Application Notes

Optimizing Method Standard HJ 834-2017 for GC-MS Semivolatiles Analysis

Increase Speed and Certainty with Highly Inert RMX-5Sil MS Columns

31 Oct 2025

young plant sprouting in soil

Key Highlights

  • Next-generation TriMax GC column deactivation creates a robust and exceptionally neutral sample flow path.
  • Maximum symmetry reduces peak tailing for a wide range of challenging semivolatiles—including acids, bases, and neutrals—improving identification and quantification.
  • Optimized GC-MS conditions speed up sample throughput while still meeting method criteria.
helpful woman on phone

Abstract

RMX-5Sil MS columns feature a groundbreaking deactivation technology that creates a highly inert surface that is broadly effective across a wide range of semivolatile compound classes. In this study, we evaluated the columns performance against the data quality criteria in Method Standard HJ 834-2017. Method criteria for system suitability, linearity, and recovery were easily met with optimized conditions that reduced sample analysis time by 20 minutes.

Introduction

Semivolatile compounds are monitored worldwide to protect human health and the environment. Analysis can be done with a variety of detectors, with GC-MS techniques being among the most popular. The People’s Republic of China Method Standard HJ 834-2017 is used for GC-MS analysis of semivolatiles in soil and sediment samples. To ensure laboratories generate accurate results, Method Standard HJ 834-2017 specifies criteria for key parameters, including system suitability, linearity, and recovery. Since semivolatiles comprise a wide range of compound chemistries, including reactive compounds, it is essential that the sample flow path be highly inert. Activity in the flow path may contribute to poor peak shape and drifting retention times, which can make it difficult to meet data quality requirements.

In this study, we demonstrate how exceptionally inert RMX-5Sil MS columns can help laboratories meet data quality objectives for Method Standard HJ 834-2017 semivolatiles analysis. RMX-5Sil MS columns undergo a unique TriMax surface deactivation that eliminates active sites (e.g., silanols) and is broadly effective across compound classes. The RMX-5Sil MS column contains a traditional 5sil polymer, so it is a direct 5sil replacement, but the highly neutral surface improves peak shape for a wide variety of compounds, making it easier to meet data quality requirements. In this study, in addition to demonstrating the effectiveness of the column for GC-MS semivolatiles analysis, we also developed optimized instrument conditions that speed up run times and improve sample throughput.

Experimental

Standard and Sample Preparation

Multicomponent semivolatiles standards were prepared at 1, 5, 10, 20, and 50 ppm in dichloromethane for calibration. Internal standards and surrogate standards were present at 40 ppm in each calibration standard.

Instrument Conditions

GC-MS semivolatiles analysis was performed following the conditions in Method Standard HJ 834-2017. In addition, semivolatiles analysis was run under optimized conditions to allow faster analysis times. Table I summarizes the instrument conditions for both approaches.

Table I: Instrument Conditions for GC-MS Semivolatiles Analysis

Parameter Method Standard HJ 834-2017 Reference Conditions Optimized Conditions
Column RMX-5Sil MS, 30 m x 0.25 mm ID x 0.25 µm (cat.# 17323) RMX-5Sil MS, 30 m x 0.25 mm ID x 0.25 µm (cat.# 17323)
Injection 1 µL, splitless, 280 °C 1 µL, splitless, 280 °C
Carrier gas Helium at 1 mL/min (constant flow) Helium at 1 mL/min (constant flow)
Oven 35 °C (hold 2 min) to 150 °C at 15 °C/min (hold 5 min) to 290 °C at 3 °C/min (hold 2 min) Faster oven conditions speed up analysis:
35 °C (hold 1 min) to 290 °C at 8 °C/min to 350 °C at 12 °C/min (hold 5 min)
Detector EI; 70 eV; 230 °C source; 150 °C quad; scan 35-450 m/z; solvent delay 5 min; 280 °Ctransfer line EI; 70 eV; 230 °C source; 150 °C quad; scan 35-450 m/z; solvent delay 2 min; 280 °Ctransfer line

Results and Discussion

System Suitability

The instrument was tuned with DFTPP in accordance with M0ethod Standard HJ 834-2017, and then system suitability was assessed by evaluating the degradation of DDT to DDE and DDD. The requirement to not exceed 15% was easily met with DDT degradation to both DDE and DDD being <1%.

