{"id":76421,"date":"2018-01-14T00:00:00","date_gmt":"2018-01-14T00:00:00","guid":{"rendered":"https:\/\/discover.restek.com\/?p=76421"},"modified":"2025-12-26T15:39:09","modified_gmt":"2025-12-26T15:39:09","slug":"gc-compounds-poor-peak-shapes-and-missing-peaks","status":"publish","type":"post","link":"https:\/\/discover.restek.com\/es\/blogs\/gnbl5495\/gc-compounds-poor-peak-shapes-and-missing-peaks","title":{"rendered":"GC compounds &#8211; poor peak shapes and missing peaks"},"content":{"rendered":"\n<p class=\"wp-block-paragraph\">One of the most common questions I am asked by customers is \u201cWhy is my peak shape so poor?\u201d Another is \u201cI do not see a peak for my compound, what should I do?\u201d<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Sometimes they are developing a new method and have no experience with the compound(s) of interest; other times things worked well in the past and now they don\u2019t. Depending upon the exact circumstances my questions\/suggestions may vary somewhat, but the first thing I usually do is to ask a few questions (Q) to help me narrow down what may be the issue.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Q1.&nbsp; Has this analysis worked well on this specific instrument in the past?&nbsp; If so, when was the last time?&nbsp; What has changed from when it worked well until now?&nbsp; Has the instrument recently been serviced or had extremely \u201cdirty\u201d samples analyzed on it?<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Q2. Has this analysis worked well on any instrument in the lab in the past?&nbsp; If so, when and on which instrument make\/model?&nbsp; Is the current GC similarly equipped to the previous model?<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Q3. Is this a new analysis&nbsp;for this lab?&nbsp; If so, what method is being followed?&nbsp; Is it a common, well-publicized method like those from EPA, ASTM, USP, NIOSH, etc., or is it an obscure method that was found searching the web?<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Q4. How does one know for certain that this particular analysis is going to work well on a specific instrument?&nbsp; In other words, have you found any data\/chromatogram\/method showing that this analysis should work well (or is even possible) on your instrument?<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Some of the questions listed above may seem somewhat repetitive, but they really are not. Depending upon the answers to the questions above, possible suggestions (S) are listed below.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">S1. If a specific instrument has provided good peak shape(s) in the past, we would suspect that it should be able to do so again.&nbsp; I suggest changing the column, cleaning the injection port, replacing the injection port liner, cleaning and\/or replacing the detector and associated consumables, verifying all gas flows, etc.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">S2. If this analysis has not been performed successfully on this particular instrument, but it has on other similarly-equipped instruments, I suggest trying to perform this analysis on the instrument where it last worked.&nbsp; Assuming this instrument is working well, if the same issues occur, there likely is something else going wrong that is not related to a single instrument.&nbsp; Verify your chemical reference standard is OK.&nbsp; Verify the instrument gases are clean.&nbsp; Look at whatever these two instruments have in common and investigate further.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">S3. If this analysis has never been performed in your lab but you are following a method, I suggest contacting the author(s) or agency that developed the method.&nbsp; It\u2019s possible they may have heard of similar issues and be able to quickly provide you a solution.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">S4. If you are not certain that your instrument is capable of performing a particular analysis, I suggest searching for chromatograms\/methods showing the analysis using a similarly-equipped instrument. You need to make sure your peak of interest isn\u2019t eluting under the solvent peak (if a solvent is involved), co-eluting with other peaks, and that you have allowed enough time for your compound to elute from the column.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">In summary, by reviewing the answers to these questions, it usually becomes easier to determine a logical approach to finding a solution. For example, if you know the analysis has worked in the past, you know it should be able to work in the future, and a methodical step-by-step&nbsp;troubleshooting&nbsp;approach will usually provide the solution.&nbsp; But if you are uncertain if a particular analysis will work using your column\/instrument\/detector, your time may be better spent&nbsp;researching&nbsp;to make certain you have chosen the proper column &amp; instrument &amp; detector.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">If you have any questions, feel free to send me an email.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">You may find the related articles helpful.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Thank you for your time.<\/p>\n\n\n\n<div style=\"height:100px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n","protected":false},"excerpt":{"rendered":"<p>One of the most common questions I am asked by customers is &ldquo;Why is my peak shape so poor?&rdquo; Another is &ldquo;I do not see a peak for my compound, what should I do?&rdquo; Sometimes they are developing a new&#8230;<\/p>\n","protected":false},"author":3,"featured_media":0,"comment_status":"closed","ping_status":"open","sticky":false,"template":"","format":"standard","meta":{"_acf_changed":false,"_kad_blocks_custom_css":"","_kad_blocks_head_custom_js":"","_kad_blocks_body_custom_js":"","_kad_blocks_footer_custom_js":"","_kadence_starter_templates_imported_post":false,"_kad_post_transparent":"","_kad_post_title":"","_kad_post_layout":"","_kad_post_sidebar_id":"","_kad_post_content_style":"","_kad_post_vertical_padding":"","_kad_post_feature":"","_kad_post_feature_position":"","_kad_post_header":false,"_kad_post_footer":false,"footnotes":"","_ppma_block_editor_authors":""},"categories":[9],"tags":[],"industries-application":[],"post-badge":[],"resource-type":[],"product-library":[],"resource-technique":[],"ppma_author":[582],"class_list":["post-76421","post","type-post","status-publish","format-standard","hentry","category-blogs"],"acf":[],"taxonomy_info":{"category":[{"value":9,"label":"Blogs"}]},"featured_image_src_large":false,"author_info":{"display_name":"Audrey Crater","author_link":"https:\/\/discover.restek.com\/es\/author\/audrey-craterrestek-com\/"},"comment_info":0,"category_info":[{"term_id":9,"name":"Blogs","slug":"blogs","term_group":0,"term_taxonomy_id":9,"taxonomy":"category","description":"","parent":0,"count":458,"filter":"raw","cat_ID":9,"category_count":458,"category_description":"","cat_name":"Blogs","category_nicename":"blogs","category_parent":0}],"tag_info":false,"authors":[{"term_id":582,"user_id":0,"is_guest":1,"slug":"alan-sensue","display_name":"Alan Sensue","avatar_url":"https:\/\/secure.gravatar.com\/avatar\/?s=96&d=mm&r=g","author_category":"1","first_name":"Alan","last_name":"Sensue","user_url":"","job_title":"","description":""}],"_links":{"self":[{"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/posts\/76421","targetHints":{"allow":["GET"]}}],"collection":[{"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/posts"}],"about":[{"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/types\/post"}],"author":[{"embeddable":true,"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/users\/3"}],"replies":[{"embeddable":true,"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/comments?post=76421"}],"version-history":[{"count":1,"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/posts\/76421\/revisions"}],"predecessor-version":[{"id":77841,"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/posts\/76421\/revisions\/77841"}],"wp:attachment":[{"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/media?parent=76421"}],"wp:term":[{"taxonomy":"category","embeddable":true,"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/categories?post=76421"},{"taxonomy":"post_tag","embeddable":true,"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/tags?post=76421"},{"taxonomy":"industries-application","embeddable":true,"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/industries-application?post=76421"},{"taxonomy":"post-badge","embeddable":true,"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/post-badge?post=76421"},{"taxonomy":"resource-type","embeddable":true,"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/resource-type?post=76421"},{"taxonomy":"product-library","embeddable":true,"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/product-library?post=76421"},{"taxonomy":"resource-technique","embeddable":true,"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/resource-technique?post=76421"},{"taxonomy":"author","embeddable":true,"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/ppma_author?post=76421"}],"curies":[{"name":"wp","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}