{"id":86362,"date":"2026-01-19T01:40:01","date_gmt":"2026-01-19T01:40:01","guid":{"rendered":"https:\/\/discover.restek.com\/senza-categoria\/trace-level-semivolatiles-analysis-an-evaluation-of-the-rmx-5sil-ms-column\/"},"modified":"2026-01-30T14:33:59","modified_gmt":"2026-01-30T14:33:59","slug":"trace-level-semivolatiles-analysis-an-evaluation-of-the-rmx-5sil-ms-column","status":"publish","type":"post","link":"https:\/\/discover.restek.com\/es\/application-notes-it\/evan5254\/trace-level-semivolatiles-analysis-an-evaluation-of-the-rmx-5sil-ms-column","title":{"rendered":"Trace-Level Semivolatiles Analysis: An Evaluation of the RMX-5Sil MS Column"},"content":{"rendered":"\n<h2 class=\"wp-block-heading\">Key Highlights<\/h2>\n\n<ul class=\"wp-block-list\">\n<li>Highly effective GC column deactivation produces an exceptionally inert sample flow path.<\/li>\n\n\n\n<li>Maximum inertness results in sharp, symmetrical peaks and single or sub-picogram instrument detection limits (IDL) for a wide range of challenging semivolatiles.<\/li>\n\n\n\n<li>Data quality objectives for calibration were easily met with \u226420% relative standard error (RSE) for all semivolatiles and lowest calibration points of 1-10 ppb for all compounds, except benzoic acid (50 ppb).<\/li>\n<\/ul>\n\n<figure class=\"wp-block-image size-large\"><a href=\"https:\/\/explore.restek.com\/RMX-contact-technical-service?utm_source=website&amp;utm_medium=button&amp;utm_campaign=RMXByRestek&amp;utm_content=EVAN5254UNV_contacttech\" target=\"_blank\" rel=\"noopener\"><img loading=\"lazy\" decoding=\"async\" width=\"1024\" height=\"170\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/\/RMX-Graphic-Ad-Contact-us-HTML-Block-TECHNICAL-SERVICE-1024x170.png\" alt=\"helpful person on the phone\" class=\"wp-image-79797\" title=\"-\" srcset=\"https:\/\/discover.restek.com\/wp-content\/uploads\/RMX-Graphic-Ad-Contact-us-HTML-Block-TECHNICAL-SERVICE-1024x170.png 1024w, https:\/\/discover.restek.com\/wp-content\/uploads\/RMX-Graphic-Ad-Contact-us-HTML-Block-TECHNICAL-SERVICE-300x50.png 300w, https:\/\/discover.restek.com\/wp-content\/uploads\/RMX-Graphic-Ad-Contact-us-HTML-Block-TECHNICAL-SERVICE-768x127.png 768w, https:\/\/discover.restek.com\/wp-content\/uploads\/RMX-Graphic-Ad-Contact-us-HTML-Block-TECHNICAL-SERVICE.png 1340w\" sizes=\"auto, (max-width: 1024px) 100vw, 1024px\" \/><\/a><\/figure>\n\n<div style=\"height:30px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n\n<figure class=\"wp-block-image size-large\"><a href=\"https:\/\/explore.restek.com\/RMX-contact-sales?utm_source=website&amp;utm_medium=button&amp;utm_campaign=RMXByRestek&amp;utm_content=EVAN5254UNV_contactsales\" target=\"_blank\" rel=\"noopener\"><img loading=\"lazy\" decoding=\"async\" width=\"1024\" height=\"170\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/\/RMX-Graphic-Ad-Contact-us-HTML-Block-SALES-1024x170.png\" alt=\"\" class=\"wp-image-79803\" title=\"-\" srcset=\"https:\/\/discover.restek.com\/wp-content\/uploads\/RMX-Graphic-Ad-Contact-us-HTML-Block-SALES-1024x170.png 1024w, https:\/\/discover.restek.com\/wp-content\/uploads\/RMX-Graphic-Ad-Contact-us-HTML-Block-SALES-300x50.png 300w, https:\/\/discover.restek.com\/wp-content\/uploads\/RMX-Graphic-Ad-Contact-us-HTML-Block-SALES-768x127.png 768w, https:\/\/discover.restek.com\/wp-content\/uploads\/RMX-Graphic-Ad-Contact-us-HTML-Block-SALES.png 1340w\" sizes=\"auto, (max-width: 1024px) 100vw, 1024px\" \/><\/a><\/figure>\n\n<div style=\"height:30px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n\n<h2 class=\"wp-block-heading\">Abstract<\/h2>\n\n<p class=\"wp-block-paragraph\">This application note evaluates the performance of an RMX-5Sil MS column for trace-level GC-MS\/MS analysis of semivolatile organic compounds. Single or sub-picogram instrument detection limits were achieved for all compounds, except benzoic acid (14.70 pg) and 2,4-dinitrophenol (11.53 pg). Calibration curve %RSE was \u226420% for all compounds, and curves ranged from 1-1000 ppb to 10-1000 ppb for all compounds, except benzoic acid, which was 50-1000 ppb.<\/p>\n\n<h2 class=\"wp-block-heading\">Introduction<\/h2>\n\n<p class=\"wp-block-paragraph\">Environmental testing laboratories routinely analyze semivolatile organic compounds (SVOCs) by GC-MS or GC-MS\/MS using methods such as EPA Method 8270E. Labs are increasingly adopting GC-MS\/MS methods because the improved selectivity of the detector allows for more sensitive analyses, such as those needed for EPA Method 3511 microextraction, which saves time and reduces solvent consumption compared to traditional sample preparation methods that can use up to a liter of sample. To achieve the full benefits of MS\/MS sensitivity, the analytical column must be highly inert to maximize peak signal-to-noise ratios. A broadly effective column deactivation is essential for ensuring good overall method performance across a wide range of compound chemistries that interact via different mechanisms with any active sites present on the column surface. In this study, an RMX-5Sil MS column was independently evaluated for analytical performance across a wide range of trace-level semivolatiles, including compounds that are known to be very challenging. Chromatographic performance and calibration ranges were assessed to evaluate the column\u2019s suitability for achieving picogram-level detection.<\/p>\n\n<h2 class=\"wp-block-heading\">Experimental<\/h2>\n\n<h3 class=\"wp-block-heading\">Standard Preparation<\/h3>\n\n<p class=\"wp-block-paragraph\">Multicomponent calibration standards were prepared in methylene chloride across a range of 1-1000 ppb (11 points) from commercially available reference standards.<\/p>\n\n<h3 class=\"wp-block-heading\">Instrument Conditions<\/h3>\n\n<p class=\"wp-block-paragraph\">Samples were run on an RMX-5Sil MS column in a 30 m, 0.25 mm ID, 0.25 \u03bcm format (cat.# 17323). A Shimadzu Nexis GC-2030 GC paired with a Shimadzu GCMS-TQ8050 NX ultra-fast mass spectrometer with EI source and UFsweeper high-efficiency collision cell was used for GC-MS\/MS semivolatiles analysis under the conditions listed below.<\/p>\n\n<p class=\"wp-block-paragraph\"><strong>Table I:<\/strong> GC-MS\/MS Method Conditions for Trace-Level Semivolatiles Analysis<\/p>\n\n<figure class=\"wp-block-table\"><table class=\"has-fixed-layout\"><tbody><tr><td>\n\t\t\t\tGas Chromatograph\n\t\t\t<\/td><td>\n\t\t\t\tShimadzu Nexis GC-2030\n\t\t\t<\/td><\/tr><tr><td>\n\t\t\t\tColumn\n\t\t\t<\/td><td>RMX-5Sil MS, 30 m x 0.25 mm ID x 0.25 \u00b5m (cat.# 17323) <\/td><\/tr><tr><td>\n\t\t\t\tFlow mode\n\t\t\t<\/td><td>Constant linear velocity (39.5 cm\/s) <\/td><\/tr><tr><td>\n\t\t\t\tInjector mode\n\t\t\t<\/td><td>\n\t\t\t\tSplit (5:1)\n\t\t\t<\/td><\/tr><tr><td>\n\t\t\t\tInjector liner\n\t\t\t<\/td><td>Topaz 3.5 mm ID single taper inlet liner w\/wool (cat.# 23336)<\/td><\/tr><tr><td>\n\t\t\t\tInjection volume\n\t\t\t<\/td><td>\n\t\t\t\t1 \u00b5L\n\t\t\t<\/td><\/tr><tr><td>\n\t\t\t\tInjector temperature\n\t\t\t<\/td><td>\n\t\t\t\t275 \u00b0C\n\t\t\t<\/td><\/tr><tr><td>\n\t\t\t\tOven program\n\t\t\t<\/td><td>\n\t\t\t\t40 \u00b0C (hold 1.5 min) to 280 \u00b0C at 20 \u00b0C \/min to 320 \u00b0C at 5 \u00b0C \/min (hold 1 min)\n\t\t\t<\/td><\/tr><tr><td>\n\t\t\t\tMS system\n\t\t\t<\/td><td>\n\t\t\t\tShimadzu GCMS-TQ8050 NX ultra-fast mass spectrometer with EI source and UFsweeper high-efficiency collision cell\n\t\t\t<\/td><\/tr><tr><td>\n\t\t\t\tIonization\n\t\t\t<\/td><td>\n\t\t\t\tElectron impact\n\t\t\t<\/td><\/tr><tr><td>\n\t\t\t\tIonization energy\n\t\t\t<\/td><td>\n\t\t\t\t70 eV\n\t\t\t<\/td><\/tr><tr><td>\n\t\t\t\tEmission current\n\t\t\t<\/td><td>\n\t\t\t\t60 mA\n\t\t\t<\/td><\/tr><tr><td>\n\t\t\t\tAcquisition mode\n\t\t\t<\/td><td>\n\t\t\t\tMRM\n\t\t\t<\/td><\/tr><tr><td>\n\t\t\t\tCollision gas\n\t\t\t<\/td><td>\n\t\t\t\tArgon\n\t\t\t<\/td><\/tr><tr><td>\n\t\t\t\tSource temperature\n\t\t\t<\/td><td>\n\t\t\t\t230 \u00b0C\n\t\t\t<\/td><\/tr><tr><td>\n\t\t\t\tGC interface temperature\n\t\t\t<\/td><td>\n\t\t\t\t300 \u00b0C\n\t\t\t<\/td><\/tr><\/tbody><\/table><\/figure>\n\n<h2 class=\"wp-block-heading\">Results and Discussion<\/h2>\n\n<h3 class=\"wp-block-heading\">Chromatographic Performance<\/h3>\n\n<p class=\"wp-block-paragraph\">Overall, the RMX-5Sil MS column produced excellent peak shapes and separations for 86 semivolatiles at 50 ppb across the chromatographic space with the last compound eluting in just under 21 minutes (Figure 1). To evaluate column inertness, peak tailing was measured at or near the low and high ends of the linear calibration range for acidic and basic compounds that are known to be particularly problematic. As shown in Figure 2, the inert surface minimized surface activity at low levels and maximized peak symmetry for easy, accurate integration, even when difficult compounds were being integrated at low concentrations. In addition, the surface did not show any negative effects on polymer selectivity, as demonstrated by good separation of the closely eluting polycyclic aromatic hydrocarbon (PAH) pair, benzo(b)fluoranthene and benzo(k)fluoranthene at 1 ppb.<\/p>\n<div class=\"wp-block-custom-chromatogram-article-top\"><div class=\"chromatogram-article-placeholder\"><div class=\"figure-heading\"><strong>Figure 1: <\/strong>GC-MS\/MS Analysis of 86 Semivolatiles on an RMX-5Sil MS Column (50 ppb, TIC)<\/div><div class='chromatogram-article-inner-full'><div class=\"chromatogram-article-inner\">\n<div class=\"wp-block-custom-chromatogram-article\"><div class=\"wp-block-custom-chromatogram-article\"><div class=\"chromatogram-image wide-image\"><img decoding=\"async\" src=\"https:\/\/ez.restek.com\/images\/cgram\/gc_ev1534.png\" alt=\"EPA 8270E Semivolatiles on RMX-5Sil MS by GC-MS\/MS (Shimadzu Chromatogram)\" title=\"-\"><\/div><p class=\"article-id\" style=\"text-align:center\"> GC_EV1534<\/p><div class=\"chromatogram-peaks\"><h4>Peaks<\/h4><table class=\"peaks col-12\">\n<thead><tr><th><\/th><th style=\"text-align: left;width: 75px\">Peaks<\/th><th style=\"text-align: center;width: 75px\">t<sub>R<\/sub> (min)<\/th><th style=\"text-align: center;width: 75px\">Transition 1<\/th><th style=\"text-align: center;width: 75px\">Collision energy 1<\/th><th style=\"text-align: center;width: 75px\">Transition 2<\/th><th style=\"text-align: center;width: 75px\">Collision energy 2<\/th><\/tr><\/thead>\n<tbody><tr><td class=\"num\">1.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for N-Nitrosodimethylamine\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/62-75-9\/N-Nitrosodimethylamine\" target=\"_blank\" rel=\"noopener\">N-Nitrosodimethylamine<\/a><\/td><td class=\"oth\">3.173<\/td><td class=\"oth\">74.00&gt;44.10<\/td><td class=\"oth\">6<\/td><td class=\"oth\">74.00&gt;42.10<\/td><td class=\"oth\">18<\/td><\/tr>\n<tr><td class=\"num\">2.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Pyridine\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/110-86-1\/Pyridine\" target=\"_blank\" rel=\"noopener\">Pyridine<\/a><\/td><td class=\"oth\">3.276<\/td><td class=\"oth\">79.10&gt;50.10<\/td><td class=\"oth\">21<\/td><td class=\"oth\">79.10&gt;52.10<\/td><td class=\"oth\">15<\/td><\/tr>\n<tr><td class=\"num\">3.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 2-Fluorophenol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/367-12-4\/2-Fluorophenol\" target=\"_blank\" rel=\"noopener\">2-Fluorophenol<\/a><\/td><td class=\"oth\">4.532<\/td><td class=\"oth\">112.00&gt;64.10<\/td><td class=\"oth\">18<\/td><td class=\"oth\">112.00&gt;92.10<\/td><td class=\"oth\">9<\/td><\/tr>\n<tr><td class=\"num\">4.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Phenol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/108-95-2\/Phenol\" target=\"_blank\" rel=\"noopener\">Phenol<\/a><\/td><td class=\"oth\">5.514<\/td><td class=\"oth\">94.00&gt;66.00<\/td><td class=\"oth\">9<\/td><td class=\"oth\">66.00&gt;40.00<\/td><td class=\"oth\">12<\/td><\/tr>\n<tr><td class=\"num\">5.<\/td><td class=\"cmpd\">Phenol-d5 Surr<\/td><td class=\"oth\">5.5<\/td><td class=\"oth\">99.00&gt;71.10<\/td><td class=\"oth\">12<\/td><td class=\"oth\">99.00&gt;69.10<\/td><td class=\"oth\">27<\/td><\/tr>\n<tr><td class=\"num\">6.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Aniline\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/62-53-3\/Aniline\" target=\"_blank\" rel=\"noopener\">Aniline<\/a><\/td><td class=\"oth\">5.554<\/td><td class=\"oth\">93.00&gt;66.10<\/td><td class=\"oth\">18<\/td><td class=\"oth\">93.00&gt;51.10<\/td><td class=\"oth\">30<\/td><\/tr>\n<tr><td class=\"num\">7.