{"id":75494,"date":"2024-08-01T16:00:00","date_gmt":"2024-08-01T16:00:00","guid":{"rendered":"https:\/\/discover.restek.com\/?p=75494"},"modified":"2026-01-08T14:25:25","modified_gmt":"2026-01-08T14:25:25","slug":"under-pressure-low-pressure-gas-chromatography-analysis-of-cannabinoid-neutrals","status":"publish","type":"post","link":"https:\/\/discover.restek.com\/es\/blogs\/under-pressure-low-pressure-gas-chromatography-analysis-of-cannabinoid-neutrals\/","title":{"rendered":"Under Pressure? Low-Pressure Gas Chromatography Analysis of Cannabinoid Neutrals"},"content":{"rendered":"\n<p class=\"wp-block-paragraph\">Cannabinoid neutrals are a group of chemical compounds naturally produced by cannabis plants. They are primarily found in the resin produced by the flowers and leaves of the cannabis plant. Some of the most well-known neutrals include delta-8- and delta-9-tetrahydrocannabinol (THC) and cannabidiol (CBD). Respectively, these compounds are known for the psychoactive effects of cannabis and the potential for therapeutic uses.<\/p>\n\n\n<div class=\"wp-block-custom-chromatogram-article-top\"><div class=\"chromatogram-article-placeholder\"><div class=\"figure-heading\"><strong>Figure 1:\u00a0<\/strong>Cannabinoid Neutrals 9 Standard on LPGC Rxi-35Sil MS<\/div><div class='chromatogram-article-inner-full'><div class=\"chromatogram-article-inner\">\n<div class=\"wp-block-custom-chromatogram-article\"><div class=\"wp-block-custom-chromatogram-article\"><div class=\"chromatogram-image regular-image\"><img decoding=\"async\" src=\"https:\/\/ez.restek.com\/images\/cgram\/gc_gn1243.png\" alt=\"Cannabinoid Neutrals 9 Standard on LPGC Rxi-35Sil MS\" title=\"-\"><\/div><p class=\"article-id\" style=\"text-align:center\"> GC_GN1243<\/p><div class=\"chromatogram-peaks\"><h4>Peaks<\/h4><table class=\"peaks col-lg-4 col-12\">\n<thead><tr><th><\/th><th style=\"text-align: left;width: 75px\">Peaks<\/th><th style=\"text-align: center;width: 50px\">t<sub>R<\/sub> (min)<\/th><th style=\"text-align: center;width: 50px\">Conc.<br \/>(&#8230;)<\/th><\/tr><\/thead>\n<tbody><tr><td class=\"num\">1.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Cannabidivarin\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/24274-48-4\/Cannabidivarin\" target=\"_blank\" rel=\"noopener\">Cannabidivarin<\/a><\/td><td class=\"oth\">4.753<\/td><td class=\"oth\">20<\/td><\/tr>\n<tr><td class=\"num\">2.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Tetrahydrocannabivarin\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/31262-37-0\/Tetrahydrocannabivarin\" target=\"_blank\" rel=\"noopener\">Tetrahydrocannabivarin<\/a><\/td><td class=\"oth\">5.013<\/td><td class=\"oth\">20<\/td><\/tr>\n<tr><td class=\"num\">3.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Cannabicyclol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/21366-63-2\/Cannabicyclol\" target=\"_blank\" rel=\"noopener\">Cannabicyclol<\/a><\/td><td class=\"oth\">5.057<\/td><td class=\"oth\">20<\/td><\/tr>\n<tr><td class=\"num\">4.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Cannabichromene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/20675-51-8\/Cannabichromene\" target=\"_blank\" rel=\"noopener\">Cannabichromene<\/a><\/td><td class=\"oth\">5.142<\/td><td class=\"oth\">20<\/td><\/tr>\n<tr><td class=\"num\">5.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Cannabidiol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/13956-29-1\/Cannabidiol\" target=\"_blank\" rel=\"noopener\">Cannabidiol<\/a><\/td><td class=\"oth\">5.189<\/td><td class=\"oth\">20<\/td><\/tr>\n<tr><td class=\"num\">6.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for d8-Tetrahydrocannabinol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/5957-75-5\/d8-Tetrahydrocannabinol\" target=\"_blank\" rel=\"noopener\">d8-Tetrahydrocannabinol<\/a><\/td><td class=\"oth\">5.367<\/td><td class=\"oth\">20<\/td><\/tr>\n<tr><td class=\"num\">7.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for d9-Tetrahydrocannabinol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/1972-08-3\/d9-Tetrahydrocannabinol\" target=\"_blank\" rel=\"noopener\">d9-Tetrahydrocannabinol<\/a><\/td><td class=\"oth\">5.433<\/td><td class=\"oth\">20<\/td><\/tr>\n<tr><td class=\"num\">8.