Blog

Column selection for USP <467> using ProEZGC

28 Feb 2024

There are a lot of decisions to be made when performing USP <467>. With three classes of solvents to potentially monitor, two phase materials to choose from, two dimensions for each phase, and the option to select and validate your column of choice, it can be daunting to start thinking about analysis. In addition, unless you are making reference standards in-house, residual solvent stocks are sub-divided into a series of mixtures to be run individually, rather than one ‘megamix’. This helps improve separations, save time, and make your lab more efficient, because one run with fewer compounds is faster and easier to analyze, but it also means adjusting methodology when it comes time to monitor a new solvent, which can cause a rift in efficiency. This is where Restek’s ProEZGC can help.

Reference standard selection:

The first challenge to overcome is what reference standards are needed. The most up-to-date stock solutions offered at Restek are:

Table 1: USP <467> reference standards available through Restek, color coordinated with Tables 3 and 5.

Cat.# Mix
36279 Class 1
36015 Class 2 Mix A
36280 Class 2 Mix B
36273 Class 2 Mix C
36016 Class 3 Mix A
36014 Class 3 Mix B

These mixes represent changes that have been implemented in chapter <467> as recently as 2020. They were also used to build our ProEZGC libraries using the recommended Rxi-624sil MS (G43) and Stabilwax (G16) phases. 

Column selection:

USP <467> recommends four possible columns, comprised of two phases: G16 and G43, and two dimensions for each phase. The Restek equivalents for these columns are as follows:

Table 2: Restek columns available in recommended USP <467> phases and dimensions.

Cat.#Mix
10624(G16) Stabilwax GC Capillary Column, 30 m, 0.32 mm ID, 0.25 µm
10625(G16) Stabilwax GC Capillary Column, 30 m, 0.53 mm ID, 0.25 µm
13870(G43) Rxi-624Sil MS GC Capillary Column, 30 m, 0.32 mm ID, 1.80 µm
13871(G43) Rxi-624Sil MS GC Capillary Column, 30 m, 0.53 mm ID, 3.00 µm

To start exploring separations on these phases and dimensions, go to ProEZGC. The simplest way to get started is to copy/paste your list of compounds either by name or CAS#, into the appropriate textbox. You can search mix by mix, or search for all mixes at once. For this example, I have searched for all compounds at once:

The first column recommended is a G16 phase. By going to the stationary phase results dropdown, we see the Rxi-624sil MS is recommended next. So here is where you get to decide how to proceed – you can move forward with either phase! But, for the sake of this article, I moved forward with the Stabilwax.

The search window for ProEZGC

Figure 1: The search window for ProEZGC. Arrows indicate the solve button, and dropdown of column phases with solvent libraries available for modeling.

The default dimensions recommended did not match the USP <467>, and I was curious why not. So, I went into the conditions tab, selected the recommended dimensions, and optimized some separations for maximum column efficiency. I then compared separations between recommended Stabilwax column dimensions, and the default dimensions offered.

ProEZGC conditions tab with default options

Figure 2: ProEZGC conditions tab with default options. Arrows indicate the conditions tab, column dimensions dropdown menu, USP <467> recommended column dimensions, oven refinement and optimization button, and radio button to optimize by efficiency.

When comparing the separations across columns, I primarily looked at separations within mixes. While there were many co-elutions between mixes, those can be easily separated by running just one mix at a time. The separations within mixes offer a more complex challenge. 

Below is a table showing the predicted separations across different Stabilwax (G16) columns, where co-elutions within mixes are boxed in red. Different colored rows represent different mixes, and red-cells highlight all co-elutions determined by ProEZGC. After optimization, the recommended column dimensions, cat.# 10624 and 10625, each had four within-mix co-elutions, while the default recommendation by ProEZGC (cat.# 10654) had only two.

Table 3: Comparison of G16 column dimensions by co-elution. Colors represent different mixes (see Table 1), red represents co-elutions, red-boxes represent within-mix co-elutions.

