- Sample Matrix: drinking water
- Sample Preparation: isotope dilution anion exchange solid-phase extraction
- Detection Technique: LC-MS/MS
Determination of Per- and Polyfluorinated Alkyl Substances in Drinking Water by Isotope Dilution Anion Exchange Solid Phase Extraction and Liquid Chromatography/Tandem Mass Spectrometry
U.S. EPA Method 533 is used for the analysis of 25 PFAS in drinking water. It combines solid-phase extraction sample preparation using weak anion exchange (WAX) cartridges with LC-MS/MS analysis on a C18 column. The method allows flexibility in some aspects, such as sample volume, LC column, LC conditions, and detector conditions, but changes to sample preservation, quality control requirements, and the extraction procedure are prohibited. In addition, the ratio of sorbent mass to aqueous sample volume cannot be reduced. The full analyte list and recommended products for PFAS analysis in drinking water by Method 533 are provided below. In addition, you can register and directly compare analyte lists and matrices across PFAS analysis methods here or visit www.restek.com/PFAS to explore our technical resources for PFAS analysis.
Analyte List for U.S. EPA Method 533 for PFAS Analysis
PFAS acronyms, compound names, and CAS numbers vary across methods and guidelines, and they also depend on compound form (e.g., neutral vs. salt). The following list uses standardized terminology to allow easy comparison across the methods and guidelines in our comprehensive guide. Please refer to U.S. EPA Method 533 for method-specific nomenclature.
| Acronym | Compound Name | CAS# |
| PFBS | Perfluoro-1-butanesulfonic acid | 375-73-5 |
| PFDA | Perfluorodecanoic acid | 335-76-2 |
| PFDoA | Perfluorododecanoic acid | 307-55-1 |
| PFHpA | Perfluoroheptanoic acid | 375-85-9 |
| PFHxS | Perfluoro-1-hexanesulfonic acid | 355-46-4 |
| PFHxA | Perfluorohexanoic acid | 307-24-4 |
| PFNA | Perfluorononanoic acid | 375-95-1 |
| PFOS | Perfluorooctanesulfonic acid | 1763-23-1 |
| PFOA | Perfluorooctanoic acid | 335-67-1 |
| PFUnA | Perfluoroundecanoic acid | 2058-94-8 |
| HFPO-DA, GenX | Hexafluoropropylene oxide dimer acid | 13252-13-6 |
| PFBA | Perfluorobutanoic acid | 375-22-4 |
| PFPeA | Perfluoropentanoic acid | 2706-90-3 |
| PFHpS | Perfluoro-1-heptanesulfonic acid | 375-92-8 |
| PFPeS | Perfluoro-1-pentanesulfonic acid | 2706-91-4 |
| 4:2 FTS | 1H,1H,2H,2H-Perfluorohexane sulfonic acid | 757124-72-4 |
| 6:2 FTS | 1H,1H,2H,2H-Perfluorooctane sulfonic acid | 27619-97-2 |
| 8:2 FTS | 1H,1H,2H,2H-Perfluorodecane sulfonic acid | 39108-34-4 |
| 9Cl-PF3ONS | 9-Chlorohexadecafluoro-3-oxanonane-1-sulfonic acid (F-53B Major) | 756426-58-1 |
| 11Cl-PF3OUdS | 11-Chloroeicosafluoro-3-oxaundecane-1-sulfonic acid (F-53B Minor) | 763051-92-9 |
| ADONA | 4,8-Dioxa-3H-perfluorononanoic acid | 919005-14-4 |
| PFEESA | Perfluoro(2-ethoxyethane)sulfonic acid | 113507-82-7 |
| NFDHA | Nonafluoro-3,6-dioxaheptanoic acid | 151772-58-6 |
| PFMPA | Perfluoro-3-methoxypropanoic acid | 377-73-1 |
| PFMBA | Perfluoro-4-methoxybutanoic acid | 863090-89-5 |
Recommended Products for U.S. EPA Method 533 for PFAS Analysis
Analytical LC Columns:
PFAS delay column: 5 µm, 50 x 3.0 mm (cat.# 580959 [custom product])
Guard Columns:
Standards:
Sample Prep:
Sample Reservoirs:
Vacuum Manifolds:
Vacuum Pumps:
General Lab Supplies:

Vous avez besoin d’aide pour vos analyses de PFAS ?
De la sélection de la méthode aux recommandations de produits, nos experts PFAS vous accompagnent pour développer vos protocoles complexes et obtenir des résultats fiables et précis.
Discover More Solutions for PFAS Analysis!
Start Saving
Time Today
Develop PFAS Methods in Minutes with our Pro EZLC Chromatogram Modeler
Start Saving Time Today
Develop PFAS Methods in Minutes with our Pro EZLC Chromatogram Modeler


