=\"_blank\" rel=\"noopener\">Trichlorfon<\/a><\/td><td class=\"oth\">2.53<\/td><td class=\"oth\">257.0<\/td><td class=\"oth\">108.9<\/td><td class=\"oth\">220.8<\/td><td class=\"oth\">173942<\/td><td class=\"oth\">63266<\/td><\/tr>\n<tr><td class=\"num\">8.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Vamidothion\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/2275-23-2\/Vamidothion\" target=\"_blank\" rel=\"noopener\">Vamidothion<\/a><\/td><td class=\"oth\">2.54<\/td><td class=\"oth\">288.0<\/td><td class=\"oth\">146.0<\/td><td class=\"oth\">118.0<\/td><td class=\"oth\">1333829<\/td><td class=\"oth\">547308<\/td><\/tr>\n<tr><td class=\"num\">9.<\/td><td class=\"cmpd\">Mevinphos isomer 1<\/td><td class=\"oth\">2.55<\/td><td class=\"oth\">241.9<\/td><td class=\"oth\">126.9<\/td><td class=\"oth\">192.9<\/td><td class=\"oth\">311274<\/td><td class=\"oth\">129961<\/td><\/tr>\n<tr><td class=\"num\">10.<\/td><td class=\"cmpd\">Mevinphos isomer 2<\/td><td class=\"oth\">2.76<\/td><td class=\"oth\">241.9<\/td><td class=\"oth\">126.9<\/td><td class=\"oth\">192.9<\/td><td class=\"oth\">74030<\/td><td class=\"oth\">29802<\/td><\/tr>\n<tr><td class=\"num\">11.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Carbaryl\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/63-25-2\/Carbaryl\" target=\"_blank\" rel=\"noopener\">Carbaryl<\/a><\/td><td class=\"oth\">3.18<\/td><td class=\"oth\">202.1<\/td><td class=\"oth\">145.0<\/td><td class=\"oth\">127.0<\/td><td class=\"oth\">39671<\/td><td class=\"oth\">11924<\/td><\/tr>\n<tr><td class=\"num\">12.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Isocarbophos\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/24353-61-5\/Isocarbophos\" target=\"_blank\" rel=\"noopener\">Isocarbophos<\/a><\/td><td class=\"oth\">3.52<\/td><td class=\"oth\">291.1<\/td><td class=\"oth\">231.1<\/td><td class=\"oth\">121.1<\/td><td class=\"oth\">33294<\/td><td class=\"oth\">11941<\/td><\/tr>\n<tr><td class=\"num\">13.<\/td><td class=\"cmpd\">Dimethomorph isomer 1<\/td><td class=\"oth\">3.96<\/td><td class=\"oth\">388.2<\/td><td class=\"oth\">300.9<\/td><td class=\"oth\">165.1<\/td><td class=\"oth\">511766<\/td><td class=\"oth\">172977<\/td><\/tr>\n<tr><td class=\"num\">14.<\/td><td class=\"cmpd\">Dimethomorph isomer 2<\/td><td class=\"oth\">4.13<\/td><td class=\"oth\">388.2<\/td><td class=\"oth\">300.9<\/td><td class=\"oth\">165.1<\/td><td class=\"oth\">877031<\/td><td class=\"oth\">328826<\/td><\/tr>\n<tr><td class=\"num\">15.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Temephos\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/3383-96-8\/Temephos\" target=\"_blank\" rel=\"noopener\">Temephos<\/a><\/td><td class=\"oth\">5.70<\/td><td class=\"oth\">467.1<\/td><td class=\"oth\">124.9<\/td><td class=\"oth\">418.9<\/td><td class=\"oth\">164310<\/td><td class=\"oth\">64751<\/td><\/tr>\n<\/tbody><\/table><\/div><div class=\"chromatogram-conditions\"><h4>Conditions<\/h4><div class=\"conditions-container container-fluid\"><div class=\"row\"><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Column<\/th><td>Raptor Inert ARC-18   (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/9314A12-T\" rel=\"noopener\">cat.# 9314A12-T<\/a>)<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Dimensions:<\/th><td>100 mm x 2.1 mm ID<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Particle Size:<\/th><td>2.7 \u00b5m<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Pore Size:<\/th><td>90 \u00c5<\/td><\/tr><tr><td><\/td><tr><th class=\"sub conditions_header\" scope=\"row\">Temp.:<\/th><td>50 \u00b0C<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Standard\/Sample<\/th><td>LC multiresidue pesticide standard #1 (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/31972\" rel=\"noopener\">cat.# 31972<\/a>)<\/td><\/tr>\n<tr><th class=\"sub conditions_header\" scope=\"row\">Diluent:<\/th><td>Water, 0.1% formic acid<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Conc.:<\/th><td>1 ng\/mL<\/td><\/tr><td><\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Inj. Vol.:<\/th><td>5 \u00b5L <\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Mobile Phase<\/th><td><\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">A:<\/th><td>Water, 2 mM ammonium formate, 0.1% formic acid <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">B:<\/th><td>Methanol, 2 mM ammonium formate, 0.1% formic acid <\/td><\/tr><tr><td><\/td><td><table class=\"cgram_ramp\"><thead><tr><th>Time (min)<\/th><th>Flow (mL\/min)<\/th><th>%A<\/th><th>%B<\/th><\/tr><\/thead><tbody><tr><td>0.00<\/td><td>0.4<\/td><td>95<\/td><td>5<\/td><\/tr><tr><td>2.00<\/td><td>0.4<\/td><td>40<\/td><td>60<\/td><\/tr><tr><td>4.00<\/td><td>0.4<\/td><td>25<\/td><td>75<\/td><\/tr><tr><td>6.00<\/td><td>0.4<\/td><td>0<\/td><td>100<\/td><\/tr><tr><td>7.50<\/td><td>0.4<\/td><td>0<\/td><td>100<\/td><\/tr><tr><td>7.51<\/td><td>0.4<\/td><td>95<\/td><td>5<\/td><\/tr><tr><td>9.00<\/td><td>0.4<\/td><td>95<\/td><td>5<\/td><\/tr><\/tbody><\/table><\/td><\/tr><\/table><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Detector<\/th><td>Shimadzu LCMS-8060<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Ion Mode:<\/th><td>ESI+ <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Mode:<\/th><td>MRM <\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Instrument<\/th><td>Shimadzu Nexera X2<\/td><\/tr><\/table><\/div><\/div><\/div><div class=\"chromatogram-related-skus\"><h4>Related SKUs<\/h4><ul><li>31972<\/li><li>9314A12-T<\/li><\/ul><\/div><div class=\"chromatogram-pdf-link\"><a href=\"https:\/\/ez.restek.com\/images\/cgram\/lc_ev0596.pdf\" target=\"_blank\" rel=\"noopener noreferrer\"><svg xmlns=\"http:\/\/www.w3.org\/2000\/svg\" width=\"18\" height=\"18\" viewBox=\"0 0 18 18\"><g data-name=\"Group 2996\"><path data-name=\"Rectangle 1246\" d=\"M0 0h18v18H0z\" style=\"fill: none;\"><\/path><\/g><g data-name=\"Group 2997\"><path data-name=\"Path 729\" d=\"M13.412 11.4v2.017H5.345V11.4H4v2.017a1.349 1.349 0 0 0 1.345 1.345h8.068a1.349 1.349 0 0 0 1.345-1.345V11.4zm-.672-2.694-.948-.948-1.741 1.735V4H8.706v5.493L6.965 7.758l-.948.948 3.361 3.361z\" transform=\"translate(-.437 -.414)\" style=\"fill: rgb(13, 123, 196);\"><\/path><\/g><\/svg>Download PDF<\/a><\/div><\/div><\/div>\n<\/div><\/div><\/div>\n\n\n<div style=\"height:25px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n\n\n<div class=\"wp-block-custom-chromatogram-article-top\"><div class=\"chromatogram-article-placeholder\"><div class=\"figure-heading\"><strong>Figure 2:<\/strong> Compared to standard LC columns, Raptor Inert LC columns provide increased peak area and height for pesticides.<\/div><div class='chromatogram-article-inner-full'><div class=\"chromatogram-article-inner\">\n<div class=\"wp-block-custom-chromatogram-article\"><div class=\"wp-block-custom-chromatogram-article\"><div class=\"chromatogram-image regular-image\"><img decoding=\"async\" src=\"https:\/\/ez.restek.com\/images\/cgram\/lc_ev0591.png\" alt=\"Comparison of Pesticides on Raptor Inert ARC-18 and Raptor ARC-18\" title=\"-\"><\/div><p class=\"article-id\" style=\"text-align:center\"> LC_EV0591<\/p><div class=\"chromatogram-related-skus\"><h4>Related SKUs<\/h4><ul><li>31972<\/li><li>9314A12<\/li><\/ul><\/div><div class=\"chromatogram-pdf-link\"><a href=\"https:\/\/ez.restek.com\/images\/cgram\/lc_ev0591.pdf\" target=\"_blank\" rel=\"noopener noreferrer\"><svg xmlns=\"http:\/\/www.w3.org\/2000\/svg\" width=\"18\" height=\"18\" viewBox=\"0 0 18 18\"><g data-name=\"Group 2996\"><path data-name=\"Rectangle 1246\" d=\"M0 0h18v18H0z\" style=\"fill: none;\"><\/path><\/g><g data-name=\"Group 2997\"><path data-name=\"Path 729\" d=\"M13.412 11.4v2.017H5.345V11.4H4v2.017a1.349 1.349 0 0 0 1.345 1.345h8.068a1.349 1.349 0 0 0 1.345-1.345V11.4zm-.672-2.694-.948-.948-1.741 1.735V4H8.706v5.493L6.965 7.758l-.948.948 3.361 3.361z\" transform=\"translate(-.437 -.414)\" style=\"fill: rgb(13, 123, 196);\"><\/path><\/g><\/svg>Download PDF<\/a><\/div><\/div><\/div>\n\n<div class=\"wp-block-custom-chromatogram-article\"><div class=\"wp-block-custom-chromatogram-article\"><div