# Stabilwax Compound Stabilwax cat.# 10624 Stabilwax cat.# 10625 Stabilwax cat.# 10654
Compound RT Res RT Res RT
1 n-Pentane 1.74 4 2.36 1.9 1.86 14.4
2 C6 1.82 2.4 2.43 1.2 2.18 7.4
3 Diethyl ether 1.87 2.4 2.47 1.2 2.37 7.4
4 Methyl-tert-butyl ether 1.99 1.2 2.57 0.6 2.84 2.9
5 n-Heptane 2.01 1.2 2.59 0.6 2.94 2.9
6 Cyclohexane 2.1 2.4 2.66 1.2 3.28 5.3
7 1,1-Dichloroethene 2.15 2.4 2.71 1.2 3.5 5.3
8 Methylcyclohexane 2.31 6.7 2.88 3.5 4.13 13.9
9 Acetone 2.65 4.4 3.12 2.5 5.5 7.1
10 Ethyl formate 2.79 0.3 3.23 0.1 6.02 0.4
11 Methyl acetate 2.79 0.3 3.24 0.1 6.06 0.4
12 Tetrahydrofuran 3.05 1.3 3.45 0.7 7.03 1.6
13 Triethylamine 3.09 1.3 3.48 0.7 7.17 1.6
14 trans-1,2-Dichloroethene 3.35 0.4 3.7 0.3 7.95 1.2
15 1,1,1-Trichloroethane 3.37 0.1 3.71 0.1 8.05 0.1
16 Carbon tetrachloride 3.38 0.1 3.72 0.1 8.07 0.1
17 Ethyl acetate 3.49 2.9 3.81 1.8 8.37 3.7
18 Methanol 3.66 1.9 3.99 1.1 8.79 2.5
19 2-Butanone 3.75 1.9 4.03 1.1 8.98 2.5
20 Isopropyl acetate 4.12 0.1 4.33 — 9.51 3.5
21 1,2-Dimethoxyethane 4.12 0.1 4.34 — 9.75 3.3
22 2-Propanol 4.29 1.3 4.48 0.8 10 1.9
23 Methylene chloride 4.36 1.3 4.54 0.8 10.14 1.9
24 Benzene 4.48 2.2 4.63 1.4 10.39 3.3
25 Ethanol 4.81 6.1 4.91 4 10.81 5.7
26 Propyl acetate 5.59 6.3 5.56 4.5 11.82 5.8
27 Trichloroethene 6.02 5.3 5.94 3.7 12.23 5.7
28 Acetonitrile 6.38 1.3 6.27 0.9 12.57 1.5
29 4-Methyl-2-pentanone 6.47 1.3 6.35 0.9 12.65 1.5
30 Isobutyl acetate 6.7 3.3 6.55 2.3 12.77 2.2
31 Chloroform 6.94 2.3 6.76 1.7 12.91 0.6
32 2-Butanol 7.1 2.3 6.9 1.7 12.94 0.6
33 Toluene 7.34 1.9 7.09 1.5 13.15 4
34 1-Propanol 7.47 1.9 7.2 1.5 13.32 4
35 1,4-Dioxane 7.96 7.3 7.58 5.2 13.66 3.2
36 1,2-Dichloroethane 8.44 — 7.94 — 13.79 —
37 Butyl acetate 8.45 — 7.95 — 13.8 —
38 2-Hexanone 8.64 3 8.09 2.1 13.95 2.1
39 Isobutanol 9.12 7.5 8.43 5.4 14.03 2.1
40 Ethylbenzene 9.95 3.5 9 2.5 14.65 1.5
41 p-Xylene 10.17 2.8 9.14 2 14.7 1.5
42 m-Xylene 10.35 2.8 9.26 2 14.75 1.5
43 1-Butanol 10.66 4.9 9.45 3.4 14.83 2.2
44 Nitromethane 11.01 4.3 9.66 3 14.95 3.5
45 Isopropylbenzene 11.26 2.2 9.81 1.5 15.17 0.1
46 Pyridine 11.39 1.5 9.88 0.9 15.17 0.1
47 2-Methoxyethanol 11.48 1.5 9.93 0.9 15.26 2.7
48 3-Methyl-1-butanol 12.19 2 10.31 1.6 15.35 2.8
49 Chlorobenzene 12.28 2 10.38 1.6 15.55 2.5
50 2-Ethoxyethanol 12.39 2.5 10.44 1.6 15.63 2.5
51 1-Pentanol 12.9 12.5 10.78 8.7 15.72 3
52 N,N-Dimethylformamide 13.83 4.9 11.46 3.2 16.6 2.4
53 Anisole 13.97 4.9 11.57 3.2 16.67 2.4
54 N,N-Dimethylacetamide 14.49 17.6 11.97 11.8 17.13 5.3
55 Acetic acid 14.95 17.6 12.34 11.5 17.29 5.3
56 Tetralin 15.36 9.9 12.66 6.6 17.93 4.1
57 Dimethyl sulfoxide 15.6 1.6 12.86 1.5 18.04 4.1
58 Formic acid 15.64 1.6 12.9 1.5 18.22 1.5
59 Ethylene glycol 15.82 8.8 13.04 5.4 18.27 1.5
60 N-Methylpyrrolidone 16.16 15.5 13.3 9.5 18.9 9
61 Formamide 16.6 18.5 13.67 12.8 19.17 9
62 Sulfolane 16.65 94.8 15.35 62.6 24.18 168