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Bis(2-chloroethyl) ether\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/111-44-4\/Bis(2-chloroethyl) ether\" target=\"_blank\" rel=\"noopener\">Bis(2-chloroethyl) ether<\/a><\/td><td class=\"oth\">5.619<\/td><td class=\"oth\">93.00&gt;63.10<\/td><td class=\"oth\">9<\/td><td class=\"oth\">95.00&gt;65.00<\/td><td class=\"oth\">6<\/td><\/tr>\n<tr><td class=\"num\">8.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 2-Chlorophenol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/95-57-8\/2-Chlorophenol\" target=\"_blank\" rel=\"noopener\">2-Chlorophenol<\/a><\/td><td class=\"oth\">5.688<\/td><td class=\"oth\">128.00&gt;64.00<\/td><td class=\"oth\">18<\/td><td class=\"oth\">128.00&gt;91.90<\/td><td class=\"oth\">15<\/td><\/tr>\n<tr><td class=\"num\">9.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 1,3-Dichlorobenzene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/541-73-1\/1,3-Dichlorobenzene\" target=\"_blank\" rel=\"noopener\">1,3-Dichlorobenzene<\/a><\/td><td class=\"oth\">5.863<\/td><td class=\"oth\">146.00&gt;111.10<\/td><td class=\"oth\">21<\/td><td class=\"oth\">146.00&gt;75.20<\/td><td class=\"oth\">30<\/td><\/tr>\n<tr><td class=\"num\">10.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 1,4-Dichlorobenzene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/106-46-7\/1,4-Dichlorobenzene\" target=\"_blank\" rel=\"noopener\">1,4-Dichlorobenzene<\/a><\/td><td class=\"oth\">5.944<\/td><td class=\"oth\">146.00&gt;111.10<\/td><td class=\"oth\">21<\/td><td class=\"oth\">146.00&gt;75.20<\/td><td class=\"oth\">30<\/td><\/tr>\n<tr><td class=\"num\">11.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Benzyl alcohol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/100-51-6\/Benzyl alcohol\" target=\"_blank\" rel=\"noopener\">Benzyl alcohol<\/a><\/td><td class=\"oth\">6.069<\/td><td class=\"oth\">79.00&gt;77.10<\/td><td class=\"oth\">12<\/td><td class=\"oth\">107.00&gt;79.10<\/td><td class=\"oth\">9<\/td><\/tr>\n<tr><td class=\"num\">12.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 1,2-Dichlorobenzene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/95-50-1\/1,2-Dichlorobenzene\" target=\"_blank\" rel=\"noopener\">1,2-Dichlorobenzene<\/a><\/td><td class=\"oth\">6.116<\/td><td class=\"oth\">146.00&gt;111.10<\/td><td class=\"oth\">21<\/td><td class=\"oth\">146.00&gt;75.20<\/td><td class=\"oth\">30<\/td><\/tr>\n<tr><td class=\"num\">13.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 2-Methylphenol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/95-48-7\/2-Methylphenol\" target=\"_blank\" rel=\"noopener\">2-Methylphenol<\/a><\/td><td class=\"oth\">6.194<\/td><td class=\"oth\">108.00&gt;77.00<\/td><td class=\"oth\">27<\/td><td class=\"oth\">108.00&gt;79.00<\/td><td class=\"oth\">18<\/td><\/tr>\n<tr><td class=\"num\">14.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 2,2&amp;apos;-oxybis(1-chloropropane)\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/108-60-1\/2,2&amp;apos;-oxybis(1-chloropropane)\" target=\"_blank\" rel=\"noopener\">2,2&#8242;-oxybis(1-chloropropane)<\/a><\/td><td class=\"oth\">6.229<\/td><td class=\"oth\">121.05&gt;77.00<\/td><td class=\"oth\">9<\/td><td class=\"oth\">121.05&gt;45.00<\/td><td class=\"oth\">6<\/td><\/tr>\n<tr><td class=\"num\">15.<\/td><td class=\"cmpd\">3 and 4-Methylphenol<\/td><td class=\"oth\">6.369<\/td><td class=\"oth\">107.10&gt;77.10<\/td><td class=\"oth\">15<\/td><td class=\"oth\">107.10&gt;79.10<\/td><td class=\"oth\">6<\/td><\/tr>\n<tr><td class=\"num\">16.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for N-Nitrosodi-N-propylamine\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/621-64-7\/N-Nitrosodi-N-propylamine\" target=\"_blank\" rel=\"noopener\">N-Nitrosodi-N-propylamine<\/a><\/td><td class=\"oth\">6.374<\/td><td class=\"oth\">130.10&gt;113.10<\/td><td class=\"oth\">4<\/td><td class=\"oth\">130.10&gt;88.10<\/td><td class=\"oth\">4<\/td><\/tr>\n<tr><td class=\"num\">17.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Hexachloroethane\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/67-72-1\/Hexachloroethane\" target=\"_blank\" rel=\"noopener\">Hexachloroethane<\/a><\/td><td class=\"oth\">6.511<\/td><td class=\"oth\">117.00&gt;81.90<\/td><td class=\"oth\">30<\/td><td class=\"oth\">119.00&gt;83.80<\/td><td class=\"oth\">33<\/td><\/tr>\n<tr><td class=\"num\">18.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Nitrobenzene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/98-95-3\/Nitrobenzene\" target=\"_blank\" rel=\"noopener\">Nitrobenzene<\/a><\/td><td class=\"oth\">6.563<\/td><td class=\"oth\">77.05&gt;51.00<\/td><td class=\"oth\">21<\/td><td class=\"oth\">123.05&gt;77.00<\/td><td class=\"oth\">15<\/td><\/tr>\n<tr><td class=\"num\">19.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Nitrobenzene-d5\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/4165-60-0\/Nitrobenzene-d5\" target=\"_blank\" rel=\"noopener\">Nitrobenzene-d5<\/a><\/td><td class=\"oth\">6.541<\/td><td class=\"oth\">82.00&gt;54.10<\/td><td class=\"oth\">18<\/td><td class=\"oth\">128.00&gt;82.10<\/td><td class=\"oth\">18<\/td><\/tr>\n<tr><td class=\"num\">20.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Isophorone\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/78-59-1\/Isophorone\" target=\"_blank\" rel=\"noopener\">Isophorone<\/a><\/td><td class=\"oth\">6.846<\/td><td class=\"oth\">82.00&gt;54.00<\/td><td class=\"oth\">9<\/td><td class=\"oth\">138.00&gt;82.00<\/td><td class=\"oth\">18<\/td><\/tr>\n<tr><td class=\"num\">21.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 2-Nitrophenol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/88-75-5\/2-Nitrophenol\" target=\"_blank\" rel=\"noopener\">2-Nitrophenol<\/a><\/td><td class=\"oth\">6.941<\/td><td class=\"oth\">139.00&gt;109.10<\/td><td class=\"oth\">9<\/td><td class=\"oth\">139.00&gt;81.00<\/td><td class=\"oth\">12<\/td><\/tr>\n<tr><td class=\"num\">22.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 2,4-Dimethylphenol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/105-67-9\/2,4-Dimethylphenol\" target=\"_blank\" rel=\"noopener\">2,4-Dimethylphenol<\/a><\/td><td class=\"oth\">6.99<\/td><td class=\"oth\">107.00&gt;77.10<\/td><td class=\"oth\">18<\/td><td class=\"oth\">122.00&gt;107.10<\/td><td class=\"oth\">18<\/td><\/tr>\n<tr><td class=\"num\">23.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Benzoic Acid\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/65-85-0\/Benzoic Acid\" target=\"_blank\" rel=\"noopener\">Benzoic Acid<\/a><\/td><td class=\"oth\">7.019<\/td><td class=\"oth\">122.10&gt;105.10<\/td><td class=\"oth\">9<\/td><td class=\"oth\">105.10&gt;77.10<\/td><td class=\"oth\">15<\/td><\/tr>\n<tr><td class=\"num\">24.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Bis(2-chloroethoxy)methane\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/111-91-1\/Bis(2-chloroethoxy)methane\" target=\"_blank\" rel=\"noopener\">Bis(2-chloroethoxy)methane<\/a><\/td><td class=\"oth\">7.102<\/td><td class=\"oth\">93.00&gt;63.10<\/td><td class=\"oth\">9<\/td><td class=\"oth\">95.00&gt;65.00<\/td><td class=\"oth\">6<\/td><\/tr>\n<tr><td class=\"num\">25.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 2,4-Dichlorophenol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/120-83-2\/2,4-Dichlorophenol\" target=\"_blank\" rel=\"noopener\">2,4-Dichlorophenol<\/a><\/td><td class=\"oth\">7.223<\/td><td class=\"oth\">162.00&gt;63.10<\/td><td class=\"oth\">33<\/td><td class=\"oth\">164.00&gt;63.10<\/td><td class=\"oth\">30<\/td><\/tr>\n<tr><td class=\"num\">26.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 1,2,4-Trichlorobenzene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/120-82-1\/1,2,4-Trichlorobenzene\" target=\"_blank\" rel=\"noopener\">1,2,4-Trichlorobenzene<\/a><\/td><td class=\"oth\">7.333<\/td><td class=\"oth\">180.00&gt;109.00<\/td><td class=\"oth\">30<\/td><td class=\"oth\">180.00&gt;145.10<\/td><td class=\"oth\">18<\/td><\/tr>\n<tr><td class=\"num\">27.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Naphthalene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/91-20-3\/Naphthalene\" target=\"_blank\" rel=\"noopener\">Naphthalene<\/a><\/td><td class=\"oth\">7.424<\/td><td class=\"oth\">128.10&gt;102.10<\/td><td class=\"oth\">20<\/td><td class=\"oth\">128.10&gt;78.00<\/td><td class=\"oth\">20<\/td><\/tr>\n<tr><td class=\"num\">28.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 2,6-Dichlorophenol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/87-65-0\/2,6-Dichlorophenol\" target=\"_blank\" rel=\"noopener\">2,6-Dichlorophenol<\/a><\/td><td class=\"oth\">7.223<\/td><td class=\"oth\">136.00&gt;108.10<\/td><td class=\"oth\">27<\/td><td class=\"oth\">136.00&gt;134.10<\/td><td class=\"oth\">27<\/td><\/tr>\n<tr><td class=\"num\">29.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 4-Chloroaniline\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/106-47-8\/4-Chloroaniline\" target=\"_blank\" rel=\"noopener\">4-Chloroaniline<\/a><\/td><td class=\"oth\">7.484<\/td><td class=\"oth\">127.00&gt;65.10<\/td><td class=\"oth\">27<\/td><td class=\"oth\">127.00&gt;92.10<\/td><td class=\"oth\">18<\/td><\/tr>\n<tr><td class=\"num\">30.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Hexachlorobutadiene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/87-68-3\/Hexachlorobutadiene\" target=\"_blank\" rel=\"noopener\">Hexachlorobutadiene<\/a><\/td><td class=\"oth\">7.59<\/td><td class=\"oth\">225.00&gt;189.80<\/td><td class=\"oth\">21<\/td><td class=\"oth\">225.00&gt;155.00<\/td><td class=\"oth\">30<\/td><\/tr>\n<tr><td class=\"num\">31.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 4-Chloro-3-methylphenol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/59-50-7\/4-Chloro-3-methylphenol\" target=\"_blank\" rel=\"noopener\">4-Chloro-3-methylphenol<\/a><\/td><td class=\"oth\">8.072<\/td><td class=\"oth\">107.00&gt;77.10<\/td><td class=\"oth\">18<\/td><td class=\"oth\">142.00&gt;107.00<\/td><td class=\"oth\">18<\/td><\/tr>\n<tr><td class=\"num\">32.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 2-Methylnaphthalene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/91-57-6\/2-Methylnaphthalene\" target=\"_blank\" rel=\"noopener\">2-Methylnaphthalene<\/a><\/td><td class=\"oth\">8.268<\/td><td class=\"oth\">142.10&gt;115.10<\/td><td class=\"oth\">28<\/td><td class=\"oth\">115.10&gt;89.00<\/td><td class=\"oth\">16<\/td><\/tr>\n<tr><td class=\"num\">33.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 1-Methylnaphthalene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/90-12-0\/1-Methylnaphthalene\" target=\"_blank\" rel=\"noopener\">1-Methylnaphthalene<\/a><\/td><td class=\"oth\">8.388<\/td><td class=\"oth\">142.10&gt;115.10<\/td><td class=\"oth\">28<\/td><td class=\"oth\">115.10&gt;89.00<\/td><td class=\"oth\">16<\/td><\/tr>\n<tr><td class=\"num\">34.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Hexachlorocyclopentadiene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/77-47-4\/Hexachlorocyclopentadiene\" target=\"_blank\" rel=\"noopener\">Hexachlorocyclopentadiene<\/a><\/td><td class=\"oth\">8.474<\/td><td class=\"oth\">237.00&gt;141.00<\/td><td class=\"oth\">27<\/td><td class=\"oth\">237.00&gt;143.00<\/td><td class=\"oth\">27<\/td><\/tr>\n<tr><td class=\"num\">35.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 2,4,6-Trichlorophenol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/88-06-2\/2,4,6-Trichlorophenol\" target=\"_blank\" rel=\"noopener\">2,4,6-Trichlorophenol<\/a><\/td><td class=\"oth\">8.613<\/td><td class=\"oth\">196.00&gt;97.00<\/td><td class=\"oth\">33<\/td><td class=\"oth\">198.00&gt;97.00<\/td><td class=\"oth\">30<\/td><\/tr>\n<tr><td class=\"num\">36.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 2,4,5-Trichlorophenol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/95-95-4\/2,4,5-Trichlorophenol\" target=\"_blank\" rel=\"noopener\">2,4,5-Trichlorophenol<\/a><\/td><td class=\"oth\">8.651<\/td><td class=\"oth\">196.00&gt;97.00<\/td><td class=\"oth\">33<\/td><td class=\"oth\">198.00&gt;97.00<\/td><td class=\"oth\">30<\/td><\/tr>\n<tr><td class=\"num\">37.