Articles

Extending the Power of Stabilized PLOT Column Technology to Process GC Analyzers

27 Oct 2020

Restek Corporation, 110 Benner Circle, Bellefonte, PA, US
* Restek Corporation, Weerhaan 9, Middelburg, The Netherlands

Recent advances in PLOT column technology that were first developed in fused silica have now been transferred to metal MXT PLOT columns. These new columns offer greater stability than conventional PLOT columns, making them a more reliable choice for continual process monitoring. Advantages include highly reproducible flow characteristics, improved layer stability, and excellent separation efficiencies.

Porous layer open tubular (PLOT) columns are useful for analyzing volatiles in petrochemical product streams, as the specialized adsorbents provide good resolution and fast analysis times. However, conventional PLOT columns suffer from poor mechanical stability, limiting their use in process analyzers which require robust columns for continual operation. Recently Restek developed new PLOT column bonding techniques that resulted in improved layer stability, consistent flow behavior, and more reproducible retention times. This technology, which was first developed for fused silica columns, is now being used with metal MXT tubing, resulting in rugged columns that outperform typical metal PLOT columns. Exceptionally stable, new MXT PLOT columns provide efficient resolution and reliable performance, making them ideal for process GC analyzers.

New Technology Increases Reliability and Reproducibility

Restek’s PLOT columns are stabilized through a proprietary process that is based on concentric adsorption layers and improved particle bonding. The resulting columns are more stable than conventional PLOT columns and have much more reproducible flow behavior. To demonstrate the superior stability of MXT PLOT columns, an MXT-Q-BOND column and a competitor’s Q type column were subjected to 240 pressure pulse cycles and the spiking observed in each analytical run was used as an indicator of particle generation or phase instability. Results demonstrate that particle generation on the vendor V column was significantly higher (Figure 1), resulting in restrictions in the column that caused a shift in retention time (Figure 2). In contrast, the MXT-Q-BOND column showed little spiking. Greater phase stability resulted in consistent flow behavior and predictable retention times (Figure 3).

In addition to more consistent flows and reliable retention times, new MXT PLOT columns show greater thermal stability and much less phase bleed than the comparable competitor product (Figure 4). Lower bleed improves sensitivity and ensures faster stabilization times. In addition, because less stationary phase is lost, retention times are more stable over the lifetime of the column.

Key Phases Available for Optimized Separations

Metal MXT columns are available for all major adsorbent types: porous polymer, molecular sieve, and alumina. Porous polymer MXT columns are highly inert and effective at separating both polar and nonpolar compounds. Volatiles are strongly retained, making these columns extremely useful for separating many target analytes. As shown in Figure 5, MXT-Q-BOND columns provide excellent chromatography for challenging solvents—even alcohols elute as sharp, symmetric peaks.

MXT-Msieve 5A columns provide efficient separation of argon and oxygen, as well as other permanent gases (Figure 6). Special coating and deactivation procedures ensure chromatographic efficiency and the integrity of the porous layer coating. The unique immobilization process guarantees that particles remain adhered to the tubing, which is particularly valuable for flow and valve switching applications and allows these procedures to be run with greater confidence.

Metal alumina columns are recommended for light hydrocarbon analysis, as alumina is one of the most selective adsorbents available and allows all C1-C5 isomers to be separated with the highest degree of resolution. As shown in Figure 7, MXT-Alumina PLOT columns provide excellent chromatographic separation of refinery gas components. Peaks are fully resolved and symmetric, showing no sign of degradation or reactivity with the metal tubing.

Summary

New MXT PLOT columns from Restek offer greater stability than conventional PLOT columns, making them a better choice for process monitoring. New bonding techniques produce columns with highly reproducible flow characteristics, improved layer stability, and excellent separation efficiencies. These robust columns produce exceptionally reproducible chromatography, providing the reliable performance needed for process GC analyzer applications.

Figure 1: New bonding technology in MXT PLOT columns results in greater phase stability and reduced particle generation, compared to conventional PLOT columns.
figure article PCAR1637 01
Figure 2: Vendor V PLOT columns release particles following pressure pulsing, forming restrictions in the column that affect flow behavior and change retention time.
Test Chromatograms Before and After Pressure Pulsing; Vendor V, Metal Q PLOT

GC_PC1185

Conditions

ColumnVendor V, Q PLOT, 25 m, 0.53 mm ID
Standard/SampleSolvent mix
Injection
Inj. Vol.:1.0 µL split
Inj. Temp.:250 °C
Split Vent Flow Rate:150 mL/min.
Oven
Oven Temp.:150 °C
Carrier GasH2, constant pressure (4 psi, 27.6 kPa)
DetectorFID @ 250 °C
InstrumentAgilent/HP6890 GC
NotesPressure pulse test = 240 cycles
Figure 3: MXT PLOT columns are exceptionally stable; flow characteristics and retention times are highly consistent and not affected by pressure pulses.
MXT-Q-BOND Columns are Highly Stable; Pressure Pulsing does not Create Particles or Alter Retention Times