The optimized GC conditions for each column are as follows:

Table 4: GC Conditions for columns in Table 3.

Column: Stabilwax, 30.00 m, 0.32 mm ID, 0.25 µm (cat.# 10624)
Carrier Gas: Helium, Constant Flow @ 1.79 mL/min
Average Velocity: 29.48 cm/sec
Outlet Pressure (abs): 14.70 psi
Oven Temp.: 30 °C (hold 5.5 min) to 70 °C @ 6.5 °C/min to 225 °C @ 26 °C/min (hold 3 min)
Column: Stabilwax, 30.00 m, 0.53 mm ID, 0.25 µm (cat.# 10625)
Carrier Gas: Helium, Constant Flow @ 2.97 mL/min
Average Velocity: 21.48 cm/sec
Outlet Pressure (abs): 14.70 psi
Oven Temp.: 30 °C (hold 6 min) to 75 °C @ 13.5 °C/min to 225 °C @ 30 °C/min (hold 3 min)
Column: Stabilwax, 30.00 m, 0.32 mm ID, 1.00 µm (cat.# 10654)
Carrier Gas: Helium, Constant Flow @ 1.79 mL/min
Average Velocity: 29.65 cm/sec
Outlet Pressure (abs): 14.70 psi
Oven Temp.: 30 °C (hold 6.5 min) to 65 °C @ 6.5 °C/min to 225 °C @ 23 °C/min (hold 7 min)

I repeated this experiment for the Rxi-624sil MS to preview what separations I would encounter on the alternative set of phases and dimensions. Below is a table showing the predicted separations across different columns, this time using the Rxi-624sil MS phase. Different colored rows represent different mixes, and red-cells highlight all co-elutions determined by ProEZGC. After optimization, the recommended column dimension cat.# 13870 has two within-mix co-elutions, and cat.# 13871 had four. The default recommendation by ProEZGC (cat.# 13868) also only had two within-mix co-elutions, but provided improved separation of isopropyl acetate and m-xylene relative to cat.# 13870.

Table 5: Comparison of G43 column dimensions by co-elution. Colors represent different mixes (see Table 1), red represents co-elutions, red-boxes represent within-mix co-elutions.