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 2-Fluorobiphenyl\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/321-60-8\/2-Fluorobiphenyl\" target=\"_blank\" rel=\"noopener\">2-Fluorobiphenyl<\/a><\/td><td class=\"oth\">8.721<\/td><td class=\"oth\">172.00&gt;151.20<\/td><td class=\"oth\">27<\/td><td class=\"oth\">172.00&gt;146.10<\/td><td class=\"oth\">27<\/td><\/tr>\n<tr><td class=\"num\">38.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 2-Chloronaphthalene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/91-58-7\/2-Chloronaphthalene\" target=\"_blank\" rel=\"noopener\">2-Chloronaphthalene<\/a><\/td><td class=\"oth\">8.862<\/td><td class=\"oth\">162.00&gt;127.10<\/td><td class=\"oth\">18<\/td><td class=\"oth\">162.00&gt;77.10<\/td><td class=\"oth\">33<\/td><\/tr>\n<tr><td class=\"num\">39.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 2-Nitroaniline\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/88-74-4\/2-Nitroaniline\" target=\"_blank\" rel=\"noopener\">2-Nitroaniline<\/a><\/td><td class=\"oth\">8.979<\/td><td class=\"oth\">138.00&gt;92.00<\/td><td class=\"oth\">15<\/td><td class=\"oth\">138.00&gt;65.10<\/td><td class=\"oth\">33<\/td><\/tr>\n<tr><td class=\"num\">40.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 1,4-Dinitrobenzene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/100-25-4\/1,4-Dinitrobenzene\" target=\"_blank\" rel=\"noopener\">1,4-Dinitrobenzene<\/a><\/td><td class=\"oth\">9.143<\/td><td class=\"oth\">168.00&gt;75.10<\/td><td class=\"oth\">30<\/td><td class=\"oth\">168.00&gt;92.00<\/td><td class=\"oth\">15<\/td><\/tr>\n<tr><td class=\"num\">41.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Dimethylphthalate\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/131-11-3\/Dimethylphthalate\" target=\"_blank\" rel=\"noopener\">Dimethylphthalate<\/a><\/td><td class=\"oth\">9.216<\/td><td class=\"oth\">163.00&gt;77.20<\/td><td class=\"oth\">15<\/td><td class=\"oth\">163.00&gt;133.10<\/td><td class=\"oth\">15<\/td><\/tr>\n<tr><td class=\"num\">42.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 1,3-Dinitrobenzene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/99-65-0\/1,3-Dinitrobenzene\" target=\"_blank\" rel=\"noopener\">1,3-Dinitrobenzene<\/a><\/td><td class=\"oth\">9.14<\/td><td class=\"oth\">168.00&gt;75.00<\/td><td class=\"oth\">30<\/td><td class=\"oth\">168.00&gt;122.00<\/td><td class=\"oth\">12<\/td><\/tr>\n<tr><td class=\"num\">43.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 2,6-Dinitrotoluene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/606-20-2\/2,6-Dinitrotoluene\" target=\"_blank\" rel=\"noopener\">2,6-Dinitrotoluene<\/a><\/td><td class=\"oth\">9.281<\/td><td class=\"oth\">165.00&gt;90.00<\/td><td class=\"oth\">15<\/td><td class=\"oth\">165.00&gt;63.10<\/td><td class=\"oth\">33<\/td><\/tr>\n<tr><td class=\"num\">44.<\/td><td class=\"cmpd\">Acenapthylene<\/td><td class=\"oth\">9.374<\/td><td class=\"oth\">152.10&gt;150.10<\/td><td class=\"oth\">28<\/td><td class=\"oth\">152.10&gt;126.10<\/td><td class=\"oth\">28<\/td><\/tr>\n<tr><td class=\"num\">45.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 3-Nitroaniline\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/99-09-2\/3-Nitroaniline\" target=\"_blank\" rel=\"noopener\">3-Nitroaniline<\/a><\/td><td class=\"oth\">9.484<\/td><td class=\"oth\">92.05&gt;65.00<\/td><td class=\"oth\">12<\/td><td class=\"oth\">138.05&gt;65.00<\/td><td class=\"oth\">27<\/td><\/tr>\n<tr><td class=\"num\">46.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Acenaphthene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/83-32-9\/Acenaphthene\" target=\"_blank\" rel=\"noopener\">Acenaphthene<\/a><\/td><td class=\"oth\">9.592<\/td><td class=\"oth\">152.10&gt;150.10<\/td><td class=\"oth\">28<\/td><td class=\"oth\">152.10&gt;126.10<\/td><td class=\"oth\">28<\/td><\/tr>\n<tr><td class=\"num\">47.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 2,4-Dinitrophenol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/51-28-5\/2,4-Dinitrophenol\" target=\"_blank\" rel=\"noopener\">2,4-Dinitrophenol<\/a><\/td><td class=\"oth\">9.617<\/td><td class=\"oth\">184.05&gt;107.00<\/td><td class=\"oth\">12<\/td><td class=\"oth\">154.05&gt;107.00<\/td><td class=\"oth\">6<\/td><\/tr>\n<tr><td class=\"num\">48.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 4-Nitrophenol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/100-02-7\/4-Nitrophenol\" target=\"_blank\" rel=\"noopener\">4-Nitrophenol<\/a><\/td><td class=\"oth\">9.689<\/td><td class=\"oth\">109.05&gt;81.00<\/td><td class=\"oth\">12<\/td><td class=\"oth\">109.05&gt;53.10<\/td><td class=\"oth\">18<\/td><\/tr>\n<tr><td class=\"num\">49.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 2,4-Dinitrotoluene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/121-14-2\/2,4-Dinitrotoluene\" target=\"_blank\" rel=\"noopener\">2,4-Dinitrotoluene<\/a><\/td><td class=\"oth\">9.778<\/td><td class=\"oth\">89.05&gt;63.10<\/td><td class=\"oth\">18<\/td><td class=\"oth\">165.05&gt;119.00<\/td><td class=\"oth\">6<\/td><\/tr>\n<tr><td class=\"num\">50.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Dibenzofuran\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/132-64-9\/Dibenzofuran\" target=\"_blank\" rel=\"noopener\">Dibenzofuran<\/a><\/td><td class=\"oth\">9.805<\/td><td class=\"oth\">168.00&gt;139.10<\/td><td class=\"oth\">24<\/td><td class=\"oth\">139.00&gt;89.10<\/td><td class=\"oth\">21<\/td><\/tr>\n<tr><td class=\"num\">51.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 2,3,5,6-Tetrachlorophenol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/935-95-5\/2,3,5,6-Tetrachlorophenol\" target=\"_blank\" rel=\"noopener\">2,3,5,6-Tetrachlorophenol<\/a><\/td><td class=\"oth\">9.904<\/td><td class=\"oth\">230.00&gt;130.90<\/td><td class=\"oth\">36<\/td><td class=\"oth\">232.00&gt;132.90<\/td><td class=\"oth\">36<\/td><\/tr>\n<tr><td class=\"num\">52.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 2,3,4,6-Tetrachlorophenol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/58-90-2\/2,3,4,6-Tetrachlorophenol\" target=\"_blank\" rel=\"noopener\">2,3,4,6-Tetrachlorophenol<\/a><\/td><td class=\"oth\">9.958<\/td><td class=\"oth\">230.00&gt;130.90<\/td><td class=\"oth\">36<\/td><td class=\"oth\">232.00&gt;132.90<\/td><td class=\"oth\">36<\/td><\/tr>\n<tr><td class=\"num\">53.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Diethylphthalate\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/84-66-2\/Diethylphthalate\" target=\"_blank\" rel=\"noopener\">Diethylphthalate<\/a><\/td><td class=\"oth\">10.093<\/td><td class=\"oth\">149.00&gt;65.00<\/td><td class=\"oth\">30<\/td><td class=\"oth\">177.00&gt;149.10<\/td><td class=\"oth\">12<\/td><\/tr>\n<tr><td class=\"num\">54.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 4-Chlorophenyl phenyl ether\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/7005-72-3\/4-Chlorophenyl phenyl ether\" target=\"_blank\" rel=\"noopener\">4-Chlorophenyl phenyl ether<\/a><\/td><td class=\"oth\">10.234<\/td><td class=\"oth\">141.00&gt;115.20<\/td><td class=\"oth\">21<\/td><td class=\"oth\">204.00&gt;141.20<\/td><td class=\"oth\">21<\/td><\/tr>\n<tr><td class=\"num\">55.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Fluorene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/86-73-7\/Fluorene\" target=\"_blank\" rel=\"noopener\">Fluorene<\/a><\/td><td class=\"oth\">10.23<\/td><td class=\"oth\">165.10&gt;163.10<\/td><td class=\"oth\">28<\/td><td class=\"oth\">165.10&gt;115.10<\/td><td class=\"oth\">28<\/td><\/tr>\n<tr><td class=\"num\">56.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 4-Nitroaniline\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/100-01-6\/4-Nitroaniline\" target=\"_blank\" rel=\"noopener\">4-Nitroaniline<\/a><\/td><td class=\"oth\">10.234<\/td><td class=\"oth\">138.00&gt;108.10<\/td><td class=\"oth\">12<\/td><td class=\"oth\">108.00&gt;80.00<\/td><td class=\"oth\">12<\/td><\/tr>\n<tr><td class=\"num\">57.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 4,6-Dinitro-2-methylphenol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/534-52-1\/4,6-Dinitro-2-methylphenol\" target=\"_blank\" rel=\"noopener\">4,6-Dinitro-2-methylphenol<\/a><\/td><td class=\"oth\">10.283<\/td><td class=\"oth\">198.05&gt;121.00<\/td><td class=\"oth\">12<\/td><td class=\"oth\">198.05&gt;53.00<\/td><td class=\"oth\">27<\/td><\/tr>\n<tr><td class=\"num\">58.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for N-Nitrosodiphenylamine (Diphenylamine)\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/86-30-6\/N-Nitrosodiphenylamine\" target=\"_blank\" rel=\"noopener\">N-Nitrosodiphenylamine (Diphenylamine)<\/a><\/td><td class=\"oth\">10.376<\/td><td class=\"oth\">169.00&gt;167.20<\/td><td class=\"oth\">27<\/td><td class=\"oth\">168.00&gt;139.00<\/td><td class=\"oth\">39<\/td><\/tr>\n<tr><td class=\"num\">59.<\/td><td class=\"cmpd\">Diphenylhydrazine<\/td><td class=\"oth\">10.429<\/td><td class=\"oth\">77.00&gt;51.20<\/td><td class=\"oth\">15<\/td><td class=\"oth\">77.00&gt;74.10<\/td><td class=\"oth\">33<\/td><\/tr>\n<tr><td class=\"num\">60.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 2,4,6-Tribromophenol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/118-79-6\/2,4,6-Tribromophenol\" target=\"_blank\" rel=\"noopener\">2,4,6-Tribromophenol<\/a><\/td><td class=\"oth\">10.524<\/td><td class=\"oth\">329.80&gt;141.00<\/td><td class=\"oth\">36<\/td><td class=\"oth\">331.80&gt;142.90<\/td><td class=\"oth\">36<\/td><\/tr>\n<tr><td class=\"num\">61.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 4-Bromophenyl phenyl ether\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/101-55-3\/4-Bromophenyl phenyl ether\" target=\"_blank\" rel=\"noopener\">4-Bromophenyl phenyl ether<\/a><\/td><td class=\"oth\">10.839<\/td><td class=\"oth\">250.00&gt;141.10<\/td><td class=\"oth\">21<\/td><td class=\"oth\">248.00&gt;141.10<\/td><td class=\"oth\">18<\/td><\/tr>\n<tr><td class=\"num\">62.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Hexachlorobenzene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/118-74-1\/Hexachlorobenzene\" target=\"_blank\" rel=\"noopener\">Hexachlorobenzene<\/a><\/td><td class=\"oth\">10.922<\/td><td class=\"oth\">283.80&gt;248.80<\/td><td class=\"oth\">24<\/td><td class=\"oth\">283.80&gt;213.80<\/td><td class=\"oth\">28<\/td><\/tr>\n<tr><td class=\"num\">63.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Pentachlorophenol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/87-86-5\/Pentachlorophenol\" target=\"_blank\" rel=\"noopener\">Pentachlorophenol<\/a><\/td><td class=\"oth\">11.159<\/td><td class=\"oth\">265.90&gt;164.90<\/td><td class=\"oth\">26<\/td><td class=\"oth\">265.90&gt;166.90<\/td><td class=\"oth\">26<\/td><\/tr>\n<tr><td class=\"num\">64.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Phenanthrene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/85-01-8\/Phenanthrene\" target=\"_blank\" rel=\"noopener\">Phenanthrene<\/a><\/td><td class=\"oth\">11.421<\/td><td class=\"oth\">178.10&gt;176.10<\/td><td class=\"oth\">28<\/td><td class=\"oth\">178.10&gt;152.10<\/td><td class=\"oth\">20<\/td><\/tr>\n<tr><td class=\"num\">65.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Anthracene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/120-12-7\/Anthracene\" target=\"_blank\" rel=\"noopener\">Anthracene<\/a><\/td><td class=\"oth\">11.484<\/td><td class=\"oth\">178.10&gt;176.10<\/td><td class=\"oth\">28<\/td><td class=\"oth\">178.10&gt;152.10<\/td><td class=\"oth\">20<\/td><\/tr>\n<tr><td class=\"num\">66.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Carbazole\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/86-74-8\/Carbazole\" target=\"_blank\" rel=\"noopener\">Carbazole<\/a><\/td><td class=\"oth\">11.68<\/td><td class=\"oth\">167.00&gt;139.20<\/td><td class=\"oth\">27<\/td><td class=\"oth\">166.00&gt;140.00<\/td><td class=\"oth\">18<\/td><\/tr>\n<tr><td class=\"num\">67.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Di-n-butylphthalate\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/84-74-2\/Di-n-butylphthalate\" target=\"_blank\" rel=\"noopener\">Di-n-butylphthalate<\/a><\/td><td class=\"oth\">12.131<\/td><td class=\"oth\">149.00&gt;93.10<\/td><td class=\"oth\">18<\/td><td class=\"oth\">149.00&gt;65.10<\/td><td class=\"oth\">24<\/td><\/tr>\n<tr><td class=\"num\">68.