Chromatographic Performance

As shown in Figure 1, the RMX-5Sil MS column produced sharp, symmetrical peaks for 64 monitored semivolatiles, six surrogates, and six internal standards (76 compounds total) at 5 ppm under both the original Method Standard HJ 834-2017 conditions and the modified GC-MS conditions. While some coelutions occurred, these compounds could easily be differentiated by their ion m/z ratios. The standard method conditions produced good results, but the >60-minute run time limits the number of samples that can be analyzed. The improved method conditions resulted in a faster 40-minute run time, which is beneficial to high-throughput labs that have many samples to analyze under tight deadlines.

Figure 1: TIC Chromatogram of Method Standard HJ 834-2017 Conditions (>60 minutes) 

HJ 834-2017 Volatiles and Semivolatiles on RMX-5Sil MS (Method Conditions)

GC_EV1533

Peaks

PeakstR (min)
1.Bis(N-methoxy-N-methylamino)methane3.989
2.Phenol, 2-fluoro-5.769
3.Phenol-d6-7.057
4.Phenol7.069
5.Bis(2-chloroethyl) ether7.194
6.Phenol, 2-chloro-7.269
7.Benzene, 1,3-dichloro-7.496
8.1,4-Dichlorobenzene-D47.579
9.Benzene, 1,4-dichloro-7.602
10.Benzene, 1,2-dichloro-7.825
11.Phenol, 2-methyl-7.956
12.Bis(2-chloro-1-methylethyl) ether7.995
13.1-Propanamine, N-nitroso-N-propyl-8.183
14.p-Cresol8.183
15.Ethane, hexachloro-8.336
16.Nitrobenzene-D58.393
17.Benzene, nitro-8.421
18.Isophorone8.795
19.Phenol, 2-nitro-8.915
20.Phenol, 2,3-dimethyl-9.002
21.Methane, bis(2-chloroethoxy)-9.146
22.Phenol, 2,4-dichloro-9.290
23.Benzene, 1,2,4-trichloro-9.427
24.Naphthalene-D89.508
25.Naphthalene9.540
PeakstR (min)
26.p-Chloroaniline9.630
27.1,3-Butadiene, 1,1,2,3,4,4-hexachloro-9.767
28.Phenol, 2-chloro-5-methyl-10.486
29.Naphthalene, 1-methyl-10.766
30.Hexachlorocyclopentadiene11.140
31.Phenol, 2,4,6-trichloro-11.412
32.Phenol, 2,4,5-trichloro-11.486
33.1,1′-Biphenyl, 2-fluoro-11.640
34.Naphthalene, 1-chloro-11.915
35.Dimethyl (2-nitroanilino)maleate12.221
36.Dimethyl phthalate12.926
37.Benzene, 2-methyl-1,3-dinitro-13.078
38.Biphenylene13.231
39.m-Nitroaniline13.680
40.Acenaphthene-d1013.813
41.Acenaphthene13.947
42.Phenol, 2,4-dinitro-14.149
43.Phenol, 4-nitro-14.626
44.Dibenzofuran14.763
45.Benzene, 1-methyl-2,4-dinitro-14.820
46.Diethyl phthalate16.434
47.Fluorene16.576
48.Benzene, 1-chloro-4-phenoxy-16.802
49.Benzene, 1-chloro-3-phenoxy-16.802
50.Phenol, 2-methyl, 4,6-dinitro-17.043
PeakstR (min)
51.Azobenzene17.714
52.Phenol, 2,4,6-tribromo-17.972
53.Benzene, 1-bromo-4-phenoxy-19.798
54.Benzene, hexachloro-19.924
55.Phenol, pentachloro-21.347
56.Phenanthrene-D1022.411
57.9H-Fluorene, 9-methylene-22.561
58.Anthracene22.911
59.Carbazole24.334
60.Dibutyl phthalate28.107
61.Fluoranthene31.814
62.Pyrene32.727
63.p-Terphenyl-d1434.881
64.Benzyl butyl phthalate39.418
65.Benz[a]anthracene42.103
66.Chrysene-D1242.103
67.Chrysene42.298
68.Bis(2-ethylhexyl) phthalate44.571
69.Phthalic acid, hept-4-yl octyl ester49.346
70.Benzo[b]fluoranthene49.728
71.Benzo[k]fluoranthene49.915
72.Benzo[a]pyrene51.738
73.Perylene – D1252.132
74.Indeno[1,2,3-cd]pyrene58.454
75.Diben[a,h]anthracene58.786
76.Benzo[ghi]perylene59.786