class=\"chromatogram-image regular-image\"><img decoding=\"async\" src=\"https:\/\/ez.restek.com\/images\/cgram\/lc_ev0593.png\" alt=\"Comparison of Pesticides on Raptor Inert ARC-18 and Raptor ARC-18 - Overlay View\" title=\"-\"><\/div><p class=\"article-id\" style=\"text-align:center\"> LC_EV0593<\/p><div class=\"chromatogram-related-skus\"><h4>Related SKUs<\/h4><ul><li>31972<\/li><li>9314A12<\/li><\/ul><\/div><div class=\"chromatogram-pdf-link\"><a href=\"https:\/\/ez.restek.com\/images\/cgram\/lc_ev0593.pdf\" target=\"_blank\" rel=\"noopener noreferrer\"><svg xmlns=\"http:\/\/www.w3.org\/2000\/svg\" width=\"18\" height=\"18\" viewBox=\"0 0 18 18\"><g data-name=\"Group 2996\"><path data-name=\"Rectangle 1246\" d=\"M0 0h18v18H0z\" style=\"fill: none;\"><\/path><\/g><g data-name=\"Group 2997\"><path data-name=\"Path 729\" d=\"M13.412 11.4v2.017H5.345V11.4H4v2.017a1.349 1.349 0 0 0 1.345 1.345h8.068a1.349 1.349 0 0 0 1.345-1.345V11.4zm-.672-2.694-.948-.948-1.741 1.735V4H8.706v5.493L6.965 7.758l-.948.948 3.361 3.361z\" transform=\"translate(-.437 -.414)\" style=\"fill: rgb(13, 123, 196);\"><\/path><\/g><\/svg>Download PDF<\/a><\/div><\/div><\/div>\n<\/div><\/div><\/div>\n\n\n<div style=\"height:50px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n\n\n\n<p class=\"wp-block-paragraph\"><strong>Table I:&nbsp;<\/strong>Pesticide peak areas and heights were up to 2X higher on Raptor Inert LC columns compared to standard columns.<\/p>\n\n\n\n<figure class=\"wp-block-table\"><table style=\"border-width:1px\"><tbody><tr><td class=\"has-text-align-left\" data-align=\"left\" rowspan=\"2\"><strong>Compound<\/strong><\/td><td class=\"has-text-align-center\" data-align=\"center\" colspan=\"3\"><strong>Peak Area<\/strong><\/td><td class=\"has-text-align-center\" data-align=\"center\" colspan=\"3\"><strong>Peak Height<\/strong><\/td><\/tr><tr><td class=\"has-text-align-left\" data-align=\"left\"><strong>Stainless Steel<\/strong><\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>Inert<\/strong><\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>Areas Ratio (Inert\/Stainless Steel)<\/strong><\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>Stainless Steel<\/strong><\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>Inert<\/strong><\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>Height Ratio (Inert\/Stainless Steel)<\/strong><\/td><\/tr><tr><td class=\"has-text-align-left\" data-align=\"left\">Methamidophos<\/td><td class=\"has-text-align-center\" data-align=\"center\">254969<\/td><td class=\"has-text-align-center\" data-align=\"center\">428941<\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>1.68<\/strong><\/td><td class=\"has-text-align-center\" data-align=\"center\">52553<\/td><td class=\"has-text-align-center\" data-align=\"center\">105189<\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>2.00<\/strong><\/td><\/tr><tr><td class=\"has-text-align-left\" data-align=\"left\">Acephate<\/td><td class=\"has-text-align-center\" data-align=\"center\">168776<\/td><td class=\"has-text-align-center\" data-align=\"center\">300642<\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>1.78<\/strong><\/td><td class=\"has-text-align-center\" data-align=\"center\">58418<\/td><td class=\"has-text-align-center\" data-align=\"center\">104729<\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>1.79<\/strong><\/td><\/tr><tr><td class=\"has-text-align-left\" data-align=\"left\">Omethoate<\/td><td class=\"has-text-align-center\" data-align=\"center\">579502<\/td><td class=\"has-text-align-center\" data-align=\"center\">892008<\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>1.54<\/strong><\/td><td class=\"has-text-align-center\" data-align=\"center\">216157<\/td><td class=\"has-text-align-center\" data-align=\"center\">337690<\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>1.56<\/strong><\/td><\/tr><tr><td class=\"has-text-align-left\" data-align=\"left\">Monocrotophos<\/td><td class=\"has-text-align-center\" data-align=\"center\">140095<\/td><td class=\"has-text-align-center\" data-align=\"center\">215810<\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>1.54<\/strong><\/td><td class=\"has-text-align-center\" data-align=\"center\">51402<\/td><td class=\"has-text-align-center\" data-align=\"center\">78425<\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>1.53<\/strong><\/td><\/tr><tr><td class=\"has-text-align-left\" data-align=\"left\">Dicrotophos<\/td><td class=\"has-text-align-center\" data-align=\"center\">340978<\/td><td class=\"has-text-align-center\" data-align=\"center\">404916<\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>1.19<\/strong><\/td><td class=\"has-text-align-center\" data-align=\"center\">135380<\/td><td class=\"has-text-align-center\" data-align=\"center\">159292<\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>1.18<\/strong><\/td><\/tr><tr><td class=\"has-text-align-left\" data-align=\"left\">Dimethoate<\/td><td class=\"has-text-align-center\" data-align=\"center\">461156<\/td><td class=\"has-text-align-center\" data-align=\"center\">807805<\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>1.75<\/strong><\/td><td class=\"has-text-align-center\" data-align=\"center\">188746<\/td><td class=\"has-text-align-center\" data-align=\"center\">342939<\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>1.82<\/strong><\/td><\/tr><tr><td class=\"has-text-align-left\" data-align=\"left\">Trichlorfon<\/td><td class=\"has-text-align-center\" data-align=\"center\">84233<\/td><td class=\"has-text-align-center\" data-align=\"center\">173942<\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>2.07<\/strong><\/td><td class=\"has-text-align-center\" data-align=\"center\">34793<\/td><td class=\"has-text-align-center\" data-align=\"center\">63266<\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>1.82<\/strong><\/td><\/tr><tr><td class=\"has-text-align-left\" data-align=\"left\">Vamidothion<\/td><td class=\"has-text-align-center\" data-align=\"center\">913264<\/td><td class=\"has-text-align-center\" data-align=\"center\">1333829<\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>1.46<\/strong><\/td><td class=\"has-text-align-center\" data-align=\"center\">354311<\/td><td class=\"has-text-align-center\" data-align=\"center\">547308<\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>1.54<\/strong><\/td><\/tr><tr><td class=\"has-text-align-left\" data-align=\"left\">Mevinphos isomer 1<\/td><td class=\"has-text-align-center\" data-align=\"center\">213632<\/td><td class=\"has-text-align-center\" data-align=\"center\">311274<\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>1.46<\/strong><\/td><td class=\"has-text-align-center\" data-align=\"center\">82105<\/td><td class=\"has-text-align-center\" data-align=\"center\">129961<\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>1.58<\/strong><\/td><\/tr><tr><td class=\"has-text-align-left\" data-align=\"left\">Mevinphos isomer 2<\/td><td class=\"has-text-align-center\" data-align=\"center\">56093<\/td><td class=\"has-text-align-center\" data-align=\"center\">74030<\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>&nbsp;1.32<\/strong><\/td><td class=\"has-text-align-center\" data-align=\"center\">29070<\/td><td class=\"has-text-align-center\" data-align=\"center\">29802<\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>1.03&nbsp;<\/strong><\/td><\/tr><tr><td class=\"has-text-align-left\" data-align=\"left\">Carbaryl<\/td><td class=\"has-text-align-center\" data-align=\"center\">43590<\/td><td class=\"has-text-align-center\" data-align=\"center\">39671<\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>0.91<\/strong><\/td><td class=\"has-text-align-center\" data-align=\"center\">14563<\/td><td class=\"has-text-align-center\" data-align=\"center\">11924<\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>0.82<\/strong><\/td><\/tr><tr><td class=\"has-text-align-left\" data-align=\"left\">Isocarbophos<\/td><td class=\"has-text-align-center\" data-align=\"center\">21587<\/td><td