Articles

Thermal Desorption Tubes vs. Canister Sampling

21 Oct 2020

Which VOC Sampling Technique is Right for You?

Thermal desorption tubes provide a complementary option to canisters for sampling VOCs. Both techniques have advantages and disadvantages and their features must be evaluated for suitability relative to the sampling environment and analytical capabilities. Enumerated below are the similarities and differences between these techniques; use this handy comparison to determine which equipment is best for you.

Similarities Between Thermal Desorption Tubes and Canisters

  • Reusable sampling device.
  • Long product lifetime.
  • Long-term sample stability.
  • Blank certification required prior to sampling.
  • Sample concentration required before GC/MS analysis.
  • Dry purge helpful to remove moisture before GC injection.
  • Ppt sensitivity.
  • Method acceptance.
  • Collection of wide range of VOCs with single device.
  • Useful for screening of unknowns.
  • Leak tightness critical to maintaining sample integrity and preventing contamination of a clean device.

Differences Between Thermal Desorption Tubes and Canisters

 Thermal Desorption TubesCanisters
MethodsUS EPA TO-17
ASTM D6196
ISO 16017, ISO 16000-6
NIOSH 2549
US EPA TO-14, TO-15
ASTM D5466
OSHA PV2120
NIOSH Protocol Draft
 World-wide acceptanceGold standard for US ambient air market
ApplicationsAmbient air, indoor air, industrial hygiene
Material emissions
Food & flavor
Chemical weapons
Ambient air, indoor air, vapor intrusion, emergency response
 C3 to C30<C3 to ~C10
HandlingLight weight for personal monitoring and general ease of useLarger and heavier; more costly to ship
SamplingActive sampling with sampling pump or diffusive sampling without pump is possible with determined diffusion coefficients for each compound.Passive sampling, no sampling pump required. Long-term sampling possible without battery to recharge.
 Integrated sampling onlyGrab & integrated sampling
 Concentrated sampleWhole air
 Proper sorbent selection recommended in methodology.N/A
 Must sample below sorbent breakthrough volumes to avoid sample loss and irreversible adsorption on sorbentN/A
 Large sample volumes >100LSample volume is function of canister size, 15L max
AnalysisTube dimensions are instrument specificCompatible with all manufacturer sample concentrators
 1 injection, more injections possible for some instrumentationMultiple sample injections
 Concentration range ppt to ppmppt to ppm
 Some sorbents prone to artifact formation.Low blanks when properly cleaned.
StorageSample storage at 4°C recommended for multi-bed tubes to prevent potential migration of compounds to more retentive sorbent which maybe difficult to recover.Room temperature
CleaningAnalytical process automatically cleans tube for reuse. Cleans as it analyzes. Conditioning/cleaning and analysis incorporated in one thermal desorption unit.Canister cleaning requires separate equipment as additional step prior to background certification and sampling.
Cost$50-130 each$200-700 each

Author

  • Restek Corporation

    Restek is a leading provider of chromatography columns, accessories, and certified reference materials. Trust Restek for reliable, high-quality analytical solutions.

    View all posts
GNOT3419