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Fluoranthene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/206-44-0\/Fluoranthene\" target=\"_blank\" rel=\"noopener\">Fluoranthene<\/a><\/td><td class=\"oth\">12.9<\/td><td class=\"oth\">202.10&gt;200.10<\/td><td class=\"oth\">30<\/td><td class=\"oth\">200.10&gt;198.10<\/td><td class=\"oth\">30<\/td><\/tr>\n<tr><td class=\"num\">69.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Benzidine\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/92-87-5\/Benzidine\" target=\"_blank\" rel=\"noopener\">Benzidine<\/a><\/td><td class=\"oth\">13.062<\/td><td class=\"oth\">184.00&gt;156.10<\/td><td class=\"oth\">24<\/td><td class=\"oth\">184.00&gt;167.10<\/td><td class=\"oth\">24<\/td><\/tr>\n<tr><td class=\"num\">70.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Pyrene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/129-00-0\/Pyrene\" target=\"_blank\" rel=\"noopener\">Pyrene<\/a><\/td><td class=\"oth\">13.18<\/td><td class=\"oth\">202.10&gt;200.10<\/td><td class=\"oth\">30<\/td><td class=\"oth\">200.10&gt;198.10<\/td><td class=\"oth\">30<\/td><\/tr>\n<tr><td class=\"num\">71.<\/td><td class=\"cmpd\">o-Terphenyl-D14<\/td><td class=\"oth\">13.377<\/td><td class=\"oth\">244.00&gt;240.10<\/td><td class=\"oth\">30<\/td><td class=\"oth\">244.00&gt;226.20<\/td><td class=\"oth\">18<\/td><\/tr>\n<tr><td class=\"num\">72.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Butylbenzylphthalate\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/85-68-7\/Butylbenzylphthalate\" target=\"_blank\" rel=\"noopener\">Butylbenzylphthalate<\/a><\/td><td class=\"oth\">14.012<\/td><td class=\"oth\">149.00&gt;65.10<\/td><td class=\"oth\">24<\/td><td class=\"oth\">149.00&gt;93.10<\/td><td class=\"oth\">18<\/td><\/tr>\n<tr><td class=\"num\">73.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Bis(2-ethylhexyl)adipate\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/103-23-1\/Bis(2-ethylhexyl)adipate\" target=\"_blank\" rel=\"noopener\">Bis(2-ethylhexyl)adipate<\/a><\/td><td class=\"oth\">14.129<\/td><td class=\"oth\">129.00&gt;55.10<\/td><td class=\"oth\">21<\/td><td class=\"oth\">129.00&gt;101.10<\/td><td class=\"oth\">9<\/td><\/tr>\n<tr><td class=\"num\">74.<\/td><td class=\"cmpd\">3&#8217;3-Dichlorobenzidine<\/td><td class=\"oth\">14.769<\/td><td class=\"oth\">212.00&gt;180.10<\/td><td class=\"oth\">24<\/td><td class=\"oth\">212.00&gt;196.20<\/td><td class=\"oth\">21<\/td><\/tr>\n<tr><td class=\"num\">75.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Benz[a]anthracene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/56-55-3\/Benz[a]anthracene\" target=\"_blank\" rel=\"noopener\">Benz[a]anthracene<\/a><\/td><td class=\"oth\">14.805<\/td><td class=\"oth\">228.10&gt;226.10<\/td><td class=\"oth\">32<\/td><td class=\"oth\">226.10&gt;224.10<\/td><td class=\"oth\">32<\/td><\/tr>\n<tr><td class=\"num\">76.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Bis(2-ethylhexyl)phthalate\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/117-81-7\/Bis(2-ethylhexyl)phthalate\" target=\"_blank\" rel=\"noopener\">Bis(2-ethylhexyl)phthalate<\/a><\/td><td class=\"oth\">14.875<\/td><td class=\"oth\">149.00&gt;65.10<\/td><td class=\"oth\">24<\/td><td class=\"oth\">167.00&gt;149.10<\/td><td class=\"oth\">15<\/td><\/tr>\n<tr><td class=\"num\">77.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Chrysene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/218-01-9\/Chrysene\" target=\"_blank\" rel=\"noopener\">Chrysene<\/a><\/td><td class=\"oth\">14.861<\/td><td class=\"oth\">228.10&gt;226.10<\/td><td class=\"oth\">32<\/td><td class=\"oth\">226.10&gt;224.10<\/td><td class=\"oth\">32<\/td><\/tr>\n<tr><td class=\"num\">78.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Di-n-octylphthalate\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/117-84-0\/Di-n-octylphthalate\" target=\"_blank\" rel=\"noopener\">Di-n-octylphthalate<\/a><\/td><td class=\"oth\">16<\/td><td class=\"oth\">149.00&gt;65.10<\/td><td class=\"oth\">24<\/td><td class=\"oth\">149.00&gt;93.20<\/td><td class=\"oth\">18<\/td><\/tr>\n<tr><td class=\"num\">79.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Benzo[b]fluoranthene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/205-99-2\/Benzo[b]fluoranthene\" target=\"_blank\" rel=\"noopener\">Benzo[b]fluoranthene<\/a><\/td><td class=\"oth\">16.691<\/td><td class=\"oth\">252.10&gt;250.10<\/td><td class=\"oth\">36<\/td><td class=\"oth\">250.10&gt;248.10<\/td><td class=\"oth\">36<\/td><\/tr>\n<tr><td class=\"num\">80.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Benzo[k]fluoranthene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/207-08-9\/Benzo[k]fluoranthene\" target=\"_blank\" rel=\"noopener\">Benzo[k]fluoranthene<\/a><\/td><td class=\"oth\">16.746<\/td><td class=\"oth\">252.10&gt;250.10<\/td><td class=\"oth\">36<\/td><td class=\"oth\">250.10&gt;248.10<\/td><td class=\"oth\">36<\/td><\/tr>\n<tr><td class=\"num\">81.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Benzo[a]pyrene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/50-32-8\/Benzo[a]pyrene\" target=\"_blank\" rel=\"noopener\">Benzo[a]pyrene<\/a><\/td><td class=\"oth\">17.338<\/td><td class=\"oth\">252.10&gt;250.10<\/td><td class=\"oth\">36<\/td><td class=\"oth\">250.10&gt;248.10<\/td><td class=\"oth\">36<\/td><\/tr>\n<tr><td class=\"num\">82.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Dibenz[a,j]acridine\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/224-42-0\/Dibenz[a,j]acridine\" target=\"_blank\" rel=\"noopener\">Dibenz[a,j]acridine<\/a><\/td><td class=\"oth\">19.26<\/td><td class=\"oth\">279.00&gt;277.10<\/td><td class=\"oth\">33<\/td><td class=\"oth\">279.00&gt;250.00<\/td><td class=\"oth\">45<\/td><\/tr>\n<tr><td class=\"num\">83.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Indeno[1,2,3-cd]pyrene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/193-39-5\/Indeno[1,2,3-cd]pyrene\" target=\"_blank\" rel=\"noopener\">Indeno[1,2,3-cd]pyrene<\/a><\/td><td class=\"oth\">19.754<\/td><td class=\"oth\">276.10&gt;274.10<\/td><td class=\"oth\">36<\/td><td class=\"oth\">274.10&gt;272.10<\/td><td class=\"oth\">36<\/td><\/tr>\n<tr><td class=\"num\">84.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Dibenz[a,h]anthracene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/53-70-3\/Dibenz[a,h]anthracene\" target=\"_blank\" rel=\"noopener\">Dibenz[a,h]anthracene<\/a><\/td><td class=\"oth\">19.813<\/td><td class=\"oth\">278.10&gt;276.10<\/td><td class=\"oth\">36<\/td><td class=\"oth\">278.10&gt;274.10<\/td><td class=\"oth\">60<\/td><\/tr>\n<tr><td class=\"num\">85.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Benzo[g,h,i]perylene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/191-24-2\/Benzo[g,h,i]perylene\" target=\"_blank\" rel=\"noopener\">Benzo[g,h,i]perylene<\/a><\/td><td class=\"oth\">20.339<\/td><td class=\"oth\">276.10&gt;274.10<\/td><td class=\"oth\">36<\/td><td class=\"oth\">274.10&gt;272.10<\/td><td class=\"oth\">36<\/td><\/tr>\n<\/tbody><\/table><\/div><div class=\"chromatogram-conditions\"><h4>Conditions<\/h4><div class=\"conditions-container container-fluid\"><div class=\"row\"><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Column<\/th><td>RMX-5Sil MS, 30 m, 0.25 mm ID, 0.25 \u00b5m (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/17323\" rel=\"noopener\">cat.# 17323<\/a>)<\/td><\/tr><td><\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Injection<\/th><td><\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Inj. Vol.:<\/th><td>1 \u00b5L split (split ratio 5:1)<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Liner:<\/th><td>Topaz 3.5 mm ID single taper inlet liner w\/wool (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/23336\" rel=\"noopener\">cat.# 23336<\/a>)<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Inj. Temp.:<\/th><td>275 \u00b0C<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\" colspan=\"2\">Oven<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Oven Temp.:<\/th><td>40 &#176;C (hold 1.5 min) to 280 &#176;C at 20 &#176;C\/min to 320 &#176;C at 5 &#176;C\/min (hold 1 min)<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Carrier Gas<\/th><td>He, constant flow<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Linear Velocity:<\/th><td>39.5 cm\/sec @ 40 \u00b0C<\/td><\/tr><\/table><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Detector<\/th><td>SRM\/MRM<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Acquisition Type:<\/th><td>SRM\/MRM <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Source Temp.:<\/th><td>230 \u00b0C<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Transfer Line Temp.:<\/th><td>300 \u00b0C<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Analyzer Type:<\/th><td>Triple Quadrupole <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Ionization Mode:<\/th><td>EI <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Collision Gas:<\/th><td>Ar <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Tune Type:<\/th><td>PFTBA <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Tune Emission Current:<\/th><td>60 \u03bcA<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Notes<\/th><td>Shimadzu Nexis GC-2030 with Shimadzu GCMS-TQ8050 NX ultra-fast mass spectrometer with EI source and UFsweeper high-efficiency collision cell<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Acknowledgement<\/th><td>Shimadzu<\/td><\/tr><\/table><\/div><\/div><\/div><div class=\"chromatogram-related-skus\"><h4>Related SKUs<\/h4><ul><li>17323<\/li><li>23336<\/li><\/ul><\/div><div class=\"chromatogram-related-searches\"><h4>Related Searches<\/h4><ul><li>8270<\/li><li>8270e<\/li><li>RMX-library<\/li><li>Semivolatiles<\/li><li>semivolatiles-gc-library<\/li><li>shimadzu<\/li><\/ul><\/div><div class=\"chromatogram-pdf-link\"><a href=\"https:\/\/ez.restek.com\/images\/cgram\/gc_ev1534.pdf\" target=\"_blank\" rel=\"noopener noreferrer\"><svg xmlns=\"http:\/\/www.w3.org\/2000\/svg\" width=\"18\" height=\"18\" viewBox=\"0 0 18 18\"><g data-name=\"Group 2996\"><path data-name=\"Rectangle 1246\" d=\"M0 0h18v18H0z\" style=\"fill: none;\"><\/path><\/g><g data-name=\"Group 2997\"><path data-name=\"Path 729\" d=\"M13.412 11.4v2.017H5.345V11.4H4v2.017a1.349 1.349 0 0 0 1.345 1.345h8.068a1.349 1.349 0 0 0 1.345-1.345V11.4zm-.672-2.694-.948-.948-1.741 1.735V4H8.706v5.493L6.965 7.758l-.948.948 3.361 3.361z\" transform=\"translate(-.437 -.414)\" style=\"fill: rgb(13, 123, 196);\"><\/path><\/g><\/svg>Download PDF<\/a><\/div><\/div><\/div>\n<\/div><\/div><\/div>\n<div style=\"height:25px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n<div class=\"wp-block-custom-chromatogram-article-top\"><div class=\"chromatogram-article-placeholder\"><div class=\"figure-heading\"><strong>Figure 2: <\/strong>The highly effective deactivation used in RMX-5Sil MS columns creates an exceptionally inert surface that produces symmetrical peaks for a wide range of semivolatiles, including acidic compounds (pentachlorophenol); basic compounds (benzidine); and closely eluting PAHs.<\/div><div class='chromatogram-article-inner-full'><div class=\"chromatogram-article-inner\">\n<div class=\"wp-block-custom-chromatogram-article\"><div class=\"wp-block-custom-chromatogram-article\"><div class=\"chromatogram-image regular-image\"><img decoding=\"async\" src=\"https:\/\/ez.restek.com\/images\/cgram\/gc_ev1535.png\" alt=\"Select EPA 8270E Semivolatiles on RMX-5Sil MS by GC-MS\/MS (Shimazu Chromatogram)\" title=\"-\"><\/div><p class=\"article-id\" style=\"text-align:center\"> GC_EV1535<\/p><div class=\"chromatogram-peaks\"><h4>Peaks<\/h4><table class=\"peaks col-lg-6 col-12\">\n<thead><tr><th><\/th><th style=\"text-align: left;width: 75px\">Peaks<\/th><th style=\"text-align: center;width: 75px\">t<sub>R<\/sub> (min)<\/th><th style=\"text-align: center;width: 75px\">Transition 1<\/th><th style=\"text-align: center;width: 75px\">Collision energy 1<\/th><th style=\"text-align: center;width: 75px\">Transition 2<\/th><th style=\"text-align: center;width: 75px\">Collision energy 2<\/th><\/tr><\/thead>\n<tbody><tr><td class=\"num\">1.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Benzidine\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/92-87-5\/Benzidine\" target=\"_blank\" rel=\"noopener\">Benzidine<\/a><\/td><td class=\"oth\">13.062<\/td><td class=\"oth\">184.00&gt;156.10<\/td><td class=\"oth\">24<\/td><td class=\"oth\">184.00&gt;167.10<\/td><td class=\"oth\">24<\/td><\/tr>\n<tr><td class=\"num\">2.