class=\"has-text-align-center\" data-align=\"center\">33294<\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>1.54<\/strong><\/td><td class=\"has-text-align-center\" data-align=\"center\">9062<\/td><td class=\"has-text-align-center\" data-align=\"center\">11941<\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>1.32<\/strong><\/td><\/tr><tr><td class=\"has-text-align-left\" data-align=\"left\">Dimethomorph isomer 1<\/td><td class=\"has-text-align-center\" data-align=\"center\">462425<\/td><td class=\"has-text-align-center\" data-align=\"center\">511766<\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>1.11<\/strong><\/td><td class=\"has-text-align-center\" data-align=\"center\">166990<\/td><td class=\"has-text-align-center\" data-align=\"center\">172977<\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>1.04<\/strong><\/td><\/tr><tr><td class=\"has-text-align-left\" data-align=\"left\">Dimethomorph isomer 2<\/td><td class=\"has-text-align-center\" data-align=\"center\">896109<\/td><td class=\"has-text-align-center\" data-align=\"center\">877031<\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>0.98&nbsp;<\/strong><\/td><td class=\"has-text-align-center\" data-align=\"center\">311657<\/td><td class=\"has-text-align-center\" data-align=\"center\">328826<\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>1.06&nbsp;<\/strong><\/td><\/tr><tr><td class=\"has-text-align-left\" data-align=\"left\">Temephos<\/td><td class=\"has-text-align-center\" data-align=\"center\">98793<\/td><td class=\"has-text-align-center\" data-align=\"center\">164310<\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>1.66<\/strong><\/td><td class=\"has-text-align-center\" data-align=\"center\">35383<\/td><td class=\"has-text-align-center\" data-align=\"center\">64751<\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>1.83<\/strong><\/td><\/tr><\/tbody><\/table><\/figure>\n\n\n\n<h2 class=\"wp-block-heading\">Mycotoxins<\/h2>\n\n\n\n<p class=\"wp-block-paragraph\">In addition to pesticides analysis, we also wanted to test whether using an inert LC column could improve results for mycotoxins analysis. Because mycotoxins are reactive compounds that can have acidic, polar, or metal-chelating groups, extensive column conditioning and equilibration is often required to obtain adequate chromatography. For these experiments, we used columns with a Biphenyl stationary phase for optimal selectivity. The results shown in Figures 3-4 and Table II clearly demonstrate that peak shape and response for mycotoxins can also be improved by using an inert column.<\/p>\n\n\n<div class=\"wp-block-custom-chromatogram-article-top\"><div class=\"chromatogram-article-placeholder\"><div class=\"figure-heading\"><strong>Figure 3:<\/strong> Raptor Inert Biphenyl columns provide excellent peak shape for mycotoxins without additional acid passivation or mobile phase additives.<\/div><div class='chromatogram-article-inner-full'><div class=\"chromatogram-article-inner\">\n<div class=\"wp-block-custom-chromatogram-article\"><div class=\"wp-block-custom-chromatogram-article\"><div class=\"chromatogram-image regular-image\"><img decoding=\"async\" src=\"https:\/\/ez.restek.com\/images\/cgram\/lc_fs0552.png\" alt=\"Multi-Mycotoxin Analysis on Raptor Inert Biphenyl\" title=\"-\"><\/div><p class=\"article-id\" style=\"text-align:center\"> LC_FS0552<\/p><div class=\"chromatogram-peaks\"><h4>Peaks<\/h4><table class=\"peaks col-12\">\n<thead><tr><th><\/th><th style=\"text-align: left;width: 75px\">Peaks<\/th><th style=\"text-align: center;width: 75px\">t<sub>R<\/sub> (min)<\/th><th style=\"text-align: center;width: 75px\">Conc.<br \/>(ng\/mL)<\/th><th style=\"text-align: center;width: 75px\">Precursor Ion<\/th><th style=\"text-align: center;width: 75px\">Product Ion <\/th><th style=\"text-align: center;width: 75px\">Peak Area<\/th><th style=\"text-align: center;width: 75px\">Peak Height<\/th><\/tr><\/thead>\n<tbody><tr><td class=\"num\">1.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Nivalenol \" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/23282-20-4\/Nivalenol \" target=\"_blank\" rel=\"noopener\">Nivalenol <\/a><\/td><td class=\"oth\">0.88<\/td><td class=\"oth\">10<\/td><td class=\"oth\">295.1<\/td><td class=\"oth\">137.1<\/td><td class=\"oth\">4182<\/td><td class=\"oth\">64495<\/td><\/tr>\n<tr><td class=\"num\">2.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Deoxynivalenol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/51481-10-8\/Deoxynivalenol\" target=\"_blank\" rel=\"noopener\">Deoxynivalenol<\/a><\/td><td class=\"oth\">1.25<\/td><td class=\"oth\">10<\/td><td class=\"oth\">297.2<\/td><td class=\"oth\">231.0<\/td><td class=\"oth\">17346<\/td><td class=\"oth\">281906<\/td><\/tr>\n<tr><td class=\"num\">3.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Fusarenon-X\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/23255-69-8\/Fusarenon-X\" target=\"_blank\" rel=\"noopener\">Fusarenon-X<\/a><\/td><td class=\"oth\">1.92<\/td><td class=\"oth\">10<\/td><td class=\"oth\">355.1<\/td><td class=\"oth\">137.1<\/td><td class=\"oth\">7668<\/td><td class=\"oth\">121790<\/td><\/tr>\n<tr><td class=\"num\">4.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 15-Acetyldeoxynivalenol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/88337-96-6\/15-Acetyldeoxynivalenol\" target=\"_blank\" rel=\"noopener\">15-Acetyldeoxynivalenol<\/a><\/td><td class=\"oth\">3.08<\/td><td class=\"oth\">10<\/td><td class=\"oth\">339.2<\/td><td class=\"oth\">137.1<\/td><td class=\"oth\">31369<\/td><td class=\"oth\">517570<\/td><\/tr>\n<tr><td class=\"num\">5.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 3-Acetyldeoxynivalenol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/50722-38-8\/3-Acetyldeoxynivalenol\" target=\"_blank\" rel=\"noopener\">3-Acetyldeoxynivalenol<\/a><\/td><td class=\"oth\">3.14<\/td><td class=\"oth\">10<\/td><td class=\"oth\">339.2<\/td><td class=\"oth\">213.1<\/td><td class=\"oth\">22613<\/td><td class=\"oth\">296396<\/td><\/tr>\n<tr><td class=\"num\">6.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Tenuazonic acid\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/610-88-8\/Tenuazonic acid\" target=\"_blank\" rel=\"noopener\">Tenuazonic acid<\/a><\/td><td class=\"oth\">4.11<\/td><td class=\"oth\">10<\/td><td class=\"oth\">198.1<\/td><td class=\"oth\">125.0<\/td><td class=\"oth\">47828<\/td><td class=\"oth\">197658<\/td><\/tr>\n<tr><td class=\"num\">7.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Altenuene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/29752-43-0\/Altenuene\" target=\"_blank\" rel=\"noopener\">Altenuene<\/a><\/td><td class=\"oth\">4.60<\/td><td class=\"oth\">10<\/td><td class=\"oth\">293.2<\/td><td class=\"oth\">257.1<\/td><td class=\"oth\">113850<\/td><td class=\"oth\">2059699<\/td><\/tr>\n<tr><td class=\"num\">8.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Alternariol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/641-38-3\/Alternariol\" target=\"_blank\" rel=\"noopener\">Alternariol<\/a><\/td><td class=\"oth\">5.27<\/td><td class=\"oth\">10<\/td><td class=\"oth\">259.0<\/td><td class=\"oth\">185.1<\/td><td class=\"oth\">73272<\/td><td class=\"oth\">1302192<\/td><\/tr>\n<tr><td class=\"num\">9.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Ergosine\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/561-94-4\/Ergosine\" target=\"_blank\" rel=\"noopener\">Ergosine<\/a><\/td><td class=\"oth\">5.28<\/td><td class=\"oth\">10<\/td><td class=\"oth\">548.4<\/td><td class=\"oth\">208.1<\/td><td class=\"oth\">486620<\/td><td class=\"oth\">9366601<\/td><\/tr>\n<tr><td class=\"num\">10.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Citrinin\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/518-75-2\/Citrinin\" target=\"_blank\" rel=\"noopener\">Citrinin<\/a><\/td><td class=\"oth\">5.46<\/td><td class=\"oth\">10<\/td><td class=\"oth\">251.2<\/td><td class=\"oth\">233.1<\/td><td class=\"oth\">1007880<\/td><td class=\"oth\">9828889<\/td><\/tr>\n<tr><td class=\"num\">11.