GC_PC1186

Conditions

ColumnMXT-Q-BOND, 30 m, 0.53 mm ID, 20 µm (cat.# 79716)
Standard/SampleSolvent mix
Injection
Inj. Vol.:1.0 µL split
Inj. Temp.:250 °C
Split Vent Flow Rate:150 mL/min.
Oven
Oven Temp.:150 °C
Carrier GasH2, constant pressure (4 psi, 27.6 kPa)
DetectorFID @ 250 °C
InstrumentAgilent/HP6890 GC
NotesPressure pulse test = 240 cycles
Figure 4: The bonding technology used in new MXT PLOT columns increases thermal tolerance, resulting in lower bleed, faster stabilization times, and higher sensitivity.
Thermal Stability Comparison of Q Type Porous Polymer Columns

GC_PC1187

Conditions

ColumnSee notes for column descriptions.
Injectionsplit
Inj. Temp.:250 °C
Split Vent Flow Rate:150 mL/min.
Oven
Oven Temp.:250 °C (hold 10 min) to 40 °C at 50 °C/min
Carrier GasH2, constant pressure (4 psi, 27.6 kPa)
DetectorFID @ 250 °C
InstrumentAgilent/HP6890 GC
NotesColumns tested:
Vendor V, metal Q type column
Restek MXT-Q-BOND

Columns were conditioned at 250 °C for the same length of time, and then bleed was compared to demonstrate the temperature stability of each column. Lower bleed columns allow higher sensitivities and faster stabilization times.
Figure 5: Sharp, symmetric solvent peaks are obtained on an MXT-Q-BOND PLOT column.
Solvents on MXT-Q-BOND

GC_PC1184

Conditions

ColumnMXT-Q-BOND, 30 m, 0.53 mm ID, 20 µm (cat.# 79716)
Standard/SampleSolvent mix
Injectionsplit
Inj. Temp.:250 °C
Split Vent Flow Rate:200 mL/min.
Oven
Oven Temp.:150 °C
Carrier GasH2, constant flow
Linear Velocity:40 cm/sec
DetectorFID
InstrumentAgilent/HP6890 GC
Figure 6: Permanent gases, including argon and oxygen, are fully separated on an MXT-Msieve 5A column using a µ-TCD.
Permanent Gases on MXT-Msieve 5A

GC_PC1182

Conditions

ColumnMXT-Msieve 5A, 30 m, 0.53 mm ID, 50 µm (cat.# 79723)
Standard/Sample
Conc.:1% in hydrogen
Injectionsplit (split ratio 50:1)
Oven
Oven Temp.:30 °C
Carrier GasH2
Detectorμ-TCD
Figure 7:  Highly inert MXT-Alumina BOND columns provide reliable baseline separation of refinery gas components.
Refinery Gas on MXT-Alumina BOND/Na<sub>2</sub>SO<sub>4</sub>

GC_PC1183

Conditions

ColumnMXT-Alumina BOND/Na2SO4, 30 m, 0.53 mm ID, 10 µm (cat.# 79714)
Standard/SampleRefinery gas
Injection
Inj. Vol.:10 µL split
Liner:2 mm split
Inj. Temp.:45 °C
Split Vent Flow Rate:80 mL/min.
Oven
Oven Temp.:45 °C (hold 1 min) to 200 °C at 10 °C/min (hold 8.5 min)
Carrier GasHe, constant pressure (10 psi, 68.9 kPa)
Linear Velocity:31 cm/sec @ 45 °C
DetectorFID @ 200 °C

Auteurs

  • Jaap de Zeeuw

    Jaap is a world-renowned chromatographer with over 40 years of experience, including 15 years with Restek and 27 years with Varian/Chrompack prior to joining Restek. Throughout his career, Jaap has focused the chromatographic challenges of industrial analysis. For his 1979 graduation from The Institute for Higher Education, where he specialized in chemistry, Jaap authored a paper titled "The Challenge of Coating Flexible Fused Silica Capillary Columns" and has since distinguished himself as an authority on the subject. Jaap has been directly involved with the creation of numerous chemically bonded columns, including the first bonded PEG column and the stabilized PLOT columns widely used in the petrochemical arena. He has also helped develop new techniques, such as fast GC-MS using vacuum GC technology (low-pressure GC or LPGC). Over his innovative career, he has filed several patents for his work. Based out of the Netherlands, Jaap is extensively published and regularly travels internationally to share his knowledge. After retiring from Restek, Jaap founded CreaVisions, where he works as a GC Consultant and teaches master classes on key GC topics as well as on creativity in science.

    View Articles
  • Rick Morehead
  • Tom Vezza
  • Gary Stidsen
PCAR1637