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Benzo[b]fluoranthene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/205-99-2\/Benzo[b]fluoranthene\" target=\"_blank\" rel=\"noopener\">Benzo[b]fluoranthene<\/a><\/td><td class=\"oth\">16.691<\/td><td class=\"oth\">252.10&gt;250.10<\/td><td class=\"oth\">36<\/td><td class=\"oth\">250.10&gt;248.10<\/td><td class=\"oth\">36<\/td><\/tr>\n<tr><td class=\"num\">3.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Benzo[k]fluoranthene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/207-08-9\/Benzo[k]fluoranthene\" target=\"_blank\" rel=\"noopener\">Benzo[k]fluoranthene<\/a><\/td><td class=\"oth\">16.746<\/td><td class=\"oth\">252.10&gt;250.10<\/td><td class=\"oth\">36<\/td><td class=\"oth\">250.10&gt;248.10<\/td><td class=\"oth\">36<\/td><\/tr>\n<tr><td class=\"num\">4.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Pentachlorophenol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/87-86-5\/Pentachlorophenol\" target=\"_blank\" rel=\"noopener\">Pentachlorophenol<\/a><\/td><td class=\"oth\">11.159<\/td><td class=\"oth\">265.90&gt;164.90<\/td><td class=\"oth\">26<\/td><td class=\"oth\">265.90&gt;166.90<\/td><td class=\"oth\">26<\/td><\/tr>\n<\/tbody><\/table><\/div><div class=\"chromatogram-conditions\"><h4>Conditions<\/h4><div class=\"conditions-container container-fluid\"><div class=\"row\"><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Column<\/th><td>RMX-5Sil MS, 30 m, 0.25 mm ID, 0.25 \u00b5m (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/17323\" rel=\"noopener\">cat.# 17323<\/a>)<\/td><\/tr><td><\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Injection<\/th><td><\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Inj. Vol.:<\/th><td>1 \u00b5L split (split ratio 5:1)<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Liner:<\/th><td>Topaz 3.5 mm ID single taper inlet liner w\/wool (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/23336\" rel=\"noopener\">cat.# 23336<\/a>)<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Inj. Temp.:<\/th><td>275 \u00b0C<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\" colspan=\"2\">Oven<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Oven Temp.:<\/th><td>40 &#176;C (hold 1.5 min) to 280 &#176;C at 20 &#176;C\/min to 320 &#176;C at 5 &#176;C\/min (hold 1 min)<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Carrier Gas<\/th><td>He, constant flow<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Linear Velocity:<\/th><td>39.5 cm\/sec @ 40 \u00b0C<\/td><\/tr><\/table><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Detector<\/th><td>SRM\/MRM<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Acquisition Type:<\/th><td>SRM\/MRM <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Source Temp.:<\/th><td>230 \u00b0C<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Transfer Line Temp.:<\/th><td>300 \u00b0C<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Analyzer Type:<\/th><td>Triple Quadrupole <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Ionization Mode:<\/th><td>EI <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Collision Gas:<\/th><td>Ar <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Tune Type:<\/th><td>PFTBA <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Tune Emission Current:<\/th><td>60 \u03bcA<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Notes<\/th><td>Shimadzu Nexis GC-2030 with Shimadzu GCMS-TQ8050 NX ultra-fast mass spectrometer with EI source and UFsweeper high-efficiency collision cell<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Acknowledgement<\/th><td>Shimadzu<\/td><\/tr><\/table><\/div><\/div><\/div><div class=\"chromatogram-related-skus\"><h4>Related SKUs<\/h4><ul><li>17323<\/li><li>23336<\/li><\/ul><\/div><div class=\"chromatogram-related-searches\"><h4>Related Searches<\/h4><ul><li>RMX-library<\/li><li>semivolatiles<\/li><\/ul><\/div><div class=\"chromatogram-pdf-link\"><a href=\"https:\/\/ez.restek.com\/images\/cgram\/gc_ev1535.pdf\" target=\"_blank\" rel=\"noopener noreferrer\"><svg xmlns=\"http:\/\/www.w3.org\/2000\/svg\" width=\"18\" height=\"18\" viewBox=\"0 0 18 18\"><g data-name=\"Group 2996\"><path data-name=\"Rectangle 1246\" d=\"M0 0h18v18H0z\" style=\"fill: none;\"><\/path><\/g><g data-name=\"Group 2997\"><path data-name=\"Path 729\" d=\"M13.412 11.4v2.017H5.345V11.4H4v2.017a1.349 1.349 0 0 0 1.345 1.345h8.068a1.349 1.349 0 0 0 1.345-1.345V11.4zm-.672-2.694-.948-.948-1.741 1.735V4H8.706v5.493L6.965 7.758l-.948.948 3.361 3.361z\" transform=\"translate(-.437 -.414)\" style=\"fill: rgb(13, 123, 196);\"><\/path><\/g><\/svg>Download PDF<\/a><\/div><\/div><\/div>\n<\/div><\/div><\/div>\n<div style=\"height:25px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n<div class=\"wp-block-image\">\n<figure class=\"aligncenter size-full\"><a href=\"https:\/\/discover.restek.com\/it\/articoli\/gnot3445-it\/strumenti-online-ezgc?utm_source=website&amp;utm_medium=button&amp;utm_campaign=RMXByRestek&amp;utm_content=EVAN5254UNV_ezgc\"><img loading=\"lazy\" decoding=\"async\" width=\"1340\" height=\"312\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/RMX-Graphic-Ad-EZGC-HTML-Block-Mock-Up.png\" alt=\"EZGC software ad\" class=\"wp-image-61667\" title=\"-\" srcset=\"https:\/\/discover.restek.com\/wp-content\/uploads\/RMX-Graphic-Ad-EZGC-HTML-Block-Mock-Up.png 1340w, https:\/\/discover.restek.com\/wp-content\/uploads\/RMX-Graphic-Ad-EZGC-HTML-Block-Mock-Up-300x70.png 300w, https:\/\/discover.restek.com\/wp-content\/uploads\/RMX-Graphic-Ad-EZGC-HTML-Block-Mock-Up-1024x238.png 1024w, https:\/\/discover.restek.com\/wp-content\/uploads\/RMX-Graphic-Ad-EZGC-HTML-Block-Mock-Up-768x179.png 768w\" sizes=\"auto, (max-width: 1340px) 100vw, 1340px\" \/><\/a><\/figure>\n<\/div>\n<div style=\"height:30px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n\n<h3 class=\"wp-block-heading\">Calibration Performance<\/h3>\n\n<p class=\"wp-block-paragraph\">As demonstrated in Table II, the combination of the Shimadzu GC-MS\/MS system and the highly inert RMX-5Sil MS analytical column allowed for outstanding trace-level sensitivity for a broad range of semivolatile compound classes. The calibration curve %RSE was \u226420% for all compounds, which met data quality objectives for calibration. In addition, R<sup>2<\/sup> was determined for a subset of representative semivolatiles and was found to be \u22650.99 for all compounds that were assessed. For all semivolatiles, the calibration curves ranged from 1-1000 ppb to 10-1000 ppb, except benzoic acid which was 50-1000 ppb. A slightly higher linear range for benzoic acid was not surprising because it does not solubilize well in 5-type phases. Finally, very low instrument detection limits were achieved and were single or sub-picogram levels for all semivolatiles, except benzoic acid (14.70 pg) and 2,4-dinitrophenol (11.53 pg).<\/p>\n\n<p class=\"wp-block-paragraph\"><strong>Table II: <\/strong>Calibration Performance for Trace-Level Semivolatiles Analysis on an RMX-5Sil MS Column<\/p>\n\n<figure class=\"wp-block-table\"><table class=\"has-fixed-layout\"><tbody><tr><th>\n\t\t\t\tCompound Name\n\t\t\t<\/th><th class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\tRetention Time (min)\n\t\t\t<\/th><th class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\tCalibration Curve %RSE\n\t\t\t<\/th><th class=\"has-text-align-center\" data-align=\"center\">R<sup>2<\/sup> <\/th><th class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\tLow Point (ppb)\n\t\t\t<\/th><th class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\tHigh Point (ppb)\n\t\t\t<\/th><th class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\tIDL (pg)\n\t\t\t<\/th><\/tr><tr><td>\n\t\t\t\tN-Nitrosodimethylamine\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t3.173\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t16.82\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\"><\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t5\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1000\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t2.14\n\t\t\t<\/td><\/tr><tr><td>\n\t\t\t\tPyridine\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t3.276\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t8.69\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t0.994\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">10<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1000\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t3.23\n\t\t\t<\/td><\/tr><tr><td>\n\t\t\t\t2-Fluorophenol\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t4.532\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t15.24\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\"><\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t5\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1000\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t0.28\n\t\t\t<\/td><\/tr><tr><td>\n\t\t\t\tPhenol\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t5.514\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t19.25\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\"><\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1000\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t0.70\n\t\t\t<\/td><\/tr><tr><td>\n\t\t\t\tPhenol-d5\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t5.5\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t17.86\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\"><\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1000\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t0.29\n\t\t\t<\/td><\/tr><tr><td>\n\t\t\t\tAniline\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t5.554\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t9.52\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t0.993\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t10\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1000\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1.43\n\t\t\t<\/td><\/tr><tr><td>\n\t\t\t\tBis(2-chloroethyl) ether\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t5.619\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t14.33\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\"><\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1000\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t0.42\n\t\t\t<\/td><\/tr><tr><td>\n\t\t\t\t2-Chlorophenol\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t5.688\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t19.78\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\"><\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1000\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1.28\n\t\t\t<\/td><\/tr><tr><td>\n\t\t\t\t1,3-Dichlorobenzene\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t5.863\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t16.96\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\"><\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1000\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t0.36\n\t\t\t<\/td><\/tr><tr><td>\n\t\t\t\t1,4-Dichlorobenzene\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t5.944\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t17.90\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\"><\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1000\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t0.77\n\t\t\t<\/td><\/tr><tr><td>\n\t\t\t\tBenzyl alcohol\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t6.069\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t17.06\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\"><\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1000\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t0.98\n\t\t\t<\/td><\/tr><tr><td>\n\t\t\t\t1,2-Dichlorobenzene\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t6.116\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t17.32\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\"><\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1000\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1.13\n\t\t\t<\/td><\/tr><tr><td>\n\t\t\t\t2-Methylphenol\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t6.194\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t19.48\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\"><\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t10\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1000\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1.95\n\t\t\t<\/td><\/tr><tr><td>\n\t\t\t\t2,2&#8242;-oxybis(1-chloropropane)\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t6.229\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t19.32\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\"><\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t10\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1000\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t4.04\n\t\t\t<\/td><\/tr><tr><td>\n\t\t\t\t3- and 