Conditions

ColumnRMX-5Sil MS, 30 m, 0.25 mm ID, 0.25 µm (cat.# 17323)
Standard/Sample
1000 ppm HJ 834-2017 VOC and SVOCs mixture 155 (LGC)
1000 ppm HJ 834-2017 substitutes mixture 156 (LGC)
1000 ppm HJ 834-2017 internal standard mixture 174 (internal standard)
Diluent:Dichloromethane
Conc.:5 µg/mL
Injection
Inj. Vol.:1 µL splitless (hold 1 min)
Liner:Topaz 4.0 mm ID single taper inlet liner w/wool (cat.# 23303)
Inj. Temp.:280 °C
Oven
Oven Temp.:35 °C (hold 2 min) to 150 °C at 15 °C/min (hold 5 min) to 290 °C at 3 °C/min (hold 2 min)
Carrier GasHe, constant flow
Flow Rate:1 mL/min @ 35 °C
DetectorMS
Mode:Scan
Scan Program:
GroupStart Time
(min)
Scan Range
(amu)
Scan Rate
(scans/sec)
12.535-4501562
Transfer Line Temp.:280 °C
Source Temp.:230 °C
Quad Temp.:150 °C
Electron Energy:70 eV
Tune Type:PFTBA
Ionization Mode:EI
InstrumentAgilent 7890B GC & 5977B MSD
Sample PreparationReference standards were diluted to 5 ppm in dichloromethane.

Figure 2: TIC Chromatogram of Optimized Method Conditions (40 minutes) 

HJ 834-2017 Volatiles and Semivolatiles on RMX-5Sil MS (Optimized Conditions)

GC_EV1532

Peaks

PeakstR (min)
1.Bis(N-methoxy-N-methylamino)methane3.941
2.Phenol, 2-fluoro-6.189
3.Phenol-d6-8.226
4.Phenol8.252
5.Bis(2-chloroethyl) ether8.417
6.Phenol, 2-chloro-8.502
7.Benzene, 1,3-dichloro-8.861
8.1,4-Dichlorobenzene-D49.01
9.Benzene, 1,4-dichloro-9.048
10.Benzene, 1,2-dichloro-9.411
11.Phenol, 2-methyl-9.726
12.Bis(2-chloro-1-methylethyl) ether9.784
13.1-Propanamine, N-nitroso-N-propyl-10.101
14.p-Cresol10.133
15.Ethane, hexachloro-10.273
16.Nitrobenzene-D510.421
17.Benzene, nitro-10.468
18.Isophorone11.137
19.Phenol, 2-nitro-11.322
20.Phenol, 2,3-dimethyl-11.561
21.Methane, bis(2-chloroethoxy)-11.822
22.Phenol, 2,4-dichloro-12.006
23.Benzene, 1,2,4-trichloro-12.23
24.Naphthalene-D812.36
25.Naphthalene12.411
PeakstR (min)
26.p-Chloroaniline12.622
27.1,3-Butadiene, 1,1,2,3,4,4-hexachloro-12.859
28.Phenol, 2-chloro-5-methyl-14.122
29.Naphthalene, 1-methyl-14.409
30.Hexachlorocyclopentadiene14.927
31.Phenol, 2,4,6-trichloro-15.284
32.Phenol, 2,4,5-trichloro-15.362
33.1,1′-Biphenyl, 2-fluoro-15.564
34.Naphthalene, 1-chloro-15.806
35.Dimethyl (2-nitroanilino)maleate16.158
36.Dimethyl phthalate16.837
37.Benzene, 2-methyl-1,3-dinitro-16.945
38.Biphenylene16.995
39.m-Nitroaniline17.381
40.Acenaphthene-d1017.45
41.Acenaphthene17.533
42.Phenol, 2,4-dinitro-17.692
43.Phenol, 4-nitro-18.011
44.Dibenzofuran18.05
45.Benzene, 1-methyl-2,4-dinitro-18.104
46.Diethyl phthalate18.972
47.Fluorene19.061
48.Benzene, 1-chloro-4-phenoxy-19.168
49.Benzene, 1-chloro-3-phenoxy-19.175
50.Phenol, 2-methyl, 4,6-dinitro-19.301
PeakstR (min)
51.Azobenzene19.626
52.Phenol, 2,4,6-tribromo-19.78
53.Benzene, 1-bromo-4-phenoxy-20.596
54.Benzene, hexachloro-20.697
55.Phenol, pentachloro-21.316
56.Phenanthrene-D1021.822
57.9H-Fluorene, 9-methylene-21.88
58.Anthracene22.025
59.Carbazole22.57
60.Dibutyl phthalate23.889
61.Fluoranthene25.396
62.Pyrene26.035
63.p-Terphenyl-d1426.717
64.Benzyl butyl phthalate28.309
65.Benz[a]anthracene29.638
66.Chrysene-D1229.684
67.Chrysene29.744
68.Bis(2-ethylhexyl) phthalate30.159
69.Phthalic acid, hept-4-yl octyl ester31.991
70.Benzo[b]fluoranthene32.608
71.Benzo[k]fluoranthene32.683
72.Benzo[a]pyrene33.417
73.Perylene – D1233.583
74.Indeno[1,2,3-cd]pyrene35.743
75.Diben[a,h]anthracene35.798
76.Benzo[ghi]perylene36.197

