FAQ

Are there any other size threads used for LC besides the 10-32?

31 Jul 2025

The 10-32 threads are probably the only ones you will use with HPLC. However, you may occasionally see ¼ x 28 or M6 (abbreviated form for M6 x 1) fittings used in laboratory settings or for lower pressure connections. The ¼ x 28 size is based on English measurement units like the 10-32. The ¼ x 28 threads are further apart and slightly larger in diameter and usually they do not withstand higher pressures as well. The M6 fittings are metric threads that are actually 6 mm in diameter and 1 mm apart.

Auteurs

  • Restek Corporation

    Restek is a leading provider of chromatography columns, accessories, and certified reference materials. Trust Restek for reliable, high-quality analytical solutions.

    View Articles
GNOT5077