4-Methylphenol\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t6.369\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t19.71\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\"><\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1000\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1.01\n\t\t\t<\/td><\/tr><tr><td>\n\t\t\t\tN-Nitrosodi-N-propylamine\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t6.374\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t5.38\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t0.997\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t10\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1000\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t3.10\n\t\t\t<\/td><\/tr><tr><td>\n\t\t\t\tHexachloroethane\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t6.511\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t15.67\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\"><\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t5\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1000\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1.36\n\t\t\t<\/td><\/tr><tr><td>\n\t\t\t\tNitrobenzene\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t6.563\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t12.10\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t0.996\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t5\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1000\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t0.81\n\t\t\t<\/td><\/tr><tr><td>\n\t\t\t\tNitrobenzene-d5\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t6.541\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t17.81\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\"><\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1000\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t0.96\n\t\t\t<\/td><\/tr><tr><td>\n\t\t\t\tIsophorone\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t6.846\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t18.92\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\"><\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1000\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t0.56\n\t\t\t<\/td><\/tr><tr><td>\n\t\t\t\t2-Nitrophenol\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t6.941\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t13.56\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t0.996\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t10\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1000\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1.26\n\t\t\t<\/td><\/tr><tr><td>\n\t\t\t\t2,4-Dimethylphenol\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t6.99\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t19.63\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\"><\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1000\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t0.36\n\t\t\t<\/td><\/tr><tr><td>\n\t\t\t\tBenzoic acid\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t7.019\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t3.84\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t0.998\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t50\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1000\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t14.70\n\t\t\t<\/td><\/tr><tr><td>\n\t\t\t\tBis(2-chloroethoxy)methane\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t7.102\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t14.01\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\"><\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1000\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t0.26\n\t\t\t<\/td><\/tr><tr><td>\n\t\t\t\t2,4-Dichlorophenol\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t7.223\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t18.99\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\"><\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1000\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t0.83\n\t\t\t<\/td><\/tr><tr><td>\n\t\t\t\t1,2,4-Trichlorobenzene\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t7.333\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t10.67\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\"><\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1000\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t0.46\n\t\t\t<\/td><\/tr><tr><td>\n\t\t\t\tNaphthalene\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t7.424\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t8.68\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\"><\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1000\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t0.73\n\t\t\t<\/td><\/tr><tr><td>\n\t\t\t\t2,6-Dichlorophenol\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t7.223\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t18.18\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\"><\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1000\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t0.62\n\t\t\t<\/td><\/tr><tr><td>\n\t\t\t\t4-Chloroaniline\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t7.484\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t13.85\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t0.998\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t5\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1000\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1.34\n\t\t\t<\/td><\/tr><tr><td>\n\t\t\t\tHexachlorobutadiene\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t7.59\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t13.96\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\"><\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1000\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t0.65\n\t\t\t<\/td><\/tr><tr><td>\n\t\t\t\t4-Chloro-3-methylphenol\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t8.072\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t16.99\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\"><\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1000\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t0.30\n\t\t\t<\/td><\/tr><tr><td>\n\t\t\t\t2-Methylnaphthalene\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t8.268\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t10.57\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\"><\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1000\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t0.63\n\t\t\t<\/td><\/tr><tr><td>\n\t\t\t\t1-Methylnaphthalene\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t8.388\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t18.50\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\"><\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1000\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t0.64\n\t\t\t<\/td><\/tr><tr><td>\n\t\t\t\tHexachlorocyclopentadiene\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t8.474\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t16.59\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t0.998\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t5\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1000\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t2.72\n\t\t\t<\/td><\/tr><tr><td>\n\t\t\t\t2,4,6-Trichlorophenol\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t8.613\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t16.69\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\"><\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t5\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1000\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1.55\n\t\t\t<\/td><\/tr><tr><td>\n\t\t\t\t2,4,5-Trichlorophenol\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t8.651\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t16.98\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\"><\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t5\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1000\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t0.76\n\t\t\t<\/td><\/tr><tr><td>\n\t\t\t\t2-Fluorobiphenyl\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t8.721\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t16.31\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\"><\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1000\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1.05\n\t\t\t<\/td><\/tr><tr><td>\n\t\t\t\t2-Chloronaphthalene\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t8.862\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t14.23\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\"><\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1000\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t0.43\n\t\t\t<\/td><\/tr><tr><td>\n\t\t\t\t2-Nitroaniline\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t8.979\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t17.82\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t0.996\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t10\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1000\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1.33\n\t\t\t<\/td><\/tr><tr><td>\n\t\t\t\t1,4-Dinitrobenzene\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t9.143\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t12.76\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t0.996\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t10\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1000\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t8.63\n\t\t\t<\/td><\/tr><tr><td>\n\t\t\t\tDimethylphthalate\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t9.216\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t18.60\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\"><\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1000\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t0.60\n\t\t\t<\/td><\/tr><tr><td>\n\t\t\t\t1,3-Dinitrobenzene\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t9.14\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t18.88\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t0.991\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t10\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1000\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t6.97\n\t\t\t<\/td><\/tr><tr><td>\n\t\t\t\t2,6-Dinitrotoluene\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t9.281\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t19.86\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t0.997\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t5\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1000\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t2.06\n\t\t\t<\/td><\/tr><tr><td>\n\t\t\t\tAcenapthylene\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t9.374\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t11.73\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\"><\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1000\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t0.37\n\t\t\t<\/td><\/tr><tr><td>\n\t\t\t\t3-Nitroaniline\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t9.484\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t18.30\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t0.995\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t10\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1000\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t2.89\n\t\t\t<\/td><\/tr><tr><td>\n\t\t\t\tAcenaphthene\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t9.592\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t9.44\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t0.991\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1000\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t2.61\n\t\t\t<\/td><\/tr><tr><td>\n\t\t\t\t2,4-Dinitrophenol\