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Ergosinine\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/596-88-3\/Ergosinine\" target=\"_blank\" rel=\"noopener\">Ergosinine<\/a><\/td><td class=\"oth\">5.46<\/td><td class=\"oth\">10<\/td><td class=\"oth\">548.4<\/td><td class=\"oth\">208.1<\/td><td class=\"oth\">496734<\/td><td class=\"oth\">8740527<\/td><\/tr>\n<tr><td class=\"num\">12.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Fumonisin B1\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/116355-83-0\/Fumonisin B1\" target=\"_blank\" rel=\"noopener\">Fumonisin B1<\/a><\/td><td class=\"oth\">5.48<\/td><td class=\"oth\">10<\/td><td class=\"oth\">722.5<\/td><td class=\"oth\">352.3<\/td><td class=\"oth\">122878<\/td><td class=\"oth\">2415567<\/td><\/tr>\n<tr><td class=\"num\">13.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Diacetoxyscirpenol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/2270-40-8\/Diacetoxyscirpenol\" target=\"_blank\" rel=\"noopener\">Diacetoxyscirpenol<\/a><\/td><td class=\"oth\">5.62<\/td><td class=\"oth\">10<\/td><td class=\"oth\">384.2<\/td><td class=\"oth\">247.1<\/td><td class=\"oth\">68139<\/td><td class=\"oth\">1208825<\/td><\/tr>\n<tr><td class=\"num\">14.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Ergotamine\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/113-15-5\/Ergotamine\" target=\"_blank\" rel=\"noopener\">Ergotamine<\/a><\/td><td class=\"oth\">5.71<\/td><td class=\"oth\">10<\/td><td class=\"oth\">582.4<\/td><td class=\"oth\">223.2<\/td><td class=\"oth\">493003<\/td><td class=\"oth\">9274155<\/td><\/tr>\n<tr><td class=\"num\">15.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Ergocornine\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/564-36-3\/Ergocornine\" target=\"_blank\" rel=\"noopener\">Ergocornine<\/a><\/td><td class=\"oth\">5.85<\/td><td class=\"oth\">10<\/td><td class=\"oth\">562.4<\/td><td class=\"oth\">268.2<\/td><td class=\"oth\">387025<\/td><td class=\"oth\">7732744<\/td><\/tr>\n<tr><td class=\"num\">16.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Ergotaminine\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/639-81-6\/Ergotaminine\" target=\"_blank\" rel=\"noopener\">Ergotaminine<\/a><\/td><td class=\"oth\">5.96<\/td><td class=\"oth\">10<\/td><td class=\"oth\">582.4<\/td><td class=\"oth\">223.2<\/td><td class=\"oth\">462119<\/td><td class=\"oth\">9237991<\/td><\/tr>\n<tr><td class=\"num\">17.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for HT-2\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/26934-87-2\/HT-2\" target=\"_blank\" rel=\"noopener\">HT-2<\/a><\/td><td class=\"oth\">6.13<\/td><td class=\"oth\">10<\/td><td class=\"oth\">447.2<\/td><td class=\"oth\">345.1<\/td><td class=\"oth\">15221<\/td><td class=\"oth\">323765<\/td><\/tr>\n<tr><td class=\"num\">18.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Ergocryptine\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/511-09-1\/Ergocryptine\" target=\"_blank\" rel=\"noopener\">Ergocryptine<\/a><\/td><td class=\"oth\">6.19<\/td><td class=\"oth\">10<\/td><td class=\"oth\">576.4<\/td><td class=\"oth\">268.2<\/td><td class=\"oth\">522204<\/td><td class=\"oth\">11360838<\/td><\/tr>\n<tr><td class=\"num\">19.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Fumonisin B3\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/1422359-85-0\/Fumonisin B3\" target=\"_blank\" rel=\"noopener\">Fumonisin B3<\/a><\/td><td class=\"oth\">6.23<\/td><td class=\"oth\">10<\/td><td class=\"oth\">706.4<\/td><td class=\"oth\">336.2<\/td><td class=\"oth\">143302<\/td><td class=\"oth\">3444421<\/td><\/tr>\n<tr><td class=\"num\">20.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Ergocristine\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/511-08-0\/Ergocristine\" target=\"_blank\" rel=\"noopener\">Ergocristine<\/a><\/td><td class=\"oth\">6.44<\/td><td class=\"oth\">10<\/td><td class=\"oth\">610.4<\/td><td class=\"oth\">223.2<\/td><td class=\"oth\">195562<\/td><td class=\"oth\">4450058<\/td><\/tr>\n<tr><td class=\"num\">21.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Fumonisin B2\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/116355-84-1\/Fumonisin B2\" target=\"_blank\" rel=\"noopener\">Fumonisin B2<\/a><\/td><td class=\"oth\">6.59<\/td><td class=\"oth\">10<\/td><td class=\"oth\">706.4<\/td><td class=\"oth\">336.2<\/td><td class=\"oth\">151719<\/td><td class=\"oth\">3869822<\/td><\/tr>\n<tr><td class=\"num\">22.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Tentoxin\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/28540-82-1\/Tentoxin\" target=\"_blank\" rel=\"noopener\">Tentoxin<\/a><\/td><td class=\"oth\">6.62<\/td><td class=\"oth\">10<\/td><td class=\"oth\">415.2<\/td><td class=\"oth\">312.2<\/td><td class=\"oth\">95175<\/td><td class=\"oth\">2131906<\/td><\/tr>\n<tr><td class=\"num\">23.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for \u03b1-Zearalenol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/36455-72-8\/\u03b1-Zearalenol\" target=\"_blank\" rel=\"noopener\">\u03b1-Zearalenol<\/a><\/td><td class=\"oth\">6.91<\/td><td class=\"oth\">10<\/td><td class=\"oth\">303.1<\/td><td class=\"oth\">285.1<\/td><td class=\"oth\">30224<\/td><td class=\"oth\">702420<\/td><\/tr>\n<tr><td class=\"num\">24.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Ergocorninine\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/564-37-4\/Ergocorninine\" target=\"_blank\" rel=\"noopener\">Ergocorninine<\/a><\/td><td class=\"oth\">6.93<\/td><td class=\"oth\">10<\/td><td class=\"oth\">562.4<\/td><td class=\"oth\">268.2<\/td><td class=\"oth\">704029<\/td><td class=\"oth\">14389283<\/td><\/tr>\n<tr><td class=\"num\">25.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Aflatoxin G2\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/7241-98-7\/Aflatoxin G2\" target=\"_blank\" rel=\"noopener\">Aflatoxin G2<\/a><\/td><td class=\"oth\">6.97<\/td><td class=\"oth\">10<\/td><td class=\"oth\">331.2<\/td><td class=\"oth\">189.0<\/td><td class=\"oth\">262824<\/td><td class=\"oth\">5274353<\/td><\/tr>\n<tr><td class=\"num\">26.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for T-2\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/50-31-7\/T-2\" target=\"_blank\" rel=\"noopener\">T-2<\/a><\/td><td class=\"oth\">7.09<\/td><td class=\"oth\">10<\/td><td class=\"oth\">489.2<\/td><td class=\"oth\">387.1<\/td><td class=\"oth\">56535<\/td><td class=\"oth\">1394735<\/td><\/tr>\n<tr><td class=\"num\">27.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Ergocryptinine\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/511-10-4\/Ergocryptinine\" target=\"_blank\" rel=\"noopener\">Ergocryptinine<\/a><\/td><td class=\"oth\">7.18<\/td><td class=\"oth\">10<\/td><td class=\"oth\">576.4<\/td><td class=\"oth\">268.2<\/td><td class=\"oth\">778972<\/td><td class=\"oth\">16765348<\/td><\/tr>\n<tr><td class=\"num\">28.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Ergocristinine\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/511-07-9\/Ergocristinine\" target=\"_blank\" rel=\"noopener\">Ergocristinine<\/a><\/td><td class=\"oth\">7.40<\/td><td class=\"oth\">10<\/td><td class=\"oth\">610.4<\/td><td class=\"oth\">223.2<\/td><td class=\"oth\">1583053<\/td><td class=\"oth\">32975663<\/td><\/tr>\n<tr><td class=\"num\">29.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Aflatoxin G1\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/1165-39-5\/Aflatoxin G1\" target=\"_blank\" rel=\"noopener\">Aflatoxin G1<\/a><\/td><td class=\"oth\">7.45<\/td><td class=\"oth\">10<\/td><td class=\"oth\">329.1<\/td><td class=\"oth\">199.7<\/td><td class=\"oth\">304389<\/td><td class=\"oth\">6102959<\/td><\/tr>\n<tr><td class=\"num\">30.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Zearalenone\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/17924-92-4\/Zearalenone\" target=\"_blank\" rel=\"noopener\">Zearalenone<\/a><\/td><td class=\"oth\">7.59<\/td><td class=\"oth\">10<\/td><td class=\"oth\">319.2<\/td><td class=\"oth\">283.1<\/td><td class=\"oth\">37162<\/td><td class=\"oth\">927455<\/td><\/tr>\n<tr><td class=\"num\">31.