# Rxi-624Sil MS Compound Rxi-624Sil MS cat.# 13870 Rxi-624Sil MS cat.# 13871 Rxi-624Sil MS cat.# 13868
RT Res RT Res RT Res
1 Ethylbenzene 2.48 22.9 3.22 15.7 2.39 28
2 Anisole 3.33 4.6 4.33 3.1 3.21 5.5
3 p-Xylene 3.57 2.4 4.64 1.6 3.44 2.9
4 1,2-Dichloroethane 3.7 2.4 4.81 1.6 3.56 2.9
5 Ethylene glycol 4.06 2.1 5.28 1.5 3.92 2.6
6 4-Methyl-2-pentanone 4.2 2.1 5.46 1.5 4.05 2.6
7 Isopropyl acetate 4.52 1.2 5.88 0.8 4.35 1.4
8 m-Xylene 4.6 1.2 5.98 0.8 4.43 1.4
9 Methylcyclohexane 4.77 2.3 6.2 1.6 4.59 2.8
10 Toluene 4.96 2.4 6.46 1.7 4.78 2.9
11 Chlorobenzene 5.15 2.4 6.7 1.7 4.96 2.9
12 Propyl acetate 5.8 0.3 7.47 0.2 5.59 0.4
13 n-Pentane 5.84 0.3 7.5 0.2 5.62 0.4
14 2-Methoxyethanol 6.57 8.2 8.06 4.9 6.29 8.4
15 Ethyl formate 7.39 6.8 8.55 5.2 6.84 8.4
16 Tetrahydrofuran 7.94 2.7 9.01 1.9 7.28 3.2
17 Isobutyl acetate 8.12 1.7 9.17 0.9 7.43 1.8
18 C6 8.22 1.5 9.24 0.8 7.51 1.6
19 1,2-Dimethoxyethane 8.32 1.5 9.31 0.8 7.58 1.6
20 2-Ethoxyethanol 8.52 1.9 9.53 0.6 7.78 1.7
21 Cyclohexane 8.66 0.5 9.58 0.6 7.85 0.9
22 Pyridine 8.66 0.5 9.64 0.8 7.89 0.9
23 Tetralin 8.84 1.4 9.84 1 8.07 1.7
24 Triethylamine 8.94 1.4 9.92 1 8.15 1.7
25 3-Methyl-1-butanol 9.06 2.2 10.02 1.4 8.25 2.5
26 Butyl acetate 9.29 0.1 10.22 0.3 8.46 0.2
27 1,4-Dioxane 9.3 0.1 10.24 — 8.47 0.2
28 Sulfolane 9.33 0.4 10.24 — 8.49 0.3
29 N,N-Dimethylacetamide 9.35 0.3 10.25 — 8.5 0.3
30 Ethyl acetate 9.37 0.3 10.3 0.7 8.53 0.7
31 n-Heptane 9.46 1.7 10.36 0.7 8.61 1.9
32 cis-1,2-Dicholoroethene 9.54 1.7 10.46 1.5 8.69 2.2
33 trans-1,2-Dichloroethene 9.7 3.5 10.61 2.2 8.84 3.9
34 Methyl-tert-butyl ether 10.02 0.4 10.9 0.9 9.13 0.9
35 Carbon tetrachloride 10.04 0.4 10.96 0.9 9.16 0.9
36 2-Hexanone 10.28 2.1 11.21 1.1 9.39 2.3
37 Diethyl ether 10.37 1.8 11.29 0.8 9.47 1.9
38 Ethanol 10.45 1.8 11.34 0.8 9.54 1.9
39 Methanol 10.68 5.9 11.6 4.1 9.77 7
40 2-Propanol 11.05 0.7 11.98 0.8 10.13 1
41 Acetone 11.08 0.7 12.03 0.4 10.17 1
42 Chloroform 11.14 1.5 12.06 0.4 10.22 1.5
43 1-Propanol 11.22 0.6 12.16 0.3 10.31 0.4
44 1-Butanol 11.25 0.6 12.19 0.3 10.32 0.4
45 1-Pentanol 11.34 2.5 12.28 1.5 10.42 3.1
46 Benzene 11.55 5.4 12.5 3.6 10.62 6.5
47 1,1,1-Trichloroethane 11.75 2.6 12.72 1.6 10.83 3.1
48 Acetonitrile 11.85 2.6 12.82 1.6 10.92 3.1
49 Methylene chloride 12 4.1 12.98 2.5 11.07 4.9
50 Formamide 12.39 1.8 13.44 1.1 11.47 2.1
51 1,1-Dichloroethene 12.46 1.8 13.51 1.1 11.55 2.1
52 Nitromethane 12.56 — 13.62 — 11.65 —
53 Isobutanol 12.56 — 13.62 — 11.65 —
54 2-Butanol 13.07 1.8 14.18 1.3 12.17 2.2
55 2-Butanone 13.14 1.7 14.26 1.1 12.24 2
56 Trichloroethene 13.2 1.7 14.33

 

1.1

12.3 2
57 Methyl acetate 14.81 18.4 16.14 12.6 13.96 22.3
58 N-Methylpyrrolidone 15.49 18.4 16.92 12.6 14.66 22.3
59 Isopropylbenzene 16.59 28.9 18.17 19.7 15.8 34.6

The optimized GC conditions for each column are as follows:

Table 6: GC Conditions for columns in Table 5.

