Articles

New Alumina Column Shows Promise for Analyzing Chlorofluorocarbons

28 Oct 2020

Restek Corporation, 110 Benner Circle, Bellefonte, PA, US
*Restek Corporation, Weerhaan 9, Middelburg, The Netherlands

Content previously published in Petro Industry News

A unique deactivation for alumina PLOT columns has recently been developed to control the high activity of alumina surfaces. Rt-Alumina BOND/CFC columns provide high retention and high selectivity for volatile halogenated hydrocarbons, without the activity that is usually observed with alumina adsorbents. Chlorofluorocarbon analysis, in particular, can benefit from this new technology.

Chlorofluorocarbons (CFCs) are a group of halogenated hydrocarbons that were heavily used as refrigerants and propellants, until their role in the depletion of the earth’s ozone layer became known. In addition to being active ozone depleting substances, CFCs are also greenhouse gases that contribute to global warming. Now, the manufacture of CFCs is prohibited in most countries and the use of recycled material is restricted. Because of this, CFC analysis has become a priority, particularly in the environmental field and for industry as companies develop alternative compounds.

CFCs are highly volatile and amenable to gas chromatography, but an extremely retentive GC column is needed for successful analysis. Thick film nonpolar capillaries can be used, but these are low in efficiency and generally do not perform optimally. Separations of CFCs have also been done with packed columns, as they can be made with adsorbents or liquid phases with a low phase ratio which allows the resolution of volatile compounds with very specific selectivity. Separations on packed columns are mainly possible due to selectivity combined with high k values; efficiency is still relatively poor as the number of theoretical plates is very low.

The introduction of chemically bonded phases and porous layer open tubular (PLOT) capillary columns allowed selectivity to be combined with a high number of theoretical plates, which provided more opportunity for resolution. Alumina, in particular, offers high retention and good selectivity for a large range of CFCs, but historically these columns have been limited by high activity. Restek recently developed the Rt-Alumina BOND/CFC column, a new alumina PLOT column which uses a unique deactivation that is quite promising for CFC analysis.

Alumina: Limited by Reactivity

In order to benefit from the retention and selectivity of alumina columns for volatiles analysis, the reactivity of the alumina must be reduced. This is usually accomplished through salt deactivations, often using sodium sulfate or potassium chloride. Even with these conventional deactivations, reactivity is often a problem. For example, using 1,2-butadiene as a test probe, column activity is revealed by the formation of reaction products (Figures 1 and 2). In contrast, the new CFC deactivation developed by Restek results in a more fully passivated surface (Figure 3). The deactivation is also quite robust, as the column can be held at the maximum temperature for alumina (200 °C) for 16 hours and no change in the retention or response of volatile halogenated hydrocarbons is seen (Figure 4).

A Promising New Alumina for CFCs

The Rt-Alumina BOND/CFC column provides excellent retention and selectivity for CFC analysis. Figures 5 and 6 show practical applications of this new technology for CFC impurities analysis. Different components are fully resolved and show no signs of decomposition. Although this new column is highly inert, residual reactivity for some mono- or di-substituted halogenated hydrocarbons remains, however the majority of these compounds can be accurately quantified from mainstream processes or in impurity analyses.

Summary

Overall, early results from our lab, as well as feedback from users in the field, show the new Rt-Alumina BOND/CFC column to be quite promising for CFC separations. The unique deactivation effectively reduces column reactivity and allows the benefits of alumina retention and selectivity to be realized for laboratories testing these priority compounds.

Figure 1: Butadiene on conventional alumina PLOT column with sodium sulfate deactivation. Butadiene, used here as an indicator of inertness, decomposes rapidly on a conventional alumina column because the sodium sulfate deactivation does not effectively prevent reactivity.
1,2-Butadiene on Na<sub>2</sub>SO<sub>4</sub> Deactivated Alumina PLOT Column

GC_PC1196

Conditions

ColumnAlumina PLOT (Na2SO4), 50 m, 0.32 mm ID
Standard/Sample1,2-butadiene
Injection
Inj. Vol.:40 µL split (split ratio 30:1)
Oven
Oven Temp.:150 °C
Carrier GasH2, constant pressure (23 psi, 158.6 kPa)
Temp.:150 °C
DetectorFID
NotesMajor decomposition of 1,2-butadiene is seen due to reactivity with a Na2SO4 deactivated alumina surface.
Figure 2: Butadiene on conventional alumina PLOT column with potassium chloride deactivation. Inertness is somewhat improved, but the column is still too reactive as evidenced by reaction product formation.
1,2-Butadiene on KCl Deactivated Alumina PLOT Column