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Cannabigerol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/25654-31-3\/Cannabigerol\" target=\"_blank\" rel=\"noopener\">Cannabigerol<\/a><\/td><td class=\"oth\">5.487<\/td><td class=\"oth\">20<\/td><\/tr>\n<tr><td class=\"num\">9.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Cannabinol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/521-35-7\/Cannabinol\" target=\"_blank\" rel=\"noopener\">Cannabinol<\/a><\/td><td class=\"oth\">5.590<\/td><td class=\"oth\">20<\/td><\/tr>\n<\/tbody><\/table><\/div><div class=\"chromatogram-conditions\"><h4>Conditions<\/h4><div class=\"conditions-container container-fluid\"><div class=\"row\"><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Column<\/th><td>LPGC Rxi-35Sil MS, includes Rxi-35mSil MS 10 m 0.32 mm 0.25 \u00b5m w\/ 5  0.15 mm ID Rxi restrictor connected via SilTite connector<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Standard\/Sample<\/th><td>Cannabinoids Neutrals 9 Standard  (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/34132\" rel=\"noopener\">cat.# 34132<\/a>)<\/td><\/tr>\n<tr><th class=\"sub conditions_header\" scope=\"row\">Diluent:<\/th><td>Methanol<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Conc.:<\/th><td>100 \u00b5g\/mL (20 \u00b5g\/mL on column)<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Injection<\/th><td><\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Inj. Vol.:<\/th><td>1 \u00b5L split (split ratio 5:1)<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Liner:<\/th><td>Topaz, splitless, single taper inlet liner, 4.0 mm x 6.5 x 78.5 (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/23303\" rel=\"noopener\">cat.# 23303<\/a>)<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Inj. Temp.:<\/th><td>290 \u00b0C<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Split Vent Flow Rate:<\/th><td>4.5 mL\/min<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\" colspan=\"2\">Oven<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Oven Temp.:<\/th><td>80 &#176;C to 300 &#176;C at 40 &#176;C\/min (hold 1 min)<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Carrier Gas<\/th><td>He, constant flow<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Flow Rate:<\/th><td>0.9 mL\/min<\/td><\/tr><\/table><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Detector<\/th><td>MS<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Mode:<\/th><td>Scan<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Scan Program:<\/th><td><\/td><\/tr><tr><td style=\"text-align:left\" colspan=\"2\"><table class=\"conditions_sim\"><thead><tr><th style=\"padding: 0 5px\">Group<\/th><th style=\"padding: 0 5px\">Start Time<br \/>(min)<\/th><th style=\"padding: 0 5px\">Scan Range<br \/>(amu)<\/th><th style=\"padding: 0 5px\">Scan Rate<br \/>(scans\/sec)<\/th><\/tr><\/thead><tbody><tr><td>1<\/td><td>1.5<\/td><td>30-550<\/td><td>5.3<\/td><\/tr><\/tbody><\/table><\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Transfer Line Temp.:<\/th><td>300 \u00b0C<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Analyzer Type:<\/th><td>Quadrupole <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Source Temp.:<\/th><td>330 \u00b0C<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Quad Temp.:<\/th><td>180 \u00b0C<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Solvent Delay Time:<\/th><td>1.5  min<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Tune Type:<\/th><td>PFTBA <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Ionization Mode:<\/th><td>EI <\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Instrument<\/th><td>Agilent 7890A GC &amp; 5975C MSD<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Sample Preparation<\/th><td>The sample was in a 2 mL short-cap, screw-thread vial (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/21143\" rel=\"noopener\">cat.# 21143<\/a>) and capped with a short-cap, screw-vial closure (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/24495\" rel=\"noopener\">cat.# 24495<\/a>).