{"id":51774,"date":"2021-12-01T13:30:00","date_gmt":"2021-12-01T13:30:00","guid":{"rendered":"https:\/\/discover.restek.com\/uncategorized\/pfas-in-air-part-2-otm-45-chromatography\/"},"modified":"2026-04-06T16:40:04","modified_gmt":"2026-04-06T16:40:04","slug":"pfas-in-air-part-2-otm-45-chromatography","status":"publish","type":"post","link":"https:\/\/discover.restek.com\/es\/blog-fr\/gnbl4768\/pfas-in-air-part-2-otm-45-chromatography","title":{"rendered":"PFAS in Air, Part 2: OTM-45 Chromatography"},"content":{"rendered":"\n<p class=\"wp-block-paragraph\">In my&nbsp;last blog&nbsp;I gave a quick intro on OTM-45, the new PFAS in source air method proposed by the EPA. While the real meat of the method is in the sample collection and extraction sections, it\u2019s worth taking some time to look at the chromatography. After all, without good chromatography sample collection and preparation are hard to judge.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">OTM-45 gives very little guidance for the LC-MS\/MS method. It states that it is fashioned after EPA 533 (one of the PFAS in drinking water methods) and gives an example solvent gradient to use (the same gradient from 533, to be exact), shown in Table 1, but otherwise instruct you to optimize resolution and peak shape.<\/p>\n\n\n\n<figure class=\"wp-block-table\"><table style=\"border-width:1px\"><thead><tr><th>Time (min)<\/th><th>% 20 mM Ammonium acetate<\/th><th>% Methanol<\/th><\/tr><\/thead><tbody><tr><td>Initial<\/td><td>95.0<\/td><td>5.0<\/td><\/tr><tr><td>0.5<\/td><td>95.0<\/td><td>5.0<\/td><\/tr><tr><td>3.0<\/td><td>60.0<\/td><td>40.0<\/td><\/tr><tr><td>16.0<\/td><td>20.0<\/td><td>80.0<\/td><\/tr><tr><td>18.0<\/td><td>20.0<\/td><td>80.0<\/td><\/tr><tr><td>20.0<\/td><td>5.0<\/td><td>95.0<\/td><\/tr><tr><td>22.0<\/td><td>5.0<\/td><td>95.0<\/td><\/tr><tr><td>25.0<\/td><td>95.0<\/td><td>5.0<\/td><\/tr><tr><td>35.0<\/td><td>95.0<\/td><td>5.0<\/td><\/tr><\/tbody><\/table><\/figure>\n\n\n\n<p class=\"wp-block-paragraph\">Table 1 \u2013 OTM-45 example HPLC gradient conditions.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">In an era of conveniences, like next-day shipping, a 35-minute run just seems too long for me. Luckily, it\u2019s easy enough to reduce that. Figure 1 shows the conditions used in this work, which are very similar to our SW-846 8327 method, with a slightly extended time at 95% B to catch some later eluters.<\/p>\n\n\n\n<figure class=\"wp-block-image\"><img decoding=\"async\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-pfas-in-air-part-2-otm-45-chromatography-01.png\" alt=\"blog-OTM-45-blog-2-chromatography-01.png\" title=\"-\"><\/figure>\n\n\n\n<p class=\"wp-block-paragraph\">Fig. 1 \u2013 HPLC-MS\/MS conditions.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Figure 2 shows the chromatogram and retention times for the method. The Force C18 column gives good peak shapes, even for the early eluting PFBA.<\/p>\n\n\n\n<figure class=\"wp-block-image\"><img decoding=\"async\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-pfas-in-air-part-2-otm-45-chromatography-02.png\" alt=\"graphical user interface\" title=\"-\"><\/figure>\n\n\n\n<p class=\"wp-block-paragraph\">Fig. 2 \u2013 Chromatogram and compound retention times for OTM-45 analysis.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">The PFBA peak shape requires careful attention to solvent choice and injection volumes, however. While OTM-45 states in section 10.3 that increasing aqueous content of the extract to better focus early eluters is not permitted, it also states in section 11.2.2 to bring up samples to 2mL after concentration using either DI water or 5% ammonium hydroxide in methanol. As Figure 3 shows, pure methanol injections can lead to split peaks for PFBA. Using an 80:20 methanol:water mix instead (matching the final solvent mix of EPA 533) gives some fronting at 5 \u00b5L, while dropping down to 3 \u00b5L gave a nice, symmetrical peak.<\/p>\n\n\n\n<figure class=\"wp-block-image\"><img decoding=\"async\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-pfas-in-air-part-2-otm-45-chromatography-03.png\" alt=\"shape, rectangle\" title=\"-\"><\/figure>\n\n\n\n<p class=\"wp-block-paragraph\">Fig. 3 \u2013 Trace of PFBA at 5 \u00b5L 100 % MeOH (left), 5 \u00b5L 80:20 MeOH:Water (middle), and 3 \u00b5L 80:20 MeOH:Water (right).