{"id":76443,"date":"2013-07-08T00:00:00","date_gmt":"2013-07-08T00:00:00","guid":{"rendered":"https:\/\/discover.restek.com\/?p=76443"},"modified":"2026-02-05T20:41:54","modified_gmt":"2026-02-05T20:41:54","slug":"need-some-help-in-choosing-the-correct-gas-regulator","status":"publish","type":"post","link":"https:\/\/discover.restek.com\/es\/blogs\/gnbl5519\/need-some-help-in-choosing-the-correct-gas-regulator","title":{"rendered":"Need some help in choosing the correct gas regulator?"},"content":{"rendered":"\n<p class=\"wp-block-paragraph\">I have talked to many customers recently who asked for some help in choosing a gas regulator.&nbsp; As a result, I decided to write this post which will specify what you need to know and how to locate the product you need.&nbsp;&nbsp; Please note that this post&nbsp;<strong>will only focus on regulators designed to be used with laboratory gas cylinders\/tanks which use CGA* fittings.<\/strong><\/p>\n\n\n\n<p class=\"wp-block-paragraph\">* = Compressed Gas Association<\/p>\n\n\n<style>.kb-image76443_b49bd6-a2:not(.kb-image-is-ratio-size) .kb-img, .kb-image76443_b49bd6-a2.kb-image-is-ratio-size{padding-top:var(--global-kb-spacing-xs, 1rem);padding-bottom:var(--global-kb-spacing-xs, 1rem);}.kb-image76443_b49bd6-a2 .kb-image-has-overlay:after{opacity:0.3;}<\/style>\n<figure class=\"wp-block-kadence-image kb-image76443_b49bd6-a2 size-medium_large\"><img loading=\"lazy\" decoding=\"async\" width=\"768\" height=\"465\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-need-some-help-in-choosing-the-correct-gas-regulator-01-768x465.jpg\" alt=\"\" class=\"kb-img wp-image-78376\" title=\"-\" srcset=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-need-some-help-in-choosing-the-correct-gas-regulator-01-768x465.jpg 768w, https:\/\/discover.restek.com\/wp-content\/uploads\/blog-need-some-help-in-choosing-the-correct-gas-regulator-01-300x182.jpg 300w, https:\/\/discover.restek.com\/wp-content\/uploads\/blog-need-some-help-in-choosing-the-correct-gas-regulator-01.jpg 861w\" sizes=\"auto, (max-width: 768px) 100vw, 768px\" \/><\/figure>\n\n\n\n<p class=\"wp-block-paragraph\">&nbsp;<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">So what do you need to know before choosing a Restek regulator?<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">&nbsp;1.&nbsp;&nbsp;<a data-uw-rm-brl=\"false\" href=\"http:\/\/www.concoa.com\/cgachart.html\" target=\"_blank\" rel=\"noopener\">CGA Fittings<\/a><\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Just knowing which gas you need to regulate is often not enough.&nbsp; For example, compressed&nbsp;<strong><em>air<\/em>&nbsp;<\/strong>cylinders\/tanks can have multiple CGA fittings, including 346, 347, 590, and 702.&nbsp; Restek only sells regulators with CGA 590.&nbsp;&nbsp;<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">2. &nbsp;Incoming gas pressure<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Always check that the regulator is able to handle the expected inlet pressure.&nbsp; Restek regulators are capable of handling 3000psi (inlet gauge displays 0 &#8211; 4000psi).<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">&nbsp;3. &nbsp;Exiting gas pressure required<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">What pressure is needed for your instrument or equipment?&nbsp;&nbsp;Many Restek regulators are designed for a maximum outlet pressure of 150psi (outlet gauge displays 30&#8243; &#8211; 200psi).<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">&nbsp;4. &nbsp;Regulator material<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Regulators from Restek are available in Chrome-Plated Brass, Stainless Steel or Sulfinert-treated.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Chrome-Plated Brass is usually chosen because of its lower cost, and stainless steel because of its resistance to corrosive gases.&nbsp; Sulfinert treatment adds extra protection and inertness to stainless steel.&nbsp;<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">&nbsp;5. &nbsp;Dual-Stage or Single Stage.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">A dual-stage regulator is used when the outlet pressure needs to be very consistent.&nbsp; You can use a single-stage regulator if the output pressure can vary somewhat.&nbsp; &nbsp;<\/p>\n\n\n\n<p class=\"wp-block-paragraph\"><a href=\"https:\/\/www.concoa.com\/applications\/how-to-choose-a-regulator\" data-type=\"link\" data-id=\"https:\/\/www.concoa.com\/applications\/how-to-choose-a-regulator\" target=\"_blank\" rel=\"noopener\">How to Choose a Regulator<\/a><\/p>\n\n\n\n<p class=\"wp-block-paragraph\">&nbsp;Once you know the answers to the questions above, you can start looking for a regulator which will meet your needs.&nbsp; Below are links which will take you directly to the products on our website.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">&nbsp;<strong>Please Note:<\/strong>&nbsp; All of our regulators are designed to connect to 1\/4 inch OD (outside diameter) tubing.