Conditions

ColumnRMX-5Sil MS, 30 m, 0.25 mm ID, 0.25 µm (cat.# 17323)
Standard/Sample
1000 ppm HJ 834-2017 VOC and SVOCs mixture 155 (LGC)
1000 ppm HJ 834-2017 substitutes mixture 156 (LGC)
1000 ppm HJ 834-2017 internal standard mixture 174 (internal standard)
Diluent:Dichloromethane
Conc.:5 µg/mL
Injection
Inj. Vol.:1 µL splitless (hold 1 min)
Liner:Topaz 4.0 mm ID single taper inlet liner w/wool (cat.# 23303)
Inj. Temp.:280 °C
Oven
Oven Temp.:35 °C (hold 1 min) to 290 °C at 8 °C/min to 350 °C at 12 °C/min (hold 5 min)
Carrier GasHe, constant flow
Flow Rate:1 mL/min @ 35 °C
DetectorMS
Mode:Scan
Scan Program:
GroupStart Time
(min)
Scan Range
(amu)
Scan Rate
(scans/sec)
12.535-4501562
Transfer Line Temp.:280 °C
Source Temp.:230 °C
Quad Temp.:150 °C
Electron Energy:70 eV
Tune Type:PFTBA
Ionization Mode:EI
InstrumentAgilent 7890B GC & 5977B MSD
Sample PreparationReference standards were diluted to 5 ppm in dichloromethane.
EZGC software ad

Quantitative Performance

Good peak shape improves signal-to-noise ratios, which allows reliable integration at low concentrations and minimizes the need for time-consuming user intervention. Accurate integration across the calibration range increases confidence that method performance criteria will be met, even for difficult reactive compounds. Method performance was assessed by evaluating linearity (unweighted, R>0.990); relative response factor (RSD <30%); and recovery of a midpoint standard (70-130% recovery at 10 ppm).

Limits of detection and quantitation were determined using generally accepted methods where LOD = 3.3 x standard deviation of the lowest calibration point/calibration slope, and LOQ = 10 x standard deviation of the lowest calibration point/calibration slope. Values were compared to the LOD and LOQ values described in the method. The method requirements show LOD and LOQ as sample concentrations. Since matrix was not used in this study, sample concentrations were converted to extract concentrations based on a theoretical sample mass of 20 g, and final extract volume of 1 mL (Table II). Without matrix, the differences between column performance can be separated from the efficacy of the sample preparation.

