Articles

Rtx-Wax: A Rugged and Reliable Choice for the Analysis of Glycols in Water

15 Oct 2020

feature EVAR3086
  • Rugged enough to withstand repeated aqueous injections.
  • Maintains symmetrical peaks even after 600 water injections.
  • Ultra-low bleed allows for accurate detection down to 0.5 ng of glycols.

Introduction

From airplane and runway deicing to hydraulic fracturing operations, glycols are common chemicals used in industry, and as a result, chemical and environmental laboratories often test samples for their presence. However, the most frequently used sample introduction technique—direct aqueous injection—puts a significant demand on the analytical column and injection port liner. The columns best suited for water injections employ polyethylene glycol (PEG) stationary phases. PEG phases offer a unique selectivity, and their polar nature is compatible with an aqueous injection. An evaluation of ethylene glycol (EG) and propylene glycol (PG) response on PEG columns from a variety of vendors was conducted under optimized conditions and is presented here.

Benchmarking

Three different columns were evaluated in this study for linearity, lifetime, and bleed using split injection under the conditions shown in the chromatograms. The optimized split injection method employed here is further detailed in application note EVAN2873.

  • Restek Columns
    • Rtx-Wax (cat.# 12455)
  • Non-Restek Columns
    • Column A
    • Column B

All columns used in this study for the analysis of glycols in water were new and of a 30 m, 0.53 mm, 1.0 µm format. They were all conditioned for an hour at their labeled maximum temperature.

Results

The results of the benchmarking study for the analysis of glycols in water are summarized in Table I and discussed in detail below.

Table I: Column Benchmarking Results

ColumnsLIFETIMELINEARITYBLEED
# of Passing Injections*Peak Symmetry
(Final Injection)
Final r2FID Response (pA)
at 240 °C
EGPGEGPG
 6000.990.920.99990.999929
Column ALower calibration limits not met, so lifetime studies not performed53
Column BLower calibration limits not met, so lifetime studies not performed60

*The experiment was designed to determine the number of passing injections up to 600 injections, at which point the experiment was terminated. Passing injections were defined as injections with ChemStation symmetry values >0.5. All injections were 1 ng on-column.

Linearity

For each column, an initial calibration curve from 0.5 ng to 100 ng on-column was collected to establish suitable linearity for the analysis of glycols in water and to evaluate peak shape at different concentrations. The two lowest levels of the calibration curve, 0.5 and 1.0 ng on-column, were not detected on competitor columns A and B because the column bleed was too high to observe EG and PG at those levels. These results disqualified columns A and B from the subsequent lifetime study.

In contrast, excellent linearity was obtained on the Rtx-Wax column for both EG and PG (Figures 1 and 2). Even after the punishing conditions of 600 splitless aqueous injections, which were used to exacerbate the potentially column-damaging effects of water, the Rtx-Wax column would easily pass a calibration check. The high response for EG and PG at low levels, which is a function of column’s low bleed characteristic, is one factor that contributed to obtaining good linearity results across the entire calibration range.

Figure 1: Linear responses were obtained for propylene glycol (0.5–100 ng on-column) analyzed on an Rtx-Wax column even after 600 water injections.
figure-article-EVAR3086-01.jpg
Figure 2: Linear responses were obtained for ethylene glycol (0.5–100 ng on-column) analyzed on an Rtx-Wax column even after 600 water injections.
figure-article-EVAR3086-02.jpg

Lifetime

For the Rtx-Wax column, the lifetime study consisted of ten 1 µL splitless injections of water followed by a 50:1 split injection of a glycols standard, delivering 1 ng on-column. The experiment was designed to repeat until the peak symmetry (values determined using ChemStation software) dropped below 0.5 or the column reached 600 water injections, whichever came first. As shown in Figure 3, the Rtx-Wax column maintained excellent peak shape, even after 600 injections. In fact, one Rtx-Wax column was subjected to more than 1600 injections in an extended lifetime experiment, and it still exhibited symmetry values greater than 0.9 at the end of the experiment.