&nbsp; If you plan to connect 1\/8 inch tubing, please read my post &#8220;<a href=\"https:\/\/discover.restek.com\/gnbl4707\/how-to-connect-18-inch-tubing-to-a-restek-gas-regulator\/\" data-type=\"post\" data-id=\"51716\">How to connect 1\/8 inch tubing to a Restek gas regulator<\/a>.&#8221;<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">If we do not offer a regulator which meets your specifications, we may be able to provide you with a custom product.&nbsp; Please complete and submit our &#8220;<a href=\"https:\/\/forms.restek.com\/223205147122947\" target=\"_blank\" rel=\"noopener\">Custom Ultra-High Purity Gas Regulator Quote Form<\/a>.&#8221;<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">General&nbsp;<a href=\"http:\/\/www.concoa.com\/faqs.html\" data-uw-rm-brl=\"false\" target=\"_blank\" rel=\"noopener\">FAQs<\/a><\/p>\n\n\n\n<p class=\"wp-block-paragraph\">I hope you have found this post helpful in choosing the correct regulator for your laboratory gas cylinders\/tanks.&nbsp; Thanks for reading.<\/p>\n\n\n\n<div style=\"height:100px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n\n\n\n<p class=\"wp-block-paragraph\"><\/p>\n","protected":false},"excerpt":{"rendered":"<p>I have talked to many customers recently who asked for some help in choosing a gas regulator. As a result, I decided to write this post which will specify what you need to know and how to locate the product&#8230;<\/p>\n","protected":false},"author":14,"featured_media":0,"comment_status":"closed","ping_status":"open","sticky":false,"template":"","format":"standard","meta":{"_acf_changed":false,"_kad_blocks_custom_css":"","_kad_blocks_head_custom_js":"","_kad_blocks_body_custom_js":"","_kad_blocks_footer_custom_js":"","_kadence_starter_templates_imported_post":false,"_kad_post_transparent":"","_kad_post_title":"","_kad_post_layout":"","_kad_post_sidebar_id":"","_kad_post_content_style":"","_kad_post_vertical_padding":"","_kad_post_feature":"","_kad_post_feature_position":"","_kad_post_header":false,"_kad_post_footer":false,"footnotes":"","_ppma_block_editor_authors":""},"categories":[9],"tags":[],"industries-application":[],"post-badge":[],"resource-type":[],"product-library":[],"resource-technique":[],"ppma_author":[582],"class_list":["post-76443","post","type-post","status-publish","format-standard","hentry","category-blogs"],"acf":[],"taxonomy_info":{"category":[{"value":9,"label":"Blogs"}]},"featured_image_src_large":false,"author_info":{"display_name":"Daniel","author_link":"https:\/\/discover.restek.com\/es\/author\/daniel-voralia\/"},"comment_info":0,"category_info":[{"term_id":9,"name":"Blogs","slug":"blogs","term_group":0,"term_taxonomy_id":9,"taxonomy":"category","description":"","parent":0,"count":458,"filter":"raw","cat_ID":9,"category_count":458,"category_description":"","cat_name":"Blogs","category_nicename":"blogs","category_parent":0}],"tag_info":false,"authors":[{"term_id":582,"user_id":0,"is_guest":1,"slug":"alan-sensue","display_name":"Alan Sensue","avatar_url":"https:\/\/secure.gravatar.com\/avatar\/?s=96&d=mm&r=g","author_category":"1","first_name":"Alan","last_name":"Sensue","user_url":"","job_title":"","description":""}],"_links":{"self":[{"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/posts\/76443","targetHints":{"allow":["GET"]}}],"collection":[{"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/posts"}],"about":[{"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/types\/post"}],"author":[{"embeddable":true,"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/users\/14"}],"replies":[{"embeddable":true,"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/comments?post=76443"}],"version-history":[{"count":4,"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/posts\/76443\/revisions"}],"predecessor-version":[{"id":87102,"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/posts\/76443\/revisions\/87102"}],"wp:attachment":[{"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/media?parent=76443"}],"wp:term":[{"taxonomy":"category","embeddable":true,"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/categories?post=76443"},{"taxonomy":"post_tag","embeddable":true,"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/tags?post=76443"},{"taxonomy":"industries-application","embeddable":true,"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/industries-application?post=76443"},{"taxonomy":"post-badge","embeddable":true,"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/post-badge?post=76443"},{"taxonomy":"resource-type","embeddable":true,"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/resource-type?post=76443"},{"taxonomy":"product-library","embeddable":true,"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/product-library?post=76443"},{"taxonomy":"resource-technique","embeddable":true,"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/resource-technique?post=76443"},{"taxonomy":"author","embeddable":true,"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/ppma_author?post=76443"}],"curies":[{"name":"wp","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}