<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">There is one downside to this method, and that\u2019s the pressure required. The 1.8 \u00b5m Force column has a peak pressure of about 7000 psi, which might be too much for some HPLC systems. Luckily, switching to a 3.0 \u00b5m Force is an option, giving you pressures closer to 3000 psi, as shown in Figure 4.<\/p>\n\n\n\n<figure class=\"wp-block-image\"><img decoding=\"async\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-pfas-in-air-part-2-otm-45-chromatography-04.png\" alt=\"chart\" title=\"-\"><\/figure>\n\n\n\n<p class=\"wp-block-paragraph\">Fig. 4 \u2013 Pressure for 3.0 \u00b5m Force column (top, maximum pressure ~ 3000 psi) and 1.8 \u00b5m Force column (bottom, maximum pressure ~ 6000 psi).<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">The change from 1.8 to 3.0 \u00b5m has little effect on the retention times. Figure 5 shows that early compounds have their retention time extended by about 6% or so, but it quickly drops off to be almost equal.<\/p>\n\n\n\n<figure class=\"wp-block-image\"><img decoding=\"async\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-pfas-in-air-part-2-otm-45-chromatography-05.png\" alt=\"chart, bar chart\" title=\"-\"><\/figure>\n\n\n\n<p class=\"wp-block-paragraph\">Fig. 5 \u2013 Retention time ratio of 3.0 \u00b5m:1.8 \u00b5m Force columns.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">The price you pay for the larger particle size is a slight decrease of the average peak height. Figure 6 shows a comparison of relative peak heights from a 20 ppb standard, and while some compounds are higher on the 3 \u00b5m, the majority show better peak height, and therefore increased signal\/noise on the 1.8 \u00b5m column. If your system can take the pressure, the 1.8 \u00b5m column will likely help you gain a bit of sensitivity.<\/p>\n\n\n\n<figure class=\"wp-block-image\"><img decoding=\"async\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-pfas-in-air-part-2-otm-45-chromatography-06.png\" alt=\"chart, bar chart\" title=\"-\"><\/figure>\n\n\n\n<p class=\"wp-block-paragraph\">Fig. 6 \u2013 Peak height ratio of 3.0 \u00b5m:1.8 \u00b5m Force columns.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">While the focus of this blog series is OTM-45, the method demonstrated here should be applicable to a number of PFAS methods that have a large compound list, such as the new draft 1633 method. In part 3 of this series I intend to cover the other key part of the instrumental analysis, calibration, so stay tuned for that.<\/p>\n\n\n\n<div style=\"height:100px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n","protected":false},"excerpt":{"rendered":"<p>In my last blog I gave a quick intro on OTM-45, the new PFAS in source air method proposed by the EPA. While the real meat of the method is in the sample collection and extraction sections, it&rsquo;s worth taking some time to look at the chromatography. After all, without good chromatography sample collection and preparation are hard to judge.<\/p>\n","protected":false},"author":27,"featured_media":0,"comment_status":"closed","ping_status":"open","sticky":false,"template":"","format":"standard","meta":{"_acf_changed":false,"_kad_blocks_custom_css":"","_kad_blocks_head_custom_js":"","_kad_blocks_body_custom_js":"","_kad_blocks_footer_custom_js":"","_kadence_starter_templates_imported_post":false,"_kad_post_transparent":"","_kad_post_title":"","_kad_post_layout":"","_kad_post_sidebar_id":"","_kad_post_content_style":"","_kad_post_vertical_padding":"","_kad_post_feature":"","_kad_post_feature_position":"","_kad_post_header":false,"_kad_post_footer":false,"footnotes":"","_ppma_block_editor_authors":""},"categories":[791],"tags":[2782],"industries-application":[2186,2185,2189],"post-badge":[],"resource-type":[],"product-library":[],"resource-technique":[2306,2310,2361],"ppma_author":[447],"class_list":["post-51774","post","type-post","status-publish","format-standard","hentry","category-blog-fr","tag-pfas","industries-application-air-monitoring-zh","industries-application-environmental-zh","industries-application-pfas-testing-zh","resource-technique-air-sampling-zh","resource-technique-liquid-chromatography-zh","resource-technique-ms-ms-zh"],"acf":[],"taxonomy_info":{"category":[{"value":791,"label":"Blog"}],"post_tag":[{"value":2782,"label":"PFAS"}],"industries-application":[{"value":2186,"label":"\u7a7a\u6c14\u76d1\u6d4b"},{"value":2185,"label":"\u73af\u5883"},{"value":2189,"label":"PFAS\u68c0\u6d4b"}],"resource-technique":[{"value":2306,"label":"\u7a7a\u6c14\u91c7\u96c6"},{"value":2310,"label":"\u6db2\u76f8\u8272\u8c31"},{"value":2361,"label":"MS\/MS"}]},"featured_image_src_large":false,"author_info":{"display_name":"Jason