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t9.617\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t7.79\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t0.995\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t10\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1000\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t11.53\n\t\t\t<\/td><\/tr><tr><td>\n\t\t\t\t4-Nitrophenol\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t9.689\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t14.01\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t0.998\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t10\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1000\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t4.39\n\t\t\t<\/td><\/tr><tr><td>\n\t\t\t\t2,4-Dinitrotoluene\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t9.778\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t14.84\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t0.999\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t10\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1000\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t4.30\n\t\t\t<\/td><\/tr><tr><td>\n\t\t\t\tDibenzofuran\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t9.805\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t9.94\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\"><\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1000\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t0.38\n\t\t\t<\/td><\/tr><tr><td>\n\t\t\t\t2,3,5,6-Tetrachlorophenol\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t9.904\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t16.86\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\"><\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t5\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1000\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t2.37\n\t\t\t<\/td><\/tr><tr><td>\n\t\t\t\t2,3,4,6-Tetrachlorophenol\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t9.958\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t17.72\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t0.997\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t5\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1000\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1.74\n\t\t\t<\/td><\/tr><tr><td>\n\t\t\t\tDiethylphthalate\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t10.093\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t15.97\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\"><\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1000\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t0.67\n\t\t\t<\/td><\/tr><tr><td>\n\t\t\t\t4-Chlorophenyl phenyl ether\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t10.234\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t7.21\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\"><\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1000\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t0.57\n\t\t\t<\/td><\/tr><tr><td>\n\t\t\t\tFluorene\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t10.23\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t14.54\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\"><\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1000\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t0.37\n\t\t\t<\/td><\/tr><tr><td>\n\t\t\t\t4-Nitroaniline\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t10.234\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t18.81\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t0.993\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t10\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1000\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t2.33\n\t\t\t<\/td><\/tr><tr><td>\n\t\t\t\t4,6-Dinitro-2-methylphenol\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t10.283\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t19.99\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t0.996\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t10\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1000\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t7.99\n\t\t\t<\/td><\/tr><tr><td>\n\t\t\t\tN-Nitrosodiphenylamine\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t10.376\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t18.03\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\"><\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1000\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t0.53\n\t\t\t<\/td><\/tr><tr><td>\n\t\t\t\tDiphenylamine\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t10.376\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t18.03\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\"><\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1000\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t0.53\n\t\t\t<\/td><\/tr><tr><td>\n\t\t\t\tDiphenylhydrazine\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t10.429\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t14.83\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\"><\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t10\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1000\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t3.39\n\t\t\t<\/td><\/tr><tr><td>\n\t\t\t\t2,4,6-Tribromophenol\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t10.524\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t18.55\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t0.997\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t5\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1000\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t4.15\n\t\t\t<\/td><\/tr><tr><td>\n\t\t\t\t4-Bromophenyl phenyl ether\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t10.839\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t15.84\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\"><\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1000\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t0.74\n\t\t\t<\/td><\/tr><tr><td>\n\t\t\t\tHexachlorobenzene\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t10.922\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t12.99\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\"><\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1000\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t0.96\n\t\t\t<\/td><\/tr><tr><td>\n\t\t\t\tPentachlorophenol\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t11.159\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t16.75\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t0.997\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t5\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1000\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1.14\n\t\t\t<\/td><\/tr><tr><td>\n\t\t\t\tPhenanthrene\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t11.421\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t13.06\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\"><\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1000\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t0.18\n\t\t\t<\/td><\/tr><tr><td>\n\t\t\t\tAnthracene\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t11.484\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t15.30\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\"><\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1000\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t0.92\n\t\t\t<\/td><\/tr><tr><td>\n\t\t\t\tCarbazole\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t11.68\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t14.68\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\"><\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1000\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t0.52\n\t\t\t<\/td><\/tr><tr><td>\n\t\t\t\tDi-n-butylphthalate\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t12.131\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t15.80\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\"><\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t5\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1000\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t0.25\n\t\t\t<\/td><\/tr><tr><td>\n\t\t\t\tFluoranthene\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t12.9\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t10.80\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\"><\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1000\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t0.26\n\t\t\t<\/td><\/tr><tr><td>\n\t\t\t\tBenzidine\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t13.062\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t19.24\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t0.998\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t10\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1000\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1.43\n\t\t\t<\/td><\/tr><tr><td>\n\t\t\t\tPyrene\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t13.18\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t10.13\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\"><\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1000\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t0.54\n\t\t\t<\/td><\/tr><tr><td>\n\t\t\t\to-Terphenyl-D14\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t13.377\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t13.45\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\"><\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1000\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t0.62\n\t\t\t<\/td><\/tr><tr><td>\n\t\t\t\tButylbenzylphthalate\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t14.012\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t15.67\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\"><\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t5\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1000\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t0.76\n\t\t\t<\/td><\/tr><tr><td>\n\t\t\t\tBis(2-ethylhexyl)adipate\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t14.129\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t18.95\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t0.994\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t10\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1000\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1.15\n\t\t\t<\/td><\/tr><tr><td>\n\t\t\t\t3&#8217;3-Dichlorobenzidine\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t14.769\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t17.70\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t0.996\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t10\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1000\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1.76\n\t\t\t<\/td><\/tr><tr><td>\n\t\t\t\tBenz[a]anthracene\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t14.805\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t11.61\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\"><\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1000\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t0.38\n\t\t\t<\/td><\/tr><tr><td>\n\t\t\t\tBis(2-ethylhexyl)phthalate\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t14.875\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t19.67\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\"><\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t5\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1000\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t0.58\n\t\t\t<\/td><\/tr><tr><td>\n\t\t\t\tChrysene\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t14.861\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t8.91\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\"><\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1000\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t0.34\n\t\t\t<\/td><\/tr><tr><td>\n\t\t\t\tDi-n-octylphthalate\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t16\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t18.78\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t0.998\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t5\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1000\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