For quantitative analysis, calibration was performed using the optimized instrument conditions. Summarized results are presented in Table II and the values for individual semivolatiles are given in Table III. All compounds (excluding internal standards and surrogates) passed linearity criteria of RRF RSD <30% and R>0.990. RSD values ranged 2.1-22.5% and R ranged 0.999-1.00. While not required for HJ 834, R2 was also evaluated because it is often employed in other methods. In this study R2 was >0.995 for each compound, ranging 0.997-1.0.

Table II: Results Summary for Method Standard HJ 834-2017 GC-MS Semivolatiles Analysis (Optimized Conditions)

RRF %RSD (<30%) R2* (>0.990) R (>0.990) LOD (Method HJ 834-2017) LOQ (Method HJ 834-2017) %Recovery
Acceptable 64 64 64 64 64 64
Unacceptable 0 0 0 0 0 0
Maximum 23% 1.000 1.000 0.89 2.98 109%
Minimum 2% 0.997 0.999 0.02 0.05 75%
*R2 is not specified in Method Standard HJ 834-2017. It is included because it is a commonly used linearity metric in other methods for GC-MS semivolatiles analysis. 

Table III: Individual Results by Compound for Method Standard HJ 834-2017 GC-MS Semivolatiles Analysis (Optimized Conditions)

