Articles

PFAS Analysis by U.S. EPA Method 8327 for Surface Water, Groundwater, and Wastewater

03 Oct 2025

feature EVAR4289
  • Sample Matrix: surface water, groundwater, and wastewater
  • Sample Preparation: method refers to SW-846 Methods 3500 and 3600
  • Detection Technique: LC-MS/MS

Per- and Polyfluorinated Alkyl Substances (PFAS) by Liquid Chromatography/Tandem Mass Spectrometry (LC-MS/MS)

U.S. EPA Method 8327 is used for the analysis of 24 PFAS in surface water, groundwater, and wastewater. It refers to SW-846 Methods 3500 and 3600 for details on an appropriate sample preparation procedure (e.g., solvent dilution or extraction), and provides example conditions for LC-MS/MS analysis on a C18 column. Method 8327 is part of the SW-846 compendium and notes that SW-846 methods are intended to be guidance methods that contain general information on how to perform an analytical procedure or technique, which a laboratory can use as a basic starting point for generating its own detailed standard operating procedure (SOP), either for its own general use or for a specific project application. The full analyte list and recommended products for PFAS analysis in surface water, groundwater, or wastewater by Method 8327 are provided below. In addition, you can register and directly compare analyte lists and matrices across PFAS analysis methods here or visit www.restek.com/PFAS to explore our technical resources for PFAS analysis.

Analyte List for U.S. EPA Method 8327 for PFAS Analysis

PFAS acronyms, compound names, and CAS numbers vary across methods and guidelines, and they also depend on compound form (e.g., neutral vs. salt). The following list uses standardized terminology to allow easy comparison across the methods and guidelines in our comprehensive guide. Please refer to U.S. EPA Method 8327 for method-specific nomenclature.

AcronymCompound NameCAS#
NEtFOSAAN-ethyl perfluorooctanesulfonamidoacetic acid2991-50-6
NMeFOSAAN-methyl perfluorooctanesulfonamidoacetic acid2355-31-9
PFBSPerfluoro-1-butanesulfonic acid375-73-5
PFDAPerfluorodecanoic acid335-76-2
PFDoAPerfluorododecanoic acid307-55-1
PFHpAPerfluoroheptanoic acid375-85-9
PFHxSPerfluoro-1-hexanesulfonic acid355-46-4
PFHxAPerfluorohexanoic acid307-24-4
PFNAPerfluorononanoic acid375-95-1
PFOSPerfluorooctanesulfonic acid1763-23-1
PFOAPerfluorooctanoic acid335-67-1
PFTeDAPerfluorotetradecanoic acid376-06-7
PFTrDAPerfluorotridecanoic acid72629-94-8
PFUnAPerfluoroundecanoic acid2058-94-8
PFBAPerfluorobutanoic acid375-22-4
PFPeAPerfluoropentanoic acid2706-90-3
PFDSPerfluoro-1-decanesulfonic acid335-77-3
PFHpSPerfluoro-1-heptanesulfonic acid375-92-8
PFPeSPerfluoro-1-pentanesulfonic acid2706-91-4
FOSAPerfluoro-1-octanesulfonamide754-91-6
4:2 FTS1H,1H,2H,2H-Perfluorohexane sulfonic acid757124-72-4
6:2 FTS1H,1H,2H,2H-Perfluorooctane sulfonic acid27619-97-2
8:2 FTS1H,1H,2H,2H-Perfluorodecane sulfonic acid39108-34-4
PFNSPerfluoro-1-nonanesulfonic acid68259-12-1

Recommended Products for U.S. EPA Method 8327 for PFAS Analysis

Analytical LC Columns:


Colonne Raptor C18 2.7µm, 50 x 2.1 mm
Colonne LC Force C18, 1,8 μm 50 mm x 2.1 mm

Delay Columns:


“Colonne-retard” PFAS HPLC, 5 µm, 50 x 2.1 mm

Guard Columns:


Filtre/précolonne UltraShield UHPLC, fritté 0.2 μm, l’unité
Support de cartouche à connexion directe EXP, livré avec un écrou et 2 ferrules
Cartouche de garde Raptor C18 EXP, 2.7 µm, L 5 x DI 2.1 mm, lot de 3

Standards:


PFAS Méthode EPA 533 Standard de dilution primaire de l’analyte (PDS) 2 µg/mL, Méthanol (1 mM NaOH)/2-Propanol (98:2), 1 mL/ampoule
Standard de dilution primaire (PDS) pour analyte PFAS méthode EPA 537.1, 2 µg/mL, Méthanol (1 mM NaOH)/2-Propanol (98:2), 1 mL/ampoule

Sample Prep:


Filtres pour seringues 30 mm, porosité 0.45 µm, nylon, rose, lot de 100
Filtres pour seringues 30 mm, porosité 0.22 µm, nylon, rose, lot de 100

Vacuum Manifolds:


Manifold sous-vide Resprep QR-24
Manifold sous-vide Resprep QR-12

Vacuum Pumps:


Pompe à vide Chemker 300 en PTFE, 20 L/min, AC110 V, 60 Hz
Pompe à vide Chemker 300 PTFE, 18 l/min, 220 V, 50 Hz

General Lab Supplies:


Flacon à volume limité à visser, polypropylène, 9 mm, 700 μl, 12 x 32 mm, lot de 100
Flacon à volume limité à visser, polypropylène, 9 mm, 700 μl, 12 x 32 mm, lot de 1000
Bouchon à visser, polyéthylène (solide), membrane de 0.254 mm, transparent, 2.0 ml, 9 mm, lot de 100
Bouchon à visser, polyéthylène (solide), membrane de 0.254 mm, transparent, 2.0 ml, 9 mm, lot de 1000
Tubes vides en polypropylène pour centrifugation de 50 ml, avec bouchons, lot de 50
Tubes vides en polypropylène pour centrifugation de 50 ml, avec bouchons, lot de 500
Seringues NORM-JECT, 10 ml, embout « luer lock », lot de 100
Seringues NORM-JECT, 5 ml, embout « luer lock », lot de 100

Discover More Solutions for PFAS Analysis!

Start Saving
Time Today

Develop PFAS Methods in Minutes with our Pro EZLC Chromatogram Modeler

Start Saving Time Today

Develop PFAS Methods in Minutes with our Pro EZLC Chromatogram Modeler

Tags

Author

  • Restek Corporation

    Restek is a leading provider of chromatography columns, accessories, and certified reference materials. Trust Restek for reliable, high-quality analytical solutions.

    View all posts
EVAR4289
Version: B