GC_PC1197

Conditions

ColumnCP-Al2O3/KCl, 50 m, 0.32 mm ID
Injection
Inj. Vol.:40 µL split (split ratio 30:1)
Oven
Oven Temp.:150 °C
Carrier GasH2, constant pressure (23 psi, 158.6 kPa)
Temp.:150 °C
DetectorFID
NotesMajor reactivity of 1,2-butadiene on a KCl deactivated alumina PLOT column.
Figure 3: Butadiene on Rt-Alumina BOND/CFC columns. These new alumina PLOT columns are significantly more inert than other alumina columns, as shown by the reduced reactivity with butadiene.
1,2-Butadiene on Rt-Alumina BOND/CFC

GC_PC1198

Conditions

ColumnRt-Alumina BOND/CFC (custom), 50 m, 0.32 mm ID, 5.0 µm
Injection
Inj. Vol.:40 µL split (split ratio 30:1)
Oven
Oven Temp.:150 °C
Carrier GasH2, constant pressure (23 psi, 158.6 kPa)
Temp.:150 °C
DetectorFID
Figure 4: Impact of high temperature activation of alumina. Highly inert Rt-Alumina BOND/CFC columns produce stable responses and retention times, even after 16 hours at maximum temperature.
Stability Test: Halogenated Hydrocarbons on Rt-Alumina BOND/CFC (Before & After 16 hrs. at 200 °C)

GC_PC1192

Conditions

ColumnRt-Alumina BOND/CFC, 30 m, 0.53 mm ID, 10 µm (cat.# 19763)
Injectionsplit (split ratio 20:1)
Liner:Single Gooseneck w/ Wool
Oven
Oven Temp.:135 °C
Carrier GasH2, constant pressure (5 psi, 34.5 kPa)
Temp.:135 °C
DetectorFID
NotesHigh temperature stability test conditions: 16 hours at 200 °C.
Figure 5: Separation of chlorofluorocarbons on an Rt-Alumina BOND/CFC column.
Impurity Analysis of 1,1,1,2-Tetrafluoroethane (CFC-134a) on Rt-Alumina BOND/CFC

GC_GN1155

Conditions

ColumnRt-Alumina BOND/CFC, 30 m, 0.53 mm ID (cat.# 19763)
Standard/Sample1,1,1,2-Tetrafluoroethane
Conc.:Neat
Injection
Inj. Vol.:500 µL split
Oven
Oven Temp.:80 °C (hold 6 min) to 140 °C at 10 °C/min (hold 2 min)
Carrier GasHe
DetectorFID
NotesGas sampling, purity analysis

Note that tailing peaks are common in CFC analyses due to overloading normally employed for this type of work.
Figure 6: Impurity analysis of CFC-134a on an Rt-Alumina BOND/CFC column. Symmetric peaks with no sign of decomposition are obtained due to high column inertness.
Chlorofluorocarbons (CFCs) on Rt-Alumina BOND/CFC

GC_PC1199

Conditions

ColumnRt-Alumina BOND/CFC, 30 m, 0.53 mm ID, 10 µm (cat.# 19763)
Injection
Inj. Vol.:500 µL split
Oven
Oven Temp.:80 °C (hold 6 min) to 140 °C at 10 °C/min (hold 2 min)
Carrier GasHe
DetectorFID
NotesGas sampling was the injection method used in this application.
AcknowledgementAndre Hahnel, Westfalen AG

Auteurs

  • Jaap de Zeeuw

    Jaap is a world-renowned chromatographer with over 40 years of experience, including 15 years with Restek and 27 years with Varian/Chrompack prior to joining Restek. Throughout his career, Jaap has focused the chromatographic challenges of industrial analysis. For his 1979 graduation from The Institute for Higher Education, where he specialized in chemistry, Jaap authored a paper titled "The Challenge of Coating Flexible Fused Silica Capillary Columns" and has since distinguished himself as an authority on the subject. Jaap has been directly involved with the creation of numerous chemically bonded columns, including the first bonded PEG column and the stabilized PLOT columns widely used in the petrochemical arena. He has also helped develop new techniques, such as fast GC-MS using vacuum GC technology (low-pressure GC or LPGC). Over his innovative career, he has filed several patents for his work. Based out of the Netherlands, Jaap is extensively published and regularly travels internationally to share his knowledge. After retiring from Restek, Jaap founded CreaVisions, where he works as a GC Consultant and teaches master classes on key GC topics as well as on creativity in science.

    View Articles
  • Tom Vezza
PCAR3242