Compound RRF %RSD (<30%) R2* R Method HJ 834-2017 LOD (ppm) Method HJ 834-2017 LOQ (ppm) LOD (ppm) LOQ (ppm) %Recovery of 10 ppm
Bis(N-methoxy-N-methylamino)methane 2% 1.000 1.000 1.60 6.40 0.36 1.22 105%
Phenol 4% 0.998 0.999 2.00 8.00 0.33 1.11 95%
Bis(2-chloroethyl) ether 3% 0.999 1.000 2.00 8.00 0.17 0.58 99%
Phenol, 2-chloro- 4% 0.999 0.999 2.00 8.00 0.16 0.55 96%
Benzene, 1,3-dichloro- 5% 0.999 0.999 1.80 7.20 0.20 0.68 100%
Benzene, 1,4-dichloro- 5% 1.000 1.000 1.20 4.80 0.19 0.62 97%
Benzene, 1,2-dichloro- 7% 1.000 1.000 1.60 6.40 0.13 0.45 96%
Phenol, 2-methyl- 5% 0.998 0.999 1.60 6.40 0.25 0.82 101%
Bis(2-chloro-1-methylethyl) ether 6% 0.999 0.999 1.60 6.40 0.34 1.13 104%
1-Propanamine, N-nitroso-N-propyl- 3% 1.000 1.000 2.00 8.00 0.43 1.43 109%
p-Cresol 6% 1.000 1.000 2.00 8.00 0.11 0.38 100%
Ethane, hexachloro- 10% 1.000 1.000 2.00 8.00 0.53 1.77 97%
Benzene, nitro- 7% 0.999 1.000 1.40 5.60 0.23 0.76 102%
Isophorone 6% 1.000 1.000 2.00 8.00 0.09 0.30 99%
Phenol, 2-nitro- 10% 1.000 1.000 2.00 8.00 0.15 0.50 96%
Phenol, 2,3-dimethyl- 6% 1.000 1.000 1.80 7.20 0.25 0.84 99%
Methane, bis(2-chloroethoxy)- 5% 1.000 1.000 1.40 5.60 0.09 0.31 100%
Phenol, 2,4-dichloro- 13% 1.000 1.000 4.00 16.00 0.09 0.30 98%
Benzene, 1,2,4-trichloro- 11% 1.000 1.000 1.80 7.20 0.16 0.54 98%
Naphthalene 9% 0.998 0.999 1.60 6.40 0.16 0.54 99%
p-Chloroaniline 8% 0.999 0.999 1.40 5.60 0.17 0.57 94%
1,3-Butadiene, 1,1,2,3,4,4-hexachloro- 5% 1.000 1.000 1.40 5.60 0.29 0.98 97%
Phenol, 2-chloro-5-methyl- 6% 1.000 1.000 1.80 7.20 0.23 0.77 93%
Naphthalene, 1-methyl- 10% 0.999 0.999 1.80 7.20 0.07 0.22 98%
Hexachlorocyclopentadiene 22% 0.998 0.999 1.20 4.80 0.27 0.92 89%
Phenol, 2,4,6-trichloro- 21% 0.999 1.000 1.20 4.80 0.13 0.44 92%
Phenol, 2,4,5-trichloro- 21% 0.999 0.999 1.60 6.40 0.27 0.91 92%
Naphthalene, 1-chloro- 15% 0.999 1.000 2.00 8.00 0.10 0.32 95%
Dimethyl (2-nitroanilino)maleate 21% 0.999 1.000 2.00 8.00 0.62 2.05 89%
Dimethyl phthalate 15% 1.000 1.000 2.00 8.00 0.12 0.39 93%
Benzene, 2-methyl-1,3-dinitro- 20% 0.998 0.999 2.00 8.00 0.56 1.87 90%
Biphenylene 18% 1.000 1.000 2.00 8.00 0.10 0.32 99%
m-Nitroaniline 22% 0.999 1.000 1.60 6.40 0.89 2.98 90%
Acenaphthene 10% 0.998 0.999 1.80 7.20 0.02 0.08 99%
Phenol, 2,4-dinitro- 16% 0.997 0.999 1.40 5.60 0.69 2.32 75%
Phenol, 4-nitro- 13% 1.000 1.000 1.60 7.20 0.23 0.75 85%
Dibenzofuran 14% 1.000 1.000 2.00 8.00 0.09 0.29 99%
Benzene, 1-methyl-2,4-dinitro- 23% 1.000 1.000 2.00 8.00 0.13 0.42 91%
Diethyl Phthalate 13% 1.000 1.000 2.00 8.00 0.05 0.18 96%
Fluorene 15% 0.999 1.000 1.80 7.20 0.04 0.14 98%
Benzene, 1-chloro-4-phenoxy- 8% 1.000 1.000 1.80 7.20 0.12 0.39 103%
Benzene, 1-chloro-3-phenoxy- 13% 1.000 1.000 4.00 16.00 0.33 1.11 95%
Phenol, 2-methyl, 4,6-dinitro- 22% 0.998 0.999 1.60 6.40 0.39 1.30 103%
Azobenzene 6% 0.999 0.999 6.00 24.00 0.15 0.50 104%
Benzene, 1-bromo-4-phenoxy- 13% 1.000 1.000 2.00 8.00 0.20 0.67 95%
Benzene, hexachloro- 8% 1.000 1.000 2.00 8.00 0.02 0.07 99%
Phenol, pentachloro- 11% 0.998 0.999 2.00 8.00 0.39 1.30 89%
9H-Fluorene, 9-methylene- 8% 0.998 0.999 2.00 8.00 0.03 0.11 101%
Anthracene 13% 1.000 1.000 4.00 16.00 0.03 0.09 104%
Carbazole 11% 0.998 0.999 2.00 8.00 0.07 0.24 96%
Dibutyl phthalate 12% 0.999 0.999 2.00 8.00 0.03 0.10 99%
Fluoranthene 10% 1.000 1.000 4.00 16.00 0.07 0.22 100%
Pyrene 9% 1.000 1.000 2.00 8.00 0.03 0.11 101%
Benzyl butyl phthalate 12% 0.998 0.999 2.00 8.00 0.20 0.68 94%
Benz[a]anthracene 5% 0.999 0.999 2.00 8.00 0.08 0.26 95%
Triphenylene 7% 0.999 1.000 2.00 8.00 0.13 0.42 106%
Bis(2-ethylhexyl) phthalate 15% 1.000 1.000 4.00 16.00 0.09 0.29 102%
Phthalic acid, hept-4-yl octyl ester 18% 1.000 1.000 2.00 8.00 0.02 0.05 103%
Benzo[b]fluoranthene 17% 0.999 1.000 2.00 8.00 0.09 0.31 96%
Benzo[k]fluoranthene 16% 1.000 1.000 4.00 16.00 0.09 0.32 102%
Benzo[a]pyrene 15% 0.999 0.999 2.00 8.00 0.08 0.26 104%
Indeno[1,2,3-cd]pyrene 11% 0.999 1.000 2.00 8.00 0.05 0.18 96%
Diben[a,h]anthracene 5% 0.999 0.999 2.00 8.00 0.10 0.35 107%
Benzo[ghi]perylene 7% 1.000 1.000 4.00 16.00 0.04 0.15 106%
*R2 is not specified in Method Standard HJ 834-2017. It is included because it is a commonly used linearity metric in other methods for GC-MS semivolatiles analysis. 