Column: Stabilwax, 30.00 m, 0.32 mm ID, 0.25 µm (cat.# 10624)
Carrier Gas: Helium, Constant Flow @ 1.79 mL/min
Average Velocity: 29.48 cm/sec
Outlet Pressure (abs): 14.70 psi
Oven Temp.: 30 °C (hold 5.5 min) to 70 °C @ 6.5 °C/min to 225 °C @ 26 °C/min (hold 3 min)
Column: Stabilwax, 30.00 m, 0.53 mm ID, 0.25 µm (cat.# 10625)
Carrier Gas: Helium, Constant Flow @ 2.97 mL/min
Average Velocity: 21.48 cm/sec
Outlet Pressure (abs): 14.70 psi
Oven Temp.: 30 °C (hold 6 min) to 75 °C @ 13.5 °C/min to 225 °C @ 30 °C/min (hold 3 min)
Column: Stabilwax, 30.00 m, 0.32 mm ID, 1.00 µm (cat.# 10654)
Carrier Gas: Helium, Constant Flow @ 1.79 mL/min
Average Velocity: 29.65 cm/sec
Outlet Pressure (abs): 14.70 psi
Oven Temp.: 30 °C (hold 6.5 min) to 65 °C @ 6.5 °C/min to 225 °C @ 23 °C/min (hold 7 min)

Overall, this comparison provided some important insight into the separations to expect when changing column dimensions. The ProEZGC default dimensions certainly seemed to provide better solutions then the USP <467> recommended dimensions for both phases. Depending on your analysis, this might save you a significant amount of trial-and-error trying to determine which of the suggested columns will work best.

If we choose to compare instead between the phases, using results from the ProEZGC recommended phases, we see that each column results in two within-mix co-elutions. Using the Stabilwax, we can expect ethyl formate and methyl acetate from Class 3 Mix A, and 1,1,1-Trichloroethane and carbon tetrachloride from Class 1 mix, to co-elute. Using the Rxi-624sil MS, we can expect isopropyl acetate and m-xylene from Class 3 Mix A, and nitromethane and isobutanol from Class 2 Mix A, to co-elute. All other co-elutions can be overcome by running one mix at a time. Depending on which solvents are more relevant to your lab, either column may be ideal.

Conclusions:

With so many options to sift through, and so many opportunities to tailor your lab efficiency, implementing USP <467> can be complicated. With the help of ProEZGC, Restek can helps your lab get running as fast as possible, and with as little troubleshooting as possible, as you build GC methods that match your labs needs.

Produits mentionnés


Solvants résiduels – Classe 1, Dans le diméthylsulfoxyde, 1 ml/ampoule
Solvants résiduels de Classe 2 – Mélange A (révision 2020), 0.35-22.5 mg/mL dans le DMSO (dimethylsulfoxide), 1 mL/ampoule
Solvants résiduels, Classe 2 – Mélange B, DMSO, 1 ml/ampoule
Solvants résiduels – Classe 2 – Mélange C, Dans le diméthylsulfoxyde, 1 ml/ampoule
Solvants résiduels – Classe 3 – Mélange A (révision 2020), 5 000 µg/ml de chaque composé dans le N,N-diméthylformamide, 1 ml/ampoule
Solvants résiduels de Classe 3 – Solution étalon B, 5000 µg/ml de chaque composé, N,N-diméthylformamide, ampoule de 1 ml
Colonne capillaire Stabilwax, L 30 m, DI 0.32 mm, 0.25 µm
Colonne capillaire Stabilwax, L 30 m, DI 0.53 mm, 0.25 µm
Colonne capillaire Stabilwax, L 30 m, DI 0.32 mm, 1.00 µm
Colonne capillaire Rxi-624Sil MS, L 30 m, DI 0.32 mm, 1.80 µm
Colonne capillaire Rxi-624Sil MS, L 30 m, DI 0.53 mm, 3.00 µm
Colonne capillaire Rxi-624Sil MS, L 30 m, DI 0.25 mm, 1.40 µm

Auteurs

  • Erica Pack, PhD

    Erica is an interdisciplinary GC applications and technologies scientist at Restek. She obtained her bachelor's degree in forensic biology from The Pennsylvania State University, and her doctorate from Virginia Tech in plant pathology, physiology, and weed science. Since joining Restek in 2021, she has worked with a wide variety of GC columns, including fused silica, MXT, PLOT, and packed columns as well as accessories, such as liners, valves, and methanizers.

    View Articles
GNBL4630