t0.49\n\t\t\t<\/td><\/tr><tr><td>\n\t\t\t\tBenzo[b]fluoranthene\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t16.691\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t11.15\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\"><\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1000\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t0.41\n\t\t\t<\/td><\/tr><tr><td>\n\t\t\t\tBenzo[k]fluoranthene\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t16.746\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t13.09\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\"><\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1000\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t0.50\n\t\t\t<\/td><\/tr><tr><td>\n\t\t\t\tBenzo[a]pyrene\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t17.338\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t12.32\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\"><\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1000\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t0.24\n\t\t\t<\/td><\/tr><tr><td>\n\t\t\t\tDibenz[a,j]acridine\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t19.26\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t19.92\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\"><\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1000\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t0.60\n\t\t\t<\/td><\/tr><tr><td>\n\t\t\t\tIndeno[1,2,3-cd]pyrene\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t19.754\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t15.17\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\"><\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1000\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t0.71\n\t\t\t<\/td><\/tr><tr><td>\n\t\t\t\tDibenz[a,h]anthracene\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t19.813\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t13.06\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\"><\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1000\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t0.56\n\t\t\t<\/td><\/tr><tr><td>\n\t\t\t\tBenzo[g,h,i]perylene\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t20.339\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t10.79\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\"><\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t1000\n\t\t\t<\/td><td class=\"has-text-align-center\" data-align=\"center\">\n\t\t\t\t0.50\n\t\t\t<\/td><\/tr><\/tbody><\/table><\/figure>\n\n<h2 class=\"wp-block-heading\">Conclusion<\/h2>\n\n<p class=\"wp-block-paragraph\">This study confirmed the suitability of RMX-5Sil MS columns for trace-level (single digit ppb) semivolatiles analysis by GC-MS\/MS. Overall, the column demonstrated inertness for a broad range of compounds chemistries (acids, bases, and neutrals) and generated excellent linear calibrations. Peak shape and symmetry were intact even at low levels where active compounds are exceptionally difficult to analyze. Nearly all compounds (98%) achieved IDLs at or below single-picogram levels; calibration ranges of 1, 5, or 10 ppb to 1000 ppb; and calibration curves with \u226420% RSE.<\/p>\n\n<h2 class=\"wp-block-heading\">References<\/h2>\n\n<ol class=\"wp-block-list\">\n<li>U.S. Environmental Protection Agency, Method 8270E, Semivolatile organic compounds by gas chromatography\/mass spectrometry, June 2018. <a href=\"https:\/\/www.epa.gov\/sites\/default\/files\/2020-10\/documents\/method_8270e_update_vi_06-2018_0.pdf\" target=\"_blank\" rel=\"noopener\">https:\/\/www.epa.gov\/sites\/default\/files\/2020-10\/documents\/method_8270e_update_vi_06-2018_0.pdf<\/a><\/li>\n\n\n\n<li>U.S. Environmental Protection Agency, Method 3511 (SW-846), Organic compounds in water by microextraction, July 2014. <a href=\"https:\/\/www.epa.gov\/hw-sw846\/sw-846-test-method-3511-organic-compounds-water-microextraction\" target=\"_blank\" rel=\"noopener\">https:\/\/www.epa.gov\/hw-sw846\/sw-846-test-method-3511-organic-compounds-water-microextraction<\/a><\/li>\n<\/ol>\n\n<div style=\"height:30px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n\n<figure class=\"wp-block-image size-large\"><a href=\"https:\/\/discover.restek.com\/it\/rmx\/?utm_source=website&amp;utm_medium=button&amp;utm_campaign=RMXByRestek&amp;utm_content=EVAN5254UNV_learn\"><img decoding=\"async\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/\/RMX-Graphic-Ad-What-Makes-RMX-Better-HTML-1024x874.jpg\" alt=\"RMX benefits graphic\" class=\"wp-image-61438\" title=\"-\"><\/a><\/figure>\n\n<div style=\"height:30px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n        <div class=\"cpb\">\n            <h3 class=\"cpb-heading\">Products Mentioned<\/h3>\n            <hr class=\"cpb-heading-underline\" \/>\n            <div class=\"cpb-list\">\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/17323\">                                Catalog No. 17323                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">RMX-5Sil MS GC Capillary Column, 30 m, 0.25 mm ID, 0.25 \u00b5m<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/17323\">View Product<\/a>\n                                                    <\/div>\n                    <\/div>\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/23336\">                                Catalog No. 23336                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">Topaz, Single Taper Inlet Liner, 3.5 mm x 5.0 x 95, for Shimadzu 17A, 2010, 2014, and 2030 GCs, w\/Quartz Wool Near Bottom, Premium Deactivation, 5-pk.<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/23336\">View Product<\/a>\n                                                    <\/div>\n                    <\/div>\n                            <\/div>\n        <\/div>\n        ","protected":false},"excerpt":{"rendered":"<p>This independent evaluation tested the suitability of RMX-5Sil MS columns for trace-level semivolatiles analysis. The column was shown to be inert and, in combination with a Shimadzu GC-MS\/MS, generated excellent calibration results across a wide range of chemical classes at pg levels of detection. <\/p>\n","protected":false},"author":70,"featured_media":68062,"comment_status":"closed","ping_status":"open","sticky":false,"template":"","format":"standard","meta":{"_acf_changed":false,"_kad_blocks_custom_css":"","_kad_blocks_head_custom_js":"","_kad_blocks_body_custom_js":"","_kad_blocks_footer_custom_js":"","_kadence_starter_templates_imported_post":false,"_kad_post_transparent":"","_kad_post_title":"","_kad_post_layout":"","_kad_post_sidebar_id":"","_kad_post_content_style":"","_kad_post_vertical_padding":"","_kad_post_feature":"","_kad_post_feature_position":"","_kad_post_header":false,"_kad_post_footer":false,"footnotes":""},"categories":[784],"tags":[2671,2738],"industries-application":[2163,2161,2166,2668],"post-badge":[29],"resource-type":[],"product-library":[2384,2372],"resource-technique":[2294,2302],"ppma_author":[2657],"class_list":["post-86362","post","type-post","status-publish","format-standard","has-post-thumbnail","hentry","category-application-notes-it","tag-GC-MSMSmethod","tag-rmx-resources","industries-application-drinking-water-wastewater-testing","industries-application-environmental","industries-application-soil-sediment-biota-testing","industries-application-semivolatiles","post-badge-new","product-library-gc-columns","product-library-gas-chromatography-products","resource-technique-gas-chromatography-gc","resource-technique-ms-ms"],"acf":[],"taxonomy_info":{"category":[{"value":784,"label":"Application Notes"}],"post_tag":[{"value":2671,"label":"GC-MS\/MS method"},{"value":2738,"label":"RMX Resources"}],"industries-application":[{"value":2163,"label":"Drinking Water &amp; Wastewater Testing"},{"value":2161,"label":"Environmental"},{"value":2166,"label":"Soil, Sediment, Biota Testing"},{"value":2668,"label":"Semivolatiles"}],"post-badge":[{"value":29,"label":"New"}],"product-library":[{"value":2384,"label":"GC Columns"},{"value":2372,"label":"Gas Chromatography Products"}],"resource-technique":[{"value":2294,"label":"Gas Chromatography (GC)"},{"value":2302,"label":"MS\/MS"}]},"featured_image_src_large":["https:\/\/discover.restek.com\/wp-content\/uploads\/feature-EVAN5254-1024x536.jpg",1024,536,true],"author_info":{"display_name":"Ramkumar Dhandapani, PhD","author_link":"https:\/\/discover.restek.com\/es\/author\/ramkumar-dhandapani-phd\/"},"comment_info":0,"category_info":[{"term_id":784,"name":"Application Notes","slug":"application-notes-it","term_group":0,"term_taxonomy_id":784,"taxonomy":"category","description":"","parent":0,"count":77,"filter":"raw","cat_ID":784,"category_count":77,"category_description":"","cat_name":"Application Notes","category_nicename":"application-notes-it","category_parent":0}],"tag_info":[{"term_id":2671,"name":"GC-MS\/MS method","slug":"GC-MS\/MS method","term_group":0,"term_taxonomy_id":2671,"taxonomy":"post_tag","description":"","parent":0,"count":1,"filter":"raw"},{"term_id":2738,"name":"RMX Resources","slug":"rmx-resources","term_group":0,"term_taxonomy_id":2738,"taxonomy":"post_tag","description":"Explore RMX Resources Now","parent":0,"count":23,"filter":"raw"}],"authors":[{"term_id":2657,"user_id":70,"is_guest":0,"slug":"ramkumar-dhandapani-phd","display_name":"Ramkumar Dhandapani, PhD","avatar_url":{"url":"https:\/\/discover.restek.com\/wp-content\/uploads\/Dhandapani_Ramkumar_Profile.jpg","url2x":"https:\/\/discover.restek.com\/wp-content\/uploads\/Dhandapani_Ramkumar_Profile.jpg"},"author_category":"1","first_name":"Ramkumar","last_name":"Dhandapani","user_url":"","job_title":"Director of Product Management","description":"Dr. Ramkumar Dhandapani\u00a0is a seasoned analytical chemist with over\u00a023 years of experience\u00a0in the chromatography industry and a Ph.D. in analytical chemistry. During his career, he has specialized in method development, validation, and the troubleshooting of chromatography methods. He has developed numerous regulatory-compliant methods across diverse sectors, including\u00a0environmental analysis, food quality and safety, pharmaceutical, fuels, and chemical industries. Currently, Dr. Dhandapani is the Director of Product Management at Restek, he is keen on innovation in chromatography and scaling breakthrough innovations\u00a0to market as commercial products."}],"_links":{"self":[{"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/posts\/86362","targetHints":{"allow":["GET"]}}],"collection":[{"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/posts"}],"about":[{"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/types\/post"}],"author":[{"embeddable":true,"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/users\/70"}],"replies":[{"embeddable":true,"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/comments?post=86362"}],"version-history":[{"count":1,"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/posts\/86362\/revisions"}],"predecessor-version":[{"id":86363,"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/posts\/86362\/revisions\/86363"}],"wp:featuredmedia":[{"embeddable":true,"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/media\/68062"}],"wp:attachment":[{"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/media?parent=86362"}],"wp:term":[{"taxonomy":"category","embeddable":true,"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/categories?post=86362"},{"taxonomy":"post_tag","embeddable":true,"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/tags?post=86362"},{"taxonomy":"industries-application","embeddable":true,"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/industries-application?post=86362"},{"taxonomy":"post-badge","embeddable":true,"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/post-badge?post=86362"},{"taxonomy":"resource-type","embeddable":true,"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/resource-type?post=86362"},{"taxonomy":"product-library","embeddable":true,"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/product-library?post=86362"},{"taxonomy":"resource-technique","embeddable":true,"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/resource-technique?post=86362"},{"taxonomy":"author","embeddable":true,"href":"https:\/\/discover.restek.com\/es\/wp-json\/wp\/v2\/ppma_author?post=86362"}],"curies":[{"name":"wp","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}