Figure 3: Propylene glycol and ethylene glycol peak shapes and retention times on an Rtx-Wax column are virtually identical, even after the column was exposed to 600 water injections.
Comparison of Glycols Before and After 600 Water Injections on Rtx-Wax

GC_EV1454

Peaks

PeakstR (min)
1.2-Butoxyethanol (IS)5.55
2.Propylene glycol6.27
3.Ethylene glycol6.41

Conditions

ColumnRtx-Wax, 30 m, 0.53 mm ID, 1.00 µm (cat.# 12455)
Standard/Sample
Glycols standard (cat.# 30471)
2-Butoxyethanol
Diluent:Water:methanol (90:10)
Conc.:50 µg/mL (1 ng on-column)
Injection
Inj. Vol.:1 µL split (split ratio 50:1)
Liner:Premium 4 mm Precision inlet liner w/wool (cat.# 23305)
Inj. Temp.:250 °C
Oven
Oven Temp.:40 °C (hold 1 min) to 250 °C at 30 °C/min
Carrier GasHe, constant flow
Flow Rate:5.7 mL/min
Linear Velocity:40 cm/sec
DetectorFID @ 250 °C
Make-up Gas Flow Rate:45 mL/min
Make-up Gas Type:N2
Hydrogen flow:40 mL/min
Air flow:450 mL/min
Data Rate:20 Hz
InstrumentAgilent/HP6890 GC

Bleed

The final test in our evaluation of PEG columns for the analysis of glycols in water was a bleed experiment. All three columns were ramped to a final temperature of 240 °C, which was the commonly achievable maximum temperature among the columns tested. The final bleed measurements showed the Rtx-Wax exhibited the lowest bleed (Table I, Figure 4). However, the maximum operating temperature for the Rtx-Wax is higher (250 °C), so it can be used with confidence when higher temperatures are required. Low column bleed allows for better sensitivity at low analyte concentrations.

Figure 4: The Rtx-Wax column exhibits the lowest bleed of the columns tested, even at temperatures that exceeded the recommended maximums (240 °C for competitor columns, 250 °C for Rtx-Wax column).
Rtx-Wax Bleed Profile Comparison

GC_EV1456

Peaks

PeakstR (min)
1.2-Butoxyethanol (IS)4.43

Conditions

ColumnRtx-Wax, 30 m, 0.53 mm ID, 1.00 µm (cat.# 12455)
Standard/Sample2-Butoxyethanol
Diluent:Water:methanol (90:10)
Conc.:200 µg/mL (10 ng on-column)
Injection
Inj. Vol.:1 µL split (split ratio 20:1)
Liner:Premium 4.0 mm ID Precision inlet liner w/wool (cat.# 23305)
Inj. Temp.:240 °C
Oven
Oven Temp.:40 °C to 240 °C at 30 °C/min (hold 10 min) to 250 °C at 30 °C/min (hold 10 min) to 260 °C at 30 °C/min (hold 10 min)
Carrier GasHe, constant flow
Linear Velocity:40 cm/sec
DetectorFID @ 240 °C
Make-up Gas Flow Rate:45 mL/min
Make-up Gas Type:N2
Hydrogen flow:40 mL/min
Air flow:450 mL/min
Data Rate:20 Hz
InstrumentAgilent/HP6890 GC
NotesAll column dimensions: 30 m, 0.53 mm ID, 1.00 μm

Conclusion

Of the polyethylene glycol-based columns evaluated in this study, the Rtx-Wax performed best in terms of linearity, lifetime, and bleed, demonstrating that even after exposure to the aggressive conditions of 600 aqueous injections it performed as well as the first injection. The Rtx-Wax is a rugged and reliable solution for the analysis of glycols in water.

Products Mentioned


Colonne capillaire Rtx-Wax, L 30 m, DI 0.53 mm, 1.00 µm

Authors

  • Corby Hilliard

    Corby Hilliard started his Restek career in the Quality Assurance department where he spent his first seven years as a QA analyst and worked his way up to a senior QA analyst. He then moved on to the Innovations department in 2009 as a GC Solutions Advanced Scientist. His experience is predominately in gas chromatography (GC) using various detectors and mass spectrometry as well as troubleshooting and method development. His primary work for Restek is generating product application data, using new and existing products in food safety, petrochemical, environmental, and the GC accessories line. He also is involved with quality and R&D as well as new product development.

    View all posts
  • Chris English

    Since 2004, Chris has managed a team of chemists in Restek's innovations laboratory who perform new product testing, method development, and applications work. Before taking the reins of our lab, he spent seven years as an environmental chemist and was critical to the development of Restek’s current line of volatile GC columns. Prior to joining Restek, he operated a variety of gas chromatographic detectors conducting method development and sample analysis. Chris holds a BS in environmental science from Saint Michael's College.

    View all posts
EVAR3086