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Alternariol monomethylether\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/26894-49-5\/Alternariol monomethylether\" target=\"_blank\" rel=\"noopener\">Alternariol monomethylether<\/a><\/td><td class=\"oth\">7.62<\/td><td class=\"oth\">10<\/td><td class=\"oth\">273.0<\/td><td class=\"oth\">199.1<\/td><td class=\"oth\">31024<\/td><td class=\"oth\">640689<\/td><\/tr>\n<tr><td class=\"num\">32.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Aflatoxin B2\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/7220-81-7\/Aflatoxin B2\" target=\"_blank\" rel=\"noopener\">Aflatoxin B2<\/a><\/td><td class=\"oth\">7.63<\/td><td class=\"oth\">10<\/td><td class=\"oth\">315.1<\/td><td class=\"oth\">287.0<\/td><td class=\"oth\">295648<\/td><td class=\"oth\">5724754<\/td><\/tr>\n<tr><td class=\"num\">33.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Aflatoxin B1\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/1162-65-8\/Aflatoxin B1\" target=\"_blank\" rel=\"noopener\">Aflatoxin B1<\/a><\/td><td class=\"oth\">8.02<\/td><td class=\"oth\">10<\/td><td class=\"oth\">313.2<\/td><td class=\"oth\">241.1<\/td><td class=\"oth\">223520<\/td><td class=\"oth\">4425821<\/td><\/tr>\n<tr><td class=\"num\">34.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Ochratoxin A\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/303-47-9\/Ochratoxin A\" target=\"_blank\" rel=\"noopener\">Ochratoxin A<\/a><\/td><td class=\"oth\">8.25<\/td><td class=\"oth\">10<\/td><td class=\"oth\">404.1<\/td><td class=\"oth\">239.0<\/td><td class=\"oth\">190060<\/td><td class=\"oth\">4411953<\/td><\/tr>\n<\/tbody><\/table><\/div><div class=\"chromatogram-conditions\"><h4>Conditions<\/h4><div class=\"conditions-container container-fluid\"><div class=\"row\"><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Column<\/th><td>Raptor Inert Biphenyl  (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/9309A12-T\" rel=\"noopener\">cat.# 9309A12-T<\/a>)<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Dimensions:<\/th><td>100 mm x 2.1 mm ID<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Particle Size:<\/th><td>2.7 \u00b5m<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Pore Size:<\/th><td>90 \u00c5<\/td><\/tr><tr><td><\/td><tr><th class=\"sub conditions_header\" scope=\"row\">Temp.:<\/th><td>60 \u00b0C<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Standard\/Sample<\/th><td><\/td><\/tr><tr><td><\/td><td>Aflatoxins standard (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/34121\" rel=\"noopener\">cat.# 34121<\/a>)<\/td><\/tr><tr><td><\/td><td>Ochratoxin A standard (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/34122\" rel=\"noopener\">cat.# 34122<\/a>)<\/td><\/tr><tr><td><\/td><\/td><\/tr>\n<tr><th class=\"sub conditions_header\" scope=\"row\">Diluent:<\/th><td>50:50 Water:methanol<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Conc.:<\/th><td>10 ng\/mL<\/td><\/tr><td><\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Inj. Vol.:<\/th><td>5 \u00b5L <\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Mobile Phase<\/th><td><\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">A:<\/th><td>Water, 0.05% formic acid  <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">B:<\/th><td>Methanol, 0.05% formic acid <\/td><\/tr><tr><td><\/td><td><table class=\"cgram_ramp\"><thead><tr><th>Time (min)<\/th><th>Flow (mL\/min)<\/th><th>%A<\/th><th>%B<\/th><\/tr><\/thead><tbody><tr><td>0.00<\/td><td>0.4<\/td><td>75<\/td><td>25<\/td><\/tr><tr><td>5.00<\/td><td>0.4<\/td><td>50<\/td><td>50<\/td><\/tr><tr><td>9.00<\/td><td>0.4<\/td><td>0<\/td><td>100<\/td><\/tr><tr><td>9.01<\/td><td>0.4<\/td><td>75<\/td><td>25<\/td><\/tr><tr><td>11.0<\/td><td>0.4<\/td><td>75<\/td><td>25<\/td><\/tr><\/tbody><\/table><\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Max Pressure:<\/th><td>440 bar<\/td><\/tr><\/table><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Detector<\/th><td>Waters Xevo TQ-S<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Ion Mode:<\/th><td>ESI+ <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Mode:<\/th><td>MRM <\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Instrument<\/th><td>Waters ACQUITY UPLC I-Class<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Notes<\/th><td><br \/><br \/><\/td><\/tr><\/table><\/div><\/div><\/div><div class=\"chromatogram-related-skus\"><h4>Related SKUs<\/h4><ul><li>34121<\/li><li>34122<\/li><li>9309A12-T<\/li><\/ul><\/div><div class=\"chromatogram-related-searches\"><h4>Related Searches<\/h4><ul><li>mycotoxin<\/li><\/ul><\/div><div class=\"chromatogram-pdf-link\"><a href=\"https:\/\/ez.restek.com\/images\/cgram\/lc_fs0552.pdf\" target=\"_blank\" rel=\"noopener noreferrer\"><svg xmlns=\"http:\/\/www.w3.org\/2000\/svg\" width=\"18\" height=\"18\" viewBox=\"0 0 18 18\"><g data-name=\"Group 2996\"><path data-name=\"Rectangle 1246\" d=\"M0 0h18v18H0z\" style=\"fill: none;\"><\/path><\/g><g data-name=\"Group 2997\"><path data-name=\"Path 729\" d=\"M13.412 11.4v2.017H5.345V11.4H4v2.017a1.349 1.349 0 0 0 1.345 1.345h8.068a1.349 1.349 0 0 0 1.345-1.345V11.4zm-.672-2.694-.948-.948-1.741 1.735V4H8.706v5.493L6.965 7.758l-.948.948 3.361 3.361z\" transform=\"translate(-.437 -.414)\" style=\"fill: rgb(13, 123, 196);\"><\/path><\/g><\/svg>Download PDF<\/a><\/div><\/div><\/div>\n<\/div><\/div><\/div>\n\n\n<div style=\"height:25px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n\n\n<div class=\"wp-block-custom-chromatogram-article-top\"><div class=\"chromatogram-article-placeholder\"><div class=\"figure-heading\"><strong>Figure 4:<\/strong> Substantial increases in peak area and height were seen for fumonisins when using a Raptor Inert Biphenyl column.<\/div><div class='chromatogram-article-inner-full'><div class=\"chromatogram-article-inner\">\n<div class=\"wp-block-custom-chromatogram-article\"><div class=\"wp-block-custom-chromatogram-article\"><div class=\"chromatogram-image wide-image\"><img decoding=\"async\" src=\"https:\/\/ez.restek.com\/images\/cgram\/lc_fs0554.png\" alt=\"Comparison of Fumonisin B Compounds on Raptor Inert Biphenyl and Raptor Biphenyl\" title=\"-\"><\/div><p class=\"article-id\" style=\"text-align:center\"> LC_FS0554<\/p><div class=\"chromatogram-related-skus\"><h4>Related SKUs<\/h4><ul><li>9309A12<\/li><\/ul><\/div><div class=\"chromatogram-pdf-link\"><a href=\"https:\/\/ez.restek.com\/images\/cgram\/lc_fs0554.pdf\" target=\"_blank\" rel=\"noopener noreferrer\"><svg xmlns=\"http:\/\/www.w3.org\/2000\/svg\" width=\"18\" height=\"18\" viewBox=\"0 0 18 18\"><g data-name=\"Group 2996\"><path data-name=\"Rectangle 1246\" d=\"M0 0h18v18H0z\" style=\"fill: none;\"><\/path><\/g><g data-name=\"Group 2997\"><path data-name=\"Path 729\" d=\"M13.412 11.4v2.017H5.345V11.4H4v2.017a1.349 1.349 0 0 0 1.345 1.345h8.068a1.349 1.349 0 0 0 1.345-1.345V11.4zm-.672-2.694-.948-.948-1.741 1.735V4H8.706v5.493L6.965 7.758l-.948.948 3.361 3.361z\" transform=\"translate(-.437 -.414)\" style=\"fill: rgb(13, 123, 196);\"><\/path><\/g><\/svg>Download PDF<\/a><\/div><\/div><\/div>\n\n<div class=\"wp-block-custom-chromatogram-article\"><div class=\"wp-block-custom-chromatogram-article\"><div class=\"chromatogram-image regular-image\"><img decoding=\"async\" src=\"https:\/\/ez.restek.com\/images\/cgram\/lc_fs0556.png\" alt=\"Comparison of Fumonisin B Compounds on Raptor Inert Biphenyl and Raptor Biphenyl\u2014Overlay View\" title=\"-\"><\/div><p class=\"article-id\" style=\"text-align:center\"> LC_FS0556<\/p><div class=\"chromatogram-related-skus\"><h4>Related SKUs<\/h4><ul><li>9309A12<\/li><\/ul><\/div><div class=\"chromatogram-pdf-link\"><a href=\"https:\/\/ez.restek.com\/images\/cgram\/lc_fs0556.pdf\" target=\"_blank\" rel=\"noopener noreferrer\"><svg xmlns=\"http:\/\/www.w3.org\/2000\/svg\" width=\"18\" height=\"18\" viewBox=\"0 0 18 18\"><g data-name=\"Group 2996\"><path data-name=\"Rectangle 1246\" d=\"M0 0h18v18H0z\" style=\"fill: none;\"><\/path><\/g><g data-name=\"Group 2997\"><path data-name=\"Path 729\" d=\"M13.412 11.4v2.017H5.345V11.4H4v2.017a1.349 1.349 0 0 0 1.345 1.345h8.068a1.349 1.349 0 0 0 1.345-1.345V11.4zm-.672-2.694-.948-.948-1.741 1.735V4H8.706v5.493L6.965 7.758l-.948.948 3.361 3.361z\" transform=\"translate(-.437 -.414)\" style=\"fill: rgb(13, 123, 196);\"><\/path><\/g><\/svg>Download PDF<\/a><\/div><\/div><\/div>\n<\/div><\/div><\/div>\n\n\n<div style=\"height:50px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n\n\n\n<p class=\"wp-block-paragraph\"><strong>Table II:<\/strong> For mycotoxins, peak height increased up to 10X and peak area increased up to 6X when using a Raptor Inert Biphenyl column compared to a standard column.