Hoisington","author_link":"https:\/\/discover.restek.com\/es\/author\/jason-hoisington\/"},"comment_info":0,"category_info":[{"term_id":791,"name":"Blog","slug":"blog-fr","term_group":0,"term_taxonomy_id":791,"taxonomy":"category","description":"","parent":0,"count":457,"filter":"raw","cat_ID":791,"category_count":457,"category_description":"","cat_name":"Blog","category_nicename":"blog-fr","category_parent":0}],"tag_info":[{"term_id":2782,"name":"PFAS","slug":"pfas","term_group":0,"term_taxonomy_id":2782,"taxonomy":"post_tag","description":"","parent":0,"count":124,"filter":"raw"}],"authors":[{"term_id":447,"user_id":27,"is_guest":0,"slug":"jason-hoisington","display_name":"Jason Hoisington","avatar_url":{"url":"https:\/\/discover.restek.com\/wp-content\/uploads\/people-hoisington-jason-01.jpg","url2x":"https:\/\/discover.restek.com\/wp-content\/uploads\/people-hoisington-jason-01.jpg"},"author_category":"1","first_name":"Jason","last_name":"Hoisington","user_url":"https:\/\/www.linkedin.com\/in\/jason-hoisington-73067653\/","job_title":"Principal Scientist, Sample Preparation","description":"Jason Hoisington received his bachelor\u2019s degree in general science with a focus on chemistry from the University of Alaska, Fairbanks. He worked for SGS Environmental for seven years in environmental soil and water testing, developing methods for the analysis of volatiles and semivolatile organics to include pesticides and polychlorinated biphenyls (PCBs). In 2012, Jason moved on to lab and application support for Dow Chemical Company, providing advanced analytical troubleshooting and method development. In 2019, Jason joined Restek and has focused on air applications."}],"_links":{"self":[{"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/posts\/51774","targetHints":{"allow":["GET"]}}],"collection":[{"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/posts"}],"about":[{"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/types\/post"}],"author":[{"embeddable":true,"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/users\/27"}],"replies":[{"embeddable":true,"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/comments?post=51774"}],"version-history":[{"count":4,"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/posts\/51774\/revisions"}],"predecessor-version":[{"id":92210,"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/posts\/51774\/revisions\/92210"}],"wp:attachment":[{"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/media?parent=51774"}],"wp:term":[{"taxonomy":"category","embeddable":true,"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/categories?post=51774"},{"taxonomy":"post_tag","embeddable":true,"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/tags?post=51774"},{"taxonomy":"industries-application","embeddable":true,"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/industries-application?post=51774"},{"taxonomy":"post-badge","embeddable":true,"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/post-badge?post=51774"},{"taxonomy":"resource-type","embeddable":true,"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/resource-type?post=51774"},{"taxonomy":"product-library","embeddable":true,"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/product-library?post=51774"},{"taxonomy":"resource-technique","embeddable":true,"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/resource-technique?post=51774"},{"taxonomy":"author","embeddable":true,"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/ppma_author?post=51774"}],"curies":[{"name":"wp","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}