<\/td><\/tr><\/table><\/div><\/div><\/div><div class=\"chromatogram-related-skus\"><h4>Related SKUs<\/h4><ul><li>34132<\/li><li>23303<\/li><li>21143<\/li><li>24495<\/li><\/ul><\/div><div class=\"chromatogram-pdf-link\"><a href=\"https:\/\/ez.restek.com\/images\/cgram\/gc_gn1243.pdf\" target=\"_blank\" rel=\"noopener noreferrer\"><svg xmlns=\"http:\/\/www.w3.org\/2000\/svg\" width=\"18\" height=\"18\" viewBox=\"0 0 18 18\"><g data-name=\"Group 2996\"><path data-name=\"Rectangle 1246\" d=\"M0 0h18v18H0z\" style=\"fill: none;\"><\/path><\/g><g data-name=\"Group 2997\"><path data-name=\"Path 729\" d=\"M13.412 11.4v2.017H5.345V11.4H4v2.017a1.349 1.349 0 0 0 1.345 1.345h8.068a1.349 1.349 0 0 0 1.345-1.345V11.4zm-.672-2.694-.948-.948-1.741 1.735V4H8.706v5.493L6.965 7.758l-.948.948 3.361 3.361z\" transform=\"translate(-.437 -.414)\" style=\"fill: rgb(13, 123, 196);\"><\/path><\/g><\/svg>Download PDF<\/a><\/div><\/div><\/div>\n<\/div><\/div><\/div>\n\n\n<p class=\"wp-block-paragraph\">This analysis was completed using low pressure gas chromatography mass spectrometry (LPGC GC-MS). LPGC is a great technique to pair with a mass spectrometer detector in the confirmation and identification of various compounds. LPGC uses a deactivated microbore column joined via a connector to a medium bore or megabore analytical column. With fast temperature programming and reduced flow rates, LPGC allows for increased optimum linear velocity. &nbsp;Using this technique can result in faster GC run times, increased sensitivity, and taller, narrower peaks.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">The cannabinoid neutrals standard was run on an LPGC kit consisting of an Rxi-35Sil MS analytical column (10 m \u00d7 0.32 mm \u00d7 0.25 \u00b5m) with a 5 m \u00d7 0.15 mm ID Rxi restrictor connected via SilTite connector. The Rxi-35Sil MS is an excellent choice for cannabinoid analysis due to its selectivity and increased thermal stability. It is also compatible with a diverse range of analytes, including polar and nonpolar compounds, making it versatile for different analytical applications.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">The image below shows the difference between a conventional GC-MS (black) analysis of cannabinoid neutrals and the LPGC (red) analysis. By using LPGC, the first eluting compound\u2014cannabidivarin (CBDV)\u2014comes out six minutes sooner than the conventional analysis times. Run times are reduced to six minutes with LPGC as the last compound\u2014cannabinol (CBN)\u2014measured against the thirteen minutes needed for a regular GC analysis. Despite the reduced separation efficiency, LPGC was able to provide satisfactory resolution for these analytes.<\/p>\n\n\n<style>.kb-image75494_98d131-76 .kb-image-has-overlay:after{opacity:0.3;}<\/style>\n<div class=\"wp-block-kadence-image kb-image75494_98d131-76\"><figure class=\"aligncenter size-full\"><img loading=\"lazy\" decoding=\"async\" width=\"960\" height=\"325\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/figure-blog-jessi-LPGC-comparison-1.png\" alt=\"\" class=\"kb-img wp-image-75496\" title=\"-\" srcset=\"https:\/\/discover.restek.com\/wp-content\/uploads\/figure-blog-jessi-LPGC-comparison-1.png 960w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure-blog-jessi-LPGC-comparison-1-300x102.png 300w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure-blog-jessi-LPGC-comparison-1-768x260.png 768w\" sizes=\"auto, (max-width: 960px) 100vw, 960px\" \/><\/figure><\/div>\n\n\n<figure class=\"wp-block-table\">\n<table class=\"has-fixed-layout\">\n<tbody>\n<tr style=\"text-align: center; background-color: #3598db; color: #ffffff;\">\n<td colspan=\"6\"><strong>Cannabinoid Neutrals\u00a0<\/strong><\/td>\n<\/tr>\n<tr style=\"background-color: #c2e0f4;\">\n<td><strong>Number<\/strong><\/td>\n<td><strong>Compound<\/strong><\/td>\n<td><strong>LPGC-MS RT<\/strong><\/td>\n<td><strong>Conventional GC-MS RT<\/strong><\/td>\n<td><strong>Primary Ion(s)<\/strong><\/td>\n<td><strong>Resolution*<\/strong><\/td>\n<\/tr>\n<tr>\n<td>1<\/td>\n<td>Cannabidivarin (CBDV)<\/td>\n<td>4.753<\/td>\n<td>11.06<\/td>\n<td>203<\/td>\n<td>8.29<\/td>\n<\/tr>\n<tr>\n<td>2<\/td>\n<td>Tetrahydrocannabivarin (THCV)<\/td>\n<td>5.013<\/td>\n<td>11.564<\/td>\n<td>271<\/td>\n<td>1.37<\/td>\n<\/tr>\n<tr>\n<td>3<\/td>\n<td>Cannabicyclol (CBL)<\/td>\n<td>5.057<\/td>\n<td>11.652<\/td>\n<td>231<\/td>\n<td>2.57<\/td>\n<\/tr>\n<tr>\n<td>4<\/td>\n<td>Cannabichromene (CBC)<\/td>\n<td>5.142<\/td>\n<td>11.793<\/td>\n<td>231, 174<\/td>\n<td>1.46<\/td>\n<\/tr>\n<tr>\n<td>5<\/td>\n<td>Cannabidiol (CBD)<\/td>\n<td>5.189<\/td>\n<td>11.9<\/td>\n<td>231, 246<\/td>\n<td>5.68<\/td>\n<\/tr>\n<tr>\n<td>6<\/td>\n<td>d8-Tetrahydrocannabinol (d8-THC)<\/td>\n<td>5.367<\/td>\n<td>12.248<\/td>\n<td>231, 314<\/td>\n<td>2.05<\/td>\n<\/tr>\n<tr>\n<td>7<\/td>\n<td>d9-Tetrahydrocannabinol (d9-THC)<\/td>\n<td>5.433<\/td>\n<td>12.373<\/td>\n<td>299, 315<\/td>\n<td>1.77<\/td>\n<\/tr>\n<tr>\n<td>8<\/td>\n<td>Cannabigerol (CBG)<\/td>\n<td>5.487<\/td>\n<td>12.451<\/td>\n<td>193<\/td>\n<td>3.28<\/td>\n<\/tr>\n<tr>\n<td>9<\/td>\n<td>Cannabinol (CBN)<\/td>\n<td>5.590<\/td>\n<td>12.677<\/td>\n<td>295<\/td>\n<td>5.60<\/td>\n<\/tr>\n<\/tbody>\n<\/table>\n<\/figure>\n\n\n<p class=\"wp-block-paragraph\">*LPGC only.\u00a0<\/p>\n\n\n\n<div style=\"height:30px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n\n\n        <div class=\"cpb\">\n            <h3 class=\"cpb-heading\">Related Products<\/h3>\n            <hr class=\"cpb-heading-underline\" \/>\n            <div class=\"cpb-list\">\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/11806\">                                Catalog No. 11806                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">LPGC Rxi-35Sil MS Column Kit<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/11806\">View Product<\/a>\n                                                    <\/div>\n                    <\/div>\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/23303\">                                Catalog No. 23303                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">Topaz, Single Taper Inlet Liner, 4.0 mm x 6.5 x 78.5, for Agilent GCs, w\/Quartz Wool, Premium Deactivation, 5-pk.<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/23303\">View Product<\/a>\n                                                    <\/div>\n                    <\/div>\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/21143\">                                Catalog No. 21143                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">Short-Cap Vial with Grad Marking Spot, 9-425 Screw-Thread, 2 mL, 9 mm, 12 x 32 (vial only), Amber, 1000-pk.<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/21143\">View Product<\/a>\n                                                    <\/div>\n                    <\/div>\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/24495\">                                Catalog No. 24495                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">Short Screw Caps, Polypropylene, Screw-Thread, PTFE\/Silicone\/PTFE Septa, Black, Preassembled, 2.0 mL, 9 mm, 100-pk.<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/24495\">View Product<\/a>\n                                                    <\/div>\n                    <\/div>\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/34132\">                                Catalog No. 34132                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">Cannabinoids Neutrals 9 Standard, 1000 \u00b5g\/mL, P&amp;T Methanol, 1 mL\/ampul<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/34132\">View Product<\/a>\n                                                    <\/div>\n                    <\/div>\n                            <\/div>\n        <\/div>\n        \n\n\n<div style=\"height:100px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n","protected":false},"excerpt":{"rendered":"<p>Cannabinoid neutrals are a group of chemical compounds naturally produced by cannabis plants. They are primarily found in the resin produced by the flowers and leaves of the cannabis plant. Some of the most well-known neutrals include delta-8- and delta-9-tetrahydrocannabinol (THC) and cannabidiol (CBD). Respectively, these compounds are known for the psychoactive effects of