Conclusion

This study demonstrated that the highly inert RMX-5Sil MS column produced method-compliant results for a wide range of challenging semivolatiles. The inertness of the column surface produced sharp, symmetrical peaks that simplified integration and resulted in excellent results for system suitability; linearity (RRF %RSD and R2); LOD; LOQ; and recovery. In addition to meeting Method Standard HJ 834-2017 requirements and producing highly linear calibration curves, the optimized GC-MS conditions that were developed here reduced sample analysis time from 60 to 40 minutes, allowing high-throughput laboratories to improve sample throughput.

RMX benefits graphic

Products Mentioned


Topaz, Single Taper Inlet Liner, 4.0 mm x 6.5 x 78.5, for Agilent GCs, w/Quartz Wool, Premium Deactivation, 5-pk.
Restek Electronic Leak Detector
Benzo[a]pyrene Standard, 1000 µg/mL, Acetone, 1 mL/ampul
Dibenz[a,h]anthracene Standard, 1000 µg/mL, Methylene Chloride, 1 mL/ampul
Indeno[1,2,3-cd]Pyrene Standard, 1000 µg/mL, Methylene Chloride, 1 mL/ampul
Benzo[g,h,i]perylene Standard, 1000 µg/mL, Methylene Chloride, 1 mL/ampul

Authors

  • Erica Pack, PhD

    Erica is an interdisciplinary GC applications and technologies scientist at Restek. She obtained her bachelor's degree in forensic biology from The Pennsylvania State University, and her doctorate from Virginia Tech in plant pathology, physiology, and weed science. Since joining Restek in 2021, she has worked with a wide variety of GC columns, including fused silica, MXT, PLOT, and packed columns as well as accessories, such as liners, valves, and methanizers.

    View all posts
  • Grace Idowu

    Dr. Grace Idowu was formerly with Restek as a principal scientist in the GC applications laboratory. She received her PhD in chemistry from the University of Manitoba, Canada; her master’s degree in environmental and petroleum geochemistry from Newcastle University, UK; and her bachelor of technology degree in pure and applied chemistry from Ladoke Akintola University, Nigeria. Grace's doctoral and post-doctoral research focused on the development and validation of GC methods for polycyclic aromatic compounds in environmental matrices. She has extensive experience developing analytical methods for GC-MS, GC-MS/MS, and GCxGC-MS, and her current research focuses primarily on persistent organic pollutants in environmental matrices.

    View all posts
  • Chris English

    Since 2004, Chris has managed a team of chemists in Restek's innovations laboratory who perform new product testing, method development, and applications work. Before taking the reins of our lab, he spent seven years as an environmental chemist and was critical to the development of Restek’s current line of volatile GC columns. Prior to joining Restek, he operated a variety of gas chromatographic detectors conducting method development and sample analysis. Chris holds a BS in environmental science from Saint Michael's College.

    View all posts
  • Ramkumar Dhandapani

    Dr. Ramkumar Dhandapani is a seasoned analytical chemist with over 23 years of experience in the chromatography industry and a Ph.D. in analytical chemistry. During his career, he has specialized in method development, validation, and the troubleshooting of chromatography methods. He has developed numerous regulatory-compliant methods across diverse sectors, including environmental analysis, food quality and safety, pharmaceutical, fuels, and chemical industries. Currently, Dr. Dhandapani is the Director of Product Management at Restek, he is keen on innovation in chromatography and scaling breakthrough innovations to market as commercial products.

    View all posts
  • Colton Myers

    Colton Myers is the R&D manager for sample preparation at Restek Corporation with over 10 years of experience in product development and application innovation, particularly in solid phase microextraction (SPME) and volatile analysis. He has made contributions across various industries, authoring multiple peer-reviewed publications. Starting his career in quality control before transitioning to the GC Innovations team, Colton now leads a team dedicated to advancing sample preparation and collection technologies. He holds a BS in chemistry from Juniata College.

    View all posts
EVAN5233