<\/p>\n\n\n\n<figure class=\"wp-block-table\"><table style=\"border-width:1px\"><tbody><tr><td rowspan=\"2\"><strong>Compound<\/strong><\/td><td class=\"has-text-align-center\" data-align=\"center\" colspan=\"3\"><strong>Peak Area<\/strong><\/td><td class=\"has-text-align-center\" data-align=\"center\" colspan=\"3\"><strong>Peak Height<\/strong><\/td><\/tr><tr><td><strong>Stainless Steel<\/strong><\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>Inert<\/strong><\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>Areas Ratio (Inert\/Stainless Steel)<\/strong><\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>Stainless Steel<\/strong><\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>Inert<\/strong><\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>Height Ratio (Inert\/Stainless Steel)<\/strong><\/td><\/tr><tr><td>Fumonisin B1<\/td><td class=\"has-text-align-center\" data-align=\"center\">32578<\/td><td class=\"has-text-align-center\" data-align=\"center\">122878<\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>3.77<\/strong><\/td><td class=\"has-text-align-center\" data-align=\"center\">399544<\/td><td class=\"has-text-align-center\" data-align=\"center\">2415567<\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>6.05<\/strong><\/td><\/tr><tr><td>Fumonisin B2<\/td><td class=\"has-text-align-center\" data-align=\"center\">23427<\/td><td class=\"has-text-align-center\" data-align=\"center\">151719<\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>6.48<\/strong><\/td><td class=\"has-text-align-center\" data-align=\"center\">383130<\/td><td class=\"has-text-align-center\" data-align=\"center\">3869822<\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>10.10<\/strong><\/td><\/tr><tr><td>Fumonisin B3<\/td><td class=\"has-text-align-center\" data-align=\"center\">29864<\/td><td class=\"has-text-align-center\" data-align=\"center\">143302<\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>4.80<\/strong><\/td><td class=\"has-text-align-center\" data-align=\"center\">472279<\/td><td class=\"has-text-align-center\" data-align=\"center\">3444421<\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>7.29<\/strong><\/td><\/tr><tr><td>Ergocristine<\/td><td class=\"has-text-align-center\" data-align=\"center\">171197<\/td><td class=\"has-text-align-center\" data-align=\"center\">195562<\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>1.14<\/strong><\/td><td class=\"has-text-align-center\" data-align=\"center\">3865898<\/td><td class=\"has-text-align-center\" data-align=\"center\">4450058<\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>1.15<\/strong><\/td><\/tr><tr><td>Ergocristinine<\/td><td class=\"has-text-align-center\" data-align=\"center\">1393116<\/td><td class=\"has-text-align-center\" data-align=\"center\">1583053<\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>1.14<\/strong><\/td><td class=\"has-text-align-center\" data-align=\"center\">29212317<\/td><td class=\"has-text-align-center\" data-align=\"center\">32975663<\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>1.13<\/strong><\/td><\/tr><tr><td>Ergotamine<\/td><td class=\"has-text-align-center\" data-align=\"center\">433635<\/td><td class=\"has-text-align-center\" data-align=\"center\">493003<\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>1.14<\/strong><\/td><td class=\"has-text-align-center\" data-align=\"center\">8149518<\/td><td class=\"has-text-align-center\" data-align=\"center\">9274156<\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>1.14<\/strong><\/td><\/tr><tr><td>Ergotaminine<\/td><td class=\"has-text-align-center\" data-align=\"center\">397370<\/td><td class=\"has-text-align-center\" data-align=\"center\">462119<\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>1.16<\/strong><\/td><td class=\"has-text-align-center\" data-align=\"center\">7885403<\/td><td class=\"has-text-align-center\" data-align=\"center\">9237991<\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>1.17<\/strong><\/td><\/tr><tr><td>Ergocryptine<\/td><td class=\"has-text-align-center\" data-align=\"center\">446481<\/td><td class=\"has-text-align-center\" data-align=\"center\">522204<\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>1.17<\/strong><\/td><td class=\"has-text-align-center\" data-align=\"center\">9671753<\/td><td class=\"has-text-align-center\" data-align=\"center\">11360839<\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>1.17<\/strong><\/td><\/tr><tr><td>Ergocryptinine<\/td><td class=\"has-text-align-center\" data-align=\"center\">658788<\/td><td class=\"has-text-align-center\" data-align=\"center\">778972<\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>1.18<\/strong><\/td><td class=\"has-text-align-center\" data-align=\"center\">13680420<\/td><td class=\"has-text-align-center\" data-align=\"center\">16765348<\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>1.23<\/strong><\/td><\/tr><tr><td>Ergocornine<\/td><td class=\"has-text-align-center\" data-align=\"center\">370509<\/td><td class=\"has-text-align-center\" data-align=\"center\">387025<\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>1.04<\/strong><\/td><td class=\"has-text-align-center\" data-align=\"center\">7248981<\/td><td class=\"has-text-align-center\" data-align=\"center\">7732744<\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>1.07<\/strong><\/td><\/tr><tr><td>Ergocorninine<\/td><td class=\"has-text-align-center\" data-align=\"center\">590167<\/td><td class=\"has-text-align-center\" data-align=\"center\">704029<\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>1.19<\/strong><\/td><td class=\"has-text-align-center\" data-align=\"center\">12052359<\/td><td class=\"has-text-align-center\" data-align=\"center\">14389283<\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>1.19<\/strong><\/td><\/tr><tr><td>Ergosine<\/td><td class=\"has-text-align-center\" data-align=\"center\">445243<\/td><td class=\"has-text-align-center\" data-align=\"center\">486620<\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>1.09<\/strong><\/td><td class=\"has-text-align-center\" data-align=\"center\">8630932<\/td><td class=\"has-text-align-center\" data-align=\"center\">9366602<\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>1.09<\/strong><\/td><\/tr><tr><td>Ergosinine<\/td><td class=\"has-text-align-center\" data-align=\"center\">439026<\/td><td class=\"has-text-align-center\" data-align=\"center\">496734<\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>1.13<\/strong><\/td><td class=\"has-text-align-center\" data-align=\"center\">7820785<\/td><td class=\"has-text-align-center\" data-align=\"center\">8740527<\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>1.12<\/strong><\/td><\/tr><tr><td>T-2<\/td><td class=\"has-text-align-center\" data-align=\"center\">43286<\/td><td class=\"has-text-align-center\" data-align=\"center\">56535<\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>1.31<\/strong><\/td><td class=\"has-text-align-center\" data-align=\"center\">1046233<\/td><td class=\"has-text-align-center\" data-align=\"center\">1394735<\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>1.33<\/strong><\/td><\/tr><tr><td>HT-2<\/td><td class=\"has-text-align-center\" data-align=\"center\">10183<\/td><td class=\"has-text-align-center\" data-align=\"center\">15221<\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>1.49<\/strong><\/td><td class=\"has-text-align-center\" data-align=\"center\">216703<\/td><td