cannabis&#8230;<\/p>\n","protected":false},"author":18,"featured_media":0,"comment_status":"closed","ping_status":"open","sticky":false,"template":"","format":"standard","meta":{"_acf_changed":false,"_kad_blocks_custom_css":"","_kad_blocks_head_custom_js":"","_kad_blocks_body_custom_js":"","_kad_blocks_footer_custom_js":"","_kadence_starter_templates_imported_post":false,"_kad_post_transparent":"","_kad_post_title":"","_kad_post_layout":"","_kad_post_sidebar_id":"","_kad_post_content_style":"","_kad_post_vertical_padding":"","_kad_post_feature":"","_kad_post_feature_position":"","_kad_post_header":false,"_kad_post_footer":false,"footnotes":"","_ppma_block_editor_authors":""},"categories":[9],"tags":[],"industries-application":[],"post-badge":[],"resource-type":[],"product-library":[],"resource-technique":[],"ppma_author":[438],"class_list":["post-75494","post","type-post","status-publish","format-standard","hentry","category-blogs"],"acf":[],"taxonomy_info":{"category":[{"value":9,"label":"Blogs"}]},"featured_image_src_large":false,"author_info":{"display_name":"Jessi Collier","author_link":"https:\/\/discover.restek.com\/es\/author\/jessi-collier\/"},"comment_info":0,"category_info":[{"term_id":9,"name":"Blogs","slug":"blogs","term_group":0,"term_taxonomy_id":9,"taxonomy":"category","description":"","parent":0,"count":457,"filter":"raw","cat_ID":9,"category_count":457,"category_description":"","cat_name":"Blogs","category_nicename":"blogs","category_parent":0}],"tag_info":false,"authors":[{"term_id":438,"user_id":18,"is_guest":0,"slug":"jessi-collier","display_name":"Jessi Collier","avatar_url":{"url":"https:\/\/discover.restek.com\/wp-content\/uploads\/people-collier-jessi.jpg","url2x":"https:\/\/discover.restek.com\/wp-content\/uploads\/people-collier-jessi.jpg"},"author_category":"1","first_name":"Jessi","last_name":"Collier","user_url":"https:\/\/www.linkedin.com\/in\/jessi-collier\/","job_title":"Advanced Scientist, GC Applied","description":"Jessi is an advanced scientist in the GC applications group where her work focuses on environmental and food contaminants. Prior to joining Restek in 2023, she worked for the Mississippi State Chemical Lab as a pesticide residue analyst, primarily testing regulatory samples, agricultural samples, environmental samples, foods, and other consumer products via GC-MS, GC-MS\/MS, GC-FID, and GC-ECD. Jessi holds a BS in chemistry and a BA in Mandarin Chinese from Mississippi State University."}],"_links":{"self":[{"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/posts\/75494","targetHints":{"allow":["GET"]}}],"collection":[{"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/posts"}],"about":[{"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/types\/post"}],"author":[{"embeddable":true,"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/users\/18"}],"replies":[{"embeddable":true,"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/comments?post=75494"}],"version-history":[{"count":1,"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/posts\/75494\/revisions"}],"predecessor-version":[{"id":75502,"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/posts\/75494\/revisions\/75502"}],"wp:attachment":[{"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/media?parent=75494"}],"wp:term":[{"taxonomy":"category","embeddable":true,"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/categories?post=75494"},{"taxonomy":"post_tag","embeddable":true,"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/tags?post=75494"},{"taxonomy":"industries-application","embeddable":true,"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/industries-application?post=75494"},{"taxonomy":"post-badge","embeddable":true,"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/post-badge?post=75494"},{"taxonomy":"resource-type","embeddable":true,"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/resource-type?post=75494"},{"taxonomy":"product-library","embeddable":true,"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/product-library?post=75494"},{"taxonomy":"resource-technique","embeddable":true,"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/resource-technique?post=75494"},{"taxonomy":"author","embeddable":true,"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/ppma_author?post=75494"}],"curies":[{"name":"wp","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}