Articles

Raptor EtG/EtS: All Matrix Interferences Resolved

21 Dec 2020

feature CFSS2741

How often are matrix interferences affecting your EtG/EtS analysis? Restek’s EtG/EtS solution consistently resolves analytes away from matrix interferences in a fast, 4-minute, dilute-and-shoot method.

  • Proven performance for accurate, reliable ethyl glucuronide (EtG) and ethyl sulfate (EtS) analysis.
  • Strong retention consistently resolves analytes from matrix interferences.
  • Long column lifetime ensures consistent performance injection after injection.
  • Fast, 4-minute, dilute-and-shoot LC-MS/MS analysis supports high sample throughput.

Switch to a Raptor EtG/EtS column when you need consistent resolution of EtG and EtS from matrix interferences and long column lifetimes.

Order yours today at www.restek.com/raptor

Column Description:

article ligand cfss2741b 01 1

Pore Size:

90 Å

Particle:

2.7 μm superficially porous particle (SPP or “core-shell” particle) silica

Surface Area:

130 m2/g

End-Cap:

Proprietary

Carbon Load:

Proprietary

USP Phase Code:

NA

Phase Category:

Proprietary

Ligand Type:

Proprietary

article profile cfss2741b 01a

Recommended Usage:

pH Range: 2.0–8.0
Maximum Temperature: 40 °C
Maximum Pressure: 600 bar/8700 psi

Properties:

  • Resolution of EtG and EtS from matrix interferences.
  • Increased retention of EtG and EtS compared
    to traditional phases.

Switch to a Raptor EtG/EtS when:

  • Other columns can’t resolve EtG and EtS from matrix components.
  • You need high-throughput EtG/EtS analysis.
  • Low-level detection limits are desired.

Definitive Results for Alcohol Monitoring

Restek has developed a new Raptor column that significantly improves EtG and EtS analysis for alcohol monitoring programs. The Raptor EtG/EtS column consistently retains and resolves these important biomarkers from matrix interferences, providing accurate, reliable results injection after injection (Figure 1). Dependable performance, a fast 4-minute analysis, and long column lifetime make this column ideal for high-throughput labs. Paired with our quality reference standards and the optimized, verified method shown below, Restek’s Raptor EtG/EtS column is part of a complete solution that will help testing labs save time and increase certainty.

Figure 1: Rugged Raptor EtG/EtS columns provide complete resolution and consistent results even after 1000 injections, allowing more samples to be analyzed between column changes.
1,000 Injection Lifetime: EtG and EtS in Urine on Raptor EtG/EtS

LC_CF0706

Peaks

PeaksPrecursor IonProduct IonProduct Ion
1.Ethyl-β-D-glucuronide-d5225.984.9
2.Ethyl-β-D-glucuronide220.884.974.8
3.Ethyl sulfate-d5129.797.7
4.Ethyl sulfate124.796.879.7

Conditions

ColumnRaptor EtG/EtS (cat.# 9325A12)
Dimensions:100 mm x 2.1 mm ID
Particle Size:2.7 µm
Pore Size:90 Å
Guard Column:UltraShield UHPLC precolumn filter, 0.2 µm frit (cat.# 25809)
Temp.:35 °C
Standard/Sample
Diluent:0.1% Formic acid in water
Conc.: A 500 ng/mL QC sample was prepared in urine. 50 µL of the sample was diluted with 950 µL of a working internal standard (25 ng/mL EtS-d5/100 ng/mL EtG-d5 in 0.1% formic acid in water). The sample was vortexed at 3500 rpm for 10 seconds to mix. The sample was then centrifuged at 3000 rpm for 5 minutes at 10 °C. The autosampler needle was adjusted to inject from the supernatant.
Inj. Vol.:10 µL
Mobile Phase
A:0.1% Formic acid in water
B:0.1% Formic acid in acetonitrile
Time (min)Flow (mL/min)%A%B
0.000.5955
2.500.56535
2.510.5955
4.000.5955
DetectorMS/MS
Ion Mode:ESI-
Mode:MRM
InstrumentHPLC
NotesReference Standards
Ethyl-β-D-glucuronide (cat.# 34101)
Ethyl-β-D-glucuronide-d5 (cat.# 34102)
Ethyl sulfate sodium salt (cat.# 34103)
Ethyl sulfate-d5 sodium salt (cat.# 34104)

Experience Selectivity Accelerated. Order the Raptor EtG/EtS LC Column today.

Related Products


Colonne Raptor EtG/EtS, 2.7 µm, Colonne LC 100 x 2.1 mm
Filtre/précolonne UltraShield UHPLC, fritté 0.2 μm, l’unité
Ethyl-β-D-glucuronide (EtG), 1000 µg/ml, méthanol, ampoule de 1 ml
Ethyl-β-D-glucuronide-d5 (EtG-d5), 1000 µg/ml, méthanol, ampoule de 1 ml
Sodium sulfate éthyl (EtS), 1000 µg/ml, méthanol, ampoule de 1 ml

Auteurs

CFSS2741
Version : C