PFAS Analysis by ASTM D8421-22 for Aqueous Samples

03 Oct 2025

feature EVAR4305
  • Sample Matrix: aqueous samples
  • Sample Preparation: filtration
  • Detection Technique: LC-MS/MS

Standard Test Method for Determination of Per- and Polyfluoroalkyl Substances (PFAS) in Aqueous Matrices by Co-solvation followed by Liquid Chromatography Tandem Mass Spectrometry (LC-MS/MS)

ASTM D8421-22 is used for the analysis of 44 PFAS in aqueous sample matrices. It combines filtration sample preparation with LC-MS/MS analysis. ASTM D8421-22 is not available from Restek, but it may be purchased from ASTM. The full analyte list and products recommended by Restek are provided below. In addition, you can register and directly compare analyte lists and matrices across PFAS analysis methods here or visit www.restek.com/PFAS to explore our technical resources for PFAS analysis.

Analyte List for ASTM D8421-22 for PFAS Analysis

PFAS acronyms, compound names, and CAS numbers vary across methods and guidelines, and they also depend on compound form (e.g., neutral vs. salt). The following list uses standardized terminology to allow easy comparison across the methods and guidelines in our comprehensive guide. Please refer to ASTM D8421-22 for method-specific nomenclature.

AcronymCompound NameCAS#
NEtFOSAAN-ethyl perfluorooctanesulfonamidoacetic acid2991-50-6
NMeFOSAAN-methyl perfluorooctanesulfonamidoacetic acid2355-31-9
PFBSPerfluoro-1-butanesulfonic acid375-73-5
PFDAPerfluorodecanoic acid335-76-2
PFDoAPerfluorododecanoic acid307-55-1
PFHpAPerfluoroheptanoic acid375-85-9
PFHxSPerfluoro-1-hexanesulfonic acid355-46-4
PFHxAPerfluorohexanoic acid307-24-4
PFNAPerfluorononanoic acid375-95-1
PFOSPerfluorooctanesulfonic acid1763-23-1
PFOAPerfluorooctanoic acid335-67-1
PFTeDAPerfluorotetradecanoic acid376-06-7
PFTrDAPerfluorotridecanoic acid72629-94-8
PFUnAPerfluoroundecanoic acid2058-94-8
HFPO-DA, GenXHexafluoropropylene oxide dimer acid13252-13-6
PFBAPerfluorobutanoic acid375-22-4
PFPeAPerfluoropentanoic acid2706-90-3
PFDSPerfluoro-1-decanesulfonic acid335-77-3
PFHpSPerfluoro-1-heptanesulfonic acid375-92-8
PFPeSPerfluoro-1-pentanesulfonic acid2706-91-4
FOSAPerfluoro-1-octanesulfonamide754-91-6
4:2 FTS1H,1H,2H,2H-Perfluorohexane sulfonic acid757124-72-4
6:2 FTS1H,1H,2H,2H -Perfluorooctane sulfonic acid27619-97-2
8:2 FTS1H,1H,2H,2H -Perfluorodecane sulfonic acid39108-34-4
8:2 FTUCA, FOUEA2H-perfluoro-2-decenoic acid70887-84-2
NMeFOSAN-Methylperfluorooctanesulfonamide31506-32-8
NEtFOSAN-ethylperfluorooctanesulfonamide4151-50-2
9Cl-PF3ONS9-Chlorohexadecafluoro-3-oxanonane-1-sulfonic acid (F-53B Major)756426-58-1
11Cl-PF3OUdS11-Chloroeicosafluoro-3-oxaundecane-1-sulfonic acid (F-53B Minor)763051-92-9
ADONA4,8-Dioxa-3H-perfluorononanoic acid919005-14-4
PFNSPerfluoro-1-nonanesulfonic acid68259-12-1
6:2 FHUEA2H-perfluoro-2-octenoic acid70887-88-6
PFEESAPerfluoro(2-ethoxyethane)sulfonic acid113507-82-7
NFDHANonafluoro-3,6-dioxaheptanoic acid151772-58-6
PFMPAPerfluoro-3-methoxypropanoic acid377-73-1
PFMBAPerfluoro-4-methoxybutanoic acid863090-89-5
NMeFOSE2-(N-methylperfluoro-1-octanesulfonamido)-ethanol24448-09-07
NEtFOSE2-(N-ethylperfluoro-1-octanesulfonamido)-ethanol1691-99-2
PFDoSPerfluorododecane sulfonate79780-39-5
3:3 FTCA3:3 Fluorotelomer carboxylic acid356-02-5
5:3 FTCA5:3 Fluorotelomer carboxylic acid914637-49-3
7:3 FTCA, FHpPA7:3 Fluorotelomer carboxylic acid or 3-perfluoroheptyl propanoic acid812-70-4
LiTFSI (LiTFSA, HQ-115)Lithium Bis(trifluoromethane)sulfonimide90076-65-6
PFPrAPentafluoropropanoic acid422-64-0

Recommended Products for ASTM D8421-22 for PFAS Analysis

Analytical LC Columns:


Colonne Raptor C18 2.7µm, 50 x 2.1 mm
Colonne LC Force C18, 1,8 μm 50 mm x 2.1 mm

Delay Columns:


“Colonne-retard” PFAS HPLC, 5 µm, 50 x 2.1 mm

PFAS delay column: 5 µm, 50 x 3.0 mm (cat.# 580959 [custom product])

Guard Columns:


Filtre/précolonne UltraShield UHPLC, fritté 0.2 μm, l’unité
Support de cartouche à connexion directe EXP, livré avec un écrou et 2 ferrules
Cartouche de garde Raptor C18 EXP, 2.7 µm, L 5 x DI 2.1 mm, lot de 3

Sample Prep:


Filtres pour seringues 25 mm, porosité 0.22 µm, polypropylène, bleu, lot de 100

General Lab Supplies:


Flacon à volume limité à visser, polypropylène, 9 mm, 700 μl, 12 x 32 mm, lot de 100
Flacon à volume limité à visser, polypropylène, 9 mm, 700 μl, 12 x 32 mm, lot de 1000
Bouchon à visser, polyéthylène (solide), membrane de 0.254 mm, transparent, 2.0 ml, 9 mm, lot de 100
Bouchon à visser, polyéthylène (solide), membrane de 0.254 mm, transparent, 2.0 ml, 9 mm, lot de 1000
Tubes vides en polypropylène pour centrifugation de 50 ml, avec bouchons, lot de 50
Tubes vides en polypropylène pour centrifugation de 50 ml, avec bouchons, lot de 500
Seringues NORM-JECT, 10 ml, embout « luer lock », lot de 100
Seringues NORM-JECT, 5 ml, embout « luer lock », lot de 100

* Note that ASTM methods refer to the use of a glass syringe. Norm-Ject plastic syringes are made from polypropylene/polyethylene and can be used as an alternative.