class=\"has-text-align-center\" data-align=\"center\">323765<\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>1.49<\/strong><\/td><\/tr><tr><td>Tentoxin<\/td><td class=\"has-text-align-center\" data-align=\"center\">70973<\/td><td class=\"has-text-align-center\" data-align=\"center\">95175<\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>1.34<\/strong><\/td><td class=\"has-text-align-center\" data-align=\"center\">1577164<\/td><td class=\"has-text-align-center\" data-align=\"center\">2131907<\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>1.35<\/strong><\/td><\/tr><tr><td>Ochratoxin<\/td><td class=\"has-text-align-center\" data-align=\"center\">173686<\/td><td class=\"has-text-align-center\" data-align=\"center\">190060<\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>1.09<\/strong><\/td><td class=\"has-text-align-center\" data-align=\"center\">4039682<\/td><td class=\"has-text-align-center\" data-align=\"center\">4411953<\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>1.09<\/strong><\/td><\/tr><tr><td>Diacetoxyscirpenol<\/td><td class=\"has-text-align-center\" data-align=\"center\">47850<\/td><td class=\"has-text-align-center\" data-align=\"center\">68139<\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>1.42<\/strong><\/td><td class=\"has-text-align-center\" data-align=\"center\">846403<\/td><td class=\"has-text-align-center\" data-align=\"center\">1208826<\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>1.43<\/strong><\/td><\/tr><tr><td>Fusarenone X<\/td><td class=\"has-text-align-center\" data-align=\"center\">3865<\/td><td class=\"has-text-align-center\" data-align=\"center\">7668<\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>1.98<\/strong><\/td><td class=\"has-text-align-center\" data-align=\"center\">60409<\/td><td class=\"has-text-align-center\" data-align=\"center\">121790<\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>2.02<\/strong><\/td><\/tr><tr><td>15-acetyl-DON<\/td><td class=\"has-text-align-center\" data-align=\"center\">17055<\/td><td class=\"has-text-align-center\" data-align=\"center\">31369<\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>1.84<\/strong><\/td><td class=\"has-text-align-center\" data-align=\"center\">269862<\/td><td class=\"has-text-align-center\" data-align=\"center\">517570<\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>1.92<\/strong><\/td><\/tr><tr><td>3-acetyldeoxyvinalenol<\/td><td class=\"has-text-align-center\" data-align=\"center\">13353<\/td><td class=\"has-text-align-center\" data-align=\"center\">22613<\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>1.69<\/strong><\/td><td class=\"has-text-align-center\" data-align=\"center\">179204<\/td><td class=\"has-text-align-center\" data-align=\"center\">296396<\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>1.65<\/strong><\/td><\/tr><tr><td>Aflatoxin G2<\/td><td class=\"has-text-align-center\" data-align=\"center\">171597<\/td><td class=\"has-text-align-center\" data-align=\"center\">262824<\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>1.53<\/strong><\/td><td class=\"has-text-align-center\" data-align=\"center\">3429501<\/td><td class=\"has-text-align-center\" data-align=\"center\">5274354<\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>1.54<\/strong><\/td><\/tr><tr><td>Aflatoxin G1<\/td><td class=\"has-text-align-center\" data-align=\"center\">224058<\/td><td class=\"has-text-align-center\" data-align=\"center\">304389<\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>1.36<\/strong><\/td><td class=\"has-text-align-center\" data-align=\"center\">4607959<\/td><td class=\"has-text-align-center\" data-align=\"center\">6102959<\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>1.32<\/strong><\/td><\/tr><tr><td>ZON<\/td><td class=\"has-text-align-center\" data-align=\"center\">25617<\/td><td class=\"has-text-align-center\" data-align=\"center\">37162<\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>1.45<\/strong><\/td><td class=\"has-text-align-center\" data-align=\"center\">656915<\/td><td class=\"has-text-align-center\" data-align=\"center\">927455<\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>1.41<\/strong><\/td><\/tr><tr><td>Aflatoxin B2<\/td><td class=\"has-text-align-center\" data-align=\"center\">159389<\/td><td class=\"has-text-align-center\" data-align=\"center\">295648<\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>1.85<\/strong><\/td><td class=\"has-text-align-center\" data-align=\"center\">3462489<\/td><td class=\"has-text-align-center\" data-align=\"center\">5724754<\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>1.65<\/strong><\/td><\/tr><tr><td>Aflatoxin B1<\/td><td class=\"has-text-align-center\" data-align=\"center\">265935<\/td><td class=\"has-text-align-center\" data-align=\"center\">223520<\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>0.84<\/strong><\/td><td class=\"has-text-align-center\" data-align=\"center\">5335576<\/td><td class=\"has-text-align-center\" data-align=\"center\">4425821<\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>0.83<\/strong><\/td><\/tr><tr><td>Alpha-zearalenol<\/td><td class=\"has-text-align-center\" data-align=\"center\">16202<\/td><td class=\"has-text-align-center\" data-align=\"center\">30224<\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>1.87<\/strong><\/td><td class=\"has-text-align-center\" data-align=\"center\">382092<\/td><td class=\"has-text-align-center\" data-align=\"center\">702420<\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>1.84<\/strong><\/td><\/tr><tr><td>Deoxynivalenol<\/td><td class=\"has-text-align-center\" data-align=\"center\">6935<\/td><td class=\"has-text-align-center\" data-align=\"center\">17346<\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>2.50<\/strong><\/td><td class=\"has-text-align-center\" data-align=\"center\">117927<\/td><td class=\"has-text-align-center\" data-align=\"center\">281906<\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>2.39<\/strong><\/td><\/tr><tr><td>Nivalenol<\/td><td class=\"has-text-align-center\" data-align=\"center\">1790<\/td><td class=\"has-text-align-center\" data-align=\"center\">4182<\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>2.34<\/strong><\/td><td class=\"has-text-align-center\" data-align=\"center\">25276<\/td><td class=\"has-text-align-center\" data-align=\"center\">64495<\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>2.55<\/strong><\/td><\/tr><tr><td>Altenuene<\/td><td class=\"has-text-align-center\" data-align=\"center\">63224<\/td><td class=\"has-text-align-center\" data-align=\"center\">113850<\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>1.80<\/strong><\/td><td class=\"has-text-align-center\" data-align=\"center\">1187958<\/td><td class=\"has-text-align-center\" data-align=\"center\">2059700<\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>1.73<\/strong><\/td><\/tr><tr><td>Alternariol monomethyl ether<\/td><td class=\"has-text-align-center\" data-align=\"center\">19537<\/td><td class=\"has-text-align-center\" data-align=\"center\">31024<\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>1.59<\/strong><\/td><td class=\"has-text-align-center\" data-align=\"center\">428922<\/td><td class=\"has-text-align-center\" data-align=\"center\">640689<\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>1.49<\/strong><\/td><\/tr><tr><td>Alternariol<\/td><td class=\"has-text-align-center\" data-align=\"center\">48204<\/td><td class=\"has-text-align-center\" data-align=\"center\">73272<\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>1.52<\/strong><\/td><td class=\"has-text-align-center\" data-align=\"center\">837410<\/td><td class=\"has-text-align-center\" data-align=\"center\">1302192<\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>1.56<\/strong><\/td><\/tr><tr><td>Citrinin<\/td><td class=\"has-text-align-center\" data-align=\"center\">499900<\/td><td class=\"has-text-align-center\" data-align=\"center\">1007880<\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>2.02<\/strong><\/td><td class=\"has-text-align-center\" data-align=\"center\">5031182<\/td><td class=\"has-text-align-center\" data-align=\"center\">9828890<\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>1.95<\/strong><\/td><\/tr><tr><td>Tenuazonic acid<\/td><td class=\"has-text-align-center\" data-align=\"center\">21503<\/td><td class=\"has-text-align-center\" data-align=\"center\">47828<\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>2.22<\/strong><\/td><td class=\"has-text-align-center\" data-align=\"center\">89293<\/td><td class=\"has-text-align-center\" data-align=\"center\">197658<\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>2.21<\/strong><\/td><\/tr><\/tbody><\/table><\/figure>\n\n\n\n<p class=\"wp-block-paragraph\">While our pesticides and mycotoxins experiments showed that columns made with inert hardware can improve chromatographic performance, the independent results from Carrera and Wuppermann are even more compelling. Based on the experiences of these well-regarded scientists, in addition to our own testing, it seems clear that Restek\u2019s Inert LC columns have many potential benefits to laboratories analyzing pesticides and mycotoxins. It will be interesting to see what other analyses may be improved as well!<\/p>\n\n\n\n<div style=\"height:100px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n\n\n        <div class=\"cpb\">\n            <h3 class=\"cpb-heading\">Products Mentioned<\/h3>\n            <hr class=\"cpb-heading-underline\" \/>\n            <div class=\"cpb-list\">\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/9309A12-T\">                                Catalog No. 9309A12-T                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">Raptor Inert Biphenyl, 2.7 \u00b5m, 100 x 2.1 mm HPLC Column<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/9309A12-T\">View Product<\/a>\n                                                    <\/div>\n                    <\/div>\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/34122\">                                Catalog No. 34122                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">Ochratoxin A Standard, 10 \u03bcg\/mL, Acetonitrile, 1 mL\/ampul<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/34122\">View Product<\/a>\n                                                    <\/div>\n                    <\/div>\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/34121\">                                Catalog No. 34121                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">Aflatoxins (B1, B2, G1, G2) Standard, 10 \u00b5g\/mL, Acetonitrile, 1 mL\/ampul<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/34121\">View Product<\/a>\n                                                    <\/div>\n                    <\/div>\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/31972\">                                Catalog No. 31972                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">LC Multiresidue Pesticide Standard #1 Organophosphorus Compounds, 100 \u00b5g\/mL, Acetonitrile, 1 mL\/ampul<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/31972\">View Product<\/a>\n                                                    <\/div>\n                    <\/div>\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/9314A12-T\">                                Catalog No. 9314A12-T                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">Raptor Inert ARC-18, 2.7 \u00b5m, 100 x 2.1 mm HPLC Column<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/9314A12-T\">View Product<\/a>\n                                                    <\/div>\n                    <\/div>\n                            <\/div>\n        <\/div>\n        \n\n\n<div style=\"height:100px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n","protected":false},"excerpt":{"rendered":"<p>Explore how inert LC column hardware improves pesticide and mycotoxin analysis.<\/p>\n","protected":false},"author":49,"featured_media":0,"comment_status":"closed","ping_status":"open","sticky":false,"template":"","format":"standard","meta":{"_acf_changed":false,"_kad_blocks_custom_css":"","_kad_blocks_head_custom_js":"","_kad_blocks_body_custom_js":"","_kad_blocks_footer_custom_js":"","_kadence_starter_templates_imported_post":false,"_kad_post_transparent":"","_kad_post_title":"","_kad_post_layout":"","_kad_post_sidebar_id":"","_kad_post_content_style":"","_kad_post_vertical_padding":"","_kad_post_feature":"","_kad_post_feature_position":"","_kad_post_header":false,"_kad_post_footer":false,"footnotes":"","_ppma_block_editor_authors":""},"categories":[9],"tags":[],"industries-application":[2230,2236,2238],"post-badge":[],"resource-type":[],"product-library":[],"resource-technique":[2326,2328],"ppma_author":[589],"class_list":["post-51727","post","type-post","status-publish","format-standard","hentry","category-blogs","industries-application-umweltbereich","industries-application-lebensmittel-getranke","industries-application-lebensmittelsicherheit","resource-technique-flussigchromatografie","resource-technique-massenspektrometrie-ms"],"acf":[],"taxonomy_info":{"category":[{"value":9,"label":"Blogs"}],"industries-application":[{"value":2230,"label":"Umweltbereich"},{"value":2236,"label":"Lebensmittel &amp; Getr\u00e4nke"},{"value":2238,"label":"Lebensmittelsicherheit"}],"resource-technique":[{"value":2326,"label":"Fl\u00fcssigchromatografie"},{"value":2328,"label":"Massenspektrometrie (MS)"}]},"featured_image_src_large":false,"author_info":{"display_name":"Hansj\u00f6rg Majer, PhD","author_link":"https:\/\/discover.restek.com\/es\/author\/hansjoerg-majer\/"},"comment_info":0,"category_info":[{"term_id":9,"name":"Blogs","slug":"blogs","term_group":0,"term_taxonomy_id":9,"taxonomy":"category","description":"","parent":0,"count":457,"filter":"raw","cat_ID":9,"category_count":457,"category_description":"","cat_name":"Blogs","category_nicename":"blogs","category_parent":0}],"tag_info":false,"authors":[{"term_id":589,"user_id":49,"is_guest":0,"slug":"hansjoerg-majer","display_name":"Hansj\u00f6rg Majer, PhD","avatar_url":"https:\/\/secure.gravatar.com\/avatar\/?s=96&d=mm&r=g","author_category":"1","first_name":"Hansj\u00f6rg","last_name":"Majer, PhD","user_url":"","job_title":"","description":""}],"_links":{"self":[{"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/posts\/51727","targetHints":{"allow":["GET"]}}],"collection":[{"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/posts"}],"about":[{"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/types\/post"}],"author":[{"embeddable":true,"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/users\/49"}],"replies":[{"embeddable":true,"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/comments?post=51727"}],"version-history":[{"count":3,"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/posts\/51727\/revisions"}],"predecessor-version":[{"id":73535,"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/posts\/51727\/revisions\/73535"}],"wp:attachment":[{"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/media?parent=51727"}],"wp:term":[{"taxonomy":"category","embeddable":true,"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/categories?post=51727"},{"taxonomy":"post_tag","embeddable":true,"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/tags?post=51727"},{"taxonomy":"industries-application","embeddable":true,"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/industries-application?post=51727"},{"taxonomy":"post-badge","embeddable":true,"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/post-badge?post=51727"},{"taxonomy":"resource-type","embeddable":true,"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/resource-type?post=51727"},{"taxonomy":"product-library","embeddable":true,"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/product-library?post=51727"},{"taxonomy":"resource-technique","embeddable":true,"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/resource-technique?post=51727"},{"taxonomy":"author","embeddable":true,"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/ppma